Skip to main content
Annals of Clinical Microbiology and Antimicrobials logoLink to Annals of Clinical Microbiology and Antimicrobials
. 2018 Oct 10;17:36. doi: 10.1186/s12941-018-0289-2

An update on gastrointestinal endoscopy-associated infections and their contributing factors

Charles Eugenio McCafferty 1,2,, Marra Jai Aghajani 1,2, David Abi-Hanna 3,4, Iain Bruce Gosbell 2, Slade Owen Jensen 1,2
PMCID: PMC6182826  PMID: 30314500

Abstract

Introduction

During clinical use, gastrointestinal endoscopes are grossly contaminated with patient’s native flora. These endoscopes undergo reprocessing to prevent infectious transmission upon future use. Endoscopy-associated infections and outbreaks have been reported, with a recent focus on the transmission of multi-drug resistant organisms. This review aims to provide an update on endoscopy-associated infections, and the factors contributing to their occurrence.

Methods

PubMed, ScienceDirect, and CINAHL were searched for articles describing gastrointestinal endoscopy-associated infections and outbreaks published from 2008 to 2018. Factors contributing to their occurrence, and the outcomes of each outbreak were also examined.

Results

This review found 18 articles, 16 of which described duodenoscope-associated infections, and the remaining two described colonoscope- and gastroscope-associated infection respectively. Outbreaks were reported from the United States, France, China, Germany, the Netherlands and the United Kingdom. The causative organisms reported were Klebsiella pneumoniae, Pseudomonas aeruginosa, Escherichia coli and Salmonella enteritidis.

Conclusions

A number of factors, including lapses in reprocessing, biofilm formation, endoscope design issues and endoscope damage, contribute to gastrointestinal endoscopy associated infection. Methods of improving endoscope reprocessing, screening for contamination and evaluating endoscope damage may be vital to preventing future infections and outbreaks.

Background

During clinical use, endoscopes are contaminated with patients’ native flora [1]. These infectious agents must be removed to prevent cross-infection during subsequent use [24]. Endoscopes usually undergo reprocessing rather than sterilization [4]. Endoscope reprocessing is performed in Australia according to the guidelines provided by the Gastroenterological Nurses College of Australia and Gastroenterological Society of Australia (GENCA/GESA) [5]. Reprocessing typically consists of 8 steps: precleaning, leak testing, manual cleaning, rinsing after cleaning, visual inspection, high-level disinfection, rinsing after high-level disinfection, and drying [5].

Endoscopes cannot be autoclaved because they are made of heat-sensitive materials [6]. There are low-temperature alternatives, including ethylene oxide (EtO) gas sterilisation, which have been adopted in response to culture-positive endoscopes and endoscopy-associated outbreaks [7, 8]. EtO gas sterilization is more expensive than standard reprocessing, requires more time, poses a risk to the health of associated staff, and has been suggested to shorten an endoscopes lifespan [7, 9, 10].

Endoscope reprocessing has been associated with a variable failure rate, one study reporting gastroscope and colonoscope reprocessing failure rates of 1.8% and 1.9% respectively [3]. There are many reasons for persistent contamination including failures in reprocessing, acquired and inherent endoscope defects, inappropriate or defective cleaning supplies, and biofilm formation [11]. Persistent endoscope contamination can facilitate endoscopy-associated infections and outbreaks [12].

The risk of infectious transmission was previously estimated to be 1 in 1.8 million endoscopic procedures [13]. This figure now appears to be a significant underestimation for many reasons, including a lack of detailed surveillance for infections following endoscopy, underreporting, and a lack of recognition of acknowledged transmissions [14]. The significance of these transmission events has increased with the emergence of multidrug-resistant organisms and their involvement in endoscopy-associated outbreaks [15, 16].

To prevent endoscopy-associated infection, it is recommended that endoscopy units perform surveillance cultures [5, 17]. In Australia, duodenoscopes, bronchoscopes, linear EUS (endoscopic ultrasound) endoscopes and automatic endoscope/flexible endoscope reprocessor (AERs/AFERs) undergo monthly testing, and all other endoscopes, including gastroscopes and colonoscopes, are tested quarterly [17]. Surveillance cultures are retrospective and act to limit rather than prevent patient exposure. As such, endoscopy-associated infections can still occur despite reprocessing and surveillance cultures. This review aimed to provide an update on published gastrointestinal endoscopy-associated infections and outbreaks, and explore the factors contributing to their occurrence.

