Abstract
It is critical to distinguish between cardiac tumors and thrombi because they require different treatment strategies. Although accurate differentiation of these cardiac masses can be challenging, computed tomography (CT) and magnetic resonance imaging (MRI) are promising tools to improve their diagnosis. This study aimed to assess the diagnostic value of a volume-based quantification strategy using dual-energy CT to differentiate between cardiac tumors and thrombi. We prospectively enrolled 41 patients who had a cardiac mass. All patients underwent electrocardiography gated dual-energy CT. Among them, 28 patients underwent late gadolinium enhancement cardiac MRI. For quantitative analysis, the following parameters of the entire cardiac masses were measured: CT attenuation values in Hounsfield units (HU), iodine concentration (mg/ml), and signal intensity (SI) ratio. A mixed effects model was used to evaluate the significance of differences in mean CT attenuation, mean iodine concentration, and SI ratios between the cardiac tumor and thrombus groups. Diagnostic performance of each parameter was evaluated by constructing a receiver operating characteristics curve. A total of 24 cardiac tumors and 19 cardiac thrombi were analyzed. The mean iodine concentration was significantly higher in tumors than in thrombi (tumors: 2.98 ± 0.23; thrombi: 1.79 ± 0.26, p = 0.002). The diagnostic performance of iodine concentration was better than that of post-contrast HU (area under the curve [AUC]: 0.77 vs. 0.51; p < 0.001), and worse than that of SI ratio (AUC: 0.89; p = 0.04) for differentiation of cardiac tumors and thrombi. Dual-energy CT using volume-based iodine measurements can differentiate between cardiac tumors and thrombi.
Introduction
Although cardiac masses are not common, they are potentially high-risk sources of embolism, so accurate and prompt diagnosis is important1. It is also clinically important to differentiate cardiac tumors from cardiac thrombi, as they are managed differently2,3.
Cardiac magnetic resonance imaging (MRI) has become a mainstream imaging modality for differential diagnosis of cardiac tumors because it provides an unrestricted field of view and superior tissue characterization without radiation exposure4,5. A previous study demonstrated that late gadolinium enhancement (LGE) cardiac MRI is a useful non-invasive modality to differentiate a cardiac tumor from a thrombus6.
Multi-detector computed tomography (CT) is increasingly being used for cardiac imaging7. However, CT lacks the superior contrast abilities of MRI and is therefore not suitable for differentiation between cardiac tumors and thrombi6,8.
Dual-energy CT using rapid kilovoltage (kVp) switching allows simultaneous acquisition of low and high kVp datasets. It is possible to differentiate iodine from other materials using the material decomposition method9. This is valuable for differentiating between iodine-enhancing lesions such as cardiac tumors, and non-enhancing lesions such as thrombi. A previous study demonstrated that dual-energy CT was useful for characterizing cardiac masses and differentiating between cardiac tumors and thrombi based on quantification of the iodine content of the mass8. However, the lack of electrocardiography synchronization in dual-energy mode and evaluation of only certain areas of the mass were limitations of the study.
The recent development of simultaneous electrocardiography synchronization in rapid kVp-switching dual-energy mode yields good quality CT images without artifacts or image degradation. In addition, newly developed software allows whole tumor segmentation, which enables whole-tumor evaluation.
The purpose of this study was to assess the diagnostic value of a volume-based quantification using dual-energy cardiac CT to differentiate between cardiac tumors and thrombi, and to compare quantitative CT values with LGE parameters.
Results
Demographic and baseline characteristics of patients
The demographic data and baseline clinical characteristics of the 41 patients are summarized in Table 1. The mean diameter of cardiac tumors was significantly larger than that of thrombi (p < 0.001, all longest and shortest diameters). Patients with thrombi had higher incidences of smoking, history of cerebrovascular disease, heart disease, and arrhythmia than those with cardiac tumors. Other clinical characteristics of the two groups were not significantly different. A total of 24 cardiac tumors and 19 cardiac thrombi were identified among the 41 patients, as two patients each had two thrombi. Among the 24 cardiac tumors, 15 were benign, and included myxoma (n = 9), cavernous hemangioma (n = 2), lipoma (n = 2), angiofibroma (n = 1), and papillary fibroelastoma (n = 1). Nine were malignant and included lymphoma (n = 4) and metastasis (n = 5). All tumors were confirmed by surgical excision or biopsy. One thrombus was confirmed by surgical excision; all other thrombi disappeared after anticoagulation therapy, which was considered proof of diagnosis.
