Abstract
Electroencephalography (EEG) and surface electromyography (sEMG) are notoriously cumbersome technologies. A typical setup may involve bulky electrodes, dangling wires, and a large amplifier unit. Adapting these technologies to numerous applications has been accordingly fairly limited. Thanks to the availability of printed electronics, it is now possible to effectively simplify these techniques. Elegant electrode arrays with unprecedented performances can be readily produced, eliminating the need to handle multiple electrodes and wires. Specifically, in this Perspective paper, we focus on the advantages of electrodes printed on soft films as manifested in signal transmission at the electrode-skin interface, electrode-skin stability, and user convenience during electrode placement while achieving prolonged use. Customizing electrode array designs and implementing blind source separation methods can also improve recording resolution, reduce variability between individuals and minimize signal cross-talk between nearby electrodes. Finally, we outline several important applications in the field of neuroscience and how each can benefit from the convergence of electrophysiology and printed electronics.
Keywords: EMG, EEG, printed electrodes, skin electronics, wearable sensors
Introduction
Surface Electrophysiology
In recent years, electroencephalography (EEG) and surface electromyography (sEMG) have been suggested for countless new applications, such as brain-machine interfaces (BMIs), neurological and psychiatric diagnostics, bio-feedback, rehabilitation, sports, and emotion detection (Griss et al., 2002; Chowdhury et al., 2013; Hwang et al., 2013; Khan and Hong, 2017). Efforts were directed toward improved signal processing methods (Hyvärinen, 2011; Farina and Holobar, 2016; Naik et al., 2016) low noise electronics and low-impedance electrodes (Sun et al., 2012; Yeo et al., 2013; Piervirgili et al., 2014; Wang et al., 2017). Until recently, electrode technology achieved relatively little progress. Most contemporary skin electrodes rely on conducting gels, a lingering bottleneck in skin electrophysiology. Gels and adhesive pastes may cause skin irritation and short circuits between adjacent electrodes.
Furthermore, signal quality gradually degrades over time as the gel dries (Searle and Kirkup, 2000; Kappenman and Luck, 2010). As an alternative, dry skin-penetration electrodes shaped as pins and barbs were suggested, utilizing micro electro mechanical systems (MEMS) techniques. Improving the electrode skin interface with applied pressure on the electrode was also explored (Searle and Kirkup, 2000; Griss et al., 2001; Liao et al., 2011; Lopez-Gordo et al., 2014). These technologies are nevertheless cumbersome and do not provide a true solution for prolonged use.
Overall, despite the clear motivation to use EEG and sEMG in many clinical and consumer settings, contemporary technology is too cumbersome for non-laboratory use. State of the art systems suffer from major technological challenges including: (a) Poor electrode quality manifested primarily by signal-to-noise ratio (SNR) reduction over time; (b) signal cross-talk between nearby electrodes resulting in limited source identification; and (c) low resolution recording (Searle and Kirkup, 2000; De Luca et al., 2010; Fiedler et al., 2014; Hug and Tucker, 2016).
Printed Electronics
Printed electronics is a key technology in a wide range of applications including displays and touch screens (Berggren et al., 2007). The growing need for flexible electronics has also contributed to vast expertise in printing materials that can accommodate bending and stretching. Printing of flexible electronics addresses the needs of huge markets, a fact which has driven its rapid development and maturation (Kamyshny and Magdassi, 2014; Sabatini et al., 2017). This resulted in the development of affordable and high performing inks, substrates, printing methods and tools, lamination processes, and more (Jiang et al., 2016). Substantial academic research is still carried out in this field, although well-established techniques along with vast expertise are already available in the industry for small to medium and large volume production.
Biocompatible materials, suitable for long-term use in direct contact with the skin, are also available commercially. Materials that can stretch to accommodate significant skin deformation during movement were developed (Jeong et al., 2013; Matsuhisa et al., 2017). Resolution achieved in printing techniques is limited compared with micro-fabrication, (Kim et al., 2011; Webb et al., 2013; Yeo et al., 2013; Madhvapathy et al., 2018) yet it is perfectly compatible with the needs of skin electrophysiology while benefiting from significantly reduced prototyping and production costs.
