Dose-compatible grating-based phase-contrast mammography on mastectomy specimens using a compact synchrotron source

Elena Eggl 1,2,, Susanne Grandl 3, Anikό Sztrόkay-Gaul 3, Martin Dierolf 1,2, Christoph Jud 1,2, Lisa Heck 1,2, Karin Burger 1,2,7, Benedikt Günther 1,2,4, Klaus Achterhold 1,2, Doris Mayr 5, Jan J Wilkens 1,2,7, Sigrid D Auweter 3, Bernhard Gleich 2, Karin Hellerhoff 3, Maximilian F Reiser 3, Franz Pfeiffer 1,2,6, Julia Herzen 1,2
PMCID: PMC6200806  PMID: 30356116

Abstract

With the introduction of screening mammography, the mortality rate of breast cancer has been reduced throughout the last decades. However, many women undergo unnecessary subsequent examinations due to inconclusive diagnoses from mammography. Two pathways appear especially promising to reduce the number of false-positive diagnoses. In a clinical study, mammography using synchrotron radiation was able to clarify the diagnosis in the majority of inconclusive cases. The second highly valued approach focuses on the application of phase-sensitive techniques such as grating-based phase-contrast and dark-field imaging. Feasibility studies have demonstrated a promising enhancement of diagnostic content, but suffer from dose concerns. Here we present dose-compatible grating-based phase-contrast and dark-field images as well as conventional absorption images acquired with monochromatic x-rays from a compact synchrotron source based on inverse Compton scattering. Images of freshly dissected mastectomy specimens show improved diagnostic content over ex-vivo clinical mammography images at lower or equal dose. We demonstrate increased contrast-to-noise ratio for monochromatic over clinical images for a well-defined phantom. Compact synchrotron sources could potentially serve as a clinical second level examination.

Introduction

Mammography is an invaluable clinical tool for the early detection of breast cancer. However, the inherently low contrast in absorption x-ray imaging for soft tissue compromises the diagnostic performance, especially in the case of dense breasts. The radiation sensitivity of breast tissue limits the allowed dose1. Recent literature reports a sensitivity between 69% and 94% and a specificity between 78% and 95% for digital mammography, depending on patient age and breast density2. The positive predictive value is between 7% and 13% in screening examinations3. These numbers explain why the field of mammography is vastly under research and a non-invasive second-level examination besides ultrasonography is needed to clarify questionable or suspicious findings and avoid unnecessary invasive procedures such as biopsies.

A strongly investigated topic is mammography with synchrotron radiation. The x-ray energy can be optimized for the best ratio of contrast to dose, thus eliminating the x-ray photons that do not give sufficient contrast (above the ideal energy) or those that deposit too much dose (below the ideal energy). A large clinical study at the SYRMEP beamline of the synchrotron Elettra (Trieste, Italy) has aimed at the patient group with inconclusive diagnosis after mammography and ultrasonography. The combination of absorption contrast and edge-enhancement resulting from phase shifts and subsequent free-space propagation enabled higher relative visibilities of abnormalities and the number of true-negative findings was significantly increased, while the dose level was lower or comparable to conventional mammography46. However, mammography with synchrotron radiation has the disadvantages of limited availability, remoteness with respect to clinics, and high costs.

A different approach to access phase information is x-ray imaging with a grating interferometer, which simultaneously provides attenuation, differential phase and dark-field signals7,8. Several studies have already successfully shown the application of phase-contrast imaging for mammography911 and especially for the classification of microcalcifications12. The dark-field signal is related to small-angle scattering and has been shown to improve the visibility of microcalcifications13 or could even help to distinguish between different types of calcifications14. Various studies carried out with dissected cancerous breasts at conventional x-ray tubes have been conducted, showing significantly higher diagnostic content provided by the multimodal images1517. However, initial investigations of this technique were performed with applied radiation doses far above radiological guidelines. Only recently, dose-compatible phase-contrast mammography images were presented18. In addition, the feasibility of installing a grating interferometer into a conventional digital mammography unit was successfully demonstrated19,20. Challenges are the low sensitivity of the interferometer due to the limited distance between the gratings, and dose compatibility.

In order to combine the advantages and avoid some disadvantages of the two approaches, we suggest to perform grating-based multimodal mammography at a compact synchrotron source based on inverse Compton scattering. These recently developed x-ray sources are compact enough to fit into regular laboratories21. They offer a monochromatic x-ray beam that is tunable in energy and partially coherent, allowing to obtain increased diagnostic content from grating-based phase-contrast and dark-field images while achieving equal image quality at lower dose compared to the aforementioned laboratory setups. In addition, monochromatic conventional absorption-based imaging or propagation-based phase-contrast imaging as done at SYRMEP is possible. Since the spatial and financial requirements of inverse Compton sources are relatively small compared to large-scale synchrotrons, the installation at medical centers appears feasible. The Munich Compact Light Source (MuCLS, Lyncean Technologies, USA) is the first commercial installation of an inverse Compton compact synchrotron source22.

