Abstract
Background:
Controlling of the renal vessels is a critical step in live donor nephrectomy (LDN). Currently, mainly 2 devices, Hem-o-Lok clips and staplers, are utilized to control vessels during LDN. Both of them have advantages and disadvantages.
Methods:
This systematic review and meta-analysis was aimed to compare the safety and the efficacy of the 2 devices and to identify the better one in LDN. A systematic search for related publications in the databases of PubMed, Medline, Embase, the Cochrane Library, and Web of Science through February 2018 was performed. Eight studies were selected and evaluated with the Newcastle-Ottawa Scale (NOS).
Results:
The meta-analysis result showed that utilization of Hem-o-Lok clips resulted in greater amount of estimated blood loss (mean differences [MD]: 40.10; 95% confidence interval [CI] 4.37–75.84) and longer time of warm ischemia (WIT) (MD: 55.61; 95% CI 36.79–74.43) than using staplers. Residual vascular length of grafts in clip group was longer than that in stapler group (MD: 2.51; 95% CI 0.78–4.24). However, the differences in primary outcomes such as device failure rate, death rate, and severe hemorrhage rate, were not significant between these 2 groups. In addition, utilization of Hem-o-Lok clips costed approximate $400 lower than staplers per patient.
Conclusion:
This study revealed that Hem-o-Lok clips and staplers have the similar function in LDN renal ligation, regarding the device failure rate, death rate, and severe hemorrhage rate. However, the surgeons would benefit from the clips in terms of the residue length of vessels, these outstanding features provide operation convenience and flexibility, such as right-sided donor nephrectomies, early vascular bifurcation, and rare vascular variation. In addition, the clips have potential economic advantages. In some developing countries, it would reduce the healthcare expenditure.
Keywords: clip, laparoscopic live donor nephrectomy, meta-analysis, stapler
1. Introduction
Laparoscopic techniques started to be widely applied in urologic surgeries since 1990s with its first application in 1991.[1] While the first live donor nephrectomy (LDN) was performed in 1995.[2] Controlling of the renal pedicle is the critical step in this surgery. Safety concerns have prompted a tremendous advance in vessel ligation devices. Now various devices are available for controlling the renal pedicles, including non-absorbable polymer locking clips (Hem-o-Lok clips), titanium clips, Endo-GIA staplers, Endo-TA staplers, and so on. These devices can be roughly divided into 2 major groups, clips and staplers. In our study, utilization of Hem-o-Lok clip, Endo-GIA stapler, and Endo-TA staplers in LDN were reviewed and analyzed.
Endo-GIA stapler became the first device being used to ligate renal vessels and have been the standard ligation tool since then. It is considered to be safe and effective.[3,4] However, several studies suggested the malfunctions of using Endo-GIA stapler.[5–8] One of disadvantages is severe hemorrhage which required conversion from laparoscopic to open surgery or even led to the death of patient. Secondly, it provides shorter length of graft vessel for anastomosis and thirdly, the medical cost is higher.[9] Therefore, surgeons have tried to secure renal pedicle with Hem-o-Lok clips since 2000 in that the clips are cheaper and can provide longer length of graft vessels.[10] However, it is also associated with severe bleeding and more severely, death of the patient[11–14] due to the slippage and dislodgement. Both the manufacturer of Hem-o-Lok clip and the US Food and Drug Administration (FDA) warned that Hem-o-Lok clip is contraindicated for renal artery ligation during LDN after 2006.[11,14,15]
Hem-o-Lok clips and staplers both showed advantages and disadvantages. To our knowledge, no studies evaluating the safety of vessel ligation devices during LDN have been published so far. This systematic review and meta-analysis study was aimed to identify the better device for vessel ligation during LDN and to provide the guidelines for clinical practice.
2. Methods
2.1. Data sources and literature search strategy
We conducted this meta-analysis in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) statement. Two reviewers performed a comprehensive literature search in the following databases: PubMed, Embase, and the Cochrane Library database. The last search was updated on February 2018. The publication language was restricted to English. Key words used were [“nephrectomy” AND (“Stapler” OR “stapling” OR “GIA” OR “TA” OR “gastrointestinal anastomosis” OR “Transfixion”) AND (“Clip” OR “clipping” OR “Hem-o-Lok” OR “Non-Transfixion”)]. The resulted literatures were further screened to exclude duplications followed by content screening which was achieved by title and abstract reading. After eligible studies were picked out by reading the title and abstract, a manual searching for more eligible publications were performed among the references of the literatures after content screening.
