Abstract
To overcome long acquisition times of Chemical Shift Imaging (CSI), a new Magnetic Resonance Spectroscopic Imaging (MRSI) technique called Reduction of Acquisition time by Partition of the sIgnal Decay in Spectroscopic Imaging (RAPID-SI) using blipped phase encoding gradients inserted during signal acquisition was developed. To validate the results using RAPID-SI and to demonstrate its usefulness in terms of acquisition time and data quantification; simulations, phantom and in vivo studies were conducted, and the results were compared to standard CSI. The method was based upon the partition of a magnetic resonance spectroscopy (MRS) signal into sequential sub-signals encoded using blipped phase encoding gradients inserted during signal acquisition at a constant time interval. The RAPID-SI technique was implemented on a clinical 3 T Siemens scanner to demonstrate its clinical utility. Acceleration of data collection was performed by inserting R (R = acceleration factor) blipped gradients along a given spatial direction during data acquisition. Compared to CSI, RAPID-SI reduced acquisition time by the acceleration factor R. For example, a 2D 16x16 data set acquired in about 17 min with CSI, was reduced to approximately 2 min with the RAPID-SI (R = 8). While the SNR of the acquired RAPID-SI signal was lower compared to CSI by approximately the factor √R, it can be improved after data pre-processing and reconstruction. Compared to CSI, RAPID-SI reduces acquisition time, while preserving metabolites information. Furthermore, the method is flexible and could be combined with other acceleration methods such as Parallel Imaging.
Introduction
Chemical Shift Imaging (CSI) [1], a technique currently available as a clinical imaging tool, is used to provide tissue metabolite maps in vivo to help characterize neurological and metabolic disease, and improve tumour treatment [2,3].
However, CSI suffers from the time cost in obtaining these metabolite maps, and its impact on the spatial resolution and Signal to Noise Ratio (SNR) [2,3]. To overcome this problem and expand its clinical use, significant effort has been made to promote high-resolution Magnetic Resonance Spectroscopic Imaging (MRSI) techniques, using fast sequences and advanced data reconstruction methods [2–8].
These methods could be grouped into techniques which use only phase encoding gradients to obtain metabolite maps such as Multi-Spin-Echo Proton Spectroscopic Imaging technique [4,5], those using the readout gradients like echo planar spectroscopic imaging (EPSI) [6], manipulations of the shapes of the RF pulses such as wavelet encoding spectroscopic imaging technique (WE-SI) [7], or exploration of the correlation between spatial and spectral information like SPICE [8]. To further reduce acquisition time, some of the techniques cited above (e.g. WE-SI, EPSI) have been combined with Parallel Imaging (PI) [9,10]. Compressed Sensing (CS) has also been combined with blipped gradients and sophisticated post-processing algorithms to accelerate data acquisition in non-proton MRSI sequences [11].
The Reduction of Acquisition time by Partition of the sIgnal Decay in Spectroscopic Imaging (RAPID-SI) technique accelerates data acquisition by inserting a series of blipped gradients during signal acquisition without the usage of the readout gradients. In this approach, the acquired magnetic resonance spectroscopy (MRS) signal is partitioned into sequential sub-signals with short time duration, each representing one k-space data point and carrying the necessary chemical shift information. However, due to the short time duration of these sequentially acquired sub-signals, both SNR and spectral resolution are reduced. To overcome this problem, the short time encoded sub-signals are extended to full time MRS signals using a pre-processing approach described in details in the theory section.
The proposed method, implemented on a 3T Siemens whole body clinical scanner, was first demonstrated on simulated data, then validated on phantoms and in vivo. The method was compared to CSI to test its usefulness in terms of acquisition time, spectral resolution and data sensitivity.
This paper is organized as follows: The first section describes the technique, followed by materials and methods section describing the simulations performed along with the acquisition of in vitro and in vivo data with the scanner configurations. The results from the simulated and real MRS data are discussed and compared to those acquired using the CSI technique. Strengths and limitations of the method as compared with other methods are summarized in the discussion section, followed by a closing remark.
Theory
Pulse sequence
The RAPID-SI technique acquires a predetermined number of localized spin echo MRS signals (AS). The decaying part of each acquired AS signal is partitioned into R successively encoded sub-signals to fill R subsequent k-space data points. The encoding is performed by insertion of R equally spaced blipped phase encoding gradients during data acquisition as shown in Fig 1. The k-space data collection accelerated by the factor R is given by:
(1) |
where the k-space variable k , with τG being the gradient blip duration) runs from to (PE is the total number of phase encoding steps, FOV is the field of view), and is the signal acquired at voxel position x in the MRSI grid during the time window tj (j runs from 1 to R), having constant duration set to the number of data points Nsub = Nfull/R (Nfull is the total number of points of the MRS signal). The time variable tj runs from 1+(j−1).Nsub to j.Nsub. The variable l runs from 0 to (AS−1), and AS is set to PE/R. The signal is given by:
(2) |
The parameter M is the number of the spectral peaks, assumed to be Lorentzian. denotes the amplitude of metabolite m at voxel position x in the MRSI grid, whereas fm, and dm, are the corresponding frequency and damping factor, respectively. The term ε(t) describes noise. Inserting Eq 2 into Eq 1, we obtain:
(3) |
where , is the phase modulated amplitude, dependent on k, and therefore specific to each s(k(l*R)+j,tj), and the constant part, , carrying the parameters dm, and fm of the spectral component m during the time window tj.
