Optimisation of Milk Protein Top-Down Sequencing Using In-Source Collision-Induced Dissociation in the Maxis Quadrupole Time-of-Flight Mass Spectrometer

Delphine Vincent 1,*, Dominik Mertens 2, Simone Rochfort 1,3
Editor: Paolo Iadarola
PMCID: PMC6278275  PMID: 30373172

Abstract

Top-down sequencing in proteomics has come of age owing to continuous progress in LC-MS. With their high resolution and broad mass range, Quadrupole Time-of-Flight (Q-ToF) hybrid mass spectrometers equipped with electrospray ionisation source and tandem MS capability by collision-induced dissociation (CID) can be employed to analyse intact proteins and retrieve primary sequence information. To our knowledge, top-down proteomics methods with Q-ToF have only been evaluated using samples of relatively low complexity. Furthermore, the in-source CID (IS-CID) capability of Q-ToF instruments has been under-utilised. This study aimed at optimising top-down sequencing of intact milk proteins to achieve the greatest sequence coverage possible from samples of increasing complexity, assessed using nine known proteins. Eleven MS/MS methods varying in their IS-CID and conventional CID parameters were tested on individual and mixed protein standards as well as raw milk samples. Top-down sequencing results from the nine most abundant proteoforms of caseins, alpha-lactalbumin and beta-lactoglubulins were compared. Nine MS/MS methods achieved more than 70% sequence coverage overall to distinguish between allelic proteoforms, varying only by one or two amino acids. The optimal methods utilised IS-CID at low energy. This experiment demonstrates the utility of Q-ToF systems for top-down proteomics and that IS-CID could be more frequently employed.

Keywords: top-down proteomics, HPLC-ESI-Q-TOF MS, cow’s milk, whey proteins and caseins, tandem MS

1. Introduction

Top-down proteomics, a term invented by Kelleher and colleagues 20 years ago [1], describes the analysis of intact proteins, either in their native form or more often in a denatured state, which allows for a characterisation of proteoforms as comprehensively as possible. Coined in 2014, the term proteoform “designates all of the different molecular forms in which the protein product of a single gene can be found, encompassing all forms of genetic variation, alternative splicing of RNA transcripts, and post-translational modifications (PTMs)” [2]. The analysis of intact proteins is now always performed using mass spectrometry (MS) and technical progresses in top-down proteomics are tightly linked to improvements made on mass analysers. The most important technological advance was the coupling of the soft ionisation technique, electrospray ionisation (ESI), to a mass spectrometer and the production of gas phase ions from large molecules [3], which led to its first application to intact proteins [4,5]. ESI generates multiply-charged protein ions of low m/z, thereby allowing the analysis of very large molecules, even on an MS platform with a limited mass range. By applying a deconvolution algorithm [6], the resulting complex multi-peak spectrum can be converted into a single peak corresponding to the true molecular weight of the molecule. Further developments in intact protein analysis quickly followed with the introduction of complex multistage mass analysers of high resolution such as linear ion trap (LIT), quadrupole time-of-flight (Q-ToF), Fourier transform (FT) ion cyclotron resonance, and orbitrap instruments.

Today, various methods of ion fragmentation are available, such as collision-induced dissociation (CID, also called collisionally activated dissociation CAD), in-source CID (IS-CID, also known as up-front CID, cone-voltage CID, and nozzle-skimmer dissociation), higher-energy collisional dissociation (HCD), electron capture dissociation (ECD), and electron transfer dissociation (ETD) (for review [7]). Different types of fragmentation modes yield different information about the structure and composition of the analyte. Tandem mass spectrometers that carry out CID to generate product ions from precursor ions have proven extremely useful for the identification and characterisation of proteins from a complex mixture. CID was first described by McLafferty and Bryce in 1967 [8] and Jennings in 1968 [9]. Parent ions collide with neutral gas atoms or molecules (typically helium, nitrogen or argon), which result in the formation of b- and y-type ions. The efficacy of a CID experiment will depend on the relative translational energy of the ion and target, the nature of the target, the number of collisions that is likely to take place, and the m/z window of the instrument. Initial top-down sequencing experiments exploited low-energy CID methods to induce protein fragmentation. However, for large molecules, CID does not produce a fragmentation pattern comprehensive enough to fully characterise proteins, but rather produces enough fragments or sequence tags to identify the protein. Additionally, if the protein contains PTMs, low-energy CID most likely will not be sufficient to localise the modified site or the PTM may be the preferred site of cleavage [10,11]. IS-CID is a proven albeit seldom utilised fragmentation mode allowing single-stage instrumentation, such as single quadrupole (Q) or ToF mass analysers, to produce spectra similar to those obtained with far more expensive hybrid instruments. In this method, invented by Katta and colleagues in 1991 [12], CID is carried out within the ion source in the high-pressure region between the capillary exit and the skimmer entrance to the Q mass spectrometer. Because there is no prior selection of the precursor ion, this does not qualify as a bona fide MS/MS experiment. By manipulating lens voltages that channel ions from the source to the mass analyser, relatively low-energy ions formed in the atmospheric pressure region of the source collide with residual background gas, usually nitrogen, in the transition region. As a result, excited ions can undergo unimolecular decomposition to produce fragment ions. As for CID, only b- and y-type ions are observed. Tandem spectra from protonated peptides produced by IS-CID or low-energy CID are comparable [13,14]. In-source dissociation has also been demonstrated for whole protein ions [15]. The time scale for IS-CID is on the order of a few hundred microseconds to a few milliseconds, which is much faster than that of CID. IS-CID fragmentation can be reproducible provided ion source parameters, such as temperature, pressure, voltage, and sample purity are tightly controlled [16].

Kelleher in 2004 [17] rigorously defines top-down proteomics as a multistep process whereby the molecular weights (MWs) of intact proteoforms are accurately measured using a high-resolution mass analyser in combination with a direct fragmentation of the protein ions using tandem MS. Top-down sequencing of intact proteins were initially performed using triple quadrupole (QQQ) instruments and IS-CID [5]. In 1990, Loo and colleagues characterised a 14-kDa bovine ribonuclease A in its native and reduced forms using CID [18]. Soon after, higher mass resolution was achieved by applying FT-MS to equine cytochrome c, porcine albumin, thioredoxin and ubiquitin [19], myoglobin [20], and carbonic anhydrase [15]. A hybrid Q-ToF instrument, also called Q-ToF [21], was invented in 1996 by Morris and colleagues [22]. Q-ToF mass spectrometers combine the quadrupole one (Q1) in which ion precursors are funnelled through and selected, the quadrupole collision cell (q) of a QQQ in which selected precursor ions are fragmented usually through CID and more recently ETD, and a reflector ToF detector in which the fragment ions are mass analysed. Spectra obtained in both full-scan (MS1) and MS/MS (MS2) modes exhibit clean spectra due to orthogonal pulsing into the ToF section, high mass accuracy along with stable isotopic resolution across the m/z range, and high fidelity meaning accurate isotopic ratios. This permits the determination of charge states and unambiguous assignment of the mono-isotopic signal of the intact molecules. This leads to accurate prediction of AA sequence and successful identification of peptides and proteins of medium MW via database searches by tightening the search parameters and augmenting the confidence in the results. Q-ToF instruments perform well for quantitative analyses [23] and for the identification of PTMs [24]. Q-ToF were demonstrated to outperform QQQ mass spectrometers owing to both their enhanced sensitivity and resolution in the region of one to two orders of magnitude [22]. This initial observation was further validated on phosphopeptides; the higher resolving power of the Q-ToF improved the selectivity and sensitivity of parent ions, thus minimising interference from other product ions and maximising sequencing results [25].

