Abstract
Previously, median effective dose (ED) of 1.6 mSv per three-dimensional rotational angiography (3DRA) has been reported. This study evaluated ED and image quality in 3DRA after implementation of a simple dose reduction protocol in pediatric catheterizations. Simple conversion factors between 3DRA ED and readily available parameters at the cathlab were determined. The dose reduction protocol consisted of frame reduction (60–30 frames/s (f/s)), active collimation of the X-ray beam, usage of a readily available low dosage program, and a pre-3DRA run check. EDs were calculated with Monte Carlo PCXMC 2.0. Three observers blindly assessed 3DRA image quality of the dose reduction and normal-dose cohort. Between October 2014 and October 2015, 84 patients (median age 4.3 years) underwent 100 3DRAs with a median ED of 0.54 mSv (0.12–2.2) using the dose reduction protocol. Median ED in the normal-dose cohort (17 3DRAs) was 1.6 mSv (1.2–4.9). Image quality in the dose reduction cohort remained excellent. Correlations between ED and dose area product (DAP) and ED and skin dose were found with a ρ of 0.82 and 0.83, respectively. ED exposure of the entire catheterization was reduced to 2.64 mSv. Introduction of a simple protocol led to 66% dose reduction in 3DRA and 79% in the entire catheterization. 3DRA image quality in this group remained excellent. In 3DRA ED correlates well with DAP and skin dose, parameters readily available at the cathlab.
Electronic supplementary material
The online version of this article (10.1007/s00246-018-1943-3) contains supplementary material, which is available to authorized users.
Keywords: Catheterization, Rotational angiography, Radiation dose reduction, Effective dose
Introduction
In pediatric cardiology imaging is essential for diagnostic and interventional purposes. For this reason, patients with congenital heart disease regularly and increasingly receive radiation over the years [1]. Radiation exposure during childhood is more harmful than during adulthood. Reasons for this are the longer life span children have and the more harmful effects radiation has on developing tissue. With increased radiation exposure, children’s lifetime cancer risks will increase [2–5]. Catheterizations contribute to the majority of radiation burden in patients with congenital heart disease [6].
Quite recently, three-dimensional rotational angiography (3DRA) has been added to the spectrum of image modalities. 3DRA is used for diagnostic and interventional reasons. 3DRA provides a real-time roadmap for anatomy-guided procedures and improves faster and simplified interventions with enhanced patients’ safety [7]. Conversion factors enabling simple estimation of effective dose (ED) for standard procedures such as 3DRA acquisition have not yet been published [8]. Many studies report dose area product (DAP), a value directly provided by the imaging system [9, 10], but this value does not quantify the radiation’s effect on patients. The effective radiation dose (ED) is the best indicator to assess the stochastic effects of radiation [3]. Furthermore, ED enables comparison between the effects of 3DRA and effects of other imaging modalities [3].
Many studies directly estimate ED from DAP [10–12]. However, the relation between DAP and ED for 3DRA has never been strongly confirmed [13]. A strong correlation will help to produce a simple formula to estimate the ED, which is more practical for daily use than the complex ED calculations.
In 2014, Peters et al. have reported a median ED of 1.6 mSv per 3DRA in only 17 pediatric patients [14]. To limit the radiation burden, the ED should be reduced to a minimum with preservation of image quality. The ED might decrease with a few simple changes in the 3DRA protocol [2, 13, 14].
Therefore, the aim of the present study is (1) to calculate the EDs after implementation of a simple dose reduction protocol in a larger group of patients and compare the results with Peters et al., (2) to evaluate imaging quality of this protocol, and (3) to further explore the correlation between DAP and ED [14].
Materials and Methods
Study Population
Patients were eligible for inclusion if they were 0–18 years of age and had undergone a cardiac catheterization procedure with 3DRA acquisition at the Wilhelmina Children’s Hospital between October 2014 and October 2015. The institutional review board approved this study and no informed consent was required. Retrospective analysis of medical records and catheterization data was performed. Parameters collected include age, weight, height, body surface area (BSA), cardiac diagnosis, and type of intervention (if applicable). Patients were grouped according to their initial diagnosis. Patient characteristics of this low-dose cohort 3DRA were compared with a patient group previously reported, undergoing a normal-dose 3DRA [14]. Reasons for exclusion from ED calculation were incomplete rotation, wrong positioning of the patient, and insufficient contrast. As contrast absorbs radiation, insufficient contrast leads to less radiation exposure. 3DRAs made with a central venous catheter or because of a non-cardiac diagnosis were excluded from image quality assessment.
