Skip to main content
Sensors (Basel, Switzerland) logoLink to Sensors (Basel, Switzerland)
. 2018 Nov 23;18(12):4106. doi: 10.3390/s18124106

Evaluation of Prestress Loss Distribution during Pre-Tensioning and Post-Tensioning Using Long-Gauge Fiber Bragg Grating Sensors

Sheng Shen 1,2,*, Yao Wang 3, Sheng-Lan Ma 4, Di Huang 1, Zhi-Hong Wu 3, Xiao Guo 3
PMCID: PMC6308587  PMID: 30477164

Abstract

Prestress loss evaluation in prestressed strands is essential for prestressed structures. However, the sensors installed outside the duct can only measure the total prestress loss. The sensors attached on strands inside the duct also have several problems, such as inadequate durability in an aggressive environment and vulnerability to damage during tensioning. This paper proposes a new installation method for long-gauge fiber Bragg grating (LFBG) sensors to prevent accidental damage. Then the itemized prestress losses were determined in each stage of the pre-tensioning and post-tensioning according to the LFBG measurements. We verified the applicability of the LFBG sensors for prestress monitoring and the accuracy of the proposed prestress loss calculation method during pre-tensioning and post-tensioning. In the pre-tensioning case, the calculated prestress losses had less deviation from the true losses than those obtained from foil-strain gauges, and the durability of the LFBG sensors was better than foil-strain gauges, whereas in post-tensioning case, the calculated prestress losses were close to those derived from theoretical predictions. Finally, we monitored prestress variation in the strand for 90 days. The itemized prestress losses at each stages of post-tensioning were obtained by the proposed calculation method to show the prospect of the LFBG sensors in practical evaluation.

Keywords: prestress monitoring, prestress loss, pre-tensioning, post-tensioning, long-gauge fiber Bragg grating, strain distribution

1. Introduction

Prestressing of steel strands provides reversal stress to counteract in-service stress partially or entirely, improve the crack resistance, and reduce the deflection of prestressed structures. Thus, the tensile stress in steel strands can be maintained over time. However, the applied prestress may decrease gradually due to various reasons, such as the stress relaxation in the steel strands, concrete creep and shrinkage, friction between the strand and duct, and deformation of anchoring devices [1]. Moreover, long-term factors including aggressive environment, pitting, stress corrosion, and hydrogen embrittlement can decrease prestressing further and trigger a fracture of the strand that gives rise to accidents, causing the degradation of the nuclear containment vessel [2], the decrease in seismic performance of the concrete frame [3], and the collapse of a bridge [4]. Therefore, measuring and evaluating prestress loss of steel strand is imperative for maintenance and accurate assessment of prestressed structures.

Current sensing techniques for prestress loss measurement can be grouped into two categories: global measurements and local measurements. The global measurements are performed via elastomagnetic sensors [5,6,7], stress waves [8,9], Anchorage-Measurement-Access system [10,11], frequency [12,13], and modal parameters [14]. A common characteristic of the global measurement techniques is that the sensors are usually placed outside anchoring devices to obtain a “global” prestress of the strand. However, this approach has critical drawbacks. First, the “global” value fails to reflect the itemized prestress losses distributing along the strand. Moreover, most global measurements are indirect, complicated in data processing, and disturbed easily by electromagnetic interference. Last, the global measurement does not provide enough sensitivity to detect stress variations that may be quite small compared to the total stress of a prestressed strand due to micro cracks.

To overcome these disadvantages, researchers have focused on the local measurement that place the sensor on the surface of the strand to directly monitor prestresses at some pre-designated points. The local measurement is applied based on some electric sensors, such as strain gauges [15,16], piezoelectric transducer actuator [17] and so on. However, local measurement is more challenging to perform than global measurement due to the following three reasons. First, if the duct has a curved profile between the ends, the sensor or connecting line may be dislocated and damaged by the friction between the strand and duct during tensioning. Second, the interspace is small between the strand and duct. Some sensors, such as the elastomagnetic sensors, cannot be installed into the duct due to their size. The third reason is that grout used to fill the duct after tensioning may trigger a short circuit to the electric sensor without resin isolation.

The above challenges could be overcome using a fiber Bragg grating (FBG) sensor that is more suitable for long-term prestress monitoring than an electric sensor due to its small size, light weight, high stability, and durability. In recent years, FBG sensors have been widely used for dynamic strain-stress and vibration measurements in bridges [18,19,20], scour monitoring [21,22], reinforcement corrosion [23,24], and leakages in concrete structures and pipelines [25,26,27,28]. Two kinds of FBG sensors have been proposed for prestress monitoring. The first type is named as “smart strand” consisting of six helical wires and a core wire embedded an FBG sensor [29,30,31,32,33]. Although the FBG sensor can accurately measure prestress in the core wire, the “smart strand” also has two drawbacks in practice. First, for cost reduction, the position of each FBG sensor in a “smart strand” is predetermined in the production stage, and the distance between adjacent sensors is usually identical. However, these predetermined positions and distances may not match the required positions and distances in practical construction. Second, because the core wire is surrounded closely by six helical wires, it is difficult to connect the embedded FBG sensor in the core wire to the optical cable used to transmit the optical signal.

An improved “smart strand” type sensor has been proposed for monitoring prestress distribution by combining the Brillouin optical time domain analysis/refectory (BOTDA/R) sensor and the FBG sensor along a single optical fiber to solve this problem [34,35]. However, the measurement error of the BOTDA/R sensor was at least dozens of micro-strain, reducing the accuracy of the monitoring data. The results of Zhang’s experiments showed that the strain measurement error of AQ8603 (produced by Ando Electric Co. Ltd., Tokyo, Japan) based on the BOTDR technique was ±130 με (1.96 σ,σ = ±65 με) [36], which is larger than the AQ8603’s nominal precision of ±50 με. The measurement error of the NBX-6000 (produced by Neubrex Co. Ltd., Kobe, Japan) based on pulse-prepump Brillouin Optical Time Domain Analysis (PPP-BOTDA) was about ±80 με (2σ, σ = ±40 με), which is larger than the NBX-6000’s nominal precision of ±25 με [37]. The second type of FBG sensor comprises a grating packaged with a metal capillary [38,39]. Then the capillary-encapsulated FBG sensors are bonded on the surface of the strand by epoxy resin to measure the elongation. The shortcomings of this type of FBG sensors also exist. First, the elastic modulus of steel is far greater than that of resin, and the mechanical strength of most epoxy resins is limited. Thus, mechanical creep may occur in epoxy, when the epoxy is stressed to beyond 50% of its tensile strength [40], and the bonding may fail if the transmitting stress reaches 80% of the ultimate tensile strength of resin [41]. Second, the sensor is proposed to install in the space between two adjacent steel wires [39]. This installation method may lead to accidental damage to the sensor caused by the dislocation of adjacent wires during strand tensioning. Finally, because the strain distributed in each wire is not identical, the measurements from the sensor for a single wire may have a remarkable difference from the true strain of the strand.

We propose that the long-gauge fiber Bragg grating (LFBG) sensor [42] could overcome the abovementioned limitations. The LFBG sensor has a sensing gauge ranging from 0.1 to 1 m suggesting that the measured strain can represent the average elongation of all wires in the gauge length. Meanwhile, the LFBG sensor is packaged by epoxy-soaked fiber reinforced polymer (FRP) material because the elastic modulus of the epoxy-soaked FRP material is less than that of steel so that the bonding can safely transmit the strain from the strand to the sensor. Moreover, the durability of the LFBG sensor has been verified by the fatigue experiment and long-term durability tests in acidic, alkaline, and salt environments [43]. The applicability of the LFBG sensor was also confirmed in practical monitoring for measuring dynamic strain [44] and dynamic displacement [45], and observing the change in the neutral axis position [46]. However, little is known on the monitoring and calculation of prestress losses based on the LFBG sensor.

This paper is organized as follows: Section 2 introduces the structure and design of an LFBG sensor for prestress monitoring in a strand and proposes the installation procedure in practical operation. Based on the strain measurements from the installed LFBG sensors, Section 3 proposes a method to calculate the itemized prestress losses in both pre-tensioning and post-tensioning. Section 4 and Section 5 demonstrate the application of the proposed method experimentally and via in-site monitoring, respectively.

2. The Design and Installation of LFBG Sensors

2.1. Introduction of the LFBG Strain Sensor

The structure of an LFBG sensor proposed by Li [42] is illustrated in Figure 1. A notable feature of this sensor is the use of an embedded and hollow polytetrafluoroethylene tube, inside which an FBG is sleeved and fixed at both ends, and the gauge length of the sensing part can be predetermined.

Figure 1.

Figure 1

The structural design of the packaged LFBG sensor proposed by Li [42].

