Skip to main content
Clinical Endoscopy logoLink to Clinical Endoscopy
. 2018 Jun 1;51(5):450–462. doi: 10.5946/ce.2018.024

Endoscopic Ultrasound-Guided Gallbladder Drainage Using a Lumen-Apposing Metal Stent for Acute Cholecystitis: A Systematic Review

Deepanshu Jain 1, Bharat Singh Bhandari 2, Nikhil Agrawal 3, Shashideep Singhal 4,
PMCID: PMC6182281  PMID: 29852730

Abstract

Surgery remains the standard treatment for acute cholecystitis except in high-risk candidates where percutaneous transhepatic gallbladder drainage (PT-GBD), endoscopic transpapillary cystic duct stenting (ET-CDS), and endoscopic ultrasound-guided gallbladder drainage (EUS-GBD) are potential choices. PT-GBD is contraindicated in patients with coagulopathy or ascites and is not preferred by patients owing to aesthetic reasons. ET-CDS is successful only if the cystic duct can be visualized and cannulated. For 189 patients who underwent EUS-GBD via insertion of a lumen-apposing metal stent (LAMS), the composite technical success rate was 95.2%, which increased to 96.8% when LAMS was combined with co-axial self-expandable metal stent (SEMS). The composite clinical success rate was 96.7%. We observed a small risk of recurrent cholecystitis (5.1%), gastrointestinal bleeding (2.6%) and stent migration (1.1%). Cautery enhanced LAMS significantly decreases the stent deployment time compared to non-cautery enhanced LAMS. Prophylactic placement of a pigtail stent or SEMS through the LAMS avoids re-interventions, particularly in patients, where it is intended to remain in situ indefinitely. Limited evidence suggests that the efficacy of EUS-GBD via LAMS is comparable to that of PT-GBD with the former showing better results in postoperative pain, length of hospitalization, and need for antibiotics. EUS-GBD via LAMS is a safe and efficacious option when performed by experts.

Keywords: Acute cholecystitis, Endoscopic ultrasound, Lumen-apposing metal stent

Introduction

Approximately 20 million Americans are estimated to have gallstones, and approximately 300,000 cholecystectomies are performed annually [1]. Approximately 1%–2% of patients with asymptomatic gallstones become symptomatic each year. Acute cholecystitis, which is the most common complication of gallstones occurs in 10% of symptomatic patients [2].

Surgery—open or laparoscopic (performed preferentially) is the standard treatment for acute cholecystitis [3]. Cholecystectomy rates were observed to increase rapidly initially, after which they stabilized in the late 1990s and currently may even be declining in the U.S [4]. Surgery is contraindicated in a few patients with advanced age and multiple comorbidities. Non-surgical gallbladder drainage (GBD) can be performed via a percutaneous transhepatic (PT) or an endoscopic route with or without an endoscopic ultrasound (EUS).

Perihepatic ascites and coagulopathy are relative contraindications for PT-GBD. PT-GBD causes postoperative pain in approximately 12% of patients [5]. Additionally, catheter obstruction, dislodgement and migration require multiple post-procedural interventions and long-term care. PT-GBD can be useful as either a bridge to surgery or occasionally is planned as destination therapy, although patients may report discomfort from external drainage tubes and reject this procedure owing to aesthetic reasons.

In the absence of EUS, endoscopic drainage of the GB has been attempted via a transpapillary approach aimed at stenting the cystic duct. This technique is feasible only if the cystic duct shows opacification on a cholangiogram and a guide-wire can be successfully passed into the cystic duct. Reportedly, the technical success rate (TSR) of this approach is approximately 91% [6]. Common complications include bleeding, stent migration or occlusion, sedation-related adverse events, and those associated with the endoscopic retrograde cholangiopancreatography (ERCP) procedure itself.

EUS-GBD is a relatively newer application of the EUS-guided approach to create an iatrogenic fistula between the GB and the gastrointestinal (GI) tract. EUS-GBD has been performed using plastic stents, nasobiliary drainage tubes, or self-expandable metal stents (SEMS); however, the high risk of stent migration and bile leakage secondary to lack of apposition of the luminal walls has been considered an obstacle to achieving safe clinical outcomes. A lumen-apposing metal stent (LAMS) is a saddle-shaped stent designed from braided nitinol wires with double-walled flanges perpendicular to the lumen on both sides to provide adequate anchorage across the non-adherent luminal structures. It is completely covered with silicone to counteract tissue ingrowth and tract leakage and to allow removability. In this article, we summarize the indications, technique, outcomes, and adverse events associated with EUS-GBD using LAMS based on our review of 10 original studies.

