Abstract
To evaluate the effectiveness of intrastromal corneal ring segment (ICRS) use in keratoconus after five years. ICRS has been widely used to correct astigmatism and improve visual acuity in keratoconus. Although the short-term outcome is well known to be effective, long-term outcome has rarely been reported. A retrospective chart review was done. A total of 30 eyes diagnosed with keratoconus and treated with INTACS (Addition Technology, Sunnyvale, CA, USA) were included. Visual acuity, refraction, indices of corneal irregularity, and higher-order aberration were evaluated at preoperative, two months, one year, three years, and five years postoperatively. Uncorrected distance visual acuity (UDVA) and spherical and spherical equivalent were improved (p < 0.05) for three years. However, they worsened (p < 0.05) at five years to preoperative values. On the other hand, corrected distance visual acuity (CDVA) was improved for five years (p < 0.05). Topographic keratometry was flattened, and corneal irregularity indices were improved at five years (all p < 0.05). Coma RMS was improved (p < 0.05) continuously for five years. ICRS has advantages in improving CDVA with topographic stabilization and decreasing coma in keratoconus for five years.
Introduction
Keratoconus is a progressive, bilateral, and non-inflammatory disease. Corneal irregular astigmatism and decrease of vision are induced by corneal thinning and protrusion1,2. Various treatment options are available. Spectacles or rigid gas-permeable contact lenses can be tried to improve visual acuity. In patients with contact-lens intolerance, intrastromal corneal ring segments (ICRSs) can be used to reduce corneal astigmatism and myopia by flattening the cornea. These days, corneal collagen crosslinking is widely performed to improve astigmatism and halt progression. For very severe cases, keratoplasty can be one option3–7.
ICRSs were first introduced to correct myopia in 19918,9. However, now they are more widely used for keratoconus and ectatic diseases than for myopia10. Corneal flattening is induced by arc shortening. Corneal segments act as a spacer in lamella. They can thicken the peripheral cornea and flatten the central cornea. This effect is known to be directly proportional to corneal thickness and inversely proportional to corneal diameter11,12.
Three types of ICRS are available: INTACS (Addition Technology, Sunnyvale, CA, USA), Kerarings (Mediphacos, Belo Horizonte, Brazil), and Ferrara rings (Mediphacos, Belo Horizonte, Brazil). INTACS is subdivided into two models: INTACS and INTACS SK. INTACS is 6.77 mm in inner diameter, with a hexagonal cross section. INTACS SK is 6.0 mm in inner diameter. It is more effective than INTACS in severe ectatic cases13. It has a round cross section to minimize glare13.
Several studies have reported that INTACS is effective in keratoconus14–20. Uncorrected distance visual acuity and spherical equivalent can be improved after INTACS implantation14–20. However, these reports mainly focused on short-term (less than one year) outcomes. Long-term outcomes of INTACS have been reported in some studies, but there is no consensus yet. Some studies have reported that topographic and refractive outcomes are improved in the long term21,22, but another study has reported regression in the long term following short-term improvement6.
INTACS can influence higher-order aberrations (HOA) and the quality of vision. However, only a few studies have reported short-term changes of HOA23,24. Long-term changes of HOA after INTACS have not been reported yet, to be best of our knowledge. Therefore, the objective of this study was to evaluate refractive and topographic parameters and HOA changes after INTACS implantation with a five-year follow up.
Patients and Methods
Inclusion and exclusion criteria
This study was performed by retrospective chart review. Eyes diagnosed as keratoconus that had received INTACS were included. The surgery was performed from January 2010 to December 2012 in Seoul St. Mary’s Hospital. Keratoconus stages 1 to 3, based on the Amsler-Krumeich classification, were included. Patients younger than 35 years at the time of surgery were included. These patients who were followed up for more than five years were included. They were evaluated on the preoperative day and at two months, one year, three years, and five years postoperatively. Eyes with a history of ocular surgery or active ocular disease other than keratoconus were excluded. Any complications, such as ring extrusion, perforation into anterior chamber, and infiltration around the ring, were also excluded. This study was approved by the Institutional Review Board (IRB) of Seoul St. Mary’s Hospital and followed the tenets of the Declaration of Helsinki. Informed consent was obtained from all participants.