Moreover, the specific design is advantageous for numerous reasons: (i) The hollow tube used to encapsulate the FBG inside can ensure the strain at each point of the fiber optic is identical, and the measurement from the FBG equals the average strain over the gauge length directly. (ii) A special epoxy resin used to recoat the FBG can effectively prevent the slippage between the bare fiber optic and the epoxy resin. Meanwhile, the strain compatibility can be achieved between the FBG and epoxy resin until the measurement attains the breaking strain. This point is important for high stress-strain measurement in practical prestress monitoring. (iii) The bonding capability of the FRP material with structural materials, such as steel and concrete, is excellent. The FRP also has an excellent long-term durability and stability to prevent degradation due to corrosion and extreme environments.

2.2. Length Design of LFBG Sensor Installed on the Strand

As shown in Figure 2, the strand is composed of a linear core wire and six helical wires. Because the gauge lengths are no longer than several centimeters, most traditional strain sensors, such as the electric resistance strain gauge and the short-gauge FBG sensor, can obtain only the strain of one wire in the strand. However, this measurement cannot represent the true elongation of the strand. Therefore, the sensing part of the LFBG sensor needs to have a length that can touch all the six helical wires in a spiral. For example, the gauge length of an LFBG sensor is about 20–25 cm for stress monitoring to a 7-wire strand. Thus, the entire length of the LFBG sensor can reach 30–35 cm considering that bonding length of each end is set to about 5 cm. Generally, the total thickness of the sensor and the surrounding epoxy resin is less than 3 mm; this thickness is remarkably less than the diameter of a wire and space between the strand and duct.

Figure 2.

Figure 2

Comparison of gauge length between LFBG sensor and the short-gauge sensor.

2.3. Installation Procedure of the LFBG Sensor

The installation procedure is designed to prevent the installed LFBG sensors from accidental failure caused by friction between adjacent strands and between the strand and duct. The restraining block is used to separate adjacent strands and provide space to the sensor. Figure 3a,b show the designed restraining blocks for a 7-strand tendon and 3-strand tendon, respectively. Every block is assembled of two symmetrical parts by connecting bolts. The central hole of the block can contain the core strand and six helical strands that are separated by six grooves. The practical sketch of the restraining blocks on a 7-strand tendon is illustrated in Figure 3c. The distance over 50–60 cm between the two restraining blocks is long enough to contain the LFBG sensors installed on the strands.

Figure 3.

Figure 3

Sketches of the restraining block for (a) 7-strand tendon; (b) 3-strand tendon; and (c) for separating adjacent strands.

As shown in Figure 4a–f, the installation procedure is outlined as follows:

  • (1)

    Mark the corresponding region on the corrugated pipe. Then let the strands pass through the marked corrugated pipe.

  • (2)

    Peel the marked region of the corrugated pipe to expose the inner tendon. Clean the surface of the exposed tendons.

  • (3)

    Install the restraining blocks and tighten the bolts.

  • (4)

    Attach the LFBG sensors on the surface of the strands. The attachment position of the sensor on each outer strand should be pointed at and close to the core strand.

  • (5)

    Let the optical cable pass through a protective sleeve and connect to the sensors.

  • (6)

    Connect the protective sleeve to the corrugated pipe and use epoxy resin to seal off the contact area. Then the protection sleeve inside which the optical cable is placed can be extended away from the corrugated pipe to the nearest vent hole or drain hole.

Figure 4.

Figure 4

The installation procedure of the LFBG sensor. (a) Place the strands through the marked pipe; (b) strip the partial corrugated pipe; (c) install the restraining blocks; (d) attach the LFBG sensors; (e) connect the sensor to the optical cable; and (f) connect the protection sleeve.

3. The Calculation Method for Itemized Prestress Losses Based on the LFBG Measurements

The discussion proposes an optimized sensor configuration for pre-tensioning and post-tensioning and gives the calculation method for itemized prestress losses in both pre-tensioning and post-tensioning based on the LFBG measurements.

3.1. The Itemized Prestress Losses

In the Chinese Code [47], the total prestress loss comprises seven itemized prestress losses named as σl1σl7. σl1 is the anchorage-seating loss. σl2 is the frictional loss containing the loss due to the friction between tendons and duct (σl2,I) and the loss due to draw-in of the wedge (σl2,II). σl3 is the loss due to the temperature difference between the tendon and the abutments in concrete curing. σl4 is the loss due to steel relaxation. σl5 is the loss due to creep and shrinkage of concrete. σl6 is the loss due to the case in which spiral prestressed rebar in annular structure, such as nuclear containment vessel, is extruded by adjacent concrete. σl7 is the loss due to elastic shortening of concrete.

These seven itemized prestress losses are broadly classified into two groups: (1) immediate reductions during prestressing of the tendons and the prestress transferring from the tendons to the concrete members; and (2) time-dependent reductions occurring gradually during the in-service life of the structures. The immediate reductions contain σl1, σl2, σl3, σl6, and σl7. σl4, and σl5 belong to the time-dependent reductions.

3.2. The Case of Pre-Tensioning

As shown in Figure 5a–c, two abutments are fixed securely at both ends of a prestressing bed, and a high-strength steel tendon is pulled between the abutments before the concrete casting. When the concrete attains the required strength for prestressing, the tendon is cut from the abutments, and the prestress is transferred from the tendon to the concrete member through the bond between them. According to the Chinese Code [47], the total loss σ is the sum of several itemized prestress losses shown in Equation (1):

σ=σl1+σl2+σl3+σl4+σl5+σl7  (1)

and the term σl2,I does not exist in pre-tensioning.

Figure 5.

Figure 5

Three stages of pre-tensioning: (a) applying prestress to tendons; (b) casting and curing of concrete member; and (c) cutting of tendon.

A pre-tensioning beam is a typical kind of simply-supported beam. As illustrated in Figure 5, the LFBG sensors can be set on these regions of the tendon as follows: (1) regions near the ends of the beam (R1 and R3); (2) region near the mid-span of the beam (R2). The reason for the former choice is that σl1 and σl2,II constitute the main part of immediate losses, and σl3 and σl4 can be considered to be uniformly distributed along the tendon. The reason for the latter is that the mid-span usually has the maximum moment under the action of daily loads.

Five stages (Stages a–e) exist in a prestressed structure from pre-tensioning to load bearing. Of those, Stages a–d are shown in Figure 5a–c, and Stage e represents the in-service stage of the structure. The measured strains at R1–R3 at Stages a–d are set as εaRi~εdRi (i = 1–3), respectively. The superscript Ri denotes the variable located at Ri (i = 1–3). The subscript a-d implies that the variable is used at Stages a-d, respectively.

At the tensioning stage (Stage a), the relationship between the tensioning force and monitored strain is:

F=AtEtεaRi (i= 13) (2)

where At and Et are the area and elastic modulus of prestressing tendon, respectively. F is the tensioning force obtained by the load cell. The value of Et is determined by a tensile test carried out in the laboratory. No prestress loss occurs at this stage.

σl1 and σl2 occur at Stage b of transferring the tensioning force from jack to prestressing bed. However, it is difficult to divide them without using special measurements, so the sum of σl1 and σl2 is shown as follows:

σl1+σl2=σl1+σl2,IIEt(εaR1εbR1)  (3)

At Stage c of concrete member curing, σl3 and σl4Ri begin to appear. Because the low-relaxation prestressing strand can finish its relaxation in several hundreds of hours, σl4Ri can be obtained entirely in Stage c. Thus, σl3 and σl4Ri can be respectively calculated by Equations (4) and (5):

σl3=EtαtΔt  (4)
σl4Ri=Et(εbRiεcRiαtΔt) (i= 1~3) (5)

At Stage d, the tendons between the beam and the abutment are cut off, and the prestress is resisted by the entire section of the beam. At this time, σl7Ri, which can be obtained by Equation (6) comes into play:

σl7Ri=Et(εcRiεdRi) (i = 1~3) (6)

At the in-service stage (Stage e), σl5Ri can be obtained by Equation (7) when the live load is not applied on the structure:

σl5Ri=Et(εdRiεeRi) (i = 1~3) (7)

Finally, substituting Equations (3)–(7) into Equation (1), the total prestress losses at different locations of the structure are obtained.

3.3. The Case of Post-Tensioning

A remarkable limitation of the pre-tensioning system is that the tendons always have to be straight. However, the post-tensioning system enables the tendons to keep a curved profile before and after tensioning. The ducts inside which the tendons are placed can be fixed to the reinforcements to remain in the desired profile. Then, once the concrete reaches the desired strength, the tendons are tensioned and anchored using external anchors rather than depending on the bond between tendon and concrete as in the pre-tensioning case. Figure 6a–b show the two stages of the post-tensioning procedure.

Figure 6.

