Effect of metal-containing topical agents on surface doses received during external irradiation

Ayumi Iyama 1,✉, Tomohiko Matsuyama 2, Eriko Matsumoto 3, Takafumi Araki 3, Satoshi Inokuchi 3, Mizuki Yamashita 3, Noritoshi Honda 4, Taiga Miyake 4, Tetsuo Saito 2, Ryo Toya 2, Yudai Kai 2, Yasuyuki Yamashita 1, Natsuo Oya 2
PMCID: PMC6251421  PMID: 30247674

Abstract

The ability of topical metal-containing agents (MCAs) to enhance radiation dermatitis remains controversial. In the present study, we evaluated increases in surface doses associated with topical agents at different application thicknesses and with MCAs versus non-metal containing agents (NMCAs). We assessed two clinically available MCAs, zinc oxide ointment (ZOO) and silver sulfadiazine cream (SSDC), and eight NMCAs. Surface doses were measured using a Markus chamber placed on a polystyrene phantom. To evaluate the role of application thickness, each agent was applied to the chamber in oil-slick (<0.1-mm), 1-mm and 5-mm layers prior to irradiation of a 10 × 10 cm field with 4-, 6- and 10-MV X-ray beams. The surface dose enhancement ratio (SDER) was calculated as the ratio of the surface dose with an agent to the dose without an agent. The SDER values for the eight NMCAs, ZOO and SSDC at an oil-slick thickness were 101.6–104.6% (mean: 103.3%), 104.5% and 105.0%, respectively, using a 6-MV X-ray beam. The corresponding values at a 1-mm thickness were 196.8–237.8% (mean: 215.7%), 229.3% and 201.4%, respectively, and those at a 5-mm thickness were 342.2–382.4% (mean: 357.9%), 357.1% and 352.6%, respectively. A similar tendency was found using 4- and 10-MV X-ray beams. The lack of a significant difference in surface dose enhancement between MCAs and NMCAs, particularly when applied in oil-slick layers, suggests that MCAs do not need to be avoided or applied in a restricted manner during radiotherapy for dosimetric reasons.

Keywords: radiation dermatitis, topical agent, surface dose, zinc oxide, silver sulfadiazine

INTRODUCTION

Radiation dermatitis is one of the most common side effects of radiation therapy. Despite the publication of clinical guidelines for skin care management during radiation therapy [1–3], sufficient evidence of the ability of the recommended measures to effectively prevent and manage radiodermatitis has not yet been provided. Recent progress in radiation therapy technology has led to a decreased incidence of severe radiation dermatitis. However, some patients still develop dermatitis with sores and bleeding when radiotherapy is administered to the head and neck, vulva, and/or in combination with chemotherapy or molecular targeted therapy, and these patients occasionally find it difficult to continue therapy.

Clinicians have often been reluctant to use metal-containing agents (MCAs) during radiotherapy, given the risk of interaction of these agents with radiation [4–6]. To our knowledge, however, no clinical trial has examined whether the use of MCAs worsens dermatitis, and some reports have even suggested the usefulness of MCAs [e.g. silver sulfadiazine cream (SSDC) [1, 3, 7, 8] and zinc oxide cream (ZOC) [8]] for radiation dermatitis. Therefore, an investigation of the dermal effects of MCAs during radiotherapy is needed.

Two previous dosimetry studies evaluated the potential effects of MCAs on surface radiation doses [9, 10]. First, Burch et al. reported no significant differences in surface doses associated with normally applied MCAs and non-metal-containing agents (NMCAs) during irradiation with a 6-MV X-ray beam [9]. Second, Fackrell et al. compared the dose effects of two MCAs (SSDC and ZOC) and one NMCA [aqueous cream (AQC)] during irradiation with a 6-MV X-ray beam, and reported that a slightly lower dose was required with ZOC than with AQC or SSDC [10]. However, neither study strictly addressed the possibility that slight differences in application thickness might affect the surface dose in the build-up region [9, 10]. Furthermore, Burch et al. evaluated clinically impractical topical agents [9]. Conclusions regarding the use of MCAs during radiotherapy in clinical practice should be made using dosimetric data associated with clinically available agents and a fixed application thickness. Accordingly, the present study aimed to examine specifically whether the presence of a metallic element would affect the surface dose, using clinically available MCAs and NMCAs at consistent application thicknesses.

