Relation between blood pressure and pulse wave velocity for human arteries

Yinji Ma a,b,1, Jungil Choi c,d,e,1, Aurélie Hourlier-Fargette c,d,e, Yeguang Xue c,d,f,g, Ha Uk Chung d,e, Jong Yoon Lee d,e, Xiufeng Wang h, Zhaoqian Xie c,d,f,g, Daeshik Kang i, Heling Wang c,d,f,g, Seungyong Han i, Seung-Kyun Kang j, Yisak Kang k, Xinge Yu l, Marvin J Slepian m, Milan S Raj d, Jeffrey B Model d, Xue Feng a,b, Roozbeh Ghaffari d,e,n, John A Rogers c,d,e,g,n,o,p,q,r,s,t,u,2, Yonggang Huang c,d,f,g,2
PMCID: PMC6217416  PMID: 30322935

Significance

Continuous, cuffless, and noninvasive blood pressure monitoring by measuring the pulse wave velocity is generally considered to be a promising technique for noninvasive measurements. Previously reported relations between blood pressure and pulse wave velocity relation involve unrealistic assumptions that do not hold for human arteries, and also rely on empirical expressions without any theoretical basis. Here, an analytical model without such assumptions or empirical expressions is established to yield a relation between blood pressure and pulse wave velocity that has general utility for future work in continuous, cuffless, and noninvasive blood pressure monitoring.

Keywords: blood pressure, pulse wave velocity, hemodynamics, arterial stiffness, artery hyperelastic model

Abstract

Continuous monitoring of blood pressure, an essential measure of health status, typically requires complex, costly, and invasive techniques that can expose patients to risks of complications. Continuous, cuffless, and noninvasive blood pressure monitoring methods that correlate measured pulse wave velocity (PWV) to the blood pressure via the Moens−Korteweg (MK) and Hughes Equations, offer promising alternatives. The MK Equation, however, involves two assumptions that do not hold for human arteries, and the Hughes Equation is empirical, without any theoretical basis. The results presented here establish a relation between the blood pressure P and PWV that does not rely on the Hughes Equation nor on the assumptions used in the MK Equation. This relation degenerates to the MK Equation under extremely low blood pressures, and it accurately captures the results of in vitro experiments using artificial blood vessels at comparatively high pressures. For human arteries, which are well characterized by the Fung hyperelastic model, a simple formula between P and PWV is established within the range of human blood pressures. This formula is validated by literature data as well as by experiments on human subjects, with applicability in the determination of blood pressure from PWV in continuous, cuffless, and noninvasive blood pressure monitoring systems.


Blood pressure is a critical and highly elusive vital sign that varies depending on emotional state, physical activity, and health status. Low and high blood pressures correspond to two disease states called hypotension and hypertension (13), respectively. Roughly 30% of the population has hypertension-related health issues (4). The traditional method to measure blood pressure relies on an inflating cuff (5) that imparts an external pressure to the arm to stop the blood flow. Releasing this external pressure allows determination of the systolic and diastolic blood pressure, as pressures that correspond to stages of initiation and unimpeded flow of blood, respectively. Ambulatory blood pressure monitoring (6) based on this scheme requires an inflating cuff and oscillometric measurement (7, 8). The possibility for tissue damage due to the repeated blocking of blood flow in such approaches limits the interval of measurement to between 15 min and 30 min (9, 10). This sampling frequency fails to offer the time resolution necessary to detect fluctuations in blood pressure caused by, for example, exercise or mood swings. Continuous blood pressure monitoring is also essential for the care of critically ill patients and is typically achieved using invasive techniques based on intraarterial pressure measurements (11). Although considered the gold standard for beat-to-beat blood pressure monitoring, such methods expose patients to risks of complications and require intensive care monitoring (12).

