Abstract
The majority of femoral fractures are surgically treated with intramedullary nails. Non-union rate is low but challenging and costly if it occurs. There have been encouraging results from the use of augmentative plating as a treatment for non-union of femoral fractures. We performed a systematic review of the literature to compare union rates, time to union and complications between exchange nailing and augmentative plating as a primary procedure following a diagnosis of femoral non-union following initial nailing. We found a total of 21 papers, which found the mean union rate of augmentative plating to be 99.8% compared to 74% (P = 2.05−12) found for exchange nailing. Times to union were comparable at 5.9 months for augmentative plating and 6.3 months for exchange nailing (P = 0.68916), and complication rate was 4% for augmentative plating compared to 20% for exchange nailing. From the evidence available, plate augmentation provides a more reliable union rate if used as the first operative intervention on a non-union of a femoral fracture compared to exchange nailing.
Level of Evidence IV Systematic review of therapeutic studies.
Keywords: Femoral non-union, Femoral fracture, Intramedullary nailing, Adjunctive plating
Background
In the western world, the majority of displaced femoral diaphyseal fractures are treated operatively using intramedullary nails (IMNs) with early return to function and a low incidence of complications [1]. When femoral non-unions do occur, the treatment options can be time consuming, challenging and expensive. Based upon a health economics study from the UK, the cost of treating a femoral non-union is at least £17,000 (US $22,000) per patient [2]. From the patients’ perspective, femoral non-unions result in on-going pain, altered gait, delayed return to work and psychosocial impairment. Therefore, treatment methods that improve the likelihood of fracture union combined with a reduced “time-to-union” are welcomed by clinicians and patients alike. Currently the most common method of treating femoral diaphyseal non-unions is to perform an exchange-nailing procedure, whereby the original IM nail is removed, and the femoral canal is reamed to stimulate the natural healing response. Reaming permits a larger diameter IMN to be inserted thus improving the mechanical stability. Despite this exchange, IM nailing is not as uniformly successful and persistent non-unions do occur.
An alternative treatment is the augmentation of the IM nail with a plate and screws, and this shows real promise at reducing the incidence of persistent non-union following IM nailing. If the clinical assessment determines the IMN has maintained its structural integrity and unlikely to fail before the fracture has healed, plate augmentation in isolation can be undertaken. One major advantage to this technique is that surgical exposure of the fracture site will permit the surgeon to remove fibrous tissue and freshen the fracture ends as a stimulus for healing; there is opportunity also for direct bone grafting at the surgeon’s discretion.
In this article, we have conducted a comprehensive systematic review of the published literature on exchange IM nailing and on plate augmentation of femoral diaphyseal fractures that were treated initially, but unsuccessfully, with IMNs.
Methods
Identification and eligibility of relevant studies
A comprehensive literature search using Medline, EMBase, Cochrane and CINAHL was conducted on the 11 April 2016, from inception to this date, identifying relevant studies using the key words and terms “femoral fracture non-union,” “exchange nailing” and “plate augmentation”. Inclusion criteria were papers written in English and limited to surgical procedures undertaken for aseptic femoral diaphyseal non-unions that had been treated with an IMN initially and with only one surgical revision procedure; infected non-unions were excluded. Other inclusion criteria were use of a measure of “time to union” by the same radiological method: this was bridging callus present on at least 3 cortices on the anteroposterior and lateral radiographs of the femur. Details of complications following surgery to treat the non-union were noted.
Data extraction
Two reviewers screened the titles and the abstracts of all potentially relevant publications independently. Full articles were critiqued, and those judged to be eligible for consideration in the study were examined in greater detail. Any controversial papers were discussed in detail between the two reviewers. If doubt still existed regarding suitability for inclusion, the final decision for inclusion or exclusion was undertaken by the senior clinician. Each paper was then reviewed in turn to extract the following data using a standardised proforma including the following headings: study design, patient age, gender, treatment method used to treat the established femoral diaphyseal non-union, percentage of patients obtaining union after the index procedure and time to union according to established radiological criteria. A proportion of the papers had included patients who had had several previous operative attempts for established non-union prior to the index intervention. Only those papers that could provide a sufficient breakdown of each patient’s treatment history undertaken prior to the index procedure were included in our analysis. Only those individuals who had undergone one procedure after developing a non-union were considered. In addition to exchange nailing, some authors described combining this technique with open bone grafting of the fracture site for some patients within their series.