MATERIALS AND METHODS

We measured the surface doses associated with 10 clinically available topical agents (i.e. creams and ointments), including two MCAs [zinc oxide ointment (ZOO) and SSDC] and eight NMCAs. ZOO contains 20% (200 mg/g) zinc oxide, and SSDC contains 1% (10 mg/g) silver sulfadiazine. We note that one of the NMCAs, a mixture of sugar and povidone-iodine, contains 3% (30 mg/g) povidone-iodine, which, although non-metallic, has a relatively high atomic number. The physical properties of all agents are listed in Table 1.

Table 1.

Comparison of the physical properties of 10 topical agents

No. Generic name Active ingredient (mass percentage) Molecular formula Brand name Specific gravity
1 White petrolatum White petrolatum (100%) C15H32 to C20H42 PROPETO® 0.815–0.880
2 Dimethyl isopropyl azulene ointment Dimethyl isopropylazulene (0.033%) C15H18 Azunol Ointment 0.033% 0.870
3 Betamethasone valerate ointment Betamethasone valerate (0.12%) C27H37FO6 Rinderon®-V 0.870
4 Lysozyme hydrochloride ointment Lysozyme hydrochloride (5%) C616H963N193O182S10·xHCL REFLAP® Ointment 5% 0.940
5 Heparinoid ointment Heparinoid (0.3%) Not applicable Hirudoid® Soft Ointment 1.000
6 Diphenhydramine cream Diphenhydramine (1%) C17H21NO RESTAMIN KOWA CREAM 1% 1.013–1.020
7 Macrogol ointment Macrogol 400 (50%) HOCH2(CH2OCH2)nCH2OHb SOLBASE ® 1.110–1.140
Macrogol 4000 (50%) n = 7–9 n = 59–84
8 Mixture of sugar and povidone-iodine Povidone-iodine (3%) (C6H9NO)n·xI U-PASTA KOWA Ointment 1.270
Sucrose (70%) C12H22O11
9 Zinc oxide ointmenta Zinc oxide (20%) ZnO Zinc Oxide Ointment ‘Hoei’ 1.064
10 Silver sulfadiazine creama Sulfadiazine silver (1%) C10H9AgN4O2S GEBEN cream 1% 0.980–0.990

aZinc oxide ointment (No. 9) and silver sulfadiazine cream (No. 10) contain metallic elements. The other agents do not contain metallic elements.bn ranging from 7 to 9 in Macrogol 400, and from 59 to 84 in Macrogol 4000.

Surface doses were measured using a Markus-type parallel plate ionization chamber (PTW, Freiburg, Germany) placed on a polystyrene phantom and an electrometer (RAMTEC Smart, Toyo Medic Co., Ltd, Tokyo, Japan) (Fig. 1). Although a previous report noted that this chamber overestimated surface doses in build-up regions [11], we did not correct for this effect because the comparisons made in this study were relative. For the experiment, each agent was applied to a thin polyethylene sheet (<0.1 mm) positioned over the chamber window, at the following thicknesses: oil-slick thickness, which is <0.1 mm, using the finger-tip unit [12]; 1 mm, which was selected as a realistic amount to apply using medical gauze; and 5 mm, which is rarely used in practice but was selected for dosimetric comparison. For the latter two thicknesses, uniform layers were created by filling an 11 × 11 × 0.1 cm or 11 × 11 × 0.5 cm well, respectively, in a polystyrene frame and smoothing the surface as much as possible (Fig. 2).

Fig. 1.

Fig. 1.

A Markus chamber set in a solid water phantom.

Fig. 2.

Fig. 2.

Specific frames were used to apply 1-mm thick and 5-mm thick layers to a thin polyethylene sheet.

The chamber window surface was set at a source-to-surface distance of 100 cm (Fig. 3). Each agent was irradiated with 100 monitor units under 4-, 6- and 10-MV photon beams in a field size of 10 × 10 cm, using an Elekta Synergy linear accelerator (Elekta Instrument AB, Stockholm, Sweden). Surface doses were measured using the protocol published by Task Group 51 of the Radiation Therapy Committee of the American Association of Physicists in Medicine [13]. Each agent was subjected to irradiation three times. The uncertainty of the three measurements was negligible (i.e. the coefficient of variation was ≤0.40%), and the mean value of the surface doses of three measurements was used for the calculations. We defined the surface dose enhancement ratio (SDER) for each topical agent at each thickness as follows: (surface dose with an agent)/(surface dose without an agent) × 100 (%).

Fig. 3.

Fig. 3.

