Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2019 Dec 1.
Published in final edited form as: Nucl Med Biol. 2018 Oct 15;67:10–14. doi: 10.1016/j.nucmedbio.2018.09.002

Automated Radiosynthesis of 5-[11C]L-Glutamine, an Important Tracer for Glutamine Utilization

Adam J Rosenberg 1,2, Michael L Nickels 1,3, Michael L Schulte 1,3, H Charles Manning 1,7
PMCID: PMC6298813  NIHMSID: NIHMS1510289  PMID: 30359787

Abstract

Introduction:

The natural amino acid L-Glutamine (Gln) is essential for both cell growth and proliferation. In addition to glucose, cancer cells utilize Gln as a carbon source for ATP production, biosynthesis, and as a defense against reactive oxygen species. The utilization of [11C]Gln has been previously reported as a biomarker for tissues with an elevated demand for Gln, however, the previous reports for the preparation of [11C]Gln were found to be lacking several crucial aspects necessary for transition to human production. Namely, the presence of unreacted precursor and the use of noncommercialized, custom built, reaction platforms. Herein, we report the development and utilization of methodology for the automated production of [11C]Gln that meets institutional criteria for human use.

Methods:

The preparation of [11C]Gln was carried out on the the GE FX2N platform. Briefly, after trapping of [11C]HCN with a solution of CsHCO3 in DMF, the [11C]CsCN was reacted with a commercially available precursor. This intermediate was then purified by HPLC and deprotected/hydrolyzed under acidic conditions. Following pH adjustment, the product was filtered to give the desired [11C]Gln as a sterile injectable. The resulting product was then analyzed for quality assurance.

Results:

Automated production by this method reliably provides over 3.7 GBq (100 mCi) of [11C]Gln. The resulting final drug product was found to have a >99% radiochemical purity, <5% of D-Gln present, no detectable impurities, and the total preparation time was roughly 45 min from the end-ofbombardment.

Conclusions:

A fast, reliable and efficient automated radiosynthesis was developed using a commercially available module. Purifications used throughout allow for both a radiochemically and chemically pure final product solution of [11C]Gln.

Keywords: Radiopharmaceuticals, [11C]Glutamine, Automated Radiosynthesis, Metabolism, PET, Carbon-11

Introduction

Positron Emission Tomography (PET) imaging using 2-deoxy-2-[18F]fluoro-D-glucose ([18F]FDG) has become a central component for cancer diagnosis and staging since the first human scan in 1976. Despite the enhanced glycolysis observed in many tumors, elevated glucose utilization also takes place in many normal tissues such as the brain, normal muscle, cardiac tissue, and inflammation.[1, 2] In addition to dependence on glucose, cancer cells can also exhibit a metabolic shift to an increased dependency on the amino acid glutamine (Gln).[3] Gln is also known to participate in protein synthesis and serves as a precursor for glutathione biosynthesis, a cellular defense mechanism against reactive oxygen species.[4, 5] Given this, quantitative measures of glutamine uptake may reflect critical processes in oncology that are difficult to measure using existing imaging metrics. PET imaging agents targeting glutamine uptake, such as [18F]-4-fluoroglutamine ([18F]4F-Gln), have been reported and used in preclinical and clinical studies in oncology.[610] However, [11C]Gln, which is chemically identical to naturally occurring Gln and has the same biological fate.[8]

As a prime example of a non-invasive molecular imaging technique, PET has become increasingly important both for direct clinical utilization and for helping to focus drug development efforts.[1114] In terms of the clinical utilization, the radiopharmaceuticals must be produced under current Good Manufacturing Practices (cGMP) which can be aided by the use of automated reaction modules. These modules, many of which are commercially available, allow for a high degree of reproducibility and standardization for the radiosynthesis. Furthermore, they allow the use of high levels of radioactivity to be utilized since they can be operated remotely while contained in a radiation shielded enclosure (i.e. “hotcell”). Through the use of automated synthesis modules the preparation of a radiopharmaceuticals can be transferred to another production facility with identical equipment with relative ease, avoiding the time-consuming and expensive redevelopment that must occur. Once the radiopharmaceutical has been produced, a series of specific quality control tests must be completed to enable the drug to be released for injection. These tests, known as release criteria, commonly include: pH, residual solvents, bacterial endotoxin, final filter integrity, chemical purity, radiochemical purity, radiochemical identity/purity, and sterility (a post-release test).[15]

