Abstract
As many studies show, there is a relation between the tissue’s mechanical characteristics and some specific diseases. Knowing this relationship would help early diagnosis or microsurgery. In this paper, a new method for measuring the viscoelastic properties of soft materials at the microscale is proposed. This approach is based on the adoption of a microsystem whose mechanical structure can be reduced to a compliant four bar linkage where the connecting rod is substituted by the tissue sample. A procedure to identify both stiffness and damping coefficients of the tissue is then applied to the developed hardware. Particularly, stiffness is calculated solving the static equations of the mechanism in a desired configuration, while the damping coefficient is inferred from the dynamic equations, which are written under the hypothesis that the sample tissue is excited by a variable compression force characterized by a suitable wave form. The whole procedure is implemented by making use of a control system.
Keywords: MEMS, tweezers, control, viscous coefficient, soft tissue
1. Introduction
The manipulation at the micro or nanoscale nowadays mainly regards the handling of small biological tissue samples or cells for a large variety of applications where their specific mechanical characteristics should be known before any possible treatment.
A specialized knowledge about the relations between the tissue sample mechanical characteristics and some specific diseases would help much in diagnosis or microsurgery, but unfortunately it is not easy to develop new experimental tests because of the lack of specialized instrumentation or of its high costs at micro or nanosize. The most typical mechanical characteristic is stiffness, although much information is expected to be contained in the viscoelastic response to a given exciting stress.
Other than simulations (for example with the aid of Finite Element Analysis) and theoretical work (Continuum Mechanics), many different experimental activities have been attempted in the recent past, such as, for example, stretch, dynamic, aspiration or indentation tests.
Mechanical stretch tests represent the most immediate way to obtain information on a biological material. Since 2000, nonlinear and viscoelastic material properties of brain tissues have been analyzed [1] and their mechanical response has been recognized to be related to strain history. In another investigation [2], mechanical properties of human skin have been characterized through uniaxial tensile testing and histomorphometric analysis. Tension tests showed that pressure ulcer tissues had significantly lower strains (at peak stress) than that characteristic of control healthy breast tissues. The mechanical properties of reconstituted tissues have been also quantitatively characterized by means of uniaxial stretch measurements [3], and the cell density and matrix organization has been studied by analyzing the strain-dependent stress and the elastic modulus of the active component. Multiaxial testing techniques showed [4] how soft biological tissues exhibit complex mechanical behaviors not easily accounted for in classic elastomeric constitutive models. In 2006, quasilinear viscoelasticity and hysteresis in porcine esophagus tissues have been investigated by means of stress-relaxation tests at various strains and Fung’s quasilinear viscoelastic model has been assessed [5]. Incremental relaxation stress tests have been also adopted to identify surface-regional mechanical variations in the porcine temporomandibular joint disc under unconfined compression [6] and a value of the coefficient of viscosity turned out to be approximately equal to 3.5 MPa-s. More recently, some efforts have been paid to characterize tissue-engineered blood vessels by analyzing their mechanical properties by means of compliance measurement, burst pressure testing, creep and cyclic testing. These investigations are helpful to develop efficient fabrication methods and to optimize tissue resistance and viscoelastic properties. For example, the mechanical properties of tissue-engineered vascular constructs have been studied by monitoring pressure and diameter variations of vascular constructs submitted to hydrostatic loading [7]. A piezoelectric device, that measures the properties of the biological tissues, has been also recently developed [8]. The system uses the properties of the piezoelectric materials [9,10,11] to measure the viscoelasticity coefficients of the tissue in the near-surface regions of the epidermis.
Dynamic and vibration based methods offer other means for characterizing materials. For example, in 2004 some dynamic tests have been performed [12] to investigate frequency-dependent mechanical properties of tissue, specifically for properties in compressive, tensile, and flexural modes over a frequency range from 0.1 to 400 Hz. Ultrasound vibrometry has provided noninvasive tools to estimate the shear moduli of tissues from the propagation velocities at multiple frequencies. For example, Voigt, Maxwell and Zener classic models have been applied to estimate shear moduli of porcine livers with different thermal damage levels and different storage times [13].
Some tests have been performed by taking advantage from robotics and control. For example, a haptics-enabled master-slave test-bed has been used [14] to understand how various modalities for feedback of interaction between a surgical tool and a soft tissue can improve the efficiency of a typical surgical task. Furthermore, a multi–axial force sensor mounted on a robotic manipulator, together with a stereo camera system, has been adopted to estimate mechanical parameters of soft tissue from sensory data collected during robotic surgical manipulation [15].
Nanoindentation has been also adopted to characterize nano and microscale mechanical properties in biomaterials [16], although this technique is usually applied to hard tissues such as bone and teeth. In 2008, considering that heart valve tissues show nonlinear, anisotropic and inhomogeneous mechanical behavior, some indentation tests have been performed [17] on linear elastic polydimethylsiloxane rubbers and compared to tensile tests on the same specimen for local characterization of nonlinear soft anisotropic tissue properties. The mechanical relaxation behavior of injured rat spinal cord tissue, at the site of mid-thoracic spinal hemisection, has been monitored post-injury using a microindentation test method in order to calculate the elastic moduli and to estimate the relaxation time constant and viscosity. This activity showed that injured tissues exhibited significantly lower stiffness and elastic modulus in comparison to uninjured control tissue [18]. Results could be used to understand some aspects of biochemical nerve regeneration. Elastic and viscoelastic properties of undifferentiated adipose-derived stem cells have been tested via atomic force microscopy and correlated with lineage-specific metabolite production [19]. Mechanical characterization of soft tissues with atomic force microscopy can be affected by tissue samples viscous effects in case of sharp tips. Hence, different responses to nanoindentation of soft biotissues with different elasticity and viscous properties have been investigated. For example, similarities with indentation of elasto-plastic materials have been ascertained and also a linear relationship between the limit indentation rate and the geometry of the AFM probe has been obtained [20].
