Abstract
Inhaled medications are the cornerstone of treatment and management of asthma and COPD. However, inhaler device errors are common among patients and have been linked with reduced symptom control, an increased risk of exacerbations, and increased healthcare utilisation. These observations have prompted GINA (Global INitiative for Asthma) and GOLD (Global initiative for chronic Obstructive Lung Disease) to recommend regular assessment of inhaler technique in a bid to improve therapeutic outcomes. To better define the relationship between device errors and health outcomes (clinical outcomes, quality of life, and healthcare utilisation) in asthma and COPD, we conducted a systematic review of the literature, with a particular focus on the methods used to assess the relationship between device errors and outcomes. Sixteen studies were identified (12 in patients with asthma, one in patients with COPD, and three in both asthma and COPD) with varying study designs, endpoints, and patient populations. Most of the studies reported that inhalation errors were associated with worse disease outcomes in patients with asthma or COPD. Patients who had a reduction in errors over time had improved outcomes. These findings suggest that time invested by healthcare professionals is vital to improving inhalation technique in asthma and COPD patients to improve health outcomes.
Introduction
Asthma and chronic obstructive pulmonary disease (COPD) are common chronic respiratory diseases that impart an economic and social burden.1–4 Rates of asthma vary between countries, affecting 1–18% of the population,5 and >10% of adults aged 40 years and over in the general population are estimated to suffer from COPD.6
Inhaled medications are the cornerstone of the treatment and management of asthma and COPD.4,5 There are many devices for the delivery of inhaled medications, including pressurised metered-dose inhalers (pMDIs), dry-powder inhalers (DPIs), soft-mist inhalers, breath-actuated MDIs, and nebulisers4,7; pMDIs and DPIs are the most commonly used.8,9 While the wide array of treatments available may be seen as positive, the large number of available devices can result in a certain amount of complexity for prescribers when teaching patients their correct use.
Effective use of inhalers requires patients to follow the prescribed inhalation technique.7,8 Errors in device use can result in suboptimal drug delivery, reducing the effective medication dose and thus compromising treatment effectiveness.9–13 In this systematic literature review, an error is defined as critical if it has an impact on the effectiveness of the drug. It has been estimated that up to 92% of patients make at least one critical error when using an inhaler, with a higher error rate reported in patients with COPD compared with those with asthma.9 The type and frequency of errors vary among devices depending on their characteristics,11,14–19 although many common errors are universal, such as failing to exhale before each inhalation and failing to hold the breath following inhalation.9 Overall, physicians tend to overestimate good inhalation technique and underestimate errors made by patients.20
It is important, therefore, that effective strategies are in place to educate patients on correct inhaler use.21 Key guideline groups such as GINA (Global INitiative for Asthma) and GOLD (Global initiative for chronic Obstructive Lung Disease) have made recommendations to monitor inhaler technique in their strategy documents4,5; for example, the most recent version of GOLD recommends the assessment and regular evaluation of inhaler technique, with the aim of improving therapeutic outcomes in patients with COPD.4 Both GINA and GOLD also recommend that a patient’s ability to use an inhaler device correctly should be integral to decision-making when choosing between available controller medications.4,5
Although there is evidence that errors in device use can affect outcomes,19 there is no comprehensive overview of the relationship between them. Therefore, this systematic review of the literature was conducted to examine the relationship between device errors and health outcomes (clinical, quality of life [QoL], and economic) in patients with asthma and COPD.
Literature search
Systematic searches of PubMed, Google, and Google Scholar were conducted in April 2017 (updated December 2017) to identify studies assessing device errors, incorrect handling, or improper technique in patients with asthma or COPD. Additional studies were identified from a recently published systematic literature review of inhaler device errors9 and by author recommendation.
PubMed was searched using the search string: (COPD OR chronic obstructive pulmonary disease OR Asthma) AND (error OR mishandling OR erroneous OR incorrect use OR incorrect technique OR improper use OR improper technique OR inadequate technique OR inadequate use OR insufficient use OR insufficient technique OR critical error OR significant error).
Google and Google Scholar were searched using the following key terms, in different combinations: (asthma or COPD device error; device error in asthma or COPD; improper technique in asthma or COPD; inadequate use in asthma or COPD; misuse or mishandling in asthma or COPD; incorrect use in asthma or COPD).
