Skip to main content
Journal of Healthcare Engineering logoLink to Journal of Healthcare Engineering
. 2018 Oct 29;2018:2376317. doi: 10.1155/2018/2376317

An Improved Fuzzy Connectedness Method for Automatic Three-Dimensional Liver Vessel Segmentation in CT Images

Rui Zhang 1, Zhuhuang Zhou 1, Weiwei Wu 2, Chung-Chih Lin 3, Po-Hsiang Tsui 4,5,6, Shuicai Wu 1,
PMCID: PMC6231381  PMID: 30510670

Abstract

In this paper, an improved fuzzy connectedness (FC) method was proposed for automatic three-dimensional (3D) liver vessel segmentation in computed tomography (CT) images. The vessel-enhanced image (i.e., vesselness image) was incorporated into the fuzzy affinity function of FC, rather than the intensity image used by traditional FC. An improved vesselness filter was proposed by incorporating adaptive sigmoid filtering and a background-suppressing item. The fuzzy scene of FC was automatically initialized by using the Otsu segmentation algorithm and one single seed generated adaptively, while traditional FC required multiple seeds. The improved FC method was evaluated on 40 cases of clinical CT volumetric images from the 3Dircadb (n=20) and Sliver07 (n=20) datasets. Experimental results showed that the proposed liver vessel segmentation strategy could achieve better segmentation performance than traditional FC, region growing, and threshold level set. Average accuracy, sensitivity, specificity, and Dice coefficient of the improved FC method were, respectively, (96.4 ± 1.1)%, (73.7 ± 7.6)%, (97.4 ± 1.3)%, and (67.3 ± 5.7)% for the 3Dircadb dataset and (96.8 ± 0.6)%, (89.1 ± 6.8)%, (97.6 ± 1.1)%, and (71.4 ± 7.6)% for the Sliver07 dataset. It was concluded that the improved FC may be used as a new method for automatic 3D segmentation of liver vessel from CT images.

1. Introduction

Hepatocellular carcinoma (HCC) is one of the most common malignancies in the world, especially in China with the fifth morbidity and the third mortality [1, 2]. Nowadays, main treatments to HCC include surgical resection, liver transplantation, and local thermal ablation [3]. Treatment planning and navigation based on medical imaging are essential for these procedures. Among different medical imaging modalities, computed tomography (CT) is commonly used for the guidance of liver tumor treatment. Three-dimensional (3D) segmentation of liver vessel is critical in CT-based liver tumor treatment planning and navigation. However, manual segmentation of liver vessel in CT images is time consuming and tedious. Thus, there is a demand for computerized 3D segmentation of liver vessel in CT images [4, 5].

Currently, computerized liver vessel segmentation techniques can be classified into region growing [68], active contour models or level sets [9], graph cuts [1012], extreme learning [13], deep learning [14], and fuzzy logic [15, 16]. However, it is still challenging to extract liver vessel in CT images, especially in those with low contrast [4]. Region growing methods [68] are simple with low computational cost, but they are sensitive to seed point location and threshold. Active contour models or level sets [9] are among mainstream vessel segmentation algorithms, but they have limitations of complex initialization and typically use speed function which implicitly assumes that images are of good contrast. Graph cuts [1012] are segmentation methods based on graph theory, but they are partly limited by the shrinking bias problem. Machine learning methods [13, 14] can take into account the diversity of liver vessel topologies and features, but they generally require plenty of training samples or long training time. Fuzzy connectedness (FC) methods [15, 16] are based on fuzzy logic. FC describes spatial connectedness between each voxel, rather than just focusing on intensity. Recently, Guo et al. [15] and Wang et al. [16] have demonstrated the potential of FC in liver vessel segmentation. However, for FC-based liver vessel segmentation in CT images, there are still issues to be addressed, including unsatisfying segmentation accuracy (especially for low-contrast CT images), requirement on multiple seeds, and sensitivity to initialization.

In this paper, an improved FC method was proposed for automatic 3D liver vessel segmentation in CT images. The vessel-enhanced image (i.e., vesselness image) was incorporated into the fuzzy affinity function of FC, rather than the intensity image used by traditional FC. An improved vesselness filter was also proposed based on the Jerman's vesselness filter [17] introduced recently. The fuzzy scene of FC was initialized by using the Otsu segmentation algorithm, and the quantity of seeds required was reduced to one which was generated automatically. The proposed method was evaluated on 40 cases of clinical CT image volumes, including low-contrast images. Experimental results demonstrate that the improved FC method can overcome the drawbacks of traditional FC and yield more satisfying segmentation performance.

2. Materials and Methods

Figure 1 shows the flow chart of the improved FC method. First, the liver volume of interest (VOI) image was obtained by using the liver mask, which could be obtained by using liver segmentation approaches [18]. The liver VOI image was then contrast enhanced by an adaptive sigmoid filtering which was initialized by K-means clustering and isotropically resampled. Subsequently, the improved vesselness filter was used to enhance the liver vessel and suppress the background, and a 3D vesselness image was obtained. Then, a 3D fuzzy scene was constructed with the 3D vesselness image by (1) incorporating the improved vesselness into the fuzzy affinity function of FC, (2) initializing the fuzzy scene by the Otsu algorithm, and (3) generating automatically one single seed. Finally, the 3D liver vessel was segmented on the basis of the 3D fuzzy scene and anisotropically resampled.

Figure 1.

Figure 1

Flowchart of the improved fuzzy connectedness method for automatic 3D segmentation of liver vessel from CT images.

2.1. Dataset

Both simulated data (n=60) and clinical CT data (n=40) were used. The synthetic dataset VascuSynth [19, 20] was provided by the Medical Image Analysis Lab, School of Computing Science, Simon Fraser University, Canada. VascuSynth contains 10 groups of data, which are publically available at http://vascusynth.cs.sfu.ca. Each group consists of 12 randomly generated images with different quantity of bifurcations. Six groups of data were randomly selected; among them, the images with bifurcations ≥11 were included in this study. Gaussian white noise was also added to the raw data. The level of the Gaussian white noise was indicated by σ2, the variance of the noise. In this study, Gaussian white noise with σ2 = 30, 45, and 60 were added.

