Abstract
Purpose
The efficacy of microfocused ultrasound with visualization (MFU-V; Ultherapy®) has been demonstrated in clinical studies and daily practice. However, data addressing skin physiology after MFU-V treatment are lacking. This observational evaluation was aimed to assess skin physiology before and after MFU-V treatment using noninvasive biophysical measurements.
Patients and methods
Twenty-two female patients with moderate-to-severe skin sagging at the jawline and submental region on the Merz Aesthetics Scale obtained a single MFU-V treatment according to protocol. Skin function measurements focused on short-term effects up to 3 days and long-term effects up to 24 weeks after treatment. Skin temperature, transepidermal water loss, skin hydration, erythema, elasticity, and skin thickness and density were evaluated under standardized conditions. Pain was assessed using a validated numeric visual analog scale.
Results
Skin temperature remained in a physiologic range and no significant increase was noted at day 3 after MFU-V treatment. Transepidermal water loss, hydration, and erythema values were fairly stable and showed no significant differences at short- and long-term measurements vs baseline. At week 4 after a single MFU-V treatment, gross and net elasticity values were significantly decreased (P=0.003 and P=0.0001, respectively), followed by significantly increased values at week 12 (P=0.015, P=0.046) and week 24 (P=0.001, P=0.049). Edema due to MFU-V treatment resolved without sequelae. For all patients, pain diminished shortly after treatment. No adverse events occurred during the 24-week follow-up period.
Conclusions
MFU-V treatment is well tolerated and it does not alter the epidermal barrier function or physiology of skin. Significant increase in the elasticity of skin was observed at 12 and 24 weeks after a single treatment, which reflects improvement in dermal tissue function. These short- and long-term effects are congruous with the mode of action of MFU-V due to a proven intrinsic tissue remodeling process.
Keywords: skin lifting, tightening, neocollagenesis, epidermal barrier, elasticity, energy-based device
Introduction
Patients with skin laxity of the face and neck often seek skin rejuvenation with noninvasive, safe, effective, and rapid cosmetic treatments. Microfocused ultrasound with visualization (MFU-V; Ulthera® System, Merz North America, Raleigh, NC, USA) is characterized by precise delivery of ultrasound energy at predefined depths with simultaneous visualization.1,2 This technology enables the induction of precise thermal coagulation points of 65°C without damaging skin surface.3 It also targets the superficial muscular aponeurotic system (SMAS) to produce focused thermal collagen denaturation and subsequent neocollagenesis.4
The MFU-V medical device is approved by the US Food and Drug Administration. According to the CE-mark, MFU-V is indicated for noninvasive lifting and sculpting of the upper face, lower face, neck, and décolleté.5 Clinical studies have demonstrated that MFU-V is effective as a noninvasive lifting and sculpting device of the upper face,6 lower face,7 neck,8,9 and décolleté.10
Patients expect noninvasive cosmetic treatments, especially those intended for facial rejuvenation, to be effective and safe with minimal recovery time. Reviews from clinical studies and post-marketing use of MFU-V suggest that energy-based devices can fulfill these desired parameters.11,12
The primary objective of this evaluation was to assess skin physiology before and both short and long term after MFU-V treatment in patients with moderate to severe skin laxity at jawline and submental region.
Patients and methods
Design
This observational, single-center, open-label safety evaluation was performed using anonymized data received through objective standardized measurements for skin physiology in routine practice. All participants were provided with detailed written and oral information and written informed consent was obtained before any treatments and assessments as suggested by local ethics committee (Ethikkommission der Ärztekammer Hamburg). The evaluation was performed in accordance with the principles of the 1975 Declaration of Helsinki. MFU-V was performed on-label for noninvasive lifting and sculpting of the lower face and submental region.
Microfocused ultrasound with visualization
MFU-V combines precise delivery of microfocused ultrasound to dermal/subdermal tissues, including the SMAS, with simultaneous visualization of the treatment area. Before delivery of microfocused ultrasound, an appropriate image ensures heat delivery to the intended area. MFU-V is available commercially and consists of a control unit with touch screen, a hand piece, and proprietary transducers (DeepSEE®, Ulthera/Merz, Mesa, AZ, USA).