Diagram of a phantom with a Markus-type chamber. SSD, source-to-surface distance.

RESULTS

Figure 4 and Table 2 present the SDER associated with each topical agent. When using a 6-MV X-ray beam, the SDER values for the eight NMCAs, ZOO and SSDC were 101.6–104.6% (mean: 103.3%), 104.5% and 105.0%, respectively, at an oil-slick thickness. The corresponding values were 196.8–237.8% (mean: 215.7%), 229.3% and 201.4%, respectively, at a 1-mm thickness and 342.2–382.4% (mean: 357.9%), 357.1% and 352.6%, respectively, at a 5-mm thickness. A similar tendency was found when using 4- and 10-MV X-ray beams (Fig. 4, Table 2).

Fig. 4.

Fig. 4.

The surface dose enhancement ratio (SDER), or the ratio of the surface dose with an agent to the dose without an agent (i.e. control) was calculated for each agent at oil-slick, 1-mm and 5-mm thicknesses using 4- (a), 6- (b), and 10-MV (c) photon beams. The surface dose without an agent (i.e. only a polyethylene sheet placed over the chamber) was used as the control. The x-axis indicates the 10 topical agents by number (see Table 2 for further details).

Table 2.

The surface dose enhancement ratio (SDER) for 10 topical agents at the indicated thicknesses

Topical agents SDER (%)
Oil-slick thickness 1-mm thickness 5-mm thickness
4 MV 6 MV 10 MV 4 MV 6 MV 10 MV 4 MV 6 MV 10 MV
NMCAsa
 1. White petrolatum 103.4 103.8 103.2 203.6 209.9 207.9 306.6 347.2 379.9
 2. Dimethyl isopropyl azulene ointment 104.3 104.1 103.3 192.2 196.8 193.4 304.4 347.3 378.0
 3. Betamethasone valerate ointment 102.7 102.4 101.9 193.4 197.9 194.1 306.3 349.7 381.7
 4. Lysozyme hydrochloride ointment 102.9 102.8 102.3 208.6 216.3 214.8 310.5 359.2 397.6
 5. Heparinoid ointment 104.8 104.6 103.4 205.8 213.7 212.8 297.9 342.2 376.3
 6. Diphenhydramine cream 102.7 102.6 102.7 218.6 228.9 230.2 313.7 362.4 405.6
 7. Macrogol ointment 103.7 103.5 103.0 221.1 232.8 234.0 322.0 382.4 436.5
 8. Mixture of sugar and povidone-iodine 101.7 101.6 100.8 225.4 237.8 240.6 316.3 375.0 428.5
MCAsa
 9. Zinc oxide ointment 104.2 104.5 104.3 217.5 229.3 234.2 303.8 357.1 411.0
 10. Silver sulfadiazine cream 104.9 105.0 103.9 195.4 201.4 199.1 304.3 352.6 391.1

aMCA, metal containing agent, NMCA, non-metal containing agent.

DISCUSSION

We observed no significant differences in surface dose enhancement between MCAs and NMCAs at any application thickness, but particularly at an oil-slick thickness (<0.1 mm) using 4-, 6- and 10-MV X-ray beams. We note that this result is highly significant, as an oil-slick thickness is generally applied in clinical practice, using the finger-tip unit [12]. We further note that although we observed slight variations in surface doses at 1-mm and 5-mm thicknesses, these variations did not always correspond to higher values with MCAs. The values obtained with SSDC, a cream containing only 1% metallic component (sulfadiazine silver), were nearly the same as or lower than those obtained with most NMCAs. Furthermore, although ZOO has a relatively high metallic concentration (20% zinc oxide), NMCAs with higher specific gravities [macrogol ointment (No. 7) and mixture of sugar and povidone-iodine (No. 8)] yielded higher surface dose increases. This correlation of dose changes with specific gravity is reasonable, as Compton scattering is the main reciprocal action associated with the high-energy X-rays used for radiotherapy [14]. Although clinicians have generally tended to restrict the use of MCAs during radiotherapy to avoid possible interactions with radiation, our results suggest that, physically, this caution is unwarranted.