Continuous, cuff less, and noninvasive blood pressure monitoring by measuring the pulse wave velocity (PWV) is generally considered to be a promising technique for continuous noninvasive measurements (1316). PWV is defined as the velocity of the propagation for the pulse wave in the artery. The Moens−Korteweg (MK) (17) + Hughes (18) Equations are generally used to relate PWV to the blood pressure P,

MKEquation:PWV=Eh02ρR0, [1a]
HughesEquation:E=E0exp(ζP), [1b]

where E, h0, and R0 are the elastic (tangent) modulus at blood pressure P and thickness and radius of the artery, respectively, ρ is the blood density, E0 is the elastic modulus at zero blood pressure, and ζ is a material coefficient of the artery. As the blood pressure P increases, the artery stiffens (i.e., increase of the tangent modulus E based on Eq. 1b), leading to an increase in PWV according to Eq. 1a. The MK Equation [1a] involves two assumptions: (i) The artery wall is thin such that it can be modeled as a thin shell, and (ii) the thickness and radius of the artery remain fixed as the blood pressure changes. For human arteries, however, these two assumptions may not hold, since the thickness-to-radius ratio h0/R0 = 0.08 to ∼0.22 (19) is beyond the limit h0/R0 < 0.05 (20) for a thin shell, and the change of the radius of a human artery can reach ∼30% due to blood pressure (19, 21). In addition, the Hughes Equation [1b] is completely empirical, without any theoretical foundation.

This paper aims to establish a relation between the PWV and blood pressure without the two assumptions involved in the MK Equation [1a], nor the empirical Hughes Equation [1b], which is replaced by the linear or nonlinear constitutive models for the artery. The results are validated by in vitro experiments using thin walled tubes of polydimethylsiloxane (PDMS), as a linear elastic material, for an artificial blood vessel. For human artery, which is well represented by the Fung hyperelastic model (21), the newly established relation between the PWV and blood pressure is much more accurate than the MK + Hughes Equations, leading to an improved understanding of the connections between blood pressure and PWV, with relevance in continuous, cuffless, and noninvasive blood pressure monitoring.

Results

The schematic diagrams in Fig. 1 show the pulse wave propagation in a human artery. The disturbances caused by beating of the heart propagate as waves along the artery at a finite velocity. For a long and straight tube (artery) containing incompressible and nonviscous blood, the PWV is related to P, the inner area of the artery (A), and ρ by (19)

PWV=AρdPdA. [2]

Fig. 1 B and C illustrates the cross-section of the artery before (initial thickness h0 and radius of R0) and after (thickness h and radius of R) the deformation induced by the blood pressure. Assumption ii in the MK Equation stated above gives the inner area of the artery fixed at A=πR02. The two assumptions i and ii in the MK Equation, together with the equilibrium of force in the artery wall, yield dP/dA=Eh0/(2πR03), and its substitution in Eq. 2 leads to the MK Equation [1a]. In the following, a PA relation is established without the two assumptions in the MK Equation.

Fig. 1.

Fig. 1.

(A) Schematic diagram of pulse wave propagation in a human artery. (B and C) The cross-sectional dimensions of the artery (B) before and (C) after deformation due to the blood pressure.

Equilibrium of force in the artery wall in the cylindrical coordinates {r,θ,z} along the artery wall requires

dσrrdr+1r(σrrσθθ)=0, [3]

where the stresses σrr and σθθ in the radial and circumferential directions are no longer uniform after assumption i in the MK Equation is relaxed; they are related to the corresponding strains via a constitutive model, such as the Fung hyperelastic model (21) for the human artery. After assumption ii in the MK Equation is relaxed, the (logarithmic) strains are obtained in terms of the (change of) inner area of the artery A. The pressure (P) can be obtained by integrating Eq. 3 from the inner radius r=R to the outer radius r=R+h after the deformation, i.e.,

P=-P0dσrr=RR+h1r(σθθσrr)dr. [4]

This equation, together with the constitutive model of the artery, gives the relation between P and A (see SI Appendix, Note 1 for details). Its substitution into Eq. 2 yields the relation between the blood pressure (P) and PWV, which is given separately in Eqs. 6, 7, 9, and 10 for the linear elastic model and Fung hyperelastic model of the artery, and is validated by the in vitro experiments.

In Vitro Experiments.

