Abstract
Piezosurgery is an innovative technique widely used for osteotomies in the field of oral and maxillofacial surgery. The surgical technique has been clinically supposed to cut mineralized bone selectively with reducing the risk of damage to adjacent soft tissues. However, none of the previous literature has reported any evidence of scientific experiments to examine performance of the piezoelectric device, i.e. the time required for cutting bone and the effect on soft tissues under the standardized conditions. This study was designed to test the hypothesis that cutting time of the piezoelectric device is longer than that of rotary instruments while the cut surface of bone is smoother and soft tissues are less damaged with piezosurgery under the standardized experimental system. We measured the time for cutting bone and soft tissues of rats with the piezoelectric device and rotary instruments. Damage to soft tissues was examined histologically, and the cut surface of bone was investigated using scanning electron microscopy. Our study demonstrated experimentally that piezosurgery provides a smooth cut bony surface with no damage to soft tissues and takes longer time to cut bone than conventional drillings. We propose that piezosurgery is beneficial for medical safety and usability.
Introduction
Piezosurgery is an innovative osteotomy technique using piezoelectric ultrasonic vibrations. Ultrasonic vibrating instruments for cutting mineralized tissue have been reported since the 1950s1,2. In 1988, Italian oral sugeon, Tomaso Vercellotti, developed a piezoelectric osteotomy device, which provided the opportunity for the widespread clinical use of piezosurgery3,4. Currently, piezoelectric devices are used widely for osteotomies, such as maxillary sinus lift, impacted mandibular third molar extraction, and bone grafting, in the field of oral and maxillofacial surgery2,4.
The major advantage of piezosurgery is its selective cutting of mineralized bone. Frequencies higher than 50 kHz are needed to cut soft tissues. The piezoelectric device is designed to produce ultrasonic microvibrations of 60–210 μm at a frequency of 25–30 kHz. Thus, piezoelectric devices, different from conventional rotary instruments and microsaws, are able to cut only mineralized tissues5–7. Therefore, piezosurgery can reduce considerably the risk of damage to adjacent soft tissues, such as blood vessels, nerves, and mucous membranes, in cases of osteotomy8–11. Other advantages of piezosurgery include reduction of overheating resulting from the generation of a cavitation effect and better visibility of the surgical area due to less bleeding2,12,13. Moreover, use of piezosurgery for extraction of the impacted mandibular third molar was reported to produce less facial swelling and trismus postoperativily compared to that of rotary osteotomy14,15. Recent studies on the healing of bone defects experimentally created by piezosurgery demonstrated no difference in the newly formed bone volume and the healing process between piezosurgery and conventional osteotomy techniques16,17. Meanwhile, piezosurgery was shown to have a longer surgical time than conventional osteotomies15,18.
Most studies on comparison of piezosurgery and conventional osteotomies have focused on postoperative outcomes, and little is known about the cutting performance of the piezoelectric device itself. With respect to the time required for cutting, surgical time in clinical cases has been investigated to date, while little information is available about experimentally measured cutting time with the piezoelectric device compared to conventional osteotomy instruments. Furthermore, few studies also have reported the effect of piezosurgery on soft tissues.
The present study was designed to test the hypothesis that cutting time of the piezoelectric device is longer than that of rotary instruments while the cut surface of bone is smoother and soft tissues are less damaged with piezosurgery under the condition of the standardized cutting force, 1.5 N, which was recommended in a previous study19. We characterized piezosurgery by comparing it to conventional drillings. We established an experimental system to assess cutting performance using a force gauge and measured the time required for cutting. We also examined the damage to soft tissues with histology and the cutting surface morphology of bone with scanning electron microscopy (SEM).
Results
Fluctuation of cutting force
Samples from six rats were cut with the piezoelectric device and conventional rotary instruments; that is, carbide, fissure, and round burs. The time for cutting is shown in the graph for typical samples (Fig. 1a–h).
The force was stable and fluctuation was limited to approximately 1.5 N until the tibia was cut through with the piezoelectric device (Fig. 1d), while the force fluctuated widely with the rotary instruments (Fig. 1a–c). Conventional drillings took a few seconds to cut through the tongue with a force of less than 1.5 N (Fig. 1e–g). In contrast, piezosurgery showed limited fluctuation of pressure for the tongue as for the tibia, but did not cut the tongue (Fig. 1h).
