3D patient imaging and retrieval analysis help understand the clinical importance of rotation in knee replacements

Arianna Cerquiglini 1,, Johann Henckel 1, Harry Hothi 1, Niccoló Rotigliano 2, Michael T Hirschmann 2, Alister J Hart 1
PMCID: PMC6208955  PMID: 29520668

Abstract

Purpose

The purpose of the present study was to correlate highly accurate CT measurements of pre-revision total knee arthroplasty (TKA) implant position with findings of retrieval analysis post-revision, to understand the clinical relevance of TKA orientation.

Methods

This study involved 53 retrieved TKA implants with pre-revision 3D-CT scans used to determine coronal (varus–valgus), sagittal (tibial slope) and rotational (internal rotation–external rotation) TKA orientation as well as tibiofemoral leg axis. Differences between femoral and tibial angles to describe the "relative rotational mismatch" were also calculated. All tibial inserts were forensically analyzed using the Hood score. Statistical analysis was performed to investigate correlations between TKA component orientation and surface damage (p < 0.05).

Results

Femoral components were found to have axial rotations mainly within ± 3° (68%), whilst 45% of the tibial components and 66% of the relative rotational mismatches were > 3° and < − 3°, respectively. The majority of femoral and tibial components (87% in both cases), as well as the femorotibial angle (70%), showed coronal orientations within ± 3°. The 64% of the tibial components showed posterior tibial slopes out of both the 0°–3° and 5°–7° ranges. There was a significant correlation between tibial slope and damage score on polyethylene tibial inserts (r = 0.2856; p = 0.0382) as well as a significant correlation between implants’ position in the axial plane and damage score on polyethylene tibial inserts (r = 0.6537, p = 0.0240).

Conclusions

This is the first study to use accurate measurements from pre-revision 3DCT to compare tibial and femoral orientation in all three planes with retrieval findings in total knee replacements. A significant correlation between implant position and polyethylene surface damage was found. These results showed the importance of optimizing component position to minimize polyethylene damage. Further analysis involving more accurate polyethylene wear measurements are fundamental to fully understand the role of components’ orientation in TKAs.

Keywords: Total knee replacement, Total knee arthroplasty, Implant orientation, Malalignment, 3D computed tomography, Polyethylene surface damage

Introduction

Total knee arthroplasty (TKA) is a common intervention with an excellent survival rate; however, up to 20% of patients have reported poor outcomes, leading to revision [1]. This phenomenon has global dimensions: the demand for revision of TKAs is projected to grow by 601% between 2005 and 2030 in the United States [2] and the same trend is also expected in Europe and Asia Pacific [3, 4].

The reasons for TKA failure are multifactorial and influenced by surgical, implant and patient factors. The most commonly reported reasons for revision are aseptic loosening, pain, infection, instability, stiffness, polyethylene wear, malposition, patellofemoral problems and dislocation/subluxation [1, 59].

A suboptimal TKA position of femoral and tibial components contributes to poor outcomes, premature polyethylene wear and “unexplained” painful TKA [7, 8, 1015]. It is well established that excessive internal rotation of the femoral component in the axial plane leads to patellar maltracking, anterior knee pain and flexion instability [7, 1012]. Some studies also found a significant correlation between malposition in the coronal plane and aseptic loosening of the implant, due to a higher amount of polyethylene wear caused by abnormal force distributions [1618]. Suboptimal flexion was found in cruciate retaining TKA with reduced posterior tibial slope [16].

Findings from retrieval studies of failed TKA implants provide valuable information on the location and potential mechanisms for TKA component damage in vivo [17, 19]. Several papers have suggested that the wear pattern is associated with different clinical and mechanical factors, such as component position, orientation and alignment [1315, 18, 2025]. For example, it has been shown that there is often more medial wear on tibial polyethylene bearings in well-positioned knees [7], and changes in orientation are known to influence this (increased medial wear in varus knees and greater lateral in valgus knees) [18, 25]. Malrotation might cause abnormal stresses and premature wear of the polyethylene components, followed by peri-prosthetic and implant loosening [7, 8, 10, 11, 18, 19]. However, the relationship between implant orientation and wear patterns in retrieved knee prostheses is still poorly understood.

The aim of this study was, for the first time, to correlate highly accurate 3D-CT measurements of pre-revision TKA position, provided by an innovative 3D imaging technique [8], with retrieval analysis findings post-revision to better understand implant orientation effects on TKA.