Figure 6

The schematic of the two stages of post-tensioning: (a) Application of tensioning to tendons; and (b) Fitting the wedge and cutting the tendon.

According to the Chinese Code [47], the total loss σ is the sum of several itemized prestress losses shown in Equation (8):

σ=σl1+σl2+σl4+σl5+σl6+σl7  (8)

As shown in Figure 6, σl6 is zero. σl2,I exists and does not equal to zero because the profile of the duct is usually curved. σl7 is zero unless the tendons are tensioned batch-wise.

The LFBG sensors can be set on the three regions (R1–R3) shown in Figure 6. In post-tensioning, the curve of the duct profile can be described as a combination of three parabolas. The two linking points between the three parabolas are key points in the design of prestressed structures. Thus, R1 and R3 are set to be near to these two linking points. R2 is set at the mid-span of the beam.

There are three stages (Stages a~c) in a post-tensioned structure from post-tensioning to load bearing. Stages a and b are the tensioning stage and the stage of anchoring, respectively. Stage c is the stage of grout curing and in-service stage of the structure that is shown in Figure 6b.

At Stage a, there are only the frictional losses σl2. In Equation (9), σl2Ri is the frictional loss in Ri:

σl2Ri=F/AtEtεaRi (i= 1~3) (9)

At Stage b, the stress variation equals σl1Ri as follows:

σl1Ri=Et(εaRiεbRi)  (10)

At Stage c, the sum of σl4Ri and σl5Ri is shown in Equation (11). It also needs to be measured without live load action:

σl4Ri+σl5Ri=Et(εbRiεcRi)  (11)

This proposed method is also suitable for prestress loss calculation of more than one tendon in pre-tensioning and post-tensioning. Moreover, it is necessary to keep a distance between the sensor-placed regions near both ends of the strand and the anchor, since the violent variation of stress may threaten the safety of the LFBG sensors in prestress releasing. Finally, all the strain measurements should be updated by temperature compensation.

4. Verification for The Prestress Loss Monitoring Using LFBG Sensor: Experiment

This experiment in this study has two main purposes: to verify the applicability of the LFBG sensor to measure the prestress of a tendon and to investigate the accuracy of the proposed calculation method of prestress loss. In order to correspond to the proposed prestress loss monitoring methods in Section 3, this experiment includes two parts: pre-tensioning test and post-tensioning test.

4.1. Pre-Tensioning Test

4.1.1. Test Design

As shown in Figure 7a, a 7-wire strand inserted into a hollow steel tube was fixed at both abutments of a prestressing bed. The tube was held by two supports and separated from the strand. The strand was placed at the center of the tube. The lengths of the strand and hollow steel tube were 3000 mm and 2500 mm, respectively. The nominal diameter and elastic modulus of the strand were 15.2 mm and 200 GPa, respectively. The outside and inside diameters of the hollow tube were 50 mm and 48 mm, respectively. Three monitored regions (R1–R3) with a uniform length of 250 mm were set from the right end to the mid-point of the strand, and the distance between adjacent regions was 250 mm. Three LFBG sensors (S1–S3) with a uniform gauge length of 250 mm were placed on R1–R3, respectively. In each region, each helical wire was attached to a foil strain-gauge (FSG) to measure the strain precisely. The numbering rule is as follows. For example, the 6 FSGs in R1 are named as E11–E16. The first number means the sensor is in R1 and the second number represents the number of the wire. Details are given in Figure 7c about the sensor placement on the wire.

Figure 7.

Figure 7

Figure 7

The sketches and photographs of the test design: (a) applying tensioning to the bare strand; (b) applying load on the cement-filled steel tube; (c) the photograph of the strand attached with sensor; (d) the photograph of tensioning and (e) the photograph of vertical load applied on the cement-filled steel tube.

The process of loading can be divided into three steps (Step I–III). At Step I, the increasing load F was applied by a jack to tension the strand through eight successive loading steps from 0 to 156 kN with an increment of 20 kN. After the final loading step, tensioning reduces from 156 kN to 149.8 kN because of the anchorage-seating loss and the loss due to draw-in of the wedge. At the beginning of Step II, the tube was filled with grout. When the strength of the grout exceeded 50 MPa (48 h after tensioning), the supports were removed. At Step III, a vertical load P was divided equivalently into two parts by a transferring steel board and applied at two points 500 mm away from both ends of the tube. P was increased with a loading step of 3 kN from 0 kN to 24 kN. Details about Step II and Step III are respectively shown in Figure 7b,d,e. The strains of R1–R3 in each helical wire at Steps I-III were measured by FSGs and LFBG sensors, respectively. In addition, the temperature was kept constant during the entire experiment to avoid the expansion or contraction of the abutments and strand. All measured data were updated by temperature compensation.

4.1.2. Results and Analysis

Table 1 gives the measured strains from the FSGs, the average values of the measured strains from the FSGs, and the measured strains from the LFBG sensors at Step I. Table 2 shows the comparison between the true stresses, the stresses calculated from the average values of the measured strains from the FSGs, and the stresses calculated from the strains obtained from the LFBG sensors. Four remarkable features are notable: First, most calculated stresses from sensor measurements are lower than the true stresses, the result ascribed to the gap between the adjacent wires. Second, the difference between the maximum and minimum strains of the six helical wires in the same region can approach or exceed 10% of the applied strain.

Table 1.

The monitored strains of the strand at Step I. (Unit: με).

Region F/kN 20 40 60 80 100 120 140 156 149.8
R1 E11 817 1487 2189 2879 3540 4201 4791 5246 5053
E12 520 1042 1656 2331 3045 3734 4400 5013 4708
E13 720 1314 1971 2670 3336 3978 4583 5124 4936
E14 671 1291 1979 2682 3381 4055 4692 5244 5080
E15 795 1576 2305 3014 3707 4374 5006 5550 5271
E16 716 1348 2028 2714 3393 4044 4665 5214 5095
Average strain(FSG) * 707 1343 2021 2715 3400 4064 4689 5232 5024
S1 713 1426 2129 2808 3524 4237 4940 5482 5270
R2 E21 695 1353 2013 2674 3356 3991 4593 5156 4929
E22 616 1254 1903 2553 3231 3857 4452 5004 4725
E23 753 1420 2076 2728 3410 4035 4631 5184 4976
E24 616 1211 1825 2456 3136 3753 4349 4892 4684
E25 705 1385 2062 2740 3460 4105 4727 5279 5149
E26 687 1329 1973 2616 3302 3911 4500 5031 4828
Average strain(FSG) 678 1325 1975 2628 3316 3942 4542 5091 4882
S2 685 1388 2035 2776 3425 4077 4814 5238 5028
R3 E31 707 1382 2080 2760 3384 4060 4716 5236 5014
E32 683 1359 2063 2756 3391 4073 4739 5265 5044
E33 705 1318 1959 2592 3187 3824 4456 4946 4826
E34 737 1425 2129 2820 3465 4135 4801 5308 4986
E35 621 1298 2004 2698 3350 4020 4690 5202 4979
E36 694 1379 2090 2788 3447 4120 4800 5307 5082
Average strain(FSG) 691 1360 2054 2736 3371 4039 4700 5211 4989
S3 711 1407 2064 2764 3500 4186 4900 5356 5128

* Average strain (FSG) represents the average value of the strains from the six FSGs obtained from the same region.

Table 2.

The measurement errors in calculated stresses from the strains obtained from the FSGs and LFBG sensors.

F/kN 20 40 60 80 100 120 140 156 149.8
True Stress/MPa 143.0 285.7 428.6 571.4 714.3 857.1 1000.0 1114.3 1070.0
R1 Stress(FSG) */MPa 141.4 268.6 404.2 543.0 680.0 812.8 937.8 1046.4 1004.8
Error/% −1.1 −6.0 −5.7 −4.8 −4.6 −6.2 −5.9 −6.1 −6.1
Stress(LFBG) **/MPa 142.6 285.2 425.8 561.6 704.8 847.4 988.0 1096.4 1054.0
Error/% −0.3 −0.2 −0.7 −1.7 −1.3 −1.1 −1.2 −1.6 −1.5
R2 Stress(FSG)/MPa 135.6 265.0 395.0 525.6 663.2 788.4 908.4 1018.2 976.4
Error/% −5.0 −7.3 −7.8 −8.0 −7.2 −8.0 −9.2 −8.6 −8.7
Stress(LFBG)/MPa 137.0 277.6 407.0 555.2 685.0 815.4 962.8 1047.6 1005.6
Error/% −4.0 −2.8 −5.0 −2.8 −4.1 −4.9 −3.7 −6.0 −6.0
R3 Stress(FSG)/MPa 138.2 272.0 410.8 547.2 674.2 807.8 940.0 1042.2 997.8
Error/% −3.5 −4.8 −4.2 −4.2 −5.6 −5.8 −6.0 −6.5 −6.7
Stress(LFBG)/MPa 142.2 281.4 412.8 552.8 700 837.2 980.0 1071.2 1025.6
Error/% −0.6 −1.5 −3.7 −3.3 −2.3 −1.9 −2.0 −3.9 −4.1

* Stress (FSG) is the stress calculated from the average strain (FSG) in Table 1. ** Stress (LFBG) is the stress calculated from the measured strains by LFBG sensors in Table 1.

