A Song of Pressure and Flow, or There and Back Again*

James P Howard a, Venkatesh L Murthy b
PMCID: PMC6329005  NIHMSID: NIHMS1003204  PMID: 29673506

It has been more than 2 decades since fractional flow reserve (FFR) was introduced, and it now plays a central role in guidelines for stable coronary artery disease (1,2). FFR is calculated as the ratio of the pressure distal to a stenosis to the pressure proximal to a stenosis during hyperemia. However, FFR’s adoption has been slow, with international uptake estimated at <10% (3). Reasons for this may include the costs and practicalities of administering adenosine, patient discomfort during hyperemia, and safety concerns in certain patient subgroups (4).

In this issue of JACC: Cardiovascular Interventions, Mejía-Rentería et al. (5) investigate the diagnostic performance of the “quantitative flow ratio” (QFR), a physiological assessment that, it is hoped, will provide a faster, safer, and cheaper surrogate measure of FFR that does not require passing of a coronary pressure wire or pharmacological hyperemia. The diagnostic performance of QFR has been investigated in small studies previously (6,7). However, because it relies on angiographic flow measurements, Mejía-Rentería et al. wished to investigate whether coronary microcirculatory parameters could affect its accuracy. The investigators now show that QFR’s diagnostic performance, when judged against FFR, does degrade in patients with coronary microcirculatory dysfunction, as high microvascular resistance causes the positive predictive value to decrease from 93% to 67%.

The method of QFR calculation used by the investigators requires 2 pieces of information. First, a 3-dimensional model of the stenosis is constructed from 2 angiographic views. Second, the flow velocity in the vessel is measured using simple “frame counting” of contrast. Using these data, the pressure drops over the length of the vessel can be simulated and the QFR calculated.

QFR therefore indirectly estimates coronary pressure ratios by measuring flow, whereas FFR uses direct measurements of coronary pressure. Given this, “fractional flow reserve” may seem a surprising name for a ratio of pressures, and indeed this gives interesting insights into the history of coronary physiology.

CORONARY FLOW MEASUREMENTS AND THE BIRTH OF CORONARY PHYSIOLOGY

Myocardial ischemia in patients with stable angina develops when myocardial oxygen demand exceeds what the coronary circulation can supply. This supply can be thought of as the volume of blood delivered to a tissue per second (cm3/s) and can be quantified simply by multiplying the blood flow velocity through the vessel (cm/s) by the vessel’s cross-sectional area (cm2).

By the early 1990s, technology had evolved sufficiently to allow velocity sensors to be mounted on angioplasty guidewires (8). By measuring coronary flow velocity at rest, and also during hyperemia, physicians could now estimate a vessel’s ability to meet the increased flow requirements during exercise. The ratio of flow velocity during hyperemia to that at rest was termed the coronary flow (velocity) reserve (CFR). Observational data showed that using a CFR cutoff of 2.0 provided superior risk stratification in patients with intermediate-severity coronary lesions than single-photon emission computer tomography (9).

THE ERA OF PRESSURE-DERIVED CORONARY PHYSIOLOGY

However, these flow-measuring guidewires could also measure pressure. Initially it was unclear how the pressure throughout the cardiac cycle could be interpreted. As in an electric circuit, pressure (or voltage) = flow (or current) × resistance, so pressure across the cardiac cycle could only become an accurate surrogate for flow if resistance across the coronary circulation were constant. Pijls et al. (10) proposed, however, that during hyperemia, resistance is low and varies minimally. They therefore published a paper covering the experimental basis for a pressure-derived estimate of flow reserve, FFR. More recently, investigators have hypothesized performing pressure measurements during the “wavefree” period of diastole, when resistance is naturally minimal and constant, providing the instantaneous wave-Free Ratio (iFR).