Materials and methods

This was a retrieval study involving 53 contemporary TKAs that had been consecutively revised at a single institution. The implants were revised from 30 female and 23 male patients with a median (range) age of 62 (42–78) years and a median time to revision of 38 (5–162) months. The reasons for revision were instability (n = 26), malposition (n = 11), patella maltracking (n = 7), aseptic loosening (n = 3), pain (n = 2), stiffness (n = 2), infection (n = 1) and arthrofibrosis (n = 1). The retrieved TKAs consisted of 40 cruciate retaining (CR) and 13 posterior stabilized (PS) TKA (Fig. 1). Table 1 summarizes the TKA specifications and patient demographics for each case.

Fig. 1.

Fig. 1

Examples of contemporary retrieved knee implants. a Posterior stabilized (PS) TKA with fix bearing, b cruciate retaining (CR) TKA with rotating bearing

Table 1.

Implant and patient demographic

Case number Design Manufacturer Patient age Gender Time to revision [months] Reason for revision
1 NexGen LCCK, PS Zimmer 69 M 29 Infection
2 Innex, CR Zimmer 65 F 71 Instability
3 PFC Sigma, CR DePuy 69 M 15 Aseptic loosening
4 PFC Sigma, CR DePuy 66 M 53 Malposition
5 Triathlon, PS Stryker 53 F 79 Instability
6 Natural Knee II, CR Zimmer 69 M 36 Instability
7 Triathlon, PS Stryker 75 M 99 Aseptic loosening
8 Triathlon, CR Stryker 66 M 72 Instability
9 BalanSys, CR Mathys 61 F 40 Patella maltracking
10 Natural Knee II, CR Zimmer 71 F 90 Patella maltracking
11 PFC Sigma, CR DePuy 46 F 20 Pain
12 TC PLUS, CR Smith&Nephew 62 M 10 Instability
13 Type LCS, CR DePuy 46 M 120 Instability
14 ATTUNE, PS DePuy 68 F 15 Instability
15 Signature, PS Zimmer 64 M 10 Malposition
16 NexGen, PS Zimmer 51 F 6 Pain
17 Type LCS, CR DePuy 51 F 22 Instability
18 Synthes-LCS, CR DePuy 71 F 24 Patella maltracking
19 NexGen, PS Zimmer 48 M 36 Instability
20 Synthes-LCS, CR DePuy 64 F 162 Instability
21 Persona, CR Zimmer 57 M 60 Patella maltracking
22 Type LCS, CR DePuy 70 F 12 Instability
23 Journey, CR Smith&Nephew 48 M 21 Instability
24 PFC Sigma, PS DePuy 53 F 45 Malposition
25 BalanSys, CR Mathys 62 F 18 Instability
26 ATTUNE, PS DePuy 78 F 5 Instability
27 Triathlon, CR Stryker 60 M 91 Instability
28 BalanSys, CR Mathys 56 M 25 Instability
29 BalanSys, CR Mathys 53 F 22 Instability
30 PFC Sigma, CR DePuy 62 F 60 Instability
31 Innex, CR Zimmer 66 F 13 Malposition
32 TC PLUS, CR Smith&Nephew 59 M 101 Stiffness
33 BalanSys, CR Mathys 42 F 51 Malposition
34 Gemini, PS Link 72 F 58 Malposition
35 PFC Sigma, CR DePuy 73 F 37 Malposition
36 PFC Sigma, PS DePuy 72 F 19 Stiffness
37 PFC Sigma, PS DePuy 63 F 61 Aseptic loosening
38 Triathlon, CR Stryker 49 F 17 Instability
39 BalanSys, CR Mathys 66 M 77 Instability
40 Colombus, CR Aesculap 59 F 60 Patella maltracking
41 BalanSys, CR Mathys 46 M 46 Malposition
42 PFC Sigma, CR DePuy 70 M 64 Instability
43 ATTUNE, CR DePuy 67 F 13 Patella maltracking
44 ATTUNE, CR DePuy 64 F 22 Instability
45 Synthes-LCS, CR DePuy 62 F 24 Instability
46 BalanSys, CR Mathys 68 M 156 Malposition
47 Natural Knee II, CR Zimmer 59 M 15 Malposition
48 BalanSys, PS Mathys 55 F 100 Instability
49 Natural Knee II, CR Zimmer 74 M 38 Malposition
50 Natural Knee II, CR Zimmer 56 M 79 Malposition
51 PFC Sigma, CR DePuy 60 M 101 Instability
52 PFC Sigma, CR DePuy 61 F 10 Instability
53 Triathlon, CR Stryker 62 F 66 Patella maltracking

All investigations were conducted in conformity with ethical principles of research, that informed consent for participation in the study was obtained and that institutional approval of the human protocol for this investigation was obtained.