Electrophysiology Meets Printed Electronics
Printing electrode arrays on flexible substrates, such as polyimide or polyester, was suggested as a mean to improve user convenience and electrode placement (on hair-free regions). Printed electrodes using various conductive inks, such as thermoplastic silver ink (Myllymaa et al., 2013a), carbon (Bareket et al., 2016; Inzelberg et al., 2018a) and poly(3,4-ethylenedioxythiophene, PEDOT) (Inzelberg et al., 2018a) were applied recently in several applications (Figure 1). Printed EEG electrode arrays, such as BrainStatus (Lepola et al., 2014; Miettinen et al., 2018) and cEEGGrid (Debener et al., 2015) opened new opportunities in neurological investigations and diagnostics. Optimal material selection for printing the electrodes also permit safe artifact-free magnetic resonance imaging (MRI) and computerized tomography (CT) scanning (Myllymaa et al., 2013a,b). In the realm of sEMG, several studies demonstrated the benefits of printing (Scalisi et al., 2015; Zucca et al., 2015; Bareket et al., 2016; Ferrari et al., 2018; Inzelberg et al., 2018a). Ultrathin sEMG electrodes can maximize the contact area, lower the contact impedance, while reducing movement artifacts (Ferrari et al., 2018; Inzelberg et al., 2018a). High density printed sEMG emerges as a non-invasive method to acquire precise information of muscle activation by increasing the electrode number and enabling data analysis schemes (Marco, 2015; Scalisi et al., 2015).
Soft Printed Electrode Arrays and Improved Signal Quality
Ideally, electrophysiological electrodes should be capable of prolonged recording time, should be flexible and stretchable to conform with the human skin, should provide user convenience and achieve high-resolution recording with no cross-talk. The mechanical properties of the ink have to satisfy two main conditions: durability in bending and stretching, and high flexibility to achieve conformity with the skin. In our investigations we have addressed this issue by identifying inks that can sustain extreme folding without breaking. This proved to be particularly important for larger EEG electrodes. The electrical conductivity of the ink is important in EEG applications when long traces resistance may reach values comparable with the skin-electrode impedance. In this case, silver traces are needed. Detailed discussion regarding ink and material properties is available in Inzelberg et al., 2018a.
Early printed arrays, used to date, were realized on polyimide or polyester. Although flexible, they are not soft enough to conform and stretch with the human skin, and as such do not resolve the need for a conducting gel. However, printing the electrodes on soft substrates, such as polyurethane, facilitated a dramatic improvement (beyond user convenience achieved with the polyimide or polyester electrodes). Principally, a conducting gel is no longer needed to establish good electrode-skin conductivity. Secondly, recording quickly stabilizes and remains stable over time. Along with these two benefits, data analysis can be used to reduce cross-talk and inter- and intra-subject variability. Moreover, soft printed electrodes allow high-resolution and array customization.
Commercially available gelled electrodes are relatively rigid and thus can record effectively from relatively flat skin areas. Their recording performances from uneven or curved skin regions (such as the face) are limited. Moreover, as the skin-electrode impedance of a gelled electrode tends to deteriorate over time, the recording quality degrades over time (Searle and Kirkup, 2000). Dry electrodes, usually in the form of stiff metal pads, do not require gel. However, they lack flexibility resulting in poor adhesion to the skin, especially in curved areas (Searle and Kirkup, 2000; Fiedler et al., 2014). Electrode arrays printed on a highly soft support, on the other hand, facilitate skin-electrode conformity at uneven skin surfaces. As soft printed electrodes are dry, they enable stable, prolonged recordings. While commercial dry electrodes interfere with user comfort, electrodes printed on highly soft substrates allow superior user convenience with minimal attention to the experimental setup (Fiedler et al., 2014; Zucca et al., 2015; Ferrari et al., 2018; Inzelberg et al., 2018a,b). This property allows prolonged recording of several hours in sEMG (Bareket et al., 2016; Inzelberg et al., 2018b) and EEG (Figure 2B; Shustak et al., in press).