With time, these more practical pressure-based measurements have gained popularity. Not only are the pressure-sensing guidewires cheaper and easier to use than the flow-measuring models, but studies showed that data can be useful for treatment decision making at the stenosis level. The loss in pressure across a stenosis is related to the flowlimiting potential of the lesion (10). This was in contrast to CFR, for which an abnormally low value could represent limitations to flow anywhere along the vessel length, including the distal microcirculation.

The data supporting FFR’s use in intermediate coronary stenoses followed over the next 2 decades. The DEFER (11), FAME (12), and FAME-2 (13) studies went on to show the safety of deferring percutaneous coronary intervention in patients with FFRs >0.75 and reduced composite endpoints (largely urgent revascularization) from intervening on lesions with FFRs < 0.80.

As with many other indexes, patients requiring clinical assessment often have values close to the clinical cutoffs of CFR and FFR, and in these patients, recent data show that CFR and FFR discordance occurs in about one-third of patients (14). Interestingly, in these discordant cases, it appears that it is CFR that better categorizes the high-risk patients rather than FFR. This begs the question: is it flow we should be trying to measure in these patients with intermediate stenoses, rather than pressure? Or even both? One might argue that the strength of pressure-derived measurements lies in their ability to identify focal pressure drops over lesions and guide revascularization strategies, whereas CFR seems to provide more accurate information on disease severity and prognostication. Perhaps by combining these 2 measures, patients can be managed to the best of our current evidence base.

A FLOW RENAISSANCE?

At first the results of this new study might appear disappointing. In patients without hemodynamically significant lesions (as judged by FFR) but with high microvascular resistance, QFR gives “false positives.” However, is this a fair assessment? Possibly not, and there are 2 reasons we might be optimistic.

First, one could argue that QFR’s utility could lie in identifying patients at low risk who do not need to progress to direct FFR measurement with a pressure wire. If so, this study’s results are encouraging, with the negative predictive value remaining as high at 87% even in patients with coronary microcirculatory dysfunction.

Second, we must remember that FFR is estimating flow only by measuring pressure, despite its name. Ironically, QFR is now trying to estimate pressure gradients by measuring flow. This may strike the reader as peculiar when one remembers the prognostic importance of CFR in CFR/FFR discordance (14). One therefore has to wonder: in patients with high microvascular resistance and lower coronary blood flow, maybe these “false positives” for QFR are not so false after all?

Conversely, we don’t currently know the significance of the rarer “false negative” QFR readings (i.e., patients with FFRs <0.80 but normal QFRs). Hopefully future studies investigating the relationship among QFR, FFR, and CFR will demonstrate whether such cases are truly failures of QFR to identify flow-limiting lesions or instead cases in which CFR is preserved and QFR identifies patients who are at lower risk (14).

We commend the investigators for this study; any new measure should first be validated against the “gold standard” by which patients are currently treated, and FFR, and more recently iFR, have been extensively validated (3). However, the investigators also acknowledge a limitation in having not stratified their results by CFR values. Such an analysis might indeed show superiority for QFR over FFR. Indeed, if QFR can accurately measure flow, it will not merely be a less invasive surrogate for FFR but rather synergistic with it, providing us with a fast, practical, safe, and reliable measure of CFR.

Acknowledgments

Dr. Howard is supported by The Wellcome Trust. Dr. Murthy is supported by grant R01HL136685 from the National, Heart, Lung, and Blood Institute; owns stock in GE, Johnsons & Johnson, and Medtronic; and receives grant support from Siemens Medical Imaging.

Footnotes

*

Editorials published in JACC: Cardiovascular Interventions reflect the views of the authors and do not necessarily represent the views of JACC: Cardiovascular Interventions or the American College of Cardiology.

