Long‐term reliability of the defibrillator lead inserted by the extrathoracic subclavian puncture

Masaya Watanabe 1, Hisashi Yokoshiki 1,2,, Hirofumi Mitsuyama 1, Kazuya Mizukami 1, Taro Tenma 1, Rui Kamada 1, Masayuki Takahashi 1, Yoshiro Matsui 3, Toshihisa Anzai 1
PMCID: PMC6174403  PMID: 30327700

Abstract

Background

As the transvenous defibrillator lead is fragile and its failure may cause a life‐threatening event, reliable insertion techniques are required. While the extrathoracic puncture has been introduced to avoid subclavian crush syndrome, the reports on the long‐term defibrillator lead survival using this approach, especially the comparison with the cephalic cutdown (CD), remain scarce. We aimed to evaluate the long‐term survival of the transvenous defibrillator lead inserted by the extrathoracic subclavian puncture (ESCP) compared with CD.

Methods

Between 1998 and 2011, 324 consecutive patients who underwent an implantable cardioverter‐defibrillator (ICD) implantation in Hokkaido University Hospital were included. ICD leads were inserted by CD from 1998 to 2003 and by contrast venography‐guided ESCP thereafter. Lead failure was defined as a nonphysiologic high‐rate oversensing with abnormal lead impedance or highly elevated sensing and pacing threshold.

Results

Of 324 patients, CD was used in 37 (11%) and ESCP in 287 patients (89%). During the median follow‐up of 6.2 years (IQR:3.2‐8.3), 7 leads (2 in CD and 5 leads in ESCP group) failed. All patients with lead failure in ESCP group were implanted with either SJM Riata (n = 1) or Medtronic Fidelis lead (n = 4). Five‐year lead survival was 93.8% (CI95%:77.3‐98.4%) in CD compared with 99.1% (CI95%:96.6‐99.8%) in ESCP group (P = 0.903). Univariate Cox regression analysis showed that the use of Fidelis or Riata lead was the strong predictor of the ICD lead failure (HR 13.8, CI95%:2.9‐96.5; P = 0.001).

Conclusions

Contrast venography‐guided extrathoracic puncture ensures the reliable long‐term survival in the transvenous defibrillator leads.

Keywords: cephalic cutdown, defibrillator lead, extrathoracic puncture, lead failure, subclavian crush syndrome

1. INTRODUCTION

Implantable cardioverter‐defibrillator (ICD) is an effective therapy for life‐threatening ventricular arrhythmias.1 Numerous clinical trials proved its efficacy, and the indication of the ICD therapy has been expanding.2, 3, 4 At the same time, complications associated with the ICD leads, the most fragile component, have been recognized.5

Cephalic vein cutdown has been used for transvenous lead insertion since the early period of the device implantation.6 While the cephalic approach is recognized as the most favorable to reduce the mechanical stress related to the insertion procedures,7, 8, 9 it requires experience and longer procedure time.10 Subclavian puncture, instead, is performed easily and briefly, whereas this method has the potential long‐term risk of conductor fractures and insulation defects developed by compression of a lead between the first rib and clavicle.11 To avoid these serious complications and provide an easier procedure, the extrathoracic approach, guided by fluoroscopic landmark,12, 13 venography14, 15 or ultrasound,16 has been introduced. While the success rate in lead insertion was preferable with a low incidence of acute complications, the report of the long‐term lead survival using the extrathoracic approach is quite limited, especially for ICD leads which are structurally complicated and therefore fragile.5, 17

In this study, we aimed to evaluate the long‐term survival of transvenous ICD leads inserted by the venography‐guided extrathoracic subclavian puncture compared with that by the cephalic vein cutdown.

2. METHODS

2.1. Study cohort

Between 1998 and 2011, 324 consecutive patients who underwent an implantable cardioverter‐defibrillator (ICD) implantation in Hokkaido University Hospital were included. All patients were treated according to the routine clinical protocol and have provided a written informed consent.

2.2. Implantation techniques

After contrast venography, an ICD pocket was created in the left pectoral region unless the venous access was not available or the patient's preference. Thereafter, a transvenous ICD lead was inserted by either cephalic vein cutdown (CD) or contrast venography‐guided extrathoracic subclavian puncture (ESCP) method.