The clinical CT image datasets, 3Dircadb and Sliver07, were used. 3Dircadb contains 20 cases of contrast-enhanced CT (CE-CT) images. 3Dircadb was provided by the Research Institute against Digestive Cancer, France, and is publically available at http://www.ircad.fr/research/3dircadb. The pixel spacing is 0.56–0.86 mm, and the slice thickness is 1–4 mm. The number of slices ranges from 64 to 502, and the in-plane resolution is 512 × 512 pixels. The gold standard of liver vessel was provided by 3Dircadb, which was manually delineated by radiologists. Sliver07 contains 30 cases of CE-CT images, including 20 training sets and 10 testing sets. The 20 cases of training data are publically available at http://www.sliver07.org and were included in this study. However, Sliver07 did not provide the gold standard of liver vessel. Therefore, radiologists were asked to manually delineate the liver vessel to serve as the gold standard for the 20 cases of training data of Sliver07. The number of slices, in-plane resolution, and interslice resolution range from 64 to 394, from 0.58 to 0.81 mm, and from 0.7 to 5.0 mm, respectively.

2.2. Improved Vesselness Filter

The multiscale Hessian matrix-based filter (vesselness filter) is commonly used for vessel enhancement [4, 17]. Classical vesselness filters were proposed by Sato et al. [21] and Frangi et al. [22]. Since then, Li et al. [23], Erdt et al. [24], and Xiao et al. [25] proposed improved methods for enhancing the vasculature. Recently, Jerman et al. [17] proposed a novel vesselness filter and demonstrated that it outperformed traditional vesselness filters. For completeness, the Jerman's filter was introduced briefly as below. Let λi, i=1,2,3 denotes the Hessian eigenvalues of a 3D image at each coordinate x. Considering the ideal eigenvalues' relationship λ2λ3 ∧ |λ2,3| > >|λ1| in vasculature, Jerman et al. [17] constructed a novel Hessian eigenvalues function to improve the enhancement performance by using a two-step piecewise compensation. In CT images, the magnitudes of λ2 and λ3 were lower at the vascular boundary or in the low-scale vessel (|λ3| ≥ |λ2| ≈ |λ1|⟶Low), which did not match the ideal Hessian eigenvalues relationship in vasculature, resulting in significant attenuation of the vesselness response. Therefore, Jerman et al. [17] performed a piecewise compensation on the eigenvalue λ3:

λρσ=λ3,if λ3>τmaxxλ3x,σ,τmaxxλ3x,σ,if 0<λ3τmaxxλ3x,σ,0,otherwise, (1)

where σ is the vessel scale and τ is a threshold between 0 and 1. In addition, traditional vesselness filters would suppress blob-like structures and obtain poor response at vascular nodes (|λ1| ≈ |λ2| ≈ |λ3|⟶High). Thus, Jerman et al. [17] compensated the ellipsoid structure conforming to the condition λ2λρ/2 > 0 to construct the final vesselness function:

υk=0,if λ20λρ0,1,if λ2λρ/2>0,λ22λρλ23/λ2+λρ3,otherwise. (2)

Jerman et al. [17] evaluated their method on clinical image datasets of lung, cerebral, and fundus vasculatures. However, for the task of liver vessel enhancement in CT images, the Jerman's filter would enhance the liver contour, liver parenchyma, and noise. Therefore, the Jerman's filter was improved by incorporating adaptive sigmoid filter for contrast enhancement and by incorporating a background-suppressing item into the vesselness function of the Jerman's filter (Equation (2)).

The adaptive sigmoid filter is defined as

Isigmoid=1+expIVOIβα1, (3)

where Isigmoid is the filtered image, IVOI is the liver VOI image, and β and α represented the intensity center and the intensity range of the vasculature. In this study, β and α were obtained adaptively by the K-means clustering (K = 5). The internal structure of IVOI was clustered into five regions with corresponding cluster centers. With the value of the intensity centers ranking from low to high, the five regions corresponded to the background, liver tumor, liver parenchyma, low-intensity vessel mixed with liver parenchyma, and high-intensity vessel, respectively. With the intensity means of the last two regions (m1 and m2), parameters β and α are calculated by

α=m2m1/2,β=m2+m1/2. (4)

The background-suppressing item, 1 − eRs2/2γ, was incorporated into Equation (2), yielding

υ=0,if λ20λρ0,1,if λ2λρ/2>0,λ22λρλ23/λ2+λρ31eRs2/2γ,otherwise, (5)

where Rs=λ12+λ22+λρ2 and γ is the background suppression coefficient, which was optimally set at λρ/3.

Finally, the vesselness response was combined by calculating the maximum response of υ in each scale σ, σ ∈ [σmin, σmax]:

Ivesselness=supυ:σminσσmax, (6)

where Ivesselness is the final vessel-enhanced image (vesselness image). The improved vesselness filter algorithm is summarized in Algorithm 1.

Algorithm 1.

Algorithm 1

Improved vesselness filter.

2.3. Improved Fuzzy Connectedness

FC involved three kinds of fuzzy relationships: fuzzy adjacency, fuzzy affinity, and fuzzy connectivity. Fuzzy affinity represented the local similarity of the voxel pair (c, d) in the entire image scene C, denoted by μκ(c, d) ∈ [0,1]:

μκc,d=μαc,dω1h1fc,fd+ω2h2fc,fd, (7)

where μα(c, d) is the fuzzy adjacency (a monotonic increasing function), and h1 and h2 are computed by

h1fc,fd=e1/2fc+fd/2m/s2,h2fc,fd=e1/2fcfdm/s2, (8)

where f(·) is the intensity of voxels; m and s are mean and standard deviation of f(·) in the VOI, respectively; and ω1 and ω2 are weight parameters, ω1+ω2=1.