Publications that did not report on clinical trials or clinical studies were excluded from the initial results. Publication titles and abstracts were then screened, and articles that did not link device errors with outcomes of asthma or COPD were excluded. Data were extracted systematically by an independent reviewer using a predefined extraction template (Supplementary Table 1). Odds ratios (ORs) were the principal measures of the relationships between errors and outcomes.
Identified studies
Of the screened publications, a total of 19 relevant articles were identified (Fig. 1). Seven of these were sourced from the database searches (PubMed [n = 6], Google/Google Scholar [n = 1]), and a further seven were identified from a recent systematic literature review.9 The remaining five articles were identified during data extraction (n = 2) and data analysis (n = 1), and as a recommendation by one of the investigators (MD) (n = 2).
Of the 19 relevant articles, three were excluded (duplicates removed [n = 2], judged by investigators to not report relevant data [n = 1]), leaving a total of 16 unique studies (Fig. 1). These studies reported associations between device errors and clinical outcomes, and were included for discussion in the review. Details of the included studies are shown in Table 1. In general, the studies examined patients with asthma and COPD separately, with most (n = 12) studying patients with asthma only (Table 1). Overall, disease severity was not specified in the studies. Most studies included 500 patients or more and focused on adults, with only a few including adolescent patients or patients of all ages (Table 1). In one study that included children below 12 years old, 20% were < 16 years old.32 In most studies, patients were recruited from outpatient clinics, mostly in European countries (Table 1). MDIs (alone or with a spacer), breath-actuated MDIs (Autohaler and Easi-Breathe inhaler [trademarks of Teva Pharmaceuticals Ltd]) and DPIs (Diskus [a trademark owned by or licensed to the GSK group of companies], Turbuhaler [trademark of AstraZeneca Ltd], HandiHaler [trademark of Boehringer Ingelheim]) were the most commonly studied inhalers (Table 1).
Table 1.
References | Study type | Country | Setting | Patient age | Sample size, n | Inhaler(s) studied | Outcomes studied | Device error definition |
---|---|---|---|---|---|---|---|---|
Asthma only | ||||||||
Giraud & Roche 200222 | Clinical cross-sectional | France | OP | > 15 years | 4078 | pMDI | Clinical, economic | Misusers vs. good users ( ≥ 1 vs. no errors), and poor coordinators vs. good coordinators ( ≥ 1 vs. no errors between actuation and inhalation) |
Molimard & Le Gros 200823 | Clinical cross-sectional | France | OP | Adults | 4362 | pMDIs, DPIs | Clinical | Patients making ≥ 1 critical errors vs. patients using inhaler correctly |
Giraud, Allaert & Magnan 201124 | Clinical cross-sectional | France | OP | Adults | 6512 | Breath-actuated MDI | Clinical | Patients with suboptimal vs. optimal technique (optimal = correctly following 7-step checklist and avoiding five possible errors) |
Natsir et al. 201325 | Clinical cross-sectional | Indonesia | OP | Adults | 60 | NR | Clinical, economic | Patients demonstrating improper inhaler use, evaluated using a checklist based on Global Initiative for Asthma |
Al-Jahdali et al. 201326 | Clinical cross-sectional | Saudi Arabia | ED | Adults | 450 | MDIs, DPIs | Clinical, economic | Proper vs. improper inhaler use (proper use = fulfilled all required steps on a device checklist over two trials of using their inhaler) |