Between 1998 and August 2003, implantation procedures were performed by a cardiovascular surgeon and an electrophysiologist. During this period, venous access with CD was obtained by a cardiovascular surgeon.

After September 2003, the procedures were performed only by the electrophysiologists and transvenous leads were inserted by ESCP. After a contrast venography, the skin incision was made perpendicular to the deltopectoral groove at the level of the coracoid process.12 Once a pocket was created beneath the pectoral fascia, the puncture using a 21 G micropuncture needle (Micropuncture Introducer Set, Cook Medical, Inc., Bloomington, IN) aiming the extrathoracic subclavian vein as it crosses the first rib was conducted.14 Coracoid process and deltopectoral groove were also used as the anatomical landmarks. The vein runs 1.5‐2 cm medial and parallel to the deltopectoral groove and it crosses the deltopectoral muscle at the level of coracoid process.12 To avoid pneumothorax and make the gentle angle between the needle and the subclavian vein, the needle was held tangential to the chest wall (Figure 1). If several attempts failed to achieve venous access, the additional injections of contrast were performed. If the vasospasm was observed, the puncture targeting the area without narrowing, normally toward lateral area, was performed, otherwise the experienced operator conducted the puncture.

Figure 1.

Figure 1

A representative case of the extrathoracic subclavian puncture (ESCP). (A) Anteroposterior (AP) (left panel) and caudal 40° views (right panel) of the contrast venography before the skin incision. (B) Radiographs demonstrating ESCP in the AP (left panel) and caudal 40° views (right panel). The circle and the arrowhead indicate the points where the needle entering the pectoral muscle and the subclavian vein as it crosses the first rib (highlighted by the dotted lines), respectively. Of note, the guide wire is advanced into the brachiocephalic vein through the needle tip with a gentle angle against the surface of the first rib

After obtaining the venous access, a lead was gently handled to reduce the mechanical stress due to the contact with the venous and heart wall. When the lead tip came out of the sheath introducer, the lead stylet was withdrawn by 2‐3 cm. Once the lead was advanced into the right ventricle (RV), active fixation was performed normally onto the RV apex.

In the measurement of ICD lead parameters, R‐wave amplitude ≥ 5 mV and a pacing threshold ≤ 2.0 V with 0.5 or 0.4 ms pulse were considered acceptable. Defibrillation threshold tests were routinely performed twice, and successful defibrillation with more than 10 J as a safety margin was required.18

2.3. Follow‐up

After discharge, all patients were followed at the outpatient ICD clinic every 4 months. Extra visits were also planned in case of ICD shocks or the serious alerts from the remote monitoring system. Evaluation of the ICD clinic consisted of a patient history and interrogation of the device for arrhythmic events. Lead malfunction was examined by measurements of the lead impedance, pacing threshold, and intracardiac sensing amplitude.

2.4. Definition of lead failure

Lead failure was classified into (a) conductor fracture, (b) insulation defect, or (c) other dysfunction by the following definitions.5 Oversensing not related to cardiac cycle accompanied by either (a) the sudden increase in lead impedance >1500 Ohms or twice the number at last follow‐up or (b) decrease < 200 ohms was defined as the conductor fracture or insulation defect, respectively. Other lead dysfunctions were defined as linear increase in lead impedance to >1500 Ohm, linear decrease in sensing R‐wave amplitude < 2 mV, or increase in pacing threshold up to a level that the treating cardiologist considered inappropriate. Acute lead dislodgement was excluded from the analysis.

2.5. Statistical analysis

Continuous variables are expressed as mean ± SD or medians with interquartile ranges (IQR) and compared using Student's t test or Mann‐Whitney U test. Categorical variables are presented as numbers and percentages (%) and compared using chi‐squared test. Freedom from lead failure was estimated by Kaplan‐Meier method and compared by log‐rank test between groups. Univariate Cox regression analysis was performed to identify risk factors for ICD lead failure. Multivariable analyses were not performed because of the limited number of patients and events. All tests were 2‐sided, and a P < 0.05 was considered statistically significant. All statistical analyses were performed with JMP version 20.0.

3. RESULTS

3.1. Patient characteristics

Cephalic vein cutdown (CD) was used in 37 (11%) and the contrast venography‐guided extrathoracic subclavian puncture (ESCP) in 287 patients (89%). Lead placement by either CD or ESCP was successfully performed in all of the study patients. None in both CD and ESCP groups received the subclavian puncture as a venous access.