In case the probed sample is non-adherent to the rigid substrate, the elastic modulus of cells and biological samples can be evaluated experimentally by means of micropipette aspiration. In 2007, intra-operative aspiration in vivo experiments assessed the mechanical response of human liver [21]. This mechanical characterization was useful to histological evaluation of tissue biopsies to make correlations between mechanical response and tissue micro-structure of normal and diseased liver and various pathologic conditions affecting the tissue samples were quantified. For example, a direct proportionality between stiffness and connective tissue percentage was ascertained. Micropipette aspiration technique has been applied to measure selectively the mechanical properties of the top layer of an intact multilayer biomaterial [22].
Mathematical modeling of biological tissues has been reviewed [23] focusing particularly on the stress-strain responses of biological tissues subjected to mechanical loading. Continuum Mechanics has been also applied to interpret deformation profile of aspirated soft tissues and so to evaluate their mechanical properties [24,25]. Micropipette aspiration has been studied by using finite element analysis simulation of the sample aspiration into the micropipette [26] and by considering also the influence of the sample adhesion. An inverse finite-element method strategy has been used in 2007 [27] to evaluate the elastic modulus and yield properties of human cortical specimens of the radial diaphysis and some numerical values have been proposed and assessed by means of isotropic linear-elastic simulations. The mechanical response of soft biological materials to nanoindentation with spherical indenters has been investigated by using finite element simulations together with three different material constitutive laws, namely, elastic, isotropic hyperelastic and anisotropic hyperelastic [28], in order to improve measurements offered by nanoindentation experiments. A computational tool has been developed [29] to describe and predict the mechanical behavior of electrospun scaffolds characterized by different microstructures. This has been useful to study the global mechanical properties of cardiovascular valve-shaped scaffolds and to understand that anisotropy is necessary to reproduce the deformation patterns observed in native heart valves.
Considering that cells are the fundamental building blocks of tissue, a more detailed comprehension of tissue mechanics can be achieved by studying the Mechanics of cells. Any modification of a cell mechanical characteristic is related to a change in its functionality, possible pathological diseases, or response to the environment. A review on cell mechanics [30] pointed out several relations between cell functionality and its mechanical response and presented theoretical models together with experimental and computational approaches. For example, Ekpenyong et al. [31] showed that some cells alter their viscoelastic properties during stem cell differentiation. The modification has been measured using an optical stretcher. Gossett et al. [32] illustrated an automated microfluidic technology capable of probing single–cell deformability at approximately 2000 cells/s. A microfluidic chip has been also proposed [33] to apply mechanical stress to live cells and measure their nuclear deformability, with the purpose of characterizing the nuclear response to physical stress. Nucleus physical plasticity has been also studied in stem cell differentiation [34]. The mechanical behavior has been studied also to characterize Embryonic stem cells (ESCs) [35]. Furthermore, optical deformability has been used to monitor the subtle changes during the progression of mouse fibroblasts and human breast epithelial cells from normal to cancerous state [36]. Finally, real-time deformability cytometry (RT-DC) has been introduced [37] for continuous cell mechanical characterization of large populations (> cells) with analysis rates greater than 100 cells/s.
Cell-specific mechanical properties have a great influence on cell stretching response. By using this concept, a method to distinguish cell types has been developed, based on stretching tests performed by positioning cells between microelectrodes and applying dielectrophoretic forces [38].
Cellular response to mechanical stimuli has been investigated in [39] where an electromagnetic cell stretching platform, based on double-sided stretching axes, has been used to introduce a cyclic and static strain pattern on a cell culture.
In the present investigation the mechanical handling of a tissue sample has been suggested as a method to test the tissue stiffness and viscous damping response. According to the suggested procedure, the handling operation is carried out by means of a microgripper that has been independently fabricated for this dedicated purpose and makes use of a new concept so–called CSFH hinge [40,41,42].
A new mechanical model for determining the tissue stiffness and viscosity has been developed, based on Kinematics and Continuum Mechanics. Furthermore, a control strategy has been proposed and simulated. As shown in literature [43,44,45], grippers and microgrippers have been extensively developed during the last decades and, more recently, new synthesis techniques [46,47,48,49] allowed designers to handle tissue also by selective, non isotropic compliance [50]. A former version of microgripper for tissue manipulation has been presented in 2015 [51], while this paper presents a new refined version of the hardware and a new control system for measuring both the stiffness and the viscosity coefficient of the tissue sample.
2. The Adopted Microsystem
A new microsystem has been specifically designed and fabricated for the purpose of this investigation. The microsystem has been also provided with a dedicated control algorithm for testing the mechanical characteristics of any biological tissue sample at the micro or nanoscale. Therefore, the control method is intended as applied to the microsystem depicted in Figure 1a,b. The actual fabrication method made use of Deep Reactive-Ion Etching applied on Silicon on Insulator wafer. The fabrication process is described in detail in [52].
Mobility is provided to the microsystem by two CSFHs, Conjugate Surface Flexure Hinges, that will be briefly introduced in the next section.