Baddar, Jayakrishnan & Al-Rawas 201427 | Clinical cross-sectional | Oman | OP | 12−72 years | 218 | NR | Clinical | Good inhaler technique (all essential steps performed accurately) vs. poor inhaler technique (any required steps missed/performed inaccurately) |
de Tarso Roth Dalcin et al. 201428 | Clinical cross-sectional | Brazil | OP | Adults | 268 | MDIs, DPIs | Clinical | Correct ( < 2 errors) vs. incorrect inhaler technique ( ≥ 2 errors) |
Giraud, Allaert & Roche 201129 | Prospective clinical (1 month follow-up) | France | Pharmacy | Adults | 727 | pMDIs, breath-actuated MDIs | Clinical | Optimal use (no errors) vs. non-optimal use ( ≥ 1 critical or non-critical error) |
Yildiz et al. 201430 | Prospective clinical, longitudinal ( ≥ 6 months follow-up) | Turkey | OP | Adults | 572 | pMDIs, DPIs | Clinical | Patients making 0–1 basic errors vs. patients making > 1 basic error |
Harnett et al. 201431 | Prospective clinical, longitudinal (3−4 months follow-up) | Ireland | OP | ≥ 16 years | 40 | pMDIs, DPIs, soft-mist inhaler | Clinical, QoL | Optimal users (no errors) vs. misusers ( ≥ 1 of 10 steps performed incorrectly) |
Levy et al. 201332 | Retrospective, database (IMPACT), cross-sectional | UK | − | All ages | 3981 | MDIs, DPIs | Clinical | Patients with correct vs. incorrect technique (incorrect technique = failure of one or more of: inspiratory flow between 10–50 L/min; correct flow for ≥ 1.5 s post-actuation; post-inspiration breath hold for ≥ 5 s) |
Price et al. 201719 | Retrospective, database (iHARP), cross-sectional | Australia, Europe | − | ≥ 16 years | 3660 | pMDIs and DPIs | Clinical | Frequency of specific errors and device-specific errors |
COPD only | ||||||||
Molimard et al. 201733 | Clinical cross-sectional | France | OP | > 40 years | 2935 | pMDIs, Respimat, DPIs | Clinical, economic | Patients with absence of error vs. presence of critical error |
Both asthma and COPD | ||||||||
Melani et al. 2011,8 | Clinical cross-sectional | Italy | OP | > 14 years | 1664 | MDIs, DPIs | Clinical, economic | Inhaler misuse (patients with presence of error or critical error) |
Maricoto et al. 201534 | Clinical cross-sectional | Portugal | OP | > 12 years | 62 | MDIs, DPIs | Clinical | Patients with ≥ 1 error, number of errors committed (0–4) |
Roggeri, Micheletto & Roggeri 201635 | Clinical cross-sectional | Italy | OP | > 14 years | 400 | NR | Economic | See above (Melani et al. 2011)8 |
COPD chronic obstructive pulmonary disease, DPI dry-powder inhaler, ED emergency department, iHARP Improving Health of At-Risk Rural Patients, IMPACT InforMing the PAthway of COPD Treatment, MDI metered-dose inhaler, NR not reported, OP outpatient, pMDI pressurised metered-dose inhaler, QoL quality of life
Examination of inhaler errors
Six studies examined critical errors (i.e., the authors specifically stated that ‘critical errors’ were evaluated).8,19,23,29,33,35 In the remaining 10 studies, the term ‘critical errors’ was not used;22,24–28,30–32,34 however, there was overlap in the actual errors across all studies. Only one study, CritiKal,19 identified associations between specific inhaler technique errors and reduced disease control, confirming those errors as critical. Some studies defined errors separately for each device used by patients.8,19,23,26,28,29 Errors were generally assessed using checklists for assessment of inhalation technique.8,19,22–24,26,28,29,31,32,34,35 Overall, a patient’s technique was judged to be poor (for both critical and non-critical errors) if any one of the required steps was missed or performed inaccurately.22,23,26–28 A number of studies in asthma or COPD identified inhaler errors that were established as critical errors in the CritiKal study; these are shown in Table 2.
Table 2.