Left ventricular ejection fraction (LVEF) was lower (49 ± 16% vs 43 ± 16%, P = 0.032), and primary prevention was more often the indication for ICD therapy (11% vs 28%, P = 0.028) in the patients treated by ESCP (ESCP group) compared to those by CD (CD group). Detail of the patient characteristics according to the lead insertion techniques is shown in Table 1.

Table 1.

Patient and ICD lead characteristics

CD ESCP P
(n = 37) (n = 287)
Age 57 ± 14 58 ± 15 0.725
Female sex, n (%) 9 (24%) 58 (20%) 0.561
Structural heart disease
Ischemic, n (%) 11 (30%) 83 (29%) 0.420
No ischemic, n (%) 22 (59%) 148 (52%)
None, n (%) 4 (11%) 56 (20%)
LVEF (%) 49 ± 16 43 ± 16 0.032
ICD indication
Primary, n (%) 4 (11%) 79 (28%) 0.028
Secondary, n (%) 33 (89%) 208 (72%)
ICD type
Single chamber, n (%) 15 (41%) 94 (33%) 0.002
Dual chamber, n (%) 21 (57%) 123 (43%)
Biventricular, n (%) 1 (3%) 70 (24%)
Lead design
Active lead fixation, n (%) 37 (100%) 287 (100%)
Dual coil, n (%) 27 (73%) 277 (97%) < 0.001
Medtronic 6949 (Fidelis), n (%) 0 34 (10%) 0.027
SJM Riata, n (%) 0 10 (3%) 0.249

CD, indicates cutdown; ESCP, extrathoracic subclavian puncture; LVEF, left ventricular ejection fraction; ICD, implantable cardioverter‐defibrillator.

3.2. ICD device and lead characteristics

Biventricular ICD was more often used in ESCP group (single chamber, dual chamber, and biventricular ICD were 41%, 57%, and 3% in CD vs 33%, 43%, and 24% in ESCP group, P = 0.002; Table 1). Dual coil leads were more often used in ESCP group (73% vs 97%, P < 0.001). Medtronic 6949 Fidelis lead or SJM 1580 Riata lead was used in 34 (10%) and 10 (3%) patients in ESCA group, respectively. Detailed ICD lead characteristics in both groups are shown in Table 2 .

Table 2.

Detailed ICD lead characteristics

Cutdown
Maker Model Model number Fixation Coil Structure Outer insulation Inner insulation Number
Guidant Endotak DSP G 0125 active dual Multilumen Silicone Silicone 4
Endotak Endurance G 0154 active dual Multilumen Silicone Silicone 1
Medtronic Sprint M 6943 active dual Multilumen Silicone Silicone 10
Sprint Quattro M 6944 active dual Multilumen Polyurethane Silicone 8
Sprint M 6945 active dual Multilumen Silicone Silicone 14
Total 37
Extrathoracic puncture
Maker Model Model number Fixation Coil Structure Outer insulation Inner insulation Number
Guidant Endotak Endurance G 0154/55 active dual Multilumen Silicone Silicone 57
Endotak Reliance G 0180 active single Multilumen Silicone Silicone 3
Endotak Reliance G 0184/85 active dual Multilumen Silicone Silicone 34
Endotak Reliance SG G 0292 active single Multilumen Silicone Silicone 1
Endotak Reliance SG G 0295 active dual Multilumen Silicone Silicone 22
Medtronic Sprint M 6945 active dual Multilumen Silicone Silicone 29
Sprint Quattro Secure M 6947 active dual Multilumen Silicone Silicone 51
Sprint Fidelis M 6949 active dual Multilumen Silicone Silicone 34
SJM Riata S 1580 active dual Multilumen Silicone Silicone 9
Riata ST S 7030 active dual Multilumen Optim Silicone 1
Durata S 7120/21/30/31 active dual Multilumen Optim Silicone 40
Durata S 7122 active single Multilumen Optim Silicone 6
Total 287

3.3. Complications associated with the lead insertion

Pneumothorax occurred in two patients (0.7%) in ESCP group while none in CD group (P = 0.259).

3.4. Follow‐up

Patients in CD group were followed for the longer period, as compared with ESCP group (median 8 [IQR 5.7‐13.6] vs 5.9 years [IQR 2.8‐8.0], P < 0.001).

