Purpose: The present study designed to evaluate entrance and exit doses for out of the radiation field in external beam radiotherapy (EBRT). The primary aim of this study was to investigate the efficiency of non tissue equivalent (NTE) thermoluminescence dosimeters (TLD) for in vivo dosimetry of out of the radiation field dose measurements.
Materials and Methods: All the measurements were performed in 10 head and neck patients (age range, 35–46 years; mean, 44 years) treated with two parallel opposed lateral fields on Bhabhatron-II TAW Telecolbalt unit (Panacea Medical Technologies, Bengaluru, India) using source to surface distance (SSD) technique. The TLDs used in this study were Nucleonix India Pvt. Ltd., CaSO4:Dy discs (13.3 mm diameter, 0.8 mm thick) and LiF: Mg, Ti chips (3.2 X 3.2 X 0.89 mm). The CaSO4:Dy discs were placed at the level of the eyes of the patient for a single right lateral treatment field only. This methodology provided the set up to assess out of field entrance and exit radiation dose to eye. TLD-100 chips were also placed exactly in the identical places in the next treatment fraction of same patient. The physical data measured were separation distance at the level of eye, distance between radiation field edge and ipsilateral right eye at SSD. The distances were calculated for radiation beam exit from isocenter at the exit surface of the patient.
Results and Discussion: The distance between radiation field edge and ipsilateral eye at SSD was measured in the range of 2.0–4.0 cm with mean 3.3 cm. The distances of separation at the level of eye, entrance and exit of edge of beam from isocenter were in the range of 11cm-15cm, 7.0cm-8.5cm, 8.0cm-9.5cm respectively. It was obvious to observe with theoretical calculations using radiation divergence property that the primary radiation beam was not passing through contralateral eye. The contralateral eye was away from the exit of edge of radiation beam in all the cases and distances were found in the range of 0.5cm-2.8cm. However, when the doses were analyzed for non tissue equivalent CaSO4:Dy discs and it was surprising to note that the exit dose to contralateral eye (Dexit) were measured 1.5 to 2.5 times higher than the entrance dose to ipsilateral eye (Dentrance).
To investigate this over-response, the doses were measured with tissue equivalent TLD-100 chips in the identical conditions. It was found that Dexit were measured 15% to 20% less than Dentrance. Thus, the possible cause for this over-response in NTE dosimeter is increase in the intensity of secondary electrons and low energy scattered photons reaching to dosimeter at the exit surface of the patient during out of field measurements. The results of this study suggested that non tissue equivalent CaSO4:Dy discs were not the dosimeter of choice for out of field exit dose measurements. One should be precautious to use non tissue equivalent CaSO4:Dy dosimeters for out of field exit dose measurements. Further study needs to be performed to deal this over-response using either appropriate correction factors or build up caps. However, studies of buildup caps should be conducted for out of field measurements. In addition, special attention should be paid to the selection of the appropriate materials for buildup caps, taking into account the impact of low energy radiation and attenuation of the beam.
Conclusion: The non tissue equivalent dosimeters were not the promising dosimeters for out of field exit dose measurements. The research outputs of this study may be helpful for the selection of the appropriate in vivo dosimeter suitable for clinical use for out of the radiation field dose measurement conditions in radiotherapy.