While the preparation of [11C]Gln has been previously reported, the methodology utilized required further development to facilitate automation and translation of the drug product from preclinical to clinical use.[7, 1618] The previously published pre-clinical preparations of [11C]Gln utilized a simple solid phase extraction protocol to purify the radiolabeled synthetic intermediate, which was never designed nor shown to be effective at removing the unreacted starting material prior to moving to the final steps of the procedure.[7, 19, 20] The presence of this impurity has the undesired consequence of lowering of the overall chemical purity of the final drug product solution. In addition to the purity concerns, all methodologies reported for the preparation of [11C]Gln have been accomplished using noncommercialized, custom built, reaction platforms. In an effort to allow for the seamless transfer of technology from production site to production site, development of this product on a commercially available platform is a high priority. Furthermore, the previously reported preparations did not address the feasibility of transitioning the [11C]Gln into a clinically ready product, nor were any of the additional preparation metrics addressed (i.e., sterility, chemical purity, pyrogenicity). Herein, we report the development and utilization of methodology for the automated production of [11C]Gln that meets cGMP criteria for human use.

Materials and Methods

All chemicals were obtained from commercial sources and were of analytical, ACS, or USP quality (Sigma-Aldrich, USA; APP Pharmaceutical, USA) and were used without further purification. (S)Tert-Butyl 2-((tert-butoxycarbonyl) amino)-4-Iodobutanoate was purchased from Advanced Biochemical Compounds (ABX, Germany) in pre-aliquoted 2 mg fractions. (S)-Tert-butyl-2-((tertbutoxycarbonyl)amino)-4-cyanobutanoate was synthesized according to literature methods.[21] The CsHCO3/18-crown-6 solution was prepared by mixing 18-crown-6 in acetonitrile (MeCN) (18-crown-6, 160 mg, 0.61 mmol dissolved 17 mL of MeCN) with CsHCO3 in water (H2O)(CsHCO3, 60 mg, 0.31 mmol dissolved in 3 mL of sterile H2O). The mixture was gently mixed to provide a clear and colorless solution, and stored at 10 °C. Solid phase extraction (SPE) cartridges (Sep-Pak® C18 Plus, HLB) were obtained from Waters (Waters® Association, MA, USA), and conditioned with 6 mL 200 proof ethanol (EtOH) followed by 20 mL of sterile H2O prior to use. AG® 11-A8 ion retardation resin was purchased from BIO-RAD, USA. All radioactivity was measured using a CAPINTEC CRC-15 Dual PET, CRC-15 PET, or CRC-25 Dual PET. The radiochemical purity and enantiomeric purity (percentages of L and D isomers) of the final product [11C]L-Gln were determined by analytical radio-HPLC with a Phenomenex® Chirex 3126 (D)-penicillamine column (150 × 4.6 mm,), mobile phase: 1 mM CuSO4 aqueous solution, flow rate 1.0 mL/min, λ = 254 nm, RT of [11C]L-Gln = 9.1 min, RT of [11C]D-Gln = 13.6 min.

Carbon-11 was produced on a commercial cyclotron (GE Healthcare, PETtrace 880) as [11C]CO2 by proton bombardment of a gas target containing a mixture of N2 and O2 gas (99% N2 with 1% O2) to induce the 14N(p,α)11C nuclear reaction. Typical irradiation current used for the bombardment was 60 μA for 20–30 minutes followed by conversion of the [11C]CO2 to [11C]HCN using the GE Chemical Process Cabinet (Procab). Briefly, the [11C]CO2 is transferred out of the target via a stainless steel delivery line, trapped at room temperature on molecular sieves (4Å), which is then heated under a stream of N2/H2 to elute the pure [11C]CO2. This gas mixture is then passed through a nickel catalyst pre-heated to 450 °C, thereby converting the [11C]CO2 to [11C]CH4 which is then mixed with ammonia gas (NH3) and immediately passed through platinum mesh pre-heated to 950 °C providing the desired [11C]HCN gas. The [11C]HCN was then delivered through a teflon delivery line to the desired reaction module for further use.