Outcome measures
The primary outcome measure was radiographic union after the index intervention. Radiographic union was defined as bridging callus across at least three cortices on anteroposterior and lateral views of the femur. The secondary outcome measure was the time to union measured in months. The tertiary outcome measure was the incidence and severity of perioperative and postoperative complications. The prevalence of use of bone grafting in both techniques was recorded.
Due to significant study heterogeneity, particularly with regard to the methods and timing of patient follow-up following the index procedure, it was deemed inappropriate to pool the data from the published articles for meta-analysis. A narrative approach was used.
Statistics
The Chi-squared test was used to compare the union rate between the two surgical treatments exchange nailing and augmentation plating. Time to union was compared using the Mann–Whitney U test between the two surgical interventions. Analysis was performed using the SPSS statistics software (SPSS Science Inc, Chicago, Illinios, USA).
Results
Our initial literature search identified 396 suitable studies for further evaluation. Removal of duplicated titles left 182 studies. Of these, 158 were omitted because the title or abstract did not fulfil the inclusion criteria. The remaining 28 studies were analysed in detail. Seven further studies were excluded subsequently because of either: (a) an inability to isolate patients that had undergone only one procedure to treat the non-union or; (b) where the time to radiographic union was not accurately recorded. The flowchart in Fig. 1 illustrates our compliance to the PRISMA method of systematic review. Twenty-one papers remained for analysis. The papers included: one cohort study comparing the two methods [3]; one randomised control trial comparing closed versus open bone-grafting techniques along with exchange nailing [4]; one cohort series comparing the effect of reaming size when exchanging the nail [5]; seven exchange-nailing case series [6–13]; one cohort series comparing augmented plating versus exchange plating [14]; and ten plate augmentation case series [15–23]. Tables 1 and 2 include details of the papers grouped according to exchange IMN and plate augmentation, respectively.
Table 1.
Year of publication | Author | Type of study | Definition of non-union | Number of femoral non-unions | Open fracture | Average age | Months until exchange nailing | Bone graft used | Months until union | Radiographic union after first procedure | Complications |
---|---|---|---|---|---|---|---|---|---|---|---|
2003 | Banaszkiewicz P. A. | Case series | – | 19 | 4 | 36 | – | 0 | 9 | 37 | 2 infected non-unions 2 broken nails |
1999 | Furlong A. J. | Case series | Pain on mobilisation Minimum of 6 months no progression on monthly radiographs for 3 months |
25 | 2 | 39 | – | 12 | 7 | 96 | Nil |
2010 | Park J. | Cohort series: Exchange nailing vs. plate augmentation | Pain on weight bearing Minimum of 6 months no progression on monthly radiographs for 3 months |
5 | – | 40 | 15 | 0 | 6 | 20 | Nil |
2009 | Shroeder J. E. | Case series | No radiographic union after 6 months and pain on weight bearing | 35 | 6 | 32 | – | 0 | 4 | 94 | Nil |
2015 | Swanson E. A. | Case series | Pain and no radiographic (plain radiograph or CT) improvement | 50 | 8 | 40 | – | 0 | 6 | 72 | Nil |
2015 | Tsang S. T. | Case series | – | 20 | – | 38 | 8 | 3 | 9 | 85 | Nil |
2000 | Weresh M. J. | Case series | Pain on weight bearing Minimum of 6 months no progression on monthly radiographs for 3 months |
19 | 0 | 39 | – | 0 | 8 | 53 | Nil |
1997 | Wu C. | Randomised controlled trial: closed vs. open bone-grafting techniques | – | 18 | – | 37 | – | – | 5 | 100 | Nil |
2007 | Wu C. | Cohort series: effect of reaming size | – | 41 | – | 34 | – | 0 | 4 | 92 | Nil |
2002 | Yu C. W. | Case series | No radiographic union after 1 year after treatment or another method is required to obtain union | 25 | 0 | 38 | 20 | 0 | 6 | 92 | Nil |
257 | 37 | 13 | 6.3 | 74 |
Table 2.