3. The Modeling Approach
As mentioned in the Introduction, many investigations presented in the literature were focused on the development of models to understand cell mechanics, such as cortical shell–liquid core models and solid models [53]. Generally, the mechanical model depends on the experimental approach, that can be based on force application techniques or force sensing techniques [30]. Also in this investigation, the mechanical model of the tissue depends on experimental techniques, as described hereafter.
A fundamental component of the developed microsystem can be represented as in Figure 2, where the mobile jaws show up in a possible configuration. A controlled-desired displacement of the right hand side jaw can be obtained by actuating the left hand side jaw, through the elastic response of a grabbed tissue sample. By virtue of the adopted CSFH Conjugate Surface Flexure Hinge it could be shown that points and are quite stationary during actuation and therefore they can be considered the rotation centers of the two jaws [54,55]. Considering also that the two jaws are rather stiffer than the tissue samples, the four-bar linkage can be adopted to model the whole microsystem, provided that the coupler link is allowed to have a variable length.
More specifically, the tissue sample is modeled by using the continuum mechanics theory, as a homogeneous and isotropic material. According to the multi-scale, three-level hierarchical approach for mechanotransduction proposed by Lim et al. [53], the model proposed in this Section can be classified at Level 1. Therefore, the tissue sample is considered as a beam element with constant cross-section (with area ) subjected to the axial load . With reference to Figure 3, the normal strain is defined as , whereas the normal stress is equal to . Assuming a linear elastic behaviour for the material, according to the Hooke’s law the stress-strain relationship can be written as , where is the Young modulus. By using the definitions of normal stress and normal strain, the previous equation can be written as
(1) |
and the stiffness coefficient of the tissue sample can be obtained as
(2) |
It is worth noticing also that the sample coefficient of viscosity is not coincident to the material viscosity , because depends on the geometry of the sample. In general is linearly proportional to , as, for example, for the simple cylindrical viscous damper [56]. Therefore, the actual value of must be related to the tissue material viscosity through a model, which involves also the sample geometry. In this contribution the sample has been modeled as a linear spring working in parallel with a viscous damper. The actual system configuration will be described by means of initial, target and incremental variables according to the following Nomenclature, which makes reference to Figure 4.
hat refers to a constant parameter, such as, for example, those referring to the initial configuration, as shown in Figure 5a;
tilde˜ refers to an actual parameter at the generic configuration, as represented in Figure 5b;
superscript refers to the desired or target parameter;
angles of bars are measured counterclockwise, starting from the positive abscissa;
is the length of vector which is split in the initial length and the deformation ;
, and refer to the variations of the angular positions of the link vectors , and , with respect to their initial position; in this way their actual absolute angular positions will be , , , respectively;
(as Figure 5a shows that they are supplementary angles)
, from geometry represented in Figure 5a;
is the common length of the two vectors and ;
is the length of the frame link ;
is the stiffness coefficient of the tissue sample;
and are the two jaws torsional stiffness, which are related to the CSFH curved beam material and geometry;
, , and are the radius, width, thickness and beam subtended angle of the CSFH flexure curved beam;
, and represent the viscous damping coefficients of the sample and of the two jaws;
and represents the two jaws moments of inertia around and , with ;
and are the tensions applied to the comb drives;
, and are the overlap angle, gap and width of the comb drive fingers;
device-handle gap (silicon oxide layer thickness);
air viscosity at 25 °C;
polar moment of area exposed to air viscous damping, calculated around the rotation points;
equivalent radius employed to model the air viscous damping.
Parameter values given in input to the developed model are shown in Table 1.
Table 1.
Parameter | Value | Parameter | Value |
---|---|---|---|
1.44 rad | m | ||
1.70 rad | m | ||
4.20 rad | m | ||
rad | , | m | |
m | , | kg | |
m | , | kgm | |
m | kg/ms | ||
m | , | kgm/rad | |
2 m | , | kgm/srad |
Air viscous damping has been used to model friction on the mobile jaws. Slide-film damping with continuum flow regime has been adjusted for the case of rotary and alternate motion.
In case of rotary motion and assuming one order of magnitude smaller than the comb drive maximum (external) radius , the viscous tangential traction on the elementary surface at the distance from the rotation axis, exerts an elementary moment . Therefore overall moment is approximately equal to . For this reason, the viscous damping coefficients and can be estimated as
(3) |
for either jaw or , where the polar moments of area and have been calculated around the rotation points and by taking into account the exact geometry of the rotating masses. The effective viscosity has been calculated as
(4) |
where is the Knudsen number, equal to the ratio of the mean free path of the molecules in air, , to the critical dimension of the device, . In our case the length is equal to m because the system works at Pa, whereas is equal to m, namely, the gap between the conjugate surfaces. Since , the continuum flow model can be accepted as the actual regime with .
4. The Adopted Electromechanical Model
Since the coupler has variable length, in this particular case the four-bar linkage has two DoF (Degrees of Freedom). Considering that two rotary comb drives give us control over and , these two variables can be assumed to be the independent ones. However, for the sake of the present investigation, the interest is in and . The former can be calculated by extracting the real and imaginary components of vector and then evaluating its module and inclination angle .