The proposed method is comparable to CSI when the number of inserted blips is one per acquired signal (R = 1). It reaches, however, its full speed when all the necessary blips are played within a single acquisition (R = PE and AS = 1). Therefore, the reduction in acquisition time is proportional to R.
The Signal to Noise Ratio per voxel of the CSI in one dimension is given by [2]:
(4) |
where A denotes the peak amplitude of a metabolite. Nx is the number of voxels along x direction, Δf is the bandwidth of the MRS spectrum, and a being a constant term contributing to the noise standard deviation [2]. The term is related to the duration of the acquisition window. Given that acceleration is performed in the one given direction (e.g. x), one can deduce the SNR per voxel of the 1D RAPID-SI method accordingly:
(5) |
which can be rewritten as:
(6) |
Combining Eq 4 and Eq 6, we obtain:
(7) |
The sensitivity of RAPID-SI in terms of SNR and acquisition time, given by [2] with Ttot being the total acquisition time, is comparable to the CSI (ΨRAPID = ΨCSI). One can increase the SNR in RAPID-SI by shortening the number of points Nfull, which could be set to the shortest T2 value of the collected metabolite signals (e.g. Nfull ≅ T2,shortest). However, reducing Nfull, lowers the acceleration factor R since a minimum data length of Nsub is required to allow for an accurate estimation of the phase modulated amplitudes . Therefore, to allow for higher acceleration and minimise the SNR loss, Nfull is generally set equal to the number of data points corresponding to the averaged T2 value of the metabolite signals (T2,avg). At the acquisition level, the data has low SNR and low spectral resolution, and sub-signal amplitudes modulated by the decaying effect due to the sequential acquisition of these sub-signals. This sequential data acquisition leads to inaccurate data analysis. To overcome this problem, the sequentially acquired short time encoded sub-signals are extended to full time MRS signals using the pre-processing approach as described below.
HLSVD-PRO based pre-processing
To recover the missing information due to the short time duration of the acquired sub-signals, the latter have to be extrapolated to full time length. We have chosen to use the HLSVD-PRO method to achieve this task [12].
To estimate dm, and fm in ψsub(m,tj) in Eq 3, a reference signal AS0 with full time length (Nfull) is acquired at the centre of k-space (k = 0) and fitted, using HLSVD-PRO, to the corresponding signal model given by:
(8) |
where the term ψfull(m,t) is described by an (Nfull X M) matrix as follows:
(9) |
The generated matrix above in Eq 9, is then time sub-divided into sequentially reduced size (Nsub X M) sub-basis functions ψsub(m,tj) according to the time occurrence tj as follows:
(10) |
Inserting Eq 10 into Eq 3, a noiseless model for the acquired s(k(l*R)+j,tj) is given by:
(11) |
Then, the complex amplitudes in Eq 11 are estimated from the corresponding acquired s(k(l*R)+j,tj) using a pseudo-inverse matrix operation [13]. The estimated amplitudes are used to substitute the amplitude vector in Eq 8 to construct the corresponding full time noiseless s(k(l*R)+j,t = Nfull) signal. Fourier Transform is then performed in both spatial and temporal domain on the PE reconstructed k-space data, including the reference signal AS0 (k = 0,t = Nfull), to obtain the metabolite maps. A schematic representation of the method is shown in Fig 1 and the algorithm used to obtain metabolite maps from the acquired data is displayed in Fig 2.
Materials and methods
Simulation
A two dimensional 16x16 MRSI data shown in Fig 3, was simulated with uniform distribution of brain metabolites like Choline (Cho), Creatine (Cr), N-acetylaspartate (NAA), Myo-Inositol (mI) and lactate (lac) using Lorentzian line shape with T2 (inversely proportional to damping factor (dm)), and resonance frequencies (fm) obtained from literature [14–18], according to Eq 2 using four singlet peaks for Cho, Cr, NAA, mI and a doublet for lactate. The amplitudes of Cho, Cr, NAA, mI and lactate resonances were set to 0.9, 1.06, 2.23, 0.64, 0.49, respectively. The following acquisition parameters were used: TE = 35 ms, TR = 2000 ms, BW = 2000 Hz, Nfull = 512. A white Gaussian noise was added to the data and varied between different SNR values from 2.5, 4.5, 6, 8, 12, 20, to 30. The SNR is defined as:
(12) |
The B0 field inhomogeneity, replicating in vivo conditions in the human brain was also incorporated into the data [19]. Five ΔB0 spatial maps (B1 –B5) were generated by varying the resonance frequencies fm and linewidth dm of the peaks in Eq 2, across all the voxels in the ROI by a factor ranging from 0 to 1 Hz in B1, 0 to 3 Hz in B2, 0 to 5 Hz in B3, 0 to 8 Hz in B4, 0 to 10 Hz in B5 (Fig 3). The FOV dimensions were set to AP = 100 mm and RL = 100 mm. The acceleration factor R was varied from 2, 4, 8, to 16.
