Skip to main content
Virology Journal logoLink to Virology Journal
. 2018 Oct 29;15:166. doi: 10.1186/s12985-018-1076-6

Validation of intra- and inter-laboratory reproducibility of the Xpert HPV assay according to the international guidelines for cervical cancer screening

Ajmal Akbari 1, Davy Vanden Broeck 1,2,3,4,, Ina Benoy 1,2,4, Elizaveta Padalko 2,5, Johannes Bogers 1,2,3,4, Marc Arbyn 6
PMCID: PMC6206920  PMID: 30373616

Abstract

Background

Cervical cancer screening with assays detecting DNA of high-risk human papillomavirus (hrHPV) types is more effective than cytology-based screening. This study completes the diagnostic accuracy assessment conducted previously within the framework of VALGENT-2 (Validation of HPV genotyping Tests) and aims to determine whether the reproducibility of Xpert HPV is in line with international validation criteria.

Methods

Validation of new hrHPV DNA assays requires demonstration of good reproducibility and non-inferior clinical accuracy for cervical precancer compared to a standard comparator assay. The international reproducibility criteria are: lower bound of 95% confidence interval of the intra- and inter-laboratory agreement regarding detection of high-risk HPV DNA exceeding 87% with kappa ≥0.5.

Results

The Xpert HPV assay showed high intra-laboratory reproducibility with an overall positivity/negativity agreement of 96.9% and a kappa of 0.925. Inter-laboratory testing showed an agreement of 97.8% with a kappa of 0.948.

Conclusions

The Xpert HPV assay fulfills the HPV test reproducibility criterion requirement for use in cervical cancer screening.

Keywords: Human papillomavirus, Real-time sequencing, Cervical cancer screening

Background

High level of evidence from large randomized trials currently exists which indicates that human papillomavirus (HPV) based assays are more effective than cytology screening to reduce the burden of severe cervical dysplasia and cancer [1]. Two HPV assays were used in the randomized trials, Hybrid Capture 2 (HC2) and GP5+/6+ polymerase chain reaction-enzyme immunoassay (PCR-EIA).

International guidelines have been widely adopted for the clinical validation of novel HPV tests. In a population of screened women of at least 30 years of age, the novel HPV test should demonstrate non-inferior sensitivity and specificity for cervical intraepithelial neoplasia grade 2+ (CIN2+) compared to the standard comparator tests (Hybrid Capture 2 (HC2) HPV DNA test or GP5+/6+ PCR EIA) [2]. In addition, the novel HPV test should demonstrate high intra- and inter-laboratory reproducibility with a lower confidence bound not less than 87% while maintaining a kappa value higher than or equal to 0.5.

Based on meta-analysis, a list was established of high-risk HPV (hrHPV) assays that fulfill the cross-sectional performance criteria for cervical cancer screening [3]. This list is continuously updated as soon as data on new HPV assays become available. This current study aims to assess the intra- and inter-laboratory reproducibility of the Xpert HPV assay.

Xpert HPV is a non-batch real-time PCR assay capable of detecting 14 hrHPV types (HPV16, − 18, − 31, − 33, − 35, − 39, − 45, − 51, − 52, − 56 -58, − 59, − 66, − 68) within the run time of 1 h. The assay utilizes six fluorescent channels for the detection of individual types of hrHPV, groups of hrHPV, and the human reference gene. Fluorescent channel 1 detects HPV 16; fluorescent channel 2 detects HPV 18 and − 45; fluorescent channel 3 detects HPV 31, − 33, − 35, − 52, − 58; fluorescent channel 4 detects HPV 51, − 59; fluorescent channel 5 detects HPV 39, − 56, − 66, − 68 and fluorescent channel 6 is used for Sample Adequacy Control (SAC).

The fulfillment of the international equivalency criteria of Xpert HPV regarding sensitivity and specificity to detect cervical precancer has been evaluated as part of a separate study (VALGENT) [7, 9].