Table 1.
Cardiac tumors (n = 24) | Cardiac thrombi (n = 17)* | P value | |
---|---|---|---|
Male | 10 (58.82) | 8 (57.14) | 0.786 |
Age (years) | 58.13 ± 17.46 | 59.47 ± 12.64 | 0.788 |
Axis size (mm) | |||
Long | 39.48 ± 14.86 | 23.46 ± 12.81 | 0.001 |
Short | 29.03 ± 10.99 | 15.13 ± 7.69 | < 0.001 |
Location | 0.133 | ||
Left atrium | 7 (29.17) | 11 (57.89) | |
Left ventricle (LV) | 3 (12.50) | 4 (21.05) | |
Right atrium | 10 (41.67) | 4 (21.05) | |
Right ventricle | 4 (16.67) | 0 (0.00) | |
Diabetes mellitus | 5 (20.83) | 6 (35.29) | 0.476 |
Hypertension | 8 (33.33) | 9 (52.94) | 0.209 |
Smoking | 5 (20.83) | 9 (52.94) | 0.033 |
Hyperlipidemia | 1 (4.17) | 2 (11.76) | 0.560 |
Prior CVA/TIA | 2 (8.33) | 8 (47.06) | 0.008 |
Heart disease† | 1 (4.17) | 8 (47.06) | 0.002 |
Arrhythmia | 6 (25.00) | 10 (58.82) | 0.029 |
Note: Values are given as mean ± SD, Data in parentheses are percentages.
*As two patients each had two thrombi, 17 patients were included in the demographics and 19 lesions were included in the location analysis.
†Heart disease includes valvular heart disease and myocardial infarction.
CVA, cerebrovascular accident; TIA, transient ischemic attack.
Quantitative analysis of dual-energy cardiac CT and MRI of cardiac tumors and thrombi
All 41 patients underwent dual-energy CT; 28 of the 41 patients underwent cardiac MRI. There were no adverse events during CT or MRI examinations. Among the 24 cases of cardiac tumors, 16 had available MRI data; among 19 cases of cardiac thrombi, 12 had available MRI data within a 2-week timeframe.
The detailed quantitative values for the cardiac tumors and thrombi are shown in Table 2. The volume of the cardiac tumors was significantly larger than that of the thrombi (24.52 ± 3.76 vs. 3.98 ± 4.23 -> 4.22, p = 0.001 -> p < 0.001). The mean CT attenuation values of the cardiac tumors and thrombi measured on post-contrast CT images were not significantly different (73.03 ± 8.78 vs. 74.50 ± 9.87 Hounsfield units (HU), respectively; p = 0.912). On iodine maps, the mean iodine concentration (mg/mL) was significantly higher in the cardiac tumors than in the thrombi (cardiac tumors: 2.98 ± 0.23; cardiac thrombi: 1.79 ± 0.26; p = 0.002). On MRI, the signal intensity (SI) ratio was also significantly different between the cardiac tumors and thrombi (6.90 ± 1.25 -> 0.12 vs. 1.43 ± 1.44 -> 0.14 respectively; p = 0.008). Representative cases of the cardiac tumors and thrombi are shown in Figs 1–3.
Table 2.
Cardiac tumors (n = 24) | Cardiac thrombi (n = 19) | P value (tumors vs. thrombi) | |||||
---|---|---|---|---|---|---|---|
Lymphoma (n = 4) | Metastasis (n = 5) | Myxoma (n = 9) | Other* (n = 6) | All (n = 24) | |||
Volume (cm3) | 32.14 ± 8.77 | 36.50 ± 7.84 | 24.24 ± 5.84 | 9.89 ± 7.16 | 24.52 ± 3.76 | 3.98 ± 4.22 | <0.001 |
CT attenuation value (HU) | 94.63 ± 21.11 | 88.60 ± 18.88 | 74.11 ± 14.08 | 44.02 ± 17.24 | 73.03 ± 8.78 | 74.50 ± 9.87 | 0.912 |
Iodine concentration (mg/mL) | 3.04 ± 0.58 | 2.96 ± 0.52 | 3.33 ± 0.39 | 2.43 ± 0.47 | 2.98 ± 0.23 | 1.79 ± 0.26 | 0.002 |
SI ratio† | 4.13 ± 2.76 | 3.31 ± 3.39 | 10.30 ± 1.95 | 5.92 ± 2.14 | 6.90 ± 0.12 | 1.43 ± 0.14 | 0.008 |
Note: Estimated mean ± SE (standard error) by the mixed effects model.
