Abstract
PURPOSE
To determine the reproducibility of quantitative susceptibility mapping at multiple sites on clinical and preclinical scanners (1.5, 3, 7 and 9.4 T) from different vendors (Siemens, GE, Philips and Bruker) for standardization of multicenter studies.
METHODS
Seven phantoms distributed from the core site, each containing five compartments with gadolinium solutions with fixed concentrations between 0.625 mM to 10 mM. Multi-echo gradient echo scans were performed at 1.5, 3, 7 and 9.4 Tesla on 12 clinical and 3 preclinical scanners. DICOM images from the scans were processed into quantitative susceptibility maps using the Laplacian Boundary Value (LBV) and MEDI+0 automatic uniform reference algorithm. Region of interest (ROI) analyses were performed by a physicist to determine agreement between results from all sites. Measurement reproducibility was assessed using regression, Bland-Altman plots and the intra-class correlation coefficient (ICC).
RESULTS
QSM maps from all scanners had similar, artifact-free visual appearance. Regression analysis showed a linear relationship between gadolinium concentrations and average QSM measurements for all phantoms (y = 350× − 0.0346, r2 > 0.99). The standard deviation of measurements increased almost linearly from 32 ppb to 230 ppb as the measured susceptibility increased from 0.26 to 3.56 ppm. A Bland-Altman plot showed the bias, upper and lower limits of agreement for all comparisons were −10, −210 and 200 ppb respectively. The ICC was 0.991 with a 95% CI (0.973, 0.99).
CONCLUSIONS
QSM shows excellent multicenter reproducibility for a large range of susceptibility values encountered in cranial and extra-cranial applications on a diverse set of scanner platforms.
Keywords: multicenter reproducibility, quantitative susceptibility mapping
INTRODUCTION
From its initial applications to endogenous brain iron quantification, quantitative susceptibility mapping (QSM) has seen increasing applications to a wide range of quantitative research on disease conditions including intracerebral hemorrhage (1–4), liver iron overload (5,6), bone mineral quantification (7) and preclinical contrast agent toxicity studies (8). These applications require confidence in the reproducibility QSM for a wider range of susceptibility values, field strengths, and vendor platforms than has been previously reported by researchers (9–13).
An important tool used in MRI research including the QSM approach is a phantom that contains a set of compartments with different gadolinium concentrations that are fixed in agarose gel (14–17). Gadolinium solutions at known concentrations are regularly employed as tissue models of susceptibility for both healthy and diseased anatomical targets because their magnetic behavior can be designed to closely mimic that of each target tissue (18). The absolute susceptibility of 0.5 mM solution of a gadolinium-based contrast agent (formulated at a standard concentration of 0.5 M), is −8.859 ppm which is close to that of most biological tissues. QSM experiments on phantoms of gadolinium have been used useful starting points for obtaining insight into various physiological processes including urine concentration by kidneys (19), cardiac flow rate (20) and cerebral perfusion (21). Gadolinium solutions can reproducibly model QSM measurements of tissue iron, which are in the form of ferritin and hemosiderin, across all field strengths, unlike iron oxides which exhibit saturation of magnetization by 1 Tesla (22). Because of regular use of such gadolinium-based phantoms for both the development and validation of QSM acquisition and post-processing algorithms, the study of reproducibility of QSM measurements in gadolinium phantoms is important.
Previous investigations that examined the reproducibility of QSM by employing phantoms are now known to have two specific shortcomings. First, studies prior to 2015 utilized specific algorithms that performed a key dipole inversion step either without regularization (9,11), or by minimization of a linear data term with L1-regularization (9), which consistently yielded moderate to severe streaking artifacts in resultant images (9,11). Subsequently developed post-processing methods have been able to more effectively suppress these streaking artifacts, most notably by minimizing a nonlinear data fidelity term with L1-regularization (23). QSM measurements from different MRI scanners are only comparable with respect to a suitably chosen reference. Susceptibility sources are usually embedded in an aqueous gel which retains phase information for more accurate QSM reconstruction. Since it is reasonable to expect that susceptibility within the aqueous gel solution is uniform, this knowledge can be employed as a constraint during QSM reconstruction to potentially improve QSM image quality. A recently developed algorithm, MEDI+0, takes advantage of this constraint by adding another L2-regularization term to the optimization equation used in the iterative approach (23), to enforce homogeneity in specified regions (24).. MEDI+0 enables the use of the aqueous gel region as a consistent automatic zero-reference and eliminates the need for manual selection of an arbitrary reference region. This feature makes MEDI+0 highly suitable for performing multicenter QSM comparisons using agarose gel phantoms. In this study, we compare the QSM estimates derived using the MEDI+0 algorithm across a fixed range of susceptibility sources in a set of simultaneously synthesized gadolinium phantoms, distributed across at seven sites on multiple vendor systems, and imaged at four field strengths.