Here, we present both absorption-only and grating-based multimodal images of freshly dissected cancerous mastectomy specimens acquired at the MuCLS at lower or equal dose compared to state-of-the-art clinical images. Furthermore, a dose study and analysis of the contrast-to-noise-ratio on a well defined sample, a mammographic accreditation phantom, was conducted. In addition, we present a comparison of the spatial resolution for experimental and clinical images. We believe that compact synchrotron sources like the MuCLS have great potential to bring benefits to clinical imaging, in particular for mammography, but also other fields, like coronary angiography23, could profit from the monochromatic, tunable x-ray beam.

Results

Four freshly dissected mastectomy specimens and a mammographic accreditation phantom were investigated with ex-vivo clinical mammography (cevAC-Mx) and both conventional absorption and grating-based multimodal monochromatic experimental mammography. Abbreviations used for the different contrast modalities in the following are given in Table 1. An overview of the CNR and the spatial resolution analysis is presented in Tables 2 and 3, respectively. The tumor characteristics for each sample are summarized in Table 4. The applied mean glandular dose (MGD) and the exposure times are also given in Table 4 for each sample and imaging modality.

Table 1.

Contrast modality abbreviations.

mAC-Mx monochromatic absorption-contrast mammography
mgbAC-Mx monochromatic grating-based absorption-contrast mammography
mgbDPC-Mx monochromatic grating-based differential phase-contrast mammography
mgbDFC-Mx monochromatic grating-based dark-field-contrast mammography
cevAC-Mx clinical ex-vivo absorption-contrast mammography
civAC-Mx clinical in-vivo absorption-contrast mammography

Table 2.

CNR calculated for dose study with mammographic accreditation phantom.

Modality MGD
[mGy]
Fibers Calcifications Tumor Masses
1 4 1 3 1 5
cevAC-Mx 2.0 2.51 1.99 36.84 13.89 5.92 0.53
mAC-Mx 1.0 3.16 0.11 30.86 11.94 8.64 1.50
mAC-Mx 1.6 3.51 1.61 38.47 15.60 10.70 2.73
mAC-Mx 2.0 4.71 1.39 44.11 16.85 12.19 2.67
mgbAC-Mx 1.8 4.46 0.15 25.32 13.25 7.14 0.89
mgbDFC-Mx 1.8 0.65 0.16 6.50 10.42 15.00 9.59
mgbAC-Mx 0.7 2.17 0.09 13.87 5.93 2.73 0.30
mgbDFC-Mx 0.7 1.25 2.55 3.59 7.77 11.84 5.58

Table 3.

Resolution calculated from power spectrum analysis.

Sample cevAC-Mx
[LP/mm]
mAC-Mx
[LP/mm]
mgbAC-Mx
[LP/mm]
mgbDFC-Mx
[LP/mm]
Phantom 3.70 ± 0.26 3.81 ± 0.11 3.55 ± 0.26 3.90 ± 0.74
I 2.52 ± 0.68 2.77 ± 0.41 3.03 ± 0.77 3.87 ± 0.70
II 1.42 ± 0.17 3.44 ± 0.16 5.30 ± 0.53
III 3.23 ± 0.68 3.37 ± 0.61 3.49 ± 0.07 3.56 ± 0.61
VI 2.94 ± 1.43 3.49 ± 0.66

Table 4.

Parameters and information for each specimen.

Specimen I II II IV Phantom
Tumor characteristics as verified by histopathology
Histological diagnosis multicentric lobular invasive carcinoma, G2, and lobular carcinoma in situ recurrent, invasive carcinoma of no specific type (NST, formerly invasive ductal), G3 invasive carcinoma of no specific type (NST, formerly invasive ductal), G2, and adjacent intraductal carcinoma (DCIS) bifocal invasive carcinoma of no specific type (NST, formerly invasive ductal), G1, and intraductal carcinoma (DCIS)
Max. tumor diameter 51 mm 47 mm 5 mm 25 mm
Sample position for ex-vivo mammography
Orientation AP AP AP CC
Compressed thickness [cm] 4.0 6.0 4.5 6.0 4.5
Clinical acquisition parameters
X-ray tube settings 27 kVp (W/Rh) 35 kVp (W/Ag) 30 kVp (W/Rh) 39 kVp (W/Ag) 28 kVp (W/Rh)
72 mAs 92 mAs 100 mAs 23 mAs 200 mAs
MGD civAC-Mx [mGy] 1.3 3.8 2.9 1.3
MGD cevAC-Mx [mGy] 0.9 2.2 1.4 1.1 2.0
Acquisition parameters MuCLS
Energy 25 keV 25 keV 25 keV 25 keV 25 keV
MGD mAC-Mx [mGy] 0.3 0.4 1.0–2.0
Exposure times [s] 3 3 10–20
MGD mgb-Mx [mGy] 0.8 0.9 0.9 0.7–1.8
Exposure times [s] 7 14 11 7–18
Stitching 4 × 4 5 × 5 5 × 5 4 × 4 2 × 2

Improved delineation of tumor lesions in phase-contrast image

Specimen I presented with a palpable mass in the right breast and skin retraction. Conventional in-vivo mammography (civAC-Mx) showed an asymmetry in the respective region (Fig. 1(e)). Ultrasound revealed inhomogeneities and several hypoechogenic lesions. Additional dynamic MRI showed an extensive infiltration with tumor branches extending close to the pectoralis muscle. Histopathology of the mastectomy specimen revealed a multicentric, invasive lobular carcinoma (G2) with extensive manifestations of a lobular carcinoma in situ.

