Abstract
Most clinical measurements of radioactivity, whether for therapeutic or imaging nuclides, rely on commercial reentrant ionization chambers (“dose calibrators”). The National Institute of Standards and Technology (NIST) maintains a battery of representative calibrators and works to link calibration settings (“dial settings”) to primary radioactivity standards. Here, we provide a summary of NIST-determined dial settings for 22 radionuclides, including several previously unreported settings. In general, current manufacturer-provided calibration settings give activities that agree with NIST standards to within a few percent.
Keywords: radioactivity standards, quantitative imaging, secondary standards, calibration factor, dial setting, dose calibrator, ionization chamber
1. Introduction
Activity measurements in the radiopharmaceutical industry rely almost exclusively on commercial reentrant ionization chambers referred to variously as “radionuclide calibrators”, “activimeters”, or “dose calibrators”. These devices, consisting of a pressurized ionization chamber and an electrometer, typically provide a readout directly in activity units, relying on a specific calibration setting or “dial setting” (DS) to convert measured current to displayed activity. In 2000, Zimmerman and Cessna reported on the methodology and results of experiments at the National Institute of Standards and Technology (NIST) to determine DSs for 11 radionuclides commonly used in nuclear medicine applications [1]. In this letter, we update those results to include new imaging and therapeutic radionuclides. We provide appropriate references, where available, and briefly describe the measurements where the determinations are otherwise unpublished.
2. Methods
In most cases, dial setting determinations are carried out using the calibration curve method [1]. A source is measured at a series of DSs covering a range that includes the setting that returns the true activity. One advantage of this method is that it can be employed prior to a precision assay of the radioactive source as long as an approximate activity can be determined to select an appropriate range of DSs for the curve. Another advantage is that this method makes it easier to evaluate uncertainties since the standard error of the fit (calculated from the residuals of the fit to the calibration curve) provides a good estimate for the uncertainty due to the method and the sensitivity to small changes in DS. This approach also allows the uncertainty to be expressed in terms of both DS units and activity.
Uncertainties on determined DSs or, analogously, on the activities measured with them are typically mostly due to the uncertainty on the primary standard. Primary activity measurements are performed at NIST using a variety of techniques, among which coincidence counting [2,3] and liquid scintillation-based efficiency tracing or triple-to-double coincidence ratio (TDCR) counting [4,5] are most common. For radionuclides detected with high counting efficiency and relatively simple decay schemes, total combined uncertainties—determined using the methodology outlined by Taylor and Kuyatt [6]—on the order of 0.5 % are often achievable. With these standards, a dose calibrator can typically be calibrated to allow activity measurements with expanded (k = 2) uncertainties on the order of 1 %.
3. Results and Discussion
Table 1 provides NIST-determined dial settings and uncertainties for 22 radionuclides. For some nuclides, DSs are given for multiple geometries. The table points to relevant references for each DS.
Table 1.
NIST-determined calibration factors in pA∙MBq−1 for the Vinten 671 Ionization Chamber (VIC) or as dial settings (DSs) for Capintec or Biodex calibrators. The calibration factors with their expanded (k = 2) uncertainties are given in bold for each specific radionuclide and geometry. The standard geometry, denoted simply as “ampoule” in the table, is the NIST 5 mL flame-sealed ampoule containing 5 mL of radioactive solution. Below each calibration factor, the % uncertainty on the measured activity that results from the expanded uncertainty on the calibration factor (UA / %) is given. The year of the NIST primary standardization upon which the DS determination is based is given, along with the reference describing the DS determination, where available. In some cases, the DS is given with a multiplier, e.g., “57(2) X 10”, indicating that the reading using DS = 57 should be multiplied by 10 to give the correct activity; in this example the expanded uncertainty on the DS is 2 DS units.