Methods

The primary investigator created the search strategy, which was reviewed by each of the authors. PubMed, ScienceDirect and CINAHL were searched for full-text articles published in English within the last 10 years, and an additional filter, “human” was applied to the PubMed search.

Searches were performed with the following keywords: “endoscop*”, “gastroscop*”, “colonoscop*”, “duodenoscop*”, “endoscopic retrograde cholangiopancreatography”, “gastrointestinal endoscop*”, “clean*”, “reprocess*”, “steril*”, “high?level disinfection”, “infect*”, “contaminat*”, “outbreak*”, and “vector*”. Across all three of the databases, a total of 1500 articles were found. Article titles were screened before two reviewers independently screened abstracts of the selected articles. Articles discussing an outbreak or transmission event related to a gastrointestinal endoscope were selected. Full-texts were then downloaded (333) and reviewed.

Articles were excluded if they did not describe an endoscopy-associated infection(s) or outbreak(s) or described an infection transmission event involving a non-gastrointestinal endoscope. Articles were included if they described case(s) of endoscopy-associated infection(s) or outbreak(s), were written in English, and were published within the last 10 years. The search strategy and study selection process are summarised in Fig. 1.

Fig. 1.

Fig. 1

Search strategy and study selection flow chart. This is a flow chart illustrating which databases were searched and the number of articles found from each database. This flow chart also documents the screening process and reasons for the exclusion of articles from the review, as well as the number of articles removed at each screening step

Results

Of the 18 articles reviewed, sixteen described outbreaks associated with duodenoscope use, and two described outbreaks associated with gastroscope and colonoscope use respectively. The causative microorganisms reported were Klebsiella pneumoniae, Pseudomonas aeruginosa, Escherichia coli, and Salmonella enteritidis.

Outbreaks related to reprocessing

Nine articles reported issues with endoscope reprocessing [1826]. Issues with drying were reported in five of these studies. Poor adherence to cleaning guidelines was reported in seven, and one article described reprocessing prior to the introduction of national standards [23].

Naas et al. [20] and Carbonne et al. [18] describe a multi-hospital outbreak of carbapenemase-producing Klebsiella pneumoniae associated with a contaminated duodenoscope [15]. Seventeen patients were exposed to the duodenoscope, ten were screened and six were found to be colonised with K. pneumoniae. A review of the reprocessing method revealed issues with drying. Following this, drying was prioritised and the frequency of duodenoscope microbiological surveillance was increased to a monthly basis.

Aumeran et al. [22] also report on a duodenoscope-associated outbreak with extended-spectrum beta lactamase-(ESBL) producing K. pneumoniae. Similarly, a reprocessing audit revealed endoscopes were not fully dried before storage and were brushed before being flushed with a detergent solution. The endoscope remained culture negative until a Tween-based solution was used. The duodenoscopes were inspected, and reprocessing adherence was mandated. An additional day was allocated for procedures to decrease the number of procedures per day.

Bajolet et al. [21] describe a gastroscope-associated outbreak in four patients with an ESBL-producing P. aeruginosa. A reprocessing audit revealed that the time dedicated to brushing/flushing of endoscope channels was less than 10 min, suction cylinders were not sterilised, a single diameter channel cleaning brush was used for all gastrointestinal endoscopes, and the drying process was inadequate. Following this, compliance with the recommended manual cleaning time, sterilisation of suction cylinders and quarterly microbiological surveillance testing was enforced.

Alrabaa et al. [19] and Sanderson et al. [24] report the infection of seven patients with carbapenemase-producing K. pneumoniae. A reprocessing audit found deviation during duodenoscope elevator channel cleaning, such that debris remained underneath the elevator piece. Culturing of this section revealed E. coli. Of 51 exposed patients, 46 underwent screening and three patients were colonised with carbapenemase-producing organisms.