The minor effects of clinically available MCAs on dosimetry may be attributable to the relatively low concentrations of metallic elements in these agents. It is generally recognized that there is a complex relationship of bolus effect, scattering, and absorption with surface dose changes associated with topical agents. A previous study of dental materials suggested that the effects of scattering and absorption during high-energy X-ray irradiation increased with the inclusion of higher density or higher atomic number substances in the irradiation field [15]. However, topical agents contain much lower concentrations of metallic elements, compared with dental materials, leading us to conclude that these agents had little effect on scattering and absorption in our study. In addition, our study suggested that the application thickness could cause a bolus effect and would have a stronger influence on the surface dose than the type of agent. However, we should note that a previous Markus-type chamber-based analysis overestimated surface doses in build-up lesions [11].

Although previous studies have also evaluated the influences of MCAs on surface doses [9, 10], we note the significance of our study, which used only clinically available topical agents and strictly determined application thicknesses. In a previous analysis, Burch et al. investigated surface dose changes associated with 15 products, both with and without high atomic number components [9], and did not detect any large increases in surface dose with normal product application. Although this result is similar to our findings, the earlier study evaluated only deodorants, powders and lotions, rather than practical topical agents (i.e. creams and ointments), and the products were not applied in specific thicknesses. In a second study, Fackrell et al. compared the dose effects of two MCAs (SSDC and ZOC) and one NMCA (AQC) [10], and observed no significant differences in dose effect between SSDC and AQC, and a slight decrease in dose when using ZOC. Although the authors used identical weights of the three agents, the agents were not applied at identical thicknesses because the authors did not consider specific gravities. Therefore, their measurements may have been insufficiently accurate, as a slight difference in thickness might affect the surface dose in a build-up region. We conclude that, when measuring surface doses, it is important to strictly determine the application thickness and thus exclude differences due to the bolus effect while evaluating the influences of the components in various agents.

Our results may increase the clinical selection of topical agents available for patients with radiation dermatitis. Both ZOO [16] and SSDC [17] are known to be beneficial for dermatitis therapy, and, in particular, zinc can promote the healing of skin and mucous membrane wounds [18]. Accordingly, ZOO is very suitable for mild dermatitis, and its early use may reduce and even prevent radiation dermatitis. Although a previous report described the use of ZOO as a treatment for Grade 2 and 3 radiation dermatitis in patients with squamous cell carcinoma of the head and neck [8], no randomized controlled trial (RCT) has demonstrated the utility of ZOO for this purpose, possibly because of concerns regarding interactions of the metallic components of ZOO with radiation. However, our study found no dosimetric reason to restrict the use of ZOO during radiotherapy, and we consider this agent to be useful in this context. Future clinical studies regarding the use of ZOO for radiation dermatitis are expected in the future. We note that several reports have also described SSDC as a useful therapeutic agent for radiation dermatitis [1, 3, 7, 8].

We must note two limitations of our study. First, we only evaluated dose changes in the vertical beam, and not in the oblique or tangent beam. However, a recent Monte Carlo simulation-based study addressed this limitation [19], and showed that a product thickness of <0.7 mm was not likely to cause a clinically meaningful dose increase in a 10 × 10 cm field across a range of gantry angles (0°–80°). No significant bolus effects associated with increased skin doses were observed in that study, particularly at a thickness of <0.1 mm (i.e. ‘oil-slick thickness’ in our study) [19]. Considering these results and our results, we suggest that no topical agent would be problematic when applied clinically at an oil-slick thickness. Second, we did not evaluate actual patients, and therefore could not consider factors such as the effect of skin temperature on agent properties, or the skin permeability, or the pharmacological effects of each agent. However, several previous RCTs demonstrated that the use of aluminum-containing antiperspirants or deodorants did not affect skin reactions during adjuvant breast radiotherapy [20, 21].

In conclusion, we observed no significant differences in surface dose enhancement between MCAs and NMCAs at any thickness, especially at the oil-slick thickness used clinically. Therefore, dosimetric restrictions on the use of MCAs during radiotherapy may be unnecessary. Although an RCT of clinically available topical agents remains to be conducted, our physical study indicated no significant influence of metallic components in topical agents.

CONFLICT OF INTEREST

The authors declare that there are no conflicts of interest.

FUNDING

None.