CritiKal study error | Number of studies that evaluated the error | ||
---|---|---|---|
Asthma studies | COPD studies | Both asthma and COPD | |
Did not remove cap/slide cover open | 422,24,26,29 | 0 | 18 |
Insufficient inspiratory effort | 519,22,26,28,29 | 133 | 134 |
Did not have head tilted such that chin is slightly upward | 219,26 | 0 | 0 |
Did not breathe out to empty lungs before inhalation | 519,22,24,26,30 | 133 | 18 |
No breath hold (or holds breath for < 3 s) | 719,22,24,26,28–30 | 133 | 28,34 |
Did not seal lips around mouthpiece | 519,24,26,28,29 | 0 | 18 |
Incorrect second dose preparation, timing, or inhalation | 119 | 0 | 0 |
Exhaled into device before inhalation | 219,30 | 133 | 0 |
Dose compromised after preparation because of shaking or tipping (DPIs only) | 219,28 | 0 | 0 |
Actuation did not correspond with inhalation, actuation before inhalation (MDI only) | 519,22,24,26,29 | 133 | 18 |
COPD chronic obstructive pulmonary disease, DPI dry-powder inhaler, MDI metered-dose inhaler
Associations between device errors and outcomes
While most studies demonstrated a cross-sectional association between errors and outcomes using several approaches, they did not distinguish between critical and non-critical errors.22,24–28,30–32,34 Most commonly, errors were defined dichotomously (i.e., any errors present vs. no errors present), with clinical outcomes compared in patients with and without inhalation errors.8,19,22–28,32,33
A less common approach was based on the number of errors (i.e., association between the number of errors and outcomes22,34). A few studies focused on specific devices and/or specific errors, and empirically determined the errors that were associated with adverse outcomes.19,22,32
Among the longitudinal analyses, two approaches were used. The first approach involved demonstrating an improvement in inhalation technique alongside an improvement in outcomes, without statistically testing the association,30 whereas the second approach demonstrated statistical associations between improved inhalation technique and improved outcomes over a period of 1 month or more.29,31 Prospective studies generally involved participant training, and before/after comparisons of inhalation technique and outcomes.29–31 However, two longitudinal studies examined the evolution in error rate over extended periods of time (3−4 and ≥ 6 months, respectively30,31).
Effect of device errors on outcome measures
Most of the studies assessed the effects of device errors on the clinical outcome of disease control in asthma and COPD (Table 1). Measures of asthma control were the Asthma Control Test (ACT; n = 6),8,26,27,30,31,34 the Asthma Control Questionnaire (ACQ; n = 3),24,29,31 study-specific questionnaires (n = 3),19,22,23 GINA category (n = 2),28,32 Control of Allergic Rhinitis and Asthma Test (n = 1),34 and any unscheduled medical intervention due to respiratory disease (n = 2).8,19 COPD control was assessed using the modified Medical Research Council (mMRC) Dyspnoea Scale (n = 2)8,34 and the number of COPD exacerbations (n = 1).33 Only one study evaluated the clinical outcome of QoL in patients with asthma, reporting the association between device errors and QoL using the Asthma Quality of Life Questionnaire (AQLQ)31 (Table 1). Six studies assessed healthcare resource utilisation (HRU) as the clinical outcome in asthma and/or COPD (Table 1). Economic outcome measures were emergency room (ER) visits (n = 6),8,22,25,26,33,35 hospitalisations (n = 2),8,35 and prescriptions (antimicrobial agents, steroids, n = 2)8,35; one study examined costs associated with increased HRU due to critical errors in both asthma and COPD patients.35
Key findings of studies in asthma and COPD are shown in Supplementary Tables 2 and 3. Data supporting an association between inhalation errors and worse disease outcomes were reported in almost all studies, and patients in the longitudinal studies who had a reduction in errors over time had improved outcomes, irrespective of endpoint. Despite differences in study designs across the evaluated studies, the magnitude of effect appeared to be similar across the different endpoints. For example, reported ORs of the relationship between device errors and worse outcomes typically ranged from 1.46 to 1.73 for ACT score,8 and 1.30 to 1.56 for study-specific questionnaires assessing asthma control.19 ORs linking critical inhaler errors to mMRC Dyspnoea score8 or severe COPD exacerbations (requiring hospitalisation/ER visit)33 were 1.10 and 1.86, respectively. Two studies failed to report an association between errors and outcomes (one in asthma patients,31 and another in the COPD cohort of a mixed asthma/COPD study34); however, both had small sample sizes (40 and 27 [COPD] patients, respectively), and were possibly underpowered.31,34
Inhaler technique, asthma outcomes and treatment adherence have all been linked to one another34; however, only five of the evaluated studies measured treatment adherence, all of which measured patient-reported adherence.19,23,27,29,33 Giraud et al.29 found that training patients in inhaler technique led to statistically significant improvements in ACQ score, patient-reported adherence, and the number of patients demonstrating optimal technique (all p < 0.001).