Reiner et al. [25] describe an outbreak of P. aeruginosa bacteraemia and sepsis in three patients following endoscopic retrograde cholangiopancreatography (ERCP) with the same duodenoscope. Cultures from 6 of the 12 endoscopes, including the implicated endoscope, grew P. aeruginosa. Fluid samples from bottles containing enzymatic solution for precleaning revealed the presence of P. aeruginosa. These bottles were all topped off by a larger bottle, which tested positive for P. aeruginosa and S. maltophila. The outbreak was terminated after the concentration of solution used was standardized, refillable enzymatic bottles were removed and replaced with single case enzymatic packs, and competencies for reprocessing personnel were repeated.

Robertson et al. [23] postulate that a patient colonised with Salmonella enteritidis underwent ERCP and contaminated a loaned duodenoscope which then transmitted the S. enteritidis spp. to three other patients. A reprocessing audit revealed issues including reuse of endoscope cleaning brushes, no dedicated sink for hand hygiene, no commissioning data for the endoscope washer disinfection (EWD), the EWD failed to provide vital information such as channel patency and failed a subsequent load dryness test. Following the outbreak, the EWD was removed from service.

Reddick [26] also reported an outbreak involving S. enteritidis, with three patients following colonoscopy with the same colonoscope. Pulse-gel field electrophoresis patterns for each isolate were identical, but the implicated colonoscope was culture negative. A reprocessing review found that clean and soiled endoscopes were kept in the same room, and endoscopes were kept in the AER when there was no hanging space available for drying.

Outbreaks associated with no reprocessing lapses or endoscope defects

Seven studies reported outbreaks with no issues in reprocessing or endoscope defects [7, 2731]. Other published outbreaks have also occurred despite no apparent lapses in reprocessing [9, 29, 32]. In four articles, implementation of EtO sterilisation limited the outbreak.

Kola et al. [28] report an outbreak of OXA-48 carbapenemase-producing K. pneumoniae (CRKP). The authors hypothesise an infected patient contaminated a duodenoscope, which then infected five other patients. A reprocessing audit revealed no deviations, but cultures performed on another endoscope grew enterococci.

Humphries et al. [30] describe an outbreak of multidrug-resistant Klebsiella pneumoniae bacteraemia and sepsis following ERCP, of which two of nine infected patients died as a result. A reprocessing review found no deviations, and the endoscopes were culture negative, except for 1–2 colonies of coagulase-negative Staphylococcus. In response, implicated scopes were retired, reprocessing was modified and duodenoscopes and linear echoendoscopes underwent EtO gas sterilization off-site. Additional duodenoscopes were purchased, reprocessing staff repeated competencies for both endoscopes and AERs, and reprocessing underwent weekly observation. Following these changes, no further outbreaks or infections were reported.

Qiu et al. [29] report an outbreak involving a duodenoscope contaminated with P. aeruginosa. The authors hypothesise a patient contaminated the duodenoscope, which then infected another two patients. The duodenoscope underwent four rounds of reprocessing but remained culture-positive. The endoscope manufacturers then sterilised the endoscope with epoxyethane, dismantled the channels, and examined them for biofilm, which was absent.

Kovaleva et al. [31] describe an outbreak of multidrug-resistant P. aeruginosa with 3 patients following ERCP. Following the second case, the duodenoscope was cultured and produced P. aeruginosa. Despite multiple rounds of reprocessing, the duodenoscope remained culture-positive. The duodenoscope was then removed from service and underwent EtO gas sterilisation, after which the duodenoscope was culture negative. Four months later, the duodenoscope was culture positive for P. aeruginosa. After inspection, structures suggestive of biofilm were identified in undamaged channels of the scope.

Kim et al. [27] report an outbreak of CRKP with two epidemiologically linked culture-negative duodenoscopes. No breaches were identified in the reprocessing method. One hundred and fifteen patients were exposed, 104 patients underwent screening and 15 were infected with CRKP. Biliary stent placement, cholangiocarcinoma and active inpatient status were each found to increase the risk of developing carbapenemase-resistant Enterobacteriaceae (CRE) infection following ERCP. The risk of clinically relevant CRE infection following ERCP was calculated at 7.7%.

Both Smith et al. [9] and Epstein et al. [7] report an outbreak of New Delhi metallo-beta-lactamase 1-(NDM-1) producing E. coli associated with contaminated duodenoscopes despite no reprocessing lapses. In response, both units implemented EtO sterilisation of duodenoscopes and no further cases were reported. Epstein et al. [7] report that the associated duodenoscopes were culture-negative.