An in vitro hemodynamic simulator is developed, as shown in Fig. 2A, to verify the theory. The simulator includes a pulsatile flow generator, an artificial blood vessel, strain sensors, pressure sensors, and a water reservoir to define the base pressure on the tube. Pressurized water controlled by solenoid valve provides pulsatile flow, while the height of the water reservoir determines the diastolic pressure in the tube. Thin PDMS tubes with various wall thicknesses and elastic properties (controlled by changing the base to curing agent mixing ratio) provide artificial blood vessels with linear elastic properties within the range of deformations studied (Fig. 2B). Two strain sensors made of carbon black-doped PDMS (CB-PDMS) (22) detect the pulses at two different positions along the tube. The time difference Δt between the arrival of the pulse at each sensor position together with the distance L between the two sensors allows calculation of PWV (Fig. 2C). Fig. 2D shows the voltage signal of the two sensors with the distance L, which gives the PWV (23)

PWV=LΔt. [5]

PDMS (base polymer: curing agent = 15:1, 17:1, and 19:1) is used to fabricate the tube for the in vitro experiment. SI Appendix, Fig. S1 shows the storage and loss moduli (E and E) of the 17:1 PDMS measured by dynamic mechanical analysis at frequencies between 0.1 Hz and ∼10 Hz. The dynamic modulus is E2+E2. The moduli of the 15:1, 17:1, and 19:1 PDMS at 10 Hz are 650, 540, and 420 kPa, respectively, for the samples used in the experiment shown in Fig. 2F. SI Appendix, Fig. S2 shows the relation between the true stress and logarithmic strain of the 17:1 PDMS measured by tensile testing, which displays good linearity for strain less than 30%. The tube in the in vitro experiment is, therefore, linear elastic with the modulus of E (i.e., the dynamic modulus E2+E2 at 0 Hz).

Fig. 2.

Fig. 2.

(A) Schematic diagram of the in vitro experimental setup. (B and C) Experimental image of (B) the PDMS tube and (C) the strain sensor. (D) Signals from the two sensors. (E) Comparison of normalized pressure versus normalized PWV for linear elastic behavior of the tube between the present model and the MK Equation. (F and G) Comparison between the present model and the in vitro experiment, without any parameter fitting, for (F) different materials and (G) different tube thickness.

The Relation Between Pressure and PWV for Linear Elastic Tube Walls.

The linear stress−strain relation for the PDMS tubes, together with Eq. 4, gives the relation between the pressure P and inner area A as (see SI Appendix, Note 1 for details)

P=E¯4[dilog(A+AwallA0+Awall)dilog(AA0)]+E¯8[ln(A+AwallA0+Awall)2ln(AA0)2], [6]

where E¯=E/(1ν2) is the plane strain modulus; ν = 0.5 is the Poisson’s ratio for PDMS; A0=πR02 and Awall=π(R0+h0)2πR02 are the inner area of the artery and the area of artery wall, respectively, without pressure; and dilog is the dilogarithm function (24). Substitution of Eq. 6 into Eq. 2 gives the PWV as

PWV=E¯A4ρ[A0A(AA0)lnAA0A0+Awall(A+Awall)(AA0)ln(A+AwallA0+Awall)]. [7]

Eqs. 6 and 7 are parametric equations for the relation between the pulse wave velocity PWV and pressure P; elimination of the intermediate variable A yields the following scaling law between the normalized PWV and pressure P:

PWVE¯ρ=g(PE¯,h0R0), [8]

where g is a nondimensional function shown in Fig. 2E. It is clear that PWV displays a strong dependence on P. For comparison, the MK Equation [1a] predicts a constant PWV (independent of the pressure), and is also shown in Fig. 2E. Fig. 2F indicates that, without any parameter fitting, the relation between PWV and P obtained from Eq. 8 agrees well with the in vitro experiments for 15:1, 17:1, and 19:1 PDMS and fixed R0 = 6.3 mm, h0 = 0.63 mm, and ρ = 1,000 kg/m3 for water. The effect of liquid viscosity is shown in SI Appendix, Note 2 and Fig. S3. Similarly, Fig. 2G shows excellent agreement with experimental results for two thicknesses (h0 = 0.63 and 0.29 mm) of the tube made of 19:1 PDMS and fixed R0 = 6.3 mm, and ρ = 1,000 kg/m3, without any parameter fitting. The experimental data all display strong dependence on the pressure, which clearly do not support the MK + Hughes Equations.

The Relation Between Blood Pressure and PWV for Human Artery Walls.