3.5. Lead survival

During the median follow‐up period of 6.2 years (IQR 3.2‐8.3), 7 ICD leads (two leads in CD and five leads in ESCP group) failed. Median time to the failure was 4.8 years (range 1.5‐8.4). The Kaplan‐Meier curve of the ICD lead survival according to the venous access is shown in Figure 2. Lead survival at 5 years was 93.8% (95% confidence interval [CI]: 77.3‐98.4%) in CD vs 99.1% (95% CI: 96.6‐99.8%) in ESCP group (P = 0.903).

Figure 2.

Figure 2

Kaplan‐Meier survival curves of transvenous defibrillator leads inserted by cephalic cutdown (CD) and the extrathoracic subclavian puncture (ESCP)

Figure 3 shows the Kaplan‐Meier curve of the ICD lead survival within the ESCP group according to the use of the recalled leads (i.e, Medtronic Fidelis and SJM Riata leads). Lead survival in the recalled leads was 94.7% and 80.7% (90.4% and 75.9% in 6949 Fidelis leads) at 5 and 10 years since implantation.

Figure 3.

Figure 3

Kaplan‐Meier survival curves of the recalled and unrecalled defibrillator leads in the patients for whom the extrathoracic subclavian puncture was used

3.6. Clinical presentation and failed lead characteristics

Clinical and lead characteristics in seven patients with failed ICD lead are shown in Table 3. All but one patient had structural heart disease and mean LVEF was 47 ± 25%.

Table 3.

Patient and lead data in the patients with ICD lead failure

Case No. Age at implant Sex Diagnosis LVEF (%) Indication Venous access Lead type Years since implant Clinical presentation Inappropriate shock Cause of failure
1 66 M HCM 70 Secondary CD Medtronic 6943 4.8 oversensing yes Unknown
2 70 F Myo 52 Secondary CD Medtronic 6945 1.5 oversensing yes Unknown
3 54 M CSA 68 Secondary ESCP SJM 1580, Riata 2.5 Pacing & sensing disorder no Conductor fracture?
4 57 F DCM 16 Primary ESCP Medtronic Fidelis 8.0 oversensing, high impedance no Conductor fracture
5 71 M OMI 42 Secondary ESCP Medtronic Fidelis 8.4 oversensing, high impedance no Conductor fracture
6 74 F OMI 67 Secondary ESCP Medtronic Fidelis 1.8 oversensing yes Conductor fracturea
7 71 M DCM 11 Primary ESCP Medtronic Fidelis 5.5 oversensing, high impedance yes Conductor fracture

HCM indicates hypertrophic cardiomyopathy; Myo, myocarditis; CSA, coronary spastic angina; CD, cutdown; ESCP, extrathoracic subclavian puncture; LVEF, left ventricular ejection fraction; DCM, dilated cardiomyopathy; OMI, old myocardial infarction.

a

Analysis of the extracted lead confirmed a conductor fracture at the distal portion.

Of note, all patients with ICD lead failure in ESCP group were treated with either SJM Riata or Medtronic Fidelis lead. In one patient with Fidelis lead failure (case 6), transvenous lead extraction was performed and a conductor fracture at the distal portion was confirmed. In all but one patient, oversensing of noise was documented. Inappropriate shocks were delivered to four patients. In one patient (case 3), unacceptable R‐wave reduction and elevation of pacing threshold were detected 2.5 years after implantation.

3.7. Risk factor for ICD lead failure

Table 4 shows the risk factors for ICD lead failure by univariate Cox regression analysis. Lead insertion by ESCP was not related to the lead failure (HR 0.46, CI 95% 0.10‐3.22; P = 0.383). The use of the recalled lead was a strong risk factor for the ICD lead failure (HR 13.8, CI 95% 2.92‐96.5, P = 0.001). In contrast to a prior study,19 higher age was a modest risk of the failure (HR 1.07, CI 95% 1.01‐1.14, P = 0.046).

Table 4.