An outbreak associated with endoscope defects

Wendorf et al. [33] describe an outbreak of AmpC E. coli associated with ERCP. An investigation found 32 cases following exclusion of secondary infection transmission and duplicate cases. A reprocessing audit revealed the unit was reprocessing above the industry standard. However, seven of the eight duodenoscopes inspected by the manufacturer had at least one critical defect; three failed a leak test during the manufacturers assessment but passed at the hospital.

An outbreak associated with endoscope design

Verfaillie et al. [32] report on an outbreak related to the design of a specific duodenoscope, the Olympus TJF-180 V, which has since been removed from clinical use. The duodenoscope had a fixed distal cap, and a sealed elevator wire channel port; sealed with an O-ring, which may not have sealed the forceps elevator axis sufficiently. Thirty patients were infected with a VIM-2-positive P. aeruginosa strain, 22 of which were infected by the duodenoscope. Following removal of the model from clinical use, the number of VIM-2-positive P. aeruginosa cases decreased.

Scanning electron microscopy (SEM) of the distal cap and its components revealed a rough surface on the O-ring, as well as sludge behind the glass that covers the light-guide lens, a crack in the fixed cap, a brown layer on the O-ring and brown staining of the frame of the distal tip. There were no breaches in adherence to the reprocessing guidelines.

Discussion

Eighteen outbreaks associated with gastrointestinal endoscopy were reviewed. The majority were associated with duodenoscopes, the distal tip of which could not be dismantled to facilitate effective cleaning [32] and contains an intricate contamination-prone elevator mechanism [34]. Recently, duodenoscopes with detachable distal caps have been developed to address this issue [35].

Outbreaks were also associated with lapses in endoscope reprocessing. Manual cleaning, which is crucial to preventing biofilm formation [36], was identified as inadequate [19, 2123] in four of the reviewed articles. In contrast, seven outbreaks occurred with no apparent reprocessing lapses, which could be explained by biofilm contamination [29]. It has been shown that biofilm can persist within an endoscope despite reprocessing [37], and this may contribute to persistent contamination. In one article, structures suggestive of biofilm were found inside the channels of a persistently contaminated duodenoscope [31]. Kovaleva specifically identified P. aeruginosa as a biofilm-producer [38], which is interesting as it was the culprit organism in three outbreaks.

The drying process is also crucial to effective endoscope reprocessing [17, 39]. Recently, the use of drying cabinets with pressurised air flow through endoscope channels has been shown to reduce microbiological load [40, 41], and has been implemented in endoscopy units [5] to ensure thorough drying.

Endoscope defects are also a potential explanation, as identified by Wendorf et al. [33]. An article by Ofstead et al. used borescopes to identify damage within endoscope channels, and found previously undetected damage and debris [42]. This damage may facilitate persistent contamination despite adherent reprocessing, and potential biofilm formation.

Furthermore, endoscope reprocessing can have a variable failure rate [3]. A study following the outbreak reported by Wendorf et al. found that despite stringent reprocessing, there was an inherent high-level disinfection failure rate of 2% [43]. Similar rates were identified by Bisset et al. for gastroscopes and colonoscopes (1.8% and 1.9% respectively) [3], and by Higa et al. for endoscopes (1.9%) [44].

The exclusion of outbreaks that were not published in a peer-reviewed journal, and the low number of published outbreaks, limit this review. Most endoscope reprocessing lapses are never reported [11] or associated infections are not recognised [14], and if an outbreak has not been contained, it may not be reported [15]. Gastmeier and Vonberg suggest in their own review of endoscopy-associated infections with Klebsiella spp. that it is very likely outbreaks may not be acknowledged because they involve commensal bacteria of the gastrointestinal tract [45]. The review is also limited by the quality of evidence; endoscopy-associated infections are usually reported as case studies or case series.

Conclusion

In conclusion, methods of improving endoscope reprocessing and screening for endoscope contamination, such as the use of adenosine triphosphate (ATP) measurement [46], may be vital to preventing future endoscopy-associated infections and outbreaks. Furthermore, endoscope reprocessing may be more effective if it is regularly reviewed to ensure adherence, and routine maintenance and inspection is crucial to preventing infection transmission.