3DRA Image Acquisition
3DRAs were obtained using the Siemens Artis Zee biplane (Siemens, Forchheim, Germany) and reconstructions were transferred to the Leonardo workstation for post-processing with Syngo DynaCT Cardiac software. All procedures were performed under general anesthesia. Rapid atrial or ventricular pacing was performed in 88 of the 100 3DRAs. Pacing frequency was increased from 180/min upwards until a reduction of 50% of the systolic blood pressure was achieved. Contrast medium was administrated to the cardiac compartment prior to the region of interest meaning the right ventricle for pulmonary imaging and the left ventricle for aortic imaging. Contrast was diluted up to 60% with saline. Contrast was injected from 2 mL/s in 3 kg neonates up to 16 mL/s in 50 kg adolescents in case of a single injection site before start of 3DRA for 5 s. When multiple injection sites were necessary, additional manual injections with diluted contrast were performed.
Dose Reduction
Compared to the study of Peters et al., the number of frames per second was reduced from 60 to 30 f/s [6]. In addition, patients were scanned with a tube voltage corresponding to a low-dose program (Table S1). All patients were scanned according to a protocol of one weight class lower than the patient’s weight. Furthermore, a thick copper filter was used to filter out low-energy photons that can cause harm and do not contribute to the image quality. Collimation was applied with a diaphragm to protect irrelevant tissue from radiation and to prevent scattering rays from causing background haze. Before the actual run, tube current was checked to be below 100 mA. If not, image plane was checked for metal artifacts and the tube current was automatically adjusted accordingly.
Calculation of ED
All data required for calculation were extracted from Artis Zee biplane and converted to Microsoft Excel (Microsoft, USA) with CareAnalytics (Siemens, Erlangen, Germany). Parameters describing the geometry of the X-ray tube, the radiation quality (tube current, filter material and thickness, and anode angle), and the patient (age, height, and weight) were imported in Monte Carlo program PCXMC 2.0 (STUK, Finland) from Microsoft Excel to calculate ED. The outcomes of the Monte Carlo stimulations are according to International Commission on Radiological Protection 103 (ICRP103) organ weighing factors [15, 16].
Image Quality Assessment
3DRA images were extracted from the Leonardo workstation and patient characteristics and cohort information were removed from the files. One experienced pediatric cardiologist and two junior researchers blindly assessed image quality independently. Pre-defined anatomical structures, necessary for diagnosis or intervention, were separately scored on a three-point scale (good = 3, moderate = 2, and poor = 1).
Statistical Analysis
Continuous values were expressed as median with a range, and categorical values as a number and percentage of the total. Differences between baseline characteristics were tested for significance using T-test or 2-tailed Mann Whitney test, for normally distributed and skewed continuous values, respectively. Significant differences for gender, diagnosis, and image quality were tested with a Chi-square test. P < 0.05 was considered to be significant. A Kruskal–Wallis test was used to compare age, weight, and ED per initial diagnosis. Spearman’s correlation testing and linear regression were performed to evaluate the association of ED with patient’s age, height, weight, BSA, skin dose, DAP, and tube current in case of non-normally distributed variables. A Fleiss kappa was calculated to test interobserver agreement of the image quality. All statistical calculations were performed using Microsoft Excel 14.6.1.
Results
Exposure Data
Table 1 summarizes characteristics at baseline of the patients undergoing low-dose and normal-dose 3DRAs. For one patient, the height was estimated, because it was not measured during hospital admission. Analysis of baseline parameters between the two groups did not show statistically significant differences. In the low-dose group, 100 runs were performed in 84 patients. Ten patients underwent pre- and post-intervention 3DRA, and four patients had a second catheterization with a 3DRA. Furthermore, two patients had an additional 3DRA for evaluation of a second intervention and further evaluation of a possible vascular ring with esophageal contrast, respectively.
Table 1.