Year of publication | Author | Type of study | Definition of non-union | Number of femoral non-unions | Open fracture | Average age | Months until augmentative plating | Bone graft | Months until union | Radiographic union after first procedure | Complications |
---|---|---|---|---|---|---|---|---|---|---|---|
2010 | Chen C. M. | Case series | – | 50 | – | 44 | – | 50 | 6 | 100 | 1 superficial infection |
2005 | Choi Y. S. | Case series | Pain on weight bearing Minimum of 6 months no progression on monthly radiographs for 3 months |
8 | 4 | 37 | 10 | 8 | 8 | 100 | Nil |
2011 | Gao K. D. | Case series | Clinical and radiographic unhealed fracture | 9 | – | 38 | 17 | 9 | 7 | 100 | Nil |
2011 | Hakeos W. M. | Case series | Pain on weight bearing Minimum of 6 months no progression on monthly radiographs for 3 months |
7 | 2 | 43 | 9 | 2 local 5 posterior iliac spine | 6 | 100 | 1 superficial infection |
2015 | Jhunjhunwala H. R. | Case series | No evidence of radiographic union for greater than 6 months | 40 | 1 | 35 | 6 | 24 | 4 | 98 | 1 infected non-union |
2014 | Jiang L. | Cohort series: Augmentation vs. exchange plating | Pain and no evidence of radiographic union greater than 6 months | 12 | – | 46 | 12 | 12 anterior iliac spine | 4 | 100 | 1 plate removal |
2012 | Lin C. J. | Case series | – | 22 | 0 | 34 | 20 | 13 anterior iliac spine | 5.5 | 100 | Nil |
2008 | Nadkarni B. | Case series | Pain and no evidence of radiographic union | 2 | 0 | 44 | 29 | 2 anterior iliac spine | 6.5 | 100 | Nil |
2010 | Park J. | Cohort series: Exchange nailing vs. plate augmentation | Pain on weight bearing Minimum of 6 months no progression on monthly radiographs for 3 months |
4 | – | 48 | 11 | 4 | 8 | 100 | Nil |
2011 | Said G. Z. | Case series | – | 14 | – | 42 | 8 | 9 anterior iliac spine | 4 | 100 | 1 superficial wound infection |
1997 | Ueng S. W. | Case series | Pain and no evidence of radiographic union | 17 | 0 | 34 | 15 | 7 | 7 | 100 | Nil |
2012 | Ye J. | Case series | 6 | 0 | 47 | 12 | 6 anterior iliac spine | 4.5 | 100 | 1 plate removal | |
191 | 41 | 13.5 | 5.9 | 99.8 |
The union rate with plate augmentation was 99.8% (190/191) compared to 74% (190/257) with exchange nailing. This difference was statistically significant (Chi-squared, p = 2.05−12). Time to radiographic union averaged 5.9 months (4–8 months) following plate augmentation in comparison with 6.3 months (4–9 months) following exchange nailing. No statistically significant difference was demonstrated using the Mann–Whitney U test (p = 0.68916).
Complications included persistence of the aseptic non-unions requiring yet further surgery and three deep infections at the site of non-union following exchange nailing. One deep infection occurred at the site of a non-union after plate augmentation. Plate augmentation was also associated with its own specific complications which included three superficial infections treated successfully with dressings and antibiotics and plate prominence requiring removal after union.
Bone grafting in the exchange-nailing cohort was used 10% of the time compared to 79% of the time in the augmented plate group. Unfortunately the data for these cases were not presented separately to enable analysis, but within the exchange-nailing group, studies that used bone grafting [5, 7, 10, 13] led to a union rate of 91%.