Sequence and hardware
The RAPID-SI sequence was accelerated in the anterior-posterior (AP) direction by inserting gradient blips during data acquisition in a PRESS sequence [20]. Regular phase encoding was applied in the Right-Left (RL) and Head-Foot (HF) directions. Water and lipid suppression were performed using CHESS and outer-volume saturation (OVS) techniques, respectively [21,22]. The sequence was implemented on a 3T Siemens whole body scanner, with a gradient strength of 45 mT/m and a nominal slew rate of 200 mT/m/s. A 32-channel receive head coil was used for data acquisition. The reference signal AS0 was acquired first from the PRESS Region of Interest (ROI) prior to encoding by turning off all the encoding gradients (k = 0,Nfull = 512).
Phantom scans
In vitro water suppressed and unsuppressed 2D RAPID-SI and CSI data were acquired from a cylindrical water solution phantom containing tubes filled with brain metabolite solutions (NAA, lac, Cho and Cr with known concentrations of 10.5 mM, 25 mM, 9.8 mM and 4.5 mM, respectively). Tube 1 contains Cr only, tube 2 contains NAA and lac, tube 3 contains lac and Cho and tube 4 containing lac and mI. The following acquisition parameters were used: TE = 35 ms, TR = 2000 ms, BW = 2000 Hz, Nfull = 512, Navg = 2, and R = 4, 8, 16. The FOV dimensions were AP = 100 mm, HF = 10 mm and RL = 100 mm for 2D acquisitions, with spatial resolution set to 16x16.
Localizer gradient echo images, with 40° flip angle, TR/TE of 20/5 ms, 10 mm thick, 256 x 128 matrix, were acquired prior to spectroscopy for the positioning of the PRESS ROI volume. Shimming was performed using Siemens software FASTMAP [23].
The water line was calculated from the unsuppressed water data and used as an internal reference to estimate the absolute concentrations of the metabolites. The absolute concentrations of the metabolites were estimated using the following equation:
(13) |
and compared to the CSI. The variable Cmetabolite is the metabolite concentration (mM), Cwater is the water concentration used as internal reference (110 M) [24], Ametabolite is the area of the metabolite peak, Awater is the area of the water peak calculated from the unsuppressed water data, np is the number of protons in a given metabolite, and T1metabolite, T2metabolite, T1water, T2water are the relaxation times of metabolites and water respectively [14–18].
Volunteer scans
Eight healthy volunteers, aged from 20 to 30 years, were scanned using the RAPID-SI sequence. Written informed consent were obtained from each participant prior to scanning. This validation study was approved by the Ethical Committee of the Ghent University Hospital. A 2D 8x8 and 16x16 brain RAPID-SI and CSI datasets were acquired using the following parameters: TE = 35 ms, TR = 2000 ms, BW = 2000 Hz, Nfull = 512, Navg = 2, and R = 4, 8. The FOV dimensions were AP = 100 mm, HF = 10 mm and RL = 100 mm for 2D acquisitions. Additional 3D (16x16x4) RAPID-SI and CSI dataset were acquired from two other volunteers using the same acquisition parameters as the 2D dataset, except for TR = 1800 ms, and Navg = 1. The resolution in the slice direction was reduced to 4 due to the long acquisition time required by CSI.
High-resolution T1 weighted gradient echo MR images (70° flip angle, TR/TE 255/2.96 ms, 30 slices, 3.0 mm thick, 256 x 192 matrix) from the entire head are acquired for PRESS volume positioning. Eight outer-volume saturation bands, placed around the PRESS ROI box, were used for lipid suppression and CHESS water suppression was performed. Shimming was performed using the automated software package FASTMAP [23].