MATERIALS AND METHODS

Two authors individually reviewed the English literature from inception through September 2017. PubMed and Google Scholar were used to identify peer-reviewed original and review articles using the following key words: “Lumen-apposing metal stents (LAMS)”, “gallbladder drainage”, and “endoscopic ultrasound”. Only studies involving humans were selected. The references cited for pertinent studies were manually searched to identify additional relevant studies. Our search results yielded 10 studies [7-16]. Indications, procedural details, technical and clinical outcomes, as well as adverse events with their management were reviewed for each study.

RESULTS

Ten original studies were included in our review article. All were retrospective studies [7-15] except 1 [16]. Three studies enrolled patients from multiple centers [11,15,16] and the rest were single-center studies [7-10,12-14]. The reported literature described studies from the US [8,10-13,15], Spain [7], Japan [9], Germany [14] and the Netherlands [16]. Details regarding all studies have been summarized in Table 1.

Table 1.

Descriptive Summary of Each Individual Study

Study Location Type of study Number of Patients Indication/Disease Age (yr) and sex Technical outcome Clinical outcome Follow up duration Stent left/re-trieved Adverse events
de la Serna-Higuera et al. (2013) [7] Spain Retrospective Case series 13 Indication- 1. Mean age-79.9 (range, 57-97) 1. Success-11/13 (84.6%) Success- 11/11 (100%) 1. Mean-100.81 days (range, 24-210) 1. Stent left in situ- 10/11 Composite procedure related AE- 2
1. Acute cholecystitis (non-surgical candidate) 2. Gender distribution- 2. Failure of insertion- 2/13 Parameter- 2. Stent retrieved-1/11(no replacement stent placed due to symptom resolution and collapse of GB wall) 1. Scant hematochezia without anemia- 1/2 (resolved with conservative Mx)
Etiology- a) F- 5 a) 1/2- Tight cobblestone GB 1. Immediate symptom relief 2. Mild right upper quadrant pain-1/2 (resolved with conservative Mx)
1. Cholelithiasis-9/13 b) M- 8 b) 1/2- Uncontrolled stent release- complete deployment into the gastric lunien- 2. Normalization of LFTs and acute phase reactants
2. Cholelithiasis+ Pancreatic cancer-2/13
3. Cholelithiasis+ C holangiocarc inoma-1/13
4. Cholangiocarcinoma-1/13
Itoi et al. (2013) [8] USA Case Report 1 Indication- 1. Age-57 Success-1/1 (100%) Success-1/1 (100%) 1 yr 1. Stent left in situ-1/1 None
1. Acute cholangitis (ERCP technically challenging) 2. Gender distribution- Parameter-
Etiology- a) F- 0 1. Clinical symptoms
1. Pancreatic head mass with concomitant duodenal and biliary obstruction b) M- 1
Itoi et al. (2014) [9] Japan Case report 1 Indication- 1. Age- 96 Success- 1/1 (100%) Success-1/1 (100%) N/A 1. Stent retrieved-1/1 (2 weeks post placement along with 20x30 mm gallstone removal with help of lithotripter) None
1. Acute cholecystitis (Non-surgical candidate) 2. Gender distribution-
a) F- 1
b) M- 0
Tharian etal. (2016) [10] USA Case report 1 Indication- 1. Age-81 Success-1/1 (100%) Success-1/1 (100%) 1 mo 1. Stent in situ-1/1 None
1. Distended GB 2. Gender distribution- Parameter-
Disease- a) F- 0 1. Clinical symptoms
1. Adenocarcinoma of GB neck b) M- 1
Irani et al. (2015) [11] USA Multicenter Retrospective 15 Indication- 1. Median age- 74 (range, 42-89) 1. Success-14/15 (93%) Success-15/15 (100%) Mean-160 (range, 39- 260) days 1. Stent in situ-15/15 Composite procedure related AE- 1
1. Acute calculous cholecystitis- 7/15 2. Gender distribution- 2. Success with assistance- 1/15 (salvaged by placing SEMS via LAMS)
2. Non calculous cholecystitis- 4/15 a) F-7 Modality-Phone calls, Clinic visits, Imaging studies 1. Post-procedure fever-1/15 (successfully treated with antibiotics)
3. Biliary obst ruction-2/15 b) M- 8
4. Gallbladder hydrops- 1/15
5. Symptomatic cholelithiasis-1/15 (All were non-surgical candidates and all refused percutaneous drainage)
Kumta et al. (2016) [12] USA Case report 1 Indication- chronic calculous cholecystitis (not a surgical candidate and refused percutaneous drainage) 1. Age- 77 years Success-1/1 (100%) Success-1/1 (100%) 6 mo 1. Stent left in situ-1/1 None
2. Gender distribution- Parameter-
a) F- 1 1. Clinical symptoms
Law et al. (2016) [13] USA Retrospective Single center Case series 7 Indication- acute calculous cholecystitis (prior percutaneous drain and poor surgical candidates) 1. Median Age- 57 (range, 32-81) Success-5/7 (71.4%) Success-7/7 (100 %) 4 mo (Interquartile range, 3.5-5.5) 1. Stent left in situ-4/7 None