Time required for cutting
The time required to cut through the tibia through is shown in Table 1. Piezosurgery took the longest (average 255.06 s), while the carbide bur took the shortest (average 3.94 s) time, and the round bur took the longest time among the conventional drillings (average 32.44 s).
Table 1.
Carbide | Fissure | Round | Piezosurgery | ||
---|---|---|---|---|---|
Tibia | Time(sec) | 4.44 | 7.86 | 30.15 | 204.01 |
3.02 | 6.15 | 23.76 | 202.06 | ||
4.56 | 6.87 | 45.1 | 345.8 | ||
3.42 | 7.76 | 39.74 | 139.65 | ||
3.99 | 9.71 | 24.28 | 238.57 | ||
4.18 | 8.99 | 31.62 | 220.29 | ||
Mean | 3.94 | 7.89 | 32.44 | 225.06 | |
SD | 0.6 | 1.31 | 8.5 | 67.9 | |
Tongue | Time(sec) | 1.53 | 1.34 | 7.37 | * |
1.19 | 1.04 | 8.34 | * | ||
1.45 | 0.96 | 9.98 | * | ||
1.38 | 0.85 | 5.77 | * | ||
0.93 | 1.01 | 6.04 | * | ||
0.77 | 0.98 | 7.81 | * | ||
Mean | 1.21 | 1.03 | 7.55 | — | |
SD | 0.3 | 0.17 | 1.55 | — |
*Time for piezosurgery was not shown because the tongue was not cut within 20 seconds.
The time required to cut through the tongue is also shown in Table 1. The fissure bur took the shortest (average 1.03 s) and the round bur took the longest (average 7.55 s) time. Piezosurgery did not cut the tongue.
The time required for cutting was analysed statistically and is shown in Fig. 1i,j. The time for piezosurgery to cut through the tibia was the longest and was significantly different from that for the carbide (P = 0.002), fissure (P = 0.002), and round (P = 0.002) burs. The time for the carbide bur was the shortest among the conventional drillings and was significantly different from that for the fissure (P = 0.001) and round (P = 0.002) burs. The round bur took longer than the fissure bur (P = 0.002).
The times for the carbide and fissure burs were not statistically different. The round bur took longer than the carbide (P < 0.001) and the fissure (P < 0.001) burs. Piezosurgery did not cut the tongue.
The region of tongue to which each instrument was applied
The region of the tongue to which each instrument was applied was examined histologically. Damage caused by the rotary instruments extended from mucous epithelia through submucous and muscular layers (Fig. 2a–c). The carbide and round burs showed large irregular (Fig. 2a) and shallow and irregular (Fig. 2c) damaged regions, respectively, and the fissure bur caused a relatively smooth surface (Fig. 2b). Piezosurgery made a dent, but did not damage any tongue epithelia or lingual papilla (Fig. 2d).
Bony regions to which each instrument was applied
SEM showed a smooth surface due to the piezoelectric device (Fig. 3d). The carbide bur caused a rough surface with bone debris (Fig. 3a). The fissure bur made fine line scratches (Fig. 2b). The round bur caused a scale-like surface (Fig. 2c).
Discussion
It has been reported clinically that piezosurgery cuts only bone but does not damage soft tissues5–7. Piezosurgery has been proposed as a minimally invasive surgical technique in cranial base and spine surgery20. To our knowledge, our study confirmed the reported clinical advantage of piezosurgery experimentally, i.e. giving no damage to soft tissues for the first time. Conventional drillings cut through the tongue within 10 s, while piezosurgery only made a dent on the tongue surface and cut none of the examined tongue samples (Table 1; Figs 1j and 2d). The tongue was not cut even though the time was extended by 200 s (data not shown). Piezosurgery is beneficial for protection of soft tissues during bone surgery.