This result implies that the strain measured in one wire only does not necessarily represent the elongation of the whole strand. Moreover, the measured strains from the LFBG sensors are between the maximum and minimum strains from the FSGs and close to the average values of the measured strains from the FSGs. This result is attributed to the fact that LFBG can acquire the average elongation of six helical wires because the sensing part of the LFBG can cover the six wires.

Finally, most of the differences between the true stresses in the strand and the calculated stresses by strains from LFBG sensors are less than 5%. The maximum error is 6.0%. However, the differences between the true stresses in the strand and the calculated stresses from the average values of the measured strains from the FSGs are over 5%. The maximum error is up to 9.2%. It appears that the errors in strains from LFBG sensors are approximately half of those obtained from FSGs. An important reason behind this effect is that the FSGs attached on the helical wires are not parallel with the tensioning direction. In summary, these features show LFBG sensor is better suited to monitor prestress in the strand than common FSG.

Table 3 gives the monitored strains from different sensors at Step II. The comparison between the true stresses and the calculated stresses are shown in Table 4. Most of the losses calculated from the strains obtained from the LFBG sensors are approximate to or less than the corresponding losses calculated from the average values of the measured strains from the FSGs. Moreover, some FGSs become invalid at the end of this step. It further shows that the traditional electrical sensor may not satisfy the requirements of “local measurement in duct” due to the lack of long-term durability.

Table 3.

The monitored strains of the strand at Step II. (Unit: με).

Time/Hour 0 1 2 3 12 24 48
F/KN 149.8 149.55 149.35 149.24 149.1 148.95 148.9
R1 E11 5053 5046 5040 5038 5032 5015 5009
E12 4708 4697 4689 4685 4674 4662 4656
E13 4936 4928 4921 4920 4918 - -
E14 5080 5066 5061 5058 5044 5027 5022
E15 5271 5265 5261 5257 5249 5234 5226
E16 5095 5087 5081 5077 5068 5052 5044
Average strain of E11–E16 5024 5015 5009 5006 4998 4998 4991
S1 5270 5260 5253 5249 5242 5238 5234
R2 E21 4929 4918 4909 4904 4893 4880 4865
E22 4725 4720 4717 4715 4709 -
E23 4976 4967 4960 4958 4951 4931 4909
E24 4684 4673 4666 4662 4657 4633 4624
E25 5149 5143 5138 5135 5130 5113 5094
E26 4828 4820 4814 4810 4801 4777 4764
Average strain of E21–E26 4882 4874 4867 4864 4857 4867 4851
S2 5028 5019 5013 5009 5003 4997 4995
R3 E31 5014 5005 4997 4993 4989 4983 4980
E32 5044 5033 5025 5018 5008 5005 5002
E33 4826 4817 4810 4808 4801 4799 4795
E34 4986 4978 4972 4970 4967 4957 4951
E35 4979 4971 4964 4961 4957 4952 -
E36 5082 5070 5063 5058 5054 5048 5044
Average strain of E31–E36 4989 4979 4972 4968 4963 4957 4954
S3 5128 5117 5112 5108 5103 5097 5094
Table 4.

The stresses and the measurement errors calculated from the strains obtained from the FSGs and LFBG sensors.

Time/Hour 1 2 3 12 24 48
True stress/MPa 1.78 3.2 4.0 5.0 6.07 6.43
R1 Stress(FSG)/MPa 1.8 3 3.6 5.2 5.2 6.6
Error/% 1.1 −6.3 −10.0 4.0 4.0 2.6
Stress(LFBG)/MPa 2 3.4 4.2 5.6 6.4 7.2
Error/% 12.4 6.3 5.0 12.0 5.4 12.0
R2 Stress(FSG)/MPa 1.6 3 3.6 5 3 6.2
Error/% −10.1 −6.3 −10.0 0 50.5 −3.6
Stress(LFBG)/MPa 1.8 3 3.8 5 6.2 6.6
Error/% 1.1 −6.3 −5.0 0 2.1 2.6
R3 Stress(FSG)/MPa 2 3.4 4.2 5.2 6.4 7
Error/% 12.4 6.3 5.0 4.0 5.4 8.9
Stress(LFBG)/MPa 2.2 3.2 4 5 6.2 6.8
Error/% 23.6 0 0 0 2.1 5.8

Before the beginning of Step III, the values of P, F, and sensors are set to zero again based on the measured data in Step II. The measured strains at each loading step of Step III are shown in Table 5. Note that the relationship between F and strand strains does not match Equation (1) anymore because the force is undertaken by not only strand but also solid grout and steel tube. There are two notable characteristics in Table 5. On the one hand, the number of damaged FSGs grows with increasing load. By contrast, all LFBG sensors can measure the strains well in the entire loading process. On the other hand, although the strains in R1–R3 should be close in theory, the difference between the measured strains from S1–S3 is less than 10% only in the case of P ≤ 12 kN. This phenomenon may be attributed to the crack in grout occurring near R1 and R2. The chief reason for that is when P = 12 kN, the measured strain is about 150 με, which is close to the threshold of the tensile strain of most concrete. This phenomenon shows the durability of LFBG sensors in the case of grout cracking in practical prestress monitoring.

Table 5.

The monitored strains of the strand at Step III. (Unit: με).

P/kN 3 6 9 12 15 18 21 24
F/kN 0.6 2.2 5.8 10.1 15.1 20.8 26.7 32.7
R1 E11 14 46 89 145 230 318 - -
E12 16 63 105 155 258 376 516 682
E13 - - - - - - - -
E14 6 34 73 123 - - - -
E15 14 48 103 - - - - -
E16 8 42 88 153 233 - - -
Average strain(FSG) 12 47 92 144 240 347 516 682
S1 12 47 97 155 252 367 508 662
R2 E21 6 32 75 130 195 279 379 482
E22 - - - - - - - -
E23 11 49 95 150 231 315 - -
E24 11 43 92 141 221 312 425 551
E25 19 59 118 172 - - - -
E26 22 66 116 174 260 365 - -
Average strain(FSG) 14 50 99 153 227 318 402 517
S2 11 47 93 149 234 335 456 578
R3 E31 8 44 91 141 206 275 373 480
E32 14 50 98 157 236 - - -
E33 9 42 81 122 - - - -
E34 19 61 - - - - - -
E35 - - - - - - - -
E36 6 33 74 124 196 273 384 -
Average strain(FSG) 11 46 86 136 213 274 379 480
S3 12 42 89 141 211 297 396 500

Based on Equations (3)–(7), Table 6 lists the itemized prestress losses calculated from the monitored strains. Because the temperature was kept constant during the test, σl3 is zero. In addition, σl7 is zero because the strand was not cut. σl5 is also zero because the interval between grout curing and loading was short. Compared with the losses calculated from the average values of the measured strains from the FSGs, the losses calculated from the measured strains from the LFBG sensors were closer to the true losses. The error in the latter is only half of that in the former. This result verifies that LFBG sensor has more accuracy than traditional FSG in prestress monitoring.

Table 6.

Comparison of different prestress losses in the pre-tensioning experiment.

Itemized Prestress Loss σl1 + σl2, II σ l2, I σ l3 σ l4 σ l5 σ l6 σ l7 Total Loss
True loss/MPa 44.3 - * 0 6.4 0 ** - 0 50.7
Loss(FSG) Value/MPa 41.6 - 0 6.6 0 - 0 48.2
Error/% −6.1 - 0 3.1 0 - 0 −4.9
Loss(LFBG) Value/MPa 42.4 - 0 6.9 0 - 0 49.3
Error/% −4.3 - 0 7.8 0 - 0 −2.8

* “-” denotes this loss does not exist in pre-tensioning case. ** “0” denotes that σl5 is small because the interval between grout curing and loading is short.