| Nuclide | 1°std | Ref. | Geometry | Instrument |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
| VIC / pA·MBq−1 |
CRC-12 | CRC-15R | CRC-25PET | CRC-35R | AL-400 | AL-500 | ||||
| F-18 | 2012 | [7] | ampoule | 10.36(8) | 450(5) | 449(4) | 455(4) | 450(4) | ||
| in % Activity | 0.8 | 1.0 | 0.7 | 0.7 | 0.7 | |||||
| Cu-62 | 1999 | [1] | ampoule | 489(8) | ||||||
| in % Activity | 1.4 | |||||||||
| Cu-62 | 1999 | [1] | 35 mL syringe | 499(6) | ||||||
| in % Activity | 0.8 | |||||||||
| Cu-64 | 2016 | [8] | ampoule | 1.952(29) | 118(2) X 2 | 135(2) X 2 | 119(2) X 2 | 53.7(7) | 54.0(8) | |
| in % Activity | 1.5 | 1.2 | 1.3 | 1.9 | 1.4 | 1.4 | ||||
| Ge-68 | 2008 | [9] | ampoule | 10.04(9) | 440(5) | 440(7) | 448(4) | 442(4) | 10.5(1) | 10.1(1) |
| in % Activity | 0.9 | 1.0 | 1.4 | 0.7 | 0.8 | 1.0 | 0.9 | |||
| Ge-68 | 2008 | [9] | 3 mL epoxy in a mock 6 mL syringe | 10.00(10) | 453(7) | 436(5) | 451(5) | 447(9) | 10.5(1) | 10.1(1) |
| in % Activity | 1.0 | 1.2 | 1.0 | 1.0 | 1.7 | 0.9 | 1.0 | |||
| Ge-68 | 2008 | 6 mL cylindrical phantom insert | 10.12(10) | |||||||
| in % Activity | 1.0 | |||||||||
| Y-90 | 1993 | [10] | Ampoule1 | 48 X 10 | 48 X 10 | 48 X 10 | ||||
| in % Activity | ||||||||||
| Y-90 | 1993 | [11] | 10 mL syringe with 3 to 9 mL-YCl3 | 57(2) X 10 | 55(2) X 10 | 54(2) X 10 | ||||
| in % Activity | ||||||||||
| Tc-99m | 1976 | [1] | ampoule | 1.247(11) | 82 | 81(2) | 91(2) | 81(2) | 37.7(3) | 37.9(3) |
| in % Activity | 0.9 | 0.9 | 0.9 | 0.9 | 0.9 | 0.9 | ||||
| Tc-99m | 1976 | [1] | 10 mL Mallinckrodt dose vial | 78 | ||||||
| in % Activity | ||||||||||
| In-111 | 1977 | ampoule | 4.081(50) | 316(5) | 315(5) | 323(5) | 313(5) | 13.6(2) | 13.3(2) | |
| in % Activity | 1.2 | 1.3 | 1.2 | 1.3 | 1.2 | 1.2 | 1.3 | |||
| Sn-117m | 1998 | [12] | 3.5 mL dose vial2 | 133(6) | ||||||
| in % Activity | ||||||||||
| I-123 | 1976 | [13] | ampoule | 1.758(18) | 306(4) | 306(4) | 321(4) | 297(4) | 13.2(1) | 13.0(1) |
| in % Activity | 1.0 | 1.4 | 1.4 | 1.4 | 1.3 | 1.4 | 1.4 | |||
| I-125 | 1989 | [1] | ampoule3 | 313 | ||||||
| in % Activity | ||||||||||
| I-131 | 1978 | [15] | ampoule4 | 3.993(36) | 156(2) | 156(4) | 174(4) | 158(4) | ||
| in % Activity | 0.90 | 0.90 | 0.90 | 0.90 | 0.90 | |||||
| Ba-133 | 1983 | [15] | ampoule4 | 4.287(59) | 628(2) | 626(3) | 640(2) | 618(2) | ||
| in % Activity | 1.4 | 1.4 | 1.4 | 1.4 | 1.4 | |||||
| Xe-133 | 1988 | [1] | ampoule | 184 | ||||||
| in % Activity | ||||||||||
| Xe-133 | 1988 | [1] | 3 mL Dupont dose vial | 181 | ||||||
| in % Activity | ||||||||||
| Gd-153 | 1989 | [1] | ampoule | 545 | ||||||
| in % Activity | ||||||||||
| Sm-153 | 1985 | ampoule5 | 0.656(28) | 248(7) | 250(7) | 276(7) | 248(7) | 16.8(4) | 16.5(4) | |
| in % Activity | 4.2 | 2.1 | 2.1 | 2.1 | 2.1 | 2.1 | 2.1 | |||
| Ho-166 | 1991 | [16] | ampoule | 683(7) X 10 | 691(8) X 10 | |||||
| in % Activity | ||||||||||
| Ho-166 | 1991 | [16] | 16 mL in 20 mL glass dose vial | 672(11) X 10 | 678(19) X 10 | |||||
| in % Activity | ||||||||||