In this paper, the vesselness image obtained by using the improved vesselness filter was used as the input of the fuzzy affinity function, rather than the intensity image used by traditional FC. The improved fuzzy affinity function, μκ′(c, d), is defined as

μκc,d=μαc,dω1h1Ivesselnessc,Ivesselnessd+ω2h2Ivesselnessc,Ivesselnessd. (9)

To adaptively set parameters m and s, the Otsu segmentation algorithm was adopted to the vesselness image. Two-threshold Otsu was used to yield a binary liver vessel mask. Parameters m and s are, respectively, set at the mean and standard deviation of the vesselness voxels belonging to the foreground of the vessel mask.

The weight parameters ω1 and ω2 are adaptively selected by using the method proposed by Pednekar et al. [26]:

ω1=h1h1+h2,ω2=1ω1. (10)

The fuzzy scene of liver vessel was initialized with one single seed generated automatically on the vesselness image, binarized by a threshold T, and anisotropically resampled to yield the final liver vessel segmentation. Figure 2 illustrates automatic selection of one single seed. In Figure 2(a), the 3D image Ivesselness was divided into several regions Rsub of 5 ∗ 5 ∗ 3 voxels. The maximum vesselness voxels at each Rsub region were selected as potential seeds (denoted by the blue points in Figure 2(b)). Then, the regions of 5 ∗ 5 ∗ 3 voxels around the potential seeds were constructed and denoted as Rseed, with each potential seed being the center of each Rseed region. The mean of the vesselness of each Rseed region was calculated. The potential seed having the largest vesselness mean in its Rseed region was automatically selected as the final single seed, which was indicated by the red point in Figure 2(b). The improved FC algorithm is summarized in Algorithm 2.

Figure 2.

Figure 2

Multiple potential seeds and one single seed generated automatically on the vesselness image. (a) Illustration for automatic seed selection. (b) Multiple potential seeds (blue) and one single seed (red) indicated on the vesselness image.

Algorithm 2.

Algorithm 2

Improved fuzzy connectedness.

2.4. Evaluation

To analyze quantitatively the performance of the proposed vessel segmentation method, evaluation metrics including accuracy, sensitivity, specificity, and Dice coefficient were used:

Accuracy=TP+TNTP+FN+TN+FP,Sensitivity=TPFN+TP,Specificity=TNTN+FP,Dice=2TP2TP+FN+FP, (11)

where TP and TN are the numbers of voxels correctly segmented as vessel and background (i.e., nonvessel), respectively; FP and FN are the numbers of voxels incorrectly segmented as vessel and background, respectively.

3. Results

Figure 3 shows the vessel segmented from the simulated data by using the improved FC method. Figure 3(a) represents the ground truth; Figure 3(b) shows the segmented vessel on the synthetic data; and Figures 3(c)3(e) show the segmented vessel on the synthetic data added with Gaussian white noise σ2 = 30, 45, and 60, respectively. The segmentation performance of the improved FC method on the synthetic dataset (n=60) is shown in Table 1, in terms of accuracy, sensitivity, specificity, and Dice coefficient. Although the sensitivity and Dice coefficient were decreased to some extent with increasing the level of Gaussian white noise, the segmentation performance was generally kept stable. It is thus indicated that the improved FC method is insensitive to Gaussian white noise.

Figure 3.

Figure 3

Segmentation of vessel from synthetic data by using the improved fuzzy connectedness method. (a) The gold standard of vessel. (b–e) The vessel segmented from the synthetic data added with Gaussian white noise σ2 = 0, 30, 45, and 60, respectively.

Table 1.

Segmentation performance of the improved fuzzy connectedness method on the synthetic dataset.

Group 1 (n=10) Group 2 (n=10) Group 3 (n=10) Group 4 (n=10) Group 5 (n=10) Group 6 (n=10)
ACC (%) SEN (%) SPE (%) DICE (%) ACC (%) SEN (%) SPE (%) DICE (%) ACC (%) SEN (%) SPE (%) DICE (%) ACC (%) SEN (%) SPE (%) DICE (%) ACC (%) SEN (%) SPE (%) DICE (%) ACC (%) SEN (%) SPE (%) DICE (%)
σ 2 = 0 99.7 ± 0.1 96.6 ± 5.1 99.7 ± 0.2 84.3 ± 3.6 99.8 ± 0.1 91.2 ± 4.9 99.8 ± 0.1 86.6 ± 2.6 99.8 ± 0.1 91.5 ± 3.5 99.8 ± 0.1 85.4 ± 3.1 99.8 ± 0.1 91.5 ± 3.5 99.8 ± 0.1 86.0 ± 3.6 99.8 ± 0.1 91.5 ± 4.0 99.9 ± 0.1 85.4 ± 2.4 99.7 ± 0.3 91.0 ± 3.9 99.9 ± 0.1 85.6 ± 3.0
σ 2 = 30 99.7 ± 0.2 86.7 ± 4.9 99.7 ± 0.2 83.8 ± 4.1 99.7 ± 0.1 87.5 ± 7.3 99.8 ± 0.1 82.2 ± 3.3 99.7 ± 0.1 89.4 ± 4.7 99.8 ± 0.1 80.7 ± 3.1 99.7 ± 0.1 89.9 ± 5.2 99.7 ± 0.1 80.7 ± 3.6 99.7 ± 0.1 90.4 ± 5.2 99.8 ± 0.1 81.0 ± 2.5 99.7 ± 0.1 88.6 ± 6.6 99.8 ± 0.1 80.2 ± 2.5
σ 2 = 45 99.5 ± 6.3 91.9 ± 4.6 99.7 ± 0.1 81.3 ± 3.0 99.7 ± 0.1 88.3 ± 6.1 99.8 ± 0.1 81.8 ± 4.1 99.7 ± 0.1 90.4 ± 4.5 99.8 ± 0.1 80.7 ± 3.1 99.7 ± 0.1 89.7 ± 5.3 99.8 ± 0.1 80.8 ± 3.5 99.7 ± 0.1 89.1 ± 5.3 99.8 ± 0.1 80.9 ± 2.8 99.7 ± 0.1 87.8 ± 6.6 99.8 ± 0.1 80.6 ± 2.7
σ 2 = 60 99.6 ± 0.1 91.4 ± 4.3 99.7 ± 0.1 80.6 ± 2.4 99.7 ± 0.1 88.2 ± 5.5 99.8 ± 0.1 82.2 ± 4.1 99.7 ± 0.1 90.3 ± 4.5 99.8 ± 0.1 81.0 ± 3.1 99.7 ± 0.1 90.2 ± 5.1 99.8 ± 0.1 81.1 ± 3.3 99.7 ± 0.1 88.6 ± 5.1 99.8 ± 0.1 81.4 ± 2.6 99.7 ± 0.1 87.2 ± 6.4 99.8 ± 0.1 81.1 ± 2.7

ACC = accuracy; SEN = sensitivity; SPE = specificity; DICE = Dice coefficient.