4.2. Post-Tensioning Test

4.2.1. Test Design

Details of the dimensions and reinforcement configuration of a simply-supported beam used in the experiment are shown in Figure 8a. The total length of the beam was 6000 mm, with a span of 5400 mm. The cross-section had a rectangular shape with 220 mm width and 450 mm depth. The compressive strength of the concrete was about 39 N/mm2. The elastic modulus and Poisson’s ratio of the concrete were 3.03 × 104 N/mm2 and 0.19, respectively. A curved duct with a diameter of 50 mm was embedded into the beam. The process of the test can be divided into two steps (Step I–II). At Step I, a 3-strand tendon was passed through the duct and tensioned by the jack with an increasing load from 0 to 120, 240, 360, 480 and 540 kN. To counteract the frictional loss, we overloaded the final tension force to 555 kN that is 3% higher than 540 kN. Because of the effect of the anchorage-seating loss σl1, the tensioning force reduced from 555 kN to 426.6 kN in the case of anchoring. At Step II, it took 72 h to observe the steel relaxation and creep of the concrete and for the strength of grout to reach 50 MPa.

Figure 8.

Figure 8

Sketches and photographs of the experiment. (a) Applied tensioning to tendons; and (b) photographs of the restraining block and tensioning.

As shown in Figure 8a, we selected five regions located at the right linking point, 1/3 span, mid-span, 2/3 span, and left linking point (R1–R5) of the top strand for monitoring. Five LFBG sensors with a uniform gauge length of 250 mm were placed on R1–R5. 30 FSGs were successively attached on R1–R5 of the six helical wires of the strand. The numbering mode for the LFBG sensors and FSGs was the same as that in the last test. More details about the sensor installation and loading are illustrated in Figure 8b.

4.2.2. Results and Analysis

Different from the pre-tensioning case, the true stresses on R1–R5 cannot be obtained due to the frictional prestress loss distribution. Therefore, the theoretical prestress loss predictions by the Chinese Code [47] are used to replace the true prestress losses in the comparisons of σl2 and σl1. Table 7 gives the monitored strains from the FSGs and LFBG sensors at Step I. The differences between the maximum and minimum strains obtained from different wires also reaches or exceeds 10% of the applied strain. Assuming the average strain is approximate to the true strain, the large dispersion in the strain measurements demonstrates that the traditional short-gauge strain sensor is not suitable in prestress monitoring. Because the extending direction of the LFBG sensors is parallel to the tensioning force, most of the strains from LFBG sensor are larger than the average strains from FSG measurements at each loading step. This phenomenon also occurs in the previous test of pre-tensioning. Using Equation (9), Table 8 shows σl2 and σl1 calculated from LFBG sensor measurements at each loading step. Figure 9a,b show the comparisons between σl2 and σl1 calculated from the measured strains from the LFBG sensors and the theoretical predictions of prestress losses by the Chinese Code [47]. The developing trend of the calculated σl2 and σl1 are close to the loss profiles of theoretical predictions although there are some deviations between the calculated values and the predictions. This result also proves that the calculations for σl2 and σl1 are correct based on the strain measurements from the LFBG sensors.

Table 7.

The monitored strains of the strand at Step I. (Unit: με).

F/kN 120 240 360 480 540 555 426.6
R1 E11 1371 2775 4193 5682 6368 6471 5214
E12 1336 2704 4111 5532 6218 6343 5111
E13 1300 2625 3993 5368 6046 6171 4807
E14 1257 2543 3786 5250 5846 5954 4968
E15 1325 2689 4161 5504 6243 6368 5125
E16 1382 2800 4229 5686 6371 6494 5227
Average strain(FSG) 1329 2689 4079 5504 6182 6300 5075
S1 1321 2689 4089 5529 6200 6336 5211
R2 E21 1357 2729 4179 5589 6289 6421 5546
E22 1318 2657 4079 5471 6132 6264 5439
E23 1264 2554 3896 5250 5932 6075 5271
E24 1236 2504 3807 5154 5825 5936 5154
E25 1289 2606 4006 5392 6079 6232 5451
E26 1361 2721 4204 5500 6236 6396 5496
Average strain(FSG) 1304 2629 4029 5393 6082 6221 5393
S2 1314 2657 4046 5461 6111 6250 5439
R3 E31 1321 2693 4125 5461 6150 6261 5461
E32 1296 2639 4046 5354 6036 6146 5375
E33 1254 2579 3893 5200 5896 6000 5246
E34 1229 2475 3829 5161 5686 5871 5164
E35 1243 2575 3950 5236 5875 5979 5229
E36 1351 2707 4115 5456 6142 6261 5432
Average strain(FSG) 1282 2611 3993 5311 5964 6086 5318
S3 1279 2607 3975 5307 5957 6061 5343
R4 E41 1282 2646 4025 5321 5986 6082 5404
E42 1279 2629 4036 5318 6007 6107 5421
E43 1204 2536 3793 5039 5646 5796 5129
E44 1189 2414 3743 5036 5579 5654 5079
E45 1218 2568 3871 5161 5807 5889 5275
E46 1329 2721 4057 5432 6114 6215 5575
Average strain(FSG) 1250 2586 3921 5218 5857 5957 5314
S4 1243 2571 3893 5182 5821 5911 5304
R5 E51 1236 2389 3861 5246 5814 5971 5689
E52 1246 2550 3936 5150 5761 5864 5618
E53 1186 2525 3900 5089 5721 5861 5518
E54 1182 2507 3789 5029 5650 5789 5554
E55 1150 2404 3257 4618 5193 5261 5018
E56 1286 2689 4097 5254 5975 6096 5731
Average strain(FSG) 1214 2511 3807 5064 5686 5807 5521
S5 1221 2525 3825 5079 5729 5829 5557
Table 8.

σl2 and σl1 calculated from the strains of the LFBG sensors at Step I. (Unit: MPa).

Itemized Losses σ l2 σ l1
P/kN 120 240 360 480 540 555 426.6
0 * - - - - - - 305.7
R1 20.8 32.2 37.2 34.2 45.0 53.8 225.0
R2 22.2 38.6 45.8 47.8 62.8 71.0 162.2
R3 29.2 48.6 60.0 78.6 93.6 108.8 143.6
R4 36.4 55.8 76.4 103.6 120.8 138.8 121.4
R5 40.8 65.0 90.0 124.2 139.2 155.2 54.4

* “0” is the zero point in the coordinate that represents the right end of the beam in Figure 8a.

Figure 9.

Figure 9

The comparisons of the two calculated losses and the predictions according to the Chinese Code [44]: (a) σl2; and (b) σl1.

At Step II, Table 9 lists the strain decrements of R1–R5 in 72 h due to (1) steel relaxation of the strands; (2) creep and shrinkage of the concrete. The sum of the corresponding prestress losses, σl4 and σl5, are listed in Table 10. Compared with the data in Table 4, the losses in Table 10 imply that the major parts of the losses are triggered by the creep and shrinkage of concrete. Also, several FSGs are observed to be invalid in this step owing to the immersion of grout. The fact that the FBG sensors can measure during 72 h of grout curing again shows the durability to the sensing part of the LFBG sensor.

Table 9.

The results of strain measurement from various sensors at Step II. (Unit: με).

Time/Hour 1 2 3 12 24 48 72
RR1 E11 5130 5111 5102 5052 5033 4999 4976
E12 5016 4988 4976 4923 4895 4862 4853
E13 4713 4687 4672 4609 4570 4544 4542
E14 4888 4868 4853 4790 4777 4752 4738
E15 5038 5012 5005 4955 4927 4899 4876
E16 5129 5097 5090 5021 4993 4955 4949
Average strain(FSG) 4985 4960 4949 4891 4865 4835 4822
S1 5141 5121 5113 5056 5026 4987 4970
RR2 E21 5450 5423 5411 5338 5329 5302 5275
E22 5348 5322 5309 5256 5231 5202 5183
E23 5182 5156 5137 5085 5057 5030 5006
E24 5068 5048 5039 4980 4963 4932 4905
E25 5362 5332 5317 5272 5256 5218 5197
E26 5402 5382 5377 5313 5293 5252 5248
Average strain of E21–E26 5302 5277 5266 5208 5189 5156 5136
S2 5366 5340 5324 5251 5223 5198 5184
RR3 E31 5407 5390 5377 5319 5310 5282 -
E32 5319 5295 5276 5245 5209 5180 5166
E33 5184 5162 5143 5097 5076 5044 5022
E34 5114 5097 5091 5041 5017 4986 -
E35 5175 5154 5144 5100 5082 5049 5029
E36 5377 5358 5350 5298 5287 5266 5246
Average strain of E31–E36 5263 5243 5230 5183 5164 5135 5116
S3 5266 5241 5231 5167 5155 5131 5118
RR4 E41 5309 5283 5270 5210 5174 5151 5126
E42 5321 5290 5281 5218 5181 5155 5121
E43 5035 5011 5004 4946 4907 4883 4848
E44 4991 4977 4964 4909 4876 4851 4825
E45 5184 5152 - - - - -
E46 5484 5454 5431 5371 5346 5319 5271
Average strain of E41–E46 5221 5195 5190 5131 5097 5072 5038
S4 5215 5191 5180 5108 5070 5021 4999
RR5 E51 5580 5544 5529 5451 5430 5392 5375
E52 5519 5507 5491 5447 5411 5380 5345
E53 5422 5406 5387 5310 5281 5252 -
E54 5454 5423 - - - - -
E55 4914 4878 4869 4808 4771 4736 4712
E56 5631 5592 5565 5562 5544 5504 5479
Average strain of E51–E56 5420 5392 5368 5316 5287 5253 5228
S5 5459 5432 5407 5363 5323 5271 5239
Table 10.