Discussion
The majority of studies identified from the literature reported an association between inhalation errors and worse disease outcomes in patients with asthma or COPD, and the magnitude of effect was similar across the different endpoints studied. Most of the studies were cross-sectional and were conducted in pulmonology clinics/outpatient departments. A few prospective, longitudinal studies, and some database analyses, were also identified. There are advantages and disadvantages to each study design that should be considered when interpreting the literature. For example, it is difficult to infer causality using a cross-sectional approach, due to lack of temporality. This approach can also be subject to recall bias (especially for HRU studies). In general, outcome data (e.g., number of hospitalisations, ER visits etc.) in the identified studies relied upon retrospective recollection by patients, and were often limited to events in the recent past (1−3 months). Furthermore, while database analyses provide readily available data, few databases capture data on device errors, and as a result, associated information on inhalation technique. Prospective studies can overcome the issue of temporality by providing more reliable data on outcomes, especially those related to HRU. However, they are operationally more challenging to conduct, and do not provide results as quickly as cross-sectional or database analyses. Longitudinal studies, on the other hand, can evaluate the long-term impact of errors and causality, and are able to address a variety of research questions. For example, longitudinal studies can be used to measure the effects of interventions that target inhaler errors.
Despite guidance by GINA, GOLD, and national asthma and COPD guidelines, inhaler errors made by patients are still prevalent, and the type and incidence of errors has not changed considerably over the past 40 years.36 As little has improved over the last decade,20 further efforts to support patients in using their inhaler(s) or an alternative approach to delivering inhaled medications is needed to resolve this ongoing issue.13 For example, investing in training time for healthcare professionals could help to enhance patient support and improve overall inhalation technique. Efforts have also been made to develop simplified regimens with fewer inhalers,13 and to provide inhalers that are more intuitive to use.37 Both strategies have the potential to reduce the number of errors made by patients, subsequently enhancing the effectiveness of the medication and associated patient outcomes.
Alternative approaches to help patients retain the skills needed to use their device correctly should also be considered. The recent development of electronic devices that attach to existing inhalers, or the development of integrated inhalers that measure inhalation characteristics, can provide direct feedback to patients on their inhaler technique. This has the potential to reduce errors, as previous studies have shown that regular feedback can improve the retention of correct inhaler technique.13
As treatment adherence has an impact on asthma outcomes,34 and an association has been demonstrated between adherence and inhaler technique (poor adherence was linked with poor technique, and good adherence with good technique),29 it is important to consider the potential for treatment adherence to amplify associations between device errors and asthma outcomes.
There is a lack of consistency in the literature regarding which errors are defined as critical. Errors should be assessed using a single checklist that is standardised as much as possible across the different devices. One unified definition for ‘critical errors’ should be used; for example, the CritiKal study catalogued errors that were related to poor asthma control.19 Studies should report errors individually for the various devices in order to identify persistent problems and better target device training. A recent systematic review and meta-analysis estimating error rates in MDIs and DPIs failed to show a difference between various inhalers due to a relatively limited body of evidence. However, it did confirm that overall and critical error rates were unacceptably high across all devices.9 Most of the errors in the CritiKal study were generic to the type of inhaler (DPI or MDI) rather than device specific.19
A limitation of this systematic review was that a meta-analysis of the literature could not be performed due to heterogeneity in study designs, outcomes, and patient populations studied, although these may become viable in the future as study designs become more uniform.
Published data from studies in asthma and COPD demonstrate the presence of an association between inhalation errors and outcomes, with an apparent relationship between a reduction in errors and improvement in outcomes, irrespective of endpoint. The magnitude of effect is similar across the different endpoints. Therefore, investment of time by healthcare professionals is vital to improving inhalation technique in asthma and COPD patients in order to improve health outcomes. The methodology used to study the relationship between inhaler errors and outcomes varies widely. Therefore, there is a need for greater standardisation of the methods used to assess inhaler errors across the spectrum of devices and outcome measures. Future research should also focus on identifying ways to improve device handling.
Electronic supplementary material
Acknowledgements
This review was funded by GSK. The authors thank Saifuddin Kharawala and Gavneet Kaur from Bridge Medical, who performed the original literature searches and data extraction. Editorial support (in the form of writing assistance, collating author comments, assembling tables, grammatical editing, fact checking, and referencing) was provided by David Mayes, MChem, of Gardiner-Caldwell Communications (Macclesfield, UK) and was funded by GSK. Autohaler and Easi-Breathe are trademarks of Teva Pharmaceutical Industries Ltd. Turbuhaler is a trademark of AstraZeneca. HandiHaler and Respimat are trademarks of Boehringer Ingelheim. Diskus is a trademark owned by or licensed to the GSK group of companies.
