. 2018 Jul 30;12(6):1241–1243. doi: 10.1177/1932296818791536

The Incremental Benefits of the Predictive Low-Glucose Suspend Function Compared to the Low-Glucose Suspend Function as Automation Against Hypoglycemia in Sensor-Augmented Pump Therapy

Pilar Isabel Beato-Víbora 1,, Estela Gil-Poch 2, Laura Galán-Bueno 2, Lucía Lázaro-Martín 1, Francisco Javier Arroyo-Díez 2
PMCID: PMC6232727  PMID: 30058373

Sensor-augmented pump therapy (SAP) has demonstrated to improve HbA1c and reduce the frequency of hypoglycemia in patients with type 1 diabetes (T1D). Different levels of automation of insulin infusion in response to sensor readings have been introduced.1-3 Our aim was to prospectively evaluate the additional benefits provided by SAP with predictive low-glucose suspend function (PLGS) in T1D patients experienced in the use of SAP with low-glucose suspend feature (LGS).

Thirty-six patients, aged 35 ± 14 years, 69% females, diabetes duration (median [IQR]) 26 years [14-33], time on SAP with LGS (median [IQR]) 20 months [15-44], 56% with history of severe hypoglycemia, already using the MiniMed-VEO®, with the LGS activated, were simultaneously upgraded to MiniMed-640G® with PLGS activated. After 6 months, while routine clinical care was maintained, all the patients were reevaluated. One month of data from both SAP systems were downloaded and glycemic variability measures were calculated. Adult patients were asked to complete a set of questionnaires (Gold and Clarke score, Hypoglycemia Fear Survey [HFS], Diabetes Quality of Life [DQoL], Diabetes Treatment Satisfaction [DTS], and Pittsburgh Sleep Quality Index [PSQI]). All patients signed an informed consent form. The protocol was approved by the Local Research Ethics Committee.

No changes in HbA1c were found at the end of follow-up (both 7.1 ± 0.6%). No severe hypoglycemia episodes or diabetic ketoacidosis episodes were reported. The changes found after the upgrade from LGS to PLGS are summarized in Table 1. In the patients with impaired awareness of hypoglycemia, 45% (n = 14), Gold Score was reduced from 5.6 ± 1.1 to 4.7 ± 1.6 (P = .022), and 3 out of the 14 patients (21%) regained hypoglycemia awareness. Of the adults, 65% (n = 20) reported being less worried about hypoglycemia, mainly at night, with PLGS and all of them agreed to keep using SAP with PLGS.

Table 1.

Changes After the Upgrade From Low-Glucose Suspend Function (LGS) to Predictive Low-Glucose Suspension (PLGS).

LGS PLGS P
Sensor data
 Mean sensor glucose (mg/dl) 153 ± 21 160 ± 21 .002
 Time <70 mg/dl (%, minutes/day) 4.1 ± 3.2
59 ± 46
3.0 ± 2.4
43 ± 35
.013
 Time <60 mg/dl (%, minutes/day) 1.8 ± 1.9
26 ± 27
1.4 ± 1.4
20 ± 2
.162
 Time <54 mg/dl (%, minutes/day) 1.0 ± 1.4
14 ± 20
0.8 ± 1.0
12 ± 14
.211
 Time <50 mg/dl (%, minutes/day) 0.7 ± 1.1
10 ± 16
0.6 ± 0.8
9 ± 12
.589
 Time >180 mg/dl (%, minutes/day) 27.0 ± 14.1
389 ± 202
32.4 ± 15.0
467 ± 216
.003
 Time >250 mg/dl (%, minutes/day) 8.9 ± 14.6
128 ± 210
7.9 ± 6.5
114 ± 94
.064
 Time >300 mg/dl (%, minutes/day) 1.8 ± 2.1
26 ± 30
2.2 ± 2.9
32 ± 42
.05
 Time of sensor use (%) 78 ± 20 82 ± 6 .051
System performance
 High alarms (number/day) 3.3 ± 3.5 5.0 ± 4.0 .002
 Low alarms (number/day) 5.1 ± 4.1 1.9 ± 4.0 .001
 “On-low” suspend events (minutes/day) 33 ± 39 15 ± 33 .023
 “On-low” suspend events (number/day) 1.03 ± 1.02 0.25 ± 0.52 .001
 “Before-low” suspend events (number/day) NA 2.6 ± 1.4 NA
 “Before-low” suspend events (minutes/day) NA 151 ± 94 NA
Glycemic variability measures
 SD (mg/dl) 54 ± 10 55 ± 1 .303
 CV (mg/dl) 35.6 ± 5.4 34.6 ± 4.5 .131
 CONGA 133.7 ± 19.4 138.7 ± 20.1 .02
 HBGI 361 ± 32 371 ± 32 .003
 MODD 58.1 ± 11.9 57.5 ± 11.4 .663
 MAGE 107 ± 25 115 ± 26 .005
 ADRR 513.7 ± 37.6 519.0 ± 3 6.0 .176
 M value 2464 ± 327 2570 ± 327 .003
 MAG 52.07 ± 11.37 51.6 ± 8.8 .676
Questionnaire scores
 Hypoglycemia Fear Survey (HFS) 50 ± 24 37 ± 25 .001
 HFS (behavior) 18 ± 8 14 ± 9 .025
 HFS (worry) 32 ± 18 24 ± 18 .01
 Diabetes Quality of Life (DQoL) 86 ± 17 84 ± 20 .640
 Diabetes Treatment Satisfaction (DTS) 31 ± 4 30 ± 4 .021
 Pittsburgh Sleep Quality Index (PSQI) > 5 (poor sleep quality), n (%) 22 (71) 15 (50) .01