Discussion
The current gold standard treatment for an established femoral diaphyseal non-union is exchange IM nailing. Our findings in a systematic review of the literature suggest there is evidence to support plate augmentation as an alternative approach. Direct comparisons in the literature between the two methods of treatment are difficult given the differences in study designs. However, there is a very high radiographic union rate achieved in all of the plate augmentation studies compared with the lower union rates of exchange nailing documented in 10 of the 11 studies included in our analysis. Where union was successful, the time to union was similar for both surgical techniques.
Most aseptic diaphyseal non-unions of the femur are thought to occur due to mechanical instability, predominantly rotational instability [24, 25]. Exchange nailing to a larger diameter nail goes some way to improving the biomechanical environment as larger diameter IMNs are torsionally stiffer. In addition, provided the intramedullary canal is not reamed excessively, better frictional contact can be achieved between the nail and surrounding bone by increasing the surface diameter of the IMN. Park et al. [26] share this opinion and have added that non-isthmal fractures decrease fracture stability since the IMN can no longer make contact with the inner aspect of the cortical bone at the level of the fracture. They postulate that fractures that occur outside the isthmus are prone to non-union since IMN-bone stability becomes increasingly dependent upon the stability offered by the proximal and distal locking screws that have significantly poorer biomechanical properties compared with the IMN itself. Another major consideration is that the majority of modern IMNs are made from titanium that have a Young’s modulus of approximately 60% of stainless steel. This alloy is the material of choice for many as it is more resistant to fatigue failure as compared with stainless steel. For this reason, the implant industry has manufactured IMNs that have very distally placed cross-screw holes without an associated high risk of IMN failure. Although these extreme distal locking options are advantageous for treating very distally placed fractures within the metaphyseal region of the femur, extreme distal locking adds no advantage when treating diaphyseal fractures. Long cross-screws are mechanically inferior to shorter ones and the degree of instability is further accentuated from the poorer hold that distally placed cross-screws have in the surrounding metaphyseal bone where the cortices are much thinner compared to the meta-diaphysis area. Titanium cross-screws are also more flexible than stainless steel screws, so the combination of increased cross-screw length and increased cross-screw flexibility can significantly reduce IMN-bone construct stability especially with respect to rotational forces [26]. Very distal locking options have perhaps led to an increased risk of non-union, and most of the papers assessed used these IMNs. Although it is recognised that IMNs are good at resisting bending forces and maintaining axial stability, their reduced rotational stability cannot be improved unless shorter and broader IM nails are used. These would permit shorter and, therefore, more mechanically stable cross-screws to be inserted. Careful consideration of the length of IMN that is required to treat each fracture type is needed without reverting to the long accepted “reinforcing the whole of the bone” concept. In this way, it may be possible to reduce the incidence of non-unions happening in the first instance and to improve the overall success rate after exchange nailing. However, plate augmentation would appear to be a promising alternative way of achieving fracture stability but avoiding the potentially negative effects to the abductor mechanism and soft tissues when exchanging an IMN.
The importance of rotational instability in femoral diaphyseal non-unions is exemplified by attempts to encourage union through dynamising femoral nails; the subsequent union rates are less than 50% and are associated with a risk of significant limb shortening by more than 2 cm [27–29]. Dynamisation increases axial loading of the fracture and should, in theory, encourage bone healing, but unless the bone ends can interlock in a way that restores rotational control, the instability and mechanical environment for healing will be made worse by the removal of one or more of the distal cross-screws.
Plate osteosynthesis used in isolation after removal of an intramedullary nail has been described in association with insertion of bone graft [30], but plates used alone are unpopular as patients have to comply with restricted weight bearing to reduce the risk of plate failure. By comparison, plate augmentation of a femur already stabilised by an IMN offers the advantages of a more favourable mechanical environment for bone healing and that of immediate weight bearing after surgery. In most of the papers reviewed, the preferred plate and screw system was a 4.5-mm compression plate, being either low contact or standard plate design. The recorded success suggests these implants are stiff enough to resist the rotational forces present at a femoral non-union and, when applied in a compression mode, may also help to limit excessive axial displacement.