Cross-talk, defined as the contamination of the sEMG signal by the myoelectric activity in adjacent muscles, is a major challenge in skin electrophysiology (Chowdhury et al., 2013; Hug and Tucker, 2016). It may cause a change in the sEMG signal shape depending on electrode location, and in many cases, a muscle can be falsely identified as generating electrical activity when it does not (Hug and Tucker, 2016). Cross-talk depends strongly on anatomical and physiological parameters. Reducing the electrode size and precisely controlling the inter-electrode distance is a direct approach to reduce cross-talk (Chowdhury et al., 2013). Nevertheless, reduced electrode dimension is beneficial as long as electrodes can record distinguishable signals, which are not located in high-density muscle regions. A more relevant approach to address cross-talk in sEMG is by applying mathematical methodologies, such as blind source separation (BSS) (Hyvärinen et al., 2001; Farina et al., 2004; Naik et al., 2008, 2016; Farina and Holobar, 2016). As printed electrodes can be designed to include high-density arrays with multiple recording sites, they are ideal for source identification. This analysis cannot be achieved with contemporary electrodes due to their relative large size, limited density, and cumbersome placement. In facial expression mapping in particular, fast independent component analysis (fastICA) algorithm is a powerful tool (Inzelberg et al., 2018b). Once fastICA is applied on the data, the relative weight at each recording site is calculated and the independent components for each event are obtained. This methodology reduces variability between repetitions and individuals as it does not rely on electrode specific location (Hyvärinen, 2011; Inzelberg et al., 2018b).
In Figure 2 we show four examples of temporary-tattoo electrode arrays designed for four different uses: (1) Sub hairlines EEG; (2) EEG, Electrooculography (EOG) and sEMG array for sleep stage [stages 1–3 and Rapid Eye Movement (REM)] recording (six hr); (3) sEMG from the cheek and eyebrow regions to detect emotional affect; and (4) High resolution (16) electrode array for sEMG of the forearm.
Novel Applications
As summarized above, printed electrodes on soft substrates offer several important benefits over conventional electrode technology. Precise electrode placement is an important part in skin electrophysiology, commonly necessitating experienced technicians, and prolonging the placement process. Replacing such a manual procedure by a quick placement of a single adhesive patch has a clear advantage in reducing time and manpower costs. Printing enables customization of electrode arrays to ideally fit specific individuals, needs and applications. The large number of electrodes integrated in a single printed array provides high-resolution recordings with reduced motion artifacts, compared with conventional electrodes (Inzelberg et al., 2018b).
Reports from the last decades addressed numerous applications of skin electrophysiology. Most of these reports were based on gel electrodes. Clearly, the number of applications of wearable electrophysiological sensors is immense. We review below several examples of applications we currently study, emphasizing the benefits of printed electronic technology for neurological applications, while discussing existing knowledge gaps for each application.
Sleep Monitoring in Neurological Evaluation
Sleep disturbances appear in many neurological diseases and are one of the most common non-motor symptoms in Parkinson’s disease (PD) (Paparrigopoulos, 2005; Postuma, 2014). REM sleep behavior disorder (RBD) is a form of parasomnia characterized by the loss of normal skeletal muscle atonia during REM sleep, with prominent motor activity accompanying dreaming. In the search for markers of the disease, RBD holds much promise, as it is highly specific for phenoconversion to PD (Postuma et al., 2013). The gold standard for diagnosis of sleep disorders is based on studies conducted overnight in a laboratory setting using polysomnography (PSG). However, since RBD is episodic in nature, and does not necessarily appear every night, it is difficult to diagnose in a single night. In addition, this method is expensive, is not feasible for widespread clinical practice, and often does not reflect everyday life conditions. An electrode array with the capability to record EEG, EOG, and sEMG signals over multiple nights in the home environment holds great promise.
We recently designed, implemented and tested such a wireless dry electrode system for sleep stage analysis. sEMG, EOG, and EEG were successfully recorded using a wireless system. Stable recordings were achieved both in a hospital environment and in a home setting (Shustak et al., in press). Utilizing our wireless recording technology (Inzelberg et al., 2017, 2018b) with a printed EEG electrode array, enabled clear separation between open versus closed eyes (Figure 2A). Sleep monitoring during a six hr session showed clear differentiation of sleep stages (Figure 2B).
Facial Expressions as a Marker of Neuro-Psychiatric Conditions
Precise mapping (spatial and temporal) of facial expressions provides a vast promise in medical assessment because many neurological and psychiatric disorders are typified by abnormal activation patterns. One example is the clinical feature of PD, hypomimia; the reduction or loss of spontaneous facial expressions (Argaud et al., 2016; Bologna et al., 2016). On the other hand, an example demonstrating excessive facial muscle activation is Tourette syndrome (TS), that includes involuntary fast, brief and repetitive facial movements in the form of tics (Brandt et al., 2016; Muth, 2017). These examples point to the necessity to objectively classify facial expressions for diagnostic and therapeutic purposes.