*Other tumor included cavernous hemangioma (n = 2), lipoma (n = 2), angiofibroma (n = 1), and papillary fibroelastoma (n = 1).
†For SI ratio, sample size was 16 for cardiac tumors and 12 for cardiac thrombi.
HU, Hounsfield units; SI, signal intensity.
Subgroup analysis of the quantitative results
Differences in quantitative values were analyzed between benign (n = 15: myxoma [n = 9], cavernous hemangioma [n = 2], lipoma [n = 2], angiofibroma [n = 1]) and malignant tumors (n = 9: lymphoma [n = 4], metastasis [n = 5]). However, there was no significant difference in parameters between the benign and malignant tumors, including HU (69.02 ± 7.21 vs. 91.28 ± 14.00, p = 0.17), iodine concentration (2.31 ± 0.22 vs. 2.99 ± 0.42, p = 0.16), or SI ratio (4.72 ± 1.19 vs. 3.80 ± 2.56, p = 0.75). According to the tumor type, there were significant differences in the iodine concentration between myxomas and thrombi (3.33 ± 0.39 vs. 1.79 ± 0.26, p = 0.002) and metastasis and thrombi (2.96 ± 0.52 vs 1.79 ± 0.26, p = 0.042). The SI ratio was significantly different between myxoma and thrombi (10.30 ± 1.95 vs 1.43 ± 0.14, p = 0.001) and other benign tumors and thrombi (5.92 ± 2.14 vs 1.43 ± 0.14, p = 0.026). However, there was no significant difference in iodine concentration or the SI ratio between other tumor subtypes. HU showed no significant difference between any tumor subtype.
Diagnostic performance of dual-energy CT and MRI
To evaluate the diagnostic performance of dual-energy CT and MRI for differentiation of cardiac tumors from thrombi, the receiver operating characteristics (ROC) curve and the area under the curve (AUC) of each parameter were evaluated.
When iodine concentration was used to differentiate cardiac tumors from thrombi, the AUC was 0.77 (95% confidence interval [CI]: 0.63–0.91). The optimal cutoff value of iodine concentration for differentiation between tumor and thrombus was 2.55 mg/mL; sensitivity and specificity were 66.70% and 79.00%, respectively.
The AUC of post-contrast HU was 0.51 (95% CI: 0.34–0.69). There was a significant difference between the AUC of iodine concentration and post-contrast HU (difference in AUC: −0.38 [95% CI: −0.49–−0.27]; p < 0.001).
The AUC of the SI ratio was 0.89 (95% CI: 0.75–1.00), which was significantly higher than that of the iodine concentration (difference of AUC: −0.13 [95% CI: −0.26–−0.004]; p = 0.043). The ROC curves of post-contrast HU, iodine concentration, and SI ratio are included in Fig. 4.
Correlation between dual-energy CT and MRI data
The correlation between the SI ratio and iodine concentration was good (0.74 [95% CI: 0.51–0.85]) in all patients; the correlation between the SI ratio and post-contrast CT attenuation value was relatively low (0.41 [95% CI: −0.24–0.87], Table 3).
Table 3.
Parameter | All patients | Cardiac tumors (n = 24) | Cardiac thrombi (n = 19) |
---|---|---|---|
Iodine concentration (mg/ml) | 0.74 (0.51, 0.85) |
0.77 (0.50, 0.92) |
0.39 (−0.03, 0.98) |
CT attenuation value (HU) | 0.41 (−0.24, 0.87) |
0.43 (−0.46, 0.96) |
0.55 (0.12, 0.98) |
SI, signal intensity; CI, confidence interval; HU, Hounsfield units.