Low dose | Normal dose [14] | |
---|---|---|
Demographic patient data | ||
Number of patients | 84 | 14 |
Male, n (%) | 39 (46) | 7 (50) |
Age (years) | 4.29 (0.0–18.8) | 3.79 (0–16.6) |
Height (cm) | 103.0 (50–176) | 101.5 (50–184) |
Weight (kg) | 16.15 (2.4–89) | 14.5 (3.4–57.5) |
BSA (m2) | 0.69 (0.19–2.10) | 0.65 (0.23–1.68) |
Patient diagnosis | ||
PA + VSD/TOF, n (%) | 23 (27.4) | 4 (28.6) |
Aortic pathology, n (%) | 20 (23.8) | 3 (21.4) |
Univentricular heart, n (%) | 19 (22.6) | 2 (14.3) |
TGA, n (%) | 6 (7.1) | 2 (14.3) |
Genetic syndrome, n (%) | 8 (9.5) | 0 (0) |
Others, n (%) | 8 (9.5) | 3 (21.4) |
Procedural data | ||
Number of 3DRAs | 100 | 17 |
Interventional procedures, n (%) | 75 (75) | 12 (71) |
Continuous variables are summarized as median and range, and categorical variables are reported as number of cases (n) and percentage
BSA body surface area, PA pulmonary atresia, TGA transposition of the great arteries, TOF tetralogy of fallot, VSD ventricular septal defect
Dose Reduction
All 3DRAs were made with 30 f/s. In 96 of the 100 3DRA, weight protocols of one weight class lower were used. Table 2 shows the radiation exposure parameters after the dose reduction.
Table 2.
Low dose N = 100 |
Normal dose [14] N = 17 |
P value | |
---|---|---|---|
Exposure parameters | |||
Tube voltage 3DRA (kV) | 70 (60–96) | 90 (90–90) | < 0.001 |
Tube current 3DRA (mA) | 228 (53–395) | 69 (26–363) | < 0.001 |
Exposure time 3DRA (ms) | 465.5 (333.2–1330) | 843.2 (452.2–868) | < 0.001 |
Skin dose 3DRA (mGy) | 9 (1.7–83) | 20,86 (10,33–90,69) | < 0.001 |
DAP 3DRA (mGy·cm2) | 1279 (150–16,987) | 3128 (1231–17,273) | < 0.001 |
Procedural time (min) | 153 (30–360) | 165 (60–540) | 0.360 |
Total fluoroscopy time (min) | 28 (0.4–121) | 22 (6.4–81.4) | 0.497 |
Effective dose | |||
3DRA (mSv) | 0.54 (0.12–2.23) | 1.62 (0.70–4.94) | < 0.001 |
Fluoroscopy (mSv) | 1.53 (0.00–25.40) | 4.4 (0.2–15.8) | 0.002 |
Angiography (mSv) | 0.45 (0.00–19.33) | 3.6 (0–79.1) | 0.009 |
Total catheterization (mSv) | 2.64 (0.27–28.13) | 12.4 (2–99.9) | < 0.001 |
Interventional group | 3.65 (0.49–28.13) | ||
Diagnostic group | 0.86 (0.27–10.09) |
Values represent median and range
Effective Dose
After applying the dose-reducing protocol, a mean and median ED per 3DRA of 0.67 mSv (± 0.44 SD) and 0.54 mSv (range 0.12–2.2), respectively, were calculated compared to a mean and median ED per 3DRA of 2.0 (± 1.1 SD) and 1.6 mSv (1.2–4.9), respectively, in the normal-dose group (Fig. 1). The reduction in ED achieved with the dose reduction protocol was highly significant (P < 0.001, 95% CI 0.82–1.32). Differences in age, weight, and 3DRA ED between diagnostic groups were not significant. Only 2 of the 100 3DRAs had EDs higher than 2.0 mSv. These two patients were aged 15 and 16 years, much higher than median age of 4.3 years. Furthermore, the total procedural ED was reduced from a median total ED 12.4 mSv in the normal-dose group to 2.64 mSv in the low-dose group (P < 0.001, 95% CI 2.18–11.28) (Table 2). In patients with a diagnostic catheterization, a median total ED of 0.86 mSv was found. Also, fluoroscopy and angiography ED were reduced, while fluoroscopy and procedure time did not differ between the two groups. DAP and skin dose correlate very well with ED (Fig. 2). Spearman’s correlations coefficients between DAP and ED and skin dose and ED of the 100 low-dose 3DRAs were ρ 0.82 and 0.83, respectively. The correlations between DAP and ED and skin dose and ED were the strongest in patients aged 1–4.99 years, ρ 0.92 and 0.93, respectively. A multiple linear regression equation was calculated to predict ED based on DAP and weight. A significant regression equation
was found (F(2,97) = 195.435, P < 0.000), with a R2 of 0.80.