The first use of LC-ESI-Q-ToF MS in top-down sequencing of intact proteins was reported by Nemeth-Cawley and Rouse in 2002 [26], who used CID to fragment ion precursors from eight known protein standards, ranging from 5 to 66 kDa. Peptide sequence tags thus obtained led to the unambiguous identification of the analysed proteins, along with the characterisation of disulphide bonds and glycosylation. While their method was successful, the authors noted some limitations: first, it could not yet be applied to complex protein mixtures without prior separation; secondly, at a low concentration, the signal-to-noise ratio was affected, thus necessitating longer MS/MS acquisition times. Based on their experience, the following year, the same group applied their validated method to characterise a recombinant immunoglobulin gamma-1 (IgG-1) fusion protein [27]. The top-down strategy allowed them to demonstrate that the recombinant protein was expressed as a full-length form as well as N- and C-processed truncated proteoforms [27]. In 2004, Ginter and colleagues exploited both IS-CID and traditional CID mode of a Q-ToF mass spectrometer, thereby achieving pseudo-MS3 levels, to top-down sequence seven known proteins spanning from 11 to 66 kDa [28]. Individual sequence tags of 10 to 26 AAs were retrieved from both N- and C-termini, and unambiguous identification of protein standards was achieved. IS-CID on its own was used to identify various purified recombinant proteins using ESI-Q-ToF MS [29]; fragmentation efficiency depended on cone voltage and y-ions formed predominantly by cleavage on the C-terminal side of nonpolar residues. An unknown 4.9 kDa recombinant peptide was completely sequenced using LC-ESI-Q-ToF MS/MS of 10,000 resolution and traditional CID fragmentation; complete y- and b-type ion series were obtained and the formation of beta-mercaptoethanol adducts was reported [30]. In 2009 Armirotti and colleagues reported 90–100% sequence coverage of horse myoglobin (17 kDa) and bovine carbonic anhydrase II (30 kDa) using LC-ESI-Q-ToF MS/MS and CID fragmentation mode [31]. They could also identify an unknown protein as superoxide dismutase (16 kDa) and locate one acetylation site. ESI CID Q-TOF MS/MS was used to identify eight intact antimicrobial peptides from Asian frog skin and locate disulphide bridges [32]. ESI-Q-ToF technology has also been employed to investigate the precise stoichiometry of protein assemblies, the interactions between subunits and the position of subunits within the complex [33].

In our laboratory, we host a maXis HD UHR-Q-ToF (60,000 resolution) with an ESI source on-line with a UHPLC 1290 Infinity Binary LC system (Agilent, Mulgrave, VIC, Australia), which we have used to develop a top-down method to analyse cow’s milk proteins [23]. This method was further implemented with CID MS/MS analysis and applied to the study of UHT milk shelf life [34]. The present study aimed at optimising MS/MS analysis by testing the effect of some of the parameters pertaining to IS-CID in combination or not with conventional CID. A total of 11 MS/MS methods were assessed on samples bearing increased complexity, namely individual milk protein standards, mixed protein standards and cow’s raw milk samples from Jersey or Holstein breeds. Milk proteins exhibit numerous PTMs, however in this initial study, we focused on optimising protein fragmentation to achieve the greatest sequence coverage possible. To this end, nine of the most abundant proteoforms of milk caseins and whey proteins, including genetic variants, were compared across samples and MS/MS methods whose efficacy was assessed based on protein sequence coverage alone. This necessitated developing processing workflows using the powerful Genedata Expressionist software to thoroughly explore and annotate the MS1 and MS2 files.

2. Results and Discussion

2.1. Sample Complexity

In this experiment, we wanted to test whether sample complexity would impact the quality of spectral acquisition. To this end, we prepared a set of samples displaying increasing proteome complexity: from simple individual milk protein standards, albeit varying in their purity level from 70% (α-CN) to 98% (β-CN), to a mixture of these standards, to the highly complex biological matrices raw milk samples that were obtained from two common cow breeds. Figure 1 describes the experimental design of this study.

Figure 1.

Figure 1

Schematic diagram of the experimental design of the study.

2.1.1. UPLC-ESI-Q-ToF MS Can Handle Complex Samples

While the raw data clearly illustrate increasing complexity with more LC-MS1 peaks detected in the milk samples compared to the individual protein standard samples (Figure 2A), the latter are more complex than first anticipated, particularly the β-CN standard, which is claimed to be 98% pure.

Figure 2.

Figure 2

LC-MS maps of the different samples (individual protein standards, mixed protein standards, Jersey milk, and Holstein milk) visualised in Genedata Refiner pre- and post-deconvolution with m/z on the x axis and LC retention time in min on the y axis. (A) LC-MS1 unprocessed raw data. (B) Deconvoluted data displaying proteins of interest in the red boxed areas. The total numbers of deconvoluted peaks detected are indicated in yellow. (C) Close-ups of the boxed areas of the individual protein standards; the nine proteoforms highlighted in red are the proteins targeted for method validation.

These standards were analysed using a bottom-up shotgun approach and yielded many protein hits [35]. The α-LA standard sample (85% pure) generated by far the simplest LC-MS1 pattern, with clear elution from 14 to 19 min and distinct m/z peaks of the charged envelope (Figure 2A).

The LC-MS1 pattern of the mixed standard sample is quite similar and as complex as the patterns of milk samples owing to the fact that the individual protein standards are far from pure. Indeed, following protein mass deconvolution and the display of protein accurate masses separated by HPLC (Figure 2B), many deconvoluted peaks appear besides the expected proteins of interest. Such a display confirms that the α-LA standard is the least complex sample with only 211 deconvoluted peaks, whereas the other individual standards display between 2194 (α-CN) and 3372 peaks (β-CN).

With 4819 deconvoluted peaks, the mixed standard sample is indeed of intermediate complexity between the individual standards and the raw milk samples, which resolved 6988 and 7554 peaks in Jersey and Holstein cows, respectively. While proteins displayed some degree of co-elution during the HPLC separation, particularly in complex samples, separation along the m/z range could isotopically resolve all proteins of less than 30 kDa, thus yielding deconvoluted monoisotopic masses (also shown in [23]).

2.1.2. The 11 MS/MS Methods Are Assessed Using a Reference Set of Known Proteins

Another aim of our study was to optimise the MS/MS method by modifying key parameters and devise objective tools to compare the different methods. We thought the best comparative approach would be a targeted one relying on a small reference set of known proteins. The proteins of interest are depicted in the boxed areas of Figure 2B which have been zoomed in in Figure 2C. We chose nine prominent proteoforms, some of them displaying allelic variations, in each of the individual standards, namely, α-CN type S1 variant B with 8 phosphorylations (αS1-CN B-8P, 23,600.3 Da), α-CN type S2 variant A with 11 phosphorylations (αS2-CN A-11P, 25,213.0 Da), β-CN variants A1 and A2 (β-CN A1 24,008.2 Da, β-CN A2 23,968.2 Da) each with five phosphorylations, κ-CN variants A and B with 1 phosphorylation (κ-CN A-1P 19,026.5 Da, κ-CN B-1P 18,993.6 Da), α-LA variant B (α-LA B 14,176.8 Da), and β-LG variants A and B (β-LG A 18,355.5 Da, β-LG B 18,269.4 Da). We have previously used these proteins to optimise a quantitative LC-MS method [23]. The full primary sequence of these proteins is known, and therefore fragmentation efficiency for top-down sequencing purpose could be assessed. This is presented in the last chapter. While not included in this article, our MS/MS method also allowed us to identify various PTMs of milk most abundant proteins such as lactosylation (+324 Da), oxidation (+16 Da), glycosylation ([Hex(1)HexNAc(1)NeuAc(2)] + 947 Da), and degradation products, as reported in [34].

2.2. Effect of IS-CID and CID on Their Own or Combined on Spectral Data

In a recent study [34], we have used conventional CID (Method 2 in the present study) to identify the degradation products of milk’s most abundant proteins occurring following UHT treatment and storage on the shelf. While this top-down sequencing method proved successful for the identification of small proteins and degradation products, we wanted to optimise it for the analysis of intact milk proteins. Our Q-ToF mass spectrometer allows us to perform both IS-CID and CID in a single experiment. The benefits of both have been stated in the introduction of this manuscript. In the present study, we have only slightly varied the parameters controlling CID fragmentation, choosing to operate it in a fully automated manner. We have mostly fine-tuned the parameters pertaining to IS-CID. We thus tested different energy values for IS-CID MS (ion funnel 1 exit) and IS-CID MS/MS (ion funnel 2 entrance) parameters. A total of 11 MS/MS methods were thus devised that employed either CID or IS-CID on their own or combined (Table 1).

Table 1.

Parameters applied on the Q-TOF mass spectrometer for the 12 MS methods.