The synthesis was carried out on a GE FX2N automated reaction module loaded as follows: vial 1: DMF (0.5 mL); vial 2: MeCN (2 mL); vial 3: 70% MeCN in 20 mM phosphate buffer (2.3 mL); vial 4: 70% MeCN in 20 mM Phosphate buffer (1.5 mL); vial 5: precursor 1 ((S)-tert-Butyl 2-((tert-butoxycarbonyl) amino)-4-Iodobutanoate) (2 mg) in DMF (0.2 mL); vial 7: MeCN (2 mL); vial 8: sterile water (4 mL); vial 10: 4:1 TFA/H2SO4 (0.2 mL); vial 11: sterile water (2 mL); vial 13: MeCN (1.5 mL); vial 14: sterile water (10 mL); Reactor 1: CsHCO3 / 18-C-6 solution (1 mL); round-bottom flask (RBF) loaded with sterile water (20 mL); product flask: 1X Phosphate Buffered Saline (PBS) (4 mL); C18 2: C18 Plus Sep-Pak; Al2O3 2: HLB SepPak Plus above a column of AG® 11-A8 Resin. The C18 Plus and HLB Plus Sep-Paks were both preconditioned with 6 mL ethanol, followed by 20 mL sterile water. The AG® 11-A8 resin column was preconditioned with 20 mL 0.5 M NaCl followed by 100 mL of sterile water.

Prior to the end of bombardment (EOB) and delivery of the [11C]HCN, a 1 mL of the CsHCO3/18crown-6 solution described above was added to reactor 1 and azeotropically dried by iterative additions (2X) and evaporations of acetonitrile (1 mL, 90 °C) under vacuum with a helium gas flow. N,NDimethylformamide (DMF) is then added to complete the trapping solution. The [11C]HCN was then delivered to the reaction module through V10 (normally closed position of the valve) with a N2 flow rate of 150 mL/min (Figure 1). Upon completion of delivery, precursor 1 was added as a solution in DMF and the reaction vessel heated at 90 °C for 8 min. The reaction mixture was then cooled by active air cooling, diluted with HPLC mobile phase (70% Acetonitrile in 20 mM aqueous phosphate buffer), and loaded onto an HPLC purification column (Phenomenex, Luna C18(2) 250 × 10 mm). The desired radioactive peak was collected into the round bottom flask containing sterile water, which was then transferred through a C18 Sep-Pak Plus cartridge to isolate the radioactivity. This reaction intermediate is then eluted from the Sep-Pak with acetonitrile (1.5 mL) into the second reaction vessel. The acetonitrile is then azeotropically dried by iterative additions (2X) and evaporations of acetonitrile (1 mL, 90 °C) under vacuum with a helium gas flow. Once completely dry, a solution of trifluoroacetic acid / sulfuric acid (4:1) was added to the reaction vessel and the mixture is heated at 90 °C for 5 min. The resulting solution was diluted with sterile water for injection (2 mL) and passed through an HLB Sep-Pak attached to a AG® 11-A8 resin packed column into the product vial, containing 1X PBS, followed by washing of the column with additional sterile water for injection. The final drug solution is then passed through a 0.22 µm sterilizing filter into the final vial, to provide the [11C]Glutamine as a ready to inject solution. The radiosynthesis requires approximately 45 minutes from the EOB to the drug product solution being deposited in the final vial.

Figure 1:

Figure 1:

Modifications (shown as orange lines and red X’s) to the GE FX2N for the production of [11C]Gln.