Figure 1.

Figure 1

Clinical mammography and monochromatic grating-based multimodal mammography for specimen I. (a) Monochromatic grating-based absorption-contrast (mgbAC-Mx), (b) differential phase-contrast (mgbDPC-Mx), and (c) dark-field contrast (mgbDFC-Mx) mammography. (d) Clinical ex-vivo absorption-contrast mammography (cevAC-Mx) in anteroposterior position. (e) Clinical in-vivo absorption-contrast mammography (civAC-Mx) of specimen I in cranio-caudal position. Tumorous lesions are indicated by red arrows, the mamilla is indicated by a light blue arrow. All images were scaled for maximum detail visibility. (f) Histopathology of the mastectomy sample showing the existence of tumorous lesions (black arrows).

Figure 1(a–d) show monochromatic grating-based absorption-contrast (mgbAC-Mx), differential phase-contrast (mgbDPC-Mx), dark-field contrast (mgbDFC-Mx) and ex-vivo clinical absorption-contrast (cevAC-Mx) mammography images of the mastectomy specimen from specimen I, measured in anteroposterior (AP) orientation. The mgbAC-Mx (a) provides improved delineation of tumor lesions (marked by red arrows) over the cevAC-Mx (d) and civAC-Mx (e) images. The mgbDPC-Mx (b) clearly shows fine tumor branches originating from the tumor and perfusing the surrounding tissue to both sides of the carcinoma. To a reduced extent, the tumor branches are also visible in mgbDFC-Mx (c). The histopathologic analysis in hematoxylin-eosin (H&E) staining proved the existence of the tumor spiculae originating from the main tumor (black arrows, (e)).

Specimen II had a history of breast cancer. While aftercare mammography and ultrasound showed a slight, ill-defined density in the scar area but no circumscribed lesion, dynamic MRI of the right breast revealed an extensive carcinoma. Histology of the mastectomy sample showed nodular infiltrations of a poorly differentiated (G3) invasive breast cancer of no special type.

Figure 2(a–c) show cevAC-Mx, mgbAC-Mx, and mgbDPC-Mx measured in an AP orientation of the specimen. While the differentiation of tumor and scar tissue is difficult in the AC images, the mgbDPC-Mx (c) depicts the tumor spiculae (red arrows) as verified by histopathology (d). The H&E stained histology image shows the infiltration of the scar tissue (pink) with tumorous cells (purple).

Figure 2.

Figure 2

Clinical mammography and monochromatic grating-based absorption-contrast and differential phase-contrast mammography for specimen II. (a) Clinical ex-vivo absorption-contrast mammography (cevAC-Mx). (b) Monochromatic grating-based absorption-contrast (mgbAC-Mx) and (c) monochromatic grating-based differential phase-contrast (mgbDPC-Mx) mammography. Red arrows indicate the tumorous region with spiculae visible especially in the mgbDPC-Mx (c). The cyan arrow points to the mamilla. All images were scaled for maximum detail visibility. (d) Histopathology of the mastectomy sample showing the scar tissue (pink) being infiltrated by tumor cells (purple).

Detection of microcalcifications at reduced dose

In Specimen III, civAC-Mx revealed a cluster of microcalcifications in the retromamillary area. Ultrasound found a hypoechoic lesion and MRI showed a corresponding suspicious mass with early contrast enhancement. Histology of the mastectomy sample revealed an invasive carcinoma of no special type, G2, and an adjacent intraductal carcinoma (DCIS), with comedonecrosis and extensive microcalcifications.

Figure 3(a–e) display cevAC-Mx, mAC-Mx, mgbAC-Mx, mgbDPC-Mx and mgbDFC-Mx measured in AP orientation. A calcification cluster that had preoperatively been clip-marked (magenta arrows) is shown in magnification in the small inlets. The mAC-Mx image (b) equally shows the extent of the microcalcification cluster compared to cevAC-Mx (a), at significantly lower dose (0.3 vs. 1.4 mGy). The existence of the microcalcification cluster was verified by histopathology (f). The invasive carcinoma revealed by histopathology could not clearly be identified by any of the imaging modalities, including civAC-Mx.

Figure 3.

Figure 3

Clinical mammography and monochromatic absorption-contrast and grating-based multimodal mammography for specimen III. (a) Clinical ex-vivo absorption-contrast mammography (cevAC-Mx). (b) Monochromatic absorption-contrast mammography (mAC-Mx). (c) Monochromatic grating-based absorption-contrast mammography (mgbAC-Mx), (d) dark-field mammography (mgbDFC-Mx) and (e) differential phase-contrast mammography (mgbDPC-Mx). Inlets show a calcification cluster that had previously been marked. The clip marker is highlighted with a magenta arrow, a light blue arrow indicates the mamilla. All images were scaled for maximum detail visibility. (f) Histopathology of the mastectomy sample showing microcalcifications.