METHODS
A phantom was constructed by embedding five latex balloons (Imperial Kaos, Tie-Not, US), each filled with a specific concentration (10, 5, 2.5, 1.25 or 0.625 mmol/L) of gadolinium (Magnevist, Berlex Laboratories, Wayne, NJ), in a 1% agarose gel solution in a 500 mL plastic container. Images of the phantom were acquired on 1.5 and 3 Tesla scanners available at the primary study site (Site A, Table 1) in a head coil using a 3D multi-echo spoiled gradient recalled-echo (SPGR) sequence with the following parameters: voxel size = 0.6 × 0.6 × 0.6 mm3, flip angle = 15°, TR = 49 ms, number of echoes = 10, first TE = 3.7 ms, echo spacing = 4.1 ms. Four additional phantoms were constructed in an identical manner, scanned at Site A and shipped to Sites B to E (Table 1) for the multi-site imaging experiment. Imaging parameters across all sites were kept similar to parameters from Site A. For imaging on small bore preclinical scanners, two dedicated phantoms were designed and constructed by embedding five balloons with gadolinium concentrations ranging from 10 to 0.625 mmol/L in a 50 mL conical centrifuge tube (Falcon, Fisher-Scientific, USA) of 1% agarose gel. This phantom was scanned on a Bruker 7T scanner (BioSpin MRI GmbH) at Site 1 using the following parameters: voxel size = 0.5 × 0.5 × 0.5 mm3, flip angle = 15°, TR = 34 ms, number of echoes = 8, first TE = 3.4 ms, echo spacing = 3.8 ms. These phantoms were then shipped to Sites E and F for imaging on preclinical scanners. The phantoms scanned on the MR platforms available at each site are shown in Table 1. The range of imaging parameters used for all scans are displayed in Table 2. All DICOM images generated from the examined scanner hardware at each site were electronically transmitted to Site A. Derived MRI DICOMs required the complex dataset that consisted of the phase map information. These were imported into a custom MATLAB software (The Mathworks, Natick, MA). For each scan, the total magnetic field map was generated from the DICOM images using a nonlinear voxel-wise fit (23) with a correction to remove residual echo-to-echo phase inconsistencies in the readout direction when present (25). Phase processing was performed using quality guided field unwrapping (26) and Laplacian Boundary Value (LBV) background field removal (27), because these methods have been shown to produce the best performance in QSM processing (12,28). The QSM map was reconstructed from the local field map using the MEDI+0 method (24), which minimized variation in the agarose gel medium of the phantom according to the expression:
(1) |
where χ is the susceptibility map, * is the convolution operation, w is noise weighting, f is the measured local field, ∇ is the gradient operator, and MG is a binary edge mask. The term, , penalizes susceptibility variation within the agarose gel medium, which can be isolated using a mask, MAGAROSE, obtained automatically by complementing a mask of the entire phantom with a mask of the gadolinium balloons defined as voxels with intensity greater than 40% of the maximum intensity of the magnitude image.. The optimal values for the regularization parameters, λ1 and λ2 were determined using L-curve plot analysis to be 10 and 1 respectively. The output of each step of the reconstruction procedure is shown in FIG 1. The mean voxel value in a 3-dimensional ROI covering the entire region of the gadolinium solution in each balloon was recorded.
Table 1.