Methods

VALGENT is an established comprehensive study design allowing validation of high-risk HPV DNA assays which potentially can be used for primary cervical cancer screening [2, 9]. It includes comparing the sensitivity and specificity for CIN2+ with a standard comparator test. Non inferior accuracy of the Xpert HPV assay compared to the GP5+/6+ PCR-assay was demonstrated in the second installment of VALGENT using samples collected from women attending the Scottish cervical cancer screening program [9]. The current study address the intra- and inter-reproducibility of the Xpert HPV assay using residual material of cervical specimens remnant after HPV testing with the AML qPCR and microscopic cytological processing. The cervical samples were collected in PreservCyt liquid based cytology media (ThinPrep®) (Hologic, San Diego, CA, USA) according to the manufacturers recommendations in agreement with European guidelines. Samples were collected between February 2016 and November 2016. Three aliquots of 1 ml of the original Preservcyt solution were prepared and stored at room temperature. The time span between specimen collection and preparation was maximum 3 days. The first two were tested independently with the Xpert HPV assay of the department of Cytopathology and Molecular Diagnostics at Algemeen Medisch Laboratorium (AML) with an interval of 4 weeks. A third aliquot was sent to laboratory of the Department of Microbiology and Immunology, Ghent University Hospital, Ghent, Belgium where it was tested again with the Xpert HPV assay.

The study set contained residual cervical cell material collected from 510 women attending cervical cancer screening in Belgium [10]. The composition was determined as recommended in the international guideline for validation of new hrHPV DNA tests [2], prescribing 30% of high-risk HPV positive specimen tested with a clinically validated assay. The study set contained aliquots form 357 hrHPV-negative and 153 hrHPV-positive women. HPV positivity was determined by the RIATOL real-time PCR targeting E6/7 genes of 18 HPV genotypes [4].

The RIATOL real-time qPCR allows genotyping of 18 HPV types for 14 hrHPV types (HPV 16, 18, 31, 33, 35, 39, 45, 51, 52, 56, 58, 59, 66, and 68), one probable hrHPV type (HPV 53), two low-risk types (HPV 6, 11) and one undetermined risk type (HPV 67). The AML HPV test has been clinically validated according to the international guidelines [5]. A detailed description of the laboratory developed validation test has been published previously. The RIATOL real-time qPCR generates a quantitative correlation by generating signal strength in the form of Ct-values (cycle threshold). The Ct-values are calculated by determining the point at which the fluorescence exceeds a threshold limit [6]. Ct-values are inversely proportional to the amount of target nucleic acid in a sample. Studies have shown that AML HPV test is highly sensitive for the detection of HPV genotypes [7].

The overall percentage of agreement was computed as the proportion of concordant results (positive + negative concordant test results) over all test results with corresponding binomial exact 95% confidence intervals. Cohen’s Kappa values were calculated and 95% CI as proposed by Fleiss [8]. Intra-reproducibility assessment was based on the two Xpert HPV testings in the AML laboratory. Inter-reproducibility assessment was based on the first testing in AML and the testing at the University of Gent. The reproducibility validation criterion for new hrHPV DNA assays usable in cervical screening was considered as fulfilled when the left 95% confidence interval (CI) bound for hrHPV concordance exceeded 87% and the kappa > 0.5.

The Ct-values per fluorescent channel were recorded during each run. To identify the consistency of these values during the intra- and inter-laboratory tests, the average, minimal and maximal difference in Ct-values of the samples were calculated per fluorescent channel.

Results

The results from hrHPV testing with Xpert HPV showed high intra-laboratory reproducibility with an agreement of 96.9% (95% CI, 95.0–98.2%) (494/510 samples) and a kappa of 0.925 (95% CI 0.888–961). The inter-laboratory test showed an excellent agreement of 97.8% (95% CI, 96.2–98.9%), 499/510 samples) with a kappa of 0.948 (95% CI 0.917–0.978). The results of intra- and inter-laboratory reproducibility are represented in Table 1.

Table 1.