Specimen IV was diagnosed with three microcalcification clusters in civAC-Mx, two of which were partially extracted in a biopsy. Histology of the mastectomy sample revealed a bifocal, well differentiated (G1) breast cancer of no special type, with an extensive intraductal carcinoma (DCIS) and microcalcifications.

Figure 4(a,b) present cevAC-Mx and mAC-Mx measured in a cranio-caudal position. Both images reveal a microcalcification cluster to equal extent, with significantly lower dose for the mAC-Mx (0.4 vs. 1.1 mGy). The histopathologic work-up (c) confirms the existence of microcalcifications. The bifocal carcinoma revealed by histopathology is obscured by the surgical margins of the mastectomy sample in images (a) and (b).

Figure 4.

Figure 4

Clinical mammography and monochromatic absorption-contrast mammography for specimen IV. (a) Clinical mammography (cevAC-Mx). (b) Monochromatic absorption-contrast mammography (mAC-Mx). Inlets show a calcification cluster. All images were scaled for maximum detail visibility. (c) Histopathology of the mastectomy specimen showing extensive microcalcifications (black arrows).

Dose-study with a mammographic accreditation phantom

In order to quantitatively compare monochromatic MuCLS mammography with a clinical mammography unit, a contrast-to-noise-ratio (CNR) analysis was performed for a mammographic accreditation phantom (Gammex, Model 156, www.sunnuclear.com/solutions/diagnostic/mammography/156phantom). The image of the accreditation phantom acquired by the conventional mammography system by automated exposure control is compared to both monochromatic absorption-contrast images and also grating-based trimodal images acquired at different mean glandular doses. A CNR analysis for different test objects is presented in Table 2. A selection of images is shown in Fig. 5.

Figure 5.

Figure 5

Dose study for the mammographic accreditation phantom. ACR guidelines require for a minimum of four fibrils, three groups of microcalcifications, and three tumor masses to be resolved. (a) Clinical mammography cevAC-Mx acquired at 2.0 mGy mean glandular dose (MGD). (b,c) Monochromatic absorption-contrast mammography (mAC-Mx) acquired at 2.0 mGy MGD (b) and 1.6 mGy MGD (c). (df) Monochromatic grating-based absorption-contrast (mbgAC-Mx) (d), differential phase-contrast (mgbDPC-Mx) (e) and dark-field-contrast (mgbDFC-Mx) (f) mammography acquired at 1.8 mGy MGD. (gi) mgbAC-Mx (g), mgbDPC-Mx (h), mgbDFC-Mx (i) acquired at 0.7 mGy MGD. All images were scaled for maximum detail visibility.

It is clearly visible that the absorption images cevAC-Mx (Fig. 5(a)), mAC-Mx (Fig. 5(b,c)), and mgbAC-Mx (Fig. 5(d)) fulfill the standard criteria of clinical image quality set by the ACR24, which require a minimum of 4 fibrils, 3 groups of microcalcifications, and 3 tumor masses to be resolved. A quantitative analysis (Table 2) demonstrates that the CNR in the mAC-Mx (b) image considerably exceeds the CNR in the cevAC-Mx (a) image at equal dose. The mAC-Mx (c) still slightly outperforms the cevAC-Mx image (a) at 20% reduced dose.

For the grating-based images, the mgbAC-Mx (Fig. 5(d)) image at slightly lower dose than cevAC-Mx (a) provides a comparable CNR. Especially for the small tumor masses, the mgbDFC-Mx (f) CNR by far outperforms the clinical image (a), where the smallest tumor mass cannot be distinguished from the background (CNR < 1). This is still the case when the dose is reduced to 35% of the clinical value (i). In addition, the mgbDPC-Mx (e), for which a quantitative analysis is not possible due to the differential nature of the signal, allows to depict all six tumor fibrils. The mgbDFC-Mx (f,i) reveals additional structures within the two largest tumor masses of the accreditation phantom.

Importantly, while the low-dose mgbAC-Mx (g) does not fulfill the ACR criteria, the combination with mgbDPC-Mx (h) and mgbDFC-Mx (i) meets or, in the case of the tumor masses, even exceeds the ACR criteria.

Resolution

A power spectrum analysis was performed for all absorption and dark-field images cevAC-Mx, mAC-Mx, mgbAC-Mx and mgbDFC-Mx (not for DPC-Mx due to the differential nature of the signal) in order to compare their resolution25. As illustrated in Table 3, the resolution is mainly in the range between 3 and 4 linepairs (LP) per mm. The analysis shows that a higher resolution was achieved in the mAC-, mgbAC- and mgbDFC-MX than in the cevAC-Mx for all investigated samples, at lower or equal dose of the monochromatic images.

