Highlights
-
•
We clarify the equivalence of body composition analysis from computed tomography images using two different software packages.
-
•
Analysis was performed using SliceOmatic and OsiriX packages on 50 patients who had undergone triphasic scans.
-
•
Body composition measures were significantly different between the two software packages, but the clinical significance of these is doubtful.
-
•
We recommend that for serial body composition analysis and for comparative purposes, the software package employed should be consistent.
Keywords: Computed tomography, Body composition, Sarcopenia, Myosteatosis, OsiriX, SliceOMatic
Abstract
Objectives
The analysis of body composition from computed tomography (CT) imaging has become widespread. However, the methodology used is far from established. Two main software packages are commonly used for body composition analysis, with results used interchangeably. However, the equivalence of these has not been well established. The aim of this study was to compare the results of body composition analysis performed using the two software packages to assess their equivalence.
Methods
Triphasic abdominal CT scans from 50 patients were analyzed for a range of body composition measures at the third lumbar vertebral level using OsiriX (v7.5.1, Pixmeo, Switzerland) and SliceOmatic (v5.0, TomoVision, Montreal, Canada) software packages. Measures analyzed were skeletal muscle index (SMI), fat mass (FM), fat-free mass (FFM), and mean skeletal muscle Hounsfield Units (SMHU).
Results
The overall mean SMI calculated using the two software packages was significantly different (SliceOmatic 51.33 versus OsiriX 53.77, P < 0.0001), and this difference remained significant for non-contrast and arterial scans. When FM and FFM were considered, again the results were significantly different (SliceOmatic 33.7 versus OsiriX 33.1 kg, P < 0.0001; SliceOmatic 52.1 versus OsiriX 54.2 kg, P < 0.0001, respectively), and this difference remained for all phases of CT. Finally, when analyzed, mean SMHU was also significantly different (SliceOmatic 32.7 versus OsiriX 33.1 HU, P = 0.046).
Conclusions
All four body composition measures were statistically significantly different by the software package used for analysis; however, the clinical significance of these differences is doubtful. Nevertheless, the same software package should be used if serial measurements are being performed.
Introduction
Computed tomography (CT) analysis of body composition to measure fat mass (FM) and fat-free mass (FFM), to calculate skeletal muscle index (SMI), and to diagnose sarcopenia and myosteatosis has become increasingly common, with literature now linking sarcopenia and myosteatosis with reduced overall survival [1], [2], decreased tolerance to chemotherapy [3], [4], and increased complications [5], [6] after surgery in patients presenting with various types of malignancy.
However, the methodology for calculating body composition from CT images is variable between studies, from the nature of the CT scan used including the vertebral level, to the use of contrast medium, to the software used to perform the analysis. The effects of the use of contrast medium in CT scanning in body composition analysis has previously been recognized to have a significant effect on results, especially the diagnosis of myosteatosis [7], [8]. Despite these inconsistencies in analysis, the results of these studies are used interchangeably, with the definition of neither sarcopenia or myosteatosis stipulating any conditions about how these derived values are calculated.
There are currently two software packages used commonly to analyze body composition from CT: SliceOmatic (TomoVision, Montreal, Canada) and OsiriX (Pixmeo, Switzerland), the results of which are also used interchangeably. One study in patients with rectal cancer [9] suggested that SliceOmatic, ImageJ (National Institutes of Health, Bethesda, MD, USA), FatSeg (Biomedical Imaging Group Rotterdam of Erasmus MC, Rotterdam, The Netherlands, using MeVisLab [Mevis Medical Solutions, Bremen, Germany]) and OsiriX analysis provide excellent levels of agreement. However, the study [9] did not consider mean skeletal muscle Hounsfield Unit (SMHU) as a surrogate for myosteatosis. The aim of the present study was to compare the SliceOmatic and OsiriX software packages and determine if there was a difference in calculated measures of body composition, namely SMI, FM, FFM and mean SMHU, using CT images.
Methods
In a single-center retrospective study, CTs from 50 patients who underwent triple-phase abdominal scans (non-contrast, arterial, and portovenous phases) between April 2014 and September 2015 were analyzed using two different software packages: SliceOmatic v5.0 and OsiriX v7.5.1. The patients were initially identified retrospectively from the Computerised Radiology Information System (CRIS v 2.09, HSS, Healthcare Systems, Mansfield, UK). The underlying pathology necessitating the CT was variable, and included trauma, suspected intraabdominal or gastrointestinal bleeding, pancreatic or hepatic pathology, and renal lesions. Three axial slices were selected from each triphasic abdominal CT (total analyzed slices in the study = 50 × 3 = 150 slices). Each slice was anatomically localized using coronal and sagittal multiplanar reformats (MPRs) to ensure it specifically lies at the third lumbar vertebra (L3). Slices were analyzed as Digital Imaging and Communication in Medicine (DICOM) images obtained from a picture archiving and communication system (PACS). Electronic data were collated for patient demographic characteristics, including height and weight data from within 1 mo of the date of the CT.