| Ho-166 | 1991 | [16] | 10 mL plastic dose vial | 703(8) X 10 | 712(4) X 10 | |||||
| in % Activity | ||||||||||
| Lu-177 | 2009 | ampoule6 | 0.344(3) | 451(5) X 10 | 449(5) X 10 | 435(5) X 10 | 450(8) X 10 | |||
| in % Activity | 0.76 | 1.0 | 0.9 | 1.0 | 1.5 | |||||
| Re-186 | 1989 | [1] | ampoule | 480 X 10 | ||||||
| in % Activity | ||||||||||
| Re-186 | 2000 | [17] | Ace Glass v-vial, 0.96 g to 2.12 g7 | 447(2) X 10 | ||||||
| in % Activity | 0.4 | |||||||||
| Re-186 | 2000 | [17] | Wheaton v-vials, 1.0 g to 2.2 g7 | 462(2) X 10 | ||||||
| in % Activity | 0.7 | |||||||||
| Re-188 | 1999 | [18] | ampoule8 | 631(4) X 10 | ||||||
| in % Activity | 0.4 | |||||||||
| Re-188 | 1999 | [18] | 5 mL SoloPak dose vial, 2 mL8 | 621(3) X 10 | ||||||
| in % Activity | 0.4 | |||||||||
| W-188 | 2002 | [19] | ampoule9 | 0.595(11) | ||||||
| in % Activity | 1.8 | |||||||||
| Ra-223 | 2013 | [20] | ampoule | 3.171(40) | 233(2) | 234(2) | 234(4) | 19.0(2) | 18.7(2) | |
| in % Activity | 1.2 | 0.65 | 0.75 | 1.1 | 0.87 | 0.87 | ||||
| Ra-223 | 2013 | [20] | ampoule | 3.18(2) | 231(5) | 231(6) | 234(5) | |||
| in % Activity | 0.63 | 1.6 | 1.9 | 1.6 | ||||||
| Ra-223 | 2013 | [20] | 20 mL dose vial | 3.19(4) | 229(5) | 231(5) | 234(5) | |||
| in % Activity | 1.3 | 1.6 | 1.6 | 1.6 | ||||||
| Ra-223 | 2013 | [20] | 2 mL syringe10 | 3.31(6) | 237(8) | 232(7) | 234(7) | |||
| in % Activity | 1.8 | 2.5 | 2.2 | 2.2 | ||||||
Reference [10] also gives DSs for several v-vial geometries; no uncertainties given.
The value and accompanying uncertainty reported here are the average and standard deviation of two determinations performed on solutions with different compositions.
See also [14] for volume effects and v-vial geometries.
The DSs reported here were determined by the “dialing-in” method [1], where the DS is tuned to give the correct activity. Uncertainties are estimated as equivalent to the uncertainty on the activity determined with the contemporaneously-determined KVIC.
The primary standard has a 2 % combined standard uncertainty; Zimmerman and Cessna found DS = 252 for the CRC-12 chamber in 2000 [1].
This DS was also found to be valid for a 10R Schott vial.
Reference [17] also gives a formula for volume dependence over a wider range.
See also [19] for recommendations on measuring mixed W-188/Re-188 solutions.
Reference [19] also gives correction factors for Capintec readings at manufacturer-recommended settings.
Reference [20] also gives results for 5 mL and 20 mL syringes with several total solution volumes.
In most cases, the settings reported here match those given in the previous NIST summary [1]. The very notable exception is for 18F. The settings reported here supplant the earlier ones, reflecting the revised NIST standard [7,21]. Another exception is for 117mSn, where Table 1 gives a value calculated by averaging the results of determinations performed with two different solution compositions. The determination for 153Sm has been updated based on new measurements, but differs only slightly from the earlier report.