3D computed tomography (CT) position

Pre-revision CT scans of both the femoral and tibial components from each patient were taken using the imperial CT protocol, which obtains the relevant bony landmarks (hip–knee–ankle) and includes specific metal artefact reduction sequences [8, 26]. From the CT images, the reference axes (anatomical and mechanical) were defined, the images standardized and the angles computed, to provide the following positioning measurements: (1) axial rotation in the transverse plane of both the femoral and tibial components, with reference to the transepicondylar and anteroposterior axes, respectively, (2) the degree of varus/valgus of single components and the tibiofemoral angle in the coronal plane and (3) the tibial component slope in the sagittal plane, Fig. 2.

Fig. 2.

Fig. 2

A 3D CT image from a patient illustrating components’ positions. a View from the transverse plane showing internal rotation of the femoral component; b view from the coronal plane showing varus angle on the femoral component; c view from the transverse plane showing external rotation of the tibial component; d view from the coronal plane showing varus angle of the tibial component; e view from the transverse plane showing tibiofemoral angle; f view from the sagittal plane showing posterior slope of the tibial component

Differences between femoral and tibial axial rotation angles to describe the relative rotational mismatch [27] between these two components in the transverse plane were also computed; the measured angles were used to estimate the magnitude of this mismatch, while the sign explained the relative position between the femoral and the tibial components (negative values meant that the femoral component was internally rotated compared to the tibial one; positive values meant that the femoral component was externally rotated compared to the tibial one).

The measured values were classified into three ranges (Fig. 3), in agreement with surgical standard aims [28, 29].

Fig. 3.

Fig. 3

Classification of the orientations in the three planes (axial, coronal and sagittal orientations)

Explant preparation

All implants were decontaminated using 10% formalin solution. The polyethylene components were then stored in a freezer at − 18 °C, to minimise the oxidation process.

Analysis of polyethylene surfaces

The tibial polyethylene inserts (n = 53) were forensically analyzed using a Keyence VHX-700F series (Keyence Co., Japan) digital microscope, with magnification from 50× up to 200×. On the articular surface, both the medial and lateral sides were divided into four sections, while the central part of the insert into two sections (Fig. 4). Each of the 10 total sections were analyzed using the Hood score, according to the presence and severity of seven modes of surface degradation (surface deformation, pitting, embedded debris, scratching, burnishing, abrasion and delamination) [30], Table 2. The maximum damage grade was 21 for a single section (grade 3 for each of the seven damage modes) and 210 for the entire surface (grade 3 for each of the seven damage modes for each of the 10 sections). Grading was performed by two different examiners. In case of disagreement, the examiners discussed the results together, to agree a final grade.

Fig. 4.

Fig. 4

Surface divisions on the articular surface of the tibial insert, showing 10 sections (from 0 to 9) defined by Hood [30]

Table 2.

Damage-grading criteria used to assess polyethylene component surface, according to the Hood score [30]

Polyethylene surfaces
Modes of surface degradation Deformation, pitting, embedded debris, scratching, burnishing, abrasion, delamination
Score
 0 Not visible
 1 < 10% of surface
 2 10–50% of surface
 3 > 50% of surface
Maximum section score 21
Maximum total score 210

Statistical analysis

Statistical analysis was performed to determine if there were any significant correlations between orientation of components and expected location of the surface damage, described in the literature [18, 20, 24, 25]. Potential associations between (1) internal relative rotational mismatch and the amount of damage on the posterior compartment in the medial side and on the anterior compartment in the lateral side, Fig. 5a; (2) external relative rotational mismatch and the amount of damage on the posterior compartment in the lateral side and on the anterior compartment in the medial side, Fig. 5b; (3) varus tibiofemoral angle and the amount of damage on the medial compartment, Fig. 5c; (4) valgus tibiofemoral angle and the amount of damage on the lateral compartment, Fig. 5d; (5) posterior slope and amount of damage on the posterior compartments, Fig. 5e were analyzed (two-tailed non-parametric Spearman correlation, p value < 0.05 was considered as significant).

Fig. 5.

Fig. 5

Localisation of expected wear (on a right knee) according to implant orientations. a An internal relative rotational mismatch is expected to increase the damage on the posterior compartment in the medial side and on the anterior compartment in the lateral side; b an external relative rotational mismatch is expected to increase the damage on the posterior compartment in the lateral side and on the anterior compartment in the medial side; c a varus tibiofemoral angle is expected to increase the damage on the medial compartment; d a valgus tibiofemoral angle is expected to increase the damage on the lateral compartment; e posterior slope is expected to decrease the damage on posterior compartments

Potential significant differences in the total Hood score between optimal and suboptimal orientation were also investigated (Mann–Whitney test, p value < 0.05 was considered as significant).