Intra and inter laboratory reproducibility

Intra-laboratory reproducibilitya
Xpert HPV – AML Run 2 Total
Xpert HPV –AML Run 1 Negative Positive
Negative 353 4 357
Positive 12 141 153
Total 365 145 510
Inter-laboratory agreementb
Xpert HPV UZ Ghent Total
Xpert HPV –AML Run 1 Negative Positive
Negative 356 1 357
Positive 10 143 153
Total 366 144 510
Inter-laboratory agreementc
Xpert HPV UZ Ghent Total
Xpert HPV –AML Run 2 Negative Positive
Negative 154 3 357
Positive 4 149 153
Total 157 153 510

aThe overall HPV test agreement 96.9% (95% CI, 95.0–98.2%) KAPPA value of 0.925 (95% CI 0.888–0.961)

bThe overall HPV test agreement 97.8% (95% CI, 96.2–98.9%) KAPPA value of 0.948 (95% CI 0.917–0.978)

cThe overall HPV test agreement 98.6% (95% CI, 97.6–99.6%) KAPPA value of 0.967 (95% CI 0.943–0.991)

Results were evaluated per fluorescent channel. The overall agreement within the intra-laboratory runs was close to 100%, with the lower confidence bound not less than 96%. The lowest overall agreement of 98% was detected in fluorescent channel 3 (HPV 31, 33, 35, 52, and 58) between the intra-laboratory run 1 and run 2. Other runs showed that the agreement within fluorescent channel 3 was slightly lower. The kappa value of the intra-laboratory reproducibility run was close to 1.

There were inconsistencies noticed in the Ct- values annotated to the sample within the three Xpert HPV runs. Table 2 represents the summary of the samples with inconsistent Ct-values in the five fluorescent channels during the intra-and inter-laboratory tests. The average difference within the Ct-values of the five fluorescent channels was 1.1 and varied between 0.1 and 3.2. The largest difference in Ct-values (MAX 8.6, 6.7, and 8.4) can be observed in fluorescent channel 3 (HPV type 31, 33, 35, 52, and 58) for the three runs, respectively.

Table 2.

Difference in Ct-values between intra- and inter- laboratory runs

N AVG Range
MIN MAX
AML run 1 versus AML run 2
 FC 1 30 1.1 0.1 3.2
 FC 2 22 0.9 0.1 3.8
 FC 3 64 1.4 0.1 8.6
 FC 4 29 1.0 0.1 3.7
 FC 5 42 1.3 0.2 4.8
AML run 1 versus Ghent run
 FC 1 31 1.2 0.1 3.3
 FC 2 21 1.3 0.1 3.9
 FC 3 63 1.3 0.1 6.7
 FC 4 29 1.3 0.1 7.1
 FC 5 40 1.3 0.1 4.2
AML run 2 versus Ghent run
 FC 1 27 1.2 0.1 3.7
 FC 2 23 1.2 0.1 5.0
 FC 3 65 1.2 0.1 8.4
 FC 4 29 1.2 0.1 6.1
 FC 5 43 1.3 0.1 6.0

FC Fluorescent channel, N Number of samples, AVG Average, MAX maximum, MIN minimum

The overall agreement within the inter-laboratory runs was close to 100%, with the lower confidence bound not less than 97% while a slightly lower agreement was found for fluorescent channel 3. The kappa value of the inter-laboratory reproducibility run was close to 1. The reproducibility of the Xpert HPV per fluorescent channel is represented in Table 3.

Table 3.

Overall agreement, lower and upper confidence bound, and the KAPPA value of each flourescent channel per run