In most cases, new DS determinations have been described in the literature. The few exceptions are addressed here:
99mTc
In November 2017, a standard NIST 5 mL flame-sealed ampoule containing 5 g of solution was obtained from the NIST Radioactivity Measurement Assurance Program (NRMAP) [22] and measured on multiple dose calibrators. The standard activity was taken from secondary standard ionization chamber (SSIC) measurements on NIST Chamber “A” (ICA) [23] using the calibration factor (“K-value”) determined in 1976 [24]. The primary standardization used coincidence counting with a pressurized proportional counter/NaI(Tl) system. The uncertainty on the standard activity (0.44 %) was the largest component in the combined standard uncertainties on the DSs.
The manufacturer-recommended DS (= 80) for the Capintec (Ramsey, NJ) CRC-15R and CRC-35R chambers resulted in activity readings biased by +1.6 % and +1.9 %, respectively [25].1The recommended DS (= 37.1) for the AtomLab-400 and AtomLab-500 chambers (manufactured by Sun Nuclear; Shirley, NY) resulted in readings biased by −1.6 % and −2.1 %, respectively [26].2
Source geometry sensitivity in dose calibrator measurements of 99mTc has been discussed previously [27] and in reference [1], DSs for the CRC-12 were given for the ampoule and 10 mL Mallinckrodt dose vial geometries. Those values are preserved in Table 1. Significantly, the new values presented for the CRC-15R and CRC-35R are consistent with the prior DS found for the CRC-12.
111In
In September 2012, four ampoules from the NRMAP were measured on multiple dose calibrators. The standard activity was taken from ICA using the K-value determined in 1977. The primary standardization used coincidence counting with a pressurized proportional counter/NaI(Tl) system. The uncertainty on the standard activity (0.55 %) was the largest component in the total combined uncertainties on the DSs.
The manufacturer-recommended DS (= 303) for the CRC-12, CRC-15R, and CRC-35R chambers [25] resulted in activity readings biased by +3.2 %, +3.4 %, and +2.4 %, respectively.
153Sm
In July and August 2013, an ampoule from the NRMAP was measured on multiple dose calibrators. The standard activity was taken from ICA using the K-value determined in 1985. The primary standardization used 4πβ liquid scintillation counting. The primary standard had a fairly large uncertainty (≈ 2 %) and a large correction (≈ 1.5 %) for the geometric drift in ICA [28] was required; a new primary standardization for this nuclide is desirable. The uncertainty on the standard activity (≈ 2 %) was the largest contributor to the combined standard uncertainties on the DSs.
For the Vinten 671 Ionization Chamber (VIC) [29] measurements, impurity corrections for 154Eu and 156Eu were based on HPGe measurements and volume-corrected IC responses communicated to us by the National Physical Laboratory (NPL; United Kingdom) [30]. The NIST- and NPL-determined calibration factors agreed to within ≈ 0.2 %. For the dose calibrators, impurity corrections could not be made since 154Eu and 156Eu responses at the 153Sm DS were not known. Over time, the effect of ingrowing impurities on DS bias was evident. To reduce bias, DS determinations relied on the earliest data acquired, where they appeared stable to < 0.1 %. The large uncertainties due to the standard activity and height correction dwarf any residual biases or underestimate of the total combined standard uncertainties.
177Lu
In 2012, NIST participated in a comparison of clinical dose calibrator measurements of 177Lu piloted by the NPL [31]. The source was delivered in a 10 mL Schott-type dose vial and NIST performed measurements on multiple dose calibrators in both the dose vial and 5 mL ampoule geometries. The standard activity was taken from ICA using the K-value determined in 2009 by anticoincidence counting [32]. The uncertainty on the standard activity (0.37 %) was the largest component in the combined standard uncertainties on the determined DSs in all cases except for the CRC-35R calibrator where the uncertainty on the reading (estimated in this case from the range of minimum and maximum readings observed over a few seconds) was larger.
In general, the results were in good agreement with the manufacturer’s recommendations, with previous NIST-determined DSs [33], and with the NPL.