There was excellent inter-observer agreement between the two radiologists regarding the measurement parameters of dual-energy CT and MRI. The intra-class correlation coefficients (ICC) for iodine concentration, SI ratio, and post-contrast HU were 0.98 (95% CI: 0.97–0.99), 0.97 (95% CI: 0.93–0.98), and 0.98 (95% CI: 0.97–0.99), respectively. Radiation exposure was estimated from the dose-length product (DLP). The calculated mean radiation dose was 4.58 mSv (DLP: 146.0–462.2 mGy × cm).
Discussion
This study aimed to evaluate the diagnostic performance of dual-energy CT and compare it with those of LGE and CT parameters for the differentiation of cardiac tumors and thrombi. We demonstrated that volume-based iodine measurements using dual-energy CT provide a feasible quantitative parameter for differentiating cardiac tumors from thrombi.
Cardiac MRI plays an important role in the assessment of cardiac masses. It allows multi-parametric characterization of cardiac masses using topography, morphology, SI, or pattern of enhancement of cardiac tumors3,10,11. However, MRI has several disadvantages, including a long scanning time, and it cannot be used to examine patients with implanted metallic devices.
LGE imaging, widely used to distinguish between viable and infarcted myocardium, allows identification of a thrombus in the absence of contrast enhancement. Several studies have demonstrated that LGE can accurately delineate a cardiac thrombus from the myocardium and chamber cavity based on tissue characteristics, irrespective of location or morphology12–15. Previous studies have demonstrated that the contrast enhancement pattern obtained from LGE can be used to differentiate between cardiac tumors and thrombi12–14. In an earlier study, cardiac tumors presented as high-SI masses compared to viable myocardium, reflecting gadolinium contrast enhancement. In comparison, thrombi presented as low-SI masses (either homogeneously black, or as a black border around a central gray zone), indicating avascular tissue6. The present study also demonstrated that LGE imaging can distinguish various cardiac tumors from thrombi, which showed higher diagnostic value than CT parameters.
CT is a widely available and highly reproducible imaging technique that can quickly, accurately, and noninvasively determine cardiac anatomy. CT has several advantages over MRI, including coverage of the whole myocardium in a short scanning time, widespread availability, and simultaneous evaluation of coronary arteries. Additionally, CT can replace MRI in cases with contraindications. Recently, dual-energy CT was proved to be a non-invasive modality that allows differentiation of cardiac myxomas from thrombi8. A previous study revealed that iodine concentration measurements were significantly higher in tumors than in thrombi, while mean CT attenuation values were not significantly different between the two mass types8. However, the limitation of that study was that only myxomas were evaluated. In addition, iodine quantification in the previous study was limited by the lack of synchronization of electrocardiography with cardiac movements using dual-energy mode. Therefore, identification of a simple quantitative parameter to differentiate cardiac tumors and thrombi within a heterogeneous cardiac mass is potentially challenging.
We conducted this study to investigate the ability of dual-energy CT to differentiate between cardiac tumors and thrombi. Iodine quantification may have measurement variability for technical (i.e., scanner, acquisition technique, patient compliance) and biological (variability of tumor blood supply) reasons16,17. In this study, we quantified the whole volume of the cardiac mass using electrocardiographic synchronization with cardiac movements using dual-energy mode.
According to our results, mean CT attenuation values were not significantly different between cardiac tumors and thrombi, while mean iodine concentration was significantly higher in cardiac tumors than in thrombi. In addition, the diagnostic performance of iodine concentration was significantly higher than that of post-contrast HU measurements for distinguishing cardiac tumors from thrombi. However, the AUC of iodine concentration was significantly lower than that of the SI ratio of LGE imaging. This result suggests that dual-energy CT is more useful for differentiation between tumors and thrombi than conventional CT. Dual-energy CT is an accurate approach for defining whether a cardiac mass is enhanced. Iodine quantification appears more accurate than standard enhancement measurements when the volume of interest (VOI) covers the entire mass. Although cardiac MRI is a reference modality for evaluation of cardiac masses10,11, dual-energy CT can play a complementary role to that of MRI in cases when MRI is not available or in which echocardiography or conventional CT results are inconclusive.