Figure 4 shows the vessel-enhanced image by using the improved vesselness filter. Figure 4(a) shows the original CT image; Figure 4(b) shows the adaptive sigmoid filtered image; Figure 4(c) shows the isotropic resampled image; and Figure 4(d) shows the improved vesselness filtered image. It can be seen that the improved vesselness filter can effectively enhance the vessel while suppressing the background. The vesselness images obtained by using the Jerman's vesselness filter and the improved vesselness filter are shown in Figure 5. The intensity of the vesselness images ranged from 0 to 1. The contrast of vessel in CT images increased from Figures 5(a)to 5(c). Note that the Jerman's vesselness filter enhanced the liver contour and almost neglected the liver vessel for the low-contrast image (Figure 5(d)). As the image contrast increased, there was still undesired enhancement at the liver contour (Figures 5(e) and 5(f)). In addition, the Jerman's vesselness filter could not effectively suppress the background (Figure 5(f)). By contrast, the improved vesselness filter successfully enhanced the vessel while suppressing the background, with little enhancement at the liver contour (Figures 5(g)5(i)).

Figure 4.

Figure 4

Illustration of the improved vesselness filter. (a) The original CT image. (b) The adaptive sigmoid filtered image of (a). (c) The isotropic resampled image of (b). (d) The improved vesselness filtered image of (c).

Figure 5.

Figure 5

Comparison of the improved vesselness filter with the Jerman's vesselness filter. (a)–(c) The original CT images. (d)–(f) The vesselness images by using the Jerman's vesselness filtering to (a)–(c), respectively. (g)–(i) The vesselness images by using the improved vesselness filtering to (a)–(c), respectively.

Figure 6 shows the liver vessel segmented by using the improved FC, depicted by yellow contour or surface. The gold standard of liver vessel is indicated by red contour or surface. The axial slices, sagittal slices, coronal slices, and 3D view are shown in Figures 6(a)6(c), 6(d)6(f), 6(g)6(i), and 6(j)-6(k), respectively. It can be seen that the proposed method yielded satisfying segmentation performance. Figure 7 shows typical CT images from the 3Dircadb (Figures 7(a)7(i)) and Sliver07 (Figures 7(j)7(r)) datasets. The original CT images in Figures 7(a)7(i) and 7(j)7(r) are used in Figures 8 and 9, respectively. For vessel segmentation, Figures 7(a)7(c) and 7(j)7(l) were of high contrast, while Figures 7(d)7(i) and 7(m)7(r) were of low contrast.

Figure 6.

Figure 6

Liver vessel segmented by using the proposed method (yellow). The gold standard of vessel is depicted in red. Each column corresponds to one case. (a)–(c) The axial slices. (d)–(f) The sagittal slices. (g)–(i) The coronal slices. (j)–(k) The 3D view.

Figure 7.

Figure 7

Typical CT images from the 3Dircadb (a)–(i) and Sliver07 (j)–(r) datasets. Each row corresponds to one case. The first, second, and third columns are axial slices, sagittal slices, and coronal slices, respectively. The original CT images in (a)–(i) and (j)–(r) are used in Figures 8 and 9, respectively. For vessel segmentation, (a)–(c) and (j)–(l) are of high contrast, while (d)–(i) and (m)–(r) are of low contrast.

Figure 8.

Figure 8

Comparison of the improved fuzzy connectedness (FC) method with traditional segmentation algorithms. Each column corresponds to one case. The original CT images of each case are shown in Figures 7(a)7(i). (a)–(c) The gold standard of liver vessel. (d)–(f) The vessel segmented by using the improved FC. (g)–(i) The vessel segmented by using traditional FC with multiple potential seeds indicated by the blue points in Figure 2(b). (j)–(l) The vessel segmented by using traditional FC with one single seed indicated by the red point in Figure 2(b). (m)–(o) The vessel segmented by using region growing with multiple potential seeds. (p)–(r) The vessel segmented by using threshold level set with multiple potential seeds.

Figure 9.

Figure 9

Comparison of the improved fuzzy connectedness (FC) method with traditional segmentation algorithms. Each column corresponds to one case. The original CT images of each case are shown in Figures 7(j)7(r). (a)–(c) The gold standard of liver vessel. (d)–(f) The vessel segmented by using the improved FC. (g)–(i) The vessel segmented by using traditional FC with multiple potential seeds indicated by the blue points in Figure 2(b). (j)–(l) The vessel segmented by using traditional FC with one single seed indicated by the red point in Figure 2(b). (m)–(o) The vessel segmented by using region growing with multiple potential seeds. (p)–(r) The vessel segmented by using threshold level set with multiple potential seeds.