2.2. Inclusion and exclusion criteria
The inclusion criteria for this study were as the following: studies comparing staplers and clips used in LDN; studies provided main outcome data evaluating the safety and reliability of the devices. If ≥2 identified studies investigated the same data source or population, the largest or the most recent study was selected.
The exclusion criteria for this study were as follows: descriptive studies without comparison between clips and staplers; review articles, meta-analysis, case reports, or conference abstracts; studies based on animal or in vitro assay; studies about open surgery; duplicated studies and repeated analyses.
2.3. Data extraction and quality assessment
The following information was extracted from each selected study independently by 2 independent reviewers (LY and HZL). Basic data included first author, purpose of surgery, surgical approach, number of patients and type of devices. Primary outcomes were device failure rate, death rate, severe hemorrhage rate, and the cost of devices. Secondary outcomes were estimated blood loss, transfusion rate, rate of open surgery conversion, reoperation rate, residual vessel length, operation duration and warm ischemia time (WIT). If a study provided both univariate and multivariate analysis results, the multivariate analysis would be selected to achieve higher accuracy. Any discrepancies were resolved by discussion to reach a consensus. Quality of selected studies was assessed by 2 independent reviewers (LY and HZL). All the studies included were retrospective cohort studies. A study with a score ≥6 was considered high-quality study after the selected publication were evaluated with the Newcastle-Ottawa Scale (NOS).[16]
2.4. Data synthesis and statistical analysis
Data were extracted and analyzed by 2 independent researchers (LY and CYT) using Review manager 5.3 (The Cochrane Collaboration, Oxford, UK). Dichotomous variables were analyzed using risk ratio (RR), while continuous variables were analyzed using the mean differences (MD) and the corresponding 95% CIs (confidence intervals). A P value <.05 was considered statistically significant. Heterogeneity was quantified using the I2. A random effects model and a fixed effects model were applied for I2 > 50% and I2 < 50%, respectively. If I2 was greater than 50%, sensitivity analysis would be performed to identify the origin of heterogeneity.
2.5. Sensitivity analysis
To assess whether an individual study had an impact on the result, sensitivity analysis was performed for all included individual studies using the random-effects model. We examined the effects of study with the heaviest weighting by removing it and observing the change in I2 of several outcomes.
3. Results
3.1. Study selection
A total of 2845 reports were found from the database search by using the described searching strategy (Fig. 1). One hundred fifty-eight duplicated articles were excluded by using Endnote X7 (Thomson Corporation, Canada). After title and abstract screening, 21 relevant studies were identified. In addition, 9 relevant studies were extracted from the references of these 21 studies, resulting in 30 publications in total. These 30 studies were read thoroughly and carefully to extract the related information. During the process, 22 articles were excluded based on the inclusion and exclusion criteria. Thus, 8 studies were finally eligible for the meta-analysis.
Figure 1.
Diagram showing literature searching process.
3.2. Study characteristics and quality assessment
All the selected studies were retrospective cohort studies. Regarding the surgical approaches, LDN in 5 studies were performed trans-peritoneally and 1 of them was used retro-peritoneally. The total number of included patients was 32,145 (clip group, 13,833; stapler group, 18,312) with one of the sample size was extremely large.[15] Usage of Hem-o-Lok clips were found in all 8 studies. For stapler usage, 4 out 8 studies used Endo-GIA stapler, other studies used TA stapler. Three studies[9,17,18] performed group comparison based on demographic characteristics and results showed no statistical difference. Besides bleeding, other existed complications included infection, ileus, and bowel injury (Table 1).[17,22] All studies were scored in accordance with the NOS and the scores ranged from 7 to 9 (Table 2).
Table 1.
Characteristics of the reviewed studies.
Table 2.
Newcastle-Ottawa Scale score of the reviewed studies.