Several limitations of this study should be acknowledged. First, a small sample size and limited histologic types of cardiac masses were included. As cardiac tumors are a rare disease entity, it was difficult to enroll a large population for this prospective study. However, according to our sample size estimation, 19 subjects in each group could have a statistical power of 0.8. Second, our contrast injection protocol may have significantly influenced the results. Few studies have investigated the role of dual-energy CT in cardiac tumor imaging and, to our knowledge, there are neither data on contrast media administration nor standard scanning protocols available. Previous studies have found that dual-energy CT performed using a double injection protocol showed both high sensitivity and high specificity for detection and differentiation of cardiac thrombi and blood stasis18. As most thrombi do not show enhancement (whereas tumors often do), a double injection protocol was used to achieve a sufficient attenuation difference between iodine-enhancing and non-enhancing lesions. Third, cardiac MRI and CT were not performed on the same day; the median time interval between the two techniques was 4.8 days. In addition, only 28 of the 41 patients underwent MRI.
In conclusion, dual-energy CT using volume-based iodine measurements can be used to differentiate between cardiac tumors and thrombi. Although dual-energy CT is not always performed in routine practice due to the radiation exposure associated with this technique, it can be a helpful complementary tool to MRI in cases when MRI is not available or in which echocardiography or conventional CT provide inconclusive results.
Materials and Methods
Patient selection
This single-center prospective observational study was approved by the institutional review board of Yonsei University Health System (approval number: 4-2012-0396). Informed consent was obtained from all patients. All studies were performed in accordance with the relevant guidelines and regulations.
From December 2012 to November 2016, 67 consecutive patients with a suspicious intracardiac mass on echocardiography or conventional CT were prospectively enrolled to undergo dual-energy CT and MRI. Patients met inclusion criteria if they were more than 19 years of age and had a cardiac mass lesion, which was defined as a lesion greater than 10 mm in the longest diameter. Exclusion criteria included: pericardial or myocardial mass (n = 3), inconclusive final results (n = 13), contrast agent allergy (n = 1), renal dysfunction (n = 2), and refusal to provide written informed consent (n = 7). Finally, 41 patients (25 men, 16 women; mean age, 59.30 years; range 19–87 years) were included. The patient selection process is summarized in Fig. 5.
Final diagnoses were based on either pathological results from surgery or biopsy, a follow-up examination including echocardiography, or MRI after anticoagulant treatment without additional treatment.
Dual-energy cardiac computed tomography (CT) examination
Dual-energy CT scans were performed with a 64-row multidetector CT scanner (Discovery CT750 HD; GE Healthcare, Chicago, IL, USA). No beta-blockers were used to regulate heart rate. The mean heart rate was 64.5 ± 9 beats per min (range 47–90 bpm) during the CT examination.
For contrast injection, we followed a double injection method that requires two injections and one scan18,19. For the first bolus injection, 50 mL of the nonionic contrast agent iopamidol (370 mg of iodine/mL, Iopamiro®, Bracco Imaging, S.p.A., Milan, Italy), was administered via power injector (Envision CT, Medrad, Warrendale, PA, USA) at 5 mL/s into the right antecubital vein. A region of interest (ROI) was plotted inside the ascending aorta and a bolus geometry curve was acquired. Curve diagrams were analyzed to determine the optimal scan delay.
The second bolus injection of 70 mL of contrast agent was administered intravenously at the same rate 180 s after the end of the first bolus injection, followed by 50 mL of contrast-saline mixture at a ratio of 30:70. The scan was begun after the optimal scan delay using the gemstone spectral imaging (GSI) mode (electrocardiography-gated dual-energy CT mode). The scan parameters were: detector collimation = 64 × 0.625 mm; gantry rotation time = 0.5 s; tube voltage = 140/80 kV; tube current = 630 mA; pitch = 1.375:1. All images were reconstructed with a 0.625-mm slice thickness.
Image analysis of dual-energy CT
Dual-energy CT data were assessed by two radiologists with 8 and 11 years of cardiac imaging experience. These radiologists were blinded to patient data. All scan images were transferred and computed in GSI Volume Viewer mode on a dedicated workstation equipped with dual-energy post-processing software (GE workstation, Volumeshare 4.4.5, GE Healthcare).