Author contributions
J.K., M.T., J.v.d.P., H.C., M.M., and M.G. contributed to the data analysis or interpretation. S.L., R.S., M.D., and L.Y. contributed to the conception or design of the review and the data analysis or interpretation. All authors drafted the work or revised it critically for important intellectual content, approved the final version, and agreed to be accountable for all aspects of the work.
Data availability
All publications identified in the systematic literature review are accessible via the public databases (PubMed, Google/Google Scholar), or via the articles’ respective journal websites. All of the included data were sourced from these publicly available articles.
Competing interests
Janwillem W.H. Kocks declares grants and personal fees from AstraZeneca, Boehringer Ingelheim, GSK and Novartis, and grants from Chiesi, Mundipharma and Teva. Henry Chrystyn has received sponsorship to carry out studies, Board Membership, consultant agreements and honoraria for presentation, from Almirall, AstraZeneca, Boehringer Ingelheim, Chiesi, GSK, Innovata Biomed, Meda, Mundipharma, Napp Pharmaceuticals, NorPharma, Novartis, Orion, Sanofi, Teva, Trudell Medical International, UCB, and Zentiva. He has also received research sponsorship from grant-awarding bodies (EPSRC and MRC). He is the owner of Inhalation Consultancy Ltd. Job van der Palen has received grant and travel funding from GSK. Mike Thomas has received personal fees from GSK, Novartis, Aerocrine, Boehringer Ingelheim, Teva, Pfizer, Roche, and Sandoz. He has also received non-financial support from GSK, Novartis, AstraZeneca, and Boehringer Ingelheim. Louisa Yates, Sarah Landis, Maurice Driessen, Raj Sharma, and Mugdha Gokhale are employed by, and hold stock in, GSK. Mathieu Molimard declares Board Membership and Consultancy with Almirall, AstraZeneca, BMS, GSK, Lundbeck, Mundipharma, Novartis Pharma, Otsuka, Pfizer, Stallergen, and Zentiva. He received payment for travel/accommodation/meeting expenses from Novartis Pharma.
Footnotes
Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Electronic supplementary material
Supplementary information accompanies the paper on the npj Primary Care Respiratory Medicine website (10.1038/s41533-018-0110-x).
References
- 1.Lozano R, et al. Global and regional mortality from 235 causes of death for 20 age groups in 1990 and 2010: a systematic analysis for the Global Burden of Disease Study 2010. Lancet. 2012;380:2095–2128. doi: 10.1016/S0140-6736(12)61728-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Ospina MB, et al. Prevalence of asthma and chronic obstructive pulmonary disease in Aboriginal and non-Aboriginal populations: a systematic review and meta-analysis of epidemiological studies. Can. Respir. J. 2012;19:355–360. doi: 10.1155/2012/825107. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.GBD. 2015 Mortality and Causes of Death Collaborators. Global, regional, and national life expectancy, all-cause mortality, and cause-specific mortality for 249 causes of death, 1980-2015: a systematic analysis for the Global Burden of Disease Study 2015. Lancet. 2016;388:1459–1544. doi: 10.1016/S0140-6736(16)31012-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.GOLD (2018). Global Strategy for the Diagnosis, Management, and Prevention of Chronic Obstructive Pulmonary Disease. http://goldcopd.org/wp-content/uploads/2017/11/GOLD-2018-v6.0-FINAL-revised-20-Nov_WMS.pdf. Accessed 13 December 2017.
- 5.Global Initiative for Asthma (GINA). 2017 GINA Report, Global Strategy for Asthma Management and Prevention. http://ginasthma.org/2017-gina-report-global-strategy-for-asthma-management-and-prevention/. Accessed 13 December 2017.