Scan acquisition
During the study period, two CT scanners were in use at Nottingham University Hospitals NHS Trust where the study was conducted. The first was Ingenuity 128 (Phillips Healthcare, Best, The Netherlands) and the second was Optima CT660 (GE Healthcare, Milwaukee, WI, USA). These were calibrated weekly to ensure that quality assurance testing was met for the Hounsfield Unit (HU) density of air (HU = –1000) and water (HU = 0). Arterial and portovenous phase scans were obtained using intravenous (IV) administration of contrast medium (100 mL fixed dose of Iopamidol, Niopam 300, Bracco, Buckinghamshire, UK). The timings of different phase scans were standardized, first with an unenhanced scan, then the arterial phase performed at 10 to 20s, and finally the portovenous scan at 65s.
Body composition analysis
The three phases of CT slice on each individual patient were analyzed by a single observer, our group having previously established high rates of interobserver reliability (SMI r2 = 0.975, P < 0.0001; mean SMHU r2 = 0.965, P < 0.0001) in the analysis of body composition variables using the techniques adopted in this study [7]. The software packages, SliceOmatic and OsiriX, were each used to calculate the cross-sectional area of skeletal muscle, visceral and subcutaneous/intramuscular adipose tissue. The different tissue types were identified by their differing radiodensities; skeletal muscle of –29 to +150 HU, visceral adipose of –150 to –50 HU and subcutaneous/intramuscular adipose of –190 to –30 HU. The mean SMHU density was also recorded for all scans analyzed.
Previously described regression equations for the calculation of whole-body FM and FFM from a single cross-sectional CT slice were used [10]:
The cross-sectional area of skeletal muscle was also transformed into the SMI by modifying it by patient height.
Statistical analysis
Statistical analysis was performed using SPSS v 22 (IBM, SPSS Statistics, Armonk, NY, USA) and GraphPad Prism v 6.0 (GraphPad, La Jolla, CA, USA). FM, FFM, SMI, and mean SMHU density values, with data checked for normality using the D'Agostino–Pearson normality test. Data were compared between different software packages using the Student's paired t test when normality was confirmed, and the Wilcoxon matched-pairs signed rank test when the data were not distributed normally. Pearson's coefficient of correlation was used to compare the body composition values calculated from the two different software packages and Bland-Altman plots used to reveal any systematic error between the analyses. All analyses were performed using two-tailed testing with a significance level set at P < 0.05.
Results
Of the 50 patients included during the study period from April 2014 to September 2015, there were 33 men and 17 women, with a mean body mass index (BMI) of 30.4 kg/m2 (SD 4).
Skeletal muscle index
Analysis of body composition by OsiriX gave a significantly greater value for SMI than scans analyzed using SliceOmatic (53.8 versus 51.3 cm2/m2, P < 0.0001) on Wilcoxon matched-pairs signed rank test, performed according to the D'Agostino–Pearson test, demonstrating a lack of normality in the data from OsiriX analysis (K2 = 7.831, P = 0.012). This difference remained between scans analyzed in non-contrast and arterial phases; however, there was no difference in scans analyzed in the portovenous phase (Table 1).
Table 1.
Non-contrast phase scan | Arterial phase scan | Portovenous phase scan | |
---|---|---|---|
Skeletal muscle index (cm2/m2) ± SD | |||
SliceOmatic | 51 ± 10.1 | 51.4 ± 10.1 | 51.6 ± 9.9 |
OsiriX | 53.3 ± 10.4 | 53.6 ± 11.1 | 54.4 ± 10.7 |
Mean difference between modalities | –2.3 ± 2.2 | –2.2 ± 3.3 | –2.7 ± 3 |
P-value | <0.0001 | <0.0001 | 0.189 |
Fat mass (kg) | |||
SliceOmatic | 34.1 ± 9.1 | 33.7 ± 8.9 | 33.5 ± 9 |
OsiriX | 33.4 ± 9 | 33 ± 8.7 | 32.8 ± 9 |
Mean difference between modalities | 0.7 ± 0.6 | 0.7 ± 0.8 | 0.7 ± 0.5 |
P-value | <0.0001 | <0.0001 | <0.0001 |
Fat-free mass (kg) | |||
SliceOmatic | 51.8 ± 11.3 | 52.1 ± 11.3 | 52.3 ± 11.3 |
OsiriX | 53.9 ± 11.7 | 54.1 ± 12.1 | 54.8 ± 11.9 |
Mean difference between modalities | –2.1 ± 2 | –2 ± 2.9 | –2.4 ± 2.7 |
P-value | <0.0001 | <0.0001 | <0.0001 |
Mean Skeletal Muscle Hounsfield Units | |||
SliceOmatic | 30.1 ± 9.3 | 33 ± 9.9 | 35.4 ± 10.2 |
OsiriX | 30.6 ± 8.6 | 32.7 ± 9.4 | 35.7 ± 10 |
Mean difference between modalities | –0.5 ± 2.2 | 0.3 ± 2.1 | –0.2 ± 2.4 |
P-value | 0.120 | 0.213 | 0.450 |
There was a significant positive correlation in SMI between analysis conducted using OsiriX and SliceOmatic software (r = 0.965, P < 0.0001) and evidence of a positive systematic bias on Bland -Altman testing (average bias = 2.432; Fig. 1).