Size (longest diameters of the long and short axes images) and volume of the intra-cardiac masses were measured. The two radiologists performed these measurements independently and inter-observer reliability was calculated. The cardiac mass locations were also recorded. Locations were classified as follows: left atrium, left ventricle, right atrium, or right ventricle.
After processing in GSI volume viewer mode, three modes of images, namely monochromatic 70-keV images (which represent conventional 120-kV post-contrast CT images), material-suppressed iodine (MSI) images (which represent conventional 120-kV pre-contrast CT images), and iodine (water) images were displayed on the workstation, side-by-side, to show the same level of the mass.
For quantitative analysis, the entire VOI of the mass was isolated using semi-automatic segmentation. If the segmented border of the mass was incorrect, the reviewers manually corrected the VOI per the border of the mass on slices that included the cardiac mass on axial monochromatic 70-keV images. After drawing a VOI covering the entire mass, the segmented VOI of the mass was propagated to the other two modes’ alternate images to cover the same VOI of the tumor across the three images (Fig. 6).
The mean CT attenuation densities of the entire cardiac masses in post-contrast CT images (monochromatic 70-keV images) and pre-contrast CT images (MSI images) were measured. In iodine (water) images, the mean iodine concentration (mg/ml) for the same VOI covering the entire tumor was also measured.
Cardiac MRI examination
MRI and dual-energy CT were performed within a 2-week timeframe. MRI was performed with a 1.5 Tesla (T) MR unit (Gyroscan Intera, Philips Medical Systems, Amsterdam, The Netherlands). LGE was performed using a segmented inversion recovery radiofrequency spoiled gradient echo (T1-TFE) sequence (TR/TE = 5.3/1.6 ms, flip angle = 15°, slice thickness = 10 mm, field of view = 360 mm, matrix = 512 × 512, number of signal average = 2), 10 min after intravenous injection of gadolinium-DTPA at a dose of 0.2 mmol/kg. The inversion time was determined using a dedicated TI-determining sequence (Look-Locker); the range was 220–300 ms.
MRI analysis
The same radiologists who performed the CT image analysis reviewed the LGE data independently using the same workstations and were again blinded to patient data; inter-observer reliability was calculated. For the quantitative analysis, the two reviewers measured the SI in the operator-defined ROI in the normal myocardium and cardiac masses on the LGE (magnitude) images; mean SI values were used in the analyses. For the cardiac masses, the ROI was drawn to cover as much of the lesion as possible on one representative image. The SI ratios of normal myocardium and cardiac masses were calculated.
Sample size consideration
From previous study data8, the expected AUC of the iodine concentration and post-contrast HU were 0.8 and 0.6, respectively. The correlation between the two measurements for each group was positively set at 0.6. A statistical power of 0.80 and type I error of 0.05 required 19 subjects per group. The sample size estimation was performed using PASS software (version 12, NCSS, LLC., Kaysville, UT, USA)20.
Statistical analysis
Categorical demographic data are expressed as numbers and percentages and were compared between the cardiac tumor and thrombus groups using the chi-square or Fisher’s exact tests. Continuous variables were expressed as means ± standard deviations. The error of mean is defined as one standard deviation. They were analyzed using a Shapiro-Wilk test to confirm a normal distribution and compared with a Student’s t-test. Differences in volume, mean post-contrast HU, mean iodine concentration, and SI ratio between the groups (cardiac tumor vs. thrombus, benign vs. malignant tumor, subtypes of tumor) were evaluated with a mixed effects model. The two groups were regarded as fixed effects, patients were considered a random effect. Correlation coefficients with 95% CI between the SI ratio and each variable were estimated based on a linear mixed model to account for replicated measurements of the two readers21. Diagnostic performance of each parameter was evaluated by an ROC curve with multi-reader and multi-case analysis using the statistical software iMRMC version 2.7 (Food and Drug Administration Center for Devices and Radiological Health, Silver Spring, MD, USA)22. The significance of differences between the AUCs of mean post-contrast HU and iodine concentration was evaluated using Delong’s method23; a different statistical method proposed by Zhou and Gatsonis24 was used to include partially paired CT data in AUC comparisons between the iodine concentration and SI ratio. We also determined the optimal cut-off value of iodine concentration to differentiate between cardiac tumors and thrombi using Youden’s index to yield the maximal sum of the sensitivity and specificity25. ICCs were used to evaluate inter-observer agreement for measurements of parameters using a two-way random effects model. Patients and radiologists were considered to be random effects. P values < 0.05 were considered significant. All statistical analyses were performed using IBM SPSS Statistics for Windows (Version 21.0, IBM Corp., Armonk, NY, USA).
