Abstract
Purposes
To develop and evaluate a Boundary Informed Electrical Properties Tomography (BIEPT) technique for high-resolution imaging of tumor electrical properties (EPs) heterogeneity.
Methods
Tumor EP distributions were inferred from a reference area external to the tumor, as well as internal EP spatial variations derived from a plurality of relative transmit B1 measurements at 7 Tesla. Edge sparsity constraint was enforced to enhance numerical stability. Phantom experiments were performed to determine the imaging accuracy and sensitivity for structures of various EP values, as well as geometrical sizes down to 1.5 mm. Numerical simulation of a realistic rodent model was employed to quantify the algorithm performance in the presence of noise. Eleven athymic rats with human breast cancer xenograft were imaged in vivo, and representative pathological samples were acquired for comparison.
Results
Reconstructed EPs of the phantoms correspond well to the ground truth acquired from dielectric probe measurements, with the smallest structure reliably detectable being 3 mm. EPs heterogeneity inside a tumor is successfully retrieved in both simulated and experimental cases. In vivo tumor imaging results demonstrate similar local features and spatial patterns to anatomical MRI and pathological slides. The imaged conductivity of necrotic tissue is higher than that of viable tissues, consistent with expectation.
Conclusion
BIEPT enables robust detection of tumor EPs heterogeneity with high accuracy and sensitivity to small structures. The retrieved quantitative EPs reflect tumor pathological features, e.g. necrosis. These results provide strong rationale to further expand studies towards pathological conditions where EPs may yield valuable, non-invasive biomarkers.
Keywords: Electrical Properties Tomography (EPT), Ultra-high-field MRI, Tumor heterogeneity, Multi-channel B1 Mapping, Electromagnetic simulation
INTRODUCTION
Electrical properties (EPs), comprised of electrical conductivity and permittivity , are intrinsic properties quantifying a tissue’s ability to conduct, dissipate, and deposit electrical energy. EPs of biological tissues are frequency-dependent, reflecting interactions of the applied electrical field with different constituents from tissue level to molecular level. EPs in the lower frequency band (<1 MHz) reflect tissue macroscopic structures, such as extracellular volume, cellularity, and cellular membrane integrity, while those in the microwave band (~1-100 GHz) correspond to the microscopic level such as overall ion concentration, water and macromolecule content, etc. In between is the radiofrequency (RF) band, in which EPs are influenced by mixed effects of the above factors (1).
Previous studies based on ex vivo dielectric probe measurements have reported deviation of tumor EPs from those of normal tissues across the electromagnetic spectrum, especially in the breast (2,3), setting the ground for EPs as potential biomarkers in cancer diagnosis and characterization. Several non-invasive imaging modalities have been proposed to retrieve tissue EPs across the spectrum, such as electrical impedance tomography (EIT) (4), magnetic resonance electrical impedance tomography (MREIT) (5), magneto-acoustic tomography with magnetic induction (MAT-MI) (6), and microwave imaging (7). Nevertheless, it remains a challenge to retrieve quantitative EP maps at high spatial resolution without injection of electric current, due to limited signal to noise ratio (SNR) of electromagnetic measurement in biological systems, as well as the ill-posed nature of image reconstruction problems derived from Maxwell’s Equations.
Alternatively, EPs can be derived by computational electrodynamics involving B1 measurements obtained from MRI in high spatial resolution. This technique, dubbed magnetic resonance-based electrical properties tomography (MREPT or EPT), exploits the spatial variation of B1 maps while alleviating the need for directly measuring E field (8–11). Compared with previous imaging modalities, MREPT substantially reduces the severity of ill-posedness and therefore represents a promising modality for non-invasive EPs imaging. The feasibility of MREPT was initially demonstrated using birdcage RF coils (12–14) and was more recently further developed using multi-channel RF coil arrays (15–21). Compared to a quadrature coil, a multi-channel coils array addresses the “transceiver phase assumption” that fails at high magnetic field (22). The plurality of unique B1 measurements from the latter also provides abundant information that contributes to improvements in characterization of tissue transition zones.
Conductivity imaging of cancer based on MREPT has been attempted in the human brain, breast, and cervix (23–25), showing results consistent with the literature based on probe measurements. However, the reconstruction methods used in these studies are based on the Helmholtz equation that describes electromagnetic wave propagation in a homogeneous medium. Consequently, tumor heterogeneity cannot be appropriately identified; Instead it contributes to reconstruction artifacts in the resulting EP values and distributions, which can potentially jeopardize quantitative measurements (26). Moreover, permittivity values are intractable from these studies due to the assumptions made about B1 in the reconstruction formalism.
A general approach to accurately image EPs heterogeneity at high spatial resolution, once developed, will provide additional information about the underlying tumor pathology. Potential correlations between this new contrast and tumor hallmarks can be exploited for improved sensitivity and specificity in cancer diagnosis, grading, and monitoring of treatment response. A preliminary framework has been reported by our group for in vivo imaging of rat cancer xenografts, which goes beyond the Helmholtz equation and takes into consideration the spatial variation of EPs in the reconstruction model (27). However, this approach requires collecting data during two consecutive MRI scanning sessions with physically flipped experimental setup executed with high precision, which is prone to operational error and consumes precious scan time. Its reconstruction was limited to a small region of interest (ROI) and not based on any calibration mechanism for quantitative EP calculation, hindering the ultimate resolution, SNR, and data interpretability across the 3D volume.
In this study, we introduce a boundary informed EPT (BIEPT) approach to quantitatively map tumor EPs with high spatial resolution, using a multi-transmit RF coil array at 7 Tesla. BIEPT demonstrates that EP distributions inside an ROI can be reliably retrieved using an external EP reference area and B1 measurements from only one experimental setup. As such, we reduce the total scan time by a factor of two, and minimize opportunities for operational error which compromises imaging resolution and accuracy. The proposed technology was systematically validated using numerical simulations and phantom experiments. EP maps, as well as anatomical T1-weighted (T1w) and T2-weighted (T2w) images, were acquired in vivo from human breast cancer xenografts on eleven athymic nude rats. Ex vivo validation was performed using a dielectric probe. Representative whole-mount histopathological slides of tumor samples were obtained for comparison.
THEORY
In this section, starting from Maxwell’s Equations, we lay out the essential steps that eventually lead to a pixel-wise inverse problem linking MRI-measurable multi-channel relative B1 to the spatial Wirtinger derivative ( ) of EPs defined on a transverse plane. Then, we propose the central optimization problem that retrieves tumor internal EP distributions from their Wirtinger derivatives, with prior information of boundary EPs and use of an L1-norm edge sparsity constraint.