Authors’ contributions

All authors contributed towards the study concept, article review and revision of the paper, and agreed to be accountable for all aspects of the work. All authors read and approved the final manuscript.

Acknowledgements

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Availability of data and materials

Not applicable.

Consent for publication

Not applicable.

Ethics approval and consent to participate

Not applicable.

Funding

Not applicable.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Abbreviations

GENCA/GESA

Gastroenterological Nurses College of Australia and Gastroenterological Society of Australia

EtO

ethylene oxide

EUS

(endoscopic ultrasound) endoscope

AER/AFER

automatic endoscope/flexible endoscope reprocessor

CINAHL

the Cumulative Index to Nursing and Allied Health Literature

ESBL

extended-spectrum beta-lactamase

ERCP

endoscopic retrograde cholangiopancreatography

EWD

endoscope washer disinfector

OXA-48

oxacillinase-48

CRE

carbapenemase-resistant enterobacteriaceae

NDM 1

new delhi metallo-beta-lactamase 1

VIM-2

verona integron-encoded metallo-beta-lactamase 2

ATP

adenosine triphosphate

Contributor Information

Charles Eugenio McCafferty, Phone: (+614) 99 777 181, Email: c.mccafferty@westernsydney.edu.au.

Marra Jai Aghajani, Email: marra.aghajani1@gmail.com.

David Abi-Hanna, Email: dr.david@abi-hanna.com.

Iain Bruce Gosbell, Email: i.gosbell@westernsydney.edu.au.

Slade Owen Jensen, Email: s.jensen@westernsydney.edu.au.