3.3. Primary outcomes
There was no significant difference between the Hem-o-Lok clips and staplers groups regarding device failure rate (risk ratio [RR]: 0.77; 95% CI: 0.51–1.16; P = .22), death rate (RR: 3.14; 95% CI: 0.35–27.84; P = .30), and severe hemorrhage rate (RR: 1.34; 95% CI: 0.26–6.86; P = .72) (Fig. 2). The average cost of Hem-o-Lok clips was 40 dollars, about 400 dollars (200–1440) lower than that of staplers per patient.[20–22]
Figure 2.
Device failure rate (A), death rate (B), and severe hemorrhage rate (C) in clip group and stapler group. Dichotomous variables were analyzed using risk ratio (RR), while continuous variables were analyzed using the mean differences (MD). 95% CIs was also calculated. A value of P < .05 was considered statistically significant. CI = confidence interval.
3.4. Secondary outcomes
The Hem-o-Lok clip group had significantly greater estimated amount of blood loss (MD: 40.10; 95% CI: 4.37–75.84; P = .03) and longer WIT than stapler group (MD: 55.61; 95% CI: 36.79–74.43; P < .001). Residual vascular length in Hem-o-Lok clip group was longer than that in stapler group (MD: 2.51; 95% CI: 0.78–4.24; P = .004). However, the differences of conversion rate (RR 0.72, 95% CI: 0.12–4.47, P = .73), transfusion rate (RR 0.74, 95% CI: 0.22–2.48, P = .63), reoperation rate (RR 6.29; 95% CI: 0.52–76.07; P = .15) and operative duration (MD: 17.45 minutes; 95% CI: 47.97–82.88; P = .60) were not statistically significant between the 2 groups (Fig. 3). Because of the heaviest weighting of the study performed by Hsi et al,[15] we excluded it and found that there was still no difference in the severe hemorrhage rate or the conversion rate. The I2 of the severe hemorrhage rate or the conversion rate dropped from 43% and 52% to 0% and 19%, respectively. This demonstrated that the heterogeneity was mainly caused by enormous differences in sample size between studies.
Figure 3.
Estimated amount of blood loss (A), length of warm ischemia time (WIT) (B), residual vascular length (C), conversion to open surgery rate (D), transfusion rate (E), reoperation rate (F), and length of operative time (G) in clip group and stapler group. Dichotomous variables were analyzed using risk ratio (RR), while continuous variables were analyzed using the mean differences (MD). 95% CIs was also calculated. A value of P < .05 was considered statistically significant. CI = confidence interval.
4. Discussion
By now, various devices have been used to secure vessels including intra-corporal knot-tying,[23,24] bipolar vascular sealing devices,[25,26] Ligasure,[27] and Harmonic Scalpel[28–30] during LDN. However, those devices are only recommended for ligating the tributaries of renal vessels.[31,32] Staplers and clips are 2 main devices used for controlling the renal pedicle. At the beginning of staplers application, experts thought staples would safer due to transfixion of vessel wall. However, Chan et al[5] reviewed 565 laparoscopic nephrectomies performed with endovascular staplers and found that device malfunction, which was defined as failure to meet its performance expectations, occurred in 10 patients (1.7%) in 2000. In addition, other studies also reported that malfunction rate of staplers ranged from 0.2% to 1.1%.[7,8] Moreover, staplers shortened the length of graft vessel and increased operative cost.[17,20,33,34] Even so, 30% surgeons still preferred to use Endo-GIA staplers.[11] Later on, clips were applied during the operation which theoretically compensated for the deficiency of staplers. However, several reports showing the death incidence[15,19] associated with clip slippage have led to the contraindication of clips for LDN. It is unknown which device has lower rate for malfunction and complication.