Site | 1.5T scanners | 3T scanners | 7T Scanners | 9.4 T scanners |
---|---|---|---|---|
A | Siemens Aera (VE11A, 64 Ch HE/NE) | Siemens Skyra (VE11A, 20 Ch HE/NE), Siemens Biograph mMR (VB20P, 32 Ch HE) | Bruker Biospin (Paravision 5.1, body coil) | |
B | Philips Achieva (5.1.7.2, 8 Ch HE) | Philips Achieva dStream (5.1.7.2, 32 Ch HE) | ||
C | GE Signa HDxT (V23, 8 Ch HE) | GE Signa HDxT (V23, 8 Ch HE) | ||
D | Siemens Aera (E11, 20 Ch HE/NE) | Siemens Prisma (E11, 20 Ch HE/NE) | ||
E | GE Signa HDxT (V23, 8 Ch HE) | GE Discovery MR750 (V25, 8 Ch HE) | ||
F | Philips Achieva (5.1.7.0, 32 Ch HE), Bruker Pharmascan (Paravision 6.0.1, body coil) | |||
G | Bruker Biospin (ParaVision 6.0.1, body coil) |
Table 2.
1.5T | 3T | 7T | 9.4 T | |
---|---|---|---|---|
Number of echoes | 6–10 | 6–10 | 8 | 7 |
Minimum echo (ms) | 1.8 – 3.7 | 1.3 – 4.3 | 3.4 | 4.0 |
ΔTE (ms) | 2.2 – 4.1 | 3.5 – 4.1 | 3.75 | 4.7 |
Slice thickness (mm) | 0.6 – 0.7 mm | 0.6 – 0.7 mm | 0.47 | 0.39 |
Bandwidth (Hz/Pixel) | 125 – 488 | 125 – 1116 | 334 – 842 | 500 |
Flip Angle | 15 | 15 | 15 | 20 |
Scan time (mins) | 6 – 8 | 6 – 8 | 5 – 12 | 20 |
Statistical Analysis
Statistical analysis was performed using Predictive Analytics Software (PASW) Statistics 18 (SPSS, Chicago, IL). Summary statistics, including mean, (μx), and standard deviation, (σx), of the QSM measurements observed at each gadolinium concentration was recorded. These measurements were subsequently used to determine if there was any change in dispersion of measurements with changes in susceptibility. A regression of the susceptibility estimates against the gadolinium concentration in each phantom was performed to estimate the molar susceptibility of gadolinium. To assess the agreement of the measurements, Bland and Altman’s method for multiple measurements per object (phantom) was applied to measurements from pairs of scanners used for each phantom (29). This method involves plotting the differences between pairs of measurements on the y-axis against their averages on the x-axis. The limits of agreement for all measurement methods (scanners) are determined from the standard deviation of the data as follows: If we designate the error and true variances in the measurements of each gadolinium concentration by multiple scanners as and respectively, then, a one-way analysis of variance (ANOVA) gives the within-group mean square ( ) and the between-group mean square ( ), where k is the number of between-group degrees of freedom. The square root of the total variance, , is then used to estimate the limits of agreement for the Bland-Altman (BA) plot.
A coefficient of reliability of measurements can be defined as . By arithmetic manipulation, this coefficient can be expressed in terms of ANOVA variables as
(2) |
Eqn (2) is also known as intra-class correlation coefficient (ICC), which is a measure of both the correlation and agreement of measurements (30). In this study, the ICC for all measurements was computed using a one-way random effect, absolute agreement, multiple raters model in PASW Statistics 18.
Another useful statistical measure is the standard deviation of the sampling error also called the standard error of measurement (SEM) which gives an estimate of the precision of a single measurement. This can be determined from the definition of the coefficient of reliability provided above to be
(3) |
and is useful for constructing confidence intervals for measurements.
RESULTS
Twenty-two QSM maps were reconstructed from GRE scans of seven synthesis-matched gadolinium phantoms performed on twelve clinical and three preclinical scanners at seven imaging sites. Individual measurements of the susceptibility in each gadolinium compartments are shown in Supporting Information Table S1. The MEDI+0 approach was employed for this study’s QSM reconstruction resulted in maps with no observable streaking artifacts and a uniform agarose gel region. Fig. 2 shows representative QSM maps generated for 3 clinical scanners at different field strengths, and for one preclinical scanner at 7T. The visual similarity of the susceptibility sources, and the uniformity of the surrounding agarose gel medium on the maps generated by MEDI+0 for all scanners can be appreciated.
