An Analysis of Foot and Ankle Device Recalls by the Food and Drug Administration

Carl Pellerin 1, Vinod Panchbhavi 2, Cory F Janney 3,
Editors: Alexander Muacevic, John R Adler
PMCID: PMC6177061  PMID: 30338198

Abstract

Introduction

Orthopaedic devices represent 12% of all medical device recalls. Products are approved through pre-market approval (PMA) or the 510(k) premarket notification process. No previous evaluation was found in the literature evaluating foot and ankle device recalls. The field of foot and ankle subspecialty has seen a rapid growth in innovation related to implants in recent years.

Methods

The Food and Drug Administration (FDA) Device Recall database was evaluated for all foot and ankle devices from 2007 through 2017 for the manufacturer, process of approval, type of implant, recall class, dates of initiation and termination of the recall, manufacturer determined reason, quantity affected, and distribution within the United States or internationally.

Results

A total of 161 products from 33 companies were identified with 158 (98.1%) approved through the 510(k) process. The most common reason for device recall was due to the device breaking intraoperatively or postoperatively. The average length of the recall was 487.5 days. 

Conclusions

Device recall is not an uncommon event with the majority of products approved through the less demanding 510(k) process. 

Keywords: foot, ankle, fda, device, recall

Introduction

Medical device recalls are common. According to the U.S. Food and Drug Administration (FDA), there were 20,093 recalls instituted by 1,641 manufacturers between November 2002 and December 2012 [1]. The incidence of recalls has not decreased over time. This unfortunate circumstance was confirmed as the FDA implemented 13,937 additional device recalls from 2013 through 2017. Of these additional product recalls, 3,202 occurred in 2017 alone [2].

The FDA is responsible for overseeing and regulating medical devices within the United States. Medical devices are typically approved through two methods: the premarket approval (PMA) process and the 510(k) premarket notification process. The PMA process has proven to be more expensive and time-consuming as it requires clinical evidence for authorization [3]. On the other hand, the 510(k) premarket notification is an accelerated process that omits medical devices from the clinical trial requirements as long as the device is “substantially equivalent” to an alternatively utilized medical device [4]. Defective medical devices have been recalled after gaining prior FDA authorization regardless of approval process utilized.

The United States medical device market has attracted a large sum of wealth and expenses throughout the years. In fact, this market was estimated to be over $133 billion in 2016 [5]. The economic impact for the numerous categories of medical devices varies significantly [6]. In the field of orthopaedics, the majority of medical devices are utilized for procedures such as joint replacement (projected U.S. market of $10.3 billion in 2018) and fracture management (projected U.S. market of $4.3 billion in 2015) [3, 7-8].

Although various medical devices have been utilized in operating rooms throughout the years, there is limited research evaluating the rationalization behind orthopaedic device recalls. Orthopaedic devices represent 12% of all medical device recalls [9]. Yet, no previous study was found in the literature examining foot and ankle devices with regards to recalls. Thus, this paper aims to compile and evaluate the reasons behind why ankle and foot orthopaedic devices have been recalled between 1 January 2007 through 31 December 2017. We hypothesized that products approved through the 510(k) approval process would have a higher recall rate.

Materials and methods

The Food and Drug Administration Device Recall database was queried for all foot and ankle devices from 1 January 2007 through 31 December 2017 by searching for the following terms: foot, ankle, toe, tibiotalar, subtalar, talus, calcaneus, navicular, cuboid, cuneiform, metatarsal, and phalanx [10]. Duplicates were removed from the search. The products were then verified as foot and ankle devices which may be used by a foot and ankle orthopaedic surgeon or podiatrist. Each recall was then used to evaluate the manufacturer, type of implant, recall class, date of initiation of the recall, date of termination of the recall, manufacturer determined reason, FDA determined reason, quantity affected, submission type, and the distribution within the United States or internationally.

Results

A total of 161 products from 33 different companies were identified (Table 1). The most common reason products were recalled was due to the device breaking intraoperatively or postoperatively, with one company recalling 78,594 suture anchors due to fractures of the hardware requiring revision surgery. Device breakage accounted for 39 of the 161 medical device recalls of implants or devices used during surgery with 12 occurring interoperatively and 26 postoperatively. One product was a prosthetic foot that could break under normal usage causing compromise to stability and function. The second-most common reason for recall was sterility concerns, which accounted for 37 recalls. Labeling issues and device design issues were also common reasons for recall, accounting for 30 and 27 recalls respectively. Employee error caused the least amount of recalls, with only one product removed from the market (Figure 1).

Table 1. Manufacturer-determined reasoning for recall of the devices from January 1, 2007 through December 31, 2017.

Recall reason Number of products recalled
Product breakage 39
Sterility issue 37
Labeling design 30
Device design 27
Manufacturing defect 22
Software design 3
Packaging issue 2
Employee error 1
Total products 161

Figure 1. Manufacturer-determined reasoning for all products recalled.