4. Discussion
Ionization chamber calibrations are an integral part of primary activity standardization efforts at NIST. For the short-lived radionuclides of interest in nuclear medicine especially, calibrated ionization chambers establish an enduring means of assaying new samples with traceability to the primary activity standard. As SSICs like the VIC establish indirect links with other metrology institutes around the world, dose calibrators link NIST standards to clinical and industrial measurements—with some caveats.
NIST always recommends that DS determinations are considered valid only for the specific calibrators and source characteristics described. Users should verify the results on their own systems. This is especially important when environmental conditions (e.g., backgrounds, local shielding) may significantly differ from those at NIST. Using the NIST methods rather than the NIST results also ensures against any larger than expected calibrator-to-calibrator variation.
In addition to suggesting that users verify all calibrations on their own instruments, NIST stresses the importance of regular quality control measurements (see user’s manuals), especially measurements of a long-lived source to monitor response constancy. Long-term trending in constancy measurements or abrupt deviations from the expected response require investigation. At best, a change in response, perhaps owing to a change in the local environment (e.g., addition of shielding) indicates that new calibration verifications are required. A change in response that cannot be explained by a change in the local environment may indicate instrument failure (e.g., damage to the electronics or a loss of chamber pressure), requiring that the calibrator be removed from service.
Another important consideration involves the practice of “sharing” calibrations between instruments. This practice is common when a user maintains multiple calibrators of the same type: one calibration verification on a specific instrument serves to establish a site-specific DS which is then shared among all instruments. This practice can be dangerous if a DS is shared between instruments from the same manufacturer but from different generations. For example, the Capintec CRC-12, CRC-15R, and CRC-35R calibrators maintained at NIST are designed to give similar responses for a given DS; however, the CRC-25PET is representative of a different series of instruments designed to handle higher activities for positron emission tomography (PET) applications, and requires a distinct DS. In fact, the CRC-25PET manual [34] includes DSs for only eight nuclides, half of which were determined with NIST-traceable sources and the other half by inter-comparison with a CRC-15R instrument. The validity of a DS should always be tested for each individual calibrator.
Occasionally, updated standards or nuclear data evaluations (see, for example, [9, 19]) result in a change in the DS recommended by instrument manufacturers or pharmaceutical companies. DS changes should only be made in response to specific instructions from pharmaceutical manufacturers, and any discrepancy between a local assay and the activity reported by the manufacturer should be investigated.
Finally, in recent years the use of long-lived calibration surrogates for some radionuclides has become a viable option. The use of 68Ge sources as surrogates for short-lived positron emission tomography (PET) radionuclides is now firmly established [9, 35]. However, some approaches, such as the use of 133Ba as a surrogate for 131I can lead to errors [15].
5. Conclusions
NIST has carried out calibrations for 22 radionuclides on different reentrant ionization chambers (“dose calibrators”) and often in multiple geometries. In most cases, calibration factors or “dial settings” (DS) are determined using the calibration curve method. This paper presents a collection of NIST-determined DSs for medically relevant radionuclides, addressing uncertainties and pointing to relevant references. The DSs are considered applicable for the NIST calibrators and the specifically described geometries; users are encouraged to check their validity on their own systems.
In general, the NIST-determined DSs are consistent with current manufacturer recommendations. This is in contrast to the situation reported in 2000 [1], when most of the considered manufacturer-provided DSs overestimated activities, in some cases by more than 20 %. In general, improved accuracy can be interpreted as reflecting the positive impact of an emphasis on traceability to common measurement standards.
Acknowledgements
We are grateful to W. Regits and K. Neal (NIST) for providing crucial data and to B. Zimmerman and B. Coursey (NIST) for input and critical readings of the manuscript.
Footnotes
Disclaimer: Certain commercial equipment, instruments, or materials are identified in this paper to foster understanding. Such identification does not imply recommendation by the National Institute of Standards and Technology, nor does it imply that the materials or equipment identified are necessarily the best available for the purpose.
Certain commercial equipment, instruments, or materials are identified in this paper to foster understanding. Such identification does not imply recommendation by the National Institute of Standards and Technology, nor does it imply that the materials or equipment identified are necessarily the best available for the purpose.