The advantage of using MEDI+0 is visible for highly magnetized gadolinium samples; that is very high concentrations at high field strengths. In Supporting Information Figure S1, the magnitude and local field map are shown together with the QSM computed using MEDI+0 and the standard MEDI algorithms. Both MEDI+0 and MEDI use an SNR weighting term (w in Eqn (1)) derived from the magnitude image. The low SNR at highest gadolinium concentration (white arrow on magnitude image) propagates into a reconstruction artifact in the QSM image generated by MEDI. The additional constraint of a Dirichlet-like boundary condition in the MEDI+0 model results in a better QSM map as can been seen in Supporting Information Figure S1.
ROI-based estimates of susceptibility values showed a linear increase as the reference Gd concentration increased for all acquired MRI scans. A linear regression of QSM estimates of susceptibility against gadolinium concentration in each phantom was performed to obtain an estimate of the molar susceptibility of gadolinium from that phantom. Fig. 3a is a dot plot showing the molar susceptibility values recorded for each phantom by different scanners. The bold horizontal line indicates the average molar susceptibility of 350 ppm L/mol obtained for all molar susceptibility estimates. This value is in reasonable agreement with previous gadolinium susceptibility measurements using MRI (17,31). Except for Phantom 2, the estimated molar susceptibility values are closely clustered showing good agreement between estimates of the molar susceptibility from a given phantom by different scanners. Possible reasons for the large dispersion observed in the data for Phantom 2, such as temperature effects, are addressed in the discussion section. The ICC value calculated from all susceptibility measurements was 0.991 with a 95% confidence interval of 0.973 to 0.999. Using this ICC value and the standard deviation of the QSM measurement at each concentration, the standard error of measurement was calculated with Eq. [2] and displayed in Fig. 3b. It showed decreasing precision of QSM measurements with increasing susceptibility strength. Fig. 3c is a plot of the differences against the averages of pairs of QSM measurements obtained using different scanners. The solid and dotted lines are the bias and limits of agreement calculated using the Bland-Altman method of multiple measurements. The bias values are close to zero for all BA comparisons showing a lack of systematic error in QSM measurements. The limits of agreements are relatively large, about ±200 ppb, and are indicative of the large standard deviation in the measurements of the highest gadolinium concentration.
DISCUSSION AND CONCLUSIONS
In this study, we analyzed 22 QSM datasets generated with the MEDI+0 automatic reference algorithm from scans of a gadolinium phantom on fifteen scanners at four field strengths and observed good reproducibility for a wide range of susceptibility values such as may be encountered in cerebral and extracranial applications of QSM. An excellent intra-class coefficient of 0.991 was obtained for all data with 95% confidence interval of 0.973 to 0.999. The standard error of measurement was estimated to increase from 3 ppb for an average measurement of 0.26 ppm to 21 ppb for an average measurement of 3.56 ppm. Bland-Altman analysis showed a bias of 10 ppb, and lower and upper limits of agreements of −210 and 200 ppb respectively.
Phantoms of gadolinium solution embedded in agarose gel are a useful starting point for the development and verification of QSM methods. As a result, investigations into the reproducibility of QSM measurements in gadolinium phantoms remain a useful area of study. However, previous studies have published maps that show streaking artifacts and lack a consistent reference. In this study, we demonstrate visual improvement in QSM images by using MEDI+0, an algorithm originally developed to enforce the homogeneity of the susceptibility map in the cerebrospinal fluid (CSF), resulting in the generation of a consistent reference and reduction of shadowing artifacts (24). MEDI+0 was used to enforce homogeneity in the region of the agarose gel to realize this improvement.
Despite good agreement in QSM measurements, the standard deviation of measurements was observed to increase with susceptibility strength resulting in decreasing precision (standard error of measurement) (Fig. 3b). This is because the standard deviation of QSM measurements is proportional to the phase noise, which is in turn inversely proportional to the SNR on the magnitude image. Increasing susceptibility values decrease the magnitude SNR from T2* loss, and consequently increase standard deviation of QSM measurements from 32.5 ppb at 0.26 ppm to 230 ppb at 3.56 ppm. From Eq. [3], the corresponding values of SEM are 3.08 to 21.9 ppb. The implication of this observation is that confidence intervals of QSM measurements widen as measured susceptibility increases. Fortuitously, the SEM is small even at a relatively high susceptibility. A possible remedy for this problem is the use of QSM algorithms developed specifically to deal with a wide range of susceptibility values (2,4).