Data processing
The acquired MRSI data was processed offline using an in-house program using Matlab (MathWorks, USA). The pre-processing software with a graphical user interface for data analysis is made available (see GitHub link in data availability statement). The reference signal, AS0 (k = 0,Nfull = 512), was processed first to estimate the signal parameters dm, and fm using HLSVD-PRO. The model order M for HLSVD-PRO was set to 10 to include the major brain metabolites. The estimated parameters were then inserted into Eq 10 to generate the sub-bases ψsub(m,tj), which were used in Eq 11, to calculate the complex amplitudes from the collected s(k(l*R)+j,tj) data. The calculated values were used in Eq 8 to produce the noiseless signals. A Hamming filter was applied on the constructed data along with AS0, prior to Fourier transform, to obtain localized spectra/metabolite maps. The sum of squares approach for coil channel combination was applied after data reconstruction. Water suppression using HLSVD-PRO was optimally achieved with model order of 7.
To accurately estimate signal parameters using HLSVD-PRO, the duration of the acquisition window was set to the average T2 values of the main brain metabolites (NAA, Cr, Cho, mI and lac) obtained at 3 Tesla, found to be approximately 250 ms [14–18] and the receiver bandwidth set to 2000 Hz, the total number of points Nfull was set to 512 data points.
As shown on the method flow chart (Figs 1 and 2), the reference signal AS0 is always used with the noiseless pre-processed s(k (k ≠ 0),t = Nfull) signals for data reconstruction. This induces noise to be present in the output localized spectra after Fourier transform.
The fitting of metabolite peaks, was achieved using HLSVD-PRO with model order of 10. All the metabolite peaks, Cho, Cr, NAA, mI and lac, in the frequency range 1 to 4 PPM were fitted. Cramér–Rao Lower Bound (CRLB) criteria for NAA, Cr and Cho peaks was used to evaluate the quality of fitting. The CRLB/amplitude < 20% was selected as a criterion to discriminate well fitted metabolites from poorly fitted ones.
Statistical analysis
A Wilcoxon signed-rank test [25] using SPSS Statistics 25, was applied to check on the difference between the median of NAA/Cr and Cho/Cr ratios of CSI and RAPID-SI techniques. A p-value of 0.05 below which the study was considered statistically significant. All analyses were conducted using SPSS Statistics 25.
Results
Simulation results
With Nfull set to 512, the highest acceptable R value was 8 with an SNR of 4.5 (Fig 4). Higher acceleration factors cause quantification errors on the estimation of metabolite peaks regardless of the SNR values and/or field inhomogeneity. This was mainly due to the very short duration of the encoded sub-signals, (e.g: 32 data points, with Nfull = 512, and R = 16). The lowest acceptable SNR, was established by the standard deviation of the estimated metabolite values, tolerated to deviate by about 20% from the expected ones (Fig 4). The effect of field inhomogeneity was insignificant on the accuracy of the data quantification when the level of field inhomogeneity (e.g: difference in full width at half maximum (FWHM) of the metabolite peaks from one voxel to another) is around 8 Hz (ΔB0 map B4). However, with a level of inhomogeneity above 8 Hz (ΔB0 map B5), significant quantification errors on estimated data were noticed (Fig 4). Therefore, to obtain accurate results with Nfull = 512, and R set to 8, the SNR should be ≥ 4.5 and the level of field inhomogeneity ≤ 8 Hz. These values were compared to those obtained from in vivo MRSI datasets collected from healthy volunteers using the same acquisition parameters and with several spatial resolutions and dimensions. The SNR in vivo was found to be 8.3 in the centre of the brain and decreases to about 6 ~ 7 at the edges of the region of interest (ROI), producing a peak broadening of about 3 Hz for the NAA, Cho and Cr resonances (S1 Fig). Therefore, the limits of the proposed method in term of SNR and field inhomogeneity were within those encountered in vivo, making RAPID-SI feasible in clinical settings. As shown in Fig 5, comparable metabolite maps for NAA, lac, Cho, Cr and mI were obtained from RAPID-SI (R = 8) with respect to the simulated data for SNR value of 6 and ΔB0 map B2.
Phantom results
Fig 6A shows the 4 tubes within the phantom with different metabolite distributions. In Fig 6R–6U and 2D 16x16 RAPID-SI and CSI in vitro results, are displayed, with comparable metabolite distributions for NAA, lac and Cho, Cr across all the tubes up to acceleration factor of 8, after pre-processing (Fig 6T). Furthermore, RAPID-SI provided accurate quantification results (Table 1) up to acceleration factor of 8, after pre-processing. However, higher than 8, the quantification accuracy degrades as seen from the spectra in Fig 6N–6Q and in the metabolite map in Fig 6U. It can also be seen in Fig 6B–6E (comparing reconstructed metabolite peaks from raw CSI data (blue) and the raw RAPID-SI data (red), zero-filled to full time length of 512 points) that the raw data acquired using RAPID-SI results in inaccurate reconstruction of the peaks due to the low number of points of the sub-signals, which is recovered after the pre-processing step as seen in Fig 6F–6I (with R = 4) and in Fig 6J–6M (with R = 8) respectively. The concentrations of the metabolites were calculated according to Eq 13. There was a significant reduction in acquisition time (17 min with CSI and about 2 min with the RAPID-SI (R = 8)).