Cohort
Patients
Twenty-two healthy women with moderate-to-severe skin laxity have been treated with MFU-V due to their clinical needs and following available treatment guidelines. No contraindication existed in any of the patients (Table 1). Subjects’ characteristics were documented before MFU-V treatment was performed. Soft-tissue laxity was assessed clinically using the five-point Merz Aesthetics Scale (MAS) for the lower face.13 Objective measurements of skin functions were taken under standardized conditions before, immediately after, and 12 and 24 weeks after MFU-V treatment.
Table 1.
Inclusion | Exclusion |
---|---|
| |
Female Age 30–65 years MAS grade 2–3 for lower face/submental skin laxity (moderate to severe) Naïve to minimally invasive or surgical cosmetic treatment Willing and able to provide informed consent Willing to attend follow-up visits |
Pregnant or lactating Open wounds or lesions in the treatment area; severe or cystic acne in the treatment area Excessive skin laxity on the lower face and neck Presence of metallic stent or active implants in the treatment area Mental illness Inability to understand the protocol or to give written consent History of minimally invasive aesthetic procedures (eg, dermal fillers, laser therapy, implants, dermabrasion, and deep facial peel) in the treatment area within the last 24 months |
Abbreviation: MFU-V, microfocused ultrasound with visualization.
Procedure
At baseline, ibuprofen (800 mg orally, once) was administered at the discretion of the treating physician and participant 1 hour before treatment. A single MFU-V treatment was performed on the lower face, submental region, and neck according to the MFU-V treatment protocol consecutively with two transducers: one at 4.0 MHz (depth of 4.5 mm, 350 lines in total) followed by one at 7.0 MHz (depth of 3.0 mm, 270 lines in total; Figure 1). Each enrolled subject completed 12-week or 24-week follow-up visits or both.
Biophysical evaluation
A number of biophysical parameters were measured to assess the skin physiology prior to the MFU-V treatment, immediately after treatment, and during each post-treatment follow-up visit at day 3, week 4, week 12, and week 24 (Figure 2). All measurements were obtained under standardized dermatological laboratory condition with an acclimatization period of at least 30 minutes in an air-conditioned room at 20°C–21°C and relative humidity of 40%–50%. Parameters assessed were in vivo transepidermal water loss (TEWL), skin hydration, erythema, viscoelasticity properties, and skin thickness and density. Measurements were taken from the same site on both cheeks using a template.
Transepidermal water loss
TEWL was assessed using a Tewametry (Tewameter® TM 300; Courage & Khazaka, Cologne, Germany). In brief, a hollow cylindrical chamber containing two hygro and temperature sensors situated at different levels above the skin surface measures relative humidity. The vapor pressure difference between the two sensor sites determines the TEWL and is given in grams per hour per square meter (g/hour/m2).14 One continuous measurement of 30 seconds was recorded at each visit at predefined localizations within the treated areas.15
Stratum corneum hydration
Stratum corneum hydration was determined using Corneometry (Corneometer® CM 825; Courage & Khazaka, Cologne, Germany). Briefly, the measurement of the electrical capacity of the stratum corneum directly correlates to hydration. The measured capacity is given in arbitrary units (CM units), ranging from 0 to 120, specifying very dry skin to <30 CM, dry skin from 30 to 40 CM and normal skin >40 CM.16 Three measurements were taken at each visit at predefined skin locations within the treated areas.
Skin redness (erythema)
Erythema was measured using mexametry (Mexameter® MX 18; Courage & Khazaka, Cologne, Germany). The Mexameter consists of 16 circularly arranged light-emitting diodes that emit green, red, and near-infrared light at three specific light wavelengths of 568, 660, and 880 nm. The Erythema index is computed based on the intensity of the absorbed and reflected green and red light for hemoglobin at wavelengths of 568 and 880 nm, respectively.
Skin elasticity
Elasticity of the skin was assessed using cutometry (Cutometer® MPA 580; Courage & Khazaka, Cologne, Germany). A defined vacuum (–450 mbar) was applied to the respective skin surface area and turned off to let the skin recover. A probe with 2 mm diameters was used. The skin properties were calculated from measurement of vertical deformation of the skin as well as timing using the software of the instrument. The parameters of interest were gross (R2) and net elasticity (R5).
Skin thickness and density
Both parameters were determined using an ultrasound scanner with a frequency of 20 MHz (DUB® 20; Taberna Pro Medicum, Lueneburg, Germany). Sonographic assessment of skin thickness was based on the intensity of reflected ultrasound waves in the respective skin area. Skin thickness was calculated based on values from amplitudes of reflected waves and depths of the skin. The B-Scan mode was used to determine skin density.