Surgical procedure
For all cases, two INTACS (6.0 mm in inner diameter) were inserted symmetrically. Implantation axis and depth were chosen according to the nomogram defined by the manufacturer using ORB scan II (Bausch & Lomb, Rochester, NY, USA).
All surgeries were performed by a single surgeon (C.K.J.). The corneal tunnel was made with a femtosecond laser (Intralase, Irvine, CA, USA) under topical anesthesia. After the tunnel was made, INTACS were inserted, and the corneal incision was sutured with 10-0 nylon. After the surgery, topical antimicrobial and topical steroid four times a day each were used for one month. The corneal suture was removed at one month postoperatively.
Visual and refractive outcome
Uncorrected distance visual acuity (UDVA), corrected distance visual acuity (CDVA), and manifest refraction (sphere and cylinder) were measured. We evaluated mean anterior simulated keratometry (anterior SimK), cylinder of anterior simulated keratometry (anterior SimKcyl), mean anterior keratometry in a 3.0-mm zone (anterior 3 mm K), cylinder of anterior keratometry in a 3.0-mm zone (anterior 3 mm Kcyl), mean posterior simulated keratometry (posterior SimK), and cylinder of posterior simulated keratometry (posterior SimKcyl) using Pentacam (Oculus, Inc., Wetzlar, Germany).
Corneal thickness
Corneal thickness was measured using Pentacam, to identify the thinnest corneal thickness (TCT), central corneal thickness (CCT), and mean peripheral corneal thickness (PCT) for the superior, nasal, inferior, and temporal corneal thickness of the 6.0-mm area.
Index of corneal irregularity
Indices of corneal irregularity provided by Pentacam1 were measured to evaluate changes of corneal irregularity. Index of surface variance (ISV, the irregularity of curvature of the anterior corneal surface), index of vertical asymmetry (IVA, asymmetry of curvature between the superior and inferior cornea), keratoconus index (KI), and minimum radius of curvature (Rmin, the smallest radius of curvature of the cornea) were evaluated.
Higher-order aberration
Corneal wavefront aberrations at 6.0 mm pupil were evaluated with Pentacam. Spherical aberration (4, 0), root mean square (RMS) value of coma (3, ±1), trefoil (3, ±3), and total corneal higher-order aberrations (HOA) (3rd to 6th order HOA) were recorded.
Statistical analysis
Statistical analysis for normality was done using a Kolmogorov-Smirnov test. A paired t-test was performed for parametric data, and a Wilcoxon signed-rank test was performed for non-parametric data to compare the preoperative and postoperative values. Repeated-measures analysis of variance (ANOVA) was used to evaluate three or more timepoints. All statistical analyses were performed using SPSS software version 22.0 (IBM Corp., Armonk, NY, USA) and p < 0.05 was considered to be statistically significant.
Results
Demographics
There were 87 eyes that received INTACS from January 2010 to December 2012. Among them, nine eyes were older than 35 years, 18 had keratoconus stage 4, and four were complicated cases (one ring extrusion, two infiltrations around the ring, and one corneal opacity); these were excluded. Another 26 eyes not followed up at at least one timepoint were also excluded. A total of 30 eyes (16 right and 14 left) of 23 consecutive patients (seven bilateral and 16 unilateral; 14 men and nine women) were included. Their mean age was 27.69 years ± 6.68 (SD) (range, 16 to 35 years). Six (20%), 13 (43.33%), and 11 (36.66%) eyes were at stages 1, 2, and 3, respectively, based on the Amsler-Krumeich classification at the time of INTACS implantation.
Visual outcome
UDVA was improved significantly at one year (p = 0.000). Such improvement (p = 0.000) was maintained up to three years postoperatively. However, UDVA became worse at five years postoperative, although not worse than the preoperative UDVA (p = 0.074) (Fig. 1A and Table 1). CDVA was improved at one year. Such improvement was maintained for five years (p = 0.008, p = 0.002, and p = 0.011 at one year, three years, and five years, respectively) (Fig. 1A and Table 1).
Table 1.