Image Quality
Blind assessment of image quality was possible in 93 of the 112 3DRAs, and those were included in our analysis. The median score in both cohorts was 3; 96% of the items had a median score of good. One anatomical structure in one patient was scored poor (1), and thus overall all images had sufficient quality for clinical decision-making. Comparing the two groups, there was a significantly better image quality in the low-dose cohort. Among the three different reviewers, fair agreement was seen, with a Fleiss kappa value of 0.34. 351 of the 390 (90%) scored items received the same score from all three reviewers. Images of a 3-month-old patient with an aortic coarctation and 3DRA ED of 0.12 mSv are shown in Fig. 3.
Discussion
With this study, we show that standardized use of traditional 2D angiography dose-lowering techniques leads to a 66% dose reduction in 3DRA with preserved (excellent) image quality.
Furthermore, a significant decrease in total procedural radiation dose was observed. Finally, the strong correlations between DAP and ED and skin dose and ED were confirmed and a simple formula to estimate 3DRA ED was proposed.
Implications
Annual average EDs are 3 mSv per person of which 80% is due to natural sources such as cosmic radiation. 0.6 mSv of annual average dose results from artificial sources such as medical exposure, atmospheric nuclear testing, and occupational exposure of which medical exposure accounts for almost 100% of the artificial ED [17]. Average annual exposure in pediatric cardiology population varies greatly. Patients with complex congenital heart disease such as a univentricular heart can have an average annual effective dose of 20 mSv solely because of ionizing radiation-producing medical examinations [4]. As a result in patients with complex heart disease, life attributable risk of cancer was 6.5% above baseline [4]. In pediatric cardiology, cardiac catheterizations contribute to 60% of the total radiation exposure. Thus, patients included in this study are exposed to higher risks of cancer development compared to the general population. Therefore, the dose reduction of the 3DRA solely with 1 mSv and almost 10 mSv for the entire procedure is highly relevant.
3DRA and Total Catheterization EDs
Low 3DRA EDs can be achieved using simple 2D angiography dose-lowering techniques and can, therefore, be easily applied in every cathlab. EDs of 3DRA have not been extensively calculated and reported in pediatric cardiology. Compared to other studies, our 3DRA ED is low. Table 3 provides an overview difference 3DRA ED values reported by previous studies. Reinke et al. suggested that an ED lower than 1 mSv is possible for 3DRA in real patients [18]. Our study is first to confirm this hypothesis and to show ED can be decreased to a minimum using simple techniques without any diagnostic image quality loss.
Table 3.
Author | Year of publication | Patients | Number of patients | Measure of center | ED (mSv) |
---|---|---|---|---|---|
Eloot [9] | 2013 | Adults | 40 | Median | 5.7 |
Wielandts [13] | 2010 | Adults | 42 | Mean | 6.6 |
De Buck [19] | 2013 | Adults | 40 (75% left atrium) | N/A | 2.6 (left atrium group) 1.2 (right atrium group) |
Haddad [10] | 2016 | Pediatric patient; age unknown | 1 | - | 1.8 |
Surendran [20] | 2017 | ≤ 2 years | 15 | Median | 1.35 |
N/A not available
Watson et al. compared computed tomographic angiography (CTA) with a diagnostic cardiac catheterization in pediatric patients. A median calculated ED of 0.74 mSv and 10.8 mSv was found for CTA and catheterization, respectively [21]. Ait-Ali et al. reported a total estimated ED for a diagnostic catheterization of 4.6 mSv in pediatric patients [22]. Our median total ED was 0.86 mSv in our diagnostic group and is comparable to the ED of CTA and much lower than the EDs of diagnostic catheterizations previously reported. Furthermore, in our study angiography ED was also significantly decreased. Thus, it seems that the excellent 3DRA image quality of this low-dose protocol reduces the need for additional angiographic imaging and the radiation burden of diagnostic catheterizations is comparable to CTA.
Image Quality
Even though 90% of the scored items received the same score, interobserver variability was fair. The small number of rating categories is the most reasonable explanation. Interestingly, image quality was better in the low-dose cohort. Enhanced operator experience and optimization of the post-processing process could explain this. Moreover, better protocols were available leading to optimal diagnostic image quality [23].