- 6.Chapman KR, et al. Epidemiology and costs of chronic obstructive pulmonary disease. Eur. Respir. J. 2006;27:188–207. doi: 10.1183/09031936.06.00024505. [DOI] [PubMed] [Google Scholar]
- 7.Laube BL, et al. What the pulmonary specialist should know about the new inhalation therapies. Eur. Respir. J. 2011;37:1308–1331. doi: 10.1183/09031936.00166410. [DOI] [PubMed] [Google Scholar]
- 8.Melani AS, et al. Gruppo Educazionale Associazione Italiana Pneumologi Ospedalieri. Inhaler mishandling remains common in real life and is associated with reduced disease control. Respir. Med. 2011;105:930–938. doi: 10.1016/j.rmed.2011.01.005. [DOI] [PubMed] [Google Scholar]
- 9.Chrystyn H, et al. Device errors in asthma and COPD: systematic literature review and meta-analysis. NPJ Prim. Care Respir. Med. 2017;27:22. doi: 10.1038/s41533-017-0016-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Newman SP, Weisz AW, Talaee N, Clarke SW. Improvement of drug delivery with a breath actuated pressurised aerosol for patients with poor inhaler technique. Thorax. 1991;46:712–716. doi: 10.1136/thx.46.10.712. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.van der Palen J, Klein JJ, Kerkhoff AH, van Herwaarden CL. Evaluation of the effectiveness of four different inhalers in patients with chronic obstructive pulmonary disease. Thorax. 1995;50:1183–1187. doi: 10.1136/thx.50.11.1183. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Osterberg L, Blaschke T. Adherence to medication. N. Engl. J. Med. 2005;353:487–497. doi: 10.1056/NEJMra050100. [DOI] [PubMed] [Google Scholar]
- 13.van Boven JFM, et al. Urging Europe to put non-adherence to inhaled respiratory medication higher on the policy agenda: a report from the First European Congress on Adherence to Therapy. Eur. Respir. J. 2017;49:1700076. doi: 10.1183/13993003.00076-2017. [DOI] [PubMed] [Google Scholar]
- 14.van der Palen J, Eijsvogel MM, Kuipers BF, Schipper M, Vermue NA. Comparison of the Diskus inhaler and the Handihaler regarding preference and ease of use. J. Aerosol Med. 2007;20:38–44. doi: 10.1089/jam.2006.0565. [DOI] [PubMed] [Google Scholar]
- 15.van der Palen J, van der Valk P, Goosens M, Groothuis-Oudshoorn K, Brusse-Keizer M. A randomised cross-over trial investigating the ease of use and preference of two dry powder inhalers in patients with asthma or chronic obstructive pulmonary disease. Expert Opin. Drug Deliv. 2013;10:1171–1178. doi: 10.1517/17425247.2013.817387. [DOI] [PubMed] [Google Scholar]
- 16.van der Palen J, Klein JJ, Schildkamp AM. Comparison of a new multidose powder inhaler (Diskus/Accuhaler) and the Turbuhaler regarding preference and ease of use. J. Asthma. 1998;35:147–152. doi: 10.3109/02770909809068202. [DOI] [PubMed] [Google Scholar]
- 17.van der Palen J, et al. Preference, satisfaction and errors with two dry powder inhalers in patients with COPD. Expert Opin. Drug Deliv. 2013;10:1023–1031. doi: 10.1517/17425247.2013.808186. [DOI] [PubMed] [Google Scholar]
- 18.Pascual S, et al. Preference, satisfaction and critical errors with Genuair and Breezhaler inhalers in patients with COPD: a randomised, cross-over, multicentre study. NPJ Prim. Care Respir. Med. 2015;25:15018. doi: 10.1038/npjpcrm.2015.18. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Price DB, et al. Inhaler errors in the CRITIKAL Study: type, frequency, and association with asthma outcomes. J. Allergy Clin. Immunol. Pract. 2017;5:1071–1081.e9. doi: 10.1016/j.jaip.2017.01.004. [DOI] [PubMed] [Google Scholar]
- 20.Molimard M, et al. Assessment of handling of inhaler devices in real life: an observational study in 3811 patients in primary care. J. Aerosol Med. 2003;16:249–254. doi: 10.1089/089426803769017613. [DOI] [PubMed] [Google Scholar]
- 21.Klijn SL, et al. Effectiveness and success factors of educational inhaler technique interventions in asthma & COPD patients: a systematic review. NPJ Prim. Care Respir. Med. 2017;27:24. doi: 10.1038/s41533-017-0022-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Giraud V, Roche N. Misuse of corticosteroid metered-dose inhaler is associated with decreased asthma stability. Eur. Respir. J. 2002;19:246–251. doi: 10.1183/09031936.02.00218402. [DOI] [PubMed] [Google Scholar]