2.Gender distribution- Success with assistance- 2/7 (salvaged with SEMS placement through LAMS) Paranieter-1. Clinical symptoms 2. Stent retrieved-3/7
a) F- 1 a) 2/3- replaced with double pigtail stent at 6 weeks and 4 mo
b) M - 6 b) 1/3- removed to allow stone extraction which spontaneously passed into ileostomy bag
Dollhopf et al. (2017) [14] Germany Retrospective Single center 75 Indication - acute calculous and acalculous cholecystitis (poor surgical candidates) 1. Median age- 75±11 (range, 41–96) 1. Success- 74/75 (98.7%) Success- 71/74 (95.9%) 1. Mean: 201±226 (range, 2–1,192) days 1. Stent left in situ- 69/75 Composite procedure related AE- 10
2. Gender distribution- 2. Failure- 1/75 (equipment malfunctioningleading to gastric perforation managed surgically) Parameter- 2. Stent retrieved- 5/75 1. Major bleeding- 1/10 (resolved with conservative management)
a) F- 39 1. Clinical symptoms 3. Stent migration- 1/75 (spontaneous, persistent cholecystogastric fistula with no clinical consequences) 2. Recurrent cholecystitis-3/10
b) M- 36 Failure- 3/74 (death on post procedure day 3, 13 and 27 secondary to worsening sepsis) a) 1/3- conservative management
b) 2/3- double pigtail stent placed via LAMS
3. Migration- 2/10
a) 1/2- proximal migration into GB at 8 m post procedure (endoscopic removal and replacement with another LAMS)
b) 1/2- intragastric migration at day 5 post procedure (endoscopic stent retrieval and closure of fistula with clip)
4. Bouveret syndrome- 1/10- 4 m post procedure (endoscopy and lithotripsy of occluding stone
5. Sepsis- 3/10- leading to death
Irani et al. (2017) [15] USA, Europe, Asia Retrospective Multicenter EUS guided GB drainage (EUS-GBD)-45 Indication - 1) Acute calculous and acalculous cholecystitis (poor surgical candidates) 1. Median age- 55 (range, 25–87) 1. Success- 44/45 (98%) 1. Success- 43/45 (96%) 215 (range, 1–621) days 1. Stent left in situ- 44/44 Composite procedure related AE- 8
2. Gender distribution 2. Success with assistance- 1/45 (salvaged with 10×60 mm fully covered biliary metal stent placed via LAMS) 2. Median pain score on post procedure Day 1- 2.5 (range, 1–9) 1. Bleeding- 2/8
a) F- 16 a) 1/2- 3 days post procedure (treated by clot evacuation and pigtail stent placement through LAMS)
b) M- 29 3. Median hospital stay post intervention- 3 (range, 1–23) b) 1/2- 6 mo post procedure (stopped spontaneously with reversal of coagulopathy)
2. Recurrent cholecystitis- 3/8 (6, 8 and 12 mo post procedure)
4. Number of re-interventions- 11 a) 1/3- treated with antibiotics
b) 2/3- endoscopic placement of pigtail stent via LAMS
3. Bile leak with peritonitis- 1/8- Day 3 post procedure (required percutaneous drain)
4. Abdominal pain- 1/8-due to food occluding trans-gastric LAMS (evacuation of food, balloon dilation of granulation overgrowth and pigtail stent placement via LAMS)
5. Sepsis- 1/8- perforated GB leading to death
Irani et al. (2017) [15] USA, Europe, Asia Percutaneous transhepatic drainage of GB (PT-GBD)- 45 1. Median age- 75 (34–94) Success- 45/45 (100%) (p=0.98) 1. Success- 41/45 (91%) (p=0.12) 265 (range, 1–1,638) days Not applicable Composite procedure related AE- 14
2.Gender distribution- 2. Median pain score on post procedure Day 1- 6.5 (range, 2–10) (p=0.001) 1. Recurrent cholecystitis- 4/14 (Trt with Abx and drain exchange)
a) F- 18 3. Median hospital stay post intervention- 9 (range, 1–121) (p=0.01) a) 1/4- drain dislodgement on post procedure day 8
b) M- 27 4. Number of re-interventions- 112 (p=0.001) b) 3/4- drain occlusion on post procedure 2, 4 and 6 mo
2. Abdominal pain without cholecystitis- 3/14 (drain occlusion Trt with exchange)
3. Cellulitis- 1/14 (Trt with oral Abx)
4. Bile leak- 3/14
a) 1/3- lead to sepsis and death
b) 2/3- additional drain placement
5. Sepsis- 2/14 (lead to death)
6. Jejunal fistula- 1/14 (allowed track to mature followed by EUS-GBD)
Walter et al. (2016) [16] Netherlands Prospective Multicenter 30 Indication- Acute calculous and acalculous cholecystitis (poor surgical candidates) 1. Median age- 85 (range, 68–97) 27/30 (90%) 26/27 (96%) 1. 298±82 days for all patients 1. Stent left in situ- 15/30 Composite procedure related AE- 6
2. Gender distribution 2. 364±82 days for patients alive at the end of study 2. Stent retrieved- 15/30 1. Recurrent cholecystitis- 2/6 (due to LAMS obstruction requiring its removal)
a) F- 19 2. Aspiration pneumonia- 1/6 (leading to death)
b) M- 11 3. Pancreatic infection- 1/6 (leading to death)
4. Melena/ thrombus in GB- 1/6 (resolved with conservative management)
5. Jaundice (hemobilia)- 1/6 (resolved with conservative management)