Separate statistical analyses on implants revised for malposition were performed and significant differences between implants with short (≤ 2 years) and long (> 2 years) implantation time were also investigated (Mann–Whitney test, p value < 0.05 was considered as significant). All statistical analyses were performed using Prism 7 (GraphPad, USA).

Results

3D-CT Position, axial plane

The 3D-CT imaging revealed that 68% of the femoral components had a rotation within ± 3°, whilst the rest showed rotations smaller than − 3°. 43% of the tibial components showed orientations ± 3°, whilst 45% and 12% had axial rotations greater than 3° and smaller than − 3°, respectively.

It was found that in 26% of cases, implants showed relative rotational mismatch within ± 3°. 66% showed internal rotations smaller than − 3° and 8% showed external rotations greater than 3°. Table 3 summarizes all the results, showing median and range values.

Table 3.

Number of cases, median and range values of orientations of the femoral and tibial components and relative rotational mismatch in the axial plane

Axial orientation Number of cases Median (range) [°]
Femoral component ± 3° 36 − 2 (− 3 to 3)
> 3°
< − 3° 17 − 6 (− 11 to − 4)
Tibial component ± 3° 23 1 (− 3 to 3)
> 3° 24 8 (4 to 19)
< − 3° 6 − 9.5 (− 12 to − 6)
Relative rotational mismatch ± 3° 14 − 1 (− 3 to 3)
> 3° 4 10 (7 to 12)
< − 3° 35 − 9 (− 22 to − 4)

3D-CT Position, coronal plane

87% of the femoral components had orientations within ± 3°, whilst 11% had angulations smaller than − 3° and the 2% showed angulation greater than 3°.

The majority of tibial components showed angulations within ± 3°, whilst 11% and 2% had angulations smaller than − 3° and greater than 3°, respectively.

Considering the tibiofemoral angle, 70% of cases showed orientations within ± 3°, whilst 21% and 9% had angulations smaller than − 3° and greater than 3°, respectively.

Table 4 summarizes all the results, showing median and range values.

Table 4.

Number of cases, median and range values of orientations of the femoral and tibial components and femorotibial angle in the coronal plane

Coronal orientation Number of cases Median (range) [°]
Femoral component ± 3° 46 0 (− 3 to 3)
> 3° 1 4 (4)
<− 3° 6 − 4 (− 7 to − 4)
Tibial component ± 3° 46 − 1 (− 3 to 3)
> 3° 1 5 (5)
< − 3° 6 − 4 (− 5 to − 4)
Femorotibial angle ± 3° 37 − 1 (− 3 to 3)
> 3° 5 5 (4 to 6)
< − 3° 11 − 4 (− 14 to − 4)

3D-CT Position, Sagittal plane

64% of the tibial components showed posterior tibial slopes that were not within 0° to 3° or 5° to 7. 36% had positions of tibial trays within 0° to 3° or 5° to 7°.

Table 5 summarizes all the results, showing median and range values.

Table 5.

Number of cases, median and range values of orientations of the tibial components in the sagittal plane: negative values mean posterior tibial slope, whilst positive values mean anterior tibial slope

Sagittal orientation Number of cases Median (range) [°]
Posterior Staibilized design − 3° to 0° 7 − 1 (− 3 to 0)
Out of − 3° to 0° 6 − 4.5 (− 9 to − 2, 4)
Cruciate Retaining design − 7° to − 5° 12 − 5 (− 7 to − 5)
Out of − 7° to − 5° 28 − 3 (− 12 to − 8, − 4 to 2)

Analysis of polyethylene surfaces

The most common modes of surface degradation were scratching, pitting and burnishing, whilst deformation and delamination were rare (Fig. 6). The median total Hood score (range) was 39 (17–72). There was no significant difference (p = 0.5459) between lateral and medial compartments (defined by the sum of 0–3 and 4–7 sections, respectively, in a right tibial insert; viceversa in a left tibial insert), that showed mean (range) damage scores of 19 (7–39) and 20 (8–41), respectively.

Fig. 6.

Fig. 6

Different modes of surface degradation found on both the articular and backside polyethylene surfaces, including scratching, pitting, burnishing and deformation, at magnification 20X

Correlations

There was a significant correlation between tibial slope and damage score on polyethylene tibial inserts: posterior tibial slope was associated with higher Hood scores on the anterior areas of the polyethylene, Fig. 7. Table 6 shows further data.

Fig. 7.

Fig. 7

Graph showing the correlation between tibial slope and Hood score. Posterior tibial slope is associated to higher Hood scores in the anterior side of the polyethylene, whilst anterior tibial slope is associated to higher Hood scores in the posterior side of the polyethylene

Table 6.