- 23.Molimard M, Le Gros V. Impact of patient-related factors on asthma control. J. Asthma. 2008;45:109–113. doi: 10.1080/02770900701815727. [DOI] [PubMed] [Google Scholar]
- 24.Giraud V, Allaert FA, Magnan A. A prospective observational study of patient training in use of the autohaler inhaler device: the Sirocco study. Eur. Rev. Med. Pharmacol. Sci. 2011;15:563–570. [PubMed] [Google Scholar]
- 25.Natsir B, Yunus F, Antariksa B, Damayanti T. Relationship between inhaler usage technique with Asthma Controlled Test (ACT) and emergency department visits: PS103. Respirology. 2013;18:110. [Google Scholar]
- 26.Al-Jahdali H, et al. Improper inhaler technique is associated with poor asthma control and frequent emergency department visits. Allergy Asthma Clin. Immunol. 2013;9:8. doi: 10.1186/1710-1492-9-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Baddar S, Jayakrishnan B, Al-Rawas OA. Asthma control: importance of compliance and inhaler technique assessments. J. Asthma. 2014;51:429–434. doi: 10.3109/02770903.2013.871558. [DOI] [PubMed] [Google Scholar]
- 28.de Tarso Roth Dalcin P, et al. Factors related to the incorrect use of inhalers by asthma patients. J. Bras. Pneumol. 2014;40:13–20. doi: 10.1590/S1806-37132014000100003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Giraud V, Allaert FA, Roche N. Inhaler technique and asthma: feasibility and acceptability of training by pharmacists. Respir. Med. 2011;105:1815–1822. doi: 10.1016/j.rmed.2011.07.004. [DOI] [PubMed] [Google Scholar]
- 30.Yildiz F, Asthma Inhaler Treatment Study Group. Asthma Inhaler Treatment Study. Importance of inhaler device use status in the control of asthma in adults: the asthma inhaler treatment study. Respir. Care. 2014;59:223–230. doi: 10.4187/respcare.02478. [DOI] [PubMed] [Google Scholar]
- 31.Harnett CM, et al. A study to assess inhaler technique and its potential impact on asthma control in patients attending an asthma clinic. J. Asthma. 2014;51:440–445. doi: 10.3109/02770903.2013.876650. [DOI] [PubMed] [Google Scholar]
- 32.Levy ML, Hardwell A, McKnight E, Holmes J. Asthma patients’ inability to use a pressurised metered-dose inhaler (pMDI) correctly correlates with poor asthma control as defined by the global initiative for asthma (GINA) strategy: a retrospective analysis. Prim. Care Respir. J. 2013;22:406–411. doi: 10.4104/pcrj.2013.00084. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Molimard M, et al. Chronic obstructive pulmonary disease exacerbation and inhaler device handling: real-life assessment of 2935 patients. Eur. Respir. J. 2017;49:16017942017. doi: 10.1183/13993003.01794-2016. [DOI] [PubMed] [Google Scholar]
- 34.Maricoto T, et al. Assessment of inhalation technique in clinical and functional control of asthma and chronic obstructive pulmonary disease. Acta Med. Port. 2015;28:702–707. doi: 10.20344/amp.5905. [DOI] [PubMed] [Google Scholar]
- 35.Roggeri A, Micheletto C, Roggeri DP. Inhalation errors due to device switch in patients with chronic obstructive pulmonary disease and asthma: critical health and economic issues. Int. J. Chron. Obstruct. Pulmon. Dis. 2016;11:597–602. doi: 10.2147/COPD.S103335. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Sanchis J, Gich I, Pedersen S. Aerosol Drug Management Improvement Team (ADMIT). Systematic review of errors in inhaler use: has patient technique improved over time? Chest. 2016;150:394–406. doi: 10.1016/j.chest.2016.03.041. [DOI] [PubMed] [Google Scholar]
- 37.Bosnic-Anticevich S, et al. The use of multiple respiratory inhalers requiring different inhalation techniques has an adverse effect on COPD outcomes. Int. J. Chron. Obstruct. Pulmon. Dis. 2017;12:59–71. doi: 10.2147/COPD.S117196. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
All publications identified in the systematic literature review are accessible via the public databases (PubMed, Google/Google Scholar), or via the articles’ respective journal websites. All of the included data were sourced from these publicly available articles.