Table 1. Mean value of expected and estimated metabolite concentration values from CSI and RAPID-SI for different acceleration factors.
NAA (mM) | Lac (mM) | Cho (mM) | Cr (mM) | |
---|---|---|---|---|
Expected | 10.5 | 25 | 9.8 | 4.5 |
CSI | 10.3 | 25.1 | 9.7 | 4.6 |
RAPID-SI (R = 4) | 10.3 | 25.2 | 9.9 | 4.6 |
RAPID-SI (R = 8) | 10.2 | 24.7 | 9.4 | 4.3 |
RAPID-SI (R = 16) | 5.1 | 13.2 | 3.3 | 1.5 |
In vivo results
The overall SNR as defined in Eq 12 was lower in RAPID-SI than CSI by approximately the factor at the acquisition level. The SNR mean ± standard deviation (SD) in CSI across the volunteers was 6.98 ± 0.2, and reduced to 2.32 ± 0.15 in RAPID-SI with (R = 8). The calculated SNR using both the methods for all the volunteers can be found in S1 Appendix.
Similar results (Fig 7E–7G) in terms of metabolite maps in vivo of Cr, NAA, Cho and mI, were obtained, demonstrating that RAPID-SI was able to preserve the spatial distribution of the metabolites while reducing acquisition time. It can also be seen when comparing reconstructed metabolite peaks from raw CSI data (blue) and the raw RAPID-SI data (red), zero-filled to full time length of 512 points (Fig 7B) that the raw data acquired using RAPID-SI results in inaccurate peaks reconstruction due to low number of data points of the sub-signals, which is recovered by the pre-processing step as displayed in Fig 7C and 7D with RAPID-SI (R = 4) and (R = 8) respectively.
The pre-processed RAPID-SI data were analysed using HLSVD-PRO and peak areas were calculated. Voxels with CRLB/amplitude < 20% were selected and their metabolite peaks were quantified. The calculated CRLB of both methods for all the volunteers was provided in S2 Appendix. Means and standard deviations of NAA/Cr and Cho/Cr ratios of both CSI and RAPID-SI (R = 8) reported in Tables 2 and 3, and calculated from the ROI (Fig 7A) across all the volunteers were comparable. The left box plot in Fig 8 displays the mean NAA/Cr ratio inside the ROI for both CSI and RAPID-SI for all the volunteers. The CSI median value was 2.01, ranging between 1.9 and 2.17, whereas with RAPID-SI (R = 8), the median value was 1.99 varying between 1.92 and 2.15. No significant difference between both methods was found based on the p-value of 0.53 using the Wilcoxon signed-rank test. Similarly, the right box plot in Fig 8 displays the mean Cho/Cr ratio inside the ROI for both techniques across all the volunteers. The CSI median value was found to be 0.90, ranging between 0.84 and 0.97, whereas with RAPID-SI (R = 8), the median value was 0.89 varying between 0.82 and 0.95. No significant difference between the two methods was found based on the p-value of 0.06.
Table 2. Mean (± standard deviation) of the in vivo NAA/Cr ratios across all the volunteers for CSI and RAPID-SI (R = 8) inside the ROI.
NAA/Cr | Subject 1 | Subject 2 | Subject 3 | Subject 4 | Subject 5 | Subject 6 | Subject 7 | Subject 8 |
---|---|---|---|---|---|---|---|---|
CSI | 2.15 ± .08 | 1.90 ± .08 | 1.92 ± .15 | 2.03 ± .11 | 2.17 ± .11 | 2.03 ± .14 | 2.00 ± .15 | 1.91 ± .22 |
RAPID-SI (R = 8) | 2.13 ± .06 | 1.92 ± .07 | 1.92 ± .17 | 2.00 ± .14 | 2.15 ± .13 | 2.05 ± .13 | 1.98 ± .14 | 1.93 ± .20 |
Table 3. Mean (± standard deviation) of the in vivo Cho/Cr ratios across all the volunteers for CSI and RAPID-SI (R = 8) inside the ROI.
Cho/Cr | Subject 1 | Subject 2 | Subject 3 | Subject 4 | Subject 5 | Subject 6 | Subject 7 | Subject 8 |
---|---|---|---|---|---|---|---|---|
CSI | 0.92 ± .05 | 0.94 ± .03 | 0.97 ± .02 | 0.86 ± .04 | 0.94 ± .03 | 0.88 ± .14 | 0.86 ± .03 | 0.84 ± .06 |
RAPID-SI (R = 8) | 0.90 ± .03 | 0.92 ± .02 | 0.95 ± .02 | 0.88 ± .04 | 0.92 ± .03 | 0.88 ± .13 | 0.84 ± .03 | 0.82 ± .07 |
To further demonstrate the usefulness of the proposed method, a 3D (16x16x4) RAPID-SI and CSI dataset, collected from two volunteers using the same acquisition parameters as the 2D, except for TR = 1800 ms, and Navg = 1. The resolution in the slice direction was reduced to 4, due to the long acquisition time required by CSI. S2 Fig displays spectra selected from one of the slices, with a zoom on the ventricle region showing the conformity of spectral content with the anatomy of both RAPID-SI and CSI. The acquisition time with the CSI was 30 min 43 sec, reduced to 3 min 50 sec with the RAPID-SI.