Acknowledgements
This research was supported by the Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education, Science and Technology (2012-R1A1A1013152, 2014R1A1A3050905).
Author Contributions
Y.J. Hong and J. Hur enrolled patients, designed the study, and wrote the manuscript. K. Han conducted statistical analyses. D.J. Im, Y.J. Kim, H.Y. Lee, Y.J. Suh, and B.W. Choi reviewed the manuscript.
Availability of Data and Material
All data generated or analyzed during this study are included in this published article.
Competing Interests
The authors declare no competing interests.
Footnotes
Publisher's note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Murtagh B, Smalling RW. Cardioembolic stroke. Curr Atheroscler Rep. 2006;8:310–316. doi: 10.1007/s11883-006-0009-9. [DOI] [PubMed] [Google Scholar]
- 2.Doufekias E, Segal AZ, Kizer JR. Cardiogenic and aortogenic brain embolism. J Am Coll Cardiol. 2008;51:1049–1059. doi: 10.1016/j.jacc.2007.11.053. [DOI] [PubMed] [Google Scholar]
- 3.Fussen S, et al. Cardiovascular magnetic resonance imaging for diagnosis and clinical management of suspected cardiac masses and tumours. Eur Heart J. 2011;32:1551–1560. doi: 10.1093/eurheartj/ehr104. [DOI] [PubMed] [Google Scholar]
- 4.Motwani M, et al. MR imaging of cardiac tumors and masses: a review of methods and clinical applications. Radiology. 2013;268:26–43. doi: 10.1148/radiol.13121239. [DOI] [PubMed] [Google Scholar]
- 5.O’Donnell DH, et al. Cardiac tumors: optimal cardiac MR sequences and spectrum of imaging appearances. AJR Am J Roentgenol. 2009;193:377–387. doi: 10.2214/AJR.08.1895. [DOI] [PubMed] [Google Scholar]
- 6.Hong YJ, et al. The usefulness of delayed contrast-enhanced cardiovascular magnetic resonance imaging in differentiating cardiac tumors from thrombi in stroke patients. Int J Cardiovasc Imaging. 2011;27(Suppl 1):89–95. doi: 10.1007/s10554-011-9961-8. [DOI] [PubMed] [Google Scholar]
- 7.Taylor AJ, et al. ACCF/SCCT/ACR/AHA/ASE/ASNC/NASCI/SCAI/SCMR 2010 appropriate use criteria for cardiac computed tomography. A report of the American College of Cardiology Foundation Appropriate Use Criteria Task Force, the Society of Cardiovascular Computed Tomography, the American College of Radiology, the American Heart Association, the American Society of Echocardiography, the American Society of Nuclear Cardiology, the North American Society for Cardiovascular Imaging, the Society for Cardiovascular Angiography and Interventions, and the Society for Cardiovascular Magnetic Resonance. J Am Coll Cardiol. 2010;56:1864–1894. doi: 10.1016/j.jacc.2010.07.005. [DOI] [PubMed] [Google Scholar]
- 8.Hong YJ, et al. Dual-energy cardiac computed tomography for differentiating cardiac myxoma from thrombus. Int J Cardiovasc Imaging. 2014;30(Suppl 2):121–128. doi: 10.1007/s10554-014-0490-0. [DOI] [PubMed] [Google Scholar]
- 9.Johnson TR, et al. Material differentiation by dual energy CT: initial experience. Eur Radiol. 2007;17:1510–1517. doi: 10.1007/s00330-006-0517-6. [DOI] [PubMed] [Google Scholar]
- 10.Esposito A, et al. CMR in the assessment of cardiac masses: primary malignant tumors. JACC Cardiovasc Imaging. 2014;7:1057–1061. doi: 10.1016/j.jcmg.2014.08.002. [DOI] [PubMed] [Google Scholar]
- 11.Esposito A, et al. CMR in assessment of cardiac masses: primary benign tumors. JACC Cardiovasc Imaging. 2014;7:733–736. doi: 10.1016/j.jcmg.2013.11.008. [DOI] [PubMed] [Google Scholar]