Parameter | Preoperative | Postoperative | P value | ||||||
---|---|---|---|---|---|---|---|---|---|
2 m | 1 yr | 3 yr | 5 yr | Pre to 2m | Pre to 1yr | Pre to 3 yr | Pre to five yr | ||
UDVA (log MAR) |
0.857 ±0.456 |
0.757 ±0.391 |
0.662 ±0.404 |
0.636 ±0.387 |
0.741 ±0.372 |
0.224 | 0.000* | 0.000* | 0.074 |
CDVA (log MAR) |
0.516 ±0.298 |
0.507 ±0.270 |
0.391 ±0.245 |
0.377 ±0.276 |
0.404 ±0.299 |
0.224 | 0.008* | 0.002* | 0.011* |
Sphere | −5.521 ±4.791 |
−3.638 ±4.661 |
−3.707 ±4.401 |
−4.021 ±3.996 |
−4.984 ±4.008 |
0.000* | 0.005* | 0.042* | 1.000 |
Spherical equivalent | −7.816 ±4.946 |
−5.664 ±4.670 |
−5.571 ±4.424 |
−5.952 ±3.973 |
−6.752 ±4.058 |
0.000* | 0.000* | 0.001* | 0.563 |
UDVA = Uncorrected distance visual acuity. CDVA = Corrected distance visual acuity. * indicates statistical significance.
Refractive outcome
Both sphere and spherical equivalent were improved at two months. Such improvement was maintained up to three years (p = 0.000, p = 0.005, and p = 0.042 at two months, one year, and three years, respectively) (Fig. 1B and Table 1). However, myopia worsened at five years to about its preoperative value (p = 1.000).
Anterior SimK was flattened at two months, and this effect was maintained for five years (p = 0.001, p = 0.000, p = 0.001, and p = 0.000 at two months, one year, three years, and five years, respectively) (Fig. 1C and Table 2). Anterior 3-mm K in Pentacam was also flattened at two months and maintained for five years (p = 0.001, p = 0.000, p = 0.005, and p = 0.001 at two months, one year, three years, and five years, respectively) (Fig. 1C and Table 2). Posterior SimK was temporarily steepened at two months. It recovered to the preoperative level at one year (p = 0.007 and p = 0.139 at two months and one year, respectively) (Fig. 1D). Types of cylinder (manifest refraction cylinder, anterior SimKcyl, anterior 3 mm Kcyl, or posterior SimKcyl) were not changed. They remained constant for five years (Fig. 1E).
Table 2.
Parameter | Preoperative | Postoperative | P value | ||||||
---|---|---|---|---|---|---|---|---|---|
2 m | 1 yr | 3 yr | 5 yr | Pre to 2 m | Pre to 1 yr | Pre to 3 yr | Pre to five yr | ||
Anterior SimK | 50.667 ±2.581 |
47.921 ±2.497 |
47.858 ±2.657 |
47.600 ±2.183 |
47.733 ±2.434 |
0.001* | 0.000* | 0.001* | 0.000* |
Anterior 3 mm K | 50.614 ±3.203 |
48.728 ±4.410 |
48.307 ±2.948 |
48.629 ±2.792 |
48.500 ±2.715 |
0.001* | 0.000* | 0.005* | 0.001* |
Posterior SimK | −8.047 ±1.176 |
−8.297 ±1.095 |
−8.242 ±1.045 |
−8.075 ±0.952 |
−8.100 ±1.104 |
0.007* | 0.139 | 1.000 | 1.000 |
Anterior SimK = Mean anterior simulated keratometry; Anterior 3 mm K = Mean anterior keratometry in 3.0 mm zone; Posterior SimK = Mean posterior simulated keratometry. * indicates statistical significance.
Corneal thickness
Thinnest corneal thickness (TCT) or central corneal thickness (CCT) was not changed for five years (Fig. 1F). Peripheral corneal thickness (PCT) was increased at two months but slightly decreased at three years (p = 0.014). However, PCT was thicker than its preoperative value for five years (p = 0.000 at two months, one year, three years, and five years) (Fig. 1F).
Index of corneal irregularity
ISV showed a decreasing tendency, reaching statistical significance at five years (p = 0.030) (Table 3). IVA decreased at three years and five years (p = 0.013 and p = 0.010, respectively) (Table 3). KI also decreased at three years. This effect was maintained for five years (p = 0.039 and p = 0.027 at three years and five years, respectively) (Table 3). Rmin was increased at one year and maintained for five years (p = 0.001, p = 0.001, and p = 0.003 at one year, three years, and five years, respectively) (Table 3).