Univariate analysis of predictors of ICD lead failures

Variable HR CI 95% P
Age↑, per 1 year 1.07 1.01‐1.14 0.046
Sex (Female) 2.48 0.49‐11.29 0.254
Structural heart disease 1.98 0.34‐37.59 0.494
Primary prevention 1.15 0.16‐5.35 0.868
LVEF 1.00 0.96‐1.05 0.957
ESCP 0.46 0.10‐3.22 0.383
Dual Coil 0.42 0.07‐7.92 0.468
Fidelis or Riata lead 13.80 2.92‐96.5 0.001

HR indicates hazard ratio; CI, confidence interval; LVEF, left ventricular ejection fraction; ESCP, extrathoracic subclavian puncture.

4. DISCUSSION

4.1. Main findings

In the present study, we demonstrate a satisfactory long‐term survival of ICD leads inserted by the contrast venography‐guided extrathoracic subclavian puncture (ESCP). Five‐year lead survival of 99.1% by ESCP was excellently comparable to that by cephalic vein cutdown (CD) of 93.8% (P = 0.903), considered as the safest venous approach.8 Of note, during the median follow‐up of 5.9 years, none in ESCP group implanted with leads other than the recalled ones had the defibrillator lead failure. These results may ensure ESCP as a reliable venous approach for the defibrillator lead insertion.

4.2. Extrathoracic puncture as a first‐line approach for lead insertion

Extrathoracic approach was introduced15 to avoid the complication related to the subclavian puncture.11, 20 While several techniques based on anatomical landmarks,12, 13 guided by contrast venography14, 15 or real‐time ultrasound,16 have been reported with the satisfactory procedure success and short‐term lead survival, the long‐term efficacy of this approach remains controversial. Aizawa et al reported the increased risk of the lead failure in ESCP compared with CD.7 In contrast, Chan et al reported that the risk of lead failure in the axially puncture was comparable with CD and significantly lower than subclavian puncture during the follow‐up > 5 years.10 Of note, in those reports, the choice of the venous accesses was at the discretion of the operator10 or ESCP was only performed if CD was failed as a fist choice.7 In addition, these reports did not only include the patients with an ICD but also those with a pacemaker. To the best of our knowledge, the present study, for the first time, demonstrated the long‐term ICD lead survival inserted by ESCP in the consecutive patients.

In our patients, five‐year lead survival treated by ESCP was 99.1%, comparable with that by CD of 93.8%. Chan et al reported, in consistent with our results, the pacemaker lead survival was 98.8% s and 97.7% with the axially puncture and CD during the follow‐up of 73.6 months (P = 0.389). Prior studies reported that the ICD lead survival was 97.5% to 85% at 5 years,21, 22, 23, 24 although most of the studies did not provide the detail in the venous access. Although simple comparison might be difficult due to various conditions among different studies, the long‐term ICD lead survival by ESCP in the present study seems satisfactory and has never been inferior to that in prior reports using any venous approach. In addition, our results and the literature guaranteed high success rate of the procedural achievement of ESCP.10, 14 We expect that our results accumulate an evidence of the efficacy of ESCP as a first‐line venous access for ICD lead implantation.

4.3. Lead failure in the recalled ICD leads

In consistent with the previous reports,19, 25 the use of the recalled lead was a risk of the lead failure in our population (Table 4). Lead survival in these fragile leads was 94.7% and 80.7% (90.4% and 75.9% in 6949 Fidelis leads) at 5 and 10 years since implantation, presenting the time‐depending increase of the lead failure.9 In a patient in whom a Fidelis lead was extracted after failure, the conductor fracture at the distal portion was confirmed. In other four patients with the failure in their recalled leads, at least, conductor fracture at the proximal portion was not observed by the X‐ray examination.

Recent multicenter study in Canada on the Fidelis lead failure9 reported that the lead failure rate in the axially vein access was higher compared with that in cutdown (23.1% vs 11.4% at 5 years). Interestingly, the failure rate in axially access was higher compared with subclavian puncture consistently from the early period since implantation (3.2% vs 1.9% and 23.1% vs 17.6% at 2 and 5 years; statistical significance was not shown), which conflicts with the previous studies reporting the lower failure rate of pacemaker leads in ESCP compared with the subclavian puncture.10, 26 Of importance, this lead has been reported extremely at risk, and the failure rate in this Canadian cohort9 was higher compared with that reported by Medtronic based on the remote monitoring (16.4% vs 11.7% at 5 years since implantation). Further analyses are needed to identify whether the use of CD instead of the ESCP contributes to the decrease of the failure in this fragile lead.