REFERENCES

  • 1.Authors/Task Force Members, Windecker S, Kolh P, et al. 2014 ESC/EACTS guidelines on myocardial revascularization. Eur Heart J 2014;35: 2541–619. [DOI] [PubMed] [Google Scholar]
  • 2.Fihn SD, Gardin JM, Abrams J, et al. 2012 ACCF/AHA/ACP/AATS/PCNA/SCAI/STS guideline for the diagnosis and management of patients with stable ischemic heart disease: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines, and the American College of Physicians, American Association for Thoracic Surgery, Preventive Cardiovascular Nurses Association, Society for Cardiovascular Angiography and Interventions, and Society of Thoracic Surgeons. J Am Coll Cardiol 2012;60:e44–164. [DOI] [PubMed] [Google Scholar]
  • 3.Götberg M, Cook CM, Sen S, Nijjer S, Escaned J, Davies JE. The evolving future of instantaneous wave-free ratio and fractional flow reserve. J Am Coll Cardiol 2017;70:1379–402. [DOI] [PubMed] [Google Scholar]
  • 4.Dattilo PB, Prasad A, Honeycutt E, Wang TY, Messenger JC. Contemporary patterns of fractional flow reserve and intravascular ultrasound use among patients undergoing percutaneous coronary intervention in the United States: insights from the National Cardiovascular Data Registry. J Am Coll Cardiol 2012;60:2337–9. [DOI] [PubMed] [Google Scholar]
  • 5.Mejía-Rentería H, Lee JM, Lauri F, et al. Influence of microcirculatory dysfunction on angiographybased functional assessment of coronary stenoses. J Am Coll Cardiol Intv 2018;11:741–53. [DOI] [PubMed] [Google Scholar]
  • 6.Tu S, Westra J,Yang, et al. Diagnostic accuracy of fast computational approaches to derive fractional flow reserve from diagnostic coronary angiography: the international multicenter FAVOR pilot study. J Am Coll Cardiol Intv 2016;9:2024–35. [DOI] [PubMed] [Google Scholar]
  • 7.Xu B, Tu S, Qiao S, et al. Diagnostic accuracy of angiography-based quantitative flow ratio measurements for online assessment of coronary stenosis. J Am Coll Cardiol 2017;70:3077–87. [DOI] [PubMed] [Google Scholar]
  • 8.Serruys PW, Di Mario C, Meneveau N, et al. Intracoronary pressure and flow velocity with sensor-tip guidewires: a new methodologic approach for assessment of coronary hemodynamics before and after coronary interventions. Am J Cardiol 1993;71:D41–53. [DOI] [PubMed] [Google Scholar]
  • 9.Chamuleau SAJ, Tio RA, de Cock CC, et al. Prognostic value of coronary blood flow velocity and myocardial perfusion in intermediate coronary narrowings and multivessel disease. J Am Coll Cardiol 2002;39:852–8. [DOI] [PubMed] [Google Scholar]
  • 10.Pijls NH, van Son JA, Kirkeeide RL, De Bruyne B, Gould KL. Experimental basis of determining maximum coronary, myocardial, and collateral blood flow by pressure measurements for assessing functional stenosis severity before and after percutaneous transluminal coronary angioplasty. Circulation 1993;87:1354–67. [DOI] [PubMed] [Google Scholar]
  • 11.Pijls NHJ, van Schaardenburgh P, Manoharan G, et al. Percutaneous coronary intervention of functionally nonsignificant stenosis. J Am Coll Cardiol 2007;49:2105–11. [DOI] [PubMed] [Google Scholar]
  • 12.Tonino PAL, De Bruyne B, Pijls NHJ, et al. Fractional flow reserve versus angiography for guiding percutaneous coronary intervention. N Engl J Med 2009;360:213–24. [DOI] [PubMed] [Google Scholar]
  • 13.De Bruyne B, Pijls NHJ, Kalesan B, et al. Fractional flow reserve–guided PCI versus medical therapy in stable coronary disease. N Engl J Med 2012;367:991–1001. [DOI] [PubMed] [Google Scholar]
  • 14.van de Hoef TP, van Lavieren MA, Damman P, et al. Physiological basis and long-term clinical outcome of discordance between fractional flow reserve and coronary flow velocity reserve in coronary stenoses of intermediate severity. Circ Cardiovasc Interv 2014;7:301–11. [DOI] [PubMed] [Google Scholar]

RESOURCES