References

  • 1.Nelson DB, Muscarella LF. Current issues in endoscope reprocessing and infection control during gastrointestinal endoscopy. World J Gastroenterol. 2006;12(25):3953–3964. doi: 10.3748/wjg.v12.i25.3953. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Ribeiro MM, de Oliveira AC. Analysis of the air/water channels of gastrointestinal endoscopies as a risk factor for the transmission of microorganisms among patients. Am J Infect Control. 2012;40(10):913–916. doi: 10.1016/j.ajic.2012.02.005. [DOI] [PubMed] [Google Scholar]
  • 3.Bisset L, Cossart YE, Selby W, West R, Catterson D, O’Hara K, et al. A prospective study of the efficacy of routine decontamination for gastrointestinal endoscopes and the risk factors for failure. Am J Infect Control. 2006;34(5):274–280. doi: 10.1016/j.ajic.2005.08.007. [DOI] [PubMed] [Google Scholar]
  • 4.Committee AQAIE. Petersen BT, Chennat J, Cohen J, Cotton PB, Greenwald DA, et al. Multisociety guideline on reprocessing flexible gastrointestinal endoscopes: 2011. Gastrointestinal endoscopy. 2011;73(6):1075–1084. doi: 10.1016/j.gie.2011.03.1183. [DOI] [PubMed] [Google Scholar]
  • 5.Gastroenterological Society of Australia (GESA). Infection control in endoscopy consenus statements. In: Australian gastroenterology week (AGW) 2017, Queensland, Australia; 2017.
  • 6.Choi HH, Cho YS. Endoscope reprocessing: update on controversial issues. Clin Endosc. 2015;48(5):356–360. doi: 10.5946/ce.2015.48.5.356. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Epstein L, Hunter JC, Arwady MA, Tsai V, Stein L, Gribogiannis M, et al. New Delhi metallo-beta-lactamase-producing carbapenem-resistant Escherichia coli associated with exposure to duodenoscopes. JAMA. 2014;312(14):1447–1455. doi: 10.1001/jama.2014.12720. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Naryzhny I, Silas D, Chi K. Impact of ethylene oxide gas sterilization of duodenoscopes after a carbapenem-resistant Enterobacteriaceae outbreak. Gastrointest Endosc. 2016;84(2):259–262. doi: 10.1016/j.gie.2016.01.055. [DOI] [PubMed] [Google Scholar]
  • 9.Smith ZL, Oh YS, Saeian K, Edmiston CE, Jr, Khan AH, Massey BT, et al. Transmission of carbapenem-resistant Enterobacteriaceae during ERCP: time to revisit the current reprocessing guidelines. Gastrointest Endosc. 2015;81(4):1041–1045. doi: 10.1016/j.gie.2014.11.006. [DOI] [PubMed] [Google Scholar]
  • 10.Rutala WA, Weber DJ. Disinfection and sterilization: an overview. Am J Infect Control. 2013;41(5 Suppl):S2–S5. doi: 10.1016/j.ajic.2012.11.005. [DOI] [PubMed] [Google Scholar]
  • 11.Dirlam Langlay AM, Ofstead CL, Mueller NJ, Tosh PK, Baron TH, Wetzler HP. Reported gastrointestinal endoscope reprocessing lapses: the tip of the iceberg. Am J Infect Control. 2013;41(12):1188–1194. doi: 10.1016/j.ajic.2013.04.022. [DOI] [PubMed] [Google Scholar]
  • 12.Herve RC. Endoscopy in the twenty-first century: minimally invasive state-of-the-art medical technology or a future main vector of hospital-acquired infections? J Hosp Infect. 2017;97(4):329–330. doi: 10.1016/j.jhin.2017.07.030. [DOI] [PubMed] [Google Scholar]
  • 13.Kimmery MBBD, Carr-Locke DL, DiMarino AJ, Jensen DM, Katon R, et al. ASGE technology assessment position paper: transmission of infection by gastrointestinal endoscopy. Gastrointest Endosc. 1993;36:885–888. doi: 10.1016/S0016-5107(93)70316-8. [DOI] [Google Scholar]
  • 14.Ofstead CL, Dirlam Langlay AM, Mueller NJ, Tosh PK, Wetzler HP. Re-evaluating endoscopy-associated infection risk estimates and their implications. Am J Infect Control. 2013;41(8):734–736. doi: 10.1016/j.ajic.2012.10.008. [DOI] [PubMed] [Google Scholar]
  • 15.O’Horo JC, Farrell A, Sohail MR, Safdar N. Carbapenem-resistant Enterobacteriaceae and endoscopy: an evolving threat. Am J Infect Control. 2016;44(9):1032–1036. doi: 10.1016/j.ajic.2016.03.029. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Rubin ZA, Murthy RK. Outbreaks associated with duodenoscopes: new challenges and controversies. Curr Opin Infect Dis. 2016;29(4):407–414. doi: 10.1097/QCO.0000000000000290. [DOI] [PubMed] [Google Scholar]