Our results showed that the death rate is not significantly different between two groups. In addition, several death incidents were not related to the devices. One donor died in Hsi study due to the rupture of the artery which is proximal to clips at junction with aorta while both clips were still in place. Another donor death in Siqueira study was caused by an unrecognized large bowel perforation and peritonitis, which was not related to stapler device. Several studies have suggested that staplers might be safer than clips since staples transfix the vessel wall without slipping while titanium clips are more likely to slip[35,36] owing to the nature of the clips that cannot transfix the vessel wall. Then, Hem-o-Lok clip was introduced to the field. It has teeth on jaws and can lock the mechanism at tips to reduce inadvertent dislodgment. Besides, there is a small space between the 2 jaws[37] which may be more secure for those wider vessels. These improvements had led to less failure of clips, which was consistent with our results that Hem-o-Lok clips and staplers were not statistically different regarding to the device failure rate and the death rate. Notably, a potential malposition or malfunction of clips could be further avoided and controlled because the vessels are generally skeletonized. When using staplers, vessels isolation is somehow less refined, leading to a more difficult control in case of malfunction. In addition, stapling close to the aorta is theoretically riskier in case of uneventful malfunctioning since it is a single shot.
There is dramatic difference between the 2 types of devices in terms of medical cost. Our results demonstrated that clips were much cheaper than staplers. Using clips would save an average of 400 dollars for each patient which would lead to greater benefits for patients in developing countries.[38] However, it should be pointed out that we didn’t make forest plot analysis for medical cost due to the lack of standard deviation data from original studies. In order to support our conclusion, we reviewed many studies and found their results were consistent with ours.[10,20–22,33,39–41]
Sufficient vessel length is important for renal transplantation as well. Meta-analysis result showed that Hem-o-Lok clips could provide longer length of vessels. Endo-GIA stapler ligates and transects vessels simultaneously,[34] leaving 3 rows of staggered staples on each side. The trimming step after the stapling would subsequently lead to the loss of vascular length. Surgeons have tried Endo-TA stapler to make up this disadvantage of staplers. Endo-TA stapler shares the same mechanism with Endo-GIA stapler with the difference that only place 3 staple rows on the donor side.[34] Sundaram et al[32] found Endo-TA stapler and Hem-o-Lok clips provided longer vessel length compared with Endo-GIA stapler. Although Meng et al[34] reported that residual vessels of Endo-GIA stapler were adequate for the subsequent anastomosis and the mean creatinine level was 1.6 mg/dL 45 weeks after surgery, longer vessels may still play a critical role in certain situations including right-sided donor nephrectomies, early vascular bifurcation, and rare vascular variation. If 2 renal arteries or veins occurred during the operation, the shorter length of vessels resulted from staplers would make anastomosis more complex and prolong the revascularization time and subsequently affect the graft function,[42] sometimes even resulting in graft loss. Therefore, in terms of health care expenditure and the operation convenience, clips obtained more significant advantage than staplers in LDN.
During renal transplantation, WIT is important and affects the graft function. The reason why clips increased WIT is that 2 more clips were applied to control vessels and it took more time for scrub nurse to reload clips.[33] However, this problem can be resolved by using 2 clip applicators.[18,33] In addition, Van and Simforoosh found slightly longer WIT (no longer than 14 minutes) had no significant effect on graft function regarding the serum creatinine levels postoperatively. This concept also got support from several other studies.[9,17,22,34,42–44]
Some surgeons prefer to use stapler owing to the allowance of staplers to en bloc ligation of renal hilum even with the theoretical risks of arteriovenous fistula (AVF)[45] which is associated with flank pain, cardiomegaly, and cardiac failure with high output. However, several retrospective studies[46,47] had confirmed that en bloc ligation was safe with no AVF postoperatively. Hemal and Mishra[48] preferred using Endo-GIA stapler when resecting pyonephrotic nonfunctioning kidneys to avoid high risk of adhesion, inadequate space, and inadvertent injury to surrounding tissue.
Some surgeons thought higher blood pressure might contribute to the dislodgement of clips. The leak-point pressure in cases of clips usage, ranging from 300 to 1800 mmHg, was much higher than that of staplers’. Several in vitro studies[30,41,45,49,50] showed that Hem-o-Lok clips and titanium clips did not slip or lead to leakage from the end of vessel cuff under the physiological pressures. In contrast, Joseph et al[51] found 4 of the 8 vascular staple lines leaked when the patients’ pressure reached 273 mmHg (237–322 mmHg) which was higher than the upper limit of normal physiological range of blood pressure. It indicated that clips may be safer than staplers for patients with supra-physiologic pressures. More in vivo and clinical studies are needed to verify this hypothesis since sometimes in vitro studies are not necessarily reflect in vivo situation.