While each calibrator has its own manual, the DS tables are consistent across the Capintec R series and across the Biodex AtomLab series. We cite just one manual from each manufacturer
References
- [1].Zimmerman BE, Cessna JT Experimental determinations of commercial ‘dose calibrator’ settings for nuclides used in nuclear medicine. Appl. Radiat. Isotop. 2000; 52, 615–619. [DOI] [PubMed] [Google Scholar]
- [2].Bobin C Primary standardization of activity using the coincidence method based on analogue instrumentation. Metrologia 2007; 44, S27–S31. [Google Scholar]
- [3].Fitzgerald R, Bailat C, Bobin C, Keightley JD Uncertainties in 4πβ-γ coincidence counting. Metrologia 2015; 52, S86–S96. [Google Scholar]
- [4].Broda R, Cassette P, Kossert K Radionuclide metrology using liquid scintillation counting. Metrologia 2007; 44, S36–S52. [Google Scholar]
- [5].Kossert K, Broda R, Cassette P, Ratel G, Zimmerman B Uncertainty determination for activity measurements by means of the TDCR method and the CIEMAT/NIST efficiency tracing technique. Metrologia 2015; 52, S172–S190. [Google Scholar]
- [6].Taylor BN, Kuyatt CE NIST Technical Note 1297: Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results. Washington, DC: US Government Printing Office, 1994. [Google Scholar]
- [7].Fitzgerald R, Zimmerman BE, Bergeron DE, Cessna JT, Pibida L, Moreira DS A new NIST primary standardization of 18F. Appl. Radiat. Isotop. 2014; 85, 77–84. [DOI] [PubMed] [Google Scholar]
- [8].Bergeron DE, Cessna JT, Fitzgerald R, Pibida L, Zimmerman BE Standardization of 64Cu activity. Appl. Radiat. Isotop. 2017; submitted. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [9].Zimmerman BE, Bergeron DE, Cessna JT Impact of recent change in the National Institute of Standards and Technology Standard for 18F on the relative response of 68Ge-based mock syringe dose calibrator standards. J. Nucl. Med. 2015; 56, 1453–1457. [DOI] [PubMed] [Google Scholar]
- [10].Coursey BM, Calhoun JM, Cessna JT Radioassays of Yttrium-90 used in nuclear medicine. Nucl. Med. Biol 1993; 20, 693–699. [DOI] [PubMed] [Google Scholar]
- [11].Zimmerman BE, Cessna JT, Millican MA Experimental determination of calibration settings for plastic syringes containing solutions of 90Y using commercial radionuclide calibrators. Appl. Radiat. Isotop. 2004; 60, 511–517. [DOI] [PubMed] [Google Scholar]
- [12].Zimmerman BE, Cessna JT, Schima FJ The standardization of the potential bone palliation radiopharmaceutical 117mSn(+4)DTPA. Appl. Radiat. Isotop. 1998; 49, 317–328. [DOI] [PubMed] [Google Scholar]
- [13].Bergeron DE, Cessna JT, Golas DB, Young RK, Zimmerman BE Dose calibrator manufacturer-dependent bias in assays of 123I. Appl. Radiat. Isotop. 2014; 90, 79–83. [DOI] [PubMed] [Google Scholar]
- [14].Zimmerman BE, Cessna JT, Dorton JA Experimental investigation of dose calibrator response for 125I brachytherapy solutions contained in 5 mL plastic syringes and 2 mL conical glass v-vials as a function of filling mass. Med. Phys. 2002; 29, 1547–1555. [DOI] [PubMed] [Google Scholar]
- [15].Zimmerman BE, Bergeron DE (Mis)use of 133Ba as a calibration surrogate for 131I in clinical activity calibrators. Appl. Radiat. Isotop. 2016; 109, 250–253. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [16].Cessna JT, Zimmerman BE, Unterweger MP, Golas DB Establishment of transfer standard for holmium-166-DOTMP. Appl. Radiat. Isotop. 2004; 60, 505–510. [DOI] [PubMed] [Google Scholar]