In fact, given , , and , the closed loop complex equation
(5) |
allows to be expressed as
(6) |
and therefore to obtain immediately its real
(7) |
and imaginary
(8) |
components. From these, the values of and can be computed, for all configuration during contact, as functions of and according to the expressions
(9) |
and
(10) |
Reminding that and multiplying the first time derivative of Equation (5)
(11) |
by , the complex equation
(12) |
is obtained, where the real components can be arranged to express the deformation velocity of the sample tissue
(13) |
which does not depend directly on .
Assuming that a jaw and its corresponding comb teeth mobile set can be approximated as a whole pseudo–rigid system, Euler’s rotation equation can be written twice, to express the dynamic balance of the left and right jaw systems during rotations, namely,
(14) |
(15) |
where the sample inertia has been assumed to be negligible.
With reference to Figure 6, by introducing the comb drives equivalent radius
(16) |
where and (see also ref. [57,58])
the torque and the capacitance at the two comb drives 2 and 4 can be expressed, respectively, as
where minus or plus holds for or , respectively.
5. The Identification of the Sample Stiffness and the Damping Coefficients
The technique proposed for measuring the damping coefficient of a sample tissue consists in exciting it with a variable compressive force which is characterized by a suitable wave form. Then, after a first phase during which the tissue is grabbed and kept firmly between the jaws, a torque is applied and the system response is measured.
5.1. Characterization of the Sample Stiffness
More precisely, if a constant torque is applied, while , with large enough to grab a sample between the jaws, keeping it firmly but without damaging it, a small variation of is obtained; the amplitude of such variation is directly related to the gripping force applied on the sample.
This effect can be obtained with a manual operation, in which the input voltage is slowly increased, in a quasi-static condition, until the desired angle is reached, or under a control system, which brings the angle to the desired value . In this paper, we will design the control scheme accordingly to the second option, as represented in Figure 7, where block represents a Regulator. In the case under study, consists of a PID (Proportional-Integral-Derivative standard regulator), and represents the control block which generates the input torque on the basis of the position error, between the desired angle and the actual one , until it goes, asymptotically, to zero.
Figure 8 depicts the steady state values of the output measured angle as a function of the unknown parameters and (In the following simulations the parameter values indicated in Table 1 have been used. The plane dimensions of the specimen are approximately m m), confirming the physical observation that it has to be dependent on the elastic coefficient only. Similar relationships can be referred to the input and the sample angles, as reported in Figure 9 and Figure 10, respectively.
In any case, at steady state, the Equations (14) and (15) assume the expressions
(17) |
(18) |
where , and . Note that, in these expressions, both and can be computed from (9) and (10), once the values of and are measured.
5.2. Characterization of Sample Characteristic Damping
Let us choose, now, in (14), , with a suitable time function described in the following; in this case, Equation (14), using Equation (17), becomes
(19) |
Some computations can be performed to put in evidence the dependency of the steady state output under a particular choice of input . Initially, the case of small and slow signals is addressed, bringing to a linearly approximated analysis useful to show, at a first approximation, the relationships between the values of the viscoelastic parameters and the measured output signal; the generalization of the results, involving both the nonlinear terms of the models and possible nonlinearities of the parameters, is then illustrated through some simulations results, obtained considering the fully nonlinear system.
Then, setting
(20) |
and, as a consequence,
(21) |
it is possible to rewrite Equation (19) as:
(22) |
Under the hypothesis of , , small enough to have , , and , the Taylor series approximation, up to the first order, can be introduced
(23) |
which, once used in Equation (22), gives the approximated expression
(24) |
Performing all the computations, the latter expression can be expanded as
(25) |
After simplifications, Equation (25) can be reordered as
(26) |
where the terms corresponding to an approximation of order one are collected and separated from the others. Neglecting once more, as done in Equation (23), all the terms of approximated order greater than one, the linear approximation of Equation (19) assumes the final expression
(27) |
The same considerations can be made for Equation (15), under zero input and once Equation (18) is considered. After easy computations one gets for the linearized approximated expression up to the first order
(28) |
The Laplace transforms of Equations (27) and (28) can be performed; recalling that given a Laplace transformable time function , with and , one has that , and , Equation (27) becomes:
(29) |
while Equation (28) is transformed into
(30) |
According to Equations (9) and (10), and can be explicitly written as functions of and computing the first order series expansion of (9) and (10):
(31) |
and
(32) |
where denotes the evaluation of function at , , i.e., . Explicitly, recalling expressions (7) and (8), one gets
(33) |
Analogous computations bring to the expressions
(34) |
(35) |
(36) |
Performing the Laplace transformation, so getting
and substituting the obtained expressions in Equations (29) and (30), after some manipulations one gets
(37) |
for Equation (29), while Equation (30) becomes:
(38) |
Such expressions, under suitable definitions of coefficients and , can be written as
or, in a more compact form,
which, by inversion, gives
(39) |
The – relationship is then described by the transfer function
(40) |
where
(41) |
and
(42) |
Taking into consideration the linear representation of the relationship between the small action of the torque and the corresponding small variation of angle , which holds under the hypothesis previously discussed, the classical theory of the steady state response of a linear system under canonical inputs can be applied.
Within this framework, recalling that feeding an asymptotically stable linear system, described by a transfer function , with a sinusoidal input, which in this case can be set as
(43) |
the output, at steady state, assumes the form
(44) |
where and are such that
(45) |
If the input torque is planned by the operator that performs the measurement, the constant value has to be fixed in order to guarantee that ; otherwise, if a closed loop control system as in Figure 11 is used, the input reference
(46) |
must be chosen in order to produce, as gripper input, the torque (43), by setting the value of suitably.