Correlation between implant orientation and Hood score in all the anatomical planes and differences in Hood scores between normal and suboptimal orientation

Correlation between orientation and Hood score in expected areas? Significant difference in the Hood score between normal and suboptimal orientation?
Axial plane Axial mismatch n.s.
r = 0.02978; p = 0.8826
n.s.
p = 0.3042
Coronal plane Tibiofemoral angle n.s.
r = − 0.1168; p = 0.4051
n.s.
p = 0.0964
Sagittal plane Tibial slope Significant
r = − 0.2856; p = 0.0382
PS → n.s. p = 0.1678
CR → n.s., p = 0.0851

n.s. not significant

Separate analyses of the group revised for malposition revealed a significant correlation between implants’ position in the axial plane and damage score on polyethylene tibial inserts, Fig. 8: internal rotation mismatch was associated with higher Hood scores on the posterior compartment in the medial side and on the anterior compartment in the lateral side, while external ones were associated with higher Hood scores on the posterior compartment in the lateral side and on the anterior compartment in the medial side, Table 7.

Fig. 8.

Fig. 8

Graph showing the correlation between relative rotational mismatch and Hood score in implants revised for malposition. Internal relative mismatches are associated to higher Hood scores on the posterior compartment in the medial side and on the anterior compartment in the lateral side, while external ones were associated with higher Hood scores on the posterior compartment in the lateral side and on the anterior compartment in the medial side

Table 7.

Correlation between implant orientation and Hood score in all the anatomical planes in implants revised for malposition

Correlation between orientation and Hood score in expected areas?
Relative rotational mismatch Significant
r = 0.6537, p = 0.0240
Femorotibial angle n.s.
r = − 0.02154, p = 0.8714
Posterior tibial slope n.s.
r = − 0.1485, p = 0.5814

n.s. not significant

There was a significant difference in total Hood score between implants with short (≤ 2 years) and long (> 2 years) implantation time.

There was a significant correlation between time to revision and total Hood score, Fig. 9.

Fig. 9.

Fig. 9

Graph showing the correlation between time to revision and hood score. Retrieval implanted for longer period show higher total Hood scores

Discussion

The most important finding of the present study was a significant correlation between relative rotational mismatch and the severity and location of polyethylene damage in implants revised for malposition. There was also a significant correlation between the tibial slope in the sagittal plane, and amount and location of surface damage, but no significant correlation was found between implant orientation and surface damage in polyethylene tibial inserts in both the axial and coronal planes. Moreover, a significant correlation was found between time to revision and hood score. These findings support a multifactorial aetiology of polyethylene wear: implant position alone cannot explain all complex mechanisms involved in the generation of surface damage in polyethylene, especially in early revised implants. This is the first retrieval study to use 3D-CT pre-revision images to measure implant position and correlate findings with results from polyethylene surface damage inspection.

3D-CT-measured implant orientation revealed that the majority of both femoral and tibial components showed optimal orientations in the axial plane, with few cases of excessive internal and external rotations of femoral and tibial components, respectively. Calculations of the relative rotational mismatch revealed that the majority were internally rotated. Regarding the coronal plane, both the femoral and tibial showed mainly optimal orientation, as well as the tibiofemoral angle. In the sagittal plane, almost the totality of the tibial trays showed suboptimal posterior slope.

The current study was the first to investigate the relationship between wear patterns in retrieved knee replacements and 3D-CT-based computed measurements in all the three anatomical planes. In particular, the orientation in the axial plane is difficult to measure with traditional imaging methods [8]; other research groups have identified position in this orientation using roentgenographic score systems [23], X-ray [18, 24] or planar CT images [12, 27, 31, 32]. Axial rotation was measured with accuracy and repeatability using an innovative 3D-CT imaging technique with a standardized, published method and correlates these measurements with results from retrievals. Interestingly, investigating implants revised only for malposition, 9 out of 12 samples showed severe cases of malposition in the axial plane: internal rotational mismatches had a median value (range) of 10 (1–22) degrees, while external rotational mismatches had a median value (range) of 9 (1–12) degrees. In these cases, a significant correlation with the Hood score was found; this could suggest that only severe malposition can lead to asymmetry in the surface damage of polyethylene tibial inserts.