The human artery walls are well characterized by the Fung hyperelastic model (21), which has the strain energy density

W=C2ea1Eθθ2+a2Ezz2C2, [9]

where Eθθ and Ezz are the Green strains in the circumferential and axial directions of the artery, respectively, and a1, a2, and C are the material parameters, which are related to the elastic modulus (at zero pressure) by E0=Ca1. Following the same analysis, but with the linear elastic model replaced by the Fung hyperelastic model for human arteries, yields parametric equations for the relation between the pulse wave velocity and pressure, similar to Eqs. 6 and 7, as (see SI Appendix, Note 1 for details)

P=14Cea2Ezz2πa1{erfi(AA02A0a1)erfi[AA02(A0+Awall)a1]}, [10]
PWV=Cea2Ezz2a1A4ρ[1A0ea1(AA0)24A021A0+Awallea1(AA0)24(A0+Awall)2]. [11]

where erfi is the imaginary error function (25). Elimination of the intermediate variable A in Eqs. 10 and 11 yields the following scaling law between the normalized pulse wave velocity PWV and blood pressure P:

PWVCea2Ezz2ρ=f(PCea2Ezz2,a1,h0R0), [12]

where f is a nondimensional function, and is shown in Fig. 3A for a1 = 0.97 (26) and h0/R0 = 0.15 (19) for the human artery. Fig. 3B examines the effect of artery stretching Ezz by comparing the limit Ezz = 0 of Eq. 12, which takes the form

PWVCρ=f(PC,a1,h0R0), [13]

to the scaling law in Eqs. 10 and 11 for a representative a2 = 2.69 (21) and Ezz = 0.1 and 0.2. The effect of artery stretching is negligible even for 20% stretching.

Fig. 3.

Fig. 3.

(A) Normalized blood pressure P versus normalized PWV for the human artery characterized by the Fung hyperelastic model. (BD) Normalized P versus normalized PWV for (B) different axial stretching of the artery, (C) different thickness-to-radius ratio h0/R0 of the artery, and (D) different SD σ for a normal distribution of h0/R0.

The scaling law in Eq. 13 degenerates to the MK Equation [1a] in the limit of low blood pressure, which gives AA0; therefore ea1(AA0)2/(4A02)=1 and ea1(AA0)2/[4(A0+Awall)2]=1. Eq. 11, at the limit Ezz = 0, then becomes PWV=Ca1Awall/[4ρ(A0+Awall)], which is identical to the MK Equation [1a] for a thin artery wall [i.e., Awall/(A0+Awall)=2h0/R0] at zero blood pressure.

The artery thickness, in general, is not a constant even for the same human artery; the thickness-to-radius ratio h0/R0 has an average of 0.15 and a variation of 40% (19). Fig. 3C shows the normalized pressure P/C versus PWV/C/ρ for h0/R0 = 0.09, 0.12, 0.15, 0.18, and 0.21, corresponding to ±20% and ±40% variations of h0/R0 = 0.15. Even for 40% variations, the curves in Fig. 3C are different by only ∼6%. For a1 = 0.97 and a normal distribution of h0/R0 with the mean 0.15 and SD σ, the mean PWV is obtained as

PWVCρ=f[PC,a1=0.97,h0R0N(0.15,σ2)], [14]

and is shown in Fig. 3D for several values of σ. The curve based on the mean h0/R0 gives an accurate relation between the PWV and blood pressure.

Fig. 4 A and B compares the present model (in Eq. 13) to the classical MK + Hughes Equations (in Eqs. 1a and 1b) for a human artery characterized by the Fung hyperelastic model with C = 39 kPa, a1 = 0.97, and h0/R0 = 0.15 (19, 26). The arterial stiffness, or the equivalent tangent modulus E, is shown in Fig. 4A versus the blood pressure P. In the range of human blood pressure (5 kPa to ∼20 kPa), the arterial stiffness is used to determine the material parameters in the Hughes Equation [1b] as E0 = 563 kPa and ζ = 0.121 kPa−1, which yields good agreement between the Hughes Equation and the present model. However, for the same range of blood pressure, Fig. 4B shows that the MK + Hughes Equations overestimate the PWV by a factor of ∼2 compared with the present model. This large discrepancy results from the large change of radius and thickness of the artery wall (>50%), which is neglected in the MK Equations (due to assumption ii) but is accounted for in the present model.

Fig. 4.

Fig. 4.