Then, it is possible to evaluate the unknown damping coefficient value by measuring the amplitude of the steady state sinusoidal behaviour of the angle after that a sinusoidal exciting torque is applied to the first joint, thanks to the known dependency of from the parameter itself. A corresponding dependency holds for the phase displacement between input and output signals.
5.3. The Adopted Operational Scheme
The full measurement scheme is depicted in Figure 11.
With reference to Figure 11 and considering that noise cannot be neglected in real applications, Type blocks are those whose outputs consist in the constant steady state values of the angles and ; noise reduction can be obtained designing as an averaging filter or, somehow equivalently, as a low pass filter with a bandwidth . Type block computes the inversion of the relationship between the amplitudes of the input and output constant values as a function of the elastic coefficient . Such a relation is represented by the gain of the system under linearity hypothesis; in a more general case, it is a nonlinear function which can be analytically or experimentally computed.
The partial scheme represented in Figure 11 by Type and blocks allows the determination of the elastic coefficient .
The remaining part is devoted to the computation of the damping coefficient . The approach is the same as for the computation of , but the operation must be performed under time variable behaviours of the signals involved. As previously introduced, inspired by the expressions in the linearly approximated case, the signals considered are sinusoidal. However, this kind of signals, can be conveniently used also when the dependency of the measured viscous coefficient on velocity is nonlinear. In fact, the use of different frequencies for the torque excitation affects directly the steady state velocities of all the mechanical components of the structure, including the sample tissue. Some of the following figures show how the value of depends on the actual working frequency. With reference to Figure 11, the steady state behaviour of the angle is given by the constant value produced by the term in plus the sinusoidal part as the response of the sinusoidal term in . Then, in order to extract the amplitude of the sinusoidal component only, which is the quantity related to the damping coefficient , the scheme proposes the preliminary computation of the maximum value of the steady state response, performed by Type block, and its comparison, performed by Type block, with the constant component computed by block .
Different choices can be adopted for Type block according to the devices and the technologies used. A first possibility is to obtain the maximum value by means of a device which can detect directly such a quantity; then a peak detector or a function envelope detector can be used. They give the quantity plus the constant contribution. In these cases, block has simply to subtract the constant term and to scale by the quantity to produce . A second possibility is represented by the computation of
(47) |
where is such that and .
From the expression of Equation (47) gives
(48) |
and in the block the computation
(49) |
has to be performed.
Analogously to Type block, Type block is devoted to the computation of the inverse function, analytically or experimentally determined, that relates the amplitude to the damping coefficient .
The generic output time evolution is reported in Figure 12, where the initial phase of the constant reference and the following effect of the sinusoidal input are well evidenced.
Numerical simulations, performed using Matlab® and Simulink® tools, can help to show the effectiveness of the proposed approach.
Figure 13 reports the ratio of the amplitude of the sinusoidal output to the corresponding input couple , for different values of the parameter and for different frequencies, keeping the value of the elastic term constant, . The great dependency of such a quantity from at all the pulses greater than rad/s confirms the effectiveness of the technique here proposed for its measurement. An equivalent result is obtained if the same ratio is depicted for different values of both the parameters and , for a fixed frequency. Figure 14 shows once again the high dependency of such a measure from .
The measurement of the input torque can be avoided, considering only angle measures. In this case, Figure 15 and Figure 16 have to be considered instead of Figure 13 and Figure 14, respectively. As expected, the results are substantially comparable, supporting this possible way as an easy alternative for the measurement scheme.
The relationships between input and output under sinusoidal signals involve also the phase delay between the two signals. Figure 17 illustrates the dependency of on the sample viscosity and the signal angular frequency for an assigned value of the sample stiffness , while Figure 18 shows how the phase delay depends on and , when the system is operated by the same pulse . The marked sensitivity of on is confirmed.
6. Conclusions
The paper has shown the feasibility of a method to detect not only the stiffness but also the viscosity coefficient of a tissue sample for the purpose of identifying diseases in living organisms. All the presented simulations showed that the new microsystem has a great potential to enable researchers to investigate over samples having very limited size, about 200 m 200 m. The new interdisciplinary and integrated approach revealed its high attitude to these kind of measurements mainly because it takes advantage from different fields, such as mechanics, modern micro–fabrication techniques and control theory. The practical utility of the proposed approach will soon be proven by carrying out an experimental campaign on real biological tissues.
Author Contributions
Paolo Di Giamberardino designed and developed the control system. Alvise Bagolini and Pierluigi Bellutti fabricated the new silicon microgripper. Matteo Verotti and Fabio Botta developed the equations adopted for the kinematic and dynamic analysis of the system. Matteo Verotti and Nicola P. Belfiore designed the mechanical components of the microgripper. Imre J. Rudas and Nicola P. Belfiore integrated the different components into a whole microsystem.
Conflicts of Interest
The authors declare no conflict of interest.