Despite the good reproducibility of QSM obtained for the statistical measures considered in this study, critical examination of the QSM values (Supporting Information Table S1), does show there can be outlier values from certain scans. The sources of this variability are difficult to trace and may be due to variations in image processing tasks such as automated mask and ROI generation. The small differences in acquisition parameters between different sites that can be seen in Table 2 did not appear to be the dominant factor in producing observed differences between QSM estimates of gadolinium. Another possible source of variability in this study was the temperature of the phantom which may have affected the measured molar susceptibility values. The theoretical value for the molar susceptibility of Gd3+ is reported to be 7.88 cm3 K mol−1(32), or 0.099 L K mol−1. For example, the molar susceptibility for Gd-DTPA at 298 K is 0.0265 cm3/mol or 332.3 ppm L/mol as estimated in (31). The change in molar susceptibility with temperature can be determined from a regression to be −1.093 ppm/M/K. Therefore, a 1 degree change in temperature can significantly affect the measured molar susceptibility. Future multicenter comparisons of molar magnetic susceptibility measurements can ensure the phantoms are kept at the same temperature prior to and during scanning.
One limitation of this study was the use of phantom replicates instead of a single phantom. The former approach was adopted to reduce the time duration of the project, the possibility of damage to, or loss of, a single phantom that is repeatedly transported from one site to another. However, the use of replicates introduces the possibility of variation between concentrations of similar gadolinium compartments in different phantoms. The QSM estimates from Site A for the replicate phantoms are shown in bold in Supporting Information Table S1. The ICC from this dataset is 0.99, which does not differ greatly from that for the entire dataset, and suggests insignificant variation between the different gadolinium phantoms. Another limitation of the study was that only DICOM images was used for reconstructing QSM images. This limited the study to scanner platforms that are known to generate DICOM images without phase artifacts. The use of the raw k-space data can allow this study to be extended to other scanner platforms that are known generate DICOMs with unreliable phase information.
In conclusion, we show that high quality QSM maps can be generated using MEDI+0 for on both preclinical and clinical scanners at several field strengths for a wide range of susceptibility values. Compared to MEDI, MEDI+0 produced improved images of the high susceptibility values at even at high field strengths. An increasing dispersion in measurements at high susceptibilities was observed, and this appeared to be the result of decreased SNR in the magnitude image, which decreases the reliability of the derived noise weighting for the data fidelity term.
Supplementary Material
References
- 1.Liu T, Surapaneni K, Lou M, Cheng L, Spincemaille P, Wang Y. Cerebral microbleeds: burden assessment by using quantitative susceptibility mapping. Radiology. 2012;262(1):269–278. doi: 10.1148/radiol.11110251. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Wei H, Dibb R, Zhou Y, Sun Y, Xu J, Wang N, Liu C. Streaking artifact reduction for quantitative susceptibility mapping of sources with large dynamic range. NMR Biomed. 2015;28(10):1294–1303. doi: 10.1002/nbm.3383. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Chang S, Zhang J, Liu T, Tsiouris AJ, Shou J, Nguyen T, Leifer D, Wang Y, Kovanlikaya I. Quantitative Susceptibility Mapping of Intracerebral Hemorrhages at Various Stages. J Magn Reson Imaging. 2016;44(2):420–425. doi: 10.1002/jmri.25143. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Liu Z, Kee Y, Zhou D, Wang Y, Spincemaille P. Preconditioned total field inversion (TFI) method for quantitative susceptibility mapping. Magn Reson Med. 2017;78(1):303–315. doi: 10.1002/mrm.26331. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Sharma SD, Hernando D, Horng DE, Reeder SB. Quantitative susceptibility mapping in the abdomen as an imaging biomarker of hepatic iron overload. Magn Reson Med. 2015;74(3):673–683. doi: 10.1002/mrm.25448. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Sharma SD, Fischer R, Schoennagel BP, Nielsen P, Kooijman H, Yamamura J, Adam G, Bannas P, Hernando D, Reeder SB. MRI-based quantitative susceptibility mapping (QSM) and R2* mapping of liver iron overload: Comparison with SQUID-based biomagnetic liver susceptometry. Magn Reson Med. 2017;78(1):264–270. doi: 10.1002/mrm.26358. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Dimov AV, Liu Z, Spincemaille P, Prince MR, Du J, Wang Y. Bone quantitative susceptibility mapping using a chemical species-specific R2* signal model with ultrashort and conventional echo data. Magn Reson Med. 2017 doi: 10.1002/mrm.26648. [DOI] [PubMed] [Google Scholar]
- 8.Hinoda T, Fushimi Y, Okada T, Arakawa Y, Liu C, Yamamoto A, Yoshida K, Miyamoto S, Togashi K. Quantitative assessment of gadolinium deposition in dentate nucleus using quantitative susceptibility mapping. J Magn Reson Imaging. 2017;45(5):1352–1358. doi: 10.1002/jmri.25490. [DOI] [PubMed] [Google Scholar]
- 9.Lin PY, Chao TC, Wu ML. Quantitative Susceptibility Mapping of Human Brain at 3T: A Multisite Reproducibility Study. AJNR Am J Neuroradiol. 2014 doi: 10.3174/ajnr.A4137. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Deh K, Nguyen TD, Eskreis-Winkler S, Prince MR, Spincemaille P, Gauthier S, Kovanlikaya I, Zhang Y, Wang Y. Reproducibility of quantitative susceptibility mapping in the brain at two field strengths from two vendors. J Magn Reson Imaging. 2015;42(6):1592–1600. doi: 10.1002/jmri.24943. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Hinoda T, Fushimi Y, Okada T, Fujimoto K, Liu C, Yamamoto A, Okada T, Kido A, Togashi K. Quantitative Susceptibility Mapping at 3 T and 1. 5 T: Evaluation of Consistency and Reproducibility. Invest Radiol. 2015;50(8):522–530. doi: 10.1097/RLI.0000000000000159. [DOI] [PubMed] [Google Scholar]
- 12.Santin MD, Didier M, Valabrègue R, Yahia Cherif L, García-Lorenzo D, Loureiro de Sousa P, Bardinet E, Lehéricy S. Reproducibility of R2 * and quantitative susceptibility mapping (QSM) reconstruction methods in the basal ganglia of healthy subjects. NMR Biomed. 2017;30(4) doi: 10.1002/nbm.3491. [DOI] [PubMed] [Google Scholar]
- 13.Lauzon ML, McCreary CR, McLean DA, Salluzzi M, Frayne R. Quantitative susceptibility mapping at 3 T: comparison of acquisition methodologies. NMR Biomed. 2017;30(4) doi: 10.1002/nbm.3492. [DOI] [PubMed] [Google Scholar]
- 14.de Rochefort L, Liu T, Kressler B, Liu J, Spincemaille P, Lebon V, Wu J, Wang Y. Quantitative susceptibility map reconstruction from MR phase data using bayesian regularization: validation and application to brain imaging. Magn Reson Med. 2010;63(1):194–206. doi: 10.1002/mrm.22187. [DOI] [PubMed] [Google Scholar]
- 15.Liu J, Liu T, de Rochefort L, Ledoux J, Khalidov I, Chen W, Tsiouris AJ, Wisnieff C, Spincemaille P, Prince MR, Wang Y. Morphology enabled dipole inversion for quantitative susceptibility mapping using structural consistency between the magnitude image and the susceptibility map. Neuroimage. 2012;59(3):2560–2568. doi: 10.1016/j.neuroimage.2011.08.082. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Liu T, Spincemaille P, de Rochefort L, Kressler B, Wang Y. Calculation of susceptibility through multiple orientation sampling (COSMOS): a method for conditioning the inverse problem from measured magnetic field map to susceptibility source image in MRI. Magn Reson Med. 2009;61(1):196–204. doi: 10.1002/mrm.21828. [DOI] [PubMed] [Google Scholar]