Consider time-harmonic electromagnetic field at angular frequency in Cartesian coordinate system. By combining Ampère’s law and Faraday’s law to eliminate the electric field, we have
[1] |
assuming that magnetic susceptibility in biological tissues. The complex electrical properties is defined as , where denotes relative permittivity to vacuum permittivity , and denotes conductivity. Equation 1 is a vector formula about in phasor terms, which contains the following three equations corresponding to directions
[2] |
is not directly measurable by MRI, but it is possible to measure its positively polarized component in the rotating frame (28). B1 and are used interchangeably in this manuscript for simplicity. Using linear combination of the first two equations in Equation 2, we have
[3] |
where . Utilizing Gauss’ Law for Magnetism, we have
[4] |
For an MRI coil comprised of long striplines, is negligible in the middle part of the coil elements compared to the transverse components (20). Taking this into consideration, the following equation can be derived based on field
[5] |
This equation holds for any transmit field from a single coil element indexed by . However, the absolute phase of , i.e. absolute transmit phase, is intractable, since the phase of MRI signal has mixed contributions from transmit field, receive field, chemical shift, magnetic susceptibility, eddy current effects, etc. Taking the phase difference between two transmit channels and , the common terms can be eliminated, resulting in a relative transmit phase distribution . The relationship between absolute and relative can be expressed as
[6] |
is measurable, whereas is the unknown absolute transmit phase of the reference channel . Combining Equation 6 with Equation 5, we have
[7] |
Cancelling out from both sides of the equation
[8] |
For each spatial location, we have the following matrix form
[9] |
Concatenating all transmit channels that share the same set of unknown and -related variables, we have a linear system of equations for each image pixel containing 4 unknowns and equations, where
[10] |
This linear system of equations is solved in a sense of least-squares to obtain , which is then used to calculate based on
[11] |
Note that is the Wirtinger derivative (29) of defined on the complex plane . In this study, we seek the numerical integral of the Wirtinger derivative defined on a simply connected two-dimensional domain with Dirichlet boundary condition as the integration constant. The central reconstruction problem inside such an ROI is expressed as an equality-constrained convex optimization problem
[12] |
where is a vectorized form of inside ; and are the spatial derivative operators along and directions, respectively; is the spatial Laplacian operator. , , are implemented using finite difference method
[13] |
where is discretization step size, is the 2D map corresponding to the vector. is a regularization parameter heuristically determined per case by L-curve method corresponding to its maximum curvature (30,31); is a subarea of with known EPs . Image sparsity is promoted by including the term that smooths homogeneous patches while preserving sharp edges.
For phantom experiments, the boundary condition is the EPs of the manually drawn boundary which falls into the background agar gel. Its EPs were measured using a dielectric probe as = 0.28 S/m, = 79 . For rat experiments, water-based ultrasound transmission gel was applied over the skin surface on top of the tumor to provide EP reference. The gel EPs were measured using the dielectric probe, which reported consistent results ( = 0.28 S/m, = 77 ) during a 4-hour time window at room temperature (22 °C) and across a temperature range of 22-58 °C (Supporting Fig. S1). The gel also serves as a high-permittivity dielectric pad that boosts local SNR (32) and helps to alleviate susceptibility induced magnetic field alteration on the surface of tissue.
For ROI boundary pixels that fall into rat muscle tissue, the following relaxed boundary constraint is used to regularize reconstruction results
[14] |
which allows unknown muscle pixels to fluctuate around the average muscle EP values , calculated from = 0.9 S/m and = 64 , based on our previous study (27) and further confirmed by probe measurement results from the animal cohort in this study (n = 8, : 0.89 ± 0.06 S/m, : 63.6 ± 2.3 ). is chosen empirically as times the boundary pixel number for all cases.
METHODS
Numerical Simulation
Radiofrequency electromagnetic simulation was performed in SEMCAD (Schmid & Partner Engineering AG, Zurich, Switzerland) based on the finite-difference-time-domain (FDTD) method. The 8-channel stripline coil array used in the experiment was numerically replicated in terms of geometrical structure and electrical circuitry. The coil was loaded with a realistic rat model (275 g female rat model, IT’IS Foundation, Zurich, Switzerland) in a similar positioning as in a real experiment. The EP values assigned to the normal composite tissues were extracted from a widely used database (33). A 20-millimeter-diameter sphere was placed on the left flank of the rat mimicking a tumor xenograft ( = 1.0 S/m, = 65 ). Inside the tumor was a necrotic core with a diameter of 8 mm ( = 1.3 S/m, = 76 ). The discretization grid with spatially variable density was generated by software based on the material and geometry of the model. Grid density was specified to be finer than 2 mm isotropic to ensure precision. The simulation was executed eight times with one channel transmitting at a time, and the input RF power was normalized to 1 Watt. The simulations were carried out on a Linux server equipped with dual high-performance GPUs (Tesla C2070, NVIDIA).
The simulated B1 field at 298 MHz from each transmit RF channel was exported to MATLAB and re-gridded using cubic splines onto an isotropic 1 mm grid. Relative B1 was calculated by combining the B1 magnitude of each channel and the relative phase with respect to the reference channel. Complex white Gaussian noise was added to the relative B1 maps to simulate various SNR levels. The standard deviation of the noise was determined by the average B1 magnitude inside the simulated object and the desirable SNR value. 100 sets of noisy data were obtained for a Monte-Carlo test.
Phantom Preparation
Agar solution containing deionized water, agar, NaCl, and NiCl2·6H2O with mass ratio of 1000:9.0:1.12:0.5 was boiled, naturally cooled down to 45°C while remaining in liquid state, and poured into a 125-mm-high and 65-mm-wide cylindrical container. The EPs were measured as = 0.30 S/m, = 79 using a dielectric probe described in a later section. Anomaly solutions with various EPs were prepared by doping deionized water with NaCl and polyvinylpyrrolidone (PVP). Glass capillary tubes and plastic straws with various diameters (from 1.5 mm to 11 mm) were filled with solutions of various EPs and inserted into the agar gel. The whole phantom was left to solidify in room temperature (22 °C) for 24 hours. Samples from all solutions were taken and their EPs measured using the dielectric probe.
MRI Methods
The experiments were conducted in a human whole-body 7 Tesla MRI system (Siemens, Erlangen, Germany) equipped with 16 × 1 kW RF amplifiers (Communication Power Corporation, Hauppauge, NY, USA). The amplifiers were driven by a phase/gain controller remotely operated by a software toolbox developed in-house based on MATLAB (34,35). The sample under test was positioned in the center of a dedicated 8-channel RF transceiver coil array, comprised of eight 100-mm-long and 12-mm-wide striplines azimuthally distributed on a 100-mm-diameter cylindrical surface (27). Each coil element was connected to a power source via a two-parameter tuning/matching capacitors network, which was adjusted before each experiment to minimized power reflection at the proton resonance frequency (297.2 MHz). Adjacent coil elements were de-coupled by a tunable capacitor so as to provide distinct B1 distributions. The coil was then connected to the MRI system via a custom transmit/receive switch box.