GB, gall bladder; LFT, liver function tests; AE, adverse events; Mx, management; ERCP, endoscopic retrograde cholangiopancreatography; N/A, not available; SEMS, self-expandable metal stent; LAMS, lumen-apposing metal stent; EUS-GBD, endoscopic ultrasound-guided gallbladder drainage; PT-GBD, percutaneous transhepatic gallbladder drainage; Trt, treatment; Abx, antibiotics.

Patient characteristics

Most patients in our study cohort were aged ≥60 years [7-16] with only a few patients aged <50 years [11,13,15]. Our study showed an even distribution of women (47.1%) and men (52.9%) in a ratio of 1:1.1 [7-16]. The retrospective studies that compared outcomes between EUS-GBD and PT-GBD showed no statistically significant intergroup differences in terms of baseline population characteristics, etiology of cholecystitis, site of obstruction, median thickness of the GB wall, median duration of cholecystitis, and antibiotic exposure prior to the planned intervention [15].

Indications

The most common indication for the procedure was acute cholecystitis [7,9,11-16] in addition to acute cholangitis [8], GB hydrops [10,11], or biliary obstruction [11]. The underlying etiology was either an isolated condition or a combination of cholelithiasis, choledocholithiasis, acalculous cholecystitis, or malignancies such as pancreatic adenocarcinoma, cholangiocarcinoma, or GB adenocarcinoma [7-16]. All such patients had been deemed poor surgical candidates and had either refused to undergo PT-GBD or preferred to undergo EUS-GBD [7-12,14-16]. A study showed that all patients who underwent PT-GBD continued to remain poor surgical candidates during follow-up and required management with internal drainage using EUS-GBD [13]. In patients with biliary obstruction, ERCP failed to relieve obstruction secondary to technically challenging anatomy [8]. In addition to patient preference, PT-GBD was considered unsafe in a few patients with concomitant perihepatic ascites, or coagulopathy, and in those who needed to continue the use of blood thinners, or were at a high risk of tube dislodgement (patients with dementia, and agitated behavior, among other such conditions) [11].

Procedural characteristics

Anesthesia

Only 4 studies described the nature of anesthesia used for the procedure, which varied from general [11,15], and monitored anesthesia [14,16] to conscious sedation [16]. These results are highly center specific. No patient-reported outcomes are available to compare procedure tolerance across different studies, although the TSR remained comparable independent of the type of anesthesia used.

Stent specifics

The LAMS is designed to allow apposition of the luminal walls of 2 non-adherent organs. The endoscopist determines the size and length of the LAMS to be used after considering the patient-specific anatomy, GB wall thickness, and diameter of in situ GB stones (to allow unobstructed passage of the stent). Details regarding stents used by different studies have been summarized in Table 2. Because of increasing experience with the use of LAMS, a few authors recommend prophylactic placement of a SEMS or a plastic pigtail stent through the LAMS to prevent future occlusion and displacement [7,11,13,15]. This practice is more useful in patients in whom the LAMS is intended to remain in situ indefinitely, to avoid the need for re-intervention. In addition to the above-mentioned studies [2], authors used SEMS to provide additional length when the proximal flange of the LAMS was noted to deploy into the extraluminal space [11,13,15] potentially salvaging a technical failure.

Table 2.