(A) The arterial stiffness (equivalent modulus) E versus the blood pressure P for a human artery characterized by the Fung hyperelastic model; the Hughes Equation is also shown, where its parameters E0 and ζ are determined by fitting the arterial stiffness within the range of human blood pressure (5 kPa to ∼20 kPa). (B) The blood pressure P versus the PWV of the human artery, given by the present model and by the MK + Hughes Equations, where the parameters E0 and ζ are determined from A. (C) The blood pressure P versus the PWV for the human artery characterized by the Fung hyperelastic model; the MK + Hughes Equations are also shown, where the parameters E0 and ζ in the Hughes Equation are determined by fitting within the range of human blood pressure (5 kPa to ∼20 kPa). (D) The artery stiffness (equivalent modulus) E versus the PWV of the human artery, given by the present model and by the Hughes Equation, where the parameters E0 and ζ are determined from C.

Another important clinical application of PWV is to determine the arterial stiffness (equivalent tangent modulus) of the artery wall as the elastic properties of arteries are affected by aging and cardiovascular diseases, therefore providing useful prognostic information (27). The blood pressure P is shown in Fig. 4C versus the PWV. In the range of human blood pressure (5 kPa to ∼20 kPa), the pressure−PWV relation is used to determine the material parameters in the Hughes + MK Equations (Eqs. 1a and 1b) as E0 = 145 kPa and ζ = 0.117 kPa−1, which yields good agreement between the MK + Hughes Equations and the present model. However, for the same range of PWV, Fig. 4D shows that the MK + Hughes Equations significantly underestimate the equivalent tangent modulus by a factor of ∼3 compared with the present model. The main reason for this large discrepancy is the same as that shown in Fig. 4 A and B.

For the human artery characterized by the Fung hyperelastic model with C = 39 kPa, a1 = 0.97, and h0/R0 = 0.15 (19, 26), the range of human blood pressure (5 kPa to ∼20 kPa) gives A/A0 = 2.46 to ∼3.55, which is relatively large such that the function erfi in Eq. 10 can be approximated by erfi(x)ex2/(πx) (28). Eqs. 10 and 11, at the limit Ezz = 0, have the asymptotes for large A/A0 (see SI Appendix, Note 3 for details),

PC2A0AA0ea1(AA0)24A02, [15]
PWV2Ca14ρAA0ea1(AA0)24A02. [16]

Eliminating the variable A yields the following relation (see SI Appendix, Note 3 for details):

lnPC+ln(1+8ρa1PWV2P1)=a116(1+8ρa1PWV2P1)2, [17]

which, as shown in Fig. 5A, is in reasonable agreement with the P versus PWV curve at large pressure, and their difference is approximately a constant (i.e., a shift along the vertical axis). The above equation suggests that the blood pressure scales with PWV2, i.e., PαPWV2, and the scaling coefficient α is approximately a constant since the logarithmic term ln(P/C) has a very weak dependence on the pressure. Accordingly, the relation between P and PWV can be represented by

P=αPWV2+β, [18]

where β represents the constant shift between the two curves in Fig. 5A, and α and β depend on the material properties and geometry of the artery (C, a1, ρ, R0, and h0) and are to be determined from the experiments.

Fig. 5.

Fig. 5.

(A) The normalized blood pressure P versus the normalized PWV given by the present model (in Eqs. 10 and 11) and its asymptote (in Eq. 17). (B and C) Comparison of P=αPWV2+β to (B) the present model (Eqs. 10 and 11) and (C) literature data (29). (D) Comparison of P=αL2/Δt2+β to the experiments of blood pressure P versus 1/Δt, where Δt is pulse arrival time.

For the human artery characterized by the Fung hyperelastic model with C = 39 kPa, a1 = 0.97, and h0/R0 = 0.15 (19, 26), the constants are α = 0.18 kPa·s2⋅m2 and β = 2.7 kPa, which show excellent agreement (Fig. 5B) with the P versus PWV relation obtained from Eqs. 10 and 11 in the range of human blood pressure (5 kPa to ∼20 kPa). Fig. 5C further compares Eq. 18 with literature data (29) of the experimental diastolic blood pressure (DBP, measured by an invasive method) versus PWV (obtained from the ear and toe pulses) during and after anesthesia for surgery. For α = 0.046 kPa·s2⋅m2 and β = 5.1 kPa, Eq. 18 agrees reasonably well with the experimental data.