The sum of σl4 and σl5 calculated from the strains obtained from the LFBG sensors. (Unit: MPa).

Itemized Losses σ l4 + σ l5
Time/hour 1 2 3 12 24 48 72
0 * 17.3 21.7 26.1 35.7 42.1 45.3 46.3
R1 14.0 18.0 19.6 31.0 37.0 44.8 48.2
R2 14.6 19.8 23.0 37.6 43.2 48.2 51.0
R3 15.4 20.4 22.4 35.2 37.6 42.4 45.0
R4 17.8 22.6 24.8 39.2 46.8 56.6 61.0
R5 19.6 25.0 30.0 38.8 46.8 57.2 63.6

* “0” is the zero point in the coordinate that represents the right end of the beam in Figure 8a.

5. Verification for the Prestress Loss Monitoring Using LFBG: In-Site Monitoring

The specific aims regarding in-site monitoring are to observe whether the LFBG sensors installed by the proposed installation procedure in Section 2.3 are valid in long-term monitoring, and to obtain the itemized prestress losses in practical post-tensioning to ensure the security of the construction.

5.1. Member Fabrication and Sensor Placement

This practical prestress loss monitoring was based on the project of Multifunctional Drama Hall of the Fuzhou Straits Cultural Art Center in Fujian Province, China. The monitored beam had a length of 18.475 m. The width and the height of the beam were 1.3 m and 0.7 m, respectively. Details of the dimensions and reinforcement configuration are shown in Figure 10. Three 7-strand tendons were passed the curved duct, and all strands in three tendons had an identical nominal diameter of 15.2 mm. The ultimate tensile strength and elastic modulus of the strand were 1860 MPa and 200 GPa, respectively. The tensioning was controlled by the oil-pressure gauge of the jack used to apply tensioning to the tendons. There are three loading steps including 10 MPa, 20 MPa, and 28 MPa shown in the oil-pressure gauge to apply the prestresses of 358.40 MPa, 728.0 MPa, and 1023.7 MPa to the tendon, respectively. The applied prestress in the tendon at the final loading step contained an overstressing of 3% to counteract the frictional loss. Due to the anchorage-seating loss σl1, the applied prestress decreased from 1023.7 MPa to 862.3 MPa.

Figure 10.

Figure 10

The schematic of the monitored beam in in-site measurement.

The six outer strands of the mid-tendon named as w1–w6 were used for prestress monitoring. As shown in Figure 10, two monitored regions were set nearby the linking points in the strands. 12 LFBG sensors were fixed in the regions. The numbering mode for the LFBG sensors is as follows. The six LFBG sensors on R1 of W1–W6 are named as S11–S16. The first number means the sensor is in R1, and the second number represents the number of the strand. The other six LFBG sensors on R2 of W1–W6 are named as S21–S26. At tensioning, the sensors measured the strain increments of each strand in each loading step. When the tensioning was finished, and the prestress began to transfer to the beam, the strain data were measured continuously. The entire monitoring lasted about 90 days, and all data were updated by temperature compensation.

Moreover, all LFBG sensors were installed by the installation method proposed in Section 2.3. Figure 11a–d show the whole process of sensor placement.

Figure 11.

Figure 11

Photographs of the LFBG sensor deployment in in-site monitoring: (a) Peeling the marked region to expose the tendon; (b) attaching the LFBG sensors on the strands; (c) connecting the sensor to the optical cable; and (d) connecting the protection sleeve to the pipe.

5.2. Results and Analysis

Table 11 lists the measured strains of the six helical strands and the average strains at each loading step in tensioning. The strains distributing in the strands are quite dispersed, and the deviation between the maximum and minimum strain can reach 50% of the average applied strains. This phenomenon implies that some strands such as w2 and w3 were still loose before the tensioning was applied. Moreover, the strand w1 may be ruptured earlier than other strands because it undertakes the extra prestress. Thus, a remarkable conclusion that can be derived from the data shown in Table 11 is that monitoring prestress in a tendon merely from the measured strains in one of the six helical wires using the traditional “short-gauge” strain sensor may ignore the uneven strain distribution in the same section. Table 12 and Table 13 give σl2 and σl1 calculated from the measured strains and their prediction based on the Chinese Code [47]. The fact that the calculated losses are close to the predictions implies that the prestress loss calculations are correct in principle based on the strain measurements from the LFBG sensors.

Table 11.

The monitored strains of the strands at tensioning. (Unit: με).

Applied Stress/MPa 358.4 728.0 1023.7 862.3
R1 S11 2456 4938 6755 6196
S12 1320 3067 4254 3606
S13 1428 2941 4325 3972
S14 1621 3461 4988 4298
S15 1913 3911 5495 4805
S16 2066 3638 5091 4611
Average strain of S11–S16 1801 3659 5151 4581
R2 S21 2361 4705 6568 6561
S22 1074 2532 3643 3628
S23 759 1771 3003 2977
S24 1945 3832 5162 5151
S25 1888 3609 5152 5134
S26 2027 3898 4949 4931
Average strain of S21–S26 1676 3391 4746 4730

Table 12.

σl2 and σl1 calculated from the strains of the LFBG sensors. (Units: MPa).

Itemized Prestress Losses σ l2 σ l1
Applied stress 358.4 728.0 1023.7 862.3
R1 w1 −120.5 −234.9 −293.5 109
w2 101.0 129.9 194.2 126.3
w3 79.9 154.5 180.3 68.9
w4 42.3 53.1 51.0 134.6
w5 −14.6 −34.6 −47.8 134.5
w6 −44.5 18.4 31.0 93.6
Average(LFBG) * 7.3 14.4 19.2 111.2
Prediction(Code) ** 7.9 16.0 22.5 108.4
R2 w1 −102.0 −189.5 −257.1 1.4
w2 149.0 234.3 313.3 2.9
w3 210.4 382.7 438.1 5.1
w4 −20.9 −19.2 17.1 2.2
w5 −9.8 24.2 19.1 3.5
w6 −36.9 −32.1 58.6 3.6
Average(LFBG) * 31.6 66.7 98.2 3.1
Prediction(Code) ** 29.4 59.7 83.9 0

* represents the average stresses or prestress losses calculated from strains obtained from the LFBG sensors. ** the values are estimated based on the Chinese Code [47].

Table 13.

The monitored strains of the strands at the in-service stage. (Unit: με).

Time R1 R2
W1 W2 W3 W4 W5 W6 Average W1 W2 W3 W4 W5 W6 Average
0 6196 3606 3972 4298 4805 4611 4581 6561 3628 2977 5151 5134 4931 4730
12 h 6052 3461 3875 4137 4681 4409 4436 6343 3454 2833 4961 4990 4800 4564
24 h 6031 3443 3857 4121 4663 4387 4417 6329 3436 2815 4950 4973 4782 4548
36 h 6020 3434 3838 4116 4654 4363 4404 6321 3427 2806 4937 4964 4771 4538
48 h 6011 3425 3819 4112 4645 4344 4393 6315 3418 2797 4923 4955 4766 4529
9 days 5959 3356 3780 4069 4582 4280 4338 6252 3353 2749 4876 4904 4714 4475
30 days 5862 3225 3717 3983 4496 4183 4244 6154 3236 2636 4776 4813 4626 4374
51 days 5810 3163 3695 3953 4459 4164 4207 6119 3209 2606 4718 4773 4599 4337
72 days 5794 3151 3686 3933 4433 4152 4192 6104 3194 2593 4694 4760 4582 4321
90 days 5784 3143 3683 3923 4422 4146 4183 6098 3181 2589 4681 4758 4570 4313

Table 13 records the strain reductions of W1–W6 due to steel relaxation of strands and creep of concrete in 90 days after anchoring. The calculated sum of the prestress losses, σl4 and σl5, is listed in Table 14 and illustrated in Figure 12.

Table 14.

The sum of σl4 and σl5 calculated from the strains obtained from the LFBG sensors. (Units: MPa).