In addition, Hem-o-Lok clips do not interfere magnetic resonance imaging (MRI) and computed tomography (CT), while metal clips are contraindicated for MRI and can lead to artifacts in CT image.
To improve the quality of our analysis and solidify the conclusion, more studies should be done. First, more data source for meta-analysis are needed. In the current study, data were extracted from only 8 studies and data about several parameters, such as death rate, estimated amount of blood loss, reoperation rate, length of operative time were extracted from only 2 out of the 8 studies. Second, all 8 studies included were retrospective studies, which may comprise some reporting bias. Third, though devices were assigned to 2 groups, staplers stapler (Endo-GIA or Endo-TA) and clips (Hem-o-Lok or titanium clip), the clips or staplers used are not exactly same. They were made by different manufactures with differences. This makes comparation between 2 devices difficult and significantly increasing the heterogenicity. Finally, several online surveys[11,12,52] revealed most of the specialists had experienced either clip slippage or stapler malfunction during surgery.
5. Conclusions:
To summarize, our meta-analysis demonstrated that Hem-o-Lok clips and staplers have the similar function in LDN renal ligation, regarding the device failure rate, death rate, and severe hemorrhage rate. However, the surgeons would benefit from the clips in terms of the residue length of vessels, these outstanding features provide operation convenience and flexibility, such as right-sided donor nephrectomies, early vascular bifurcation, and rare vascular variation. In addition, the clips have potential economic advantages. In some developing countries, it would reduce the healthcare expenditure.
Author contributions
Conceptualization: Yu Liu, Zhongli Huang, Banghua Liao, Kunjie Wang, Hong Li.
Critical revision of the manuscript: Kunjie Wang, Hong Li.
Data acquisition: Yu Liu, Zhongli Huang.
Data analysis: Yu Liu, Zhongli Huang, Yuntian Chen, Banghua Liao, Deyi Luo, Xiaoshuai Gao.
Data curation: Yu Liu, Zhongli Huang, Deyi Luo.
Drafting of the manuscript: Yu Liu.
Formal analysis: Yu Liu, Zhongli Huang, Xiaoshuai Gao.
Funding acquisition: Banghua Liao.
Investigation: Xiaoshuai Gao.
Methodology: Yu Liu, Zhongli Huang, Yuntian Chen, Banghua Liao, Xiaoshuai Gao.
Project administration: Yu Liu, Deyi Luo, Kunjie Wang.
Resources: Yu Liu, Zhongli Huang.
Software: Zhongli Huang, Yuntian Chen, Deyi Luo, Xiaoshuai Gao.
Study design: Yu Liu, Zhongli Huang, Yuntian Chen, Kunjie Wang, Hong Li.
Supervision: Yu Liu, Yuntian Chen, Banghua Liao, Deyi Luo, Kunjie Wang, Hong Li.
Validation: Zhongli Huang, Yuntian Chen, Banghua Liao, Deyi Luo, Kunjie Wang, Hong Li.
Visualization: Deyi Luo, Hong Li.
Writing – original draft: Yu Liu.
Writing – review & editing: Banghua Liao, Deyi Luo, Kunjie Wang, Hong Li.
Footnotes
Abbreviations: CI = confidence interval, LDN = live donor nephrectomy, MD = mean differences, PRISMA = the Preferred Reporting Items for Systematic Reviews and Meta-Analyses, RR = risk ratio, WIT = warm ischemia time.
YL, ZH, and YC have contributed equally to this work and should be considered as co-first authors.
Ethical approval: The meta-analysis was based on published retrospective studies. So, for this type of study ethical approval and formal consent are not required.
This study was funded by The National Natural Science Fund of China (81770703) and 1.3.5 project for disciplines of excellence, West China Hospital, Sichuan University (ZY2016104).
The authors declare no competing interests.