References
- 1.Morrison B., III, Meaney D.F., McIntosh T.K. Mechanical characterization of an in vitro device designed to quantitatively injure living brain tissue. Ann. Biomed. Eng. 1998;26:381–390. doi: 10.1114/1.61. [DOI] [PubMed] [Google Scholar]
- 2.Edsberg L.E., Cutway R., Anain S., Natiella J.R. Microstructural and mechanical characterization of human tissue at and adjacent to pressure ulcers. J. Rehabil. Res. Dev. 2000;37:463–471. [PubMed] [Google Scholar]
- 3.Wakatsuki T., Kolodney M.S., Zahalak G.I., Elson E.L. Cell mechanics studied by a reconstituted model tissue. Biophys. J. 2000;79:2353–2368. doi: 10.1016/S0006-3495(00)76481-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Sacks M.S., Sun W. Multiaxial mechanical behavior of biological materials. Annu. Rev. Biomed. Eng. 2003;5:251–284. doi: 10.1146/annurev.bioeng.5.011303.120714. [DOI] [PubMed] [Google Scholar]
- 5.Yang W., Fung T.C., Chian K.S., Chong C.K. Viscoelasticity of esophageal tissue and application of a QLV model. J. Biomech. Eng. 2006;128:909–916. doi: 10.1115/1.2372473. [DOI] [PubMed] [Google Scholar]
- 6.Allen K.D., Athanasiou K.A. Viscoelastic characterization of the porcine temporomandibular joint disc under unconfined compression. J. Biomech. 2006;39:312–322. doi: 10.1016/j.jbiomech.2004.11.012. [DOI] [PubMed] [Google Scholar]
- 7.Levesque P., Gauvin R., Larouche D., Auger F.A., Germain L. A computer-controlled apparatus for the characterization of mechanical and viscoelastic properties of tissue-engineered vascular constructs. Cardiovasc. Eng. Technol. 2011;2:24–34. doi: 10.1007/s13239-011-0033-y. [DOI] [Google Scholar]
- 8.Dagdeviren C., Shi Y., Joe P., Ghaffari R., Balooch G., Usgaonkar K., Gur O., Tran P.L., Crosby J.R., Meyer M., et al. Conformal piezoelectric systems for clinical and experimental characterization of soft tissue biomechanics. Nature Materials. 2015;14:728–736. doi: 10.1038/nmat4289. [DOI] [PubMed] [Google Scholar]
- 9.Botta F., Marx N., Gentili S., Schwingshackl C., Di Mare L., Cerri G., Dini D. SPIE Smart Structures and Materials + Nondestructive Evaluation and Health Monitoring. International Society for Optics and Photonics; San Diego, CA, USA: 2012. Optimal placement of piezoelectric plates for active vibration control of gas turbine blades: experimental results; p. 83452H. [Google Scholar]
- 10.Botta F., Dini D., Schwingshackl C., di Mare L., Cerri G. Optimal placement of piezoelectric plates to control multimode vibrations of a beam. Adv. Acoust. Vib. 2013;2013 doi: 10.1155/2013/905160. [DOI] [Google Scholar]
- 11.Botta F., Marx N., Schwingshackl C., Cerri G., Dini D. A wireless vibration control technique for gas turbine blades using piezoelectric plates and contactless energy transfer; Proceedings of the ASME Turbo Expo; San Antonio, TX, USA. 3–7 June 2013. [Google Scholar]
- 12.Kiss M.Z., Varghese T., Hall T.J. Viscoelastic characterization of in vitro canine tissue. Phys. Med. Biol. 2004;49:4207–4218. doi: 10.1088/0031-9155/49/18/002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Chen K., Yao A., Zheng E.E., Lin J., Zheng Y. Shear wave dispersion ultrasound vibrometry based on a different mechanical model for soft tissue characterization. J. Ultrasound Med. 2012;31:2001–2011. doi: 10.7863/jum.2012.31.12.2001. [DOI] [PubMed] [Google Scholar]
- 14.Tavakoli M., Aziminejad A., Patel R., Moallem M. Multi-sensory force/deformation cues for stiffness characterization in soft-tissue palpation; Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society; New York, NY, USA. 31 August–3 September 2006; pp. 837–840. [DOI] [PubMed] [Google Scholar]
- 15.Boonvisut P., Çavuşoǧlu M.C. Estimation of soft tissue mechanical parameters from robotic manipulation data. IEEE/ASME Trans. Mech. 2013;18:1602–1611. doi: 10.1109/TMECH.2012.2209673. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Ebenstein D.M., Pruitt L. Nanoindentation of biological materials. Nano Today. 2006;1:26–33. doi: 10.1016/S1748-0132(06)70077-9. [DOI] [Google Scholar]
- 17.Cox M.A.J., Driessen N.J.B., Boerboom R.A., Bouten C.V.C., Baaijens F.P.T. Mechanical characterization of anisotropic planar biological soft tissues using finite indentation: Experimental feasibility. J. Biomech. 2008;41:422–429. doi: 10.1016/j.jbiomech.2007.08.006. [DOI] [PubMed] [Google Scholar]
- 18.Saxena T., Gilbert J., Stelzner D., Hasenwinkel J. Mechanical characterization of the injured spinal cord after lateral spinal hemisection injury in the rat. J. Neurotrauma. 2012;29:1747–1757. doi: 10.1089/neu.2011.1818. [DOI] [PubMed] [Google Scholar]