A hybrid B1 mapping method was used to retrieve the B1 magnitude (36), which combines (i) a reference scan based on 3D actual flip-angle imaging (AFI) (37) with all channels transmitting together in large-flip-angle regime, and (ii) a series of multi-slice gradient echo (GRE) images in small-flip-angle regime pulsing through each channel sequentially. AFI parameters are FA = 60°, TR = 120/20 ms, TE = 3.23 ms, voxel size 1.0 × 1.0 × 2.0 mm3, spatial coverage 256 × 168 × 96 mm3, duration 12.4 min; GRE parameters are FA = 15°, TR = 126 ms, TE = 3.3 ms, in-plane resolution 1.0 × 1.0 mm2, FOV 256 × 168 mm2, 12 slices, 2 mm slice thickness, 8 averages, total duration 28.3 min. Relative B1 phase was acquired using another set of GRE with the same parameters except FA = 60° and average number = 3, with duration of 10.6 min. Before the acquisition of the high-resolution B1 data, scout low-resolution images of GRE and AFI were acquired for B1 shimming and calibration purposes, with duration of ~10 min. The total B1 mapping session is approximately 60 min.
T1w images in resolution of 0.5 × 0.5 × 1.0 mm3 were obtained by taking the ratio of images from a 3D magnetization prepared rapid acquisition gradient echo (3D-MPRAGE) sequence and a 3D small-flip angle gradient echo (GRE) sequence, so as to cancel out undesirable weightings such as transmit and receive B1, T2*, and proton density (38). T2w images were based on a turbo spin echo (TSE) sequence in 0.25 × 0.25 × 0.5 mm3 resolution, with shimmed B1 in the tumor to maximize excitation homogeneity.
In Vivo Experiments
All animal procedures and care were approved by the University of Minnesota Institutional Animal Care and Use Committee (IACUC) in accordance with federally approved guidelines.
Animal was initially anesthetized in an induction chamber with 5% isoflurane mixed with 70% medical air and 30% oxygen, then maintained during imaging via a nose cone (Kent Scientific) at 1-2% isoflurane. Animal heating was achieved by a heating pad during preparation and hot water surrounding the animal holder during MR imaging. The temperature of water flowing through tubing around the animal was controlled by a thermostat integrated with water bath circulator (NESLAB RTE 7, Thermo Scientific). The respiratory rate and body temperature were monitored by an MR-compatible small animal monitoring and gating system (SA Instruments) during the procedures. The animal was positioned to lie on its side with the implanted tumor facing upwards. The orientation and the central axial slices of the tumor were labeled using an extra-fine felt-tip marker. The skin area around the tumor was shaved, and ultrasound transmission gel (Aquasonic 100, Parker Lab. Inc.) was applied to form a layer of approximately 10 mm on top of the tumor. For terminal studies, the animal was euthanized by carbon dioxide inhalation after the scan.
Additional animal handling procedures may be found in the supporting information (Supporting Method S2).
Dielectric Probe Measurement
Measurement was performed using Keysight 85070E dielectric probe kit connected to an E4991A network analyzer. The probe, shaped like a planar disk, has a diameter of 18 mm. Measurement of samples smaller than 20 mm is not recommended by the manufacturer due to compromised accuracy. Prior to measurement, the system was calibrated with air, shorting block, and deionized water each time it was set up. Immediately after euthanasia of the animal, a central cut was performed to expose the maximum cross-section of the tumor, where the open-ended coaxial probe was positioned and held still for 5 seconds during one trial. Measurement was repeated 3~5 times for each sample. Since the measurements were completed in a short period right after sacrifice, we consider the rat body temperature (37 °C) as the measurement temperature.
Among the eleven animals involved in the study, six were both imaged and probed; Five were imaged but not probed, including three animals with tumor smaller than 20 mm, and two animals whose tumor contains substantial liquefaction that hinders a successful measurement. Two additional animals, in which tumor did not successfully grow, were used only for probe measurement of muscle tissue.
EPs Reconstruction
Noisy simulation data and experimental data were smoothed using a low-pass Gaussian filter with a kernel size of 5 × 5 × 3 pixels and standard deviation of 1.2 × 1.2 × 0.8 pixels. The spatial Laplacian and gradient of relative B1 were taken using a Savitzky-Golay filter with a kernel size of 5 × 5 × 3. The pixel-wise inverse problem (Equ. 9) was solved in the least-squares sense in MATLAB using backslash operation.
For phantoms, target maps were generated from manual segmentation of pilot images followed by assignment to each patch the measured EP values obtained from the probe. During BIEPT reconstruction, EPs on the phantom boundary were given as prior information based on probe measurement of the agar background ( = 0.30 S/m, = 79 ). For simulation and in vivo experiments, the “gel region” was contoured and assigned EP values based on probe measurement ( = 0.28 S/m, = 77 ). Then a connected ROI was contoured that contains the whole tumor, part of the gel region and muscle tissue. For in vivo data, bone and tendon pixels were carefully excluded due to unreliable B1 measurement caused by the weak MR signal.
The central optimization problem (Equ. 12) was generated for each 2-dimensional (2D) ROI and solved using CVX, a MATLAB compatible package for specifying and solving convex programs (39). It takes approximately 3 seconds to solve for each axial slice on a desktop with 3.4 GHz GPU and 32 GB RAM. Conductivity and relative permittivity were then computed from as
[15] |
Single slice EP results were interpolated using cubic splines to yield an in-plane resolution of 0.5 × 0.5 mm2. The reconstructed EPs inside ROI are shown in color maps overlapped onto structural MRI.
A flowchart for the relative B1 mapping and BIEPT reconstruction is shown in Figure 1.
RESULTS
Phantom Validation Results
Reconstruction results of the resolution phantom are shown in Figure 2. Compared to the target maps, the locations and shapes of the tube anomalies are faithfully reflected by the reconstructed EP images. The central profiles of the reconstructed tubes show successful preservation of the conductivity and permittivity contrasts in structures ≥ 3 mm. Stable plateaus can be observed for structures ≥ 7 mm. To separate the reconstruction error due to B1 mapping from that due to BIEPT algorithm, this phantom was replicated in an electromagnetic simulation to acquire accurate B1 for EP reconstruction. The reconstruction results using the same filters and regularization parameters are shown in Supporting Fig. S2. Similar smoothing effects on tube boundaries can be observed due to filtering in the algorithm, but imaging accuracy in the center of the large tubes is improved thanks to the use of accurate B1.