Technical Details of Procedure across Each Study

Study Location Type of study Number of Patients Site of approach (gastric or duodenum) Needle size Stent specifics Accessory equipment Anesthesia Procedure duration
de la Serna-Higuera et al. (2013) [7] Spain Retrospective Case series 13 1. Transgastric- 12/13 19 G A. LAMS- 1. Linear echoendoscope N/A N/A
2. Transduodenal- 1/13 1. 10×10 mm- 7/11 2. 0.035 inch guidewire
2. 15×10 mm- 4/11 3. 8.5 F Cystotome
3. Flange diameter- 20 mm (11/11) 4. 4 mm biliary balloon dilator
B. Coaxial SEMS within LAMS- 4/11 5. 10 mm balloon dilator
Itoi et al. (2013) [8] USA Case Report 1 Transgastric- 1/1 19 G A. LAMS 1. Linear echoendoscope N/A N/A
1. 10×10 mm- 1/1 2. 4 mm balloon dilator
2. Flange diameter- 20 mm (1/1)
Itoi et al. (2014) [9] Japan Case Report 1 Transgastric- 1/1 N/A A. LAMS N/A N/A N/A
1. Diameter- 15 mm
Tharian et al. (2016) [10] USA Case Report 1 Transduodenal- 1/1 19 G N/A 1. Linear echoendoscope N/A N/A
2. 0.025 inch guidewire
3. 4 mm balloon dilator
Irani et al. (2015) [11] USA Multicenter Retrospective 15 Transduodenal- 14/15 19 G A. LAMS 1. Linear echoendoscope General anesthesia 1. Median- 38 (range, 15–52) min
Transgastric- 1/15 1. 10×10 mm- 12/15 2. 0.025 inch visiglidewire or 0.035 inch jagwire
2. 15×10 mm- 3/15
3. Flange Diameter- 3. 4 mm balloon dilator or 6 F/7 F tapered dilator
a)21 mm- 12/15
b) 24 mm- 3/15 4. Over the wire needle knife or 10 F cystotome
B. Stent through LAMS
1. 7 F × 4 cm double pig tail stent- 6/15 (prophylactic)
2. 10×6 cm fully covered biliary metal stent- 1/15
Kumta et al. (2016) [12] USA Case Report 1 Transduodenal- 1/1 N/A A. LAMS with cautery 1. Linear echoendoscope N/A N/A
1. 15×10 mm 2. Guidewire
3. Balloon dilator
Law et al. (2016) [13] USA Retrospective Single center Case series 7 Transduodenal- 7 19 G A. LAMS (diameter/length) 1. Linear echoendoscope N/A N/A
1. 10×10 mm- 5/7 2. 450 cm- biliary guidewire
2. 15×10 mm- 2/7 3. Cystostome (10 Fr) or balloon dilator
B. Stent through LAMS- 5/7
1. Double pig tail stent (7 F × 4 cm)- 3/7
2. Biliary SEMS (10×6 cm) + double pig tail stent (7 F × 4 cm)- 1/7
3. Biliary SEMS (10 mm × 6 cm)+biliary SEMS (10 mm × 4 cm)- 1/7
Dollhopf et al. (2017) [14] Germany Retrospective Single center 75 Transduodenal- 38 1. 19 G- 32/75 (42.7%) A. LAMS (diameter/length) with cautery 1. Linear echoendoscope 1. Anesthesia monitored Mean- 26 (range, 8–60) min
Transgastric- 36 2. No needle use- 43/75 (LAMS cautery) 1. 10×10 mm- 65/75 2. 0.035 inch guidewire
Transjejunal- 1 2. 15×10 mm- 7/75
3. 8×8 mm- 2/75
4. 6×8 mm- 1/75
Irani et al. (2017) [15] USA, Europe, Asia Retrospective Multicenter EUS-GBD-45 1. Transduodenal- 32/45 19 G A. LAMS 1. Linear echoendoscope General anesthesia-40/45 Mean- 28 (range, 18–52) min
2. Transgastric- 13/45 1. 10×10 mm- 37/45 2. 0.035 inch Jagwire or 0.025 inch visiglide wire
2. 15×10 mm- 8/45
3. Flange diameter- 21 or 24 mm 3. 4 mm biliary balloon dilator
4. Saddle length- 10 mm (45/45) 4. 10 F cystostome or needle knife or blank
B. Stent through LAMS-
1. Plastic pigtail stent- 24/45
PT-GBD-45 Percutaneous transhepatic N/A 8 or 10 F self locking pigtail catheter N/A General anesthesia-5/45 (p<0.0001) Mean- 22 (range, 12–30) min (p=0.02)
Walter et al. (2016) [16] Netherlands Prospective Multicenter 30 Transduodenal- 19/30 19 G A. LAMS 1. Linear echoendoscope 1. Monitored anesthesia (propofol)- 4/30 Median- 15 (range, 13–110) min
Transgastric- 11/30 1. 10×10 mm- 13/30 2. 0.035 inch guidewire 2. Concious sedation (fentanyl and midazolam)- 26/30
2. 15×10 mm- 17/30 3. Cystostome or balloon dilator

LAMS, lumen-apposing metal stent; SEMS, self-expandable metal stent; N/A, not available; EUS-GBD, endoscopic ultrasound-guided gallbladder drainage; PT-GBD, percutaneous transhepatic gallbladder drainage.