  • 17.Gastroenterological Society of Australia (GESA), Gastroenterological Nurses College of Australia (GENCA). Infection control in endoscopy, 3rd ed. Mulgrave; 2010.
  • 18.Carbonne A, Thiolet JM, Fournier S, Fortineau N, Kassis-Chikhani N, Boytchev I, et al. Control of a multi-hospital outbreak of KPC-producing Klebsiella pneumoniae type 2 in France, September to October 2009. Euro surveill Bull Europeen sur les maladies transmissibles = European communicable disease bulletin. 2010;15(48):19734. doi: 10.2807/ese.15.48.19734-en. [DOI] [PubMed] [Google Scholar]
  • 19.Alrabaa SF, Nguyen P, Sanderson R, Baluch A, Sandin RL, Kelker D, et al. Early identification and control of carbapenemase-producing Klebsiella pneumoniae, originating from contaminated endoscopic equipment. Am J Infect Control. 2013;41(6):562–564. doi: 10.1016/j.ajic.2012.07.008. [DOI] [PubMed] [Google Scholar]
  • 20.Naas T, Cuzon G, Babics A, Fortineau N, Boytchev I, Gayral F, et al. Endoscopy-associated transmission of carbapenem-resistant Klebsiella pneumoniae producing KPC-2 beta-lactamase. J Antimicrob Chemother. 2010;65(6):1305–1306. doi: 10.1093/jac/dkq117. [DOI] [PubMed] [Google Scholar]
  • 21.Bajolet O, Ciocan D, Vallet C, de Champs C, Vernet-Garnier V, Guillard T, et al. Gastroscopy-associated transmission of extended-spectrum beta-lactamase-producing Pseudomonas aeruginosa. J Hosp Infect. 2013;83(4):341–343. doi: 10.1016/j.jhin.2012.10.016. [DOI] [PubMed] [Google Scholar]
  • 22.Aumeran C, Poincloux L, Souweine B, Robin F, Laurichesse H, Baud O, et al. Multidrug-resistant Klebsiella pneumoniae outbreak after endoscopic retrograde cholangiopancreatography. Endoscopy. 2010;42(11):895–899. doi: 10.1055/s-0030-1255647. [DOI] [PubMed] [Google Scholar]
  • 23.Robertson P, Smith A, Anderson M, Stewart J, Hamilton K, McNamee S, et al. Transmission of Salmonella enteritidis after endoscopic retrograde cholangiopancreatography because of inadequate endoscope decontamination. Am J Infect Control. 2017;45(4):440–442. doi: 10.1016/j.ajic.2016.11.024. [DOI] [PubMed] [Google Scholar]
  • 24.Sanderson R, Braithwaite L, Ball L, Ragan P, Eisenstein L. An outbreak of carbapenem-resistant Klebsiella pneumoniae infections associated with endoscopic retrograde cholangiopancreatography (ERCP) procedures at a hospital. Am J Infect Control. 2010;38(5):e141. doi: 10.1016/j.ajic.2010.04.191. [DOI] [Google Scholar]
  • 25.Reiner S. Investigation of a cluster of genomically identical Pseudomonas aeruginosa blood isolates following endoscopic retrograde cholangiopancreatography in a gastroenterology laboratory. Am J Infect Control. 2008;36(5):E198. [Google Scholar]
  • 26.Reddick E. Investigation of salmonellosis outbreak following a hospital endoscopy: a public health case study. Can J Infect Control. 2017;32(3):156–159. [Google Scholar]
  • 27.Kim S, Russell D, Mohamadnejad M, Makker J, Sedarat A, Watson RR, et al. Risk factors associated with the transmission of carbapenem-resistant Enterobacteriaceae via contaminated duodenoscopes. Gastrointest Endosc. 2016;83(6):1121–1129. doi: 10.1016/j.gie.2016.03.790. [DOI] [PubMed] [Google Scholar]
  • 28.Kola A, Piening B, Pape UF, Veltzke-Schlieker W, Kaase M, Geffers C, et al. An outbreak of carbapenem-resistant OXA-48—producing Klebsiella pneumonia associated to duodenoscopy. Antimicrob Resist Infect Control. 2015;4:8. doi: 10.1186/s13756-015-0049-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Qiu L, Zhou Z, Liu Q, Ni Y, Zhao F, Cheng H. Investigating the failure of repeated standard cleaning and disinfection of a Pseudomonas aeruginosa-infected pancreatic and biliary endoscope. Am J Infect Control. 2015;43(8):e43–e46. doi: 10.1016/j.ajic.2015.04.204. [DOI] [PubMed] [Google Scholar]
  • 30.Humphries RM, Yang S, Kim S, Muthusamy VR, Russell D, Trout AM, et al. Duodenoscope-related outbreak of a carbapenem-resistant Klebsiella pneumoniae identified using advanced molecular diagnostics. Clin Infect Dis. 2017;65(7):1159–1166. doi: 10.1093/cid/cix527. [DOI] [PubMed] [Google Scholar]