- 19.González-Cruz R.D., Fonseca V.C., Darling E.M. Cellular mechanical properties reflect the differentiation potential of adipose-derived mesenchymal stem cells. Proc. Natl. Acad. Sci. USA. 2012;109:E1523–E1529. doi: 10.1073/pnas.1120349109. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Ficarella E., Lamberti L., Papi M., De Spirito M., Pappalettere C. Mechanics of Biological Systems and Materialsz. Volume 6. Springer International Publishing; Cham, Switzerland: 2017. Viscohyperelastic calibration in mechanical characterization of soft matter; pp. 33–37. [Google Scholar]
- 21.Mazza E., Nava A., Hahnloser D., Jochum W., Bajka M. The mechanical response of human liver and its relation to histology: An in vivo study. Med. Image Anal. 2007;11:663–672. doi: 10.1016/j.media.2007.06.010. [DOI] [PubMed] [Google Scholar]
- 22.Zhao R., Sider K.L., Simmons C.A. Measurement of layer-specific mechanical properties in multilayered biomaterials by micropipette aspiration. Acta Biomater. 2011;7:1220–1227. doi: 10.1016/j.actbio.2010.11.004. [DOI] [PubMed] [Google Scholar]
- 23.Choi D.K. Mechanical characterization of biological tissues: Experimental methods based on mathematical modeling. Biomed. Eng. Lett. 2016;6:181–195. doi: 10.1007/s13534-016-0222-6. [DOI] [Google Scholar]
- 24.Nava A., Mazza E., Kleinermann F., Avis N.J., McClure J., Bajka M. Evaluation of the mechanical properties of human liver and kidney through aspiration experiments. Technol. Health Care. 2004;12:269–280. [PubMed] [Google Scholar]
- 25.Nava A., Mazza E., Furrer M., Villiger P., Reinhart W.H. In vivo mechanical characterization of human liver. Med. Image Anal. 2008;12:203–216. doi: 10.1016/j.media.2007.10.001. [DOI] [PubMed] [Google Scholar]
- 26.Boudou T., Ohayon J., Arntz Y., Finet G., Picart C., Tracqui P. An extended modeling of the micropipette aspiration experiment for the characterization of the Young’s modulus and Poisson’s ratio of adherent thin biological samples: Numerical and experimental studies. J. Biomech. 2006;39:1677–1685. doi: 10.1016/j.jbiomech.2005.04.026. [DOI] [PubMed] [Google Scholar]
- 27.Bosisio M.R., Talmant M., Skalli W., Laugier P., Mitton D. Apparent Young’s modulus of human radius using inverse finite-element method. J. Biomech. 2007;40:2022–2028. doi: 10.1016/j.jbiomech.2006.09.018. [DOI] [PubMed] [Google Scholar]
- 28.Valero C., Navarro B., Navajas D., García-Aznar J.M. Finite element simulation for the mechanical characterization of soft biological materials by atomic force microscopy. J. Mech. Behav. Biomed. Mater. 2016;62:222–235. doi: 10.1016/j.jmbbm.2016.05.006. [DOI] [PubMed] [Google Scholar]
- 29.Argento G., Simonet M., Oomens C.W.J., Baaijens F.P.T. Multi-scale mechanical characterization of scaffolds for heart valve tissue engineering. J. Biomech. 2012;45:2893–2898. doi: 10.1016/j.jbiomech.2012.07.037. [DOI] [PubMed] [Google Scholar]
- 30.Rodriguez M.L., McGarry P.J., Sniadecki N.J. Review on cell mechanics: Experimental and modeling approaches. Appl. Mech. Rev. 2013;65:060801. doi: 10.1115/1.4025355. [DOI] [Google Scholar]
- 31.Ekpenyong A., Whyte G., Chalut K., Pagliara S., Lautenschläger F., Fiddler C., Paschke S., Keyser U., Chilvers E., Guck J. Viscoelastic properties of differentiating blood cells are fate- and function-dependent. PLoS ONE. 2012;7 doi: 10.1371/journal.pone.0045237. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Gossett D.R., Tse H.T.K., Lee S.A., Ying Y., Lindgren A.G., Yang O.O., Rao J., Clark A.T., Di Carlo D. Hydrodynamic stretching of single cells for large population mechanical phenotyping. Proc. Natl. Acad. Sci. USA. 2012;109:7630–7635. doi: 10.1073/pnas.1200107109. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Hodgson A., Verstreken C., Fisher C., Keyser U., Pagliara S., Chalut K. A microfluidic device for characterizing nuclear deformations. Lab Chip. 2017;17:805–813. doi: 10.1039/C6LC01308B. [DOI] [PubMed] [Google Scholar]
- 34.Pajerowski J.D., Dahl K.N., Zhong F.L., Sammak P.J., Discher D.E. Physical plasticity of the nucleus in stem cell differentiation. Proc. Natl. Acad. Sci. USA. 2007;104:15619–15624. doi: 10.1073/pnas.0702576104. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Pagliara S., Franze K., McClain C., Wylde G., Fisher C., Franklin R., Kabla A., Keyser U., Chalut K. Auxetic nuclei in embryonic stem cells exiting pluripotency. Nat. Mater. 2014;13:638–644. doi: 10.1038/nmat3943. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Guck J., Schinkinger S., Lincoln B., Wottawah F., Ebert S., Romeyke M., Lenz D., Erickson H., Ananthakrishnan R., Mitchell D., et al. Optical deformability as an inherent cell marker for testing malignant transformation and metastatic competence. Biophys. J. 2005;88:3689–3698. doi: 10.1529/biophysj.104.045476. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Otto O., Rosendahl P., Mietke A., Golfier S., Herold C., Klaue D., Girardo S., Pagliara S., Ekpenyong A., Jacobi A., et al. Real-time deformability cytometry: On-the-fly cell mechanical phenotyping. Nat. Methods. 2015;12:199–202. doi: 10.1038/nmeth.3281. [DOI] [PubMed] [Google Scholar]