Mode Step Parameter Method 1 Method 2 Method 3 Method 4 Method 5 Method 6 Method 7 Method 8 Method 9 Method 10 Method 11 Method 12
MS Source End plate offset 500 V 500 V 500 V 500 V 500 V 500 V 500 V 500 V 500 V 500 V 500 V 500 V
MS Source Capillary 4500 V 4500 V 4500 V 4500 V 4500 V 4500 V 4500 V 4500 V 4500 V 4500 V 4500 V 4500 V
MS Source Nebuliser 1.5 bar 1.5 bar 1.5 bar 1.5 bar 1.5 bar 1.5 bar 1.5 bar 1.5 bar 1.5 bar 1.5 bar 1.5 bar 1.5 bar
MS Source Dry gas 8 L/min 8 L/min 8 L/min 8 L/min 8 L/min 8 L/min 8 L/min 8 L/min 8 L/min 8 L/min 8 L/min 8 L/min
MS Source Temperature 190 °C 190 °C 190 °C 190 °C 190 °C 190 °C 190 °C 190 °C 190 °C 190 °C 190 °C 190 °C
MS Source Waste (min) 0–2.5 0–2.5 0–2.5 0–2.5 0–2.5 0–2.5 0–2.5 0–2.5 0–2.5 0–2.5 0–2.5 0–2.5
MS Source Source (min) 2.5–40 2.5–40 2.5–40 2.5–40 2.5–40 2.5–40 2.5–40 2.5–40 2.5–40 2.5–40 2.5–40 2.5–40
MS Transfer Funnel RF 400 Vpp 400 Vpp 400 Vpp 400 Vpp 400 Vpp 400 Vpp 400 Vpp 400 Vpp 400 Vpp 400 Vpp 400 Vpp 400 Vpp
MS Transfer IS-CID energy 0 eV 0 eV 0 eV 3 eV 30 eV 50 eV 70 eV 5 eV 0 eV 3 eV 0 eV 10 eV
MS Transfer Transfer multipole RF 400 Vpp 400 Vpp 400 Vpp 400 Vpp 400 Vpp 400 Vpp 400 Vpp 400 Vpp 400 Vpp 400 Vpp 400 Vpp 400 Vpp
MS Quadrupole Ion energy 5 eV 5 eV 5 eV 5 eV 5 eV 5 eV 5 eV 5 eV 5 eV 5 eV 5 eV 5 eV
MS Quadrupole Low mass 300 m/z 300 m/z 300 m/z 300 m/z 300 m/z 300 m/z 300 m/z 300 m/z 300 m/z 300 m/z 300 m/z 300 m/z
MS Collision cell Collision energy 10 eV 10 eV 10 eV 10 eV 10 eV 10 eV 10 eV 10 eV 10 eV 10 eV 10 eV 10 eV
MS Collision cell Collision RF (Vpp) 1800 1800 1800 1800 1800 1800 1800 1800 1800 1800 1800 1800
MS Collision cell Transfer time 120 μs 120 μs 120 μs 120 μs 120 μs 120 μs 120 μs 120 μs 120 μs 150 μs 120 μs 120 μs
MS Collision cell Pre-pulse storage 10 μs 10 μs 10 μs 10 μs 10 μs 10 μs 10 μs 10 μs 10 μs 15 μs 10 μs 10 μs
MS MS1 Ion polarity positive positive positive positive Positive Positive positive positive positive positive positive positive
MS MS1 Mass range (m/z) 600–3000 600–3000 600–3000 200–2000 200–2000 200–2000 200–2000 200–2000 200–2000 200–2000 200–2000 200–2000
MS MS1 Summation 15,000 15,000 15,000 18,321 18,321 18,321 18,321 18,321 18,321 18,321 18,321 18,321
MS MS1 Rolling average mode 2 2 2 2 2 2 2 2 2 2 2 2
in source IS-CID IS-CID NO NO NO YES YES YES YES YES YES YES YES YES
in source IS-CID IS-CID MS (funnel 1 exit) 3 eV 30 eV 50 eV 70 eV 5 eV 3 eV 3 eV 3 eV 10 eV
in source IS-CID IS-CID MS/MS (funnel 2 entry) 15 eV 10 eV 10 eV 10 eV 20 eV 10 eV 30 eV 10 eV 15 eV
in source IS-CID Acquisition time factor 1 1 1 1 1 1 1 1 1
MS/MS MS2 Auto MS/MS mode OFF ON ON OFF ON ON ON ON ON ON ON OFF
MS/MS CID Acquisition NO YES YES NO YES YES YES YES YES YES YES NO
MS/MS CID MS/MS low (per 1000 sum.) 104 cts 104 cts 104 cts 104 cts 104 cts 104 cts 104 cts 104 cts 104 cts
MS/MS CID MS/MS high (per 1000 sum.) 106 cts 106 cts 106 cts 106 cts 106 cts 106 cts 106 cts 106 cts 106 cts
MS/MS MS2 Preference activated NO NO NO NO NO NO NO NO YES NO NO NO
MS/MS MS2 Charge state preferred range 7–15
MS/MS MS2 Exclude singly YES
MS/MS Precursor ions Exclude (m/z) 100–600 100–600 100–600 100–600 100–600 100–600 100–600 100–600 100–600
MS/MS Precursor ions No. of precursors 2 2 2 2 2 3 4 4 4
MS/MS Threshold Absolute (per 1000 sum.) 1153 cts 170 cts 10 cts 10 cts 10 cts 10 cts 10 cts 10 cts 10 cts
MS/MS Threshold Absolute 17,302 cts 2550 cts 183 cts 183 cts 183 cts 183 cts 183 cts 183 cts 183 cts
MS/MS Active exclusion ON/OFF ON ON ON ON ON ON ON ON ON
MS/MS Active exclusion Exclude after 3 spectra 3 spectra 1 spectra 1 spectra 1 spectra 1 spectra 1 spectra 1 spectra 1 spectra
MS/MS Active exclusion Release after 5 min 5 min 5 min 5 min 5 min 5 min 5 min 5 min 5 min

In summary, Methods 2 and 3 employed CID only, Methods 4 and 12 employed IS-CID only, and Methods 5–11 employed IS-CID within the ESI source/ion transfer region, followed by CID within the collision cell.

Figure 3 displays the LC-MS patterns for each of the 12 methods using the mixed protein standards sample.

Figure 3.

Figure 3

LC-MS maps of the mixed protein standards sample visualised in Genedata Refiner with the scanned mass range (200–2000 m/z) on the x axis and LC retention time (3–28 min) on the y axis across all 12 methods following data processing. White dots represent true MS/MS (MS2) events using conventional CID fragmentation. AE, active exclusion.

Table 2 compares all 12 methods based on LC-MS and LC-MS/MS observations

Table 2.

Number of spectral peaks, clusters, groups and MS/MS spectra found in the mixed standard sample across all 12 methods using Genedata Refiner workflow.

Method 1 2 3 4 5 6 7 8 9 10 11 12
LC separation (min) 22.5 22.5 22.5 22.5 22.5 22.5 22.5 22.5 22.5 22.5 22.5 22.5
LC separation (s) 1350 1350 1350 1350 1350 1350 1350 1350 1350 1350 1350 1350
Duty cycle (s) 2.7 13.7 13.7 5.6 13.7 13.7 13.8 19.1 24.6 24.6 24.6 5.5
No. of full MS1 scans 493 99 99 243 99 99 98 71 55 55 55 247
No. of precursors 2 2 2 2 2 3 4 4 4
(a) Theor. no. of MS2 events 0 197 198 0 197 198 196 212 219 219 220 0
(b) MS1 peaks (300–3000 m/z) 8728 8402 5184 7141 4509 4415 3692 3321 3171 1816 3115 6614
(c) MS1 peaks (700–3000 m/z) 8711 8387 5184 7124 4503 4415 3599 3040 3169 1628 3113 6599
(d) MS1 peaks (300–700 m/z) 17 15 0 17 6 0 93 281 2 188 2 15
(d)/(b) (%) 0.2 0.2 0.0 0.2 0.1 0.0 2.5 8.5 0.1 10.4 0.1 0.2
MS1 clusters 1069 1044 601 869 551 536 530 654 386 405 371 769
MS1 groups 424 426 240 346 234 231 282 487 170 319 159 296
(e) Observed No. of MS2 spectra 0 56 170 0 196 196 196 210 218 218 218 0
(e)/(a) (%) 28 86 99 99 100 99 99 99 99
(e)/(b) (%) 0.7 3.3 4.3 4.4 5.3 6.3 6.9 12.0 7.0
MS2 m/z range 730–1636 644–1854 643–1955 703–1955 657–1966 609–1913 607–1774 609–1966 634–1928
600–700 m/z 1 3 2 18 3 20 4
700–800 m/z 2 2 5 5 5 14 4 19 8
800–900 m/z 4 12 20 20 15 22 26 27 29
900–1000 m/z 20 53 45 44 21 34 56 36 54
1000–1100 m/z 28 77 53 51 47 23 63 24 57
1100–1200 m/z 5 29 26 35 24 30 18 23
1200–1300 m/z 2 19 21 26 26 20 29 18
1300–1400 m/z 2 3 7 17 14 8 7 2
1400–1500 m/z 2 2 5 4 11 2 10 5
1500–1600 m/z 2 3 4 5 8 3 9 3
1600–1700 m/z 2 6 4 4 7 7 1 8 7
1700–1800 m/z 3 4 3 3 1 2 5 5
1800–1900 m/z 3 4 4 7 7 4 1
1900–2000 m/z 2 2 2 1 2 2

2.2.1. Precursor Intensity Threshold Is a Key Parameter for CID Fragmentation

Methods 2 and 3 (CID only) only differed in the intensity threshold for precursor selection, being almost 7 times higher in Method 2 (2550 counts) relative to Method 3 (17,302 counts) (Table 1). This allows precursor ions of low signal intensity to undergo CID fragmentation. As expected, this did not affect the LC-MS1 data; however, it did result in many more MS/MS events when Method 3 was used. If we consider the mixed standard sample as an example, Method 2 resulted in 8402 ions with only 56 (0.7%) MS2 spectra, whereas Method 3 yielded 5184 peaks and 170 (3.3%) MS2 spectra (Table 2). Consequently, the top-down sequencing efficiency varied between Methods 2 and 3, as will be discussed in the following section.