Figures 8 and 9 show the comparison of the improved FC method with traditional segmentation algorithms, including traditional FC [27], region growing [27], and threshold level set [27]. Figures 8(a)8(c) and 9(a)9(c) show the gold standard of liver vessel. Figures 8(d)8(f) and 9(d)9(f) show the vessel segmented by using the improved FC method. Figures 8(g)8(i) and 9(g)9(i) show the vessel segmented by using traditional FC with multiple potential seeds indicated by the blue points in Figure 2(b). Figures 8(j)8(l) and 9(j)9(l) show the vessel segmented by using threshold level set with multiple potential seeds. Traditional FC with one single seed could not segment completely the liver vessel (Figures 8(j)8(l) and 9(j)9(l)). When multiple potential seeds were used for traditional FC, the segmentation performance show the vessel segmented by using traditional FC with one single seed indicated by the red point in Figure 2(b). Figures 8(m)8(o) and 9(m)9(o) show vessel segmented by using region growing with multiple potential seeds. Figures 8(p)8(r) and 9(p)9(r) shows improved, but it was still unsatisfying (Figures 8(g)8(i) and 9(g)9(i)). For region growing and threshold level set with multiple potential seeds, both undersegmentation and oversegmentation of vessel occurred (Figures 8(m)8(r) and 9(m)9(r)). It is interesting to discuss the segmentation performance on low-contrast cases shown in Figures 7(d)7(i) and 7(m)7(r). If a part of the main vessel was low-contrast, it would be totally unsegmented, as indicated by the black arrows in Figures 8(h), 8(n), 8(q) and 9(h), 9(n), 9(q). When the peripheral vessel was low-contrast, it would be merged (Figures 8(h) and 8(q)) or missed (Figures 9(h), 9(n), 9(d)), as indicated by blue arrows. Even in the high-contrast images shown in Figures 7(a)7(c) and 7(j)7(l), part of the vessel was segmented falsely (Figures 8(m) and 8(p)) and the periphery vessel was not segmented (Figures 8(g) and 9(m), 9(p)), as indicated by green arrows. By contrast, the proposed method was capable to segment completely the liver vessel, even for the low-contrast images.

To compare further the improved vesselness filter with the Jerman's vesselness filter, the liver vessel segmented by using the improved FC method on the basis of the Jerman's vesselness filtering, rather than the improved vesselness filtering, is shown in Figure 10. Figures 10(a) and 10(b) show that the Jerman's vesselness filter falsely enhanced the liver contour. Figure 10(c) shows that the Jerman's vesselness filter could not effectively suppress the background (nonvessel) tissues. Quantitative comparison of the improved FC (with one single seed) with traditional FC, region growing, and threshold level set (with multiple seeds) on the 3Dircadb (n=20) and Sliver07 (n=20) datasets are listed in Table 2 and shown in Figure 11, in terms of accuracy, sensitivity, specificity, and Dice coefficient. It can be observed that the improved FC outperformed traditional FC, region growing, and threshold level set. The average accuracy, sensitivity, specificity, and Dice coefficient of the improved FC method were, respectively, (96.4 ± 1.1)%, (73.7 ± 7.6)%, (97.4 ± 1.3)%, and (67.3 ± 5.7)% for the 3Dircadb dataset and (96.8 ± 0.6)%, (89.1 ± 6.8)%, (97.6 ± 1.1)%, and (71.4 ± 7.6)% for the Sliver07 dataset.

Figure 10.

Figure 10

The liver vessel segmented by using the improved fuzzy connectedness method on the basis of the Jerman's vesselness filtering, rather than the improved vesselness filtering. (a) and (b) The Jerman's vesselness filter falsely enhances the liver contour. (c) The Jerman's vesselness filter fails to effectively suppress the background (nonvessel) tissues.

Table 2.

Segmentation performance of the improved fuzzy connectedness (with one single seed), traditional fuzzy connectedness, region growing, and threshold level set (with multiple potential seeds).

Improved fuzzy connectedness Fuzzy connectedness Region growing Threshold level set
ACC (%) SEN (%) SPE (%) DICE (%) ACC (%) SEN (%) SPE (%) DICE (%) ACC (%) SEN (%) SPE (%) DICE (%) ACC (%) SEN (%) SPE (%) DICE (%)
3Dircadb (n=20) 96.4 ± 1.1 73.7 ± 7.6 97.4 ± 1.3 67.3 ± 5.7 82.2 ± 21.9 67.1 ± 25.5 82.5 ± 24.7 40.4 ± 16.2 84.1 ± 21.1 66.1 ± 25.0 86.4 ± 23.4 42.4 ± 15.6 94.3 ± 4.5 51.5 ± 20.3 97.0 ± 5.2 54.0 ± 13.9
Sliver07 (n=20) 96.8 ± 0.6 89.1 ± 6.8 97.6 ± 1.1 71.4 ± 7.6 82.9 ± 25.5 85.0 ± 17.2 83.1 ± 26.7 45.1 ± 18.8 90.6 ± 12.6 84.1 ± 19.2 90.7 ± 13.6 56.2 ± 19.4 95.3 ± 2.9 82.5 ± 6.9 95.3 ± 3.2 67.5 ± 6.7

ACC = accuracy; SEN = sensitivity; SPE = specificity; DICE = Dice coefficient.

Figure 11.

Figure 11

Comparison of the improved fuzzy connectedness (with one single seed) with traditional fuzzy connectedness, region growing, and threshold level set (with multiple potential seeds). (a) and (b) show the accuracy, sensitivity, specificity, and Dice coefficient for the 3Dircadb and Sliver07 datasets, respectively. ACC = accuracy; SEN = sensitivity; SPE = specificity; DICE = Dice coefficient; IFC = improved fuzzy connectedness; FC = fuzzy connectedness; RG = region growing; LS = threshold level set.

4. Discussion

4.1. Significance of This Study

3D liver vessel segmentation is critical in computer-assisted liver tumor treatment planning and navigation. FC is an emerging method for image segmentation. However, traditional FC obtained unsatisfying performance for liver vessel segmentation in CT images, and it required multiple seeds and was sensitive to initialization. To address these issues, an improved FC method was proposed in this paper. Our method was fully automatic. The main contributions of this study were as follows. The Jerman's vesselness filter was improved by incorporating adaptive sigmoid filtering and a background-suppressing item. The improved vesselness filter effectively enhanced the vessel and suppressed the background. The improved vesselness response was incorporated into the fuzzy affinity function, increasing the segmentation performance of FC. The fuzzy scene was initialized by two-threshold Otsu with one single seed, reducing the number of seeds and the sensitivity to initialization in traditional FC.

4.2. Implementation Details of the Algorithms

The algorithms described in this paper were implemented by using C++ and the Insight Segmentation and Registration Toolkit (ITK) (http://itk.org) [27]. The following ITK classes were mainly used:

  1. The improved vesselness filter was implemented on the basis of the class itk::HessianToObjectnessMeasureImageFilter.

  2. The improved FC method was implemented on the basis of the class itk::SimpleFuzzyConnectednessScalarImageFilter. This class was also used for the traditional FC segmentation.