In our review, approximately 50% of the operations (48.2%) were performed using cautery enhanced LAMS [12,14,15] and the rest were performed using non-cautery enhanced LAMS (51.8%) [7-11,13,15,16].

Site of puncture

Most studies did not explain their choice of a transgastric vs. a transduodenal or transjejunal approach. This decision is left to the discretion of the endoscopist after assessing the patient’s anatomy to determine the site of maximal direct apposition between the GB wall and the GI tract to allow LAMS placement. LAMS placement in porcine models is known to cause mature fistula tract formation within 4–5 weeks [17]. The LAMS tends to cause significant tissue overgrowth, which serves as a drawback during stent removal. Walter et al. proposed that the retroperitoneal location of the duodenum allows stable tract formation compared to the stomach where peristaltic movements lead to a higher degree of tissue reaction consequently leading to tissue overgrowth [16].

In our review comprising 189 patients, 75 (39.7%) underwent transgastric [7-9,11,14-16], 113 (59.8%) underwent transduodenal [7,10-16], and 1 (0.5%) patient underwent transjejunal puncture [14] to access the GB.

Technique

All authors used an oblique/forward viewing therapeutic linear array echoendoscope to visualize the GB. A standard 19-gauge needle was used to puncture the gastric antrum or the duodenal bulb to access the GB wall, and entry into the GB was confirmed using real-time ultrasound and color Doppler imaging. Next, the contrast agent was injected to obtain fluoroscopic images of the biliary tree. A standard biliary guide-wire was coiled into the GB lumen. The tract was then dilated over the guide-wire using a balloon dilator or cystotome. A preloaded stent was then advanced over the guide-wire. The distal flange of the stent was first deployed into the GB and abutted against the wall under ultrasonographic or fluoroscopic guidance. Next, the proximal flange of the stent was deployed into the duodenum or the stomach under direct visualization, thereby performing cholecystoenterostomy or cholecystogastrostomy, respectively. The stent position was confirmed endoscopically or fluoroscopically and via aspiration of biliary contents. If a cautery enhanced LAMS is being used, the puncture site, dilation of the tract and stent deployment can all be performed simultaneously, thereby potentially decreasing the procedure time.

Procedure time

Only 4 studies reported procedure times [11,14-16], which varied widely between a minimum of 8 min [14] and a maximum of 110 min [16]. The mean procedure time varied from 22–38 min [11,14,15]. The mean EUS-GBD procedure time was 6 min longer than that of a PT-GBD (p=0.02) [15]. Dollhopf et al. reported a significantly (p=0.04) shorter stent deployment time with the use of cautery enhanced LAMS (3.1 min) vs. non-cautery enhanced LAMS (7.7 min) [14].

Technical outcome

The TSR was determined by calculating the percentage of patients who underwent successful LAMS deployment across the GB wall with drainage into the stomach or the small bowel. The composite TSR (without SEMS) for EUS-GBD was 95.2% (180/189) [7-16]. The LAMS could not initially be deployed in 3 patients; however, a SEMS was successfully placed through the LAMS to bridge the gap across the GB wall to overcome this difficulty [11,13,15]. The composite TSR (with SEMS) for EUS-GBD was 96.8% (183/189) [7-16]. The individual TSR (without SEMS) varied between 71.4% [13] and 100% [8-10,12]. The individual TSR (with SEMS) varied between 84.6% [7] and 100% [8-13,15]. Reasons for failure (n=6) could be attributed to a stiff cobblestone GB (1/6) [7], uncontrolled stent release (1/6) [7], equipment malfunction leading to gastric perforation (1/6) [14], and unknown etiology (3/6) [16]. A study performed by Irani et al. showed that the TSR was comparable between EUS-GBD and PT-GBD (p=0.98) [15].

Clinical outcomes

The clinical success rate (CSR) was measured by calculating the percentage of patients who showed significant improvement in clinical, laboratory, or radiological parameters post EUS-GBD via LAMS (with or without SEMS). The composite CSR was 96.7% (177/183) [7-16]. The individual CSR varied between 95.9% [14] and 100% [7-13]. Among the 6 patients who showed a lack of response, GB wall perforation was considered the possible etiology in 2 [14,15], and no obvious reason could be detected in the remaining 4 patients other than coexistent comorbidities [14-16]. The CSR of EUS-GBD (96%) was higher than that of PT-GBD (91%), although this result was not statistically significant (p=0.12) [15]. Notably, the authors demonstrated a statistically significant decrease in the postoperative pain on day 1 (p=0.001), a shorter median length of hospitalization post-intervention (p=0.01), as well as fewer re-interventions (p=0.001) with EUS-GBD than with PT-GBD [15].

Adverse events

Sepsis

Lack of resolution of acute cholecystitis despite stent placement resulted in progression to sepsis, necrosis of the GB wall, perforation, and death [14,15,16]. These complications have been presented in detail under Clinical Outcomes.