Correlations with ED
3DRA ED could be estimated with simple values as DAP and weight, which are readily available in every cathlab. Previously, Wielandts et al. found no relationship in adult patients between DAP and ED [13]. However, Peters et al. in children and Eloot et al. in adults found comparable Pearson’s correlation coefficients (r) of 0.87 and 0.92, respectively [9, 14]. Recently, Haddad et al. reported Pearson’s correlation coefficients ranging from 0.67 (aged > 15 years) to 0.98 (aged < 5 years) in 100 phantoms [10]. This confirms our finding that DAP seems to predict the ED best in the youngest patients. 3DRA is a standardized procedure and therefore DAP is likely to correlate well to ED. Correlations between ED and skin dose have not been widely published. Only from Peters et al. study a Pearson’s correlation coefficient of 0.91 can be derived, which is slightly higher than our Pearson’s correlation coefficient of 0.85.
Strengths and Limitations
One of the strengths of this study is the large group of patients. This is the first study reporting 3DRA EDs from 100 patients. A fair comparison could be made with previously reported 3DRA EDs from the same center where patients underwent the same 3DRA protocol except for the above-mentioned factors. Calculation of ED was done according to the current standard [3]. A limitation of this study is the heterogenous patient group and, therefore, patients could not be directly matched to patients from the control cohort. However, the anatomical roadmap required for the procedures does not differ enormously, nor can the 3DRA procedure. This would not affect 3DRA ED values greatly. Second, ED calculations have been done based on phantom models corrected for height and weight, but not for the exact age and gender. Third, our study was based on certain protocols and equipment, which not every institution might use. This makes comparison and reproducibility harder.
Conclusions
Standardized use of traditional 2D angiography dose-lowering techniques leads to a 66% dose reduction in 3DRA with preserved image quality and a significant decrease in total procedural radiation dose. DAP and skin dose are reliable predictors of ED. The usage of the dose-reducing steps described in this study is strongly advised.
Electronic supplementary material
Below is the link to the electronic supplementary material.
Acknowledgements
The authors would like to thank Stichting Hartekind for the financial support and Rens Wientjes for his support with extracting radiation data.
Abbreviations
- 3DRA
Three-dimensional rotational angiography
- DAP
Dose area product
- ED
Effective dose
Funding
This study was funded by Stichting Hartekind, the Netherlands.
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical Approval
This study was approved by our hospitals’ review board.
Informed Consent
No informed consent was required.
References
- 1.Ladouceur VB, Lawler PR, Gurvitz M, et al. Exposure to low-dose ionizing radiation from cardiac procedures in patients with congenital heart disease: 15-year data from a population-based longitudinal cohort. Circulation. 2016;133:12–20. doi: 10.1161/CIRCULATIONAHA.115.019137. [DOI] [PubMed] [Google Scholar]
- 2.Bacher K, Bogaert E, Lapere R, et al. Patient-specific dose and radiation risk estimation in pediatric cardiac catheterization. Circulation. 2005;111:83–89. doi: 10.1161/01.cir.0000151098.52656.3a. [DOI] [PubMed] [Google Scholar]
- 3.Khong PL, Ringertz H, Donoghue V, et al. General aspects of radiological protection in paediatric diagnostic imaging. Ann ICRP. 2013;42:21–28. doi: 10.1016/j.icrp.2012.10.001. [DOI] [PubMed] [Google Scholar]
- 4.Johnson JN, Hornik CP, Li JS, et al. Cumulative radiation exposure and cancer risk estimation in children with heart disease. Circulation. 2014;130:161–167. doi: 10.1161/CIRCULATIONAHA.113.005425. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Raissadati A, Nieminen H, Haukka J, et al. Late causes of death after pediatric cardiac surgery: a 60-year population-based study. J Am Coll Cardiol. 2016;68:487–498. doi: 10.1016/j.jacc.2016.05.038. [DOI] [PubMed] [Google Scholar]
- 6.Glatz AC, Purrington KS, Klinger A, et al. Cumulative exposure to medical radiation for children requiring surgery for congenital heart disease. J Pediatr. 2014;164:789–794.e10. doi: 10.1016/j.jpeds.2013.10.074. [DOI] [PubMed] [Google Scholar]