Figure 1

The average length of the active recall time from initiation to termination was 487.5 days. The smallest number of recalled products in commerce were two sizes of Mitek SpiraLok anchors (DePuy Mitek Inc., Raynham, Massachusetts, US) with zero distributed that were recalled secondary to concerns for hardware failure. The largest number of recalled products were Synthes screws (Synthes Products LLC, West Chester, Pennsylvania) (cancellous, cortical, and cancellous of various sizes) with 19,497,844 screws recalled due to labeling issues.

Of the 161 medical devices that were recalled, 158 (98.1%) were approved through the 510(k) premarket notification process. Conversely, only one product was authorized through the premarket approval (PMA) process. The submission procedure was not indicated for the remaining two medical devices (Figure 2).

Figure 2. FDA approval process utilized by the foot and ankle devices recalled during the selected timeframe.

Figure 2

FDA - Food and Drug Administration.

Discussion

During the recall process, it is the manufacturers’ duty to notify surgeons. The surgeons should then identify the patients affected and notify them, making sure they understand the reasons for the recall as well as what the risks are to the patient [11]. Although the surgeon may be concerned about litigation, the cases are categorized as product liability cases and generally are not considered the surgeon’s fault [12]. These patients should be followed frequently but prophylactic surgery is generally not recommended [13].

The overwhelming majority of recalls involved products approved through the 510(k) premarket notification process. This process does not involve the same stringent standards as the PMA application process. The PMA process requires clinical evidence before approval is granted, often requiring evidence on the level of randomized clinical trials or prospective data compared with historical controls [3-4, 14]. This study further supports the idea that products approved through the PMA process are less likely to be recalled [1, 15]. There is a lack of clinical data with products approved through the 510(k) process that may cause the devices to carry a greater risk to patients. With regards to total hip arthroplasty implants, it is estimated that only 15% of these devices had data on clinical effectiveness published [16].

The 510(k) process allows a company to fast-track approval of a device by claiming it “substantially equivalent” to a device that has already been approved and marketed. Claiming that one device is equivalent to another device may be illogical even when the predicate comes from the same company [17-18]. This process may also lead to devices being approved on the predicate from another device that was itself approved through the 510(k) process, sometimes leading to decades between the original device and subsequent predicate devices without an increase in evidence of the product working [17].

Products breaking and sterility concerns were the most likely reasons for recall. Issues with products breaking intraoperatively or postoperatively may have been determined previously if the products had undergone the more demanding PMA process for approval. Sterility issues are often related to packaging which may not have been determined through a different testing protocol. 

This study does have several limitations. We focused only on implants utilized within foot and ankle orthopaedic surgery by attempting to identify these products within the FDA database by searching for a wide array of terms within the database. If a product did not register one of the keywords we utilized, we may have missed some of the devices that were recalled. We also were unable to identify the approval process for two devices: the SYMBIONIC LEG (Ossur Americas Inc., Foothill Ranch, California, US) and Centurion Sterile Rubber Bands (#84 Rubber Bands Reorder EB84, Centurion Medical Products Corporation, Howell, Michigan, US). The SYMBIONIC LEG is an integrated prosthetic leg for transfemoral and knee disarticulations that was recalled due to a software design error that failed to notify patients of a low battery before shutting down. The Centurion Sterile Rubber Bands were mislabeled as “latex free” erroneously. These rubber bands did, in fact, contain natural latex rubber and were recalled to prevent patients from having an allergic reaction.

Another limitation is that although the FDA-determined reasons were provided in the database, this may not correlate exactly with the manufacturer’s reason for initiating a recall. Therefore, we chose to focus on the manufacturer’s reasoning for initiating the recall. We were also unable to ascertain the costs associated with the initiation of these recalls or the direct effect on patients as this data is not readily available within the database.

Conclusions

A large proportion of implants used for foot and ankle surgery approved through 501(k) process were recalled when compared to the proportion of implants approved through the more stringent PMA process. Surgeons should stay abreast of recalls amongst the products that they use frequently. Although recalls are not often considered to be the fault of the operating surgeon, the patient’s best interest is served with close follow up and making sure the patients is knowledgeable regarding the issues associated with the products themselves. Streamlining the PMA process or changing the 510(k) process may enable manufacturers to improve upon the safety of their devices.

The content published in Cureus is the result of clinical experience and/or research by independent individuals or organizations. Cureus is not responsible for the scientific accuracy or reliability of data or conclusions published herein. All content published within Cureus is intended only for educational, research and reference purposes. Additionally, articles published within Cureus should not be deemed a suitable substitute for the advice of a qualified health care professional. Do not disregard or avoid professional medical advice due to content published within Cureus.

The views expressed in this article are those of the author(s) and do not necessarily reflect the official policy or position of the Department of the Navy, Department of Defense, or the United States Government. LCDR Cory Janney is a military service member (or employee of the U.S. Government). This work was prepared as part of his official duties. Title 17, USC, §105 provides that ‘Copyright protection under this title is not available for any work of the U.S. Government.’ Title 17, USC, §101 defines a U.S. Government work as a work prepared by a military service member or employee of the U.S. Government as part of that person’s official duties.

Vinod Panchbhavi declare(s) personal fees from Stryker. Vinod Panchbhavi declare(s) Editor for a Journal from Lippincott Williams & Wilkins (. Editor for Techniques in Foot & Ankle Surgery