References
- [1].Clayman RV, Kavoussi LR, Figenshau RS, et al. Laparoscopic nephroureterectomy: initial clinical case report. J Laparoendosc Surg 1991;1:343–9. [DOI] [PubMed] [Google Scholar]
- [2].Ratner LE, Ciseck LJ, Moore RG, et al. Laparoscopic live donor nephrectomy. Transplantation 1995;60:1047–9. [PubMed] [Google Scholar]
- [3].Devra AK, Patel S, Shah SA. Laparoscopic right donor nephrectomy: endo TA stapler is safe and effective. Saudi J Kidney Dis Transpl 2010;21:421–5. [PubMed] [Google Scholar]
- [4].Ko EY, Castle EP, Desai PJ, et al. Utility of the endovascular stapler for right-sided laparoscopic donor nephrectomy: a 7-year experience at Mayo Clinic. J Am Coll Surg 2008;207:896–903. [DOI] [PubMed] [Google Scholar]
- [5].Chan D, Bishoff JT, Ratner L, et al. Endovascular gastrointestinal stapler device malfunction during laparoscopic nephrectomy: early recognition and management. J Urol 2000;164:319–21. [PubMed] [Google Scholar]
- [6].Kwazneski D, 2nd, Six C, Stahlfeld K. The unacknowledged incidence of laparoscopic stapler malfunction. Surg Endosc 2013;27:86–9. [DOI] [PubMed] [Google Scholar]
- [7].Breda A, Veale J, Liao J, et al. Complications of laparoscopic living donor nephrectomy and their management: the UCLA experience. Urology 2007;69:49–52. [DOI] [PubMed] [Google Scholar]
- [8].Deng DY, Meng MV, Nguyen HT, et al. Laparoscopic linear cutting stapler failure. Urology 2002;60:415–9. discussion 419-420. [DOI] [PubMed] [Google Scholar]
- [9].Liu KL, Chiang YJ, Wang HH, et al. Techniques of vascular control in laparoscopic donor nephrectomy. Transplant Proc 2008;40:2342–4. [DOI] [PubMed] [Google Scholar]
- [10].Jellison FC, Shah SK, Mashni JW, Jr, et al. Vessel length following laparoscopic donor nephrectomy: impact of vascular ligation technique on allograft vessel length. J Endourol 2008;22:973–7. [DOI] [PubMed] [Google Scholar]
- [11].Friedman AL, Peters TG, Jones KW, et al. F at al and nonfatal hemorrhagic complications of living kidney donation. Ann Surg 2006;243:126–30. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [12].Janki S, Verver D, Klop KW, et al. Vascular management during live donor nephrectomy: an online survey among transplant surgeons. Am J Transplant 2015;15:1701–7. [DOI] [PubMed] [Google Scholar]
- [13].Dekel Y, Mor E. Hem-o-lok clip dislodgment causing death of the donor after laparoscopic living donor nephrectomy. Transplantation 2008;86:887. [DOI] [PubMed] [Google Scholar]
- [14].Meng MV. Reported failures of the polymer self-locking (Hem-o-lok) clip: review of data from the Food and Drug Administration. J Endourol 2006;20:1054–7. [DOI] [PubMed] [Google Scholar]
- [15].Hsi RS, Ojogho ON, Baldwin DD. Analysis of techniques to secure the renal hilum during laparoscopic donor nephrectomy: review of the FDA database. Urology 2009;74:142–7. [DOI] [PubMed] [Google Scholar]
- [16].Wells G, Shea B, O’Connell J. The Newcastle-Ottawa Scale (NOS) for Assessing the Quality of Nonrandomised Studies in Meta-analyses, Vol 7; 2014. [Google Scholar]
- [17].Bittner JG, 4th, Sajadi K, Brown JA. Comparison of renal artery occlusion techniques in hand-assisted laparoscopic living donor nephrectomy. J Endourol 2009;23:933–7. [DOI] [PubMed] [Google Scholar]
- [18].Chueh SC, Wang SM, Lai MK. Use of Hem-o-lok clips effectively lengthens renal vein during laparoscopic live donor nephrectomy. Transplant Proc 2004;36:2623–4. [DOI] [PubMed] [Google Scholar]