The main characteristic degradation features found on polyethylene inserts were scratching, pitting and burnishing, with no significant difference between medial and lateral sides. This symmetric wear pattern was in contrast to the predominantly asymmetric patterns found in previous studies [18, 20, 2224]. This discrepancy might be caused by the subjective nature of the widely used Hood score. Indeed, surface damage was found to be only a moderate predictor of wear in polyethylene [33]. Moreover, our components were revised relatively early with median time to revision of 37 months (5–162). The main reasons for revision were instability, malposition and patella maltracking. It is most likely that all patients had poor clinical function and, therefore, low levels of use. This would imply a lower gait cycle number (steps or joint cycles of use per year) and well below the average of 0.9–1.4 million gait cycles per annum. This relative lack of use may also explain why wear patterns did not correlate with components’ positions in all the anatomical planes. Furthermore, the multifactorial nature of the wear pattern mechanisms has to be taken into consideration: this includes a combination of surgical, implant and patient variables. For example, the precision in determining the rotational alignment during a total knee replacement procedure is crucial; correct component positioning can prevent post-cam impingement and reduce the polyethylene wear [32]. Knee replacement design is an important factor that can influence wear [10, 11, 18, 2023 27, 18]. It has been shown that a new posterior stabilized design can lead to a reduction in polyethylene surface damage and peg deformation [34]. It has also previously been reported that more active patients show greater wear in polyethylene inserts [35].

A considerable number of limitations have to be considered. First, the evaluation of wear pattern was performed using visual scoring for plastic components. Although previous papers [6, 8, 10, 23, 24] used these methods and demonstrated that they can give an acceptable estimation about the quality of the damage, they may be not accurate enough to measure the amount of wear [33]. Recently, the utility of alternative techniques, such as a coordinate measure machine (CMM), laser and micro-CT scanning [24, 25], was demonstrated. Further analyses involving these types of tool may lead to new findings and results, which in combination with 3D CT might allow us to fully understand the role of components’ orientation in total knee replacements and to define a complete “safe zone” for TKA. Future studies should recruit greater number of patients and include a greater number of implants of a single design.

These results showed the importance of optimizing component position to minimize polyethylene damage.

Conclusion

This is the first study to use accurate measurements from pre-revision 3D-CT to compare tibial and femoral orientation in all three planes with retrieval findings in 53 total knee replacements. Our results revealed a significant correlation between implant axial alignment, and the severity and location of polyethylene damage in implants revised for malposition as well as between tibial slope and damage score on polyethylene tibial inserts.

Author contributions

AC was involved in study design, data collection, results interpretation and writing the manuscript. JH, HH, NR, MTH and AH were involved in study design, results interpretation and writing the manuscript.

Funding

The senior author received funding from the British Orthopaedic Association through an industry consortium of nine manufacturers: DePuy International Ltd, Zimmer GmbH, Smith & Nephew UK Ltd, Biomet UK Ltd, JRI Ltd, Finsbury Orthopaedics Ltd, Corin Group PLC, Mathys Orthopaedics Ltd, and Stryker UK Ltd.

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical approval

All investigations were conducted in conformity with ethical principles of research and that institutional approval of the human protocol for this investigation was obtained.

Informed consent

Informed consent for participation in the study was obtained.

Contributor Information

Arianna Cerquiglini, Email: arianna.cerquiglini.15@ucl.ac.uk.

Johann Henckel, Email: j.henckel@ucl.ac.uk.

Harry Hothi, Email: h.hothi@ucl.ac.uk.

Niccoló Rotigliano, Email: Niccolo.Rotigliano@ksbl.ch.