- 7.Stenger A, Dittrich S, Glöckler M. Three-dimensional rotational angiography in the pediatric Cath Lab: optimizing aortic interventions. Pediatr Cardiol. 2016;37:528–536. doi: 10.1007/s00246-015-1310-6. [DOI] [PubMed] [Google Scholar]
- 8.Hill KD, Frush DP, Han BK, et al. Radiation safety in children with congenital and acquired heart disease: a scientific position statement on multimodality dose optimization from the image gently alliance. JACC Cardiovasc Imaging. 2017;10:797–818. doi: 10.1016/j.jcmg.2017.04.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Eloot L, Bacher K, Steenbeke F, et al. Three-dimensional rotational X-ray acquisition technique is reducing patients’ cancer risk in coronary angiography. Catheter Cardiovasc Interv. 2013;82:419–427. doi: 10.1002/ccd.24879. [DOI] [PubMed] [Google Scholar]
- 10.Haddad L, Waller BR, Johnson J, et al. Radiation protocol for three-dimensional rotational angiography to limit procedural radiation exposure in the pediatric cardiac catheterization lab. Congenit Heart Dis. 2016;11:637–646. doi: 10.1111/chd.12356. [DOI] [PubMed] [Google Scholar]
- 11.Corredoira E, Vañó E, Ubeda C, Gutiérrez-Larraya F. Patient doses in paediatric interventional cardiology: impact of 3D rotational angiography. J Radiol Prot. 2015;35:179–195. doi: 10.1088/0952-4746/35/1/179. [DOI] [PubMed] [Google Scholar]
- 12.Manica JLL, Borges MS, de Medeiros RF, et al. A comparison of radiation dose between standard and 3D angiography in congenital heart disease. Arq Bras Cardiol. 2014;103:131–137. doi: 10.5935/abc.20140118. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Wielandts JY, De Buck S, Ector J, et al. Three-dimensional cardiac rotational angiography: effective radiation dose and image quality implications. Europace. 2010;12:194–201. doi: 10.1093/europace/eup394. [DOI] [PubMed] [Google Scholar]
- 14.Peters M, Krings G, Koster M, et al. Effective radiation dosage of three-dimensional rotational angiography in children. Europace. 2015;17:611–616. doi: 10.1093/europace/euu207. [DOI] [PubMed] [Google Scholar]
- 15.Valentin RJ. The 2007 Recommendations of the International Commission on Radiological Protection. Ann ICRP. 2007;37:81–123. doi: 10.1016/j.icrp.2006.06.001. [DOI] [PubMed] [Google Scholar]
- 16.Servomaa A. Organ dose calculation in medical X ray examinations by the program PCXMC. Radiat Prot Dosim. 1998;80:213–219. doi: 10.1093/oxfordjournals.rpd.a032509. [DOI] [Google Scholar]
- 17.UNSCEAR (2010) Sources and effects of ionizing radiation: sources Annex B. Exposures of the public and workers from various sources of radiation
- 18.Reinke G, Halbfaß J, Dittrich S, et al. Three-dimensional rotational angiography in congenital heart disease: estimation of radiation exposure. Open J Radiol. 2013;3:124–129. doi: 10.4236/ojrad.2013.33020. [DOI] [Google Scholar]
- 19.De Buck S, Alzand BSN, Wielandts JY, et al. Cardiac three-dimensional rotational angiography can be performed with low radiation dose while preserving image quality. Europace. 2013;15:1718–1724. doi: 10.1093/europace/eut140. [DOI] [PubMed] [Google Scholar]
- 20.Surendran S, Waller BR, Elijovich L, et al. Use of 3-D digital subtraction rotational angiography during cardiac catheterization of infants and adults with congenital heart diseases. Catheter Cardiovasc Interv. 2017;90:618–625. doi: 10.1002/ccd.27180. [DOI] [PubMed] [Google Scholar]
- 21.Watson TG, Mah E, Joseph Schoepf U, et al. Effective radiation dose in computed tomographic angiography of the chest and diagnostic cardiac catheterization in pediatric patients. Pediatr Cardiol. 2013;34:518–524. doi: 10.1007/s00246-012-0486-2. [DOI] [PubMed] [Google Scholar]
- 22.Ait-Ali L, Andreassi MG, Foffa I, et al. Cumulative patient effective dose and acute radiation-induced chromosomal DNA damage in children with congenital heart disease. Heart. 2010;96:269–274. doi: 10.1136/hrt.2008.160309. [DOI] [PubMed] [Google Scholar]
- 23.Starmans NLP, Krings GJ, Molenschot MMC, et al. Three-dimensional rotational angiography in children with an aortic coarctation. Neth Heart J. 2016;24:666–674. doi: 10.1007/s12471-016-0899-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.