Table 3.
Parameter | Preoperative | Postoperative | P value | ||||||
---|---|---|---|---|---|---|---|---|---|
2 m | 1 yr | 3 yr | 5 yr | Pre to 2m | Pre to 1yr | Pre to 3 yr | Pre to five yr | ||
ISV | 100.43 ±34.798 |
104.17 ±32.417 |
98.36 ±32.993 |
91.36 ±28.397 |
89.64 ±28.659 |
0.366 | 1.000 | 0.150 | 0.030* |
IVA | 0.951 ±0.515 |
1.060 ±0.516 |
1.020 ±0.491 |
0.888 ±0.451 |
0.876 ±0.452 |
0.169 | 0.744 | 0.013* | 0.010* |
KI | 1.259 ±0.155 |
1.265 ±0.143 |
1.255 ±0.158 |
1.218 ±0.145 |
1.215 ±0.140 |
1.000 | 1.000 | 0.039* | 0.027* |
Rmin | 5.637 ±0.512 |
5.764 ±0.606 |
5.888 ±0.569 |
5.971 ±0.480 |
5.957 ±0.506 |
0.233 | 0.001* | 0.001* | 0.003* |
ISV = Index of surface variance; IVA = Index of vertical asymmetry; KI = Keratoconus index; Rmin = Minimum radius of curvature. * indicates statistical significance.
Higher order aberration
Spherical aberration was negatively increased at two months. However, it recovered to the preoperative level at one year (p = 0.000 and p = 1.000 at two months and one year, respectively) (Fig. 2). Coma RMS continuously decreased from two months to five years (p = 0.000, p = 0.007, p = 0.002, and p = 0.005 at two months, one year, three years, and five years, respectively) (Fig. 2 and Table 4). Otherwise, trefoil RMS or HOA RMS did not change for five years.
Table 4.
Parameter | Preoperative | Postoperative | P value | ||||||
---|---|---|---|---|---|---|---|---|---|
2 m | 1 yr | 3 yr | 5 yr | Pre to 2m | Pre to 1yr | Pre to 3 yr | Pre to five yr | ||
Spherical aberration | −1.461 ±1.708 |
−1.743 ±0.737 |
−1.425 ±0.944 |
−1.440 ±0.771 |
−1.388 ±1.081 |
0.000* | 1.000 | 1.000 | 1.000 |
Trefoil RMS | 0.416 ±0.235 |
0.661 ±0.457 |
0.656 ±0.421 |
0.845 ±0.425 |
0.753 ±0.548 |
0.267 | 1.000 | 0.726 | 0.141 |
Coma RMS | 2.484 ±1.548 |
2.161 ±1.269 |
1.937 ±1.150 |
1.821 ±1.821 |
1.642 ±1.163 |
0.000* | 0.007* | 0.002* | 0.005* |
HOA RMS | 14.541 ±1.664 |
15.344 ±1.275 |
14.365 ±1.112 |
13.474 ±1.052 |
12.448 ±1.044 |
0.357 | 0.268 | 0.248 | 0.071 |
RMS = Root mean square; HOA = Higher-order aberration. * indicates statistical significance.
Discussion
Long-term follow-up study of INTACS is rare. Change over time has rarely been reported. Also, published studies have analyzed only refractive and topographic findings but not higher-order aberrations6,21,22, which can influence visual quality after corneal surgery. Its importance has been emphasized these days. Therefore, we looked for changes in refractive, topographic, and aberrational outcomes over time.
UDVA was improved at one year and maintained constant for three years. However, it decreased at five years to preoperative level (Fig. 1A and Table 1), perhaps because of refraction change. After the surgery, sphere and spherical equivalent were improved at two months and maintained for three years. However, myopia was increased at five years close to its preoperative value (Fig. 1B). Increase of myopia at five years can decrease UDVA.