Bone grafting was used more widely in the plating group as compared to the exchange-nailing group most likely because the non-union site was exposed during this procedure. Unfortunately, these patients’ data were never separated or analysed except for one paper [7] where a small improvement in union time was found but not statistically significant.
The complication rates for exchange nailing were 20% compared to 4% for plate augmentation. However, if failure to achieve satisfactory bone union is excluded, plate augmentation is associated with a 4% complication compared with 1% for exchange nailing. This may relate to the risk of infection that accompanies larger surgical approaches that communicate with the site of non-union. In comparison, revision nailing is not a benign procedure; often removal of the nail and insertion of another can lead to substantial damage to the abductor muscles and their insertions.
The limitations to this study include it being a systematic review of low-level evidence studies, mostly level IV case series, involving relatively small numbers of patients. The fractures are heterogenous. The definition of non-union differed in the papers which may have led to some over- or under-treatment of the non-union. Using radiographic time to union is notoriously difficult as opinions vary and the timing of clinical reviews is not consistent. The majority of the papers failed to perform patient-reported outcome measures.
Conclusion
The results of plate augmentation are at least as good as and perhaps better than exchange IM nailing for treating established femoral diaphyseal non-unions that had been primarily stabilised using an IMN. The time to union following the plate augmentation or exchange nailing is similar.
Bone grafting around the fracture site can be an adjunct for successful union.
Both techniques are associated with a low incidence of complications. Although postoperative infection is more common following plate augmentation, the incidence of serious deep-seated infection would appear to be no higher when compared with exchange IMN.
Authors contribution
G. Medlock and I. Stevenson contributed to all aspects of production of the manuscript. A. Johnstone contributed to all parts of production except data acquisition. All have been involved with drafting and critical review and have approved the final version for production.
Funding
This study received no funding.
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical approval
This was a review of other studies involving humans and the authors sort ethical approval appropriately.
Informed consent
No informed consent was required as this was a systematic review of other papers.
Contributor Information
Gareth Medlock, Phone: 0345 456 6000, Email: gareth.medlock@nhs.net.
Iain M. Stevenson, Email: iainstevenson@nhs.net
Alan J. Johnstone, Email: a.johnstone@nhs.net
References
- 1.Winquist RA, Hansen ST, Jr, Clawson DK. Closed intramedullary nailing of femoral fractures. A report of five hundred and twenty cases. 1984. J Bone Joint Surg Am. 2001;83-A(12):1912. doi: 10.2106/00004623-200112000-00021. [DOI] [PubMed] [Google Scholar]
- 2.Kanakaris NK, Giannoudis PV. The health economics of the treatment of long-bone non-unions. Injury. 2007;38(Suppl 2):S77–84. doi: 10.1016/S0020-1383(07)80012-X. [DOI] [PubMed] [Google Scholar]
- 3.Park J, Kim SG, Yoon HK, Yang KH. The treatment of nonisthmal femoral shaft nonunions with im nail exchange versus augmentation plating. J Orthop Trauma. 2010;24(2):89–94. doi: 10.1097/BOT.0b013e3181b8dafd. [DOI] [PubMed] [Google Scholar]