Charge state preference aside, in this study we did not attempt to optimise CID conditions and took full advantage of the automatic MS/MS mode available on our instrument with default parameters as it proved efficient [34]. Using CID on its own, Nemeth-Cawley and colleagues successfully top-down sequenced unknown intact proteins purified by affinity and size-exclusion chromatography from mammalian cells using an ESI-Q-ToF MS/MS strategy [27].

2.2.2. Ion Funnel 1 Energy Has Little Impact on IS-CID Fragmentation

Methods 4 and 12 (IS-CID only) differed only in the energy applied to the ion funnel 1 exit during the IS-CID MS1 step. Method 4 applied 3 eV and Method 12 applied 10 eV at ion funnel 1 (Table 1). LC-MS patterns displayed only slight changes, with Method 4 yielding 7141 peaks, 869 clusters, and 346 groups, and Method 12 producing 6614 peaks, 769 clusters, and 296 groups (Table 2 and Figure 3). Protein annotation by top-down sequencing was also comparable, as will be described in the next chapter.

2.2.3. Charge State Preference Does Not Influence Precursor Selection

Methods 9 and 11 (IS-CID + CID) only differed in their charge state preferred range. No charge state was specified in Method 11 (Table 1). In Method 9, a preference for the most abundant ions bearing 7‒15 charges was applied to assess whether it would favour the selection of highly protonated ions for CID fragmentation. This parameter had no impact on the selection process of parent ions since both methods generated similar numbers of peaks, clusters and groups and, most importantly, an identical number of MS2 spectra (218) (Table 2). This indicated that the automatic MS2 mode of the Q-ToF aptly selected highly charged protein ions for subsequent CID fragmentation.

2.2.4. A Long Duty Cycle Minimises MS/MS Events

To maximise signal sensitivity and minimise background noise, high summation values were utilised, which resulted in duty cycles that lasted at least 2.7 s (MS1 only—Method 1) (Table 2). Furthermore, these duty cycles were extended upon MS/MS analyses in a method-specific fashion (from 5.5 s in Methods 4 and 12, 13.7 s in Methods 2–3 and 5–6, to 24.6 s in Methods 9–11), which restricted the number of full MS1 scans (Table 2). Consequently, it limited how many ions could be selected as parents for the CID process during the LC run.

Under our LC-MS/MS conditions, protein separation occurred from 2.5 to 25 min (22.5 min or 1350 s) with two to four precursors selected per duty cycle. The maximum number of MS2 events were computed and reported in Table 2. Apart from the CID-only methods, which only realised 28% (56/197 in Method 2) and 86% (170/198 in Method 3) of their potential due to high threshold requirements, all the other methods delivered 99–100% of the expected number of MS2 events (Table 2). However, these numbers are low. Instruments of high resolution offering a faster scanning capability would therefore be extremely advantageous.

In our study, the slow scanning rate was partially alleviated by resorting to IS-CID only (Methods 4 and 12), whose fast fragmentation mode allowed for shorter duty cycles (5.5 s) and up to 247 scans during the 22.5-min separation (Table 2).

2.2.5. High Energy Fragmentation Produces Lower m/z Ions

Overall, low-energy fragmentation was applied at the ion funnel 1 exit during the IS-CID MS mode (3, 5 or 10 eV), except for three methods for which high energy was applied. Methods 5–7 applied 30, 50, and 70 eV, respectively (Table 1). The consequences of high-energy IS-CID become visible upon reaching 70 eV (Method 7), which generates many more ions of less than 700 m/z, indicative of protein fragmentation. For instance, in the mixed standard sample, Method 1 (MS1 only) resulted in 8728 peaks, with the vast majority above 700 m/z (8711, 99.8%). Method 5 (30 eV) produced 4509 peaks in total with only 6 (0.1%) below 700 m/z. Conversely, Method 7 (70 eV) produced 3692 peaks in total, of which 93 (2.5%) were of less than 700 m/z (Table 2).

Similarly, low-energy fragmentation (10 or 15 eV) was applied at the ion funnel 2 entrance during the IS-CID MS/MS mode, except for two methods for which greater energy was delivered. Methods 8 and 10 applied 20 and 30 eV, respectively (Table 1). Increasing the energy level during the IS-CID MS/MS step proved even more effective at fragmenting intact proteins within the ESI source, as more spectral peaks of low m/z were created (Figure 3). For instance, in the mixed standard sample, applying 20 eV (Method 8) or 30 eV (Method 10) resulted in the proportion of peaks below 700 m/z reaching 8.5% (281/3321 peaks) and 10.4% (188/1816 peaks) for Methods 8 and 10, respectively (Table 2). This resulted in ions of lower m/z being selected as precursors for a subsequent CID process. In the mixed standard sample, only three (1.5%) MS2 spectra resulted from precursors of less than 700 m/z using Method 5, while 20 (9.2%) MS2 spectra were acquired from precursors of less than 700 m/z using Method 10 (Table 2). However, as will be reported in the next chapter, such enhanced in-source fragmentation did not benefit the top-down sequencing output.

2.2.6. CID Versus IS-CID

While CID parameters mostly revolve around ion features (e.g., intensity threshold, charge state, number of precursors) that allow for relatively controlled and robust conditions, IS-CID parameters themselves pertain to factors operating globally within the ESI source and the ion transfer region, the most important one being the DC voltage applied, which affects signal sensitivity. Owing to its less controllable nature, the IS-CID fragmentation process is more complex and far less predictable than MS/MS results. Therefore, a prerequisite of IS-CID efficacy is good chromatographic resolution with little or no background contamination [16].

Our ESI-Q-ToF mass spectrometer establishes IS-CID fragmentation not within the ESI source per se but more accurately within the ion transfer area called funnel 2. The two funnel-staged ion transfer region separates the ions from the drying gas and solvent, as well as transfers these ions, with minimal losses, to the quadrupole stage. The first and second funnel stages are separated by a DC plate. By increasing the DC potentials of funnel 1, the ions are accelerated into funnel 2, which in turn activates IS-CID. This ingenious, albeit unpredictable, fragmentation mode was invented 30 years ago and reported to favour highly charged species at lower energetic collision (i.e., cone voltage) akin to conventional CID process [5]. As with CID, as the ions’ m/z increase, the IS-CID efficacy decreases. That was confirmed in our experiment. Parent ions were predominantly sampled from the 800–1300 m/z range, where the most intense ions resolved, and only a handful of precursors arose from 1600 m/z and above (Table 2).

IS-CID in combination with conventional CID on a ESI-Q-ToF system was tested at increasing cone voltage conditions (from 45 to 90 V) by Ginter and colleagues (2004) to top-down sequence a mixture of seven protein standards; they reported that some proteins necessitated higher voltage than others for in-source fragmentation to occur [28]. In a different study where only IS-CID was applied to generate MS2 spectra using a ESI-Q-ToF instrument, source cone voltage varied from 20 to 70 V and was evaluated on a sample mixture of 13 protein standards [29]. The authors also concluded that IS-CID efficient fragmentation greatly depended on cone voltage; in their experience, 40–60 V proved the most optimal range.

2.3. Top-Down Sequencing of Milk Proteins

In the present study, a total of 11 MS/MS methods (Methods 2–12) were compared and the ultimate validation criterium was the coverage depth resulting from top-down sequencing annotation. The nine most abundant proteins known from milk were chosen to systematically assess AA sequence coverage across the 11 methods using Genedata Expressionist program. These proteoforms were αS1-CN B-8P, αS2-CN A-11P, β-CN A1, β-CN A2, κ-CN A-1P, κ-CN B-1P, α-LA B, β-LG A, and β-LG B. Table 3 documents the number of AAs top-down sequenced for each protein from either individual or mixed standards, or milk samples from Jerseys or Holsteins.

Table 3.

Number of AAs top-down sequenced across the 11 MS2 methods from the nine most abundant milk proteins with the individual standards, mixed standards, and milk samples.