SI Appendix, Fig. S4A, Inset shows a multimodal wearable sensor called the BioStamp (MC10 Inc.) (30), which contains electrodes and an optical sensor for collection of electrocardiogram (ECG) (31) and photoplethysmograph (PPG) (32) signals simultaneously. The BioStamp mounts on the torso in the subclavicular region (SI Appendix, Fig. S4A) or, alternatively, on the posterior side overlaying the scapula (SI Appendix, Fig. S4B). These two positions allow collection of ECG and PPG signals concurrently, and exploit the temporal relationship between these two signals to compute pulse arrival times (PVW=L/Δt: pulse arrival distance, L, pulse arrival time, Δt). Evaluations of the correlations between these Δt measurements and blood pressure rely on episodic measurements of blood pressure with a conventional cuff device with the subject instrumented with the BioStamp. The Δt and DBP versus time are measured during the postexercise period. The 1/Δt and DBP both decrease during the first 300 s (after sprinting), then approach a stable resting value as shown in SI Appendix, Fig. S5. Their relation can be described by substituting Eq. 5 into Eq. 18,

P=αL2Δt2+β. [19]

The 1/Δt and DBP response during the first 300 s are shown in Fig. 5D, together with Eq. 19 and αL2 = 0.064 kPa·s2 and β = 5.8 kPa, which shows reasonable agreement with the experimental data. In general, once β and α in Eq. 18 (or αL2 in Eq. 19) are determined, then the continuous, cuffless, and noninvasive blood pressure can be monitored by measuring the PWV.

Conclusions

This paper establishes a relation between the blood pressure P and pulse wave velocity PWV that does not rely on the Hughes Equation or on assumptions in the MK Equation. This relation degenerates to the MK Equation in a regime of extremely low blood pressures. An in vitro hemodynamic simulator is developed to collect PWV and pressure data using liquid flow through a PDMS (with linear stress−strain relation) tube. These in vitro experiments show that the PWV depends strongly on pressure, unlike expectations based on the MK equation but in excellent, quantitative agreement with the newly established relation without any parameter fitting. For human arteries, which are well characterized by the Fung hyperelastic model, a simple formula P=αPWV2+β is established within the range of human blood pressure. This formula is validated by literature data as well as by experiments on human subjects, and can be used to determine the blood pressure from the measured PWV in continuous, cuffless, and noninvasive blood pressure monitoring.

Methods

Hemodynamic Simulator.

A pressurized bottle filled with water driven by a 12-V solenoid valve (Adafruit Industries) produced pulse wave flow by opening of a given repeatable pressure in the water reservoir. Two factory-calibrated pressure sensors (HHP886; OMEGA Engineering) with measurement accuracy of ±1.5% were located before strain sensor #1 and after strain sensor #2 as shown in Fig. 2A. Strain sensors were placed on the surface of a PDMS tube at a specific distance. Resistance difference during pulse wave by data acquisition system at 1-kHz sampling rate (Powerlab 8/35; ADInstruments) provided the detection of the peak of resistance change from strain sensors. The water reservoir controlled the diastolic pressure in the tube by adjusting the water height. The water generated from the pulse generator flowed out from the tube to maintain the pressure in the tube before and after pulse generation.

Fabrication of Thin CB-PDMS Strain Sensor.

Spin coating 30:1 PDMS (Sylgard 184 Silicone Elastomer; Dow Corning) at 1,000 rpm on a Si wafer generated a substrate. Spin coating polyimide (PI2545; HD Microsystems) at 3,000 rpm for 30 s followed by baking at 110 °C for 1 min, 150 °C for 4 min, and 250 °C for 5 min produced a thin layer. Spin coating AZ4620 (AZ Electronic Materials) at 2,000 rpm for 30 s and developing generated a mold for the strain sensor. Fabrication of thin CB-PDMS began with mixing 25 wt% carbon black (VULCAN XC72R; Cabot Corporation) and 30:1 PDMS. A doctor blade method formed a thin CB-PDMS layer in the opening region in the mold. After baking at 70 °C for 2 h, immersion in acetone removed the photoresist to leave only the patterned CB-PDMS on a PI layer. Spin coating and baking at 70 °C for 2 h of 30:1 PDMS generated a uniform encapsulation layer.

Fabrication of Thin PDMS Tube.

Pouring a precursor to PDMS (Sylgard 184 Silicone Elastomer; Dow Corning) with a specific mixing ratio into the inside of clear poly(ethylene terephthalate)-glycol tube with 12.7-mm inner diameter (McMaster-Carr) to cover all of the inside of the tube and curing it overnight while held in a vertical position at room temperature at 20 °C generated one layer of thin PDMS. The tube was then reversed before pouring a second layer to reach an approximately homogeneous thickness along the length of the tube (thickness variation is less than 10%). The number of repetitions of this process determined the thickness of the tubing. A week of additional curing process reached a stable state of the PDMS in terms of elastic properties.