Discussion
The proposed technique was compared to CSI in terms of acquisition time, spectral resolution and data sensitivity. Both RAPID-SI and CSI, use the same encoding principle (phase encoding), but differ in their time of application. The CSI performs the encoding prior to signal acquisition, whereas the RAPID-SI achieves the same task during data collection to segment the signal into sequential short time duration encoded sub-signals. This results in an SNR decrease by approximately the factor at the acquisition level (Eq 7), which is compensated during data pre-processing. The amount of noise in the output is minimal due to the use of noiseless full time pre-processed signals estimated from the corresponding acquired sub-signals along with the acquired reference signal at k = 0. The amplitudes are accurately estimated as demonstrated by simulation, phantom and in vivo results.
Other techniques like the Multi-Spin-Echo Proton Spectroscopic Imaging technique, EPSI, SPICE and others have been developed to accelerate conventional MRSI. However, in the Multi-Spin-Echo Proton Spectroscopic Imaging technique [4,5], spectral resolution is reduced because of the shortened readout and spatial blurring occurs due to signal modulation caused by T2 relaxation. This technique is most useful for metabolites having long T2 values such as NAA and Choline and the maximum acceleration factor is also typically only 2 to 3. In a conventional EPSI sequence [6], the acceleration factor is higher. For the example of a 2D 16x16 data, the acquisition time can be accelerated by a factor of 16 and even higher for higher resolution but at the penalty of reduced SNR by a factor determined by the dwell time of acquisition. It also requires high performance gradient systems with high slew rate. The SPICE technique [8] has been proposed to improve the spectral resolution and SNR as compared to EPSI with reduced acquisition time but still it requires the sophisticated gradient system for data acquisition along with sophisticated post processing method. RAPID-SI uses an acceleration factor of 8 for any spatial resolution and recovers the metabolite information. Higher acceleration factor reduces the performance of the technique, with no requirement of high performance gradients, used in other techniques such as EPSI.
The short time duration of the encoded sub-signals are expected to carry similar frequency and damping factors of these metabolites. The use of a reference signal AS0 (k = 0,Nfull = 512) is therefore sufficient to accurately estimate the values of the frequency and damping factors, which are used in the data reconstruction process. It is to be noted that data analysis and quantification can only be performed once the pre-processing is performed, without which the data will be very low in SNR and with poor spectral resolution leading to inaccurate data analysis. Therefore, the pre-processing is a mandatory step and must follow data acquisition for accurate data quantification.
One should notice that because the reference signal is the sum of the signals from all the voxels at the centre of k-space, the estimated frequency and damping factor values for each metabolite is the average value across all the voxels. The B0 inhomogeneity may cause discrepancy of these values, which is reflected on the corresponding estimated average. However, this average value should remain constant across the encoded sub-signals, since each encoded sub-signal is also the sum of the sub-signals of all the voxels but at a given k-space value. This is comparable to CSI, where the collected signals are the sum of all the voxels at each encoding step with frequency and damping factor averaged. The well-known Lorentzian model has been chosen to represent the spectral lines. However, the method is flexible to incorporate any other signal model (e.g. Gaussian, Voigt), and therefore, any source of signal deformation (e.g. field inhomogeneity) could be included in the signal model to improve the method outcome. In our pre-processing software, we have chosen the HLSVD method, which is a reliable method and proven to provides accurate results. Other analysis procedures such as LCModel [26], or jMRUI [27], have to be adapted to be used as a substitute in the pre-processing step. These methods can also be used for data quantification following the pre-processing step.
The duration of the acquisition window was set to the average T2 values of the main brain metabolites (NAA, Cr, Cho, mI, and lac) obtained at 3 Tesla, which is approximately 250 ms. This is found to preserve peak information in all the collected sub-signals, especially those acquired at high k space values (edges of k-space), and therefore, provide accurate estimation of signal parameters using the pre-processing algorithm. As a consequence, the suitable Nfull value was found to be 512 data points for a receiver bandwidth equal to 2 kHz, and the best acceleration factor was 8. Higher acceleration factor is possible (R = 16). However, this reduces the number of points in Nsub (inversely proportional to R), and degrades the data quality, by affecting the quantification results as shown in the simulation and in vitro results.