Agreement 95% CI Kappa Kappa 95% CI
Lower bound Upper bound Lower bound Upper bound
Intra-laboratory run 1versusIntra-laboratory run 2
 Fluorescent channel 1 99.4% 98.8% 100% 0.99 97.0% 100%
 Fluorescent channel 2 99.6% 99.1% 100% 0.99 97.8% 100%
 Fluorescent channel 3 98.0% 96.8% 99.2% 0.95 92.4% 98.2%
 Fluorescent channel 4 99.2% 98.5% 100% 0.98 96.3% 100%
 Fluorescent channel 5 98.8% 97.9% 99.8% 0.97 95.0% 99.4%
Intra-laboratory run 1 versusInter-laboratory (Ghent)
 Fluorescent channel 1 99.6% 99.1% 100% 0.99 97.8% 100%
 Fluorescent channel 2 99.6% 99.1% 100% 0.99 97.8% 100%
 Fluorescent channel 3 98.8% 97.9% 99.8% 0.97 94.9% 99.4%
 Fluorescent channel 4 99.2% 98.5% 100% 0.98 96.3% 100%
 Fluorescent channel 5 98.6% 97.6% 99.6% 0.97 94.3% 99.1%
Intra-laboratory run 2 versusInter-laboratory (Ghent)
 Fluorescent channel 1 99.4% 98.8% 100% 0.99 97.0% 100%
 Fluorescent channel 2 100% 100% 100% 1.00 100% 100%
 Fluorescent channel 3 98.8% 97.9% 99.8% 0.97 95.0% 99.4%
 Fluorescent channel 4 99.2% 98.5% 100% 0.98 96.3% 100%
 Fluorescent channel 5 99.8% 99.4% 100% 1.00 98.6% 10%

CI Confidence interval, Fluorescent channel FC 1 detects HPV16, FC2 detects HPV18 and HPV45, FC3 detects HPV31,33,35,52,58, FC4 detects HPV51,59, FC5 detects HPV 39,56,66,68

Discussion and conclusions

During the intra- and inter- laboratory reproducibility testing 1530 Xpert HPV assay runs were conducted. Sixteen samples that were identified as INVALID by the Xpert HPV assay. These samples were re-tested and the results of the assay turned out to be valid after re-testing. This implies that the internal control at the initial testing was indicating assay failure rather than sample failure. Xpert HPV has shown excellent reproducibility but it also has an error rate of approximately 1%. An error rate of 1% can be significant when large numbers of samples are analyzed with Xpert HPV.

Xpert HPV showed excellent overall intra-laboratory reproducibility with an agreement of 96.9% and a kappa of 0.925). The overall reproducibility was close to 100% for HPV 16 and HPV 18/45, confirming reliability of Xpert HPV for the main high-risk types. Besides that the inter-laboratory test showed an excellent agreement of 97.8% with a kappa of 0.948 and great reproducibility statistics for HPV16 and HPV18/45. A small number of discrepancies were found, but almost all of them appeared to be linked to detection limit, Ct-value or borderline issues.

Our data confirm that the intra-laboratory reproducibility and inter-laboratory agreement of Xpert HPV largely exceeds the requirements formulated in the international guidelines for the validation of a novel hrHPV test with respect to its use in primary cervical cancer screening. Our study completes the VALGENT-2 clinical validation study which demonstrated a relative sensitivity and relative specificity of hrHPV testing with Xpert HPV compared to GP5+/6+ PCR of 0.984 (95% CI 0.931–1.040, pnon-inferiority [pni] = 0.019) and 1.006 (95% CI 0.997–1.016, pni < 0.0001), respectively, among screened women aged 30 or older [9]. Considering the VALGENT findings with the reproducibility results from the current study, we can conclude that the Xpert HPV can be added to the list of assays that fulfill international criteria for primary cervical cancer screening [3]. In addition, Xpert HPV is hereby the only point of care test fulfilling these criteria.

Acknowledgments

Funding

Support was provided by the molecular diagnostic company Cepheid Inc. for methodological and statistical work as foreseen in VALGENT Network (Arbyn et al., J clinical Virol, 2016). VALGENT is an independent researcher induced research project where manufacturers can have their HPV assays evaluated, under the condition that they provided equipment, kits and to cover costs for laboratory work and statistical analysis. Manufacturers cannot influence publication of manuscripts.

Availability of data and materials

Data sets can be shared upon request.

Abbreviations

CIN2+

Cervical intraepithelial neoplasia grade 2+

Ct

Cycle threshold

DNA

Deoxyriboynucleic acid

EIA

Enzyme immunoassay

HC2

Hybrid capture 2

HPV

Human papillomavirus

HrHPV

High risk human papillomavirus

PCR

Polymerase chain reaction

SAC

Sample adequacy control

Authors’ contributions

AA Conduction of practical work. DVB Conduction of practical work and protocol development. IB Conduction of practical work. EP Conduction of practical work. JB Protocol development. MA Statistical Analysis and protocol development. All authors approved and read the final manuscript.