  3. The K-means clustering was implemented by using the class itk::Statistics::ScalarImageKmeansImageFilter.

  4. The class itk::SigmoidImageFilter was used for sigmoid filtering.

  5. The classes itk::ResampleImageFilter and itk::IdentityTransform were used for isotropic resampling.

  6. The class itk::ConfidenceConnectedImageFilter was used for region growing segmentation.

  7. The class itk::ThresholdSegmentationLevelSetImageFilter was used for threshold level set segmentation.

Average run time of the proposed algorithm was 200 s for 3Dircadb and 210 s for Sliver07. The improved vesselness filtering took approximately 30 s. The improved FC segmentation also took nearly 30 s. Each of the isotropic resampling and anisotropic resampling took around 60 s.

4.3. Sensitivity of the Proposed Algorithm to Key Algorithmic Parameters

Sensitivity analysis of key algorithmic parameters in Algorithms 1 and 2 was performed. The vessel scales σmin and σmax were set on the basis of the findings of Luu et al. [4]. Here, two key algorithmic parameters were analyzed: the threshold τ in the improved vesselness filter and the threshold T in the improved FC. The threshold τ in vesselness filter determined the degree of piecewise compensation on the eigenvalue λ3. In theory, the smaller the threshold τ is, the more enhancement on the vessel boundary would be obtained; however, a too small τ is prone to cause undersegmentation. The threshold T in FC determined the degree of undersegmentation or oversegmentation. A too small T caused undersegmentation, while a too large T resulted in oversegmentation. The value of T from 0.01 to 0.09 was tested, as the segmented vasculature would be incomplete when T > 0.1. For the compromise between undersegmentation and oversegmentation, the value of T was firstly fixed to 0.05 to analyze the sensitivity of the proposed algorithm to the threshold τ. Figure 12 shows the average accuracy, sensitivity, specificity, and Dice coefficient of the proposed method on 10 cases randomly selected from the 3Dircadb dataset. The threshold τ ranged from 0.1 to 0.9 (T = 0.05). The accuracy and Dice coefficient reached peak when τ was optimally set at 0.6. Then, the value of τ was fixed to 0.6 to analyze the sensitivity of the proposed algorithm to the threshold T. Figure 13 shows the segmentation performance of the proposed method on the 10 randomly selected cases, with T ranging from 0.01 to 0.09 (τ=0.6). Based on the maximum value of the accuracy and Dice coefficient, the parameter T was optimally set at 0.05.

Figure 12.

Figure 12

Evaluation of the segmentation performance of the improved fuzzy connectedness method on 10 cases randomly selected from the 3Dircadb dataset for the values of T ranging from 0.1 to 0.9 (T = 0.05). The value of T is optimally set at 0.6. ACC = accuracy; SEN = sensitivity; SPE = specificity; DICE = Dice coefficient.

Figure 13.

Figure 13

Evaluation of the segmentation performance of the improved fuzzy connectedness method on 10 cases randomly selected from the 3Dircadb dataset for the values of T ranging from 0.01 to 0.09 (T = 0.6). The value of T is optimally set at 0.05. ACC = accuracy; SEN = sensitivity; SPE = specificity; DICE = Dice coefficient.

4.4. Comparison with Related Work

Table 3 shows a comparison of the improved FC method with related work in terms of segmentation method, dataset, number of cases, automation, precision, and run time. For the run time of the proposed method, it should be noted that each of the improved vesselness filtering and the improved FC segmentation only took around 30 s.

Table 3.

Comparison of the proposed method with related work.

Author Year Method Dataset # Automation Precision (%) Run time (s)
Oliveira et al. [7] 2011 RG Sliver07 20 Auto
Luu et al. [4] 2015 RG Clinical CTA 51 Auto ACC = 86.2; SEN = 85.1; SPE = 92.3
Esneault et al. [10] 2010 GC Clinical CTA 1 Auto 10–100
Zeng et al. [12] 2017 GC Clinical CTA 6 Auto ACC = 97.7; SEN = 79.8; SPE = 98.6 390
Sangsefidi et al. [11] 2018 GC 3Dircadb/Clinical CTA 7 Auto DICE = 74.0 560
Shang et al. [9] 2011 LS Clinical CTA 20 Auto SEN = 91.0 480
Ahmadi et al. [28] 2016 FCC Sliver07 20 Auto ACC = 91.0; SEN = 94.1; SPE = 83.6 27.1
Zeng et al. [13] 2016 ML Clinical CTA 6 Auto ACC = 98.1; SEN = 74.2; SPE = 99.3 0.05–0.1
Guo et al. [15] 2015 FC Clinical CTA 4 Semi 112.5
Wang et al. [16] 2016 FC Clinical CTA 3 Semi 22
Huang et al. [14] 2018 DL 3Dircadb 20 Auto ACC = 97.1; SEN = 74.3; SPE = 98.3; DICE = 67.5 230
Ours 2018 IFC 3Dircadb
Sliver07
20
20
Auto
Auto
ACC = 96.4; SEN = 73.7; SPE = 97.4; DICE = 67.3
ACC = 96.8; SEN = 84.4; SPE = 97.6; DICE = 71.4
200
210

Evaluation by the number of vascular nodes; CTA = computed tomography angiography; RG = region growing; GC = graph cuts; LS = level set; FCC = fuzzy C-means clustering; ML = machine learning; FC = fuzzy connectedness; DL = deep learning; IFC = improved fuzzy connectedness; ACC = accuracy; SEN = sensitivity; SPE = specificity; DICE = Dice coefficient; Auto = automatic; Semi = semiautomatic.