- 4.Wu CC, Chen WJ. Treatment of femoral shaft aseptic nonunions: comparison between closed and open bone-grafting techniques. J Trauma. 1997;43(1):112–116. doi: 10.1097/00005373-199707000-00026. [DOI] [PubMed] [Google Scholar]
- 5.Wu CC. Exchange nailing for aseptic nonunion of femoral shaft: a retrospective cohort study for effect of reaming size. J Trauma. 2007;63(4):859–865. doi: 10.1097/01.ta.0000233663.24838.76. [DOI] [PubMed] [Google Scholar]
- 6.Banaszkiewicz PA, Sabboubeh A, McLeod I, Maffulli N. Femoral exchange nailing for aseptic non-union: not the end to all problems. Injury. 2003;34(5):349–356. doi: 10.1016/S0020-1383(02)00191-2. [DOI] [PubMed] [Google Scholar]
- 7.Furlong AJ, Giannoudis PV, DeBoer P, Matthews SJ, MacDonald DA, Smith RM. Exchange nailing for femoral shaft aseptic non-union. Injury. 1999;30(4):245–249. doi: 10.1016/S0020-1383(99)00073-X. [DOI] [PubMed] [Google Scholar]
- 8.Shroeder JE, Mosheiff R, Khoury A, Liebergall M, Weil YA. The outcome of closed, intramedullary exchange nailing with reamed insertion in the treatment of femoral shaft nonunions. J Orthop Trauma. 2009;23(9):653–657. doi: 10.1097/BOT.0b013e3181a2a337. [DOI] [PubMed] [Google Scholar]
- 9.Swanson EA, Garrard EC, Bernstein DT, O’Connor DP, Brinker MR. Results of a systematic approach to exchange nailing for the treatment of aseptic femoral nonunions. J Orthop Trauma. 2015;29(1):21–27. doi: 10.1097/BOT.0000000000000166. [DOI] [PubMed] [Google Scholar]
- 10.Tsang ST, Mills LA, Baren J, Frantzias J, Keating JF, Simpson AH. Exchange nailing for femoral diaphyseal fracture non-unions: risk factors for failure. Injury. 2015;46(12):2404–2409. doi: 10.1016/j.injury.2015.09.027. [DOI] [PubMed] [Google Scholar]
- 11.Weresh MJ, Hakanson R, Stover MD, Sims SH, Kellam JF, Bosse MJ. Failure of exchange reamed intramedullary nails for ununited femoral shaft fractures. J Orthop Trauma. 2000;14(5):335–338. doi: 10.1097/00005131-200006000-00005. [DOI] [PubMed] [Google Scholar]
- 12.Wu CC, Shih CH, Chen WJ, Tai CL. Treatment of ununited femoral shaft fractures associated with locked nail breakage: comparison between closed and open revision techniques. J Orthop Trauma. 1999;13(7):494–500. doi: 10.1097/00005131-199909000-00006. [DOI] [PubMed] [Google Scholar]
- 13.Yu CW, Wu CC, Chen WJ. Aseptic nonunion of a femoral shaft treated using exchange nailing. Chang Gung Med J. 2002;25(9):591–598. [PubMed] [Google Scholar]
- 14.Jiang L, Pan Z, Zheng Q. Augmentation plating in hypertrophic non-union after nail fixation of femoral shaft fracture compared with exchange plating. Chin Med J (Engl) 2014;127(16):2900–2905. [PubMed] [Google Scholar]
- 15.Chen CM, Su YP, Hung SH, Lin CL, Chiu FY. Dynamic compression plate and cancellous bone graft for aseptic nonunion after intramedullary nailing of femoral fracture. Orthopedics. 2010;33(6):393. doi: 10.3928/01477447-20100429-18. [DOI] [PubMed] [Google Scholar]
- 16.Choi YS, Kim KS. Plate augmentation leaving the nail in situ and bone grafting for non-union of femoral shaft fractures. Int Orthop. 2005;29(5):287–290. doi: 10.1007/s00264-005-0668-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Hakeos WM, Richards JE, Obremskey WT. Plate fixation of femoral nonunions over an intramedullary nail with autogenous bone grafting. J Orthop Trauma. 2011;25(2):84–89. doi: 10.1097/BOT.0b013e3181dfbb33. [DOI] [PubMed] [Google Scholar]