Unlike UDVA, CDVA was improved at one year and maintained for five years (Fig. 1A), perhaps because of topographic regularization. Anterior 3 mm K measured at the corneal center 3.0 mm using Pentacam was flattened at two months and maintained for five years (Fig. 1C). Anterior SimK was also flattened at two months and kept constant for five years (Fig. 1C). Indices of corneal irregularity were also improved. ISV, the index of irregularity of anterior surface curvature, gets higher when the cornea is becoming irregular. In our study, ISV gradually decreased, reaching statistical significance at five years (Table 3). High IVA indicates high vertical asymmetry of curvature. This index also decreased from three to five years (Table 3). KI, representing the severity of keratoconus, also decreased from three to five years (Table 3). Rmin, the inverse of corneal steepness, was expected to decrease in keratoconus. However, it was significantly increased from one year to five years (Table 3).
Cylinder has been reported to be decreased after INTACS implantation in many other studies6,22,23,25. Only one study has reported that there is no difference from the preoperative cylinder value16. Interestingly, no types of cylinders were changed throughout five years in our study (Fig. 1E). Although the cornea was significantly flattened after INTACS implantation (Fig. 1C), the cylinder was not affected at all. Thus, although INTACS can flatten the cornea, it might not have enough power to decrease the cylinder itself.
PCT was increased at two months but slightly decreased at three years, although it was still thicker than it was preoperatively (680.21 µm and 661.69 µm at one year and three years, respectively, p = 0.014) (Fig. 1F). Ezra et al.26 have reported abnormal accumulation of fibrotic extracellular matrix components and proteinases near INTACS, suggesting ongoing lysis and remodeling of corneal stroma. Although PCT was thickened for five years because of the spacer effect of INTACS, the slight decrease of PCT in the current study might be related to remodeling of corneal stroma. Otherwise, TCT or CCT was not changed throughout five years (Fig. 1F).
In HOA, only coma RMS continuously decreased throughout five years (Fig. 2), perhaps because of flattening and decreased corneal irregularity (Figs 1C, 3 and Table 3). HOA change after INTACS implantation in keratoconus has rarely been reported. Furthermore, long-term HOA change after INTACS implantation has not been sufficiently evaluated yet. Pinero et al. have reported that coma-like aberration is decreased in keratoconus after KeraRing implantation at three months23. Coma-like aberration has also been reported to be decreased in post-LASIK ectasia after INTACS implantation at two years24. Similarly, coma RMS decreased after surgery, and coma RMS continuously decreased for five years in the present study. Coma has been demonstrated to have a negative effect on visual acuity because of optical blur27. Therefore, decrease of coma along with topographic regularization might have contributed to improvement of CDVA in this study.
For the posterior cornea, posterior SimK was temporarily steepened at two months, although it recovered to the preoperative level with a small change (−8.047 to −8.297 from preoperative to postoperative two months, respectively, p = 0.007) (Fig. 1D). Posterior SimKcyl was not changed throughout five years (Fig. 1E). INTACSs were implanted into deep corneal stroma. This might have influenced posterior corneal shape. However, in this study, there was no significant change in posterior cornea for five years.
A corneal suture was made after INTACS implantation to prevent ring extrusion or intrastromal infection. This suture can influence the corneal surface. However, a corneal suture was removed after one month, and we analyzed the outcome after two months postoperatively. The effect of the suture on the cornea seemed to be negligible; so we did not take it into account when analyzing outcomes.
The limitation of this study is its small sample size. For this reason, we cannot categorize keratoconus stages from 1 to 4 separately. Given the lack of data, we could not evaluate whether eyes were progressing or not before the surgery. Further studies are needed to overcome these limitations.
In summary, we found that UDVA and refractive outcome were improved for three years. Although spherical and spherical equivalent showed myopic regression, CDVA was improved for five years because of topographic regularization and decrease of coma. Otherwise, there was no thinning of corneal thickness for five years.
In conclusion, ICRS has the advantage of improving CDVA with topographic stabilization. It can improve coma in stages 1 to 3 keratoconus for five years.
Acknowledgements
This work was supported by a grant (2016R1A6A1A03010528) of the Basic Science Research Program through the National Research Foundation (NRF) funded by the Ministry of Education, Republic of Korea.
