Feasibility of Using a Leadless Patch Monitor in Community Cohort Studies: The Multiethnic Study of Atherosclerosis

Michael J Zhang a, Nicholas S Roetker b, Aaron R Folsom b, Alvaro Alonso c, Susan R Heckbert d, Lin Y Chen a
PMCID: PMC6219902  NIHMSID: NIHMS989367  PMID: 30225944

Asymptomatic and undetected “silent” atrial fibrillation (AF) is an important public health problem because it is often detected only after a stroke 1 and its prevalence in the general population is not precisely known2. The Zio Patch (Zio) is a simple device that can continuously record heart rhythm for 14 days3. Although it is most commonly applied in clinic, data are sparse on the reliability of patient self-application. We therefore conducted a randomized controlled trial to evaluate whether a community-based cohort can self-apply the Zio, thus reducing the burden of an in-person visit.

The Multi-Ethnic Study of Atherosclerosis (MESA) 4 is a cohort study of 6,814 men and women from six US communities evaluating progression of subclinical cardiovascular disease. We enrolled 45 participants older than 65 (Table 1) from the University of Minnesota field center.

Table 1.

Baseline Characteristics.

All
N=45
Group 1
(self-
application)
N=30
Group 2
(clinic-
application)
N=15
Age, (years) (SD) 71.6 (6.2) 72.2 (5.9) 70.3 (6.8)
Female sex 19 (42) 11 (37) 8 (53)
Hispanic ethnicity* 14 (31) 9 (30) 5 (33)
Hypertension 23 (51) 15 (50) 8 (53)
Diabetes 8 (18) 6 (20) 2 (13)
Coronary heart disease 2 (4) 2 (7) 0 (0%)

Data are presented as number (%) of participants unless otherwise stated.

SD, standard deviation

*

Only Hispanic and non-Hispanic white participants were enrolled at the MN field center

We randomized 30 participants to self-application (Group 1) using written instructions and as-needed telephone assistance and 15 participants to in-office application by MESA staff (Group 2). We compared the Zio data using wear time (from device activation to the last recorded analyzable signal) and analyzable time fraction (proportion of total wear time that the ECG signal was sufficiently free of noise to be interpretable).

Table 2 displays the wear time and analyzable time fraction for the two study groups.

Table 2.

Performance Metrics of Zio Patch in MESA. Participants were randomized to self-application vs. clinic-application. Group 1 participants self-applied the Zio Patch. Group 2 participants received Zio Patches in-office.

Group 1
N=30
Group 2
N=15
p-value
Wear time (days)
 Mean (SD) 13.29 (2.27) 13.44 (1.06) 0.76
 Median (25th, 75th percentile) 13.86 (13.70, 13.99) 13.89 (13.09, 13.99) 0.80
 ≥48 hours 97% 100%
 ≥6 days 97% 100%
 ≥13 days 93% 80% 0.31§
Analyzable time fraction
  Mean (SD) 94% (12%) 97% (4%) 0.11
 Median (25th, 75th percentile) 98% (94%, 99%) 99% (98%, 100%) 0.07

Differences tested using two-sample t-tests

Wilcoxon rank sum tests

§

Fisher’s exact test

There was no statistical difference in mean wear time or analyzable time fraction between the Group 1 and 2. Although we excluded patients with previously documented AF, we nevertheless diagnosed two participants (4%) with AF. The most common adverse reaction was skin irritation; three participants removed their Zio as a result, and one participant self-administered diphenhydramine and left the Zio in place.

Our results suggest that self-application of the Zio is equivalent to in-office application. It also suggests it is feasible to conduct community-based cohort studies involving ambulatory event monitoring with self-applied Zio to detect silent AF in the general population. Because the Zio is a small, leadless, self-contained device with straightforward installation, the likelihood of a patient successfully self-applying the Zio is inherently high.

The primary limitation of this study was its small sample size. We only enrolled participants from a single MESA study site; it also did not encompass all the ethnicities represented in the entire cohort. Finally, our study did not have the power to correlate success of self-application with specific functional capacity of the participant. Thus, future studies with larger sample sizes can help identify specific patient factors that predict successful Zio self-application.

The overall implication of this study is that participants in community-based cohort studies can apply the Zio by themselves at home. If ever new clinical data emerge or widespread consensus forms to support screening for AF 5, self-application of the Zio is a potential screening method.

Acknowledgements:

We thank Jacqueline Munoz for her assistance with coordinating this study.

Sources of Funding: Alvaro Alonso is supported by American Heart Association grant 16EIA26410001. Susan R. Heckbert is supported by NIH grant R01HL127659. Lin Y. Chen is supported by NIH grant R01HL126637 and R01HL141288.

Footnotes

Disclosures: No conflicts of interest to disclose. The views expressed in this manuscript are those of the authors and do not necessarily represent the views of the National Heart, Lung, and Blood Institute; the National Institutes of Health; or the U.S. Department of Health and Human Services.

References

  • 1.Sposato LA, Cipriano LE, Saposnik G, Ruiz Vargas E, Riccio PM, Hachinski V. Diagnosis of atrial fibrillation after stroke and transient ischaemic attack: a systematic review and meta-analysis. Lancet Neurol 2015;14(4):377–387. [DOI] [PubMed] [Google Scholar]
  • 2.Barbarossa A, Guerra F, Capucci A. Silent Atrial Fibrillation: A Critical Review. J Atr Fibrillation. 2014;7(3):1138. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Turakhia MP, Hoang DD, Zimetbaum P, et al. Diagnostic utility of a novel leadless arrhythmia monitoring device. Am J Cardiol 2013;112(4):520–524. [DOI] [PubMed] [Google Scholar]
  • 4.Bild DE, Bluemke DA, Burke GL, et al. Multi-Ethnic Study of Atherosclerosis: objectives and design. Am J Epidemiol 2002;156(9):871–881. [DOI] [PubMed] [Google Scholar]
  • 5.Svennberg E, Engdahl J, Al-Khalili F, Friberg L, Frykman V, Rosenqvist M. Mass Screening for Untreated Atrial Fibrillation: The STROKESTOP Study. Circulation. 2015;131(25):2176–2184. [DOI] [PubMed] [Google Scholar]

RESOURCES