- 17.de Rochefort L, Brown R, Prince MR, Wang Y. Quantitative MR susceptibility mapping using piece-wise constant regularized inversion of the magnetic field. Magn Reson Med. 2008;60(4):1003–1009. doi: 10.1002/mrm.21710. [DOI] [PubMed] [Google Scholar]
- 18.Erdevig HE, Russek SE, Carnicka S, Stupic KF, Keenan KE. Accuracy of magnetic resonance based susceptibility measurements. AIP Advances. 2017;7(5):056718. [Google Scholar]
- 19.Xie L, Layton AT, Wang N, Larson PE, Zhang JL, Lee VS, Liu C, Johnson GA. Dynamic contrast-enhanced quantitative susceptibility mapping with ultrashort echo time MRI for evaluating renal function. Am J Physiol Renal Physiol. 2016;310(2):F174–182. doi: 10.1152/ajprenal.00351.2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.de Rochefort L, Nguyen T, Brown R, Spincemaille P, Choi G, Weinsaft J, Prince MR, Wang Y. In vivo quantification of contrast agent concentration using the induced magnetic field for time-resolved arterial input function measurement with MRI. Med Phys. 2008;35(12):5328–5339. doi: 10.1118/1.3002309. [DOI] [PubMed] [Google Scholar]
- 21.Xu B, Spincemaille P, Liu T, Prince MR, Dutruel S, Gupta A, Thimmappa ND, Wang Y. Quantification of cerebral perfusion using dynamic quantitative susceptibility mapping. Magn Reson Med. 2014 doi: 10.1002/mrm.25257. [DOI] [PubMed] [Google Scholar]
- 22.Barick KC, Aslam M, Lin Y-P, Bahadur D, Prasad PV, Dravid VP. Novel and efficient MR active aqueous colloidal Fe3O4 nanoassemblies. Journal of Materials Chemistry. 2009;19(38):7023–7029. [Google Scholar]
- 23.Liu T, Wisnieff C, Lou M, Chen W, Spincemaille P, Wang Y. Nonlinear formulation of the magnetic field to source relationship for robust quantitative susceptibility mapping. Magn Reson Med. 2013;69(2):467–476. doi: 10.1002/mrm.24272. [DOI] [PubMed] [Google Scholar]
- 24.Liu Z, Spincemaille P, Yao Y, Zhang Y, Wang Y. MEDI+0: Morphology enabled dipole inversion with automatic uniform cerebrospinal fluid zero reference for quantitative susceptibility mapping. Magn Reson Med. 2017 doi: 10.1002/mrm.26946. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Spincemaille P, Alexey D, Yi W. Correction of Residual Echo-to-Echo Phase Inconsistencies in Readout Phase Corrected Multi-Echo Gradient Echo for Quantitative Susceptibility Mapping. Medical University of Graz; Austria: 2016. [Google Scholar]
- 26.Cusack R, Papadakis N. New robust 3-D phase unwrapping algorithms: application to magnetic field mapping and undistorting echoplanar images. Neuroimage. 2002;16(3 Pt 1):754–764. doi: 10.1006/nimg.2002.1092. [DOI] [PubMed] [Google Scholar]
- 27.Zhou D, Liu T, Spincemaille P, Wang Y. Background field removal by solving the Laplacian boundary value problem. NMR Biomed. 2014;27(3):312–319. doi: 10.1002/nbm.3064. [DOI] [PubMed] [Google Scholar]
- 28.Fortier V, Levesque IR. Phase processing for quantitative susceptibility mapping of regions with large susceptibility and lack of signal. Magn Reson Med. 2017 doi: 10.1002/mrm.26989. [DOI] [PubMed] [Google Scholar]
- 29.Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17(4):571–582. doi: 10.1080/10543400701329422. [DOI] [PubMed] [Google Scholar]
- 30.Koo TK, Li MY. A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research. J Chiropr Med. 2016;15(2):155–163. doi: 10.1016/j.jcm.2016.02.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Weis J, Nilsson S, Ericsson A, Wikström M, Sperber GO, Hemmingsson A. Measurement of magnetic susceptibility and MR contrast agent concentration. Magn Reson Imaging. 1994;12(6):859–864. doi: 10.1016/0730-725x(94)92026-5. [DOI] [PubMed] [Google Scholar]
- 32.Kahn O. Molecular magnetism. Vol. 1993. VCH Publishers, Inc; USA: 1993. p. 393. [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.