Discussion

We presented conventional absorption-contrast as well as grating-based multimodal mammography images of freshly dissected mastectomy samples acquired with a monochromatic x-ray beam at a compact synchrotron source. Comparison with clinical ex-vivo mammography images showed equal diagnostic quality at lower dose or superior diagnostic quality at equal dose of the monochromatic images. Tumorous lesions could be identified significantly better in the phase-contrast modality as verified by histopathology (Specimens I, II). Microcalcification clusters were revealed to an equal extent as in the clinical mammography image, at a significantly reduced dose for the monochromatic images (Specimens III, IV). A power spectrum analysis showed that the monochromatic absorption-contrast and dark-field-contrast images (mAC-Mx, mgbAC-Mx, mgbDFC-Mx) could achieve better resolution than the clinical image (cevAC-Mx) at lower or equal dose, proving that the small source size of the MuCLS is beneficial for the resolution.

A dose study conducted with a mammographic accreditation phantom showed a superior CNR of various test objects at equal mean glandular dose and a higher or comparable CNR at 20% reduced dose for the monochromatic mammography images compared to the clinical one. The grating-based phase-contrast and dark-field image yield an enhanced diagnostic content for the fibers and tumor masses, respectively. The dark-field image outperforms the clinical image for the tumor masses even at a 65% lower dose.

The presented results demonstrate that mammography at a compact synchrotron source like the MuCLS has great potential to improve the diagnostic quality of mammography. The compact size and limited financial requirements compared to large-scale synchrotron source allow for envisioning the installation at hospitals in the future. Thus, the implementation of a compact synchrotron could enable a second-level examination in the case of inconclusive diagnosis (as performed at the SYRMEP beamline46), in addition with the possibility to choose between conventional, but monochromatic absorption imaging and grating-based trimodal imaging.

The main limitation of our study was the different orientation (anteroposterior instead of cranio-caudal or mediolateral-oblique) for three of the four specimens compared to clinical in-vivo mammography. The different orientation was necessary in order to avoid artifacts originating from surgical resection margins. Therefore, the ex-vivo images are not fully comparable to in-vivo images concerning orientation and sample thickness.

The limited sizes of beam and gratings required scanning of the samples, resulting in an increased scan time incompatible with clinical applications. Only recently, the feasibility of manufacturing stitched gratings in order to cover larger fields of view was demonstrated26. The field of compact synchrotron sources is constantly developing22,27,28 and considerable improvements and further developments of the technology with regard to increased stability, flux and beam size can be expected. An increased field of view can be achieved by employing a larger exit aperture. The currently available 4 mrad opening angle is only limited by the diameter of the exit aperture. By simply replacing the exit aperture by one with a larger diameter, the beam size at the experimental station would increase. In addition, other potential clinical applications will profit especially from an extension of the energy range.

Concerning the mean glandular dose applied in mammography, our results show that dose-compatible grating-based mammography is feasible at a monochromatic source. The used setup even has the potential for further dose reduction by thinning the grating wafers from the currently used 500 μm thick Si substrate down to 200 μm and by modifying the setup to allow positioning of the sample to behind the phase grating.

In conclusion, we believe that monochromatic x-rays provided by compact synchrotron sources like the MuCLS, the first commercially installed source of this type, can enable significant improvements in diagnostic x-ray imaging. Mammography could benefit from enhanced diagnostic image content and improved resolution at a reduced dose. In addition, as the x-ray energy of these sources is tunable, also other applications, like for instance coronary angiography23, could benefit from the implementation of compact synchrotron sources in a clinical setting.

Methods

Study protocol

The study was conducted in accordance with the Declaration of Helsinki and was approved by the local ethics committee (Ethikkommission of the Ludwig-Maximilian-University, Munich, project number 240-10, date of permission 26/08/2010, amendment 30/05/2012). Inclusion criteria were a histologically proven breast cancer in preoperative core biopsy with a recommendation for mastectomy according to gynecological guidelines or the patient’s wish for mastectomy. Participants gave written informed consent before participation after adequate explanation of the study protocol. Indication for breast surgery followed recommendation of the interdisciplinary tumor board of the University of Munich breast center.

Preoperative diagnostics

Preoperative diagnostics included clinical breast examination, standard two-view digital mammography in cranio-caudal (CC) and mediolateral-oblique (MLO) projections (Hologic Selenia Dimensions, Bedford, USA) using a standard breast compression paddle and high resolution B-mode ultrasound (standard linear transducer 13.5 MHz, Siemens Acuson Antares, Siemens Healthcare, Germany). Additional MRI was performed in three cases by using a dedicated sensitivity-encoding-enabled bilateral breast coil with a 1.5- or 3.0-Tesla system.

Ex-vivo mammography

The mastectomy samples were intraoperatively marked with surgical sutures for 3D orientation. The samples were fixed within a metal-framed sample holder to afford adequate breast compression. The position was cranio-caudal, or anteroposterior in case a cranio-caudal position was impossible due to the shape of the dissected specimen. A digital ex-vivo mammography image was acquired at a clinical mammography unit (Hologic Selenia Dimensions) with a pixel size of 70 × 70 μm2.