Time R1 R2
W1 W2 W3 W4 W5 W6 Average W1 W2 W3 W4 W5 W6 Average
12 h 28.1 28.3 18.9 31.4 24.2 39.4 28.4 42.5 33.9 28.1 37.1 28.1 25.5 32.5
24 h 32.2 31.8 22.4 34.5 27.7 43.7 32.1 45.2 37.4 31.6 39.2 31.4 29.1 35.7
36 h 34.3 33.5 26.1 35.5 29.4 48.4 34.5 46.8 39.2 33.3 41.7 33.2 31.2 37.6
48 h 36.1 35.3 29.8 36.3 31.2 52.1 36.8 48.0 41.0 35.1 44.5 34.9 32.2 39.3
9 days 46.2 48.8 37.4 44.7 43.5 64.5 47.5 60.3 53.6 44.5 53.6 44.9 42.3 49.9
30 days 65.1 74.3 49.7 61.4 60.3 83.5 65.7 79.4 76.4 66.5 73.1 62.6 59.5 69.6
51 days 75.3 86.4 54 67.3 67.5 87.2 73.0 86.2 81.7 72.3 84.4 70.4 64.7 76.6
72 days 78.4 88.7 55.8 71.2 72.5 89.5 76.0 89.1 84.6 74.9 89.1 72.9 68.1 79.8
90 days 80.3 90.3 56.4 73.1 74.7 90.7 77.6 90.3 87.2 75.7 91.7 73.3 70.4 81.4

Figure 12.

Figure 12

The increment of the sum of σl4 and σl5 for 90 days: (a) R1; and (b) R2.

There are three features that may be observed in Table 14 and Figure 12. First, the phenomenon that all LFBG sensors can trace the stress variation proves that the LFBG has enough durability for long-term prestress loss monitoring, and the proposed installation method is valid. Second, the increment of the sum of σl4 and σl5 in each strand has a similar trend. It proves the stability of the LFBG in measurements. Finally, almost 50% losses are completed in 48 h and the monitored data become stable after the 51st day. This feature shows that the prestress loss monitoring should be applied at an early stage of prestressing and last until the loss data tends to stabilize.

6. Conclusions

We demonstrated the deployment of the LFBG sensor for prestress loss monitoring and Evaluation. An appropriate length design of LFBG sensor and the installation method were proposed. Then we showed the calculation methods for the itemized prestress losses in pre-tensioning and post-tensioning. The applicability of the LFBG for prestress loss monitoring was verified by experiments in the laboratory and in-site monitoring. From the results and discussions, the following conclusions can be drawn:

  • (1)

    An appropriate gauge length for LFBG sensor is at least 25 cm for prestress loss monitoring in the strand because the gauge can obtain the average strain by covering the six helical wires.

  • (2)

    Severe frictions between the strand and duct and the grout crack can bring accidental damage to LFBG sensor. The proposed installation method can prevent the LFBG sensor from these ruptures effectively occurring at not only tendon tensioning but also structure loading. The durability and stability of the LFBG sensor are proved to be better than those of traditional FSGs.

  • (3)

    The proposed calculation method acquired the itemized prestress losses at different stages of applying pretension accurately. Our results from the experiments including the cases of pre-tensioning and post-tensioning showed that the losses calculated from the measured strains of the LFBG sensors were more precise compared to those calculated from traditional FSGs. Moreover, from the in-site monitoring, we obtained the uneven stress distribution in different strands, measured the immediate losses at tensioning, and traced the time-dependent losses for 90 days. Thus, this calculation method can be easy to apply in the itemized prestress losses monitoring.

  • (4)

    Compared with the traditional electrical sensor, the LFBG sensor is proved to have better durability for long-term prestress loss monitoring in practice, especially in the case of grout cracking and aggressive environment.

Acknowledgments

The authors thank Zhi-ping Zhang and Yun-peng Zhang for their assistance with the experiments. The authors also express their gratitude for the anonymous reviewers for their constructive comments and suggestions.

Author Contributions

The theoretical derivations were proposed by S.S. and D.H. The installation method was developed by Y.W. The experiments and in-site monitoring were designed and carried out by Z.-H.W. and X.G. The paper was revised and finalized by S.-L.M.

Funding

This work is supported by the National Natural Science Foundation of China (No. 51808119), the Technique Project of Fuzhou (Grant No. 2017-G-74), The Education and Scientific Research Project for Young and Middle-aged Teachers of Fujian Province (Grant No. JAT160044), the Transportation and Communication Science and Technology Project of Fujian Province (Grant No. 201536), the Major Research Projects of the Hebei Provincial Education Department(Grant No. ZD2015060) and the Open Fund of Fujian Provincial Key Laboratory of Advanced Technology and Informatization in Civil Engineering(Research on monitoring technology for prestress loss of steel strand in duct of building based on long gage fiber Bragg grating sensor).

Conflicts of Interest

The authors declare no conflict of interest.