Recurrent cholecystitis

Among the 177 patients who showed initial clinical resolution of symptoms, 9 (5.1%) presented with a repeat attack of acute cholecystitis. This condition usually occurs secondary to either stent dislodgement (usually in the early post-procedural period before the fistula matures into a tract), or occlusion (an early or late post-procedural complication) secondary to gallstones, food bolus, clots, or tissue overgrowth over time. Patients may improve with conservative management (antibiotics) alone [14,15] or may require endoscopic debridement with placement of a pigtail stent through the LAMS or complete removal of the LAMS. A few authors recommend prophylactic placement of pigtail stents through the LAMS at the time of initial procedure to prevent this complication [7,11,13,15]. A few authors also recommend performing endoscopic lavage with/without stone extraction via the LAMS after placement to prevent future attacks [7].

Bleeding

Intraprocedural bleeding is an expected outcome because the procedure involves the creation of a transmural fistula between the GI tract and the GB wall. Bleeding was easily controlled in most patients and was not reported as a complication. Among 189 patients, 5 (2.6%) reported post-procedural GI bleeding [7-16], which in 4/5 patients were managed with conservative treatment alone (reversal of coagulopathy, administration of intravenous fluids, and other such interventions) [7,14-16]. In 1 patient, endoscopic evaluation revealed a clot without active bleeding, which was treated with clot evacuation and placement of a pigtail stent through the LAMS [15]. One patient developed hemobilia, which improved with supportive care alone [16].

Stent migration

Stent migration is a relatively rare complication with LAMS placement compared to the placement of SEMS or pigtail catheters.

Among the 183 patients who underwent EUS-GBD via LAMS, only 2 (1.1%) showed stent migration [7-16]. One event occurred on post-procedure day 5 (managed with stent retrieval and clip closure of the fistula) and the other occurred 8 months post procedure (managed with LAMS replacement) [14].

Miscellaneous

These complications noted in the composite study cohort (n=189) included abdominal pain in 1.1% of patients (1 patient developed stent occlusion from a food bolus requiring endoscopic debridement and pigtail stent placement, and the other improved with supportive care alone) [7,15], fever in 0.5% (improved with supportive care) [11], bile leak with peritonitis in 0.5% (managed with percutaneous drain placement) [15], Bouveret syndrome in 0.5% (treated with endoscopic lithotripsy and stone removal) [14], aspiration pneumonia in 0.5% (eventually leading to death) [16] and pancreatic infection in 0.5% (eventually leading to death) [16].

CONCLUSIONS

The absence of an external drainage tube and widespread applicability in patients with coagulopathy or ascites make EUS-GBD using LAMS an attractive option for patients with acute cholecystitis in whom surgery is contraindicated. The procedure demonstrates a high TSR and CSR, although this evidence is based on the few high-volume centers where only a small number of experts perform this procedure. A learning curve has been suggested; however, the minimum training requirements are yet to be determined. Complications although rare, are associated with the risk of worsening the clinical situation in already sick patients who are poor surgical candidates. Based on the available evidence, EUS-GBD using LAMS is a safe and efficacious alternative when performed by experts in high-risk surgical patients with acute cholecystitis.

Footnotes

Conflicts of Interest:The authors have no financial conflicts of interest.