Reconstructed EPs from the contrast phantom are shown and fitted to the target values in Figure 3. A good correlation between the reconstructed and target EP values is observed from the imaging results as well as linear fitting results, with ≤ 5% conductivity error and ≤ 12% permittivity error across the physiologically relevant EP ranges (0.3 S/m ≤ ≤ 1.6 S/m, 50 ≤ ≤ 80 ). Conductivity imaging exhibits superior robustness to permittivity imaging, in the sense of image distortion and fitting accuracy. This phenomenon is consistent with previous studies (40,41,20) and will be discussed later.
Numerical Simulation Results
The simulation setup and single-channel absolute B1 maps are shown in Figure 4a. Relative B1 maps were measured and used to calculate EPs. Reconstruction results from two SNR levels, 150 and 50, are shown in Figure 4b. At both SNR levels, the simulated necrotic core with higher conductivity and permittivity can be clearly identified. The Monte-Carlo simulation of 100 trials demonstrates successful differentiation of EP differences among the three compartments on both noise levels, with less than 10% reconstruction error for both conductivity and permittivity. Nevertheless, contrast differences among the three compartments are smaller, due to the smoothing effects of the de-noise filter, as well as contrast leakage effects from the spatial integration process.
Effects of regularization parameter selection are shown in Figure 5 using simulated B1 with SNR = 50. The value corresponding to the maximum curvature of the L-curve gives a reasonable tradeoff between data fidelity and sparsity. The reconstruction results are insensitive to the exact value of as long as it is in the “bending knee” region of the L-curve. Similarly, the results are robust against selection when it is close to the empirically chosen value marked in red.
Animal Experiment Results
Bulk EP values obtained from BIEPT and the dielectric probe are shown in Figure 6. For each individual animal as well as grouped results, BIEPT reports average EP values close to those from the probe measurement. Probed conductivity is slightly higher than that obtained by BIEPT, potentially due to bleeding when cutting the tumor. The bulk EPs are found not to be significantly dependent on tumor size, which is consistent with previous finding in conductivity imaging of breast cancer patients (23). However, it can be observed that considerable EP variation resides in each imaging case, with standard deviation as large as 20% of the average values. These variations are less prominent in the values obtained from the physical probe measurements, which is expected since the latter report the bulk average of the entire tumor sample without spatial information, considering that the size of the physical probe is about the size of the tumor.
To demonstrate the capability of BIEPT to image tumor heterogeneity, we compare the reconstructed EP images with corresponding T1w, T2w images and whole-mount histopathology slides. Two representative cases are shown in Figures 7 and 8, respectively.
In the first case, the necrotic cores indicated by purple and red arrows show opposite signal change on T1w and T2w images compared to solid tumor (Fig. 7a). It is known that opposite T1w and T2w contrast variations can be observed within regions that are identified as necrotic in nature, due to two competing factors, paramagnetic ions and water content, with the former decreasing T1 and T2 and the latter increasing them (42). Yet both factors synergistically contribute to locally increased conductivity, as shown as distinguished tumor sub-regions on the conductivity map. On the other hand, water content contributes to permittivity elevation while ion concentration has minimal effect on the latter, leading to the contrast between these sub-regions on the reconstructed permittivity map. Hematoxylin and eosin (H&E) stain confirms the presence of three coagulative necrotic areas (Fig. 7b). Tumor necrosis is defined as irreversible cell and nuclear changes such as karyolysis, vacuolization of the nucleus and/or cytoplasm, and dissolution of the membrane integrity. As a result of these morphologic changes, necrotic areas will appear as “pink” areas on H&E stain within the tissue that are readily conspicuous at lower magnification. A central cut was made through the tumor along the green dashed line, allowing for dielectric probe measurement of the bulk tumor EPs, found as 1.01 ± 0.04 S/m and 65.8 ± 4.4 , comparable to the BIEPT reported EP values of 0.95 ± 0.14 S/m and 66.0 ± 4.0 .
In the second case, a large necrotic core appears in the second quartile of the tumor with two distinct areas showing clear contrast on T2w image (Fig. 8a, red and purple arrows). On T1w image, one area exhibits significant hypointensity (red arrow), while the other shows marginal hyperintensity (purple arrow). Nevertheless, both of these two areas show conductivity increase while discrepancy in permittivity is preserved. The H&E slide shows more than 80% necrosis with massive karyolysis and cell membrane loss in the left sub-region (Fig. 8b, purple arrow). Neutrophil and fibroblast infiltration is observed that removes water from tissue, in keeping with the underlying process of healing and scar formation, which may contribute to relatively low free water content. Figure 7a also shows successful detection of two 2-mm necrotic cores with elevated EPs, labeled by blue arrows, while their contrast on T1w and T2w images are less discernable. These two necrotic cores merge as a single hot spot on EP images due to the inherent smoothing properties of the reconstruction process. Probe measurement of the tumor reported 1.12 ± 0.10 S/m and 72.1 ± 2.9 , comparable to the BIEPT measured EP values as 0.96 ± 0.08 S/m and 71.0 ± 4.3 .
Despite the fact that each axial slice was reconstructed independently, consistent 3D EPs imaging results can be obtained thanks to the gel region that provides a reliable ground truth across all slices. Multi-slice EPs images are shown in the support information (Supporting Animation S4). A longitudinal study was also performed to monitor tumor growth in a 1.5-month time window, showing in high-resolution EP pattern evolution as the tumor grows (Fig. 9).
DISCUSSION
In this study, we have proposed a Boundary Informed Electrical Properties Tomography (BIEPT) technique for high-resolution electrical conductivity and permittivity mapping from a boundary reference area and EPs’ Wirtinger derivative computed using a plurality of relative transmit B1 measurements from a 7 Tesla MRI system, with enhanced numerical stability from edge sparsity regularization. This method does not require physical flipping of the imaging setup or image registration of multiple datasets, thus representing a more generalized methodology compared to gEPT in the context of multi-transmission. Simulation and phantom results demonstrated its high accuracy across physiologically relevant EP ranges and sensitivity to anomalies as small as 3 mm, under varied noisy conditions. In vivo tumor imaging experiments were performed on eleven rats with human breast cancer xenograft, referenced to externally applied ultrasound gel. We observed association of local conductivity elevation with tumor necrosis. We also noticed potential correlation of permittivity with additional tumor pathological characteristics such as healing and scar formation, yet the confidence level of permittivity is lower due to its susceptibility to image distortions and noise. Nevertheless, combination of the spatial distributions of both may give rise to a useful biomarker in clinical settings, such as tumor staging and treatment evaluation.