Quality Control

Radiochemical and chemical purities were analyzed by analytical HPLC ([11C]Gln RT = 9.1 min)(1 mM CuSO4 (aq), 1 mL/min, 254 nm). The chemical purity of the product was >95%. Analysis for residual organic solvents was carried out using an Agilent 7890B GC (Agilent Technologies Inc., USA) with a capillary column (length 30 m, ID 0.520 mm, DB-WAX 1.0 μm, Agilent Technologies Inc., USA). Apyrogenicity (Charles River Laboratories Inc., USA) tests were performed in-house to ensure that doses of [11C]Gln contained <8.75 endotoxin units (EU) per ml. ColorpHast® pH indicator strips (EMD Chemicals Inc., USA) were used to determine pH of the final product. Final filter integrity testing was performed on the sterilizing filter by standard bubble-point along with sterility testing via direct inoculation into growth media. Radionuclidic purity was confirmed by half-life determination of the final product. Stability testing of the [11C]Gln product was performed at periodic times over two hours, testing for radiochemical purity by HPLC and pH

Results & Discussions

Automated Radiosynthesis Optimization

The radiosynthesis of [11C]Gln is shown in Scheme 1. Our approach was based upon two previous reports of the radiosynthesis of [11C]Gln, as well as a recently reported preparation of [18F]4-F-Gln.[7, 9, 20] Previously [11C]Gln was prepared by alkylation with [11C]HCN of (S)-tert-butyl 2-((tert-butoxycarbonyl) amino)-4-Iodobutanoate, followed by a C18 Sep-Pak trap and release, deprotection/partial hydrolysis with TFA/H2SO4 and finally desalting to provide the final product.[7] This method was then improved upon by Qu and co-workers, with a systematic examination of the trapping and deprotection conditions.[20] Unfortunately, the process that was reported did not allow or account for the purification of the desired product away from the unreacted precursors or any other non-radioactive impurities that were generated throughout the reaction process.

Scheme 1:

Scheme 1:

Initial automated synthetic pathway for the radiosynthesis of [11C]Gln.

Our initial attempts at automation of the literature process started from the optimized procedure, directly applying the procedure on a GE FX2N. Unfortunately in our hands the yields were quite low, typically providing as little as 0.93 GBq (25 mCi) of final product from 59.2 GBq (1600 mCi) of trapped [11C]HCN. Furthermore, the use of TFA/H2SO4 for the second reaction step caused rapid degradation of several permanent components of the reaction module, namely the PEEK connections which were exposed to the solution.

Zhang and co-workers, examining the related automation radiosynthesis of [18F]4-FGln developed a process using the GE FX2N.[9] Small modifications to the modules standard flow-paths allowed for the HPLC purification of the intermediate, and subsequent acidic deprotection to produce the desired final product. We applied this method to the synthesis of [11C]Gln, with a similar reconfiguration of a GE FX2N (Figure 1), that allows for HPLC purification of the intermediate, subsequent trap & release, acidic deprotection, and reaction to give the final product. To overcome the hardware destruction caused by the TFA/H2SO4 solution we replaced the PEEK connection between Vial 10 and Valve 34 with a custommade teflon replacement of exactly the same dimensions as the PEEK connector and removed the PEEK filter located inside of the connector. This teflon replacement has proven to be resistant to the acidic corrosion with over fifty runs to-date. We also added an HLB Sep-Pak Plus in-line with the ionretardation column. The addition of the HLB Sep-Pak Plus added to the inlet of the desalting column was found to be necessary to aid in lowering the overall flow-rate of the product solution through the desalting column, avoiding an over-pressurization scenario and providing lower flow-rate through the resin thereby providing a consistent removal of the acid used in the previous step. This, in theory, has the added benefit of further sequestration of any organic impurities that result from the deprotection performed in the prior step. The final modification was to change the formulation of the final product from 100% water, to 0.4X phosphate-buffered saline. This modification ensures the final product consistently has a pH that is acceptable for human administration.

With the addition of the HPLC purification of the intermediate, not only were we able to produce a chemically pure product, but the radiochemical yield was significantly improved to over 70% decay corrected and to 15% non-decay-corrected (based upon the [11C]HCN trapped). The identity and radiochemical purity of final product was confirmed by HPLC (RT = 9.1 min), with the only detectable radioactive impurity being [11C]D-Gln (RT = 13.6 min). (Figure 2) Furthermore, the process was analyzed by ion-pair HPLC chromatography to determine the level of [11C]L-Glutamate byproduct produced; < 1% [11C]L-Glutamate was observed under these conditions. (Supplementary Figure 1) The results of the three consecutive qualification runs are shown in Table 1.