Firstly, the proposed method was compared with related work that used the 20 cases of the Sliver07 training dataset. Oliveira et al. [7] used region growing for liver vessel segmentation, but they only performed visual assessment for the segmentation. Ahmadi et al. [28] segmented liver vessel by using fuzzy C-means clustering and initialized the parameters by the genetic algorithm. Though the run time was shorter, the training process was more complex, and the accuracy and specificity of Ahmadi et al. [28] were lower than those of the proposed method. Then, the proposed method was compared with related work that used the 3Dircadb dataset. Huang et al. [14] segmented liver vessel on the 20 cases of 3Dircadb by using the 3D U-Net network. Their method reduced the need for the quantity of training data, but it required long training time (48 h). The accuracy, sensitivity, specificity, and Dice coefficient of Huang et al. [14] were slightly higher than those of the proposed method. Sangsefidi et al. [11] employed graph cuts for segmenting liver vessel, but they evaluated their method on only few cases of 3Dircadb.

Finally, the proposed method was compared with related work that used clinical data other than Sliver07 and 3Dircadb. These studies mostly used CT angiography (CTA) images, which were specific CT for vasculature with clear vascular boundary. However, in the context of computer-assisted liver tumor treatment planning and navigation, CE-CT images may be used more commonly, as liver tumors could be observed in CE-CT images. Though region growing methods had relatively higher operation efficiency, they are depended on the number and distribution of seeds, resulting in unsatisfied segmentation performance even in high-contrast CTA images [4]. Graph cuts and level set methods would take long time to segment liver vessel [912]. Esneault et al. [10] just showed the segmentation on one case of data, and they reported that the segmented vascular branches needed to be registered, which would take more time. Zeng et al. [12] reported that their method only achieved good performance on high-contrast CTA images, so their method might be restricted in practical applications when only low-contrast CT images are available. The similar issue existed in Shang et al. [9] and Zeng et al. [13]. Shang et al. [9] evaluated the sensitivity by the number of vascular nodes (denoted as SEN∗ in Table 3), but this evaluation metric may not be rigorous. In comparison with Guo et al. [15] and Wang et al. [16] which increased the time efficiency of traditional FC, this study focused on improving the segmentation performance and reducing the number of seeds and the sensitivity to initialization. In addition, our method did not require manual interaction to select the seed.

4.5. Limitations and Future Work

One limitation of this study is the small number of clinical data with the gold standard (40 cases). More clinical data may be used in future (if possible) to further verify the performance of the proposed method. In addition, the algorithmic steps of isotropic resampling and anisotropic resampling are time consuming, each taking around 60 s. This limitation may be overcome in future work.

5. Conclusions

An improved FC method was presented for automatic liver vessel segmentation in CT volumetric images. The Jerman's vesselness filter was improved by incorporating adaptive sigmoid filtering and a background-suppressing item. The improved vesselness filter effectively enhanced the liver vessel while suppressing the background. The improved vesselness response was incorporated into the fuzzy affinity function of FC. The fuzzy scene was initialized by two-threshold Otsu with one single seed generated automatically, reducing the number of seeds and the sensitivity to initialization in traditional FC. The improved FC method was evaluated on 40 cases of clinical CT volumetric images. Experimental results showed that the proposed liver vessel segmentation strategy could achieve better segmentation performance than traditional FC, region growing, and threshold level set. It is concluded that the proposed algorithm may be used as a new method for automatic 3D liver vessel segmentation in CT images.

Acknowledgments

This work was supported by the National Natural Science Foundation of China (Grant Nos. 61801312 and 71661167001), Beijing Natural Science Foundation (Grant No. 4184081), China Postdoctoral Science Foundation (Grant No. 2017M620566), Postdoctoral Research Fund of Chaoyang District, Beijing (Grant No. 2017ZZ-01-03), and Basic Research Fund of Beijing University of Technology.

Data Availability

The VascuSynth dataset is publically available at http://vascusynth.cs.sfu.ca/. The 3Dircadb dataset is publically available at http://www.ircad.fr/research/3dircadb. The training data of the Sliver07 dataset are publically available at http://www.sliver07.org/.

Conflicts of Interest

The authors declare that they have no conflicts of interest.