Author Contributions

Conceptualization: Deepanshu Jain

Data curation: DJ, Bharat Singh Bhandari, Nikhil Agrawal

Formal analysis: DJ

Investigation: DJ, Shashideep Singhal

Methodology: DJ

Project administration: DJ, SS

Resources: DJ

Supervision: DJ, SS

Validation: DJ, SS

Visualization: DJ

Writing-original draft: DJ, BSB, NA

Writing-review&editing: DJ, SS

REFERENCES

  • 1.Hassler KR, Jones MW. Treasure Island (FL): StatPearls Publishing; c2017. Gallbladder, cholecystectomy, laparoscopic [Internet] [updated 2017 Oct 16; cited 2018 Apr 12]. Available from: https://www.ncbi.nlm.nih.gov/books/NBK448145/ [Google Scholar]
  • 2.Friedman GD, Raviola CA, Fireman B. Prognosis of gallstones with mild or no symptoms: 25 years of follow-up in a health maintenance organization. J Clin Epidemiol. 1989;42:127–136. doi: 10.1016/0895-4356(89)90086-3. [DOI] [PubMed] [Google Scholar]
  • 3.Garden OJ, Bradbury AW, Forsythe JLR. Principles and practice of surgery. 4th ed. Edinburgh: Churchill Livingstone; 2002. [Google Scholar]
  • 4.Zacks SL, Sandler RS, Rutledge R, Brown RS., Jr A population-based cohort study comparing laparoscopic cholecystectomy and open cholecystectomy. Am J Gastroenterol. 2002;97:334–340. doi: 10.1111/j.1572-0241.2002.05466.x. [DOI] [PubMed] [Google Scholar]
  • 5.Itoi T, Sofuni A, Itokawa F, et al. Endoscopic transpapillary gallbladder drainage in patients with acute cholecystitis in whom percutaneous transhepatic approach is contraindicated or anatomically impossible (with video) Gastrointest Endosc. 2008;68:455–460. doi: 10.1016/j.gie.2008.02.052. [DOI] [PubMed] [Google Scholar]
  • 6.Widmer J, Alvarez P, Sharaiha RZ, et al. Endoscopic gallbladder drainage for acute cholecystitis. Clin Endosc. 2015;48:411–420. doi: 10.5946/ce.2015.48.5.411. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.de la Serna-Higuera C, Pérez-Miranda M, Gil-Simón P, et al. EUS-guided transenteric gallbladder drainage with a new fistula-forming, lumen-apposing metal stent. Gastrointest Endosc. 2013;77:303–308. doi: 10.1016/j.gie.2012.09.021. [DOI] [PubMed] [Google Scholar]
  • 8.Itoi T, Binmoeller K, Itokawa F, Umeda J, Tanaka R. Endoscopic ultrasonography-guided cholecystogastrostomy using a lumen-apposing metal stent as an alternative to extrahepatic bile duct drainage in pancreatic cancer with duodenal invasion. Dig Endosc. 2013;25 Suppl 2:137–141. doi: 10.1111/den.12084. [DOI] [PubMed] [Google Scholar]
  • 9.Itoi T, Itokawa F, Tsuchiya T, Kurihara T, Tanaka R. Transgastric large gallstone extraction through a lumen-apposing metal stent in a patient with acute cholecystitis. Gastrointest Endosc. 2014;79:547. doi: 10.1016/j.gie.2013.12.010. [DOI] [PubMed] [Google Scholar]
  • 10.Tharian B, Varadarajulu S, Hawes R. Drainage of obstructed gallbladder with use of lumen-apposing metal stent. Gastrointest Endosc. 2016;83:460–461. doi: 10.1016/j.gie.2015.08.016. [DOI] [PubMed] [Google Scholar]
  • 11.Irani S, Baron TH, Grimm IS, Khashab MA. EUS-guided gallbladder drainage with a lumen-apposing metal stent (with video) Gastrointest Endosc. 2015;82:1110–1115. doi: 10.1016/j.gie.2015.05.045. [DOI] [PubMed] [Google Scholar]
  • 12.Kumta NA, Lordello Passos M, Rodela Silva GL, Novikov A, Kahaleh M. Endoscopic ultrasound-guided transmural gallbladder drainage with a lumen-apposing metal stent using an electrocautery enhanced delivery system. Endoscopy. 2016;48:E327. doi: 10.1055/s-0042-117634. [DOI] [PubMed] [Google Scholar]
  • 13.Law R, Grimm IS, Stavas JM, Baron TH. Conversion of percutaneous cholecystostomy to internal transmural gallbladder drainage using an endoscopic ultrasound-guided, lumen-apposing metal stent. Clin Gastroenterol Hepatol. 2016;14:476–480. doi: 10.1016/j.cgh.2015.10.026. [DOI] [PubMed] [Google Scholar]
  • 14.Dollhopf M, Larghi A, Will U, et al. EUS-guided gallbladder drainage in patients with acute cholecystitis and high surgical risk using an electrocautery-enhanced lumen-apposing metal stent device. Gastrointest Endosc. 2017;86:636–643. doi: 10.1016/j.gie.2017.02.027. [DOI] [PubMed] [Google Scholar]
  • 15.Irani S, Ngamruengphong S, Teoh A, et al. Similar efficacies of endoscopic ultrasound gallbladder drainage with a lumen-apposing metal stent versus percutaneous transhepatic gallbladder drainage for acute cholecystitis. Clin Gastroenterol Hepatol. 2017;15:738–745. doi: 10.1016/j.cgh.2016.12.021. [DOI] [PubMed] [Google Scholar]
  • 16.Walter D, Teoh AY, Itoi T, et al. EUS-guided gall bladder drainage with a lumen-apposing metal stent: a prospective long-term evaluation. Gut. 2016;65:6–8. doi: 10.1136/gutjnl-2015-309925. [DOI] [PubMed] [Google Scholar]
  • 17.Binmoeller KF, Shah J. A novel lumen-apposing stent for transluminal drainage of nonadherent extraintestinal fluid collections. Endoscopy. 2011;43:337–342. doi: 10.1055/s-0030-1256127. [DOI] [PubMed] [Google Scholar]

Articles from Clinical Endoscopy are provided here courtesy of Korean Society of Gastrointestinal Endoscopy

RESOURCES