Protein Variant PTM Code Mo. Mass (Da) Length (AAs) RT (min) Sample 1 Method 2 Method 3 Method 4 Method 5 Method 6 Method 7 Method 8 Method 9 Method 10 Method 11 Method 12
αS1-CN B 8P αS1-CN B 8P 23,600.3 199 15–18 I 196 196 195 194 195 190 70 195 7 194 195
αS2-CN A 11P αS2-CN A 11P 25,213.0 207 6.8–8.3 I 14 6 83 6 6 6 6 6 0 6 83
β-CN A1 5P β-CN A1 5P 24,008.2 209 21–22 I 204 204 197 205 205 203 197 205 95 205 195
β-CN A2 5P β-CN A2 5P 23,968.2 209 21.5–23 I 202 202 201 203 203 201 186 203 92 203 200
κ-CN A 1P κ-CN A 1P 19,026.5 169 6.5–8.5 I 160 160 165 157 157 156 119 157 31 157 165
κ-CN B 1P κ-CN B 1P 18,993.6 169 8.5–10 I 162 162 157 160 159 157 133 161 59 161 154
α-LA B α-LA B 14,176.8 123 15–18 I 98 97 84 94 100 98 90 97 86 97 83
β-LG A β-LG A 18,355.5 162 22–24 I 162 162 161 155 155 153 153 162 151 162 161
β-LG B β-LG B 18,269.4 162 20–22 I 159 158 159 153 154 149 155 159 146 159 158
αS1-CN B 8P αS1-CN B 8P 23,600.3 199 15–18 M 178 178 192 179 177 173 91 178 52 181 192
αS2-CN A 11P αS2-CN A 11P 25,213.0 207 6.8–8.3 M 121 123 150 111 116 108 52 115 8 115 151
β-CN A1 5P β-CN A1 5P 24,008.2 209 21–22 M 171 172 168 162 167 156 79 163 27 165 174
β-CN A2 5P β-CN A2 5P 23,968.2 209 21.5–23 M 201 201 201 203 201 200 182 202 43 203 201
κ-CN A 1P κ-CN A 1P 19,026.5 169 6.5–8.5 M 108 109 122 106 109 93 41 106 14 107 124
κ-CN B 1P κ-CN B 1P 18,993.6 169 8.5–10 M 122 125 112 120 120 110 33 122 10 121 116
α-LA B α-LA B 14,176.8 123 15–18 M 86 85 96 83 82 77 37 83 21 86 94
β-LG A β-LG A 18,355.5 162 22–24 M 158 157 156 158 158 157 135 157 62 157 155
β-LG B β-LG B 18,269.4 162 20–22 M 111 112 126 106 107 96 42 106 20 111 125
αS1-CN B 8P αS1-CN B 8P 23,600.3 199 15–18 J 194 195 198 195 195 189 137 194 97 195 198
αS2-CN A 11P αS2-CN A 11P 25,213.0 207 6.8–8.3 J 92 140 177 122 110 97 18 114 20 126 177
β-CN A1 5P β-CN A1 5P 24,008.2 209 21–22 J 206 206 205 206 206 205 191 206 110 206 204
β-CN A2 5P β-CN A2 5P 23,968.2 209 21.5–23 J 199 199 201 200 200 199 188 199 106 199 201
κ-CN A 1P κ-CN A 1P 19,026.5 169 6.5–8.5 J 82 132 151 121 107 104 12 110 30 119 154
κ-CN B 1P κ-CN B 1P 18,993.6 169 8.5–10 J 164 164 157 164 163 164 146 164 78 163 157
α-LA B α-LA B 14,176.8 123 15–18 J 120 120 121 120 120 111 62 120 38 120 121
β-LG A β-LG A 18,355.5 162 22–24 J 154 154 153 154 154 154 139 154 73 154 153
β-LG B β-LG B 18,269.4 162 20–22 J 158 156 156 158 156 154 134 156 76 156 155
αS1-CN B 8P αS1-CN B 8P 23,600.3 199 15–18 H 191 191 197 183 184 177 133 185 89 183 197
αS2-CN A 11P αS2-CN A 11P 25,213.0 207 6.8–8.3 H 172 181 192 179 179 172 65 178 88 180 193
β-CN A1 5P β-CN A1 5P 24,008.2 209 21–22 H 204 204 200 204 204 204 181 204 170 204 201
β-CN A2 5P β-CN A2 5P 23,968.2 209 21.5–23 H 206 207 203 207 207 207 198 207 168 207 203
κ-CN A 1P κ-CN A 1P 19,026.5 169 6.5–8.5 H 147 150 165 147 153 133 54 151 72 150 165
κ-CN B 1P κ-CN B 1P 18,993.6 169 8.5–10 H 89 97 118 91 86 82 11 89 17 95 118
α-LA B α-LA B 14,176.8 123 15–18 H 118 119 119 106 105 94 52 107 23 104 119
β-LG A β-LG A 18,355.5 162 22–24 H 157 159 156 158 158 155 136 156 97 157 155
β-LG B β-LG B 18,269.4 162 20–22 H 162 162 159 162 162 160 138 162 128 162 158

I, individual standard; M, mixed standard; J, Jersey milk; H, Holstein milk.

Results were consistent from sample to sample, particularly for the proteins that responded very well to top-down sequencing such as β-CN A2 with a coefficient of variation (CV) as low as 0.4%.

For ease of interpretation, the values reported in Table 2 were converted to percentages and displayed as histograms (Figure 4).

Figure 4.

Figure 4

Comparison of the 11 MS/MS methods based on the AA sequence coverage in percent of the nine most prominent milk proteins. (A) Cumulative coverage in the individual standards. (B) Cumulative coverage in the mixed standards. (C) Cumulative coverage in the Jersey milk sample. (D) Cumulative coverage in the Holstein milk sample. (E) Illustration of AA sequence coverage for one particular protein, β-LG A, in increasingly complex samples. (F) Averaged sequence coverage for all eight proteins sorted by method efficacy; the fragmentation modes are reported as well.

2.3.1. Reproducibility and Protein Specificity

While being both protein- and method-specific, top-down sequencing results were reproducible across samples of increasing complexity, from the least (individual standards, Figure 4A) and mildly (mixed standards, Figure 4B) complex, to the most complex samples (raw milk, Figure 4C–D). The sample reproducibility was also demonstrated on one particular protein, β-LG A, which exhibited CV values spanning from 1.0% (Method 7) to 1.9% (Methods 9 and 12), omitting Methods 8 and 10, which were not only the least efficient but also the least reproducible (Figure 4E). As observed in previous studies [28,29], top-down sequencing success is protein-dependent. In our conditions, αS2-CN A-10P was the least responsive, possibly due to its elevated number of phosphorylation sites, while β-LGs and β-CNs were the most responsive.

2.3.2. Low-Energy Fragmentation Drives Top-Down Sequencing Efficacy

The histograms in Figure 4 also clearly demonstrate which methods were suitable for top-down sequencing and which were not. In Figure 4F, the methods have been sorted in descending order of AA sequence coverage. Under our conditions, the best methods were Methods 4 and 12 (90.6% sequence coverage on average), closely followed by Methods 3 (88.6%), 11 (87.4%), 9 (86.8%), 5 and 2 (86.7%), 6 (86.6%), and 7 (83.6%). The worst-performing methods were 8 (60.2%) and 10 (38.6%). Based on our observations, we can conclude that the methods that resorted to IS-CID on its own with little energy applied, in combination with CID or not, were best suited to top-down sequence prominent milk proteins.

Optimal conditions in our case involved applying 3 eV (Method 4) or 10 eV (Method 12) to the ion funnel 1 exit during the IS-CID MS step and 15 eV to the ion funnel 2 entrance during the IS-CID MS/MS step. What seemed to be detrimental for our purposes was using too much energy (20 eV in Method 8 and 30 eV in Method 10) during the IS-CID MS/MS step. Moreover, we do not recommend applying an energy level higher than 10 eV during the IS-CID MS step, as was tested in Methods 5 (30 eV), 6 (50 eV), and 7 (70 eV).

2.3.3. AA Position Has No Influence on Fragmentation

In an attempt to further explain how CID-based fragmentation operated under our conditions, we aligned side by side all the AA sequences of the nine proteins of interest for each of the 11 MS/MS methods and blackened the AAs that were successfully top-down sequenced. For each protein, the sequenced AAs were counted across the methods and, as a result, we generated a score from 0 (not sequenced at all) to 11 (sequenced in all methods). Based on this scoring system, the corresponding position in the sequence was coloured red (the higher the score, the darker the shade). This is represented in Supplementary Figure S1. This display allowed for the quick visualisation of regions that remained unfragmented under our various conditions. Such recalcitrant regions were scattered along the protein sequence, suggesting that the AA position had no influence over fragmentation efficiency.