Figure 2:

Figure 2:

Typical HPLC chromatogram of [11C]Gln product. The top trace (panel A) is the UV absorbance at 254 nm of the [12C]Gln standard; the bottom trace (panel B) is a representative radioactive trace of a final drug product solution.

Table 1:

Results of the Qualification Runs administration as part of two first-in-man clinical trials (NCT03275974 & NCT03263429).

Quality Control Test Requirements for Pass Qual Run 1 Qual Run 2 Qual Run 3
Appearance Clear, colorless, no
particulates
Clear, colorless, no
particulates
Clear, colorless, no
particulates
Clear, colorless, no
particulates
Filter Integrity (Bubble
Point)
Meets pressure specified
by Manufacturer
Pass Pass Pass
pH 4.5 – 8.5 6.5 7.0 7.0
Radiochemical Purity (%) > 90% 97.48% 99.85% 97.00%
Radiochemical Identity
(% difference from
Reference Std)
< 5% 1.99% 0.13% 0.66%
Amount of Product N/A 4.31 GBq (116.6 mCi) 2.84 GBq (76.7 mCi) 4.68 GBq (126.6 mCi)
Residual Acetonitrile
Level, ppm
< 400 3.2 23.3 24.5
Residual DMF Level, ppm < 800 0.0 0.0 0.0
Radionuclidic Identity
(t1/2, minutes)
19.38 – 21.42 min 20.36 min 20.33 min 20.38 min
Bacterial Endotoxin
Levels (EU/mL)
< 8.75 EU/mL < 1.00 EU/mL < 1.22 EU/mL < 1.00 EU/mL
Sterility (Observed
growth after 14 d)
No growth observed in
14 days
No Growth No Growth No Growth

Conclusion

We have successfully developed an automated radiosynthesis of [11C]Gln using a commercially available reaction platform and commercially available materials. Addition of the HPLC purification of the intermediate allows removal of unreacted precursor as well as any other impurities; in addition to allowing for an increased yield and production of [11C]Gln in suitable quantities for clinical use. This allows production of pure [11C]Gln without fear of contamination with the difficult to detect impurities. The final [11C]Gln passes all clinically required release criteria and has been approved for human

Supplementary Material

1

Scheme 2:

Scheme 2:

Optimized Radiosynthesis of [11C]Gln.

Acknowledgements

The authors would like to acknowledge Yiu-Yin Cheung PhD for experimental assistance. The authors wish to acknowledge research support from the Vanderbilt Ingram Cancer Center Support Grant (National Institutes of Health (NIH) National Cancer Institute (NCI) P30CA068485); the Kleberg Foundation; a Vanderbilt Trans-Institutional Program (TIPS) Award to the Vanderbilt Center for Molecular Probes; and the Vanderbilt Digestive Disease Research Center (NIH National Institute of Diabetes and Digestive and Kidney Diseases (NIDDK) P30DK058404); Shared Instrumentation Grant (NIH, 1S10OD019963).