Various clinical trials in the published literature have been reviewed systematically to evaluate the clinical efficacy of peizosurgery with meta-analyses. A systematic review indicated that piezosurgery can reduce the risk of the Schneiderian membrane perforation in maxillary sinus floor augmentation compared to conventional rotary osteotomy21. Another review examined outcomes for the surgical extraction of mandibular third molars and found a lower risk of neurological complications with the piezoelectric device than with conventional rotary burs22. A recent report showed less postoperative neurosensory disturbance after orthognathic surgery with piezosurgery compared to conventional techniques23. These meta-analyses suggested that piezosurgery does not damage soft tissues, which has been supported by our results of the present study.
The operative field of osteotomies and bone biopsies in piezosurgery is said to have bleeding from surrounding soft tissues reduced significantly and the operative field almost free of blood during the cutting compared to conventional drillings, where blood is moved in and out of the cutting area and visibility is decreased4. It has been supposed to be due to the cavitation effect created by the cooling fluid distribution and/or by the type of vibration the instrument generates, in which the blood is essentially washed away. Our results of the experiments demonstrated that piezosurgery does not damage soft tissues and provided the scientific supports for the clinical empirical concepts for the first time.
Previous reports suggested that piezosurgery took longer to cut bone compared to conventional drillings, most of which described the time required for clinical surgery but did not compare the performance of the instruments experimentally15,18. Therefore, our study established the original experimental model to measure the time required for cutting standardized samples (e.g., tibia and tongue of 10-week-old rats) and compared results among the instruments. Our results indicated that conventional drilling cut through the tibia within one minute, while piezosurgery took longer than three minutes (Table 1; Fig. 1i). Our study also demonstrated experimentally that piezosurgery took longer to cut bone compared to conventional drillings.
The force to cut the tibia fluctuated widely with conventional drillings, although we attempted to keep a constant pressure of approximately 1.5 N (Fig. 1a–c). This may be due to fluctuation made by the burs while they kept contact with the samples. The round bur might have fluctuated widely at the beginning of cutting and the fluctuation decreased later when the bur fit the defect it made. In contrast, fluctuation of piezosurgery was limited compared to the conventional drillings (Fig. 1d), which suggests that piezosurgery is able to cut bone with a stable force with little fluctuation of the insert tip of the piezoelectric device.
The present study examined the bony regions to which each instrument was applied using SEM. As a result, piezosurgery caused a smooth surface without bone debris (Fig. 3) compared to other conventional drillings, which supports a report of equivalent rabbit experiments made by Maurer and colleagues24. It has not been known how piezosurgery affects viability and function of cellular components. The viability and differentiation potency of derived osteogenic cells showed no difference between autogenous bone chips harvested with piezosurgery and those with conventional drillings25. Osteocytes were maintained in the cut region made by piezosurgery and conventional drillings26. Previous studies suggested that there are no significant differences in healing of osteotomy sites between piezosurgery and conventional drillings16,17. The smear layer left on the bone surface by piezosurgery may even impair bone healing27. Preparation of the smooth cut surface may not contribute to maintaining cell viability and differentiation potency, and further investigations are required to provide better understanding.
There are some limitations to this study. Firstly, the cutting force in the experiments was 1.5 N and the force larger than 1.5 N was not applied. The larger cutting force may damage the tongue mucosa and reduce the cutting time. Secondly, we used only the OT-7 tip for piezosurgery. The use of different tips may cause different results of damage to the mucosa and cutting time. Other conditions of cutting force and tips should be examined for cutting time and damage given to the mucosa in the future.
Our study demonstrated experimentally cutting time of the piezoelectric device is statically longer than that of conventional rotary instruments while piezosurgery provides a smooth cut bony surface with no damage to soft tissues under the condition of the standardized cutting force, 1.5 N. Based on these results, we propose that piezosurgery is beneficial for medical safety and usability.
Methods
Animals and sample preparation
All experimental procedures conformed to “Regulations for Animal Experiments and Related Activities at Tohoku University”, and were reviewed by the Institutional Laboratory Animal Care and Use Committee of Tohoku University, and finally approved by the President of the University.
We obtained 10-week-old male Wistar rats weighing 210–230 g from the SLC Corporation (Japan SLC, Inc., Hamamatsu, Japan). They were euthanized with an overdose of isoflurane by inhalation. Tibias and tongues were resected as samples.