We recently demonstrated a unique ability to differentiate between facial muscle activations using printed electrode arrays (Inzelberg et al., 2018b). Furthermore, the ability to record spontaneous, commanded and mimicry facial expressions, utilizing both screen- and inkjet-printed electrodes comprising both the cheek and the eyebrows regions in a single integrated array, was also implemented (Figure 2C; Inzelberg et al., 2018a).
Limb Muscle Activation in Neurology
Abnormal limb muscle activation reflects a multitude of pathological conditions that encompass diseases of the muscles, peripheral nerves and central nervous system, as well as orthopedics and trauma medicine. These pathologies are manifested by various clinical symptoms, including muscle weakness, change in muscle tone, co-activation patterns, as well as changes in the electromyography (EMG) signals. One such example is the change in activation of the first dorsal interosseous (FDI) muscle of the hand, resulting from lesions of the ulnar nerve (Kimura, 2013). In such cases, contemporary recordings rely on invasive needle EMG to reach single muscle specificity. Quantitative recordings from the FDI using printed electrode arrays can be readily achieved (Bareket et al., 2016; Inzelberg et al., 2018a). Another example is muscle atrophy, as in amyotrophic lateral sclerosis (ALS), where multiple muscles are recorded using needle EMG (Kimura, 2013). BSS is of importance in such an application as it can differentiate between muscle activations using a single integrated printed patch non-invasively (Figure 2D).
The Future: Objective Physiological Evaluation for Everyday Clinics
With the increase of life expectancy, neurological disorders are becoming one of the greatest societal and economical challenges of our era. Despite the immensity of the problem, clinical evaluation still relies heavily on subjective judgment of an experienced physician. Patients who lack access to a high quality specialist evaluation may be undiagnosed or misdiagnosed. In both cases, they miss an opportunity for early diagnosis that can significantly affect disease management. Physicians also rely on patient self-reporting to assess symptoms that may not be apparent during a visit to the clinic. Moreover, symptom severity is almost impossible to quantify, therefore the efficacy of drugs is difficult to assess. Finally, latent clinical problems, which may already indicate disease onset, are very difficult to be identified. Printed electrodes on soft substrates together with advanced analysis schemes of the acquired data provide a simple and cheap tool for objective mapping of neurophysiological abnormalities. Additional sensors such as temperature and skin conductivity may further enhance the performances of printed films. Improving both printed technology and data analysis methodology, such as BSS (Urigüen and Garcia-Zapirain, 2015; Negro et al., 2016; Inzelberg et al., 2018b), classification (Debener et al., 2015; Zhang et al., 2016) and machine learning (Wang et al., 2014; Amin et al., 2015; Gokgoz and Subasi, 2015) will impact diagnosis, evaluation of treatment efficacy and enhance research of neuro-psychiatric disorders. Ideally, such systems will enable automatic feedback and screening of normal versus pathological performances.
Summary
Printed electrodes can be used to perform high fidelity EEG and sEMG recordings and have countless applications in the medical and consumer markets. Further development may transform EEG and sEMG into a widely used diagnostic method in neurology, psychology and marketing. Further advancement of printed electronics may allow in the future the integration of additional capabilities beyond the sensing elements, such as amplifiers, temperature sensors, energy harvesting elements, chemical sensing and more.
Ethics Statement
All experiments on human skin were conducted on volunteers in accordance with relevant guidelines and regulations under approval from the Institutional Ethics Committee Review Board at Tel Aviv University. Written informed consent was obtained from all subjects.
Author Contributions
LI and YH wrote the manuscript. YH supervised the project.
Conflict of Interest Statement
The authors declare financial interest in a new company (under formation) which will hold the licensing rights of the temporary-tattoo technology described in this paper.
Acknowledgments
The authors thank many colleagues and collaborators for fruitful and inspiring discussions. In particular, they thank Anat Mirelman, Firas Fahoum, Maarten de Vos, Talma Hendler, Tom Schonberg, and Gerardo Hernandez-Sosa. Moshe David-Pur, Shiran Shustak, Stanislav Steinberg, Neomi Singer, Maayan Doron, and David Rand participated in designing and testing the printed electrodes for various applications.