Fat mass
FM calculated by OsiriX was significantly lower than that calculated by SliceOmatic (33.1 versus 33.7 kg, P < 0.0001) as calculated by the Student's paired t test, as the data were demonstrated to be normally distributed, and this difference was seen when all individual phase data were analyzed (Table 1).
The correlation between FM analysis using OsiriX and SliceOmatic was significant (r = 0.997, P < 0.0001) and Bland-Altman testing revealed no evidence of a systematic bias (average bias = –0.680; Fig. 2).
Fat-free mass
Analysis of FFM using the two software packages demonstrated significantly greater values with OsiriX analysis than with SliceOmatic (54.2 versus 52.1 kg, P < 0.0001) as calculated by the Student's paired t test as the data were demonstrated to be normally distributed. This finding remained consistent in slices analyzed in non-contrast, arterial, and portovenous phases (Table 1).
There was a significant positive correlation between analysis of FFM performed using OsiriX and SliceOmatic software packages (r = 0.977, P < 0.0001) and there was evidence of a systematic bias on Bland-Altman testing (average bias = 2.16; Fig. 3).
Mean skeletal muscle Hounsfield units
The mean SMHU density was overall significantly higher when analyzed using OsiriX rather than SliceOmatic software (33.1 versus 32.7 HU, P = 0.046) as calculated by the Student's paired t test as the data were demonstrated to be normally distributed. However, when the individual phases of CT were compared, there were no significant differences between OsiriX and SliceOmatic (Table 1).
There was a significant positive correlation in the mean SMHU between the two software packages (r = 0.976, P < 0.0001) and no evidence of any systematic bias (average bias = 0.360; Fig. 4).
Discussion
This study provides evidence of the relative clinical equivalence of analysis of body composition measures analysed by two different software packages, namely OsiriX and SliceOmatic. However, statistically significantly greater SMI, FFM, and mean SMHU values and significantly lower FFM were demonstrated when the analyses were performed with OsiriX compared with SliceOmatic. There was significant positive correlation for all measures when the two software packages were compared, although Bland-Altman testing revealed evidence of a significant systematic bias when analyzing SMI and FFM. The results of the present study are similar to those of the previously published comparison of OsiriX, SliceOmatic, ImageJ, and FatSeg [9], which found that body composition in terms of cross-sectional muscle area, visceral adipose tissue area, and subcutaneous adipose tissue area had excellent levels of agreement, suggesting that the results of analysis with the different software packages could be used interchangeably. However, this study suggested evidence of a systematic bias in the analysis of SMI and FFM, which should be considered when comparing results of body composition analysis performed using different software packages. The previous study [9], however, did not include myosteatosis, as calculated by the mean SMHU value, which is being increasingly used in body composition analysis. Additionally, the present study considered the different phases of abdominal CT (non-contrast, arterial, and portovenous), which were not considered by the previous literature; indeed no statement is made regarding the phase of CT scan considered by the previous study [9].
Although the results of the present study demonstrate statistically significant differences in body composition variables by software package used for analysis, the clinical significance of several of these outcomes is doubtful. The mean SMHU was different by just 0.4 HU, much less than the difference in SMHU between different phases of CT scan (in OsiriX analysis a difference of 5.1 HU was seen between non-contrast and portovenous scans and 5.3 HU in SliceOmatic analysis). This discrepancy in radiodensity of skeletal muscle has been documented previously [7] and its clinical relevance questioned. Therefore, with such a small difference, this is very unlikely to have a significant effect on the diagnosis of myosteatosis. Similarly, the difference between software packages was minimal in FM analysis, with an overall difference of 0.7 kg, which represents just 1.8% of the overall mass from OsiriX analysis. The difference was more pronounced in SMI and FFM analysis, with a difference of 2.5 cm2/m2 (4.6%) and 2.1 kg (3.9%) respectively, which are more likely to represent a clinically relevant difference. This difference in body composition variables has not been demonstrated previously, and the results of body composition analysis using OsiriX and SliceOmatic software packages are used interchangeably within the literature.