All cow’s milk caseins are heavily phosphorylated. There was no visible pattern associated with phosphorylation sites (highlighted in yellow) either, apart from αS2-CN A-10P, for which five phosphorylated serine residues (S at positions 7, 13, 31, 56, and 143) resisted fragmentation. PTMs were not investigated in this study. Milk proteins are also glycosylated, as has been well documented for κ-CN, for instance [36]. Glycosylation sites could account for areas difficult to fragment, but we could not test this hypothesis in the present study.

2.3.4. Hydrophobicity Affects Fragmentation

To further elucidate whether AAs themselves influenced fragmentation efficiency, we converted the AA counts from Supplementary Figure S1 into percentages, sorted them according to their highest fragmentation efficiency (meaning sequenced across all 11 MS/MS methods) and plotted them (Supplementary Figure S2). Underneath the chart we listed the physical properties attributed to AAs (ThermoFisher Scientific website) to assist us in finding a pattern.

In this representation, leucine, glycine, cysteine, tryptophan, histidine, and isoleucine displayed the best response to top-down sequencing with a success rate above 17% across 11 MS/MS methods (Supplementary Figure S2). These AAs are moderately to highly hydrophobic. Conversely, threonine, asparagine and arginine showed the lowest success rate (less than 9% across the 11 methods). Such AAs are hydrophilic. Our results suggest that AA hydrophobicity level has an impact on fragmentation efficiency.

2.3.5. High Sequence Coverage Is Critical for Allelic Variants

High sequence coverage is a prerequisite to successfully distinguishing between protein allelic variants such as β-CN A1 and A2, which only vary at position 67 (H to P); κ-CN A and B, which vary at positions 136 (T to I) and 148 (D to A); and β-LG A and B, which vary at positions 64 (D to G) and 118 (V to A). The mixed standard sample, Method 10, and, in the case of κ-CNs, Method 8 failed to deliver such relevant information since AAs at the aforementioned positions were not sequenced (Supplementary Figure S1).

Interestingly, some genetic variants proved more amenable to top-down sequencing than others. For instance, in the mixed standard sample processed with Method 12, β-CN A2 (97%) was more thoroughly sequenced than β-CN A1 (83%). Likewise, β-LG A (96%) displayed greater sequence coverage than β-LG B (77%); this was also observed for κ-CN A (73%) and κ-CN B (68%) (Table 3 and Figure S1). In Holstein milk samples, we managed to reach 100% sequence coverage of β-LG B in six out of 11 MS/MS methods. Complete top-down sequencing was also achieved for β-LG A in individual standard samples in four MS/MS methods (Table 3).

In the early days of top-down proteomics on ESI-Q-ToF platforms, the resulting short sequence tags from C- and N-termini from small to medium-sized proteins were just long enough to unambiguously identify the protein family, but not their allelic variants [26,27,28,29,30]. Almost complete top-down sequencing was achieved in 2009 with 17 kDa horse myoglobin and 30 kDa bovine carbonic anhydrase II being 92% and 96% sequenced, respectively [31]. Since then and to our knowledge, Q-ToF instruments have not been employed for top-down sequencing of intact proteins from complex biological matrices because they have been superseded by more expensive platforms such as FT mass spectrometers, offering better power resolution and alternative fragmentation methods like ETD, HCD, and ECD.

3. Materials and Methods

3.1. Materials and Sample Preparation

The experimental design is schematised in Figure 1. Protein standards purchased from Sigma-Aldrich (Castle Hill, NSW, Australia) were alpha-casein (α-CN) from bovine milk (C6780-250MG, 70% pure), beta-casein (β-CN) from bovine milk (C6905-250MG, 98% pure), kappa-casein (κ-CN) from bovine milk (C0406-250MG, 70% pure), alpha-lactalbumin (α-LA) from bovine milk (L5385-25MG, 85% pure), and beta-lactoglobulin (β-LG) from bovine milk (L3908-250MG, 90% pure). These lyophilised standards were prepared as described in [23]. Briefly, they were fully solubilised at a 10 mg/mL concentration in 50% MilliQ (MerckMillipore, Bayswater, VIC, Australia) water/50% solution A (0.1 M Bis-Tris, 6 M Guanidine-HCl, 5.37 mM sodium citrate tribasic dehydrate, and 20 mM DTT). A volume of 50% acetic acid to reach 1% acetic acid final concentration was added to the standards. A 0.1-mL aliquot of the solubilised standard was transferred into a 100-μL glass insert placed in a 2-mL glass vial for immediate analysis by LC-MS. A standard mixture was prepared by mixing individual standards in the following proportions to account for various ionisation efficiency [23]: 25% α-CN, 25% κ-CN, 20% β-LG, 20% β-CN, and 10% α-LA.

Milk collection from Holstein-Friesian cows (coded H) and Jersey cows (coded J) was described in [35]. Milk sample preparation was described in [23]. Briefly, 0.5 mL of cold skim milk was transferred into a 1.5-mL tube and 0.5 mL of Solution A was added. A 0.02 mL volume of 50% acetic acid (1% acetic acid final concentration, pH 5.8) was then added. A 0.1-mL aliquot of the milk protein extract was transferred into a 100-μL glass insert placed in a 2-mL glass vial for immediate analysis by LC-MS.

3.2. HPLC Separation of Intact Proteins

The separation of intact proteins by UHPLC 1290 Infinity Binary LC system (Agilent) through a Aeris™ WIDEPORE XB-C8 (3.6 μm particle size, 200 Å pore size, 150 × 2.1 mm dimensions, C8 reverse phase core-shell silica from Phenomenex (Lane Cove, NSW, Australia) column at 75 °C was described in [23]. In brief, 3 μL of sample was injected and separated as followed: starting conditions 20% B, ramping to 28% B in 2.5 min, ramping to 40% B in 27.5 min, ramping to 99% B in 1 min and held for 4 min, lowering to 20% B in 0.1 min, equilibration at 20% B for 4.9 min. Mobile phase flow rates was 200 μL/min. Mobile phase A contained ACN with 0.1% FA and 0.02% trifluoroacetic acid (TFA), and mobile phase B contained H2O with 0.1% FA and 0.02% TFA. The diode array detector (DAD) spectrum was acquired from 190 to 400 nm. The pressure limit was set at 600 bars.

3.3. MS1 Analysis

MS1 analysis was described in [23]. The UHPLC was on-line with a maXis HD UHR-Q-ToF (60,000 resolution) fitted with a standard ESI Apollo-source (BrukerDaltonikGmbh, Preston, VIC, Australia). To ensure mass accuracy, a Na-formate solution was infused continuously at 0.1 mL/h and the first 2.5 min of each run were used to re-calibrate masses post-acquisition. Each 40-min run was thus segmented as follows: 2.5 min to waste and the following 37.5 min to source. Capillary voltage was set at 4500 V. The nebuliser was set at 1.5 bar. The dry gas was set at 8 L/min. The dry temperature was set at 190 °C. The transfer funnel RF and multipole RF were set at 400 Vpp; no IS-CID energy was applied. The quadrupole ion energy was 5 eV, the collision cell energy was 10eV, and the collision RF 1800 Vpp. The ion cooler transfer time was 120 μs, with a prepulse storage of 10 μs and a RF of 400 Vpp. The ion polarity was positive and scan mode was MS. The rolling average mode was activated and set at 2. This MS1 method is called Method 1 hereafter.

3.4. MS2 Analyses and Top-Down Sequencing

MS/MS experiments were performed using the same LC parameters and MS1 parameters on the Q-TOF mass spectrometer as described above. A total of 11 MS2 methods were tested; their parameters are indicated in Table 1. These methods are referred to as Method 2 to Method 12 hereafter. LC-MS files were visualised using Bruker Compass DataAnalysis version 4.2. MS2 spectra were annotated using Bruker Biotools version 3.2 and SequenceEditor version 3.2. The retrieval of AA sequences was detailed in [23].