Michael T. Hirschmann, Email: michael.hirschmann@unibas.ch

Alister J. Hart, Email: a.hart@ucl.ac.uk

References

  • 1.National Joint Registry for England and Wales 14th Annual Report (2017) http://www.njrcentre.org.uk/njrcentre/Portals/0/Documents/England/Reports/14th_annual_report. Accessed 25 Oct 2017
  • 2.Kurtz S, Ong K, Lau E, Mowat F, Halpern M. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am. 2007;89(4):780–785. doi: 10.2106/JBJS.F.00222. [DOI] [PubMed] [Google Scholar]
  • 3.The Australian Orthopaedic Association National Joint Replacement Registry (AOANJRR), annual report 2017. https://aoanjrr.sahmri.com
  • 4.Robertsson O, Bizjajeva S, Fenstad AM, et al. Knee arthroplasty in Denmark, Norway and Sweden: a pilot study from the Nordic Arthroplasty Register Association. Acta Orthop. 2010;81(1):82–89. doi: 10.3109/17453671003685442. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Callaghan JJ, O’roukre MR, Saleh K. Why knees fail: lessons learned. J Arthroplasty. 2004;19(4 Suppl 1):31–34. doi: 10.1016/j.arth.2004.02.015. [DOI] [PubMed] [Google Scholar]
  • 6.Dalury DF, Pomeroy DL, Gorab RS, Adams MJ. Why are total knee arthroplasties being revised? J Arthroplasty. 2013;28(8 Suppl):120–121. doi: 10.1016/j.arth.2013.04.051. [DOI] [PubMed] [Google Scholar]
  • 7.Hirschmann MT, Becker R. The unhappy knee replacement. A comprehensive review and management guide. Roland Becker Editors. New York: Springer; 2015. [Google Scholar]
  • 8.Hirschmann MT, Konala P, Amsler F, Iranpour F, Friederich NF, Cobb JP. The position and orientation of total knee replacement components. J Bone Joint Surg Br. 2011;93(5):629–633. doi: 10.1302/0301-620X.93B5.25893. [DOI] [PubMed] [Google Scholar]
  • 9.Schroer WC, Berend KR, Lombardi AV, Barnes CL, Bolognesi MP, Berend ME, Ritter MA, Nunley RM. Why are total knees failing today? Etiology of total knee revision in 2010 and 2011. J Arthroplasty. 2013;28(8 Suppl):116–119. doi: 10.1016/j.arth.2013.04.056. [DOI] [PubMed] [Google Scholar]
  • 10.Armstrong AD, Brien HJ, Dunning CE, King GJ, Johnson JA, Chess DG. Patellar position after total knee arthroplasty: influence of femoral component malposition. J Arthroplasty. 2003;18(4):458–465. doi: 10.1016/S0883-5403(03)00145-1. [DOI] [PubMed] [Google Scholar]
  • 11.Berger R. A., Crossett L. S., Jacobs J. J., Rubash H. E. Malrotation Causing Patellofemoral Complications After Total Knee Arthroplasty. Clinical Orthopaedics and Related Research. 1998;356:144–153. doi: 10.1097/00003086-199811000-00021. [DOI] [PubMed] [Google Scholar]
  • 12.Nicoll D, Rowley DI. Internal rotational error of the tibial component is a major cause of pain after total knee replacement. J Bone Joint Surg Br. 2010;92(9):1238–4124. doi: 10.1302/0301-620X.92B9.23516. [DOI] [PubMed] [Google Scholar]
  • 13.Slevin O, Amsler F, Hirschmann MT. No correlation between coronal alignment of total knee arthroplasty and clinical outcomes: a prospective clinical study using 3D-CT. Knee Surg Sports Traumatol Arthrosc. 2016;25(12):3892–3900. doi: 10.1007/s00167-016-4400-y. [DOI] [PubMed] [Google Scholar]
  • 14.Slevin O, Hirschmann A, Schiapparelli FF, Amsler F, Huegli RW, Hirschmann MT, Hirschmann MT. Neutral alignment leads to higher knee society scores after total knee arthroplasty in preoperatively non-varus patients: a prospective clinical study using 3D-CT. Knee Surg Sports Traumatol Arthrosc. 2017 doi: 10.1007/s00167-017-4744-y. [DOI] [PubMed] [Google Scholar]
  • 15.Slevin O, Schmid FA, Schiapparelli FF, Rasch H, Amsler F, Hirschmann MT. Coronal femoral TKA position significantly influences in vivo patellar loading in unresurfaced patellae after primary total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc. 2017;25(11):3605–3610. doi: 10.1007/s00167-017-4627-2. [DOI] [PubMed] [Google Scholar]
  • 16.Chambers AW, Wood AR, Kosmopoulos V, Sanchez HB, Wagner RA. Effect of posterior tibial slope on flexion and anterior-posterior tibial translation in posterior cruciate-retaining total knee arthroplasty. J Arthroplasty. 2016;31(1):103–106. doi: 10.1016/j.arth.2015.08.027. [DOI] [PubMed] [Google Scholar]
  • 17.Di Laura A, Whittaker R, Hothi H, Kwon YM, Skinner J, Hart A. Dual mobility vs standard bearings: what is the effect on taper junction? Bone Joint J. 2016;98(sup 11):3. [Google Scholar]
  • 18.Wasielewski RC, Galante JO, Leighty RM, Roseneberg AG. Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical consideration during total knee arthroplasty. Clin Orthop Relat Res. 1994;299:31–43. [PubMed] [Google Scholar]