Osteotomy instruments
For piezoelectric osteotomies, Piezosurgery Touch® with an OT-7 tip (Mectron, Carasco, Italy) was used with a frequency of 30 kHz (special mode) under 20 mL/min irrigation. For conventional drilling, VOLVERE Vmax® (Nakanishi, Inc., Kanuma, Japan) with carbide, fissure and round burs was used at 20 × 1000 rpm under irrigation.
Measurement of cutting time
A force gauge (ZTS-50N; Imada Co., Ltd., Toyohashi, Japan) and a force-time graphing software (Force Recorder Standard; Imada Co., Ltd., Toyohashi, Japan) were used to measure the time required for cutting samples. A force gauge was equipped with a handpiece holder set at a distance of 5 cm from an axis of the force gauge to the tip of the insert tip or burs (Fig. 1). Using each instrument, tibias were cut transversely proximally at a site 10 mm away from the junction of the tibia with the fibula. Tongues were divided into two halves at the midline, and halves were cut transversely at a site 10 mm from the apex. The time required for cutting through tibias and tongues at approximately 1.5 N with the piezoelectric device or rotary instruments was measured.
Statistical analysis
Statistical analysis was performed with SPSS 22.0 (IBM Corp., Armonk, NY, USA) to compare the time required for cutting samples between the piezoelectric device and rotary instruments. Data were compared by one-way analysis of variance (ANOVA) followed by the Games-Howell test. The significance level was set at 0.05.
Histology
The tongue samples were fixed in 4% paraformaldehyde in 0.1 M phosphate buffer, pH 7.4. After dehydration through a graded series of ethanol solutions, samples were embedded in paraffin. Sections 5μm thick were cut and processed for hematoxylin-eosin (H-E) staining.
Surface morphology by SEM
Proximal and distal epiphyses of tibias were cut and bone marrow was removed with a wire 0.3 mm in diameter. After irrigation of the bone marrow cavity with physiological saline, tibias were cut transversely proximally at a site 10 mm away from the junction of the tibia with the fibula by piezosurgery or conventional drilling. Eventually, 3–4 mm long specimens were prepared for SEM. Specimens were fixed in 2% paraformaldehyde with 2% glutaraldehyde in 0.1 M phosphate buffer, pH 7.4, dehydrated, replaced with isoamyl acetate, and dried using a critical point dryer (JCPD-5; JEOL, Tokyo, Japan). The dried specimens were coated with platinum and the surface morphology was examined by scanning electron microscope (SEM) (JSM-6390LA, JEOL, Tokyo, Japan).
Acknowledgements
This work was supported by JSPS KAKENHI Grant Numbers JP15K11002, JP15K11003, JP17K17081. We thank Mr. Yasuto Mikami, Division of Craniofacial Development and Regeneration, Tohoku University Graduate School of Dentistry, for his excellent assistance.
Author Contributions
Y.O. performed the experiments and A.H. instructed him in performing the experiments. M.N., the corresponding author designed the study and drafted the main manuscript text. T.T. and Y.S. contributed to interpretation of the data and revised the manuscript. All authors reviewed the manuscript.
Data Availability Statement
All data generated or analysed during this study are included in this published article.
Competing Interests
The authors declare no competing interests.