Measurement of Elastic Properties.

Elastic properties of each PDMS tube were measured using a RSA3 dynamic mechanical analyzer, within a few hours after the pulse wave velocity measurement to avoid any aging effect.

Supplementary Material

Supplementary File
pnas.1814392115.sapp.pdf (927.5KB, pdf)

Acknowledgments

Y.M. and X.F. acknowledge support from the National Basic Research Program of China Grant 2015CB351900 and National Natural Science Foundation of China Grants 11402135 11625207, 11320101001, and 11222220. Y.H. acknowledges support from National Science Foundation Grants 1400169, 1534120, and 1635443.

Footnotes

The authors declare no conflict of interest.

This article contains supporting information online at www.pnas.org/lookup/suppl/doi:10.1073/pnas.1814392115/-/DCSupplemental.

References

  • 1.Lewington S. Prospective studies collaboration. Age-specific relevance of usual blood pressure to vascular mortality: A meta-analysis of individual data for one million adults in 61 prospective studies. Lancet. 2002;360:1903–1913, and erratum (2003) 361:1060. doi: 10.1016/s0140-6736(02)11911-8. [DOI] [PubMed] [Google Scholar]
  • 2.Méndez AS, et al. Risk factors for orthostatic hypotension: Differences between elderly men and women. Am J Hypertens. 2018;31:797–803. doi: 10.1093/ajh/hpy050. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Wright JT, Jr, et al. SPRINT Research Group A randomized trial of intensive versus standard blood-pressure control. N Engl J Med. 2015;373:2103–2116. doi: 10.1056/NEJMoa1511939. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Lewington S, et al. China Kadoorie Biobank Consortium The burden of hypertension and associated risk for cardiovascular mortality in China. JAMA Intern Med. 2016;176:524–532. doi: 10.1001/jamainternmed.2016.0190. [DOI] [PubMed] [Google Scholar]
  • 5.McLaughlin AR. A modified erlanger sphygmomanometer. Science. 1928;67:72–73. doi: 10.1126/science.67.1725.72. [DOI] [PubMed] [Google Scholar]
  • 6.Punzi HA. Why ambulatory blood pressure monitoring? Am J Health Syst Pharm. 1998;55:S12–S16. doi: 10.1093/ajhp/55.suppl_3.S12. [DOI] [PubMed] [Google Scholar]
  • 7.Ukawa T, et al. Novel non-invasive method of measurement of endothelial function: Enclosed-zone flow-mediated dilatation (ezFMD) Med Biol Eng Comput. 2012;50:1239–1247. doi: 10.1007/s11517-012-0955-z. [DOI] [PubMed] [Google Scholar]
  • 8.Chandrasekhar A, et al. Smartphone-based blood pressure monitoring via the oscillometric finger-pressing method. Sci Transl Med. 2018;10:eaap8674. doi: 10.1126/scitranslmed.aap8674. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Pickering TG, et al. Recommendations for blood pressure measurement in humans and experimental animals: Part 1: Blood pressure measurement in humans: A statement for professionals from the subcommittee of professional and public education of the American Heart Association council on high blood pressure research. Circulation. 2005;111:697–716. doi: 10.1161/01.CIR.0000154900.76284.F6. [DOI] [PubMed] [Google Scholar]
  • 10.Fuke S, Suzuki T, Nakayama K, Tanaka H, Minami S. Blood pressure estimation from pulse wave velocity measured on the chest. Conf Proc IEEE Eng Med Biol Soc. 2013;2013:6107–6110. doi: 10.1109/EMBC.2013.6610946. [DOI] [PubMed] [Google Scholar]
  • 11.Berkelmans GFN, Kuipers S, Westerhof BE, Spoelstra-de Man AME, Smulders YM. Comparing volume-clamp method and intra-arterial blood pressure measurements in patients with atrial fibrillation admitted to the intensive or medium care unit. J Clin Monit Comput. 2018;32:439–446. doi: 10.1007/s10877-017-0044-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Scheer B, Perel A, Pfeiffer UJ. Clinical review: Complications and risk factors of peripheral arterial catheters used for haemodynamic monitoring in anaesthesia and intensive care medicine. Crit Care. 2002;6:199–204. doi: 10.1186/cc1489. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Peter L, Noury N, Cerny M. A review of methods for non-invasive and continuous blood pressure monitoring: Pulse transit time method is promising? IRBM. 2014;35:271–282. [Google Scholar]