Previous human cancer studies based on either an invasive probe or non-invasive MREPT approaches have reported differences in mean EP values between benign and malignant tissues (3,43,24), tumors of low and high grades (25), as well as tumors in early and late stages (2). Despite the established correlation between EPs and those clinically meaningful tumor classifications, the underlying contrast mechanism and tumor biology remain unclear, partially due to the limited spatial resolution of the previous approaches and the practice of evaluating bulk EPs across large tumors (> 20 mm). To address this pivotal question, we developed BIEPT to retrieve high-resolution EP images from an animal tumor model in vivo. The observed association of EP changes with tumor necrosis may provide an explanation for the previously reported EP deviations in late-stage, high-grade, malignant tumors. Moreover, local feature identification and heterogeneity quantification of EPs derived from high-resolution image modality are potentially more sensitive biomarkers than bulk mean EPs in terms of tumor characterization (44,45).
Successful detection of both spontaneous and induced tumor necrosis is expected to have impacts on clinical practice. Spontaneous necrosis is often observed in highly invasive tumors with aggressive proliferation, due to multiple factors including deficiency of local blood perfusion that disrupts pH equilibrium, oxygen and nutrient supply, as well as removal of metabolic wastes (46). Necrotic tumors secrete stimulatory chemokines and cytokines which in turn facilitate tumor growth and infiltration (47,48). In practice, excessive necrosis is a feature of aggressive tumors with unfavorable prognosis in terms of early recurrence and mortality (49,50), and necrotic tumors are found to be more resistant to many types of chemotherapy as well as radiation therapy (51). Therefore, mapping the geometry and EPs of tumor necrosis can potentially provide valuable information to help guide clinical decisions and optimize individualized treatment plan. On the other hand, many successful cancer therapies, including chemo- and radiation-therapy approaches, induce intra-tumor necrosis when they are effective, and this typically takes place well before substantial changes in tumor volumetric numbers. Therefore, imaging tumor necrosis may provide not only early feedback on the effectiveness of the administered therapy but also partial response alerts. The proposed EPs based necrosis imaging technique is based on endogenous tissue contrast, representing a new avenue to be explored for monitoring early therapeutic response as well as a suboptimal therapy.
Based on our current findings of tumor necrosis, more investigations are warranted to explore the association of EP changes with other tumor signatures, such as angiogenesis and hypoxia, as well as tumor response to treatment. These biological effects typically happen on a much smaller spatial scale compared to necrosis, and they are expected to have more subtle impacts on EPs. Therefore, detection of these effects may have a better chance on dedicated animal scanners with higher gradient power for spatial encoding and higher magnetic field for improved SNR (52,53). In addition, animal tumor imaging in situ, i.e. in its original site, should be explored for improved modeling of tumor behavior, as well as patient studies to evaluate EPs as clinically meaningful biomarkers.
We include in this paper a single case of tumor growth monitoring over a 1.5-month time window. Despite the limited sample size, this case study demonstrates consistent spatial-temporal EP patterns in comparison to the anatomical T1w and T2w images. It provides an initial indication in favor of the reliability of the technology and as such may provide strong rationale to justify a larger sample number in longitudinal monitoring to eventually achieve biological significance.
In this study, BIEPT has been demonstrated to produce reconstruction results close to the ground truth in both simulation and phantom experiments. However, systematic reconstruction error can be spotted which mainly comes from two sources: B1 mapping and BIEPT algorithm. There is small residual bias in the measured B1, such as proton density and T1, which is amplified by its spatial differentiation and propagates to the reconstructed EPs. Therefore, improved imaging accuracy can be achieved in the simulation results using accurate B1 (Supporting Fig. 2) compared to the experimental results using measured B1 (Fig. 2). Supporting Fig. 2 also shows that BIEPT alone does not seem to significantly compromise the reconstruction accuracy in large structures ≥ 6 mm when using accurate B1. Its effect is mainly smoothing out small structures, due to the low-pass filters and operational kernels involved that are unavoidable in a successful EP reconstruction. Previous studies have shown that, to the leading order, permittivity reconstruction is more sensitive to the accuracy of B1 magnitude while conductivity reconstruction is more sensitive to the accuracy of B1 phase (40,54,9,11). The potential sources of bias in our B1 mapping method have more significant impacts on its magnitude than its phase, which partially accounts for the larger reconstruction error observed in permittivity than conductivity (Fig. 2). Using simulated B1, conductivity and permittivity can be reconstructed at a similarly high-accuracy level (Supporting Fig. 2).
Current BIEPT presents other limitations that need to be addressed in future work. First of all, the current multi-channel B1 mapping technique requires about one hour of acquisition time to produce 1.0 × 1.0 × 2.0 mm3 maps in 8 channels with satisfactory SNR. While providing high-quality B1 maps in multi-channel arrays, the hybrid B1 mapping method utilized in this work has not been specifically optimized for time efficiency; alternative B1 mapping methods recently proposed, such as Bloch-Siegert method (55), DREAM (56), Large Dynamic Range B1 mapping (57), could be implemented for comparison. Furthermore, a high count of receive channels positioned close to the imaged body could also be employed additionally to a transmit array to improve SNR, instead of a transceiver array which has inherently compromised receive performance compared with receive-only arrays. Another limitation in the current study stems from the fact that BIEPT is tested in a regularly shaped, simply-connected ROI, which serves our purpose well to image a tumor xenograft surrounded by the water-based gel. In cases where the ROI is irregularly shaped or has cavities, e.g. imaging near the bone or in the bowel, the robustness of BIEPT needs to be addressed. It is expected that prior knowledge about the average EPs of the imaging area and regularization methods will be helpful in those situations. BIEPT also requires an external reference area to infer the underlying EPs of a tumor, which would not be practical in most clinical settings. Nevertheless, EPs of normal body tissues have been well characterized, including some that have a relatively large area of reasonable homogeneity (e.g. muscle, liver, etc.), and may be used as a reference to BIEPT when a target tumor is contiguous to them. EP values of these normal tissues can be either extracted from the widely used literature by Gabriel et al. based on ex vivo measurements (33), or calculated in vivo using B1 based on the Helmholtz equation.
As a technology to retrieve tissue EPs from the B1 field instead of the full magnetic field which cannot be directly measured by MRI, EPT intrinsically involves assumptions about the field distribution such as negligible to reduce the unknown variables in the central equation to a solvable level. This is especially important when the spatial variation of EPs is also considered in the EPT problem. Several studies have utilized the assumption and shown its validity in the central part of a quadrature volume coil (14,58,59) as well as a stripline coil element (20). However, this assumption posts limitation to the types of RF coil that can be used and the spatial imaging coverage. On the other hand, a series of numerical procedures need to be followed for a successful reconstruction of the EP contrast. These procedures include design the of low-pass filters, Laplacian and gradient kernels, spatial difference methods, regularization strategies, etc. The main goal of these procedures is to suppress measurement noise in B1 and to achieve stability in the reconstruction results, at a price of imaging resolution and accuracy as well as algorithm complexity. There is a tradeoff among these metrics of imaging quality, which has been investigated from various aspects in previous work (53,60). In the future, technical innovations are called for, potential based on different imaging principles such as scattering field theory (59,61,62), to bypass a subset of the aforementioned field assumptions and approximations in order to further generalize the EPT technology and to bring its capability to the next level.