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

Bibliography

  • [1].Tagliabue L and Del Sole A. Appropriate use of positron emission tomography with [18F]fluorodeoxyglucose for staging of oncology patients. European Journal of Internal Medicine;25:6–11. [DOI] [PubMed] [Google Scholar]
  • [2].Ide M Cancer screening with FDG-PET. The quarterly journal of nuclear medicine and molecular imaging : official publication of the Italian Association of Nuclear Medicine (AIMN) [and] the International Association of Radiopharmacology (IAR), [and] Section of the So 2006;50:23–7. [PubMed] [Google Scholar]
  • [3].Wise DR and Thompson CB. Glutamine addiction: a new therapeutic target in cancer. Trends Biochem Sci 2010;35:427–33. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [4].Wellen KE, Lu C, Mancuso A, Lemons JMS, Ryczko M, Dennis JW, et al. The hexosamine biosynthetic pathway couples growth factor-induced glutamine uptake to glucose metabolism. Genes & Development 2010;24:2784–99. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [5].Hassanein M, Hight MR, Buck JR, Tantawy MN, Nickels ML, Hoeksema MD, et al. Preclinical Evaluation of 4-[18F]Fluoroglutamine PET to Assess ASCT2 Expression in Lung Cancer. Mol Imaging Biol 2016;18:18–23. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [6].Lieberman BP, Ploessl K, Wang L, Qu W, Zha Z, Wise DR, et al. PET imaging of glutaminolysis in tumors by 18F-(2S,4R)4-fluoroglutamine. J Nucl Med 2011;52:1947–55. [DOI] [PubMed] [Google Scholar]
  • [7].Qu W, Oya S, Lieberman BP, Ploessl K, Wang L, Wise DR, et al. Preparation and Characterization of l-[5–11C]-Glutamine for Metabolic Imaging of Tumors. Journal of Nuclear Medicine 2012;53:98–105. [DOI] [PubMed] [Google Scholar]
  • [8].Jeitner TM, Kristoferson E, Azcona JA, Pinto JT, Stalnecker C, Erickson JW, et al. Fluorination at the 4 position alters the substrate behavior of l-glutamine and l-glutamate: Implications for positron emission tomography of neoplasias. Journal of Fluorine Chemistry 2016;192:58–67 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [9].Zhang X, Basuli F, Shi Z-D, Xu B, Blackman B, Choyke PL, et al. Automated synthesis of [18F](2S,4R)-4-fluoroglutamine on a GE TRACERlab™ FX-N Pro module. Applied Radiation and Isotopes 2016;112:110–4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [10].Zhou R, Pantel AR, Li S, Lieberman BP, Ploessl K, Choi H, et al. [18F](2S,4R)4-Fluoroglutamine PET Detects Glutamine Pool Size Changes in Triple-Negative Breast Cancer in Response to Glutaminase Inhibition. Cancer Res 2017;77:1476–84. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [11].Gambhir SS. Molecular imaging of cancer with positron emission tomography. Nature Reviews Cancer 2002;2:683. [DOI] [PubMed] [Google Scholar]
  • [12].Matthews Paul M, Rabiner Eugenii A, Passchier J, and Gunn Roger N. Positron emission tomography molecular imaging for drug development. British Journal of Clinical Pharmacology 2011;73:175–86. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [13].Piel M, Vernaleken I, and Rösch F. Positron Emission Tomography in CNS Drug Discovery and Drug Monitoring. Journal of Medicinal Chemistry 2014;57:9232–58. [DOI] [PubMed] [Google Scholar]
  • [14].Lewis DY, Soloviev D, and Brindle KM. Imaging Tumor Metabolism Using Positron Emission Tomography. The Cancer Journal 2015;21:129–36. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [15]. United States Pharmacopeia, Chapter <823>.
  • [16].Rosenberg A, Nickels M, Schulte M, and Manning HC. Automated Clinical Product of [11C]Glutamine on Commercial Synthesis Modules. Journal of Nuclear Medicine 2017;58:332.27587706 [Google Scholar]
  • [17].Padakanti P, Schmitz A, Lee s, Mankoff D, and Mach R synthesis of carbon-11 labelled L-glutamine on Synthra HCN synthesis module. Journal of Nuclear Medicine 2017;58:1025.28473593 [Google Scholar]
  • [18].Nasr K, Xin Y, Rydberg N, Hallgren R, Cai H, and Sun X. Improved Automated Production for Clinical Use of 11C-L-Glutamine. Journal of Nuclear Medicine 2017;58:872. [Google Scholar]
  • [19].Antoni G, Omura H, Ikemoto M, Moulder R, Watanabe Y, and Långström B. Enzyme catalysed synthesis of L‐[4‐11C]aspartate and L‐[5‐11C]glutamate. Journal of Labelled Compounds and Radiopharmaceuticals 2001;44:287–94. [Google Scholar]
  • [20].Gleede T, Riehl B, Shea C, Kersting L, Cankaya AS, Alexoff D, et al. Investigation of SN2 cyanation for base-sensitive substrates: an improved radiosynthesis of l-[5–11C]-glutamine. Amino Acids 2015;47:525–33. [DOI] [PubMed] [Google Scholar]
  • [21].Wu Z, Zha Z, Li G, Lieberman BP, Choi SR, Ploessl K, et al. [18F](2S,4S)-4-(3-Fluoropropyl)glutamine as a Tumor Imaging Agent. Molecular Pharmaceutics 2014;11:3852–66. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

1

RESOURCES