3.5. Protein Annotation in Genedata Expressionist

The data files obtained following LC-MS analysis using MS2 methods 2 to 12 were curated in the Refiner MS module of Genedata Expressionist® version 11.0 with the following parameters: (1) Conversion of MS/MS to primary MS. (2) Chromatogram chemical noise subtraction using moving average algorithm and a five-scan window. (3) Spectrum smoothing using a moving average algorithm and a five-point m/z window. (4) Chromatogram peak detection using a five-scan summation window, a minimum peak size of four scans with boundaries merge strategy and five-point maximum merge distance, a curvature-based peak detection with 70% intensity threshold and inflection points boundary determination. (5) Chromatogram isotope clustering using a 0.2 min and 20 ppm tolerances, a peptide isotope shaping with 1‒8 protons, a 0.6 log-ration maximum distance, mono-isotopic computation, linear charge dependency with a minimum size ratio of 0.65. (6) Singleton filtering. (7) Peptide mapping with a 50 ppm tolerance, unspecific enzyme, three missed cleavage maximum, 10 AAs minimum length, and variable modifications, as explained hereafter. The peptide mapping was performed using a text file containing all 49 AA sequences in FASTA format of the bovine allelic variants of milk caseins, alpha-lactalbumin, and beta-lactoglobulin (Farrell et al., 2004) with the following variable modifications: pyro-Glu (N-term Q) and Phospho (ST). Some aspects of this workflow are illustrated in Figure S3.

Protein deconvolution was also performed on Method 1 files in the Refiner MS module of Genedata Expressionist® version 11.0 with the following parameters. (1) Data sweep to remove UV data. (2) Chromatogram chemical noise subtraction using moving average algorithm, with a subtraction method, a 71 scan window, and 60% quantile. (3) Chromatogram lock mass using the Na-formate ion series. (4) Retention time range restriction from 4 to 25 min. (5) Intact protein activity using the harmonic suppression deconvolution method with 0.02 Da steps and 5-30 kDa masses. (6) Intensity thresholding using a clipping method with an intensity of 50. (7) Spectrum baseline subtraction using a 99% quantile and 30 kDa m/z window. (8) Chromatogram retention time alignment using a pairwise alignment-based tree scheme with 50-scan search interval and a gap penalty of 1.

All MS1 and MS2 files are available from the stable public repository MassIVE at the following URL: http://massive.ucsd.edu/ProteoSAFe/datasets.jsp with the accession number MSV000082070.

3.6. Validation and Computational Analyses

Top-down sequencing results were visualised and validated in Genedata Expressionist® version 11.0 Refiner MS module within the Peptide Mapping activity and all its subsidiary tabs as well as using Bruker Biotools version 3.2 and SequenceEditor version 3.2, as exemplified in Figure S3.

Top-down sequencing annotations (including b- and y-ions series) for each of the nine proteins of interest in every sample and each of the 11 MS/MS methods were exported from Refiner MS module as .csv files. The .csv files contained the list of peptides identified for each protein along with the peptide AA lengths and their position in the protein sequence they matched to. The .csv files were imported into Microsoft® Excel 2016 software for further processing. Coverage information was derived from the length and position of the identified peptides relative to their matching protein sequence to produce Table 3 and Figure 4. AA responsiveness to MS2 fragmentation was obtained using the positions of the first and last AA of the identified peptides. These analyses produced Figures S1 and S2.

4. Conclusions

In this experiment, we have fine-tuned IS-CID parameters with and without traditional CID using a LC-ESI-Q-ToF system in order to optimise the top-down sequencing of milk intact proteins. The most efficient methods utilised IS-CID on its own at low energy. Full (100%) sequence covering could be obtained on β-LGs, and overall enough coverage was achieved to distinguish between allelic proteoforms, varying only by one or two AAs. Such information is highly biologically relevant as alleles influence function and phenotypic features. In bovine milk, for instance, 12 allelic variants of beta-caseins have been identified across different breeds and populations [37]. The most common variants, A1 and A2, differ at amino acid position 67 with histidine in A1 and proline in A2 milk, as a result of a single nucleotide difference. One application of this research would be screening for the presence of the A1 variant in cow’s milk samples.

Abbreviations

Abbreviation Definition
AA Amino acid
ACN Acetonitrile
AE Active exclusion
CAD Collisionally activated dissociation
CID Collision-induced dissociation
CV Coefficient of variation
DAD Diode array detector
DC Direct current
ECD Electron capture dissociation
ESI Electrospray ionisation
ETD Electron transfer dissociation
eV Electron-volt
FA Formic acid
FT Fourier transform
HCD Higher-energy collisional dissociation
HPLC High-performance liquid chromatography
IS-CID In-source CID
kDa KiloDalton
keV Kiloelectron-volt
LC-MS Liquid chromatography‒mass spectrometry
LIT Linear ion trap
m/z Mass-to-charge ratio
MS Mass spectrometry
MS/MS Tandem MS
MS1 Full scan MS
MS2 Tandem MS
MW Molecular weight
ppm Parts per million
PTM Post-translational modification
Q Quadrupole
QQQ Triple quadrupole
Q-ToF Quadrupole time-of-flight
RF Radiofrequency
TFA Trifluoroacetic acid
ToF Time-of-flight
UHT Ultra-high temperature
V Volt
α-LA Alpha-lactalbumin
αS1-CN Alpha-S1-casein
αS2-CN Alpha-S2-casein
β-CN Beta-casein
β-LG Beta-lactoglobulin
κ-CN Kappa-casein

Supplementary Materials

The following are available online, Figure S1: Top-down sequencing responsiveness of AAs for each of the eight most abundant proteins from the mixed standard sample, Figure S2: CID-friendly amino acids from top-down sequenced milk proteins, Figure S3: Data processing of MS/MS files using Genedata Refiner and DataAnalysis/Biotools exemplified on the Holstein milk sample processed using Method 3.

Author Contributions

Conceptualisation, D.V.; methodology, D.V. and D.M.; software, D.V. and D.M.; validation, D.V.; formal analysis, D.V.; investigation, D.V.; resources, S.R.; data curation, D.V.; writing—original draft preparation, D.V.; writing—review and editing, D.V., D.M., and S.R.; visualisation, D.V.; supervision, S.R.; project administration, D.V.; funding acquisition, S.R.

Funding

This research received no external funding.

Conflicts of Interest

The authors declare no conflict of interest.

Footnotes

Sample Availability: Samples of the compounds are not available from the authors.