CONCLUSION
We conclude that tumor EPs heterogeneity can be inferred by the proposed boundary informed electrical properties tomography (BIEPT) technique based on unique transmit B1 images measured from a multi-channel RF coil, using an external reference region and image edge sparsity constraint. Similar spatial patterns amongst tumor EP distributions, anatomical T1w and T2w images and pathology have been observed in animal experiments. Association of altered local EP values with tumor necrosis has been reported. Further investigations will be dedicated to improving and generalizing BIEPT for identification of EPs-related pathological features, as well as their evaluation as biomarkers in cancer patient care.
Supplementary Material
Acknowledgments
This work was supported in part by NIH EB017069, NIH EB014353, NIH MH114233, NIH EB021027, NIH NS096761, NIH HL117664, NIH EB015894, NIH NS076408, NIH RR026783, NIH EB008389, NSF CBET-1450956, NSF CBET-1264782, and WM KECK Foundation. The authors thank Dr. Gregor Adriany for MRI hardware assistance; Drs. Michael Garwood, Patrick Bolan and Michael Nelson for their inputs on cancer imaging; Drs. Long Yu and Ting Yang for discussions on numerical analysis; Mr. Kai Yu for inputs on phantom construction; Ms. Jiaoyue Liu for artwork assistance and Mr. Christopher Cline for manuscript editing.
References
- 1.Schwan HP, Foster KR. RF-field interactions with biological systems: electrical properties and biophysical mechanisms. Proc IEEE. 1980;68:104–113. doi: 10.1109/PROC.1980.11589. [DOI] [Google Scholar]
- 2.Swarup A, Stuchly SS, Surowiec A. Dielectric properties of mouse MCA1 fibrosarcoma at different stages of development. Bioelectromagnetics. 1991;12:1–8. doi: 10.1002/bem.2250120102. [DOI] [PubMed] [Google Scholar]
- 3.Joines WT, Zhang Y, Li C, Jirtle RL. The measured electrical properties of normal and malignant human tissues from 50 to 900 MHz. Med Phys. 1994;21:547–550. doi: 10.1118/1.597312. [DOI] [PubMed] [Google Scholar]
- 4.Metherall P, Barber DC, Smallwood RH, Brown BH. Three-dimensional electrical impedance tomography. Nature. 1996;380:509–512. doi: 10.1038/380509a0. [DOI] [PubMed] [Google Scholar]
- 5.Woo EJ, Seo JK. Magnetic resonance electrical impedance tomography (MREIT) for high-resolution conductivity imaging. Physiol Meas. 2008;29:R1. doi: 10.1088/0967-3334/29/10/R01. [DOI] [PubMed] [Google Scholar]
- 6.Xu Y, He B. Magnetoacoustic tomography with magnetic induction (MAT-MI) Phys Med Biol. 2005;50:5175. doi: 10.1088/0031-9155/50/21/015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Fear EC, Li X, Hagness SC, Stuchly MA. Confocal microwave imaging for breast cancer detection: localization of tumors in three dimensions. IEEE Trans Biomed Eng. 2002;49:812–822. doi: 10.1109/TBME.2002.800759. [DOI] [PubMed] [Google Scholar]
- 8.Haacke EM, Petropoulos LS, Nilges EW, Wu DH. Extraction of conductivity and permittivity using magnetic resonance imaging. Phys Med Biol. 1991;36:723–734. doi: 10.1088/0031-9155/36/6/002. [DOI] [Google Scholar]
- 9.Katscher U, van den Berg CAT. Electric properties tomography: biochemical, physical and technical background, evaluation and clinical applications. NMR Biomed. 2017;30:e3729. doi: 10.1002/nbm.3729. [DOI] [PubMed] [Google Scholar]
- 10.Zhang X, Liu J, He B. Magnetic-resonance-based electrical properties tomography: a review. Biomed Eng IEEE Rev In. 2014;7:87–96. doi: 10.1109/RBME.2013.2297206. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Liu J, Wang Y, Katscher U, He B. Electrical properties tomography based on B1 maps in MRI: principles, applications, and challenges. IEEE Trans Biomed Eng. 2017;64:2515–2530. doi: 10.1109/TBME.2017.2725140. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Wen H. Proc SPIE. San Diego, USA: 2003. Noninvasive quantitative mapping of conductivity and dielectric distributions using RF wave propagation effects in high-field MRI; pp. 471–477. [DOI] [Google Scholar]
- 13.Katscher U, Voigt T, Findeklee C, Vernickel P, Nehrke K, Dossel O. Determination of electric conductivity and local SAR via B1 mapping. IEEE Trans Med Imaging. 2009;28:1365–1374. doi: 10.1109/TMI.2009.2015757. [DOI] [PubMed] [Google Scholar]
- 14.Zhang X, Zhu S, He B. Imaging Electric Properties of Biological Tissues by RF Field Mapping in MRI. IEEE Trans Med Imaging. 2010;29:474–481. doi: 10.1109/TMI.2009.2036843. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Katscher U, Findeklee C, Voigt T. B1-based specific energy absorption rate determination for nonquadrature radiofrequency excitation. Magn Reson Med. 2012;68:1911–1918. doi: 10.1002/mrm.24215. [DOI] [PubMed] [Google Scholar]
- 16.Zhang X, Schmitter S, Van de Moortele P, Liu J, He B. From complex mapping to local SAR estimation for human brain MR imaging using multi-channel transceiver coil at 7T. IEEE Trans Med Imaging. 2013;32:1058–1067. doi: 10.1109/TMI.2013.2251653. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Zhang X, de Moortele P-FV, Schmitter S, He B. Complex B1 mapping and electrical properties imaging of the human brain using a 16-channel transceiver coil at 7T. Magn Reson Med. 2013;69:1285–1296. doi: 10.1002/mrm.24358. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Sodickson D, Alon L, Deniz C, Ben-Eliezer N, Cloos M, Sodickson L, Collins C, Wiggins G, Novikov D. Proc Intl Soc Mag Reson Med. 21. Salt Lake City, USA: 2013. Generalized Local Maxwell Tomography for Mapping of Electrical Property Gradients and Tensors; p. 3372. [Google Scholar]
