Published in final edited form as: Rev Sci Instrum. 2018 May;89(5):056101. doi: 10.1063/1.5023906

A 3D-printed alkali metal dispenser

E B Norrgard 1,a), D S Barker 2, J A Fedchak 2, N Klimov 2, J Scherschligt 2, S Eckel 2
PMCID: PMC6350244  NIHMSID: NIHMS1516087  PMID: 29864797

Abstract

We demonstrate and characterize a source of Li atoms made from direct metal laser sintered titanium. The source’s outgassing rate is measured to be 5 (2) · 10−7 Pa L s−1 at a temperature T = 330 °C, which optimizes the number of atoms loaded into a magneto-optical trap. The source loads ≈ 107 7Li atoms in the trap in ≈ 1s. The loaded source weighs 700mg and is suitable for a number of deployable sensors based on cold atoms.


Numerous emerging quantum technologies are being adapted from laboratory-scale experiments to deployable real-world sensors and space-based missions14. Cold atoms are at the heart of many of these applications, including interferometers3 and ultra-precise atomic clocks5,6. Recently, we have begun a program to develop a device based on trapped cold atoms which is simultaneously a primary standard and an absolute sensor of vacuum7,8.

Translating these cold atom-based technologies into deployable sensors requires an atomic source which is scaleable, lightweight, and suitable for ultra-high vacuum (UHV, < 10−6 Pa). Miniature vapor cells are well suited for sensors using atoms which have substantial vapor pressure at room temperature (such as Rb)9,10. However, many applications use atoms which must be heated to several hundred degrees Celsius to produce a substantial vapor pressure, including Li, Sr, and Yb2,5,6.

Several groups have reported direct loading of a magneto-optical trap (MOT) from effusive sources2,1113 such as ovens or alkali metal dispensers (AMDs). Some commercial AMDs are known to outgas at levels not suitable for ultra-high vacuum12 when operating. This can be mitigated by including a reducing agent, but these typically limit the alkali yield to ≈10mg. Alternatively, the small features and thin walls (for high electrical resistance) of an AMD makes additive manufacturing an attractive construction technique. Direct metal laser sintered (DMLS) titanium has recently been demonstrated to be low outgassing and suitable for in-vacuum components14 and vacuum flanges15.

In this note, we report on a lithium AMD made from DMLS grade 5 titanium (6% Al, 4% V by weight). The AMD holds ≈ 100mg of Li and is used to directly load a MOT suitable for use in cold atom experiments. The measured outgassing rate is 5(2) · 10−7 Pa L s−1 for optimal MOT loading conditions (uncertainties in this paper are the uncorrelated combination of 1σ statistical and systematic uncertainties), limited by outgassing from the loaded lithium metal.

Figure 1 shows the 3D-printed AMD, which consists of two parts: a tube (5.1 mm diameter, 13.8 mm long, 0.13 mm wall) and a plug (1.2 mm thick), which snugly fits into the open end of the tube. Both the tube and plug have 0.25 mm thick, 5.1 mm wide tabs with a 2.4 mm diameter clearance hole for mounting and electrical connections. A 5.1 mm long, 0.25 mm wide slit in the tube roughly directs the exiting alkali atoms orthogonal to the plane of the tabs. The combined tube and plug have a measured mass of 584 (2) mg and designed total surface area (not including surface roughness) of 6.8 cm2.

FIG. 1.

FIG. 1.

(Color online) Photograph of the 3D-printed titanium cap (left) and tube (right).

The AMD is loaded with metallic lithium and inserted into the vacuum chamber while under an argon-purged atmosphere. We loaded the AMD with seven pellets of natural isotopic abundance Li, which we estimate to total 100mg. The plug fits tightly into the tube to secure the Li. The AMD is mounted on oxygen-free, high-conductivity Cu electrical feedthroughs 35mm from the center of the MOT.

The AMD is resistively heated by a current of typically 10A to 15A. Upon initial warm up, the resistance of the AMD dropped by nearly a factor of two, presumably as the (relatively conductive) Li melted and came into better electrical contact with the tube. We therefore characterize the AMD in terms of the power P dissipated across the AMD. With subsequent operations, we have noticed further decreases in the AMD resistance at the 1% level, and noticed a