Footnotes
Funding. This research was supported in part by an ERC (Funmania), Teva Pharmaceutical and an Israel Science Foundation (1355/17), Israeli Ministry of Economy (NanoEdge) and Zimin fund grants. LI was partially supported by the Sagol School of Neuroscience, Tel Aviv University, Israel.
References
- Amin H. U., Malik A. S., Ahmad R. F., Badruddin N., Kamel N., Hussain M., et al. (2015). Feature extraction and classification for EEG signals using wavelet transform and machine learning techniques. Australas. Phys. Eng. Sci. Med. 38 139–149. 10.1007/s13246-015-0333-x [DOI] [PubMed] [Google Scholar]
- Argaud S., Delplanque S., Houvenaghel J.-F., Auffret M., Duprez J., Vérin M., et al. (2016). Does facial amimia impact the recognition of facial emotions? An EMG study in Parkinson’s disease. PLoS One 11:e0160329. 10.1371/journal.pone.0160329 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Bareket L., Inzelberg L., Rand D., David-Pur M., Rabinovich D., Brandes B., et al. (2016). Temporary-tattoo for long-term high fidelity biopotential recordings. Sci. Rep. 6:25727. 10.1038/srep25727 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Berggren M., Nilsson D., Robinson N. D. (2007). Organic materials for printed electronics. Nat. Mater. 6 3–5. 10.1038/nmat1817 [DOI] [PubMed] [Google Scholar]
- Bologna M., Berardelli I., Paparella G., Marsili L., Ricciardi L., Fabbrini G., et al. (2016). Altered kinematics of facial emotion expression and emotion recognition deficits are unrelated in Parkinson’s disease. Front. Neurol. 7:230. 10.3389/fneur.2016.00230 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Brandt V. C., Patalay P., Bäumer T., Brass M., Münchau A. (2016). Tics as a model of over-learned behavior-imitation and inhibition of facial tics. Mov. Disord. 31 1155–1162. 10.1002/mds.26607 [DOI] [PubMed] [Google Scholar]
- Chowdhury R., Reaz M., Ali M., Bakar A., Chellappan K., Chang T. (2013). Surface electromyography signal processing and classification techniques. Sensors 13 12431–12466. 10.3390/s130912431 [DOI] [PMC free article] [PubMed] [Google Scholar]
- De Luca C. J., Donald Gilmore L., Kuznetsov M., Roy S. H. (2010). Filtering the surface EMG signal: movement artifact and baseline noise contamination. J. Biomech. 43 1573–1579. 10.1016/j.jbiomech.2010.01.027 [DOI] [PubMed] [Google Scholar]
- Debener S., Emkes R., De Vos M., Bleichner M. (2015). Unobtrusive ambulatory EEG using a smartphone and flexible printed electrodes around the ear. Sci. Rep. 5:16743. 10.1038/srep16743 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Farina D., Févotte C., Doncarli C., Merletti R. (2004). Blind separation of linear instantaneous mixtures of nonstationary surface myoelectric signals. IEEE Trans. Biomed. Eng. 51 1555–1567. 10.1109/TBME.2004.828048 [DOI] [PubMed] [Google Scholar]
- Farina D., Holobar A. (2016). Characterization of human motor units from surface EMG decomposition. Proc. IEEE 104 353–373. 10.1109/JPROC.2015.2498665 [DOI] [Google Scholar]
- Ferrari L. M., Sudha S., Tarantino S., Esposti R., Bolzoni F., Cavallari P., et al. (2018). Ultraconformable temporary tattoo electrodes for electrophysiology. Adv. Sci. 5:1700771. 10.1002/advs.201700771 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Fiedler P., Haueisen J., Jannek D., Griebel S., Zentner L., Vaz F., et al. (2014). Comparison of three types of dry electrodes for electroencephalography. Acta Imeko 3 33–37. 10.21014/acta [DOI] [Google Scholar]
- Gokgoz E., Subasi A. (2015). Comparison of decision tree algorithms for EMG signal classification using DWT. Biomed. Signal Process. Control 18 138–144. 10.1016/j.bspc.2014.12.005 [DOI] [Google Scholar]