Ethics approval and consent to participate

A favorable ethical opinion was provided by the West of Scotland Research Ethics Committee 4. Reference number: 11/WS/0038.

Consent to publication

Not applicable

Competing interests

The authors declare that they have no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Contributor Information

Ajmal Akbari, Email: ajmal.akbari@gmail.com.

Davy Vanden Broeck, Email: Davy.vandenbroeck@aml-lab.be.

Ina Benoy, Email: ina.benoy@aml-lab.be.

Elizaveta Padalko, Email: elizaveta.padalko@uzgent.be.

Johannes Bogers, Email: john-paul.bogers@aml-lab.be.

Marc Arbyn, Email: Marc.Arbyn@sciensano.be.

References

  • 1.Arbyn M, Ronco G, Anttila A, et al. Evidence regarding human papillomavirus testing in secondary prevention of cervical cancer. Vaccine. 2012;30(Suppl 5):F88–F99. doi: 10.1016/j.vaccine.2012.06.095. [DOI] [PubMed] [Google Scholar]
  • 2.Meijer CJLM, Berkhof H, Heideman DA, Hesselink AT, Snijders PJF. Validation of high-risk HPV tests for primary cervical screening. J Clin Virol. 2009;46:S1–S4. doi: 10.1016/S1386-6532(09)00540-X. [DOI] [PubMed] [Google Scholar]
  • 3.Arbyn M, Snijders PJF, Meijer CJLM, et al. Which high-risk HPV assays fulfil criteria for use in primary cervical cancer screening? Clin Microbiol Infect. 2015;21:817–826. doi: 10.1016/j.cmi.2015.04.015. [DOI] [PubMed] [Google Scholar]
  • 4.Micalessi IM, Boulet GA, Bogers JJ, Benoy IH, Depuydt CE. High-throughput detection, genotyping and quantification of the human papillomavirus using real-time PCR. Clin Chem Lab Med. 2012;50(4):655–661. doi: 10.1515/cclm.2011.835. [DOI] [PubMed] [Google Scholar]
  • 5.Depuydt CE, Benoy IH, Beert JF, Criel AM, Bogers JJ, Arbyn M. Clinical validation of a type-specific real-time quantitative human papillomavirus PCR against the performance of hybrid capture 2 for the purpose of cervical cancer screening. J Clin Microbiol. 2012;50(12):4073–4077. doi: 10.1128/JCM.01231-12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Heid CA, Stevens J, Livak KJ, Williams PM. Real time quantitative PCR. Genome Res. 1996;6(10):986–994. doi: 10.1101/gr.6.10.986. [DOI] [PubMed] [Google Scholar]
  • 7.Arbyn Marc, Depuydt Christophe, Benoy Ina, Bogers Johannes, Cuschieri Kate, Schmitt Markus, Pawlita Michael, Geraets Daan, Heard Isabelle, Gheit Tarik, Tommasino Massimo, Poljak Mario, Bonde Jesper, Quint Wim. VALGENT: A protocol for clinical validation of human papillomavirus assays. Journal of Clinical Virology. 2016;76:S14–S21. doi: 10.1016/j.jcv.2015.09.014. [DOI] [PubMed] [Google Scholar]
  • 8.Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. Psychol Bull. 1979;86(2):420–428. doi: 10.1037/0033-2909.86.2.420. [DOI] [PubMed] [Google Scholar]
  • 9.Cuschieri K, Geraets D, Cuzick J, Cadman L, Moore C, Vanden Broeck D, et al. Performance of a cartridge based assay for the detection of clinically significant HPV infection - lessons from VALGENT (validation of HPV genotyping tests) J Clin Microbiol. 2016;54:2337–2347. doi: 10.1128/JCM.00897-16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Arbyn M, Van Oyen H. Cervical cancer screening in Belgium. Eur J Cancer. 2000;36:2191–2197. doi: 10.1016/S0959-8049(00)00308-7. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

Data sets can be shared upon request.


Articles from Virology Journal are provided here courtesy of BMC

RESOURCES