Working principle of the MuCLS

The MuCLS is a compact synchrotron based on inverse Compton scattering and was developed and manufactured by Lyncean Technologies Inc., USA. As a relativistic electron bunch collides with a laser pulse, x-ray photons are produced. In good approximation for head-on collision and backscattering, the energy Ex of the emitted x-rays is given by Ex = 4γ2EL, where γ = Ee/E0 with E0 the rest energy and Ee the total energy of the electron, and EL is the energy of the laser photons. The electrons circulate in a miniature storage ring and collide with the laser pulse stored in a bow-tie enhancement cavity upon each revolution (cf. Fig. 6)21,22,29.

Figure 6.

Figure 6

Schematic drawing of the Munich Compact Light Source (MuCLS) and the experimental setup. X-rays are generated in the process of inverse Compton scattering in a laser-electron storage ring design (the linear accelerator section is not drawn here). The grating interferometer is located approximately 16 m from the interaction point. A two-grating interferometer in the first Talbot order was used. The freshly dissected breast specimen is placed in a dedicated sample holder for reasonable sample compression.

The produced x-rays are collimated to a 4 mrad cone. The x-ray beam is monochromatic, partially coherent, and the energy is tunable between 15 and 35 keV by adjusting the electron energy. The MuCLS currently offers - on a daily basis - a flux exceeding 2.4 · 1010 photons per second (at 35 keV), and a source size of less than 50 × 50 μm2. At a distance of about 16 m from the interaction point, the beam has an elliptic shape of 62 mm × 74 mm22.

Mammography at the MuCLS

Maintaining the position of the breast specimen within the sample holder in order to ensure comparability, monochromatic mammography images were acquired at the MuCLS. A Talbot interferometer was set up approximately 16 m from the interaction point of the MuCLS, as shown schematically in Fig. 6. The inter-grating distance was ~25 cm, the grating periods were p1 = 4.9 μm and p2 = 5.0 μm for phase and analyzer grating, respectively. The signals were retrieved from phase stepping as described in8. Conventional absorption-contrast images were acquired without the grating interferometer. A Dexela 1512 flatpanel detector (PerkinElmer Inc., USA) with a Gd2O2S scintillator was used, yielding an effective pixel size of 71 × 71 μm2.

Scanning the sample and afterwards stitching of several images was required to cover the whole sample, using linear ramps to blend the overlapping areas. Table 4 summarizes the acquisition parameters for each scan. The exposure times in Table 4 are given for one Stitching-scan. The MuCLS x-ray energy was tuned to 25 keV for all samples. The interferometer visibility was approximately 45–50% for all measurements.

Dose calculation

The MGD automatically registered by the clinical mammography unit for clinical in vivo- and ex vivo-images is given in Table 4 for each sample. It is calculated based on the x-ray tube settings and the compressed sample thickness. The MGD value for the cevAC-Mx images was corrected when an incorrect sample thickness was assumed due to the shape of the sample holder. A correction factor was calculated based on the half value layer of Aluminum for the respective acquisition settings and the assumed and correct breast thickness, from tabulated values by30,31.

For the monochromatic mammography performed at the MuCLS, the MGD was calculated using monoenergetic normalized glandular dose coefficients DgN(E) (with E the x-ray energy) as tabulated by Boone et al.32, taking into account the MuCLS spectrum22, and summing over all energy bins E:

MGD=EK(E)[mGy]0.114[RmGy](DgN(E)[mGyR]). 1

The formula for the MGD was adapted from32 to be used with air kerma K instead of the older unit exposure, as the fitted equation for exposure given by32 is incorrect33. The DgN (E) values were selected according to the compressed thickness of the dissected breasts, and assuming a 50%/50% distribution of glandular and adipose tissue. The air kerma for the MuCLS beam K per energy bin E can be calculated for known photon flux Φ at the sample position and x-ray spectrum34:

K(E)=EΦ(E)(μen/ρ)air(E), 2

where (μen/ρ)air (E) is the mass energy attenuation coefficient of air35. The photon flux per energy bin Φ(E) was calculated from a single photon counting Pilatus 200 K detector (Dectris Ltd., Switzerland), taking into account the x-ray spectrum and the efficiency of the Si sensor of the detector36. The calculation of the air kerma was validated using a soft x-ray ionization chamber (Model 34013, PTW Freiburg GmbH, Germany) and values agreed within the measurement uncertainties of ±10%. A scintillation counter was calibrated to calculated air kerma prior to each sample measurement and the scintillation counts were logged for each measurement frame, thus yielding an exact measure of air kerma for each scan.

Image Analysis

The CNR was calculated according to the definition

CNR=S1¯S2¯σBG, 3

where S1¯ and S2¯ are the average signals in two regions of interest (ROI) which should be compared, and σBG is the standard deviation within a larger ROI located in the background region.

The resolution of the images was determined by analyzing their power spectra25. The images were Fourier transformed and a Gaussian filter was applied on the squared norm. The resolution of the images is then given by the maximal spatial frequency where the spectral power of the signal equals twice the spectral power of the noise baseline, taking into account the effective pixel size of clinical and experimental images.