N = 36. Data are expressed as mean ± standard deviation, unless otherwise indicated.

Few studies have specifically addressed the advantages of PLGS in patients already experienced with the use of LGS. Biester et al4 showed a 59% reduction in time <70 mg/dl in 6 children, as a result of an unintentional study protocol deviation. Zhong et al5 retrospectively analyzed anonymized data from 851 users who had switched from SAP with LGS to SAP with PLGS, finding a 40% reduction in the time spent <70 mg/dl. Gómez et al6 found a reduction in the frequency of severe hypoglycemia.

Our study has several strengths. The analysis of glycemic variability and patient-reported outcomes give a comprehensive assessment of the advantages and weaknesses of the PLGS. Testing the system in patients already benefiting from automation against hypoglycemia, with an optimal glycemic control and a low rate of hypoglycemia, has challenged the possibilities of the new system. Also, the impact of diabetes technology on sleep quality has been previously very scarcely studied.

In summary, we present a prospective study evaluating the benefits of SAP with PLGS compared to SAP with LGS, showing a reduction in hypoglycemia frequency along with an improvement in patients’ fear of hypoglycemia and the sleep quality, while an optimal and stable glycemic control was maintained.

Footnotes

Abbreviations: ADRR, Average Daily Risk Range; CONGA, Continuous Overall Net Glycaemic Sction; CV, coefficient of variation; DQoL, Diabetes Quality of Life; DTS, Diabetes Treatment Satisfaction; HBGI, High Blood Glucose Index; HFS, Hypoglycemia Fear Survey; IQR, interquartile range; LGS, low-glucose suspend; MAG, Mean Absolute Glucose; MAGE, Mean Amplitude of Glycaemic Excursions; MODD, Mean Of Daily Differences; PLGS, predictive low-glucose suspend; PSQI, Pittsburgh Sleep Quality Index; SAP, sensor-augmented pump therapy; SD, standard deviation; SMBG, self-monitoring of blood glucose; T1D, type 1 diabetes.

Declaration of Conflicting Interests: The author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Funding: The author(s) received no financial support for the research, authorship, and/or publication of this article.

ORCID iD: Pilar Isabel Beato-Víbora Inline graphic https://orcid.org/0000-0003-4075-4969

References

  • 1. Buckingham BA, Raghinaru D, Cameron F, et al. In Home Closed Loop Study Group. Predictive low-glucose insulin suspension reduces duration of nocturnal hypoglycemia in children without increasing ketosis. Diabetes Care. 2015;38:1197-1204. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Battelino T, Nimri R. Prevention of hypoglycemia with predictive low glucose insulin suspension in children with type 1 diabetes: a randomized controlled trial. Diabetes Care. 2017;40:764-770. [DOI] [PubMed] [Google Scholar]
  • 3. Abraham MB, Nicholas JA, Smith GJ, et al. Reduction in hypoglycemia with the predictive low-glucose management system: a long-term randomized controlled trial in adolescents with type 1 diabetes. Diabetes Care. 2018;41:303-310. [DOI] [PubMed] [Google Scholar]
  • 4. Biester T, Kordonouri O, Holder M, et al. “Let the algorithm do the work”: reduction of hypoglycemia using sensor-augmented pump therapy with predictive insulin suspension (SmartGuard) in pediatric type 1 diabetes patients. Diabetes Technol Ther. 2017;19:173-182. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Zhong A, Choudhary P, McMahon C, et al. Effectiveness of automated insulin management features of the MiniMed(®) 640g sensor-augmented insulin pump. Diabetes Technol Ther. 2016;18:657-663. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Gómez AM, Henao DC, Imitola A, et al. Efficacy and safety of sensor-augmented pump therapy (SAPT) with predictive low-glucose management in patients diagnosed with type 1 diabetes mellitus previously treated with SAPT and low glucose suspend [published online ahead of print June 15, 2018]. Endocrinol Diabetes Nutr. doi: 10.1016/j.endinu.2018.03.009. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Diabetes Science and Technology are provided here courtesy of Diabetes Technology Society