References

  • 1.Ferlay J., Soerjomataram I., Dikshit R., et al. Cancer incidence and mortality worldwide: sources, methods and major patterns in GLOBOCAN 2012. International Journal of Cancer. 2015;136(5):E359–E386. doi: 10.1002/ijc.29210. [DOI] [PubMed] [Google Scholar]
  • 2.Chen W., Zheng R., Baade P. D., et al. Cancer statistics in China. CA: A Cancer Journal for Clinicians. 2016;66(2):115–132. doi: 10.3322/caac.21338. [DOI] [PubMed] [Google Scholar]
  • 3.Lee M. W., Raman S. S., Asvadi N. H., et al. Radiofrequency ablation of hepatocellular carcinoma as bridge therapy to liver transplantation: a 10-year intention-to-treat analysis. Hepatology. 2017;65(6):1979–1990. doi: 10.1002/hep.29098. [DOI] [PubMed] [Google Scholar]
  • 4.Luu H. M., Klink C., Moelker A., et al. Quantitative evaluation of noise reduction and vesselness filters for liver vessel segmentation on abdominal CTA images. Physics in Medicine and Biology. 2015;60(10):3905–3926. doi: 10.1088/0031-9155/60/10/3905. [DOI] [PubMed] [Google Scholar]
  • 5.Moghbel M., Mashohor S., Mahmud R., Saripan M. I. B. Review of liver segmentation and computer assisted detection/diagnosis methods in computed tomography. Artificial Intelligence Review. 2018 doi: 10.1007/s10462-017-9550-x. In press. [DOI] [Google Scholar]
  • 6.Selle D., Preim B., Schenk A., Peitgen H. O. Analysis of vasculature for liver surgical planning. IEEE Transactions on Medical Imaging. 2002;21(11):1344–1357. doi: 10.1109/tmi.2002.801166. [DOI] [PubMed] [Google Scholar]
  • 7.Oliveira D. A., Feitosa R. Q., Correia M. M. Segmentation of liver, its vessel and lesions from CT images for surgical planning. BioMedical Engineering OnLine. 2011;10(1):p. 30. doi: 10.1186/1475-925x-10-30. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Wang Y., Fang B., Pi J., et al. Automatic multi-scale segmentation of intrahepatic vessel in CT images for liver surgery planning. International Journal of Pattern Recognition and Artificial Intelligence. 2013;27(1) doi: 10.1142/s0218001413570012.1357001 [DOI] [Google Scholar]
  • 9.Shang Y., Deklerck R., Nyssen E., et al. Vascular active contour for vessel tree segmentation. IEEE Transactions on Biomedical Engineering. 2011;58(4):1023–1032. doi: 10.1109/tbme.2010.2097596. [DOI] [PubMed] [Google Scholar]
  • 10.Esneault S., Lafon C., Dillenseger J. L. Liver vessel segmentation using a hybrid geometrical moments/graph cuts method. IEEE Transactions on Biomedical Engineering. 2010;57(2):276–283. doi: 10.1109/tbme.2009.2032161. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Sangsefidi N., Foruzan A. H., Dolati A. Balancing the data term of graph-cuts algorithm to improve segmentation of hepatic vascular structures. Computers in Biology and Medicine. 2017;93:117–126. doi: 10.1016/j.compbiomed.2017.12.019. [DOI] [PubMed] [Google Scholar]
  • 12.Zeng Y. Z., Zhao Y. Q., Tang P., et al. Liver vessel segmentation and identification based on oriented flux symmetry and graph cuts. Computer Methods and Programs in Biomedicine. 2017;150:31–39. doi: 10.1016/j.cmpb.2017.07.002. [DOI] [PubMed] [Google Scholar]
  • 13.Zeng Y. Z., Zhao Y. Q., Liao M., et al. Liver vessel segmentation based on extreme learning machine. Physica Medica. 2016;32(5):709–716. doi: 10.1016/j.ejmp.2016.04.003. [DOI] [PubMed] [Google Scholar]
  • 14.Huang Q., Sun J., Ding H., Wang X., Wang G. Robust liver vessel extraction using 3D U-Net with variant dice loss function. Computers in Biology and Medicine. 2018;101:153–162. doi: 10.1016/j.compbiomed.2018.08.018. [DOI] [PubMed] [Google Scholar]
  • 15.Guo X., Huang S., Fu X., Huang X. Vascular segmentation in hepatic CT images using adaptive threshold fuzzy connectedness method. Biomedical Engineering Online. 2015;14(1):p. 57. doi: 10.1186/s12938-015-0055-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Wang L., Li D., Huang S. An improved parallel fuzzy connected image segmentation method based on CUDA. Biomedical Engineering Online. 2016;15(1):p. 56. doi: 10.1186/s12938-016-0165-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Jerman T., Pernuš F., Likar B., Špiclin Ž. Enhancement of vascular structures in 3D and 2D angiographic images. IEEE Transactions on Medical Imaging. 2016;35(9):2107–2118. doi: 10.1109/tmi.2016.2550102. [DOI] [PubMed] [Google Scholar]
  • 18.Wu W., Zhou Z., Wu S., Zhang Y. Automatic liver segmentation on volumetric CT images using supervoxel-based graph cuts. Computational and Mathematical Methods in Medicine. 2016;2016 doi: 10.1155/2016/9093721.9093721 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Jassi P., Hamarneh G. Vascusynth: vascular tree synthesis software. Insight Journal. 2011:1–12. http://hdl.handle.net/10380/3260. [Google Scholar]
  • 20.Hamarneh G., Jassi P. VascuSynth: simulating vascular trees for generating volumetric image data with ground-truth segmentation and tree analysis. Computerized Medical Imaging and Graphics. 2010;34(8):605–616. doi: 10.1016/j.compmedimag.2010.06.002. [DOI] [PubMed] [Google Scholar]
  • 21.Sato Y., Nakajima S., Shiraga N., et al. Three-dimensional multi-scale line filter for segmentation and visualization of curvilinear structures in medical images. Medical Image Analysis. 1998;2(2):143–168. doi: 10.1016/s1361-8415(98)80009-1. [DOI] [PubMed] [Google Scholar]
  • 22.Frangi A. F., Niessen W. J., Vincken K. L., Viergever M. A. Multiscale vessel enhancement filtering. Proceedings of International Conference on Medical Image Computing and Computer-Assisted Intervention; Octomber 1998; Cambridge, MA, USA. Springer; pp. 130–137. [Google Scholar]
  • 23.Li Q., Sone S. Selective enhancement filters for nodules, vessels, and airway walls in two‐and three‐dimensional CT scans. Medical Physics. 2003;30(8):2040–2051. doi: 10.1118/1.1581411. [DOI] [PubMed] [Google Scholar]
  • 24.Erdt M., Raspe M., Suehling M. Medical Imaging and Augmented Reality. Heidelberg, Germany: Springer; 2008. Automatic hepatic vessel segmentation using graphics hardware; pp. 403–412. [Google Scholar]
  • 25.Xiao C., Staring M., Shamonin D., et al. A strain energy filter for 3D vessel enhancement with application to pulmonary CT images. Medical Image Analysis. 2011;15(1):112–124. doi: 10.1016/j.media.2010.08.003. [DOI] [PubMed] [Google Scholar]
  • 26.Pednekar A., Kakadiaris I. A., Kurkure U. Adaptive fuzzy connectedness-based medical image segmentation. Proceedings of Indian Conference on Computer Vision; December 2002; Graphics & Image Processing, Ahmedabad, India. [Google Scholar]
  • 27.Ibanez L., Schroeder W., Ng L., Cates J. The ITK Software Guide. Bethesda, Maryland: U.S. National Library of Medicine; 2005. [Google Scholar]
  • 28.Ahmadi K., Karimi A., Fouladi N. B. New technique for automatic segmentation of blood vessels in CT scan images of liver based on optimized fuzzy-means method. Computational and Mathematical Methods in Medicine. 2016;2016 doi: 10.1155/2016/5237191.5237191 [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The VascuSynth dataset is publically available at http://vascusynth.cs.sfu.ca/. The 3Dircadb dataset is publically available at http://www.ircad.fr/research/3dircadb. The training data of the Sliver07 dataset are publically available at http://www.sliver07.org/.


Articles from Journal of Healthcare Engineering are provided here courtesy of Wiley

RESOURCES