Footnotes
Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Horton JE, Tarpley TM, Jr., Wood LD. The healing of surgical defects in alveolar bone produced with ultrasonic instrumentation, chisel, and rotary bur. Oral surgery, oral medicine, and oral pathology. 1975;39:536–546. doi: 10.1016/0030-4220(75)90192-9. [DOI] [PubMed] [Google Scholar]
- 2.Stubinger S, Stricker A, Berg BI. Piezosurgery in implant dentistry. Clinical, cosmetic and investigational dentistry. 2015;7:115–124. doi: 10.2147/CCIDE.S63466. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Magrin GL, Sigua-Rodriguez EA, Goulart DR, Asprino L. Piezosurgery in Bone Augmentation Procedures Previous to Dental ImplantSurgery: A Review of the Literature. The open dentistry journal. 2015;9:426–430. doi: 10.2174/1874210601509010426. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Pavlikova G, et al. Piezosurgery in oral and maxillofacial surgery. International journal of oral and maxillofacial surgery. 2011;40:451–457. doi: 10.1016/j.ijom.2010.11.013. [DOI] [PubMed] [Google Scholar]
- 5.Labanca M, Azzola F, Vinci R, Rodella LF. Piezoelectric surgery: twenty years of use. The British journal of oral & maxillofacial surgery. 2008;46:265–269. doi: 10.1016/j.bjoms.2007.12.007. [DOI] [PubMed] [Google Scholar]
- 6.Schlee M, Steigmann M, Bratu E, Garg AK. Piezosurgery: basics and possibilities. Implant dentistry. 2006;15:334–340. doi: 10.1097/01.id.0000247859.86693.ef. [DOI] [PubMed] [Google Scholar]
- 7.Robiony M, Polini F, Costa F, Vercellotti T, Politi M. Piezoelectric bone cutting in multipiece maxillary osteotomies. Journal of oral and maxillofacial surgery: official journal of the American Association of Oral and Maxillofacial Surgeons. 2004;62:759–761. doi: 10.1016/j.joms.2004.01.010. [DOI] [PubMed] [Google Scholar]
- 8.Schaeren S, et al. Assessment of nerve damage using a novel ultrasonic device for bone cutting. Journal of oral and maxillofacial surgery: official journal of the American Association of Oral and Maxillofacial Surgeons. 2008;66:593–596. doi: 10.1016/j.joms.2007.03.025. [DOI] [PubMed] [Google Scholar]
- 9.Beziat JL, Bera JC, Lavandier B, Gleizal A. Ultrasonic osteotomy as a new technique in craniomaxillofacial surgery. International journal of oral and maxillofacial surgery. 2007;36:493–500. doi: 10.1016/j.ijom.2007.01.012. [DOI] [PubMed] [Google Scholar]
- 10.Stubinger S, Kuttenberger J, Filippi A, Sader R, Zeilhofer HF. Intraoral piezosurgery: preliminary results of a new technique. Journal of oral and maxillofacial surgery: official journal of the American Association of Oral and Maxillofacial Surgeons. 2005;63:1283–1287. doi: 10.1016/j.joms.2005.05.304. [DOI] [PubMed] [Google Scholar]
- 11.Kotrikova B, et al. Piezosurgery–a new safe technique in cranial osteoplasty? International journal of oral and maxillofacial surgery. 2006;35:461–465. doi: 10.1016/j.ijom.2005.12.006. [DOI] [PubMed] [Google Scholar]
- 12.Landes CA, et al. Critical evaluation of piezoelectric osteotomy in orthognathic surgery: operative technique, blood loss, time requirement, nerve and vessel integrity. Journal of oral and maxillofacial surgery: official journal of the American Association of Oral and Maxillofacial Surgeons. 2008;66:657–674. doi: 10.1016/j.joms.2007.06.633. [DOI] [PubMed] [Google Scholar]
- 13.Laird WR, Walmsley AD. Ultrasound in dentistry. Part 1–Biophysical interactions. Journal of dentistry. 1991;19:14–17. doi: 10.1016/0300-5712(91)90030-3. [DOI] [PubMed] [Google Scholar]
- 14.Barone A, et al. A randomized clinical evaluation of ultrasound bone surgery versus traditional rotary instruments in lower third molar extraction. Journal of oral and maxillofacial surgery: official journal of the American Association of Oral and Maxillofacial Surgeons. 2010;68:330–336. doi: 10.1016/j.joms.2009.03.053. [DOI] [PubMed] [Google Scholar]
- 15.Sortino F, Pedulla E, Masoli V. The piezoelectric and rotatory osteotomy technique in impacted third molar surgery: comparison of postoperative recovery. Journal of oral and maxillofacial surgery: official journal of the American Association of Oral and Maxillofacial Surgeons. 2008;66:2444–2448. doi: 10.1016/j.joms.2008.06.004. [DOI] [PubMed] [Google Scholar]