- 38.Guido I., Jaeger M., Duschl C. Dielectrophoretic stretching of cells allows for characterization of their mechanical properties. Eur. Biophys. J. 2011;40:281–288. doi: 10.1007/s00249-010-0646-3. [DOI] [PubMed] [Google Scholar]
- 39.Kamble H., Vadivelu R., Barton M., Boriachek K., Munaz A., Park S., Shiddiky M., Nguyen N.T. An electromagnetically actuated double-sided cell-stretching device for mechanobiology research. Micromachines. 2017;8:256. doi: 10.3390/mi8080256. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Verotti M., Crescenzi R., Balucani M., Belfiore N.P. MEMS-based conjugate surfaces flexure hinge. J. Mech. Des. 2015;137:012301. doi: 10.1115/1.4028791. [DOI] [Google Scholar]
- 41.Belfiore N.P., Broggiato G.B., Verotti M., Balucani M., Crescenzi R., Bagolini A., Bellutti P., Boscardin M. Simulation and construction of a MEMS CSFH based microgripper. Int. J. Mech. Control. 2015;16:21–30. [Google Scholar]
- 42.Balucani M., Belfiore N.P., Crescenzi R., Verotti M. The development of a MEMS/NEMS-based 3 D.O.F. compliant micro robot. Int. J.Mech. Control. 2011;12:3–10. [Google Scholar]
- 43.Belfiore N.P., Pennestrì E. An atlas of linkage-type robotic grippers. Mech. Mach. Theory. 1997;32:811–833. doi: 10.1016/S0094-114X(97)00006-2. [DOI] [Google Scholar]
- 44.Dochshanov A., Verotti M., Belfiore N.P. A comprehensive survey on microgrippers design: operational strategy. J. Mech. Des. 2017;139:070801. doi: 10.1115/1.4036352. [DOI] [Google Scholar]
- 45.Verotti M., Dochshanov A., Belfiore N.P. A comprehensive survey on microgrippers design: mechanical structure. J. Mech. Des. 2017;139:060801. doi: 10.1115/1.4036351. [DOI] [Google Scholar]
- 46.Verotti M., Belfiore N.P. Isotropic compliance in E(3): Feasibility and workspace mapping. J. Mech. Robot. 2016;8:061005. doi: 10.1115/1.4032408. [DOI] [Google Scholar]
- 47.Verotti M., Masarati P., Morandini M., Belfiore N.P. Isotropic compliance in the Special Euclidean Group SE(3) Mech. Mach. Theory. 2016;98:263–281. doi: 10.1016/j.mechmachtheory.2015.12.002. [DOI] [Google Scholar]
- 48.Belfiore N.P., Simeone P. Inverse kinetostatic analysis of compliant four-bar linkages. Mech. Mach. Theory. 2013;69:350–372. doi: 10.1016/j.mechmachtheory.2013.06.008. [DOI] [Google Scholar]
- 49.Belfiore N.P., Verotti M., Di Giamberardino P., Rudas I. Active joint stiffness regulation to achieve isotropic compliance in the euclidean space. J. Mech. Robot. 2012;4 doi: 10.1115/1.4007307. [DOI] [Google Scholar]
- 50.Verotti M., Dochshanov A., Belfiore N.P. Compliance synthesis of CSFH MEMS-based microgrippers. J. Mech. Des. 2017;139:022301. doi: 10.1115/1.4035053. [DOI] [Google Scholar]
- 51.Cecchi R., Verotti M., Capata R., Dochshanov A., Broggiato G., Crescenzi R., Balucani M., Natali S., Razzano G., Lucchese F., et al. Development of micro-grippers for tissue and cell manipulation with direct morphological comparison. Micromachines. 2015;6:1710–1728. doi: 10.3390/mi6111451. [DOI] [Google Scholar]
- 52.Bagolini A., Ronchin S., Bellutti P., Chiste M., Verotti M., Belfiore N.P. Fabrication of novel MEMS microgrippers by deep reactive ion etching with metal hard mask. IEEE J. Microelectromechanical Syst. 2017;26:926–934. doi: 10.1109/JMEMS.2017.2696033. [DOI] [Google Scholar]
- 53.Lim C., Zhou E., Quek S. Mechanical models for living cells—A review. J. Biomech. 2006;39:195–216. doi: 10.1016/j.jbiomech.2004.12.008. [DOI] [PubMed] [Google Scholar]
- 54.Verotti M. Analysis of the center of rotation in primitive flexures: Uniform cantilever beams with constant curvature. Mech. Mach. Theory. 2016;97:29–50. doi: 10.1016/j.mechmachtheory.2015.10.007. [DOI] [Google Scholar]
- 55.Verotti M. Effect of initial curvature in uniform flexures on position accuracy. Mech. Mach. Theory. 2018;119:106–118. doi: 10.1016/j.mechmachtheory.2017.08.021. [DOI] [Google Scholar]
- 56.Rao S.S. Mechanical Vibrations. Addison-Wesley; Reading, MA, USA: 1993. [Google Scholar]
- 57.Tu C.C., Fanchiang K., Liu C.H. 1 × N rotary vertical micromirror for optical switching applications. Proc. SPIE. 2005;5719:14–22. [Google Scholar]
- 58.Hou M.T.K., Huang J.Y., Jiang S.S., Yeh J.A. In-plane rotary comb-drive actuator for a variable optical attenuator. J. Micro/Nanolithography MEMS MOEMS. 2008;7 doi: 10.1117/1.3013547. [DOI] [Google Scholar]