- 19.Liu J, Zhang X, de Moortele P-FV, Schmitter S, He B. Determining electrical properties based on B1 fields measured in an MR scanner using a multi-channel transmit/receive coil: a general approach. Phys Med Biol. 2013;58:4395. doi: 10.1088/0031-9155/58/13/4395. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Liu J, Zhang X, Schmitter S, Van de Moortele P-F, He B. Gradient-based electrical properties tomography (gEPT): A robust method for mapping electrical properties of biological tissues in vivo using magnetic resonance imaging. Magn Reson Med. 2015;74:634–646. doi: 10.1002/mrm.25434. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Liu J, de Moortele PFV, Zhang X, Wang Y, He B. Simultaneous Quantitative Imaging of Electrical Properties and Proton Density From Maps Using MRI. IEEE Trans Med Imaging. 2016;35:2064–2073. doi: 10.1109/TMI.2016.2547988. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Van Lier ALHMW, Raaijmakers A, Voigt T, Lagendijk JJW, Luijten PR, Katscher U, van den Berg CAT. Electrical Properties Tomography in the Human Brain at 1.5, 3, and 7T: A Comparison Study. Magn Reson Med. 2014;71:354–363. doi: 10.1002/mrm.24637. [DOI] [PubMed] [Google Scholar]
- 23.Shin J, Kim MJ, Lee J, Nam Y, Kim M, Choi N, Kim S, Kim D-H. Initial study on in vivo conductivity mapping of breast cancer using MRI. J Magn Reson Imaging. 2015;42:371–378. doi: 10.1002/jmri.24803. [DOI] [PubMed] [Google Scholar]
- 24.Balidemaj E, de Boer P, van Lier ALHMW, Remis RF, Stalpers LJA, Westerveld GH, Nederveen AJ, van den Berg CAT, Crezee J. In vivo electric conductivity of cervical cancer patients based on B1 maps at 3T MRI. Phys Med Biol. 2016;61:1596. doi: 10.1088/0031-9155/61/4/1596. [DOI] [PubMed] [Google Scholar]
- 25.Tha KK, Katscher U, Yamaguchi S, et al. Noninvasive electrical conductivity measurement by MRI: a test of its validity and the electrical conductivity characteristics of glioma. Eur Radiol. 2018;28:348–355. doi: 10.1007/s00330-017-4942-5. [DOI] [PubMed] [Google Scholar]
- 26.Duan S, Xu C, Deng G, Wang J, Liu F, Xin SX. Quantitative analysis of the reconstruction errors of the currently popular algorithm of magnetic resonance electrical property tomography at the interfaces of adjacent tissues. NMR Biomed. 2016;29:744–750. doi: 10.1002/nbm.3522. [DOI] [PubMed] [Google Scholar]
- 27.Liu J, Shao Q, Wang Y, Adriany G, Bischof J, Van de Moortele P-F, He B. In vivo imaging of electrical properties of an animal tumor model with an 8-channel transceiver array at 7 T using electrical properties tomography. Magn Reson Med. 2017;78:2157–2169. doi: 10.1002/mrm.26609. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Hoult DI. The principle of reciprocity in signal strength calculations—A mathematical guide. Concepts Magn Reson. 2000;12:173–187. doi: 10.1002/1099-0534(2000)12:4<173::AID-CMR1>3.0.CO;2-Q. [Google Scholar]
- 29.Freitag E, Busam R. Complex Analysis. Springer Science & Business Media; 2006. [Google Scholar]
- 30.Hansen PC, Danmarks Tekniske Universitet, Institut for Matematisk Modellering . The L-curve and its use in the numerical treatment of inverse problems. Lyngby: IMM, Department of Mathematical Modelling, Technical Universityof Denmark; 1999. [Google Scholar]
- 31.Sohrabpour A, Lu Y, Worrell G, He B. Imaging brain source extent from EEG/MEG by means of an iteratively reweighted edge sparsity minimization (IRES) strategy. NeuroImage. 2016;142:27–42. doi: 10.1016/j.neuroimage.2016.05.064. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Yang QX, Mao W, Wang J, Smith MB, Lei H, Zhang X, Ugurbil K, Chen W. Manipulation of image intensity distribution at 7.0 T: passive RF shimming and focusing with dielectric materials. J Magn Reson Imaging. 2006;24:197–202. doi: 10.1002/jmri.20603. [DOI] [PubMed] [Google Scholar]
- 33.Gabriel S, Lau RW, Gabriel C. The dielectric properties of biological tissues: III. Parametric models for the dielectric spectrum of tissues. Phys Med Biol. 1996;41:2271–2293. doi: 10.1088/0031-9155/41/11/003. [DOI] [PubMed] [Google Scholar]
- 34.Van de Moortele P-F, Akgun C, Adriany G, Moeller S, Ritter J, Collins CM, Smith MB, Vaughan JT, Uğurbil K. B1 destructive interferences and spatial phase patterns at 7 T with a head transceiver array coil. Magn Reson Med. 2005;54:1503–1518. doi: 10.1002/mrm.20708. [DOI] [PubMed] [Google Scholar]
- 35.Metzger GJ, Snyder C, Akgun C, Vaughan T, Ugurbil K, Van de Moortele P-F. Local B1+ shimming for prostate imaging with transceiver arrays at 7T based on subject-dependent transmit phase measurements. Magn Reson Med. 2008;59:396–409. doi: 10.1002/mrm.21476. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Van de Moortele PF, Snyder C, DelaBarre L, Adriany G, Vaughan JT, Ugurbil K. Proc Intl Soc Mag Reson Med. 15. Berlin, Germany: 2007. Calibration Tools for RF Shim at Very High Field with Multiple Element RF Coils: from Ultra Fast Local Relative Phase to Absolute Magnitude B1+ Mapping; p. 1676. [Google Scholar]
- 37.Yarnykh VL. Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field. Magn Reson Med. 2007;57:192–200. doi: 10.1002/mrm.21120. [DOI] [PubMed] [Google Scholar]
- 38.Van de Moortele P-F, Auerbach EJ, Olman C, Yacoub E, Uğurbil K, Moeller S. T1 weighted brain images at 7 Tesla unbiased for Proton Density, T2* contrast and RF coil receive B1 sensitivity with simultaneous vessel visualization. NeuroImage. 2009;46:432–446. doi: 10.1016/j.neuroimage.2009.02.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Grant MC, Boyd SP. Recent Advances in Learning and Control. Springer; London: 2008. Graph implementations for nonsmooth convex programs; pp. 95–110. [Google Scholar]