4.4. Predictor of lead failure

Older age, contrary to the prior studies,27 was a modest risk of lead failure by univariate analysis in the present study. In addition, neither higher LVEF nor the absence of the structural heart disease indicating the higher physical activity was associated with the risk of the lead failure. While the difference in the patient or lead characteristics was a potential reason for the discrepancy with the prior studies, it may also be explained that the extrathoracic approach attenuated the mechanical stress of the defibrillator lead at the proximal portion and contributed to decrease the lead failure in those with higher physical activity.

4.5. Limitations

We admit several limitations in our retrospective observational study conducted in a single center. Firstly, the number of the patients treated with CD was smaller and those patients were followed longer compared with the patients with ECSP approach. Secondly, patient characteristics were different between the two groups mainly because ESCP group included more patients underwent ICD therapy as a primary prevention. Thirdly, the ICD lead model was different between the two groups, which may have influenced the lead survival. Finally, the influence of the venous access on the recalled leads was not evaluated as all these recalled leads were inserted by ESCP.

5. CONCLUSIONS

The present study demonstrated an excellent long‐term survival of the transvenous ICD leads inserted by the contrast venography‐guided extrathoracic subclavian puncture in the consecutive patients, which was comparable with the cephalic vein cutdown. Further studies enrolling larger number of the patients are required to determine whether the extrathoracic puncture is eligible as a first‐line venous access for ICD lead implantation.

CONFLICT OF INTEREST

Authors declare no conflict of interests for this article.

Watanabe M, Yokoshiki H, Mitsuyama H, et al. Long‐term reliability of the defibrillator lead inserted by the extrathoracic subclavian puncture. J Arrhythmia. 2018;34:541–547. 10.1002/joa3.12107