The encoding efficiency of the proposed technique is subject to the T2 values of the metabolites. Given that the k-space is filled during the acquisition window, the number of k-space data points used to spatially encode each metabolite signal is proportional to its corresponding T2 value. Signal components with long T2 are sampled by several data points in k-space, whereas those with short T2 will have less points to characterize them. This may cause a spatial blurring (voxel bleeding) of the metabolite maps with short T2. To overcome this problem, a deconvolution between a sinc-like function with width at half maximum (WHM) inversely proportional to the T2 values of the blurred metabolite signals and their corresponding maps was performed to recover the original spatial distribution of the metabolites.
Several k-space trajectories are possible (e.g. Cartesian, radial, and spiral) using this method. For the sake of simplicity, the Cartesian scheme was chosen, starting from one edge of the k space, which coincides with the origin of the decaying part of the spin echo and moving towards the other edge while recording the sub-signals. In this manner, a balance in term of signal intensity, and therefore an SNR level between the acquired sub-signals is maintained to allow for better estimation of the signal parameters when applying the pre-processing procedure. Other k-space paths can also be explored in future and compared in terms of acquisition time, data sensitivity and reconstruction.
The developed method was tested and validated by performing simulated, phantom and in vivo experiments. As noticed from simulation, the performance of method drops at very low SNR values (SNR = 4.5) and high level of field inhomogeneity (more than 8Hz). However, if these severe conditions are encountered, the acceleration factor is lowered. Compared to CSI, this method achieves a reduction in acquisition time, which is proportional to R, while providing accurate metabolite quantification results. Furthermore, the method is flexible to be combined with other acceleration methods, such as parallel imaging.
It will be more beneficial if a larger comparative study is conducted and other high speed techniques such as PEPSI, and SPICE methods are used, in order to compare their performance in terms of acquisition time, SNR, resolution, hardware requirements and domain of application with the RAPID-SI. In the future, we plan to extend this study for clinical applications along with comparison to other fast spectroscopic imaging techniques.
Supporting information
Acknowledgments
The authors would like to thank Dr. Sabine Van Huffel from University of Leuven for the use of the HLSVD software.
Data Availability
All relevant data are within the paper. The pre-processing software with a graphical user interface for doing data analysis along with with one volunteer's anonymized raw dataset is included at https://github.com/souravUniGent/RAPID-SI. Other raw dataset and pulse sequence source code will be made available on request to the first author: Sourav.Bhaduri@UGent.be.
Funding Statement
This work has been supported by the Flemish funding agency (FWO: Fonds voor Wetenschappelijk Onderzoek) grant number: G086912N.
References
- 1.Maudsley A, Hilal S, Perman W, Simon H. Spatially resolved high resolution spectroscopy by “four-dimensional” NMR. J Magn Reson. 1983;51: 147–152. [Google Scholar]
- 2.Pohmann R, von Kienlin M, Haase A. Theoretical evaluation and comparison of fast chemical shift imaging methods. J Magn Reson. 1997;129: 145–160. 10.1006/jmre.1997.1245 [DOI] [PubMed] [Google Scholar]
- 3.Posse S, Otazo R, Dager SR, Alger J. MR spectroscopic imaging: principles and recent advances. J Magn Reson Imaging. 2013;37: 1301–1325. 10.1002/jmri.23945 [DOI] [PubMed] [Google Scholar]
- 4.Duyn JH, Frank JA, Moonen CTW. Incorporation of Lactate Measurement in Multi-Spin-Echo Proton Spectroscopic Imaging. Magn Reson Med. 1995;33: 101–107. [DOI] [PubMed] [Google Scholar]
- 5.Duyn JH, Moonen CTW. Fast proton spectroscopic imaging of human brain using multiple spin-echoes. Magn Reson Med. 1993;30: 409–414. [DOI] [PubMed] [Google Scholar]
- 6.Posse S, Tedeschi G, Risinger R, Ogg R, Le Bihan D. High speed 1H spectroscopic imaging in human brain by echo planar spatial-spectral encoding. Magn Reson Med. 1995;33: 34–40. [DOI] [PubMed] [Google Scholar]