- Griss P., Enoksson P., Tolvanen-Laakso H. K., Merilainen P., Ollmar S., Stemme G. (2001). Micromachined electrodes for biopotential measurements. J. Microelectromech. Syst. 10 10–16. 10.1109/84.911086 [DOI] [Google Scholar]
- Griss P., Tolvanen-Laakso H. K., Merilainen P., Stemme G. (2002). Characterization of micromachined spiked biopotential electrodes. IEEE Trans. Biomed. Eng. 49 597–604. 10.1109/TBME.2002.1001974 [DOI] [PubMed] [Google Scholar]
- Hug F., Tucker K. (2016). “Surface electromyography to study muscle coordination,” in Handbook of Human Motion, eds Müller B., Wolf S. I., Brueggemann G.-P., Deng Z., McIntosh A., Miller F., et al. (Cham: Springer International Publishing; ), 1–21. 10.1007/978-3-319-30808-1_184-1 [DOI] [Google Scholar]
- Hwang H.-J., Kim S., Choi S., Im C.-H. (2013). EEG-based brain-computer interfaces: a thorough literature survey. Int. J. Hum. Comput. Interact. 29 814–826. 10.1080/10447318.2013.780869 [DOI] [Google Scholar]
- Hyvärinen A. (2011). Testing the ICA mixing matrix based on inter-subject or inter-session consistency. Neuroimage 58 122–136. 10.1016/j.neuroimage.2011.05.086 [DOI] [PubMed] [Google Scholar]
- Hyvärinen A., Karhunen J., Erkki O. (2001). Independent Component Analysis. Hoboken, NJ: John Wiley & Sons; 10.1002/0471221317 [DOI] [Google Scholar]
- Inzelberg L., David Pur M., Steinberg S., Rand D., Farah M., Hanein Y. (2017). “Wireless electronic-tattoo for long-term high fidelity facial muscle recordings,” in Proceedings of the SPIE 10194. Micro-and Nanotechnology Sensors, Systems, and Applications IX, eds George T., Dutta A. K., Islam M. S. (Bellingham, WA: SPIE; ), 10.1117/12.2263522 [DOI] [Google Scholar]
- Inzelberg L., Pur M. D., Schlisske S., Rödlmeier T., Granoviter O., Rand D., et al. (2018a). Printed facial skin electrodes as sensors of emotional affect. Flex. Print. Electron. 3:045001 10.1088/2058-8585/aae252 [DOI] [Google Scholar]
- Inzelberg L., Rand D., Steinberg S., David-Pur M., Hanein Y. (2018b). A wearable high-resolution facial electromyography for long term recordings in freely behaving humans. Sci. Rep. 8:2058. 10.1038/s41598-018-20567-y [DOI] [PMC free article] [PubMed] [Google Scholar]
- Jeong J. W., Yeo W. H., Akhtar A., Norton J. J. S., Kwack Y. J., Li S., et al. (2013). Materials and optimized designs for human-machine interfaces via epidermal electronics. Adv. Mater. 25 6839–6846. 10.1002/adma.201301921 [DOI] [PubMed] [Google Scholar]
- Jiang J., Bao B., Li M., Sun J., Zhang C., Li Y., et al. (2016). Fabrication of transparent multilayer circuits by inkjet printing. Adv. Mater. 28 1420–1426. 10.1002/adma.201503682 [DOI] [PubMed] [Google Scholar]
- Kamyshny A., Magdassi S. (2014). Conductive nanomaterials for printed electronics. Small 10 3515–3535. 10.1002/smll.201303000 [DOI] [PubMed] [Google Scholar]
- Kappenman E. S., Luck S. J. (2010). The effects of electrode impedance on data quality and statistical significance in ERP recordings. Psychophysiology 47 888–904. 10.1111/j.1469-8986.2010.01009.x [DOI] [PMC free article] [PubMed] [Google Scholar]
- Khan M. J., Hong K.-S. (2017). Hybrid EEG–fNIRS-based eight-command decoding for BCI: application to quadcopter control. Front. Neurorobot. 11:6. 10.3389/fnbot.2017.00006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kim D.-H., Lu N., Ma R., Kim Y.-S., Kim R.-H., Wang S., et al. (2011). Epidermal electronics. Science 333 838–843. 10.1126/science.1206157 [DOI] [PubMed] [Google Scholar]