REFERENCES

  • 1. Wong RK, Bensadoun RJ, Boers-Doets CB et al. Clinical practice guidelines for the prevention and treatment of acute and late radiation reactions from the MASCC Skin Toxicity Study Group. Support Care Cancer 2013;21:2933–48. [DOI] [PubMed] [Google Scholar]
  • 2. Chan RJ, Larsen E, Chan P. Re-examining the evidence in radiation dermatitis management literature: an overview and a critical appraisal of systematic reviews. Int J Radiat Oncol Biol Phys 2012;84:e357–62. [DOI] [PubMed] [Google Scholar]
  • 3. McQuestion M. Evidence-based skin care management in radiation therapy: clinical update. Semin Oncol Nurs 2011;27:e1–17. [DOI] [PubMed] [Google Scholar]
  • 4. Harper JL, Franklin LE, Jenrette JM et al. Skin toxicity during breast irradiation: pathophysiology and management. South Med J 2004;97:989–93. [DOI] [PubMed] [Google Scholar]
  • 5. Salvo N, Barnes E, van Draanen J et al. Prophylaxis and management of acute radiation-induced skin reactions: a systematic review of the literature. Curr Oncol 2010;17:94–112. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Hassey KM, Rose CM. Altered skin integrity in patients receiving radiation therapy. Oncol Nurs Forum 1982;9:44–50. [PubMed] [Google Scholar]
  • 7. Hemati S, Asnaashari O, Sarvizadeh M et al. Topical silver sulfadiazine for the prevention of acute dermatitis during irradiation for breast cancer. Support Care Center 2012;20:1613–8. [DOI] [PubMed] [Google Scholar]
  • 8. Bernier J, Bonner J, Vermorken JB et al. Consensus guidelines for the management of radiation dermatitis and coexisting acne-like rash in patients receiving radiotherapy plus EGFR inhibitors for the treatment of squamous cell carcinoma of the head and neck. Ann Oncol 2008;19:142–9. [DOI] [PubMed] [Google Scholar]
  • 9. Burch SE, Parker SA, Vann AM et al. Measurement of 6-MV X-ray surface dose when topical agents are applied prior to external beam irradiation. Int J Radiat Oncol Phys 1997;38:447–51. [DOI] [PubMed] [Google Scholar]
  • 10. Fackrell D, Kirby D, Sanghera P et al. The effect of silver sulfadiazine and zinc oxide creams on dose distribution during radiotherapy. J Radiother Pract 2015;14:111–6. [Google Scholar]
  • 11. Mellenberg DE., Jr. Determination of build-up region over-response corrections for a Markus-type chamber. Med Phys 1990;17:1041–4. [DOI] [PubMed] [Google Scholar]
  • 12. Long CC, Finlay AY. The finger-tip unit—a new practical measure. Clin Exp Dermatol 1991;16:444–7. [DOI] [PubMed] [Google Scholar]
  • 13. Almond PR, Biggs PJ, Coursey BM et al. AAPM’s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys 1999;26:1847–70. [DOI] [PubMed] [Google Scholar]
  • 14. Franzel W, Gerlach R. The irradiation action on human dental tissue by X-rays and electrons—a nanoindenter study. Z Med Phys 2009;19:5–10. [DOI] [PubMed] [Google Scholar]
  • 15. Reitemeier B, Reitemeier G, Schmidt A et al. Evaluation of a device for attenuation of electron release from dental restorations in a therapeutic radiation field. J Prosthet Dent 2002;87:323–7. [DOI] [PubMed] [Google Scholar]
  • 16. Maitra AK, Dorani B. Role of zinc in post-injury wound healing. Arch Emerg Med 1992;9:122–4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Spann CT, Tutrone WD, Weinberg JM et al. Topical antibacterial agents for wound care: a primer. Dermatol Surg 2003;29:620–6. [DOI] [PubMed] [Google Scholar]
  • 18. Gupta M, Mahajan VK, Mehta KS et al. Zinc therapy in dermatology: a review. Dermatology Res Pract 2014;2014:709152. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Morley L, Cashell A, Sperduti A et al. Evaluating the relevance of dosimetric considerations to patient instructions regarding skin care during radiation therapy. J Radiother Pract 2014;13:294–301. [Google Scholar]
  • 20. Lewis L, Carson S, Bydder S et al. Evaluating the effects of aluminum-containing and non-aluminum containing deodorants on axillary skin toxicity during radiation therapy for breast cancer: a 3-armed randomized controlled trial. Int J Radiat Oncol Biol Phys 2014;90:765–71. [DOI] [PubMed] [Google Scholar]
  • 21. Watson LC, Gies D, Thompson E et al. Randomized control trial: evaluating aluminum-based antiperspirant use, axilla skin toxicity, and reported quality of life in women receiving external beam radiotherapy for treatment of Stage 0, I, and II breast cancer. Int J Radiat Oncol Biol Phys 2012;83:e29–34. [DOI] [PubMed] [Google Scholar]