References

  • 1.Nawy E. Prestressed Concrete: A Fundamental Approach. 5th ed. Pearson Education Asia Ltd.; Singapore: Chongqing University Press; Chongqing, China: 2006. pp. 59–81. [Google Scholar]
  • 2.Smith J. Capacity of Prestressed Concrete Containment Vessels with Prestressing Loss. Sandia National Laboratories; Albuquerque, NM, USA: 2001. Sandia Report. [Google Scholar]
  • 3.Mo Y.L., Hwang W.L. The effect of prestress losses on the seismic response of prestressed concrete frames. Comput. Struct. 1996;59:1013–1020. doi: 10.1016/0045-7949(95)00348-7. [DOI] [Google Scholar]
  • 4.Woodward R.J., Williams F.W. Collapse of Ynys-y-Gwas Bridge, West Glamorgan. Proc. Inst. Civ. Eng. 1988;84:635–669. doi: 10.1680/iicep.1989.3660. [DOI] [Google Scholar]
  • 5.Sumitro S., Jarosevic A., Wang M.L. Elasto-magnetic sensor utilization on steel cable stress measurement; Proceedings of the 1st Fib Congress; Osaka, Japan. 13–19 October 2002; pp. 13–19. [Google Scholar]
  • 6.Cho K., Park S.Y., Cho J.R., Kim S.T., Park Y.H. Estimation of Prestress Force Distribution in the Multi-Strand System of Prestressed Concrete Structures. Sensors. 2015;15:14079–14092. doi: 10.3390/s150614079. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Cappello C., Zonta D., Laasri H.A., Glisic B., Wang M. Calibration of Elasto-Magnetic Sensors on In-Service Cable-Stayed Bridges for Stress Monitoring. Sensors. 2018;18:466. doi: 10.3390/s18020466. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Chen R.H.L., Wissawapaisal K. Application of Wigner-Ville Transform to Evaluate Tensile Forces in Seven-Wire Prestressing Strands. J. Eng. Mech. 2001;127:1206–1214. doi: 10.1061/(ASCE)0733-9399(2001)127:6(599). [DOI] [Google Scholar]
  • 9.Chen R.H.L., Wissawapaisal K. An ultrasonic method for measuring tensile forces in a seven-wire prestressing strand. AIP Conf. Proc. 2002;615:1295. doi: 10.1063/1.1472945. [DOI] [Google Scholar]
  • 10.Fang Z., Wang J. Vertical prestressing loss in the bos girder of long-span PC continuous bridges. China Civ. Eng. J. 2006;39:78–84. [Google Scholar]
  • 11.Xiao Y., Zhang Y., Lu J., Liu Y., Cheng W. Experimental analysis on pre-stress friction loss of crushed limestone sand concrete beams. Appl. Sci. 2018;8 doi: 10.3390/app8050683. [DOI] [Google Scholar]
  • 12.Kim J.T., Yun C.B., Ryu Y.S., Cho H.M. Identification of prestress-loss in PSC beams using modal information. Struct. Eng. Mech. 2004;17:467–482. doi: 10.12989/sem.2004.17.3_4.467. [DOI] [Google Scholar]
  • 13.Kim J.T., Park J.H., Hong D.S., Cho H.M., Na W.B. Vibration and impedance monitoring for prestress-loss prediction in PSC girder bridges. Smart Struct. Syst. 2009;5:81–94. doi: 10.12989/sss.2009.5.1.081. [DOI] [Google Scholar]
  • 14.Ho D.D., Kim J.T., Stubbs N., Park W.S. Prestress-force estimation in PSC girder using modal parameters and system identification. Adv. Struct. Eng. 2012;15:997–1012. doi: 10.1260/1369-4332.15.6.997. [DOI] [Google Scholar]
  • 15.Yang Y., Myers J.J. Prestress loss measurements in Missouri’s first fully instrumented HPC bridge; Proceedings of the 84th Annual Meeting of the Transportation Research Board; Washington, DC, USA. 11–13 January 2005; pp. 1–19. [Google Scholar]
  • 16.Guo X., Liu D., Huang P., Zheng X. Prestress loss of CFL in a prestressing process for strengthening RC beams. Int. J. Polym. Sci. 2017;2017:3832468. doi: 10.1155/2017/3832468. [DOI] [Google Scholar]
  • 17.Bartoli I., Salamone S., Phillips R., Scalea F.L., Sikorsky C.S. Use of interwire ultrasonic leakage to quantify loss of prestress in multiwire tendons. J. Eng. Mech. 2011;137:324–333. doi: 10.1061/(ASCE)EM.1943-7889.0000227. [DOI] [Google Scholar]
  • 18.Ye X.W., Su Y.H., Xi P.S. Statistical Analysis of Stress Signals from Bridge Monitoring by FBG System. Sensors. 2018;18:491. doi: 10.3390/s18020491. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Xiao F., Chen G.S., Hulsey J.L. Monitoring Bridge Dynamic Responses Using Fiber Bragg Grating Tiltmeters. Sensors. 2017;17:2390. doi: 10.3390/s17102390. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Hu D.T., Guo Y.X., Chen X.F., Zhang C.R. Cable Force Health Monitoring of Tongwamen Bridge Based on Fiber Bragg Grating. Appl. Sci. 2017;7:384. doi: 10.3390/app7040384. [DOI] [Google Scholar]
  • 21.Zhao X., Li W., Song G., Zhu Z., Du J. Scour monitoring system for subsea pipeline based on active thermometry: Numerical and experimental studies. Sensors. 2013;13:1490–1509. doi: 10.3390/s130201490. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Kong X., Ho S.C.M., Song G., Cai C.S. Scour Monitoring System Using Fiber Bragg Grating Sensors and Water-Swellable Polymers. ASCE J. Bridge Eng. 2017;22:04017029. doi: 10.1061/(ASCE)BE.1943-5592.0001062. [DOI] [Google Scholar]
  • 23.Li W., Ho S.C.M., Song G. Corrosion detection of steel reinforced concrete using combined carbon fiber and fiber Bragg grating active thermal probe. Smart Mater. Struct. 2016;25:045017. doi: 10.1088/0964-1726/25/4/045017. [DOI] [Google Scholar]
  • 24.Li W., Xu C., Ho S.C.M., Wang B., Song G. Monitoring concrete deterioration due to reinforcement corrosion by integrating acoustic emission and FBG strain measurements. Sensors. 2017;17:657. doi: 10.3390/s17030657. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Hou Q., Ren L., Jiao W., Zou P., Song G. An improved negative pressure wave method for natural gas pipeline leak location using FBG based strain sensor and wavelet transform. Math. Probl. Eng. 2013;2013:278794. doi: 10.1155/2013/278794. [DOI] [Google Scholar]
  • 26.Ho S.C.M., Ren L., Li H.N., Song G. A fiber Bragg grating sensor for detection of liquid water in concrete structures. Smart Mater. Struct. 2013;22:055012. doi: 10.1088/0964-1726/22/5/055012. [DOI] [Google Scholar]
  • 27.Hou Q., Jiao W., Ren L., Cao H., Song G. Experimental study of leakage detection of natural gas pipeline using FBG based strain sensor and least square support vector machine. J. Loss Prev. Process. Ind. 2014;32:144–151. doi: 10.1016/j.jlp.2014.08.003. [DOI] [Google Scholar]
  • 28.Feng Q., Kong Q., Huo L., Song G. Crack detection and leakage monitoring on reinforced concrete pipe. Smart Mater. Struct. 2015;24:115020. doi: 10.1088/0964-1726/24/11/115020. [DOI] [Google Scholar]
  • 29.Kim J.M., Kim H.W., Park Y.H., Yang I.H., Kim Y.S. FBG sensors encapsulated into 7-wire steel strand for tension monitoring of a prestressing tendon. Adv. Struct. Eng. 2012;15:907–917. doi: 10.1260/1369-4332.15.6.907. [DOI] [Google Scholar]
  • 30.Kim S.T., Park Y.H., Park S.Y., Cho K., Cho J.R. A Sensor-Type PC Strand with an Embedded FBG Sensor for Monitoring Prestress Forces. Sensors. 2015;15:1060–1070. doi: 10.3390/s150101060. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Cho K., Kim S.T., Cho J.R., Park Y.H. Estimation of tendon force distribution in prestressed concrete girders using smart strand. Appl. Sci. 2017;7:1319. doi: 10.3390/app7121319. [DOI] [Google Scholar]
  • 32.Shin K.J., Lee S.C., Kim Y.Y., Kim J.M., Park S. Construction condition and damage monitoring of post-Tensioned PSC girders using embedded sensors. Sensors. 2017;17:1843. doi: 10.3390/s17081843. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Kim J.M., Kim C.M., Choi S.Y., Lee B.Y. Enhanced strain measurement range of an FBG sensor embedded in seven-wire steel strands. Sensors. 2017;17:1654. doi: 10.3390/s17071654. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Zhou Z., He J., Chen G., Ou J. A smart steel strand for the evaluation of prestress loss distribution in post-tensioned concrete structures. J. Intell. Mater. Syst. Struct. 2009;20:1901–1912. doi: 10.1177/1045389X09347021. [DOI] [Google Scholar]
  • 35.Lan C., Zhou Z., Ou J. Full-scale prestress loss monitoring of damaged RC structures using distributed optical fiber sensing technology. Sensors. 2012;12:5380–5394. doi: 10.3390/s120505380. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Zhang H., Wu Z.S. Performance evaluation of BOTDR-based distributed fiber optic sensors for crack monitoring. Struct. Health Monit. 2008;7:143–156. doi: 10.1177/1475921708089745. [DOI] [Google Scholar]
  • 37.Zhang H., Wu Z.S. Performance Evaluation of PPP-BOTDA-Based Distributed Optical Fiber Sensors. Int. J. Distrib. Sens. Netw. 2012;2012:414692. doi: 10.1155/2012/414692. [DOI] [Google Scholar]
  • 38.Mckeeman I., Fusiek G., Perry M., Johnston M., Saafi M., Niewczas P., Walsh M., Khan S. First-time demonstration of measuring concrete prestress levels with metal packaged fibre optic sensors. Smart Mater. Struct. 2016;25:095051. doi: 10.1088/0964-1726/25/9/095051. [DOI] [Google Scholar]
  • 39.Perry M., Yan Z., Sun Z., Zhang L., Niewczas P., Johnston M. High Stress monitoring of prestressing tendons in nuclear concrete vessels using fiber optic sensors. Nucl. Eng. Des. 2014;268:35–40. doi: 10.1016/j.nucengdes.2013.12.038. [DOI] [Google Scholar]
  • 40.Majda P., Skrodzewicz J. A modified creep model of epoxy adhesive at ambient temperature. Int. J. Adhes. Adhes. 2009;29:396–404. doi: 10.1016/j.ijadhadh.2008.07.010. [DOI] [Google Scholar]
  • 41.Li E., Xi J., Chicharo J.F., Liu T., Li X., Jiang J., Li L., Wang Y., Zhang Y. SPIE Smart Structures, Devices, and Systems II. Volume 5649. SPIE; Bellingham, WA, USA: 2005. The Experimental evaluation of FBG sensor for strain measurement of prestressed steel strand; pp. 463–469. [Google Scholar]
  • 42.Li S., Wu Z. Development of distributed long-gage fiber optic sensing system for structural health monitoring. Struct. Health Monit. 2007;6:133–145. doi: 10.1177/1475921706072078. [DOI] [Google Scholar]
  • 43.Wu Z. Master’s Thesis. Southeast University; Nanjing, China: May 27, 2013. Durability Study of Fiber Bragg Grating Strain Sensor Packaged with Fiber in Structural Health Monitoring. [Google Scholar]
  • 44.Hong W., Wu Z.S., Yang C.Q., Wan C., Wu G. Investigation on the damage identification of bridges using distributed long-gauge dynamic macrostrain response under ambient excitation. J. Intell. Mater. Syst. Struct. 2012;23:85–103. doi: 10.1177/1045389X11430743. [DOI] [Google Scholar]
  • 45.Wang T., Tang Y.S. Dynamic displacement monitoring of flexural structures with distributed long-gage macro-strain sensors. Adv. Mech. Eng. 2017;9 doi: 10.1177/1687814017698885. [DOI] [Google Scholar]
  • 46.Tang Y.S., Ren Z.D. Dynamic Method of Neutral Axis Position Determination and Damage Identification with Distributed Long-Gauge FBG Sensors. Sensors. 2017;17:411. doi: 10.3390/s17020411. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Ministry of Housing and Urban-Rural Development of the People’s Republic of China . Code for Design of Prestressed Concrete Structures. China Architecture & Building Press; Beijing, China: 2016. [Google Scholar]

Articles from Sensors (Basel, Switzerland) are provided here courtesy of Multidisciplinary Digital Publishing Institute (MDPI)

RESOURCES