- Kimura J. (2013). Electrodiagnosis in Diseases of Nerve and Muscle. Oxford: Oxford University Press; 10.1093/med/9780199738687.001.0001 [DOI] [Google Scholar]
- Lepola P., Myllymaa S., Töyräs J., Muraja-Murro A., Mervaala E., Lappalainen R., et al. (2014). Screen-printed EEG electrode set for emergency use. Sens. Actuators A Phys. 213 19–26. 10.1016/j.sna.2014.03.029 [DOI] [Google Scholar]
- Liao L. D., Wang I. J., Chen S. F., Chang J. Y., Lin C. T. (2011). Design, fabrication and experimental validation of a novel dry-contact sensor for measuring electroencephalography signals without skin preparation. Sensors 11 5819–5834. 10.3390/s110605819 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Lopez-Gordo M., Sanchez-Morillo D., Valle F. (2014). Dry EEG electrodes. Sensors 14 12847–12870. 10.3390/s140712847 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Madhvapathy S. R., Ma Y., Patel M., Krishnan S., Wei C., Li Y., et al. (2018). Epidermal electronic systems for measuring the thermal properties of human skin at depths of up to several millimeters. Adv. Funct. Mater. 28:1802083 10.1002/adfm.201802083 [DOI] [Google Scholar]
- Marco G. (2015). “Low cost inkjet printing for the fast prototyping of surface EMG detection systems,” in Proceedings of the IEEE International Symposium on Medical Measurements and Applications (MeMeA), (Rome: IEEE; ), 79–83. 10.1109/MeMeA.2015.7145176 [DOI] [Google Scholar]
- Matsuhisa N., Inoue D., Zalar P., Jin H., Matsuba Y., Itoh A., et al. (2017). Printable elastic conductors by in situ formation of silver nanoparticles from silver flakes. Nat. Mater. 16 834–840. 10.1038/nmat4904 [DOI] [PubMed] [Google Scholar]
- Miettinen T., Myllymaa K., Muraja-Murro A., Westeren-Punnonen S., Hukkanen T., Töyräs J., et al. (2018). Screen-printed ambulatory electrode set enables accurate diagnostics of sleep bruxism. J. Sleep Res. 27 103–112. 10.1111/jsr.12536 [DOI] [PubMed] [Google Scholar]
- Muth C. C. (2017). Tics and tourette syndrome. JAMA 317:1592. 10.1001/jama.2017.0547 [DOI] [PubMed] [Google Scholar]
- Myllymaa S., Lepola P., Hukkanen T., Oun A., Mervaala E., Toyras J., et al. (2013a). “Novel screen printed electrode set for routine EEG recordings in patients with altered mental status,” in Proceedings of the 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), (Osaka: IEEE; ), 6724–6727. 10.1109/EMBC.2013.6611099 [DOI] [PubMed] [Google Scholar]
- Myllymaa S., Lepola P., Töyräs J., Hukkanen T., Mervaala E., Lappalainen R., et al. (2013b). New disposable forehead electrode set with excellent signal quality and imaging compatibility. J. Neurosci. Methods 215 103–109. 10.1016/j.jneumeth.2013.02.003 [DOI] [PubMed] [Google Scholar]
- Naik G. R., Kumar D. K., Palaniswami M. (2008). “Addressing source separation and identification issues in surface EMG using blind source separation,” in Proceedings of the 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, (Vancouver: IEEE; ), 1124–1127. 10.1109/IEMBS.2008.4649358 [DOI] [PubMed] [Google Scholar]
- Naik G. R., Suviseshamuthu S. E., Gobbo M., Acharyya A., Nguyen H. T. (2016). Principal component analysis applied to surface electromyography: a comprehensive review. IEEE Access. 4 1–17. 10.1109/ACCESS.2016.2593013 [DOI] [Google Scholar]
- Negro F., Muceli S., Castronovo A. M., Holobar A., Farina D. (2016). Multi-channel intramuscular and surface EMG decomposition by convolutive blind source separation. J. Neural Eng. 13:026027. 10.1088/1741-2560/13/2/026027 [DOI] [PubMed] [Google Scholar]
- Paparrigopoulos T. J. (2005). REM sleep behaviour disorder: clinical profiles and pathophysiology. Int. Rev. Psychiatry 17 293–300. 10.1080/09540260500104540 [DOI] [PubMed] [Google Scholar]