Acknowledgements

The authors acknowledge financial support through the DFG Cluster of Excellence Munich-Centre for Advanced Photonics (MAP, DFG EXC-158), the Center for Advanced Laser Applications (CALA), the German Federal Ministry of Education and Research (BMBF), the DFG Gottfried Wilhelm Leibniz program and the European Research Council (ERC, H2020, AdG 695045). This work was supported by the German Research Foundation (DFG) and the Technical University of Munich within the funding programme Open Access Publishing. This work was carried out with the support of the Karlsruhe Nano Micro Facility (KNMF, www.knmf.kit.edu), a Helmholtz Research Infrastructure at Karlsruhe Institute of Technology (KIT). The authors would like to thank the staff of Lyncean Technologies Inc. for their technical support.

Author Contributions

E.E., J.H., S.G., A.S., F.P., S.D.A. and K.H. conceived the experiment. E.E., S.G., C.J., M.D., B.Gu. and A.S. performed the experiment. E.E., S.G., A.S. and L.H. analyzed the data. S.G., A.S., S.D.A., K.H. and M.F.R. contributed the samples. D.M. provided the histopathology. E.E., K.B. and J.W. performed the dose calculation. M.D., K.A., B.G. and F.P. provided and prepared the experimental setup. E.E. and S.G. wrote the manuscript, with significant input from all coauthors.

Competing Interests

The authors declare no competing interests.