Skin temperature
Skin temperature was measured before, 30 minutes and 3 days after the treatment using a thermistor (113050, Rochester, Inc. Rochester, NY, USA). Temperature is measured by placing the half inch diameter surface probe of the thermistor on the skin until equilibrium is reached (defined by a <0.2°C change in temperature of >15 seconds).
Safety
Tolerability and safety assessments were performed immediately after the MFU-V treatment and on follow-up visits. Patients’ subjective pain assessment using a visual analog scale (VAS; 11 points: 0= no pain, 10= worst pain imaginable) was performed immediately after treatment. At each follow-up visit, all patients were examined for edema and adverse events.
Statistical analysis
Statistical analyses were carried out with SPSS® v 22.0 (IBM, Armonk, NY, USA). Kolmogorov–Smirnov test was used to assess normal distribution of the data. A parametric t-test for paired samples was applied on normally distributed data, and the nonparametric Wilcoxon signed-rank test was used when the data did not appear normally distributed. P-values of ≤0.05 were considered as statistically significant, P-values of ≤0.01 were considered as highly significant and very highly significant when the P-value was P≤0.0001.
Results
Demographics
Twenty-two Caucasian women (mean age, 52.32±9.32 years; Fitzpatrick skin type, I–III; body mass index, 20–30 kg/m2) with moderate-to-severe skin laxity (mean MAS score, 2.8±0.87) were treated with MFU-V and eligible for evaluation.
Safety
No adverse events occurred during the course of the evaluation, and no patients withdrew from the evaluation due to an adverse event. Dermal edema in the treatment area was observed 3 days after MFU-V but resolved completely by week 4. Sonography illustrated a representative example of the course of edema (Figure 3).
Self-assessment of pain during MFU-V treatment of the lower face and submental region revealed minimal-to-moderate pain (VAS score, 1–5) for 15 cases, with 7 patients reporting minimal pain and 6 patients reporting moderate pain. For all patients, pain diminished shortly after treatment.
Skin temperature
Mean skin temperature was significantly increased immediately after treatment (P<0.05) from 31.536°C ± 1.369°C at baseline to 32.323°C ± 1.315°C. Measurements at day 3 after MFU-V treatment showed decreased mean values of 31.468°C ± 1.08°C (Figure 4).
Transepidermal water loss
Evaporimetry demonstrated physiologic values before, immediately after, and 3 days after treatment (Figure 5), as well as at 4 and 12 weeks after MFU-V treatment for long-term assessment (Figure 6). At baseline, mean values were 4.555±1.651 g/hour/m2. They were increased slightly immediately after MFU-V to 5.295±2.145 g/hour/m2 and declined to 4.755±2.175g/hour/m2 (day 3), 3.685±1.525 g/hour/m2 (4 weeks) and 4.702±1.517 g/hour/m2 (12 weeks). There were no statistically significant increases for TEWL after MFU-V treatment for either short- or long-term follow-up.
Hydration
Skin hydration (shown in Corneometer Units) showed a significant decrease over 12 weeks, but values remained fairly stable within the physiologic range before (54.31±12.69), 4 weeks (52.64±8.47), and 12 weeks (48.43±9.94) after MFU-V treatment (Figure 7).
Erythema
The mean erythema level (shown in Mexameter Units) at baseline was 313.70±76.11, and was increased slightly immediately after treatment to 326.45±72.84 and decreased at day 3 after treatment to 321.36±70.19 (Figure 8). Erythema values were further decreased at week 4 (293.63±69.51 MU) and at week 12 (290.31±62.34 MU; Figure 9). Erythema values did not demonstrate significant changes compared to baseline for short-term or long-term evaluation.
Elasticity
A single MFU-V treatment targeting two preselected depths of 4.5 and 3.0 mm led to a significant increase in net skin elasticity at 12 and 24 weeks, respectively, compared with that at baseline (P<0.05). After 4 weeks, the net elasticity was significantly lower than that at baseline, suggesting physiologic restructuring of collagen tissue. Skin gross elasticity also showed decreased values at week 4, but significantly increased values at weeks 12 and 24 (Figure 10).