- Piervirgili G., Petracca F., Merletti R. (2014). A new method to assess skin treatments for lowering the impedance and noise of individual gelled Ag–AgCl electrodes. Physiol. Meas. 35 2101–2118. 10.1088/0967-3334/35/10/2101 [DOI] [PubMed] [Google Scholar]
- Postuma R. B. (2014). Prodromal Parkinson’s disease - Using REM sleep behavior disorder as a window. Park. Relat. Disord. 20 S1–S4. 10.1016/S1353-8020(13)00400-8 [DOI] [PubMed] [Google Scholar]
- Postuma R. B., Gagnon J. F., Montplaisir J. (2013). Rapid eye movement sleep behavior disorder as a biomarker for neurodegeneration: the past 10 years. Sleep Med. 14 763–767. 10.1016/j.sleep.2012.09.001 [DOI] [PubMed] [Google Scholar]
- Sabatini V., Farina H., Ortenzi M. A. (2017). Conductive inks based on methacrylate end-capped poly(3,4-ethylenedioxythiophene) for printed and flexible electronics. Polym. Eng. Sci. 57 491–501. 10.1002/pen.24502 [DOI] [Google Scholar]
- Scalisi R. G., Paleari M., Favetto A., Stoppa M., Ariano P., Pandolfi P., et al. (2015). Inkjet printed flexible electrodes for surface electromyography. Org. Electron. Phys. Mater. Appl. 18 89–94. 10.1016/j.orgel.2014.12.017 [DOI] [Google Scholar]
- Searle A., Kirkup L. (2000). A direct comparison of wet, dry and insulating bioelectric recording electrodes. Physiol. Meas. 21 271–283. 10.1088/0967-3334/21/2/307 [DOI] [PubMed] [Google Scholar]
- Shustak S., Inzelberg L., Steinberg S., Rand D., David Pur M., Hillel I., et al. (in press). Home monitoring of sleep with a temporary-tattoo EEG, EOG and EMG electrode array. J. Neural Eng. [DOI] [PubMed] [Google Scholar]
- Sun M., Jia W., Liang W., Sclabassi R. J. (2012). “A low-impedance, skin-grabbing, and gel-free EEG electrode,” in Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, (Honolulu, HI: IEEE; ), 1992–1995. 10.1109/EMBC.2012.6346347 [DOI] [PubMed] [Google Scholar]
- Urigüen J. A., Garcia-Zapirain B. (2015). EEG artifact removal - State-of-the-art and guidelines. J. Neural Eng. 12:031001. 10.1088/1741-2560/12/3/031001 [DOI] [PubMed] [Google Scholar]
- Wang K., Parekh U., Pailla T., Garudadri H., Gilja V., Ng T. N. (2017). Stretchable dry electrodes with concentric ring geometry for enhancing spatial resolution in electrophysiology. Adv. Healthc. Mater. 6:1700552. 10.1002/adhm.201700552 [DOI] [PubMed] [Google Scholar]
- Wang X.-W., Nie D., Lu B.-L. (2014). Emotional state classification from EEG data using machine learning approach. Neurocomputing 129 94–106. 10.1016/j.neucom.2013.06.046 [DOI] [Google Scholar]
- Webb R. C., Bonifas A. P., Behnaz A., Zhang Y., Yu K. J., Cheng H., et al. (2013). Ultrathin conformal devices for precise and continuous thermal characterization of human skin. Nat. Mater. 12 938–944. 10.1038/nmat3755 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yeo W.-H., Kim Y.-S., Lee J., Ameen A., Shi L., Li M., et al. (2013). Multifunctional epidermal electronics printed directly onto the skin. Adv. Mater. 25 2773–2778. 10.1002/adma.201204426 [DOI] [PubMed] [Google Scholar]
- Zhang Y., Zhou G., Jin J., Zhao Q., Wang X., Cichocki A. (2016). Sparse bayesian classification of EEG for brain–computer interface. IEEE Trans. Neural Networks Learn. Syst. 27 2256–2267. 10.1109/TNNLS.2015.2476656 [DOI] [PubMed] [Google Scholar]
- Zucca A., Cipriani C., Sudha, Tarantino S., Ricci D., Mattoli V., et al. (2015). Tattoo conductive polymer nanosheets for skin-contact applications. Adv. Healthc. Mater. 4 983–990. 10.1002/adhm.201400761 [DOI] [PubMed] [Google Scholar]