  • 31.Kovaleva J, Meessen NE, Peters FT, Been MH, Arends JP, Borgers RP, et al. Is bacteriologic surveillance in endoscope reprocessing stringent enough? Endoscopy. 2009;41(10):913–916. doi: 10.1055/s-0029-1215086. [DOI] [PubMed] [Google Scholar]
  • 32.Verfaillie CJ, Bruno MJ, Voor in ‘t Holt AF, Buijs JG, Poley JW, Loeve AJ, et al. Withdrawal of a novel-design duodenoscope ends outbreak of a VIM-2-producing Pseudomonas aeruginosa. Endoscopy. 2015;47(6):493–502. doi: 10.1055/s-0034-1391886. [DOI] [PubMed] [Google Scholar]
  • 33.Wendorf KA, Kay M, Baliga C, Weissman SJ, Gluck M, Verma P, et al. Endoscopic retrograde cholangiopancreatography-associated AmpC Escherichia coli outbreak. Infect Control Hosp Epidemiol. 2015;36(6):634–642. doi: 10.1017/ice.2015.66. [DOI] [PubMed] [Google Scholar]
  • 34.Humphries RM, McDonnell G. Superbugs on duodenoscopes: the challenge of cleaning and disinfection of reusable devices. J Clin Microbiol. 2015;53(10):3118–3125. doi: 10.1128/JCM.01394-15. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Medical P. Video Duodenoscopes ED34-i10T. Japan: PENTAX Medical; 2018. https://www.pentaxmedical.com/pentax/en/100/1/Video-Duodenoscopes-ED34-i10T. Accessed Mar 2018.
  • 36.Asge Standards Of Practice C. Banerjee S, Shen B, Nelson DB, Lichtenstein DR, Baron TH, et al. Infection control during GI endoscopy. Gastrointest Endosc. 2008;67(6):781–790. doi: 10.1016/j.gie.2008.01.027. [DOI] [PubMed] [Google Scholar]
  • 37.Ren-Pei W, Hui-Jun X, Ke Q, Dong W, Xing N, Zhao-Shen L. Correlation between the growth of bacterial biofilm in flexible endoscopes and endoscope reprocessing methods. Am J Infect Control. 2014;42(11):1203–1206. doi: 10.1016/j.ajic.2014.07.029. [DOI] [PubMed] [Google Scholar]
  • 38.Kovaleva J. Infectious complications in gastrointestinal endoscopy and their prevention. Best Pract Res Clin Gastroenterol. 2016;30(5):689–704. doi: 10.1016/j.bpg.2016.09.008. [DOI] [PubMed] [Google Scholar]
  • 39.Alfa MJ. Current issues result in a paradigm shift in reprocessing medical and surgical instruments. Am J Infect Control. 2016;44(5 Suppl):e41–e45. doi: 10.1016/j.ajic.2016.01.020. [DOI] [PubMed] [Google Scholar]
  • 40.Pineau L, Villard E, Duc DL, Marchetti B. Endoscope drying/storage cabinet: interest and efficacy. J Hosp Infect. 2008;68(1):59–65. doi: 10.1016/j.jhin.2007.10.012. [DOI] [PubMed] [Google Scholar]
  • 41.Grandval P, Hautefeuille G, Marchetti B, Pineau L, Laugier R. Evaluation of a storage cabinet for heat-sensitive endoscopes in a clinical setting. J Hosp Infect. 2013;84(1):71–76. doi: 10.1016/j.jhin.2013.01.013. [DOI] [PubMed] [Google Scholar]
  • 42.Ofstead CL, Wetzler HP, Eiland JE, Heymann OL, Held SB, Shaw MJ. Assessing residual contamination and damage inside flexible endoscopes over time. Am J Infect Control. 2016;44(12):1675–1677. doi: 10.1016/j.ajic.2016.06.029. [DOI] [PubMed] [Google Scholar]
  • 43.Ross AS, Baliga C, Verma P, Duchin J, Gluck M. A quarantine process for the resolution of duodenoscope-associated transmission of multidrug-resistant Escherichia coli. Gastrointest Endosc. 2015;82(3):477–483. doi: 10.1016/j.gie.2015.04.036. [DOI] [PubMed] [Google Scholar]
  • 44.Higa JT, Gluck M, Ross AS. Duodenoscope-associated bacterial infections: a review and update. Curr Treat Options Gastroenterol. 2016;14(2):185–193. doi: 10.1007/s11938-016-0088-9. [DOI] [PubMed] [Google Scholar]
  • 45.Gastmeier P, Vonberg RP. Klebsiella spp. in endoscopy-associated infections: we may only be seeing the tip of the iceberg. Infection. 2014;42(1):15–21. doi: 10.1007/s15010-013-0544-6. [DOI] [PubMed] [Google Scholar]
  • 46.McCafferty CE, Abi-Hanna D, Aghajani MJ, Micali GT, Lockart I, Vickery K, et al. The validity of adenosine triphosphate (ATP) measurement in detecting endoscope contamination. J Hosp Infect. 2018 doi: 10.1016/j.jhin.2018.08.004. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Not applicable.


Articles from Annals of Clinical Microbiology and Antimicrobials are provided here courtesy of BMC

RESOURCES