- 12.Weinsaft JW, et al. Contrast-enhanced anatomic imaging as compared to contrast-enhanced tissue characterization for detection of left ventricular thrombus. JACC Cardiovasc Imaging. 2009;2:969–979. doi: 10.1016/j.jcmg.2009.03.017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Weinsaft JW, et al. Detection of left ventricular thrombus by delayed-enhancement cardiovascular magnetic resonance prevalence and markers in patients with systolic dysfunction. J Am Coll Cardiol. 2008;52:148–157. doi: 10.1016/j.jacc.2008.03.041. [DOI] [PubMed] [Google Scholar]
- 14.Paydarfar D, et al. In vivo magnetic resonance imaging and surgical histopathology of intracardiac masses: distinct features of subacute thrombi. Cardiology. 2001;95:40–47. doi: 10.1159/000047342. [DOI] [PubMed] [Google Scholar]
- 15.Gomes AS, Lois JF, Child JS, Brown K, Batra P. Cardiac tumors and thrombus: evaluation with MR imaging. AJR Am J Roentgenol. 1987;149:895–899. doi: 10.2214/ajr.149.5.895. [DOI] [PubMed] [Google Scholar]
- 16.Kimura H, et al. Fluctuations in red cell flux in tumor microvessels can lead to transient hypoxia and reoxygenation in tumor parenchyma. Cancer Res. 1996;56:5522–5528. [PubMed] [Google Scholar]
- 17.Hakime A, et al. CT perfusion for determination of pharmacologically mediated blood flow changes in an animal tumor model. Radiology. 2007;243:712–719. doi: 10.1148/radiol.2433052048. [DOI] [PubMed] [Google Scholar]
- 18.Hur J, et al. Cardioembolic stroke: dual-energy cardiac CT for differentiation of left atrial appendage thrombus and circulatory stasis. Radiology. 2012;263:688–695. doi: 10.1148/radiol.12111691. [DOI] [PubMed] [Google Scholar]
- 19.Hur J, et al. Dual-enhanced cardiac CT for detection of left atrial appendage thrombus in patients with stroke: a prospective comparison study with transesophageal echocardiography. Stroke. 2011;42:2471–2477. doi: 10.1161/STROKEAHA.110.611293. [DOI] [PubMed] [Google Scholar]
- 20.Hanley JA, McNeil BJ. A method of comparing the areas under receiver operating characteristic curves derived from the same cases. Radiology. 1983;148:839–843. doi: 10.1148/radiology.148.3.6878708. [DOI] [PubMed] [Google Scholar]
- 21.Hamlett A, Ryan L, Serrano-Trespalacios P, Wolfinger R. Mixed models for assessing correlation in the presence of replication. J Air Waste Manag Assoc. 2003;53:442–450. doi: 10.1080/10473289.2003.10466174. [DOI] [PubMed] [Google Scholar]
- 22.Gallas BD. One-shot estimate of MRMC variance: AUC. Acad Radiol. 2006;13:353–362. doi: 10.1016/j.acra.2005.11.030. [DOI] [PubMed] [Google Scholar]
- 23.DeLong ER, DeLong DM, Clarke-Pearson DL. Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics. 1988;44:837–845. doi: 10.2307/2531595. [DOI] [PubMed] [Google Scholar]
- 24.Zhou XH, Gatsonis CA. A simple method for comparing correlated ROC curves using incomplete data. Stat Med. 1996;15:1687–1693. doi: 10.1002/(SICI)1097-0258(19960815)15:15<1687::AID-SIM324>3.0.CO;2-S. [DOI] [PubMed] [Google Scholar]
- 25.Youden WJ. Index for rating diagnostic tests. Cancer. 1950;3:32–35. doi: 10.1002/1097-0142(1950)3:1<32::AID-CNCR2820030106>3.0.CO;2-3. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
All data generated or analyzed during this study are included in this published article.