  • 19.Metthies AK, Henckel J, Skinner JA, Hart AJ. A retrieval analysis of explanted Durom metal-on-metal hip arthroplasties. Hip Int. 2011;21(06):724–731. doi: 10.5301/HIP.2011.8885. [DOI] [PubMed] [Google Scholar]
  • 20.Currier JH, Bill MA, Mayor MB. Analysis of wear asymmetry in a series of 94 retrieved polyethylene tibial bearings. J Biomech. 2005;38(2):367–375. doi: 10.1016/j.jbiomech.2004.02.016. [DOI] [PubMed] [Google Scholar]
  • 21.Flannery M, Jones E, Birkinshaw C. Compliant layer knee bearings. Part II: preliminary wear assessment. Wear. 2010;269(5–6):331–338. doi: 10.1016/j.wear.2010.04.002. [DOI] [Google Scholar]
  • 22.Liau JJ, Cheng CK, Huang CH, Lo WH. The effect of malalignment on stresses in polyethylene component of total knee prostheses—a finite element analysis. Clin Biomech (Bristol Avon) 2002;17(2):140–146. doi: 10.1016/S0268-0033(01)00109-7. [DOI] [PubMed] [Google Scholar]
  • 23.Lokte PA, Ecker ML. Influence of positioning of Prosthesis in total knee replacement. J Bone Joint Surg Am. 1977;59(1):77–79. doi: 10.2106/00004623-197759010-00013. [DOI] [PubMed] [Google Scholar]
  • 24.Pang HN, Bin Abd Razak HR, Jamiecson P, Teeter MG, Naudie DD, MacDonald SJ. Factors affecting wear of constrained polyethylene tibial inserts in total knee arthroplasty. J Arthroplasty. 2016;31(6):1340–1345. doi: 10.1016/j.arth.2015.12.011. [DOI] [PubMed] [Google Scholar]
  • 25.Srivastava A, Lee GY, Steklov N, Colwell CW, Jr, Ezzet KA, D’Lima DD. Effect of tibial component varus on wear in total knee arthroplasty. Knee. 2012;19(5):560–563. doi: 10.1016/j.knee.2011.11.003. [DOI] [PubMed] [Google Scholar]
  • 26.Henckel J, Richards R, Lozhkin K, Harris S, Rodriguez y Baena FM, Barrett ARW, Cobb JP. Very low-dose computed tomography for planning and outcome measurement in knee replacement. The imperial knee protocol. J Bone Joint Surg. 2006;88-B:1513–1518. doi: 10.1302/0301-620X.88B11.17986. [DOI] [PubMed] [Google Scholar]
  • 27.Harman MK, Banks SA, Kirschner S, Lützner J. Prosthesis alignment affects axial rotation motion after total knee replacement: a prospective in vivo study combining computed tomography and fluoroscopic evaluations. BMC Musculoskelet Disord. 2012;13:206. doi: 10.1186/1471-2474-13-206. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Gromov K, Korchi M, Thomsen MG, Husted H, Troelsen A. What is the optimal alignment of the tibial and femoral components in knee arthroplasty? An overview of the literature. Acta Orthop. 2014;85(5):480–487. doi: 10.3109/17453674.2014.940573. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Jeffery RS, Morris RW, Denham RA. Coronal alignment after total knee replacement. J Bone Joint Surg Br. 1991;73(5):709–714. doi: 10.1302/0301-620X.73B5.1894655. [DOI] [PubMed] [Google Scholar]
  • 30.Hood RW, Wright TM, Burstein AH. Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses. J Biomed Mater Res. 1983;17:829–842. doi: 10.1002/jbm.820170510. [DOI] [PubMed] [Google Scholar]
  • 31.Ferrara F, Cipriani A, Rapisarda S, Iacobucci M, Magarelli N, Leone A, Bonomo L. Assessment of implant position after total knee arthroplasty by dual-energy computed tomography. Acta Radiol. 2016;57(5):612–619. doi: 10.1177/0284185115595656. [DOI] [PubMed] [Google Scholar]
  • 32.Yau WP, Chiu KY, Tang WM. How precise is the determination of rotational alignment of the femoral prosthesis in total knee arthroplasty: an in vivo study. J Arthroplasty. 2007;22(7):1042–1048. doi: 10.1016/j.arth.2006.12.043. [DOI] [PubMed] [Google Scholar]
  • 33.Knowlton CB, Bhutani P, Wimmer MA. Relationship of surface damage appearance and volumetric wear in retrieved TKR polyethylene liners. J Biomed Mater Res Part B. 2016;105(7):2053–2059. doi: 10.1002/jbm.b.33684. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Gilbert SL, Rana AJ, Lipman JD, Wright GH, Westrich GH. Design changes improve contact patterns and articular surface damage in total knee arthroplasty. Knee. 2014;21(6):1129–1134. doi: 10.1016/j.knee.2014.07.022. [DOI] [PubMed] [Google Scholar]
  • 35.McClung CD, Zahiri CA, Higa JK, Amstutz HC, Schmalzried TP. Relationship between body mass index and activity in hip or knee arthroplasty patients. J Orthop Res. 2000;18(1):35–39. doi: 10.1002/jor.1100180106. [DOI] [PubMed] [Google Scholar]

Articles from Knee Surgery, Sports Traumatology, Arthroscopy are provided here courtesy of Springer