- [19].Goh YS, Cheong PS, Lata R, et al. A necessary step toward kidney donor safety: the transition from locking polymer clips to transfixion techniques in laparoscopic donor nephrectomy. Transplant Proc 2014;46:310–3. [DOI] [PubMed] [Google Scholar]
- [20].Kaushik M, Bagul A, Yates PJ, et al. Comparison of techniques of vascular control in laparoscopic donor nephrectomy: the leicester experience. Transplant Proc 2006;38:3406–8. [DOI] [PubMed] [Google Scholar]
- [21].Izaki H, Fukumori T, Takahashi M, et al. Clinical research of renal vein control using Hem-o-lok clips in laparoscopic nephrectomy. Int J Urol 2006;13:1147–9. [DOI] [PubMed] [Google Scholar]
- [22].Siqueira TM, Jr, Mitre AI, Simoes FA, et al. A cost-effective technique for pure laparoscopic live donor nephrectomy. Int Braz J Urol 2006;32:23–8. discussion 28-30. [DOI] [PubMed] [Google Scholar]
- [23].Chiu AW, Wu ST, Sun GH, et al. Novel intracorporeal vascular knot applier for large vessel ligation in laparoscopic surgery: experience in pigs and preliminary clinical use. J Endourol 2003;17:73–8. [DOI] [PubMed] [Google Scholar]
- [24].Chon CH, Chung SY, Ng CS, et al. Looped silk tie: surgical technique for management of the renal vein during laparoscopic live-donor nephrectomy. J Endourol 2005;19:401–5. [DOI] [PubMed] [Google Scholar]
- [25].Heniford BT, Matthews BD, Sing RF, et al. Initial results with an electrothermal bipolar vessel sealer. Surg Endosc 2001;15:799–801. [DOI] [PubMed] [Google Scholar]
- [26].Sengupta S, Webb DR. Use of a computer-controlled bipolar diathermy system in radical prostatectomies and other open urological surgery. ANZ J Surg 2001;71:538–40. [DOI] [PubMed] [Google Scholar]
- [27].Constant DL, Florman SS, Mendez F, et al. Use of the LigaSure vessel sealing device in laparoscopic living-donor nephrectomy. Transplantation 2004;78:1661–4. [DOI] [PubMed] [Google Scholar]
- [28].Landman J, Kerbl K, Rehman J, et al. Evaluation of a vessel sealing system, bipolar electrosurgery, harmonic scalpel, titanium clips, endoscopic gastrointestinal anastomosis vascular staples and sutures for arterial and venous ligation in a porcine model. J Urol 2003;169:697–700. [DOI] [PubMed] [Google Scholar]
- [29].Helal M, Albertini J, Lockhart J, et al. Laparoscopic nephrectomy using the harmonic scalpel. J Endourol 1997;11:267–8. [DOI] [PubMed] [Google Scholar]
- [30].Siperstein AE, Berber E, Morkoyun E. The use of the harmonic scalpel vs conventional knot tying for vessel ligation in thyroid surgery. Arch Surg 2002;137:137–42. [DOI] [PubMed] [Google Scholar]
- [31].Kennedy JS, Stranahan PL, Taylor KD, et al. High-burst-strength, feedback-controlled bipolar vessel sealing. Surg Endosc 1998;12:876–8. [DOI] [PubMed] [Google Scholar]
- [32].Sundaram CP, Bargman V, Bernie JE. Methods of vascular control during laparoscopic donor nephrectomy. J Endourol 2006;20:467–9. discussion 469-470. [DOI] [PubMed] [Google Scholar]
- [33].Bernie JE, Sundaram CP, Guise AI. Laparoscopic vascular control techniques in donor nephrectomy: effects on vessel length. JSLS 2006;10:141–4. [PMC free article] [PubMed] [Google Scholar]
- [34].Meng MV, Freise CE, Kang SM, et al. Techniques to optimize vascular control during laparoscopic donor nephrectomy. Urology 2003;61:93–7. discussion 97-98. [DOI] [PubMed] [Google Scholar]
- [35].Baek M, Chun H, Oh SJ, et al. Open conversion from laparoscopic nephrectomy: slippage of surgical clips ligating the renal artery affected by atherosclerosis. J Urol 2004;171:333–4. [DOI] [PubMed] [Google Scholar]