This study was conducted retrospectively. However, all scans were performed on individual patients at the same time, so although the hydration status was not known, it would be consistent for all scans and, therefore, would not have an effect on these results. Height and weight data were not always available from the date of the scan, which may render the calculation of body composition measures less accurate.
Further work on body composition analysis is necessary to standardize the methodology used to calculate clinical body composition outcomes including the presence of sarcopenia and myosteatosis. This should include muscle biopsy samples of the rectus abdominis at the L3 vertebral level to correlate radiologic and histologic analysis of skeletal muscle.
To our knowledge, this is the first study to investigate the analysis of body composition variables including myosteatosis by software package of analysis, and has demonstrated statistically significant differences in values in all outcomes. Although some statistically significant differences were demonstrated between the two software packages, these are unlikely to be clinically relevant. However, given the demonstrable differences in body composition measures, it is suggested that the two packages should not be used interchangeably for clinical or research purposes.
Footnotes
This work was supported by the Medical Research Council (grant number MR/K00414 X/1), Arthritis Research UK (grant number 19891). The funders had no role in the design, execution, or write up of the study. KER was funded by a research fellowship from the European Society for Clinical Nutrition and Metabolism (ESPEN). IAM has received research funding from Mars Inc. and serves on the advisory board of IKEA for unrelated work. DNL has received unrestricted research funding and speaker's honoraria from Fresenius Kabi, BBraun, and Baxter Healthcare for unrelated work. He has also served on advisory boards for Baxter Healthcare and AbbVie in the past. The authors have no conflicts of interest to declare.
References
- 1.Rollins KE, Tewari N, Ackner A, Awwad A, Madhusudan S, Macdonald IA. The impact of sarcopenia and myosteatosis on outcomes of unresectable pancreatic cancer or distal cholangiocarcinoma. Clin Nutr. 2016;35:1103–1109. doi: 10.1016/j.clnu.2015.08.005. [DOI] [PubMed] [Google Scholar]
- 2.Tan BH, Birdsell LA, Martin L, Baracos VE, Fearon KC. Sarcopenia in an overweight or obese patient is an adverse prognostic factor in pancreatic cancer. Clin Cancer Res. 2009;15:6973–6979. doi: 10.1158/1078-0432.CCR-09-1525. [DOI] [PubMed] [Google Scholar]
- 3.Choi Y, Oh DY, Kim TY, Lee KH, Han SW, Im SA. Skeletal muscle depletion predicts the prognosis of patients with advanced pancreatic cancer undergoing palliative chemotherapy, independent of body mass index. PLoS One. 2015;10 doi: 10.1371/journal.pone.0139749. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Miyamoto Y, Baba Y, Sakamoto Y, Ohuchi M, Tokunaga R, Kurashige J. Negative impact of skeletal muscle loss after systemic chemotherapy in patients with unresectable colorectal cancer. PLoS One. 2015;10 doi: 10.1371/journal.pone.0129742. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Levolger S, van Vugt JL, de Bruin RW, JN IJ. Systematic review of sarcopenia in patients operated on for gastrointestinal and hepatopancreatobiliary malignancies. Br J Surg. 2015;102:1448–1458. doi: 10.1002/bjs.9893. [DOI] [PubMed] [Google Scholar]
- 6.Joglekar S, Nau PN, Mezhir JJ. The impact of sarcopenia on survival and complications in surgical oncology: a review of the current literature. J Surg Oncol. 2015;112:503–509. doi: 10.1002/jso.24025. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Rollins KE, Javanmard-Emamghissi H, Awwad A, Macdonald IA, Fearon KCH, Lobo DN. Body composition measurement using computed tomography: does the phase of the scan matter? Nutrition. 2017;41:37–44. doi: 10.1016/j.nut.2017.02.011. [DOI] [PubMed] [Google Scholar]
- 8.Boutin RD, Kaptuch JM, Bateni CP, Chalfant JS, Yao L. Influence of IV contrast administration on CT measures of muscle and bone attenuation: implications for sarcopenia and osteoporosis evaluation. AJR Am J Roentgenol. 2016;207:1046–1054. doi: 10.2214/AJR.16.16387. [DOI] [PubMed] [Google Scholar]
- 9.van Vugt JL, Levolger S, Gharbharan A, Koek M, Niessen WJ, Burger JW. A comparative study of software programmes for cross-sectional skeletal muscle and adipose tissue measurements on abdominal computed tomography scans of rectal cancer patients. J Cachexia Sarcopenia Muscle. 2017;8:285–297. doi: 10.1002/jcsm.12158. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Mourtzakis M, Prado CM, Lieffers JR, Reiman T, McCargar LJ, Baracos VE. A practical and precise approach to quantification of body composition in cancer patients using computed tomography images acquired during routine care. Appl Physiol Nutr Metab. 2008;33:997–1006. doi: 10.1139/H08-075. [DOI] [PubMed] [Google Scholar]