REFERENCES

  • 1. Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators . A comparison of antiarrhythmic‐drug therapy with implantable defibrillators in patients resuscitated from near‐fatal ventricular arrhythmias. N Engl J Med. 1997;337:1576–1583. [DOI] [PubMed] [Google Scholar]
  • 2. Moss AJ, Hall WJ, Cannom DS, et al. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. Multicenter automatic defibrillator implantation trial investigators. N Engl J Med. 1996;335:1933–1940. [DOI] [PubMed] [Google Scholar]
  • 3. Moss AJ, Zareba W, Hall WJ, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002;346:877–883. [DOI] [PubMed] [Google Scholar]
  • 4. Bardy GH, Lee KL, Mark DB, et al. Amiodarone or an implantable cardioverter‐defibrillator for congestive heart failure. N Engl J Med. 2005;352:225–237. [DOI] [PubMed] [Google Scholar]
  • 5. Haqqani HM, Mond HG. The implantable cardioverter‐defibrillator lead: principles, progress, and promises. Pacing Clin Electrophysiol. 2009;32:1336–1353. [DOI] [PubMed] [Google Scholar]
  • 6. Furman S. Venous cutdown for pacemaker implantation. Ann Thorac Surg. 1986;41:438–439. [DOI] [PubMed] [Google Scholar]
  • 7. Aizawa Y, Negishi M, Kashimura S, et al. Predictive factors of lead failure in patients implanted with cardiac devices. Int J Cardiol. 2015;199:277–281. [DOI] [PubMed] [Google Scholar]
  • 8. Parsonnet V, Roelke M. The cephalic vein cutdown versus subclavian puncture for pacemaker/ICD lead implantation. Pacing Clin Electrophysiol. 1999;22:695–697. [DOI] [PubMed] [Google Scholar]
  • 9. Birnie DH, Parkash R, Exner DV, et al. Clinical predictors of fidelis lead failure: report from the Canadian heart rhythm society device committee. Circulation. 2012;125:1217–1225. [DOI] [PubMed] [Google Scholar]
  • 10. Chan NY, Kwong NP, Cheong AP. Venous access and long‐term pacemaker lead failure: comparing contrast‐guided axillary vein puncture with subclavian puncture and cephalic cutdown. Europace. 2017;19:1193–1197. [DOI] [PubMed] [Google Scholar]
  • 11. Roelke M, O'Nunain SS, Osswald S, Garan H, Harthorne JW, Ruskin JN. Subclavian crush syndrome complicating transvenous cardioverter defibrillator systems. Pacing Clin Electrophysiol. 1995;18:973–979. [DOI] [PubMed] [Google Scholar]
  • 12. Belott PH. Blind axillar venous access. Pacing Clin Electrophysiol. 1999;22:1085–1089. [DOI] [PubMed] [Google Scholar]
  • 13. Gardini A, Benedini G. Blind extrathoracic subclavian venipuncture for pacemaker implant: a 3‐year experience in 250 patients. Pacing Clin Electrophysiol. 1998;21:2304–2308. [DOI] [PubMed] [Google Scholar]
  • 14. Calkins H, Ramza BM, Brinker J, et al. Prospective randomized comparison of the safety and effectiveness of placement of endocardial pacemaker and defibrillator leads using the extrathoracic subclavian vein guided by contrast venography versus the cephalic approach. Pacing Clin Electrophysiol. 2001;24:456–464. [DOI] [PubMed] [Google Scholar]
  • 15. Ramza BM, Rosenthal L, Hui R, et al. Safety and effectiveness of placement of pacemaker and defibrillator leads in the axillary vein guided by contrast venography. Am J Cardiol. 1997;80:892–896. [DOI] [PubMed] [Google Scholar]
  • 16. Seto AH, Jolly A, Salcedo J. Ultrasound‐guided venous access for pacemakers and defibrillators. J Cardiovasc Electrophysiol. 2013;24:370–374. [DOI] [PubMed] [Google Scholar]
  • 17. Swerdlow CD, Kalahasty G, Ellenbogen KA. Implantable cardiac defibrillator lead failure and management. J Am Coll Cardiol. 2016;67:1358–1368. [DOI] [PubMed] [Google Scholar]
  • 18. Mizukami K, Yokoshiki H, Mitsuyama H, et al. Predictors of high defibrillation threshold in patients with implantable cardioverter‐defibillator using a transvenous dual‐coil lead. Circ J. 2015;79:77–84. [DOI] [PubMed] [Google Scholar]
  • 19. Hauser RG, Maisel WH, Friedman PA, et al. Longevity of sprint fidelis implantable cardioverter‐defibrillator leads and risk factors for failure: implications for patient management. Circulation. 2011;123:358–363. [DOI] [PubMed] [Google Scholar]
  • 20. Parsonnet V, Bernstein AD, Lindsay B. Pacemaker‐implantation complication rates: an analysis of some contributing factors. J Am Coll Cardiol. 1989;13:917–921. [DOI] [PubMed] [Google Scholar]
  • 21. Kleemann T, Becker T, Doenges K, et al. Annual rate of transvenous defibrillation lead defects in implantable cardioverter‐defibrillators over a period of >10 years. Circulation. 2007;115:2474–2480. [DOI] [PubMed] [Google Scholar]
  • 22. Dorwarth U, Frey B, Dugas M, et al. Transvenous defibrillation leads: high incidence of failure during long‐term follow‐up. J Cardiovasc Electrophysiol. 2003;14:38–43. [DOI] [PubMed] [Google Scholar]
  • 23. Kron J, Herre J, Renfroe EG, et al. Lead‐ and device‐related complications in the antiarrhythmics versus implantable defibrillators trial. Am Heart J. 2001;141:92–98. [DOI] [PubMed] [Google Scholar]
  • 24. Eckstein J, Koller MT, Zabel M, et al. Necessity for surgical revision of defibrillator leads implanted long‐term: causes and management. Circulation. 2008;117:2727–2733. [DOI] [PubMed] [Google Scholar]
  • 25. Epstein AE, Baker JH 2nd, Beau SL, Deering TF, Greenberg SM, Goldman DS. Performance of the st jude medical riata leads. Heart Rhythm. 2009;6:204–209. [DOI] [PubMed] [Google Scholar]
  • 26. Olgun H, Karagoz T, Celiker A, Ceviz N. Patient‐ and lead‐related factors affecting lead fracture in children with transvenous permanent pacemaker. Europace. 2008;10:844–847. [DOI] [PubMed] [Google Scholar]
  • 27. Morrison TB, Rea RF, Hodge DO, et al. Risk factors for implantable defibrillator lead fracture in a recalled and a nonrecalled lead. J Cardiovasc Electrophysiol. 2010;21:671–677. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Arrhythmia are provided here courtesy of Japanese Heart Rhythm Society