- [36].Maartense S, Heintjes RJ, Idu M, et al. Renal artery clip dislodgement during hand-assisted laparoscopic living donor nephrectomy. Surg Endosc 2003;17:1851. [DOI] [PubMed] [Google Scholar]
- [37].Chan DY, Su LM, Kavoussi LR. Rapid ligation of renal hilum during transperitoneal laparoscopic nephrectomy. Urology 2001;57:360–2. [DOI] [PubMed] [Google Scholar]
- [38].Kumar A, Chaudhary H, Srivastava A, et al. Laparoscopic live-donor nephrectomy: modifications for developing nations. BJU Int 2004;93:1291–5. [DOI] [PubMed] [Google Scholar]
- [39].Sureka SK, Gaur P, Srivastava A, et al. Safe vascular control in laparoscopic donor nephrectomy: an experience of more than 1400 cases. Ind J Transplant 2016;10:111. [Google Scholar]
- [40].Chibber PJ, Shah HN. Are titanium clips for control of the renal hilar vessels as unsafe as generally presumed? Surg Laparosc Endosc Percutan Tech 2006;16:276–80. [DOI] [PubMed] [Google Scholar]
- [41].Jellison FC, Baldwin DD, Berger KA, et al. Comparison of nonabsorbable polymer ligating and standard titanium clips with and without a vascular cuff. J Endourol 2005;19:889–93. [DOI] [PubMed] [Google Scholar]
- [42].Abreu SC, Goldfarb DA, Derweesh I, et al. Factors related to delayed graft function after laparoscopic live donor nephrectomy. J Urol 2004;171:52–7. [DOI] [PubMed] [Google Scholar]
- [43].Van der Merwe A, Heyns CF. Retroperitoneoscopic live donor nephrectomy: review of the first 50 cases at Tygerberg Hospital, Cape Town, South Africa. S Afr J Surg 2014;52:53–6. [DOI] [PubMed] [Google Scholar]
- [44].Simforoosh N, Basiri A, Shakhssalim N, et al. Effect of warm ischemia on graft outcome in laparoscopic donor nephrectomy. J Endourol 2006;20:895–8. [DOI] [PubMed] [Google Scholar]
- [45].Kerbl K, Chandhoke PS, Clayman RV, et al. Ligation of the renal pedicle during laparoscopic nephrectomy: a comparison of staples, clips, and sutures. J Laparoendosc Surg 1993;3:9–12. [DOI] [PubMed] [Google Scholar]
- [46].Schatloff O, Ramon J, Lindner U, et al. Is postoperative arteriovenous fistula still a concern after en bloc stapling of the renal hilum during laparoscopic nephrectomy? J Endourol 2009;23:639–43. [DOI] [PubMed] [Google Scholar]
- [47].Resorlu B, Oguz U, Polat F, et al. Comparative analysis of pedicular vascular control techniques during laparoscopic nephrectomy: en bloc stapling or separate ligation? Urol Int 2015;94:79–82. [DOI] [PubMed] [Google Scholar]
- [48].Hemal AK, Mishra S. Retroperitoneoscopic nephrectomy for pyonephrotic nonfunctioning kidney. Urology 2010;75:585–8. [DOI] [PubMed] [Google Scholar]
- [49].Elliott SP, Joel AB, Meng MV, et al. Bursting strength with various methods of renal artery ligation and potential mechanisms of failure. J Endourol 2005;19:307–11. [DOI] [PubMed] [Google Scholar]
- [50].Papaioannou T, Daykhovsky L, Grundfest WS. Safety evaluation of laparoscopically applied clips. J Laparoendosc Surg 1996;6:99–107. [DOI] [PubMed] [Google Scholar]
- [51].Joseph J, Leung YY, Eichel L, et al. Comparison of the Ti-knot device and Hem-o-lok clips with other devices commonly used for laparoscopic renal-artery ligation. J Endourol 2004;18:163–6. [DOI] [PubMed] [Google Scholar]
- [52].McGregor TB, Patel P, Sener A, et al. Vascular control during laparoscopic kidney donation. Can J Surg 2017;60:150–1. [DOI] [PMC free article] [PubMed] [Google Scholar]