Discussion
The effectiveness of MFU-V treatment has been demonstrated in several clinical studies6–10 and its safety profile has been assessed.12 However, prior to this evaluation, no structured data addressed the influence of MFU-V treatment on epidermal/dermal physiologic and biomechanical parameters.
Furthermore, maintenance of skin integrity after treatments for the lifting and tightening of skin is challenging, because damage to the superficial skin layers is unwanted, whereas neocollagenesis in deeper layers is desired. This clinical assessment focused on evaluating short- and long-term physiologic effects of a single MFU-V treatment on skin of the lower face and submental region.
In the present evaluation we provided, for the first time, data for biophysical skin assessments scheduled around MFU-V treatment. We found that after MFU-V treatment, the essential protective skin functions remained in the physiologic range and parameters representing skin tightening were improved. Skin temperature increased within a physiologic range immediately after MFU-V treatment, but there was no significant increase at day 3 after treatment. This observation suggests that MFU-V treatment results in heat being introduced precisely at small thermal coagulation points at intended predefined depths of 4.5 and 3.0 mm.
The epidermal barrier was not disturbed after MFU-V treatment as TEWL, skin temperature, skin hydration, and erythema values did not change significantly either within 3 days of MFU-V treatment or within a long-term follow-up at 24 weeks. These results suggest that this MFU-V technology delivers microfocused ultrasound in a transcutaneous manner without damaging the skin surface.
Net and gross elasticity, as measured by the firmness and involution of the skin, were significantly improved at weeks 12 and 24. These findings are consistent with clinical studies demonstrating that MFU-V skin-tightening effects most often occur 6 months post treatment, which coincides with neocollagenesis and collagen conversion.17 These data also demonstrate that MFU-V can be a promising and suitable treatment not only for patients with slight loss of elasticity and skin contour of the face, but also for patients with moderate to severe loss of facial skin elasticity.
Dermal edema was assessed as a parameter of safety in MFU-V treatment. The course of edema resolution might be used as an indicator for wound healing in subsequent combined minimal invasive or surgical cosmetic treatments. The data of this structured assessment suggest that dermal edema might be present 3 days after treatment, but diminished completely by 4 weeks after treatment. Based on these results, waiting for at least 4 weeks before performing any subsequent skin tightening treatments may be prudent. According to consensus recommendations for combined aesthetic treatments, MFU-V might be an effective and safe initial treatment in a combined aesthetic plan.18
Conclusion
MFU-V treatment of the lower face and submental region did not alter skin barrier and skin physiology. However, for the first time, we have demonstrated significant improvement of viscoelastic properties of the skin 12 and 24 weeks post treatment, thereby likely reflecting physiologic dermatologic basis for lifting and tightening of treated areas.
Footnotes
Author contributions
MK conducted the evaluation performing visits and instrumental evaluations. ATN and CEN participated in data analyses. SH, JLB, and MK participated in writing, critically revising review, and approving the manuscript. All authors contributed toward data analysis, drafting and revising the paper, gave approval of the final version to be published and agree to be accountable for all aspects of the work.
Disclosure
MK has received research support and has conducted clinical trials for Merz Pharmaceuticals GmbH and has acted as a speaker and/or investigator for Merz, Kythera, Q-Med/Gal-derma, and Pierre Fabre. SH and JLB are employees of Merz Pharmaceuticals. ATN and CEN are scientific employees at the University of Hamburg. The authors report no other conflicts of interest in this work.
