. Author manuscript; available in PMC: 2019 Jan 15.
Published in final edited form as: Wear. 2017 Oct 16;394-395:195–202. doi: 10.1016/j.wear.2017.10.008

Table 1.

Comparison of methods to measure surface wear in acetabular liners.

Study Method Accuracy
(µm)
Scan Time
(min)
Number of
pointsa
Present study Optical Scanner 2.1 10 ~ 1.8 million
Lord et al. [15] CMM 0.9 20 3024–4104
Wang et al. [17] CMM 1.9 50 3720
Teeter et al. [20] Micro-CT 50b 95 Not given
Bowden et al. [18] Micro-CT 74b 240 Not given
Yun et al. [30] Laser Scanner Not given 25–34 Not given
a

Measured on the acetabular liner.

b

Accuracy based on spatial resolution of the Micro-CT system.