- 7.Serrai H, Senhadji L. Acquisition time reduction in magnetic resonance spectroscopic imaging using discrete wavelet encoding. J Magn Reson. 2005;177: 22–30. 10.1016/j.jmr.2005.07.006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Lam F, Ma C, Clifford B, Johnson CL, Liang ZP. High-resolution (1) H-MRSI of the brain using SPICE: Data acquisition and image reconstruction. Magn Reson Med. 2016;76: 1059–1070. 10.1002/mrm.26019 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Fu Y, Serrai H. Fast magnetic resonance spectroscopic imaging (MRSI) using wavelet encoding and parallel imaging: in vitro results. J Magn Reson. 2011;211: 45–51. 10.1016/j.jmr.2011.03.019 [DOI] [PubMed] [Google Scholar]
- 10.Posse S, Otazo R, Tsai SY, Yoshimoto AE, Lin FH. Single-Shot MR Spectroscopic Imaging with Partial Parallel Imaging. Magn Reson Med. 2009;61: 541–547. 10.1002/mrm.21855 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Hu S, Lustig M, Chen AP, Crane J, Kerr A, Kelley DAC, et al. Compressed sensing for resolution enhancement of hyperpolarized 13C flyback. J Magn Reson. 2008;192: 258–264. 10.1016/j.jmr.2008.03.003 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Laudadio T, Mastronardi N, Vanhamme L, Van Hecke P, Van Huffel S. Improved Lanczos Algorithms for Blackbox MRS Data Quantitation. J Magn Reson. 2002;157: 292–297. [DOI] [PubMed] [Google Scholar]
- 13.Hansen PC. Regularization Tools version 4.0 for Matlab 7.3. Numer Algorithms. 2007;46: 189–194. [Google Scholar]
- 14.Isobe T, Matsumura A, Anno I, Yoshizawa T, Nagatomo Y, Itai Y, et al. Quantification of cerebral metabolites in glioma patients with proton MR spectroscopy using T2 relaxation time correction. Magn Reson Imaging. 2002;20: 343–349. [DOI] [PubMed] [Google Scholar]
- 15.Ernst T, Kreis R, Ross BD. Absolute Quantitation of Water and Metabolites in the Human Brain. I. Compartments and Water. J Magn Reson B. 1993;102: 1–8. [Google Scholar]
- 16.Matsumura A, Isobe T, Takano S, Kawamura H, Anno I. Non-invasive quantification of lactate by proton MR spectroscopy and its clinical applications. Clin Neurol Neurosurg. 2005;107: 379–384. 10.1016/j.clineuro.2004.10.009 [DOI] [PubMed] [Google Scholar]
- 17.Bottomley PA, Foster TH, Argersinger RE, Pfeifer LM. A review of normal tissue hydrogen NMR relaxation times and relaxation mechanisms from 1–100 MHz: dependence on tissue type, NMR frequency, temperature, species, excision, and age. Med Phys. 1984;11: 425–448. 10.1118/1.595535 [DOI] [PubMed] [Google Scholar]
- 18.Mlynarik V, Gruber S, Moser E. Proton T1 and T2 relaxation times of human brain metabolites at 3 Tesla. NMR Biomed. 2001;14: 325–331. [DOI] [PubMed] [Google Scholar]
- 19.Maudsley AA, Govind V, Arheart KL. Associations of age, gender and body mass with 1H MR-observed brain metabolites and tissue distributions. NMR Biomed. 2012;25: 580–93. 10.1002/nbm.1775 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Klose U. Measurement sequences for single voxel proton MR spectroscopy. Eur J Radiol. 2008;67: 194–201. 10.1016/j.ejrad.2008.03.023 [DOI] [PubMed] [Google Scholar]
- 21.Haase A, Frahm J, Hanicke W, Matthaei D. 1H NMR chemical shift selective (CHESS) imaging. Phys Med Biol. 1985;30: 341–344. [DOI] [PubMed] [Google Scholar]
- 22.Duyn JH, Matson GB, Maudsley A, Weiner MW. 3D phase encoding 1H spectroscopic imaging of human brain. Magnetic Resonance Imaging. 1992;10: 315–319. [DOI] [PubMed] [Google Scholar]
- 23.Gruetter R. Automatic, localized in vivo adjustment of all first- and second-order shim coils. Magn Reson Med. 1993;29: 804–811. [DOI] [PubMed] [Google Scholar]
- 24.Serrai H, Clayton DB, Senhadji L, Zuo C, Lenkinski RE. Localized proton spectroscopy without water suppression: removal of gradient induced frequency modulations by modulus signal selection. J Magn Reson. 2002;154: 53–59. 10.1006/jmre.2001.2462 [DOI] [PubMed] [Google Scholar]
- 25.Aban IB, George B. Statistical Considerations for Preclinical Studies. Exp Neurol. 2015;270: 82–87. 10.1016/j.expneurol.2015.02.024 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Provencher SW. Automatic quantitation of localized in vivo 1H spectra with LCModel. NMR Biomed. 2001;14: 260–264. [DOI] [PubMed] [Google Scholar]
- 27.Dcf Stefan D, Cesare F, Andrasescu A, Popa E, Lazariev A, Vescovo E, et al. Quantitation of magnetic resonance spectroscopy signals: The jMRUI software package. vol. 20; 2009. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
All relevant data are within the paper. The pre-processing software with a graphical user interface for doing data analysis along with with one volunteer's anonymized raw dataset is included at https://github.com/souravUniGent/RAPID-SI. Other raw dataset and pulse sequence source code will be made available on request to the first author: Sourav.Bhaduri@UGent.be.