- [17].Zimmerman BE, Pipes DW Experimental determination of dose calibrator settings and study of associated volume dependence in v-vials for rhenium-186 perrhenate solution sources. J. Nucl. Med. Technol 2000; 28, 264–270. [PubMed] [Google Scholar]
- [18].Zimmerman BE, Cessna JT, Unterweger MP, Li AN, Whiting JS, Knapp FF A new experimental determination of the dose calibrator setting for 188Re. J. Nucl. Med. 1999; 40, 1508–1516. [PubMed] [Google Scholar]
- [19].Zimmerman BE, Cessna JT, Unterweger MP The standardization of 188W/188Re by 4πβ liquid scintillation spectrometry with the CIEMAT/NIST 3H-standard efficiency tracing method. Appl. Radiat. Isotop. 2002; 56, 315–320. [DOI] [PubMed] [Google Scholar]
- [20].Bergeron DE, Cessna JT, Zimmerman BE Secondary standards for 223Ra revised. Appl. Radiat. Isotop. 2015; 101, 10–14. [DOI] [PubMed] [Google Scholar]
- [21].Bergeron DE, Cessna JT, Coursey BM, Fitzgerald R, Zimmerman BE A review of NIST primary activity standards for 18F: 1982 to 2013. J. Res. Natl. Inst. Stand. Technol. 2004; 119, 371–396. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [22].Cessna JT, Golas DB The NIST radioactivity measurement assurance program for the radiopharmaceutical industry. Appl. Radiat. Isotop. 2012; 70, 2227–2231. [DOI] [PubMed] [Google Scholar]
- [23].Calhoun JM NBS Special Publication 250–10: Radioactivity Calibrations with the “4π” Gamma Ionization Chamber and Other Radioactivity Calibration Capabilities. Washington, DC: US Government Printing Office, 1987. [Google Scholar]
- [24].Ayres RL, Hirshfeld AT Radioactivity Standardization of 99mTc and 99Mo. Int. J. App. Radiat. Isotop. 1982; 33, 835–841. [DOI] [PubMed] [Google Scholar]
- [25].Capintec Inc. Radioisotope calibrator owner’s manual, Rev. E. Ramsey, NJ, 1986. [Google Scholar]
- [26].Biodex Medical Systems, Inc. AtomLab 500 Dose Calibrator Operation and Service Manual, Rev B. Shirley, NY, 2012. [Google Scholar]
- [27].Calhoun JM, Golas DB, Harris SG Effects of varying geometry on dose calibrator response: Cobalt-57 and Technetium-99m. J. Nucl. Med. 1987; 28, 1478–1483. [PubMed] [Google Scholar]
- [28].Fitzgerald R NIST Ionization Chamber “A” Sample-Height Corrections. J. Res. Natl. Inst. Stand. Technol. 2012; 117, 80–95. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [29].Woods MJ, Callow WJ, Christmas P The NPL Radionuclide Calibrator – Type 271. Int. J. Nucl. Med. Biol. 1983; 10, 127–132. [Google Scholar]
- [30].Keightley J, personal communication.
- [31].Fenwick A, Baker M, Ferreira K, Keightley J Comparison of 90Y and 177Lu measurement capability in UK and European hospitals. Appl. Radiat. Isotop. 2014; 87, 10–13. [DOI] [PubMed] [Google Scholar]
- [32].Zimmerman BE, Altzitzoglou T, Antohe A, Arinc A, Bakhshandeiar E, Bergeron DE, et al. Results of an international comparison for the activity measurement of 177Lu. Appl. Radiat. Isotop. 2012; 70, 1825–1830. [DOI] [PubMed] [Google Scholar]
- [33].Zimmerman BE, Unterweger MP, Brodack JW The standardization of 177Lu by 4πβ liquid scintillation spectrometry with 3H-standard efficiency tracing. Appl. Radiat. Isotop. 2001; 54, 623–631. [DOI] [PubMed] [Google Scholar]
- [34].Capintec Inc. CRC-25PET Owner’s Manual, Rev. B. Ramsey, NJ, 2007. [Google Scholar]
- [35].Zimmerman BE, Cessna JT Development of a traceable calibration methodology for solid 68Ge/68Ga sources used as a calibration surrogate for 18F in radionuclide activity calibrators. J. Nucl. Med. 2010; 51, 448–453. [DOI] [PubMed] [Google Scholar]