Author Contributions
Study concept and design, M.J.K. and C.K.J. Data collection, M.J.K., Y.S.B., Y.S.Y. and W.J.W. Analysis and interpretation of data, M.J.K., C.K.J. Writing of the manuscript, M.J.K. and C.K.J. Critical revision of the manuscript, M.J.K., Y.S.B., Y.S.Y., W.J.W. and C.K.J. Statistical expertise, M.J.K., Y.S.B., Y.S.Y. and W.J.W. Administrative, technical, or material support, M.J.K., Y.S.B., Y.S.Y., W.J.W. and C.K.J. Supervision, C.K.J.
Data Availability
The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.
Competing Interests
The authors declare no competing interests.
Footnotes
Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Amanzadeh K, Elham R, Jafarzadepur E. Effects of single-segment Intacs implantation on visual acuity and corneal topographic indices of keratoconus. J Curr Ophthalmol. 2017;29:189–193. doi: 10.1016/j.joco.2016.10.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Avni-Zauberman N, Rootman DS. Cross-Linking and Intracorneal Ring Segments–Review of the Literature. Eye Contact Lens. 2015;41:254. doi: 10.1097/ICL.0000000000000169. [DOI] [PubMed] [Google Scholar]
- 3.Snibson GR. Collagen cross-linking: a new treatment paradigm in corneal disease - a review. Clin Exp Ophthalmol. 2010;38:141–153. doi: 10.1111/j.1442-9071.2010.02228.x. [DOI] [PubMed] [Google Scholar]
- 4.Pinero DP, Alio JL. Intracorneal ring segments in ectatic corneal disease - a review. Clin Exp Ophthalmol. 2010;38:154–167. doi: 10.1111/j.1442-9071.2010.02197.x. [DOI] [PubMed] [Google Scholar]
- 5.Barnett M, Mannis MJ. Contact lenses in the management of keratoconus. Cornea. 2011;30:1510–1516. doi: 10.1097/ICO.0b013e318211401f. [DOI] [PubMed] [Google Scholar]
- 6.Vega-Estrada A, Alio JL, Plaza-Puche AB. Keratoconus progression after intrastromal corneal ring segment implantation in young patients: Five-year follow-up. J Cataract Refract Surg. 2015;41:1145–1152. doi: 10.1016/j.jcrs.2014.08.045. [DOI] [PubMed] [Google Scholar]
- 7.Bromley JG, Randleman JB. Treatment strategies for corneal ectasia. Curr Opin Ophthalmol. 2010;21:255–258. doi: 10.1097/ICU.0b013e32833a8bfe. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Burris TE. Intrastromal corneal ring technology: results and indications. Curr Opin Ophthalmol. 1998;9:9–14. doi: 10.1097/00055735-199808000-00003. [DOI] [PubMed] [Google Scholar]
- 9.Schanzlin DJ, Asbell PA, Burris TE, Durrie DS. The intrastromal corneal ring segments. Phase II results for the correction of myopia. Ophthalmology. 1997;104:1067–1078. doi: 10.1016/S0161-6420(97)30183-3. [DOI] [PubMed] [Google Scholar]
- 10.Rabinowitz YS. Intacs for keratoconus. Curr Opin Ophthalmol. 2007;18:279–283. doi: 10.1097/ICU.0b013e3281fc94a5. [DOI] [PubMed] [Google Scholar]
- 11.Burris TE, Ayer CT, Evensen DA, Davenport JM. Effects of intrastromal corneal ring size and thickness on corneal flattening in human eyes. Refract Corneal Surg. 1991;7:46–50. [PubMed] [Google Scholar]
- 12.Patel S, Marshall J, Fitzke FW., III Model for deriving the optical performance of the myopic eye corrected with an intracorneal ring. J Refract Surg. 1995;11:248–252. doi: 10.3928/1081-597X-19950701-08. [DOI] [PubMed] [Google Scholar]
- 13.Kymionis GD, Bouzoukis DI, Portaliou DM, Pallikaris IG. New INTACS SK implantation in patients with post-laser in situ keratomileusis corneal ectasia. Cornea. 2010;29:214–216. doi: 10.1097/ICO.0b013e3181a26b72. [DOI] [PubMed] [Google Scholar]