- 16.Esteves JC, et al. Dynamics of bone healing after osteotomy with piezosurgery or conventional drilling - histomorphometrical, immunohistochemical, and molecular analysis. Journal of translational medicine. 2013;11:221. doi: 10.1186/1479-5876-11-221. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Ma L, Stubinger S, Liu XL, Schneider UA, Lang NP. Healing of osteotomy sites applying either piezosurgery or two conventional saw blades: a pilot study in rabbits. International orthopaedics. 2013;37:1597–1603. doi: 10.1007/s00264-013-1908-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Rullo R, Addabbo F, Papaccio G, D’Aquino R, Festa VM. Piezoelectric device vs. conventional rotative instruments in impacted third molar surgery: relationships between surgical difficulty and postoperative pain with histological evaluations. Journal of cranio-maxillo-facial surgery: official publication of the European Association for Cranio-Maxillo-Facial Surgery. 2013;41:e33–38. doi: 10.1016/j.jcms.2012.07.007. [DOI] [PubMed] [Google Scholar]
- 19.Hennet P. Piezoelectric Bone Surgery: A Review of the Literature and Potential Applications in Veterinary OromaxillofacialSurgery. Frontiers in veterinary science. 2015;2:8. doi: 10.3389/fvets.2015.00008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Schaller Bernhard J., Gruber R, Merten H A., Kruschat Thomas, Schliephake H, Buchfelder Michael, Ludwig H C. Piezoelectric Bone Surgery: A Revolutionary Technique for Minimally Invasive Surgery in Cranial Base and Spinal Surgery? Technical Note. Operative Neurosurgery. 2005;57:E410. doi: 10.1227/01.NEU.0000176700.77461.C9. [DOI] [PubMed] [Google Scholar]
- 21.Jordi C, Mukaddam K, Lambrecht JT, Kuhl S. Membrane perforation rate in lateral maxillary sinus floor augmentation using conventional rotating instruments and piezoelectric device-a meta-analysis. International journal of implant dentistry. 2018;4:3. doi: 10.1186/s40729-017-0114-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Badenoch-Jones EK, David M, Lincoln T. Piezoelectric compared with conventional rotary osteotomy for the prevention of postoperative sequelae and complications after surgical extraction of mandibular third molars: a systematic review and meta-analysis. The British journal of oral & maxillofacial surgery. 2016;54:1066–1079. doi: 10.1016/j.bjoms.2016.07.020. [DOI] [PubMed] [Google Scholar]
- 23.Pagotto LEC, de Santana Santos T, de Vasconcellos SJA, Santos JS, Martins-Filho PRS. Piezoelectric versus conventional techniques for orthognathic surgery: Systematic review and meta-analysis. Journal of cranio-maxillo-facial surgery: official publication of the European Association for Cranio-Maxillo-Facial Surgery. 2017;45:1607–1613. doi: 10.1016/j.jcms.2017.06.011. [DOI] [PubMed] [Google Scholar]
- 24.Maurer P, et al. Micromorphometrical analysis of conventional osteotomy techniques and ultrasonic osteotomy at the rabbit skull. Clinical oral implants research. 2008;19:570–575. doi: 10.1111/j.1600-0501.2007.01516.x. [DOI] [PubMed] [Google Scholar]
- 25.Chiriac G, Herten M, Schwarz F, Rothamel D, Becker J. Autogenous bone chips: influence of a new piezoelectric device (Piezosurgery) on chip morphology, cell viability and differentiation. Journal of clinical periodontology. 2005;32:994–999. doi: 10.1111/j.1600-051X.2005.00809.x. [DOI] [PubMed] [Google Scholar]
- 26.Heinemann F, et al. Experimental and histological investigations of the bone using two different oscillating osteotomy techniques compared with conventional rotary osteotomy. Annals of anatomy = Anatomischer Anzeiger: official organ of the Anatomische Gesellschaft. 2012;194:165–170. doi: 10.1016/j.aanat.2011.10.005. [DOI] [PubMed] [Google Scholar]
- 27.Baek KW, et al. A comparative investigation of bone surface after cutting with mechanical tools and Er:YAG laser. Lasers in surgery and medicine. 2015;47:426–432. doi: 10.1002/lsm.22352. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
All data generated or analysed during this study are included in this published article.