- 40.Voigt T, Katscher U, Doessel O. Quantitative conductivity and permittivity imaging of the human brain using electric properties tomography. Magn Reson Med. 2011;66:456–466. doi: 10.1002/mrm.22832. [DOI] [PubMed] [Google Scholar]
- 41.Marques JP, Sodickson DK, Ipek O, Collins CM, Gruetter R. Single acquisition electrical property mapping based on relative coil sensitivities: A proof-of-concept demonstration. Magn Reson Med. 2014;74:185–195. doi: 10.1002/mrm.25399. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Jakobsen I, Kaalhus O, Lyng H, Rofstad EK. Detection of necrosis in human tumour xenografts by proton magnetic resonance imaging. Br J Cancer. 1995;71:456–461. doi: 10.1038/bjc.1995.93. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Kim S-Y, Shin J, Kim D-H, Kim MJ, Kim E-K, Moon HJ, Yoon JH. Correlation between conductivity and prognostic factors in invasive breast cancer using magnetic resonance electric properties tomography (MREPT) Eur Radiol. 2015;26:2317–26. doi: 10.1007/s00330-015-4067-7. [DOI] [PubMed] [Google Scholar]
- 44.Swanton C. Intratumor heterogeneity: evolution through space and time. Cancer Res. 2012;72:4875–4882. doi: 10.1158/0008-5472.CAN-12-2217. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.O’Connor JPB, Rose CJ, Waterton JC, Carano RAD, Parker GJM, Jackson A. Imaging intratumor heterogeneity: role in therapy response, resistance, and clinical outcome. Clin Cancer Res. 2015;21:249–257. doi: 10.1158/1078-0432.CCR-14-0990. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Weinberg R. The Biology of Cancer. Second. Garland Science; New York: 2013. [Google Scholar]
- 47.Vakkila J, Lotze MT. Inflammation and necrosis promote tumour growth. Nat Rev Immunol. 2004;4:641–648. doi: 10.1038/nri1415. [DOI] [PubMed] [Google Scholar]
- 48.Proskuryakov SY, Gabai VL. Mechanisms of tumor cell necrosis. Curr Pharm Des. 2010;16:56–68. doi: 10.2174/138161210789941793. [DOI] [PubMed] [Google Scholar]
- 49.Gilchrist KW, Gray R, Fowble B, Tormey DC, Taylor SG. Tumor necrosis is a prognostic predictor for early recurrence and death in lymph node-positive breast cancer: a 10-year follow-up study of 728 eastern cooperative oncology group patients. J Clin Oncol. 1993;11:1929–1935. doi: 10.1200/JCO.1993.11.10.1929. [DOI] [PubMed] [Google Scholar]
- 50.Barker FG, Davis RL, Chang SM, Prados MD. Necrosis as a prognostic factor in glioblastoma multiforme. Cancer. 1996;77:1161–1166. doi: 10.1002/(sici)1097-0142(19960315)77:6<1161::aid-cncr24>3.0.co;2-z. doi: 10.1002/(SICI)1097-0142(19960315)77:6<1161::AID-CNCR24>3.0.CO;2-Z. [DOI] [PubMed] [Google Scholar]
- 51.Sutherland RM, Rasey JS, Hill RP. Tumor biology. Am J Clin Oncol. 1988;11:253–274. doi: 10.1097/00000421-198806000-00004. [DOI] [PubMed] [Google Scholar]
- 52.Budinger TF, Bird MD, Frydman L, et al. Toward 20 T magnetic resonance for human brain studies: opportunities for discovery and neuroscience rationale. Magn Reson Mater Phys Biol Med. 2016;29:617–639. doi: 10.1007/s10334-016-0561-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Lee S-K, Bulumulla S, Hancu I. Theoretical investigation of random noise-limited signal-to-noise ratio in MR-based electrical properties tomography. IEEE Trans Med Imaging. 2015;34:2220–32. doi: 10.1109/TMI.2015.2427236. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Van Lier ALHMW, Brunner DO, Pruessmann KP, Klomp DWJ, Luijten PR, Lagendijk JJW, van den Berg CAT. B 1+ Phase mapping at 7 T and its application for in vivo electrical conductivity mapping. Magn Reson Med. 2012;67:552–561. doi: 10.1002/mrm.22995. [DOI] [PubMed] [Google Scholar]
- 55.Sacolick LI, Sun L, Vogel MW, Dixon WT, Hancu I. Fast radiofrequency flip angle calibration by Bloch–Siegert shift. Magn Reson Med. 2011;66:1333–1338. doi: 10.1002/mrm.22902. [DOI] [PubMed] [Google Scholar]
- 56.Nehrke K, Börnert P. DREAM—a novel approach for robust, ultrafast, multislice B1 mapping. Magn Reson Med. 2012;68:1517–1526. doi: 10.1002/mrm.24158. [DOI] [PubMed] [Google Scholar]
- 57.Padormo F, Hess AT, Aljabar P, Malik SJ, Jezzard P, Robson MD, Hajnal JV, Koopmans PJ. Large dynamic range relative B1+ mapping. Magn Reson Med. 2016;76:490–499. doi: 10.1002/mrm.25884. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Hafalir FS, Oran OF, Gurler N, Ider YZ. Convection-Reaction Equation Based Magnetic Resonance Electrical Properties Tomography (cr-MREPT) IEEE Trans Med Imaging. 2014;33:777–793. doi: 10.1109/TMI.2013.2296715. [DOI] [PubMed] [Google Scholar]
- 59.Balidemaj E, van den Berg C, Trinks J, Lier A, Nederveen A, Stalpers L, Crezee H, Remis R. CSI-EPT: A Contrast Source Inversion Approach for Improved MRI-Based Electric Properties Tomography. IEEE Trans Med Imaging. 2015;34:1788–1796. doi: 10.1109/TMI.2015.2404944. [DOI] [PubMed] [Google Scholar]
- 60.Mandija S, Sbrizzi A, Katscher U, Luijten PR, van den Berg CAT. Error analysis of helmholtz-based MR-electrical properties tomography. Magn Reson Med. 2017;80:90–100. doi: 10.1002/mrm.27004. [DOI] [PubMed] [Google Scholar]
- 61.Schmidt R, Webb A. A new approach for electrical properties estimation using a global integral equation and improvements using high permittivity materials. J Magn Reson. 2016;262:8–14. doi: 10.1016/j.jmr.2015.11.002. [DOI] [PubMed] [Google Scholar]
- 62.Serralles J, Polimeridis A, Vaidya M, Haemer G, White J, Sodickson D, Daniel L, Lattanzi R. Proc Intl Soc Mag Reson Med. 24. Singapore: 2016. Global Maxwell Tomography: a novel technique for electrical properties mapping without symmetry assumptions or edge artifacts; p. 2993. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.