  • 14.Sharma M, et al. Cuff-less and continuous blood pressure monitoring: A methodological review. Technologies. 2017;5:21. [Google Scholar]
  • 15.Boutry CM, et al. A sensitive and biodegradable pressure sensor array for cardiovascular monitoring. Adv Mater. 2015;27:6954–6961. doi: 10.1002/adma.201502535. [DOI] [PubMed] [Google Scholar]
  • 16.Katsuura T, et al. 2017. Wearable pulse wave velocity sensor using flexible piezoelectric film array. 2017 IEEE Biomedical Circuits and Systems Conference (BioCAS) (Inst Electr Electron Eng, New York), abstr 7208.
  • 17.Bramwell JC. The velocity of the pulse wave in man. Proc R Soc B. 1922;93:298–306. [Google Scholar]
  • 18.Hughes DJ, Babbs CF, Geddes LA, Bourland JD. Measurements of Young’s modulus of elasticity of the canine aorta with ultrasound. Ultrason Imaging. 1979;1:356–367. doi: 10.1177/016173467900100406. [DOI] [PubMed] [Google Scholar]
  • 19.Fung YC. Biomechanics: Circulation. Springer; New York: 1997. [Google Scholar]
  • 20.Timoshenko S. Theory of Plates and Shells. McGraw-Hill; London: 1940. [Google Scholar]
  • 21.Fung YC. Biomechanics: Mechanical Properties of Living Tissues. Springer; New York: 1993. [Google Scholar]
  • 22.Lu NS, Lu C, Yang SX, Rogers J. Highly sensitive skin-mountable strain gauges based entirely on elastomers. Adv Funct Mater. 2012;22:4044–4050. [Google Scholar]
  • 23.Dagdeviren C, et al. Conformable amplified lead zirconate titanate sensors with enhanced piezoelectric response for cutaneous pressure monitoring. Nat Commun. 2014;5:4496. doi: 10.1038/ncomms5496. [DOI] [PubMed] [Google Scholar]
  • 24.Maximon LC. The dilogarithm function for complex argument. Proc R Soc A. 2003;459:2807–2819. [Google Scholar]
  • 25.Morris RM, Leach PGL. Symmetry reductions and solutions to the Zoomeron equation. Phys Scr. 2015;90:015202. [Google Scholar]
  • 26.Chamiot-Clerc P, Copie X, Renaud JF, Safar M, Girerd X. Comparative reactivity and mechanical properties of human isolated internal mammary and radial arteries. Cardiovasc Res. 1998;37:811–819. doi: 10.1016/s0008-6363(97)00267-8. [DOI] [PubMed] [Google Scholar]
  • 27.Laurent S, et al. European Network for Non-invasive Investigation of Large Arteries Expert consensus document on arterial stiffness: Methodological issues and clinical applications. Eur Heart J. 2006;27:2588–2605. doi: 10.1093/eurheartj/ehl254. [DOI] [PubMed] [Google Scholar]
  • 28.Acton FS. Numerical Methods That Work. Harper Row; Washington, DC: 1970. [Google Scholar]
  • 29.Chen Y, Wen C, Tao G, Bi M. Continuous and noninvasive measurement of systolic and diastolic blood pressure by one mathematical model with the same model parameters and two separate pulse wave velocities. Ann Biomed Eng. 2012;40:871–882. doi: 10.1007/s10439-011-0467-2. [DOI] [PubMed] [Google Scholar]
  • 30.Liu Y, et al. Intraoperative Monitoring of neuromuscular function with soft, skin-mounted wireless devices. npj Digital Med. 2018;1:19. doi: 10.1038/s41746-018-0023-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Jang KI, et al. Self-assembled three dimensional network designs for soft electronics. Nat Commun. 2017;8:15894. doi: 10.1038/ncomms15894. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Li H, et al. Epidermal inorganic optoelectronics for blood oxygen measurement. Adv Healthc Mater. 2017;6:1601013. doi: 10.1002/adhm.201601013. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary File
pnas.1814392115.sapp.pdf (927.5KB, pdf)