- 14.Fahd DC, Alameddine RM, Nasser M, Awwad ST. Refractive and topographic effects of single-segment intrastromal corneal ring segments in eyes with moderate to severe keratoconus and inferior cones. J Cataract Refract Surg. 2015;41:1434–1440. doi: 10.1016/j.jcrs.2014.10.037. [DOI] [PubMed] [Google Scholar]
- 15.Hashemian MN, et al. Outcomes of Single Segment Implantation of Conventional Intacs versus Intacs SK for Keratoconus. J Ophthalmic Vis Res. 2014;9:305–309. doi: 10.4103/2008-322X.143359. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Zare MA, Mehrjardi HZ, Afarideh M, Bahrmandy H, Mohammadi SF. Visual, Keratometric and Corneal Biomechanical Changes after Intacs SK Implantation for Moderate to Severe Keratoconus. J Ophthalmic Vis Res. 2016;11:17–25. doi: 10.4103/2008-322X.180698. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Rho CR, et al. Changes in anterior and posterior corneal parameters in patients with keratoconus after intrastromal corneal-ring segment implantation. Curr Eye Res. 2013;38:843–850. doi: 10.3109/02713683.2013.788723. [DOI] [PubMed] [Google Scholar]
- 18.Zare MA, Hashemi H, Salari MR. Intracorneal ring segment implantation for the management of keratoconus: safety and efficacy. J Cataract Refract Surg. 2007;33:1886–1891. doi: 10.1016/j.jcrs.2007.06.055. [DOI] [PubMed] [Google Scholar]
- 19.Shetty R, et al. Intacs in advanced keratoconus. Cornea. 2008;27:1022–1029. doi: 10.1097/ICO.0b013e318172fc54. [DOI] [PubMed] [Google Scholar]
- 20.Ertan A, Kamburoglu G. Intacs implantation using a femtosecond laser for management of keratoconus: Comparison of 306 cases in different stages. J Cataract Refract Surg. 2008;34:1521–1526. doi: 10.1016/j.jcrs.2008.05.028. [DOI] [PubMed] [Google Scholar]
- 21.Bedi R, Touboul D, Pinsard L, Colin J. Refractive and topographic stability of Intacs in eyes with progressive keratoconus: five-year follow-up. J Refract Surg. 2012;28:392–396. doi: 10.3928/1081597X-20120509-01. [DOI] [PubMed] [Google Scholar]
- 22.Kymionis GD, et al. Long-term follow-up of Intacs in keratoconus. Am J Ophthalmol. 2007;143:236–244. doi: 10.1016/j.ajo.2006.10.041. [DOI] [PubMed] [Google Scholar]
- 23.Pinero DP, Alio JL, Teus MA, Barraquer RI, Uceda-Montanes A. Modeling the intracorneal ring segment effect in keratoconus using refractive, keratometric, and corneal aberrometric data. Invest Ophthalmol Vis Sci. 2010;51:5583–5591. doi: 10.1167/iovs.09-5017. [DOI] [PubMed] [Google Scholar]
- 24.Pinero DP, Alio JL, Uceda-Montanes A, El Kady B, Pascual I. Intracorneal ring segment implantation in corneas with post-laser in situ keratomileusis keratectasia. Ophthalmology. 2009;116:1665–1674. doi: 10.1016/j.ophtha.2009.05.030. [DOI] [PubMed] [Google Scholar]
- 25.Hashemi H, Amanzadeh K, Miraftab M, Asgari S. Femtosecond-assisted intrastromal corneal single-segment ring implantation in patients with keratoconus: a 12-month follow-up. Eye Contact Lens. 2015;41:183–186. doi: 10.1097/ICL.0000000000000106. [DOI] [PubMed] [Google Scholar]
- 26.Maguen E, et al. Alterations of extracellular matrix components and proteinases in human corneal buttons with INTACS for post-laser in situ keratomileusis keratectasia and keratoconus. Cornea. 2008;27:565–573. doi: 10.1097/ICO.0b013e318165b1cd. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Applegate RA, Sarver EJ, Khemsara V. Are all aberrations equal? J Refract Surg. 2002;18:S556–S562. doi: 10.3928/1081-597X-20020901-12. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The datasets generated during and/or analyzed during the current study are available from the corresponding author on reasonable request.