References

  • 1.Kelleher N.L., Lin H.Y., Valaskovic G.A., Aaserud D.J., Fridriksson E.K., McLafferty F.W. Top down versus bottom up protein characterization by tandem high-resolution mass spectrometry. J. Am. Chem. Soc. 1999;121:806–812. doi: 10.1021/ja973655h. [DOI] [Google Scholar]
  • 2.Toby T.K., Fornelli L., Kelleher N.L. Progress in Top-Down Proteomics and the Analysis of Proteoforms. Annu. Rev. Anal. Chem. 2016;9:499–519. doi: 10.1146/annurev-anchem-071015-041550. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Dole M., Mack L.L., Hines R.L., Mobley R.C., Ferguson L.D., Alice M.B. Molecular Beams of Macroions. J. Chem. Phys. 1968;49:2240–2249. doi: 10.1063/1.1670391. [DOI] [Google Scholar]
  • 4.Fenn J.B., Mann M., Meng C.K., Wong S.F., Whitehouse C.M. Electrospray ionization for mass spectrometry of large biomolecules. Science. 1989;246:64–71. doi: 10.1126/science.2675315. [DOI] [PubMed] [Google Scholar]
  • 5.Loo J.A., Udseth H.R., Smith R.D. Collisional Effects on the Charge Distribution of Ions from Large Molecules, Formed by Electrospray-ionization Mass Spectrometry. Rapid Commun. Mass Spectrom. 1988;2:207–210. doi: 10.1002/rcm.1290021006. [DOI] [Google Scholar]
  • 6.Mann M., Meng C.K., Fenn J.B.F. Interpreting Mass Spectra of Multiply Charged Ions. Anal. Chem. 1989;61:1702–1708. doi: 10.1021/ac00190a023. [DOI] [Google Scholar]
  • 7.Jones A.W., Cooper H.J. Dissociation techniques in mass spectrometry-based proteomics. Analyst. 2011;136:3419–3429. doi: 10.1039/c0an01011a. [DOI] [PubMed] [Google Scholar]
  • 8.McLafferty F.W., Bryce T.A. Metastable-ion characteristics: Characterization of isomeric molecules. Chem. Commun. 1967;0:1215–1217. doi: 10.1039/c19670001215. [DOI] [Google Scholar]
  • 9.Jennings K.R. Collision-induced decompositions of aromatic molecular ions. Int. J. Mass Spectrom. Ion. Phys. 1968;1:227–235. doi: 10.1016/0020-7381(68)85002-8. [DOI] [Google Scholar]
  • 10.Garcia B.A. What does the future hold for Top Down mass spectrometry? J. Am. Soc. Mass Spectrom. 2010;21:193–202. doi: 10.1016/j.jasms.2009.10.014. [DOI] [PubMed] [Google Scholar]
  • 11.Pamreddy A., Panyala N.R. Top-down proteomics: Applications, recent developments and perspectives. J. Appl. Bioanal. 2016;2:52–75. doi: 10.17145/jab.16.009. [DOI] [Google Scholar]
  • 12.Katta V., Chowdhury S.K., Chait B.T. Use of a single-quadrupole mass spectrometer for collision-induced dissociation studies of multiply charged peptide ions produced by electrospray ionization. Anal. Chem. 1991;63:174–178. doi: 10.1021/ac00002a016. [DOI] [PubMed] [Google Scholar]
  • 13.Bure C., Lange C. Comparison of dissociation of ions in an electrospray source, or a collision cell in tandem mass spectrometry. Curr. Org. Chem. 2003;7:1613–1624. doi: 10.2174/1385272033486332. [DOI] [Google Scholar]
  • 14.Van Dongen W.D., van Wijk J.I., Green B.N., Heerma W., Haverkamp J. Comparison between collision induced dissociation of electrosprayed protonated peptides in the up-front source region and in a low-energy collision cell. Rapid Commun. Mass Spectrom. 1999;13:1712–1716. doi: 10.1002/(SICI)1097-0231(19990915)13:17<1712::AID-RCM703>3.0.CO;2-8. [DOI] [PubMed] [Google Scholar]
  • 15.Senko M.W., Beu S.C., McLafferty F.W. High-resolution tandem mass spectrometry of carbonic anhydrase. Anal. Chem. 1994;66:415–418. doi: 10.1021/ac00075a016. [DOI] [PubMed] [Google Scholar]
  • 16.Parcher J.F., Wang M., Chittiboyina A.G., Khan I.A. In-source collision-induced dissociation (IS-CID): Applications, issues and structure elucidation with single-stage mass analyzers. Drug Test. Anal. 2018;10:28–36. doi: 10.1002/dta.2249. [DOI] [PubMed] [Google Scholar]
  • 17.Kelleher N.L. Top-down proteomics. Anal. Chem. 2004;76:197A–203A. doi: 10.1021/ac0415657. [DOI] [PubMed] [Google Scholar]
  • 18.Loo J.A., Edmonds C.G., Smith R.D. Primary sequence information from intact proteins by electrospray ionization tandem mass spectrometry. Science. 1990;248:201–204. doi: 10.1126/science.2326633. [DOI] [PubMed] [Google Scholar]
  • 19.Loo J.A., Quinn J.P., Ryu S.I., Henry K.D., Senko M.W., McLafferty F.W. High-resolution tandem mass spectrometry of large biomolecules. Proc. Natl. Acad. Sci. USA. 1992;89:286–289. doi: 10.1073/pnas.89.1.286. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Beu S.C., Senko M.W., Quinn J.P., McLafferty F.W. Improved fourier-transform ion-cyclotron-resonance mass spectrometry of large biomolecules. J. Am. Soc. Mass Spectrom. 1993;4:190–192. doi: 10.1016/1044-0305(93)85077-B. [DOI] [PubMed] [Google Scholar]
  • 21.Shevchenko A., Chernushevich I., Ens W., Standing K.G., Thomson B., Wilm M., Mann M. Rapid ‘de novo’ peptide sequencing by a combination of nanoelectrospray, isotopic labeling and a quadrupole/time-of-flight mass spectrometer. Rapid Commun. Mass Spectrom. 1997;11:1015–1024. doi: 10.1002/(SICI)1097-0231(19970615)11:9<1015::AID-RCM958>3.0.CO;2-H. [DOI] [PubMed] [Google Scholar]
  • 22.Morris H.R., Paxton T., Dell A., Langhorne J., Berg M., Bordoli R.S., Hoyes J., Bateman R.H. High sensitivity collisionally-activated decomposition tandem mass spectrometry on a novel quadrupole/orthogonal-acceleration time-of-flight mass spectrometer. Rapid Commun. Mass Spectrom. 1996;10:889–896. doi: 10.1002/(SICI)1097-0231(19960610)10:8<889::AID-RCM615>3.0.CO;2-F. [DOI] [PubMed] [Google Scholar]
  • 23.Vincent D., Elkins A., Condina M.R., Ezernieks V., Rochfort S. Quantitation and Identification of Intact Major Milk Proteins for High-Throughput LC-ESI-Q-TOF MS Analyses. PLoS ONE. 2016;11:e0163471. doi: 10.1371/journal.pone.0163471. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Domon B., Aebersold R. Mass spectrometry and protein analysis. Science. 2006;312:212–217. doi: 10.1126/science.1124619. [DOI] [PubMed] [Google Scholar]
  • 25.Steen H., Kuster B., Mann M. Quadrupole time-of-flight versus triple-quadrupole mass spectrometry for the determination of phosphopeptides by precursor ion scanning. J. Mass Spectrom. 2001;36:782–790. doi: 10.1002/jms.174. [DOI] [PubMed] [Google Scholar]
  • 26.Nemeth-Cawley J.F., Rouse J.C. Identification and sequencing analysis of intact proteins via collision-induced dissociation and quadrupole time-of-flight mass spectrometry. J. Mass Spectrom. 2002;37:270–282. doi: 10.1002/jms.281. [DOI] [PubMed] [Google Scholar]
  • 27.Nemeth-Cawley J.F., Tangarone B.S., Rouse J.C. “Top Down” characterization is a complementary technique to peptide sequencing for identifying protein species in complex mixtures. J. Proteome Res. 2003;2:495–505. doi: 10.1021/pr034008u. [DOI] [PubMed] [Google Scholar]
  • 28.Ginter J.M., Zhou F., Johnston M.V. Generating protein sequence tags by combining cone and conventional collision induced dissociation in a quadrupole time-of-flight mass spectrometer. J. Am. Soc. Mass Spectrom. 2004;15:1478–1486. doi: 10.1016/j.jasms.2004.07.004. [DOI] [PubMed] [Google Scholar]
  • 29.Johnson R.W.J., Ahmed T.F., Miesbauer L.J., Edalji R., Smith R., Harlan J., Dorwin S., Walter K., Holzman T. Protein fragmentation via liquid chromatography-quadrupole time-of-flight mass spectrometry: The use of limited sequence information in structural characterization. Anal. Biochem. 2005;341:22–32. doi: 10.1016/j.ab.2005.03.009. [DOI] [PubMed] [Google Scholar]
  • 30.Deepalakshmi P.D. Top-down approach of completely sequencing a 4.9 kDa recombinant peptide using quadrupole time-of-flight mass spectrometry. Rapid Commun. Mass Spectrom. 2006;20:3747–3754. doi: 10.1002/rcm.2797. [DOI] [PubMed] [Google Scholar]
  • 31.Armirotti A., Benatti U., Damonte G. Top-down proteomics with a quadrupole time-of-flight mass spectrometer and collision-induced dissociation. Rapid Commun. Mass Spectrom. 2009;23:661–666. doi: 10.1002/rcm.3925. [DOI] [PubMed] [Google Scholar]
  • 32.Liu J., Jiang J., Wu Z., Xie F. Antimicrobial peptides from the skin of the Asian frog, Odorrana jingdongensis: De novo sequencing and analysis of tandem mass spectrometry data. J. Proteomics. 2012;75:5807–5821. doi: 10.1016/j.jprot.2012.08.004. [DOI] [PubMed] [Google Scholar]
  • 33.Boeri Erba E. Investigating macromolecular complexes using top-down mass spectrometry. Proteomics. 2014;14:1259–1270. doi: 10.1002/pmic.201300333. [DOI] [PubMed] [Google Scholar]
  • 34.Raynes J.K., Vincent D., Zawadzki J.L., Savin K., Logan A., Mertens D., Williams R.P.W. Investigation of age gelation UHT Milk. Beverages. 2018 submitted for publication. [Google Scholar]
  • 35.Vincent D., Ezernieks V., Elkins A., Nguyen N., Moate P.J., Cocks B.G., Rochfort S. Milk Bottom-Up Proteomics: Method Optimization. Front. Genet. 2015;6:360. doi: 10.3389/fgene.2015.00360. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.O’Riordan N., Kane M., Joshi L., Hickey R.M. Structural and functional characteristics of bovine milk protein glycosylation. Glycobiology. 2014;24:220–236. doi: 10.1093/glycob/cwt162. [DOI] [PubMed] [Google Scholar]
  • 37.Farrell H.M., Jr., Jimenez-Flores R., Bleck G.T., Brown E.M., Butler J.E., Creamer L.K., Hicks C.L., Hollar C.M., Ng-Kwai-Hang K.F., Swaisgood H.E. Nomenclature of the proteins of cows’ milk—Sixth revision. J. Dairy Sci. 2004;87:1641–1674. doi: 10.3168/jds.S0022-0302(04)73319-6. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials


Articles from Molecules are provided here courtesy of Multidisciplinary Digital Publishing Institute (MDPI)

RESOURCES