Footnotes

Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Perry N, et al. European guidelines for quality assurance in breast cancer screening and diagnosis. Ann Oncol. 2008;19:614–622. doi: 10.1093/annonc/mdm481. [DOI] [PubMed] [Google Scholar]
  • 2.Stout NK, et al. Benefits, harms, and costs for breast cancer screening after US implementation of digital mammography. J. Natl. Cancer Inst. 2014;106:dju092. doi: 10.1093/jnci/dju092. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Venkatesan A, Chu P, Kerlikowske K, Sickles EA, Smith-Bindman R. Positive predictive value of specific mammographic findings according to reader and patient variables. Radiology. 2009;250:648–657. doi: 10.1148/radiol.2503080541. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Castelli E, et al. Mammography with Synchrotron Radiation: First Clinical Experience with Phase-Detection Technique. Radiology. 2011;259:684–694. doi: 10.1148/radiol.11100745. [DOI] [PubMed] [Google Scholar]
  • 5.Longo R, et al. Clinical study in phase- contrast mammography: image-quality analysis. Philos. Trans. Roy. Soc., A. 2014;372:20130025. doi: 10.1098/rsta.2013.0025. [DOI] [PubMed] [Google Scholar]
  • 6.Olivo A, Castelli E. X-ray phase contrast imaging: From synchrotrons to conventional sources. Riv. Nuovo Cimento Soc. Ital. Fis. 2014;37:467–508. [Google Scholar]
  • 7.Pfeiffer F, Weitkamp T, Bunk O, David C. Phase retrieval and differential phase-contrast imaging with low-brilliance X-ray sources. Nat. Phys. 2006;2:258–261. doi: 10.1038/nphys265. [DOI] [Google Scholar]
  • 8.Pfeiffer F, et al. Hard-X-ray dark-field imaging using a grating interferometer. Nat. Mat. 2008;7:134–137. doi: 10.1038/nmat2096. [DOI] [PubMed] [Google Scholar]
  • 9.Stampanoni M, et al. Toward clinical differential phase contrast mammography: preliminary evaluations and image processing schemes. J. Instrum. 2013;8:C05009. doi: 10.1088/1748-0221/8/05/C05009. [DOI] [Google Scholar]
  • 10. Olivo, A. et al. Low-dose phase contrast mammography with conventional x-ray sources 40, 090701 (2013). [DOI] [PubMed]
  • 11.Tanaka Toyohiko, Honda Chika, Matsuo Satoru, Noma Kazuo, Oohara Hiromu, Nitta Norihisa, Ota Shinichi, Tsuchiya Keiko, Sakashita Yoko, Yamada Aya, Yamasaki Michio, Furukawa Akira, Takahashi Masashi, Murata Kiyoshi. The First Trial of Phase Contrast Imaging for Digital Full-Field Mammography Using a Practical Molybdenum X-Ray Tube. Investigative Radiology. 2005;40(7):385–396. doi: 10.1097/01.rli.0000165575.43381.48. [DOI] [PubMed] [Google Scholar]
  • 12.Wang Z, et al. Non-invasive classification of microcalcifications with phase-contrast x-ray mammography. Nature communications. 2014;5:3797. doi: 10.1038/ncomms4797. [DOI] [PubMed] [Google Scholar]
  • 13.Michel T, et al. On a dark-field signal generated by micrometer-sized calcifications in phase-contrast mammography. Phys. Med. Biol. 2013;58:2713–2732. doi: 10.1088/0031-9155/58/8/2713. [DOI] [PubMed] [Google Scholar]
  • 14.Scherer K, et al. Improved Diagnostics by Assessing the Micromorphology of Breast Calcifications via X-Ray Dark-Field Radiography. Sci. Rep. 2016;6:36991. doi: 10.1038/srep36991. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Hauser N, et al. A study on mastectomy samples to evaluate breast imaging quality and potential clinical relevance of differential phase contrast mammography. Invest. Radiol. 2014;49:131–137. doi: 10.1097/RLI.0000000000000001. [DOI] [PubMed] [Google Scholar]
  • 16.Scherer K, et al. Bi-directional x-ray phase-contrast mammography. PLoS ONE. 2014;9:e93502. doi: 10.1371/journal.pone.0093502. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Grandl S, et al. Improved visualization of of breast cancer features in multifocal carcinoma using phase-contrast and dark-field mammography: an ex-vivo study. Eur. J. Radiol. 2015;25:3659–3668. doi: 10.1007/s00330-015-3773-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Scherer K, et al. Toward Clinically Compatible Phase-Contrast Mammography. PloS one. 2015;10:e0130776. doi: 10.1371/journal.pone.0130776. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Roessl E, Daerr H, Koehler T, Martens G, van Stevendaal U. Clinical boundary conditions for grating-based differential phase-contrast mammography. Philos. Trans. Roy. Soc., A. 2014;372:1–7. doi: 10.1098/rsta.2013.0033. [DOI] [PubMed] [Google Scholar]
  • 20.Koehler T, et al. Slit-scanning differential x-ray phase-contrast mammography: Proof-of-concept experimental studies. Med. Phys. 2015;42:1959–1965. doi: 10.1118/1.4914420. [DOI] [PubMed] [Google Scholar]
  • 21.Huang Z, Ruth RD. Laser-electron storage ring. Phys. Rev. Lett. 1998;80:976–979. doi: 10.1103/PhysRevLett.80.976. [DOI] [Google Scholar]
  • 22.Eggl E, et al. The Munich Compact Light Source: Initial performance measures. J. Synchrotron Radiat. 2016;23:1137–1142. doi: 10.1107/S160057751600967X. [DOI] [PubMed] [Google Scholar]
  • 23.Eggl E, et al. Mono-energy coronary angiography with a compact synchrotron source. Sci. Rep. 2017;7:42211. doi: 10.1038/srep42211. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Hendrick, R., Bassett, L. & Botsco, M. Mammography quality control manual. American College of Radiology, Preston (1999).
  • 25.Modregger P, Lübbert D, Schäfer P, Köhler R. Spatial resolution in Bragg-magnified X-ray images as determined by Fourier analysis. Phys. Status Solidi A. 2007;204:2746–2752. doi: 10.1002/pssa.200675685. [DOI] [Google Scholar]
  • 26.Schröter TJ, et al. Large field-of-view tiled grating structures for x-ray phase-contrast imaging. Rev. Sci. Instrum. 2017;88:029901. doi: 10.1063/1.4975395. [DOI] [PubMed] [Google Scholar]
  • 27.Variola, A. The ThomX Project. In 2nd International Particle Accelerator Conference (IPAC11), 1903–1905 (2011).
  • 28.Kuroda R, et al. Quasi-monochromatic hard X-ray source via laser Compton scattering and its application. Nucl. Instrum. Methods Phys. Res., Sect. A. 2011;637:S183–S186. doi: 10.1016/j.nima.2010.04.001. [DOI] [Google Scholar]
  • 29.Loewen, R. A Compact Light Source: Design and Technical Feasibility Study of a Laser-Electron Storage Ring X-ray Source. SLAC-Report 632, Stanford University, USA (2003).
  • 30.Dance DR. Monte Carlo calculation of conversion factors for the estimation of mean glandular breast dose. Physics in Medicine and Biology. 1990;35:1211–1219. doi: 10.1088/0031-9155/35/9/002. [DOI] [PubMed] [Google Scholar]
  • 31.Dance DR, Skinner CL, Young KC, Beckett JR, Kotre CJ. Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol. Physics in Medicine and Biology. 2000;45:3225–3240. doi: 10.1088/0031-9155/45/11/308. [DOI] [PubMed] [Google Scholar]
  • 32.Boone JM. Normalized glandular dose (DgN) coefficients for arbitrary X-ray spectra in mammography: computer-fit values of Monte Carlo derived data. Med. Phys. 2002;29:869–875. doi: 10.1118/1.1472499. [DOI] [PubMed] [Google Scholar]
  • 33.Nosratieh A, et al. Mean glandular dose coefficients (DgN) for x-ray spectra used in contemporary breast imaging systems. Phys. Med. Biol. 2015;60:7179–7190. doi: 10.1088/0031-9155/60/18/7179. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Schlegel, W. & Bille, J. Medizinische Physik 2 (Springer Verlag, 1999).
  • 35.Buhr H, Büermann L, Gerlach M, Krumrey M, Rabus H. Measurement of the mass energy-absorption coefficient of air for x-rays in the range from 3 to 60 keV. Phys. Med. Biol. 2012;57:8231–8247. doi: 10.1088/0031-9155/57/24/8231. [DOI] [PubMed] [Google Scholar]
  • 36.Donath T, et al. Characterization of the PILATUS photon-counting pixel detector for X-ray energies from 1.75 keV to 60 keV. J. Phys.: Conf. Ser. 2013;425:62001. [Google Scholar]

Articles from Scientific Reports are provided here courtesy of Nature Publishing Group