- 18.Jhunjhunwala HR, Dhawale AA. Is augmentation plating an effective treatment for non-union of femoral shaft fractures with nail in situ? Eur J Trauma Emerg Surg. 2015;Apr:29. doi: 10.1007/s00068-015-0534-8. [DOI] [PubMed] [Google Scholar]
- 19.Lin CJ, Chiang CC, Wu PK, Chen CF, Huang CK, Su AW, et al. Effectiveness of plate augmentation for femoral shaft nonunion after nailing. J Chin Med Assoc. 2012;75(8):396–401. doi: 10.1016/j.jcma.2012.06.008. [DOI] [PubMed] [Google Scholar]
- 20.Nadkarni B, Srivastav S, Mittal V, Agarwal S. Use of locking compression plates for long bone nonunions without removing existing intramedullary nail: review of literature and our experience. J Trauma-Inj Infect Crit Care. 2008;65(2):482–486. doi: 10.1097/TA.0b013e31817c9905. [DOI] [PubMed] [Google Scholar]
- 21.Said GZ, Said HG, El-Sharkawi MM. Failed intramedullary nailing of femur: open reduction and plate augmentation with the nail in situ. Int Orthop. 2011;35(7):1089–1092. doi: 10.1007/s00264-010-1192-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Ueng SW, Chao EK, Lee SS, Shih CH. Augmentative plate fixation for the management of femoral nonunion after intramedullary nailing. J Trauma. 1997;43(4):640–644. doi: 10.1097/00005373-199710000-00013. [DOI] [PubMed] [Google Scholar]
- 23.Ye J, Zheng Q. Augmentative locking compression plate fixation for the management of long bone nonunion after intramedullary nailing. Arch Orthop Trauma Surg. 2012;132(7):937–940. doi: 10.1007/s00402-012-1497-4. [DOI] [PubMed] [Google Scholar]
- 24.Park K, Kim K, Choi YS. Comparison of mechanical rigidity between plate augmentation leaving the nail in situ and interlocking nail using cadaveric fracture model of the femur. Int Orthop. 2011;35(4):581–585. doi: 10.1007/s00264-010-0983-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Ma X, Wang M, Zhang B, Wang Z, Lang Y, Gong X. Anti-rotational plates with intramedullary nailing stabilize femoral shaft fracture nonunion: a biomechanical study. Int J Surg Res Pract. 2015;2(2):2–3. doi: 10.23937/2378-3397/1410034. [DOI] [Google Scholar]
- 26.Park J, Yang KH. Indications and outcomes of augmentation plating with decortication and autogenous bone grafting for femoral shaft nonunions. Injury. 2013;44(12):1820–1825. doi: 10.1016/j.injury.2013.02.021. [DOI] [PubMed] [Google Scholar]
- 27.Pihlajamaki HK, Salminen ST, Bostman OM. The treatment of nonunions following intramedullary nailing of femoral shaft fractures. J Orthop Trauma. 2002;16(6):394–402. doi: 10.1097/00005131-200207000-00005. [DOI] [PubMed] [Google Scholar]
- 28.Wu CC. The effect of dynamization on slowing the healing of femur shaft fractures after interlocking nailing. J Trauma. 1997;43(2):263–267. doi: 10.1097/00005373-199708000-00010. [DOI] [PubMed] [Google Scholar]
- 29.Wu CC, Chen WJ. Healing of 56 segmental femoral shaft fractures after locked nailing. Poor results of dynamization. Acta Orthop Scand. 1997;68(6):537–540. doi: 10.3109/17453679708999022. [DOI] [PubMed] [Google Scholar]
- 30.Ring D, Jupiter JB, Sanders RA, Quintero J, Santoro VM, Ganz R, et al. Complex nonunion of fractures of the femoral shaft treated by wave-plate osteosynthesis. J Bone Joint Surg Br. 1997;79(2):289–294. doi: 10.1302/0301-620X.79B2.6886. [DOI] [PubMed] [Google Scholar]
- 31.Moher D, Liberati A, Tetzlaff J, Altman DG, The PRISMA Group Preferred reporting item for systematic reviews and meta-analyses: the PRISMA statement. PLoS Med. 2009;6(6):e1000097. doi: 10.1371/journal.pmed1000097. [DOI] [PMC free article] [PubMed] [Google Scholar]