References
- 1.Laubach HJ, Makin IR, Barthe PG, Slayton MH, Manstein D. Intense focused ultrasound: evaluation of a new treatment modality for precise microcoagulation within the skin. Dermatol Surg. 2008;34(5):727–734. doi: 10.1111/j.1524-4725.2008.34196.x. [DOI] [PubMed] [Google Scholar]
- 2.Minkis K, Alam M. Ultrasound skin tightening. Dermatol Clin. 2014;32(1):71–77. doi: 10.1016/j.det.2013.09.001. [DOI] [PubMed] [Google Scholar]
- 3.White WM, Makin IR, Slayton MH, Barthe PG, Gliklich R. Selective transcutaneous delivery of energy to porcine soft tissues using Intense Ultrasound (IUS) Lasers Surg Med. 2008;40(2):67–75. doi: 10.1002/lsm.20613. [DOI] [PubMed] [Google Scholar]
- 4.White WM, Makin IR, Barthe PG, Slayton MH, Gliklich RE. Selective creation of thermal injury zones in the superficial musculoaponeurotic system using intense ultrasound therapy: a new target for noninvasive facial rejuvenation. Arch Facial Plast Surg. 2007;9(1):22–29. doi: 10.1001/archfaci.9.1.22. [DOI] [PubMed] [Google Scholar]
- 5.Ulthera® System [Instruction for use] Merz North America; Raleigh, NC: 2018. [Google Scholar]
- 6.Chan NP, Shek SY, Yu CS, Ho SG, Yeung CK, Chan HH. Safety study of transcutaneous focused ultrasound for non-invasive skin tightening in Asians. Lasers Surg Med. 2011;43(5):366–375. doi: 10.1002/lsm.21070. [DOI] [PubMed] [Google Scholar]
- 7.Oni G, Hoxworth R, Teotia S, Brown S, Kenkel JM. Evaluation of a microfocused ultrasound system for improving skin laxity and tightening in the lower face. Aesthet Surg J. 2014;34(7):1099–1110. doi: 10.1177/1090820X14541956. [DOI] [PubMed] [Google Scholar]
- 8.Alam M, White LE, Martin N, Witherspoon J, Yoo S, West DP. Ultrasound tightening of facial and neck skin: a rater-blinded prospective cohort study. J Am Acad Dermatol. 2010;62(2):262–269. doi: 10.1016/j.jaad.2009.06.039. [DOI] [PubMed] [Google Scholar]
- 9.Werschler WP, Werschler PS. Long-term efficacy of micro-focused ultrasound with visualization for lifting and tightening lax facial and neck skin using a customized vectoring treatment method. J Clin Aesthet Dermatol. 2016;9(2):27–33. [PMC free article] [PubMed] [Google Scholar]
- 10.Fabi SG, Goldman MP, Dayan SH, Gold MH, Kilmer SL, Hornfeldt CS. A prospective multicenter pilot study of the safety and efficacy of microfocused ultrasound with visualization for improving lines and wrinkles of the décolleté. Dermatol Surg. 2015;41(3):327–335. doi: 10.1097/DSS.0000000000000322. [DOI] [PubMed] [Google Scholar]
- 11.Day D. Microfocused ultrasound for facial rejuvenation: current perspectives. Res Rep Focus Ultrasound. 2014;2:13–17. [Google Scholar]
- 12.Hitchcock TM, Dobke MK. Review of the safety profile for microfocused ultrasound with visualization. J Cosmet Dermatol. 2014;13(4):329–335. doi: 10.1111/jocd.12111. [DOI] [PubMed] [Google Scholar]
- 13.Narins RS, Carruthers J, Flynn TC, et al. Validated assessment scales for the lower face. Dermatol Surg. 2012;38(2 Spec No):333–342. doi: 10.1111/j.1524-4725.2011.02247.x. [DOI] [PubMed] [Google Scholar]
- 14.Humbert P, Fanian F, Maibach HI, Agache A. Agache’s Measuring the Skin. 2nd ed. New York: Springer; 2017. [Google Scholar]
- 15.du Plessis J, Stefaniak A, Eloff F, et al. International guidelines for the in vivo assessment of skin properties in non-clinical settings: Part 2. transepidermal water loss and skin hydration. Skin Res Technol. 2013;19(3):265–278. doi: 10.1111/srt.12037. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Heinrich U, Koop U, Leneveu-Duchemin MC, et al. Multicentre comparison of skin hydration in terms of physical-, physiological- and product-dependent parameters by the capacitive method (Corneometer CM 825) Int J Cosmet Sci. 2003;25(1–2):45–53. doi: 10.1046/j.1467-2494.2003.00172.x. [DOI] [PubMed] [Google Scholar]
- 17.Fabi SG, Goldman MP. Retrospective evaluation of micro-focused ultrasound for lifting and tightening the face and neck. Dermatol Surg. 2014;40(5):569–575. doi: 10.1111/dsu.12471. [DOI] [PubMed] [Google Scholar]
- 18.Carruthers J, Burgess C, Day D, et al. Consensus recommendations for combined aesthetic interventions in the face using botulinum toxin, fillers, and energy-based devices. Dermatol Surg. 2016;42(5):586–597. doi: 10.1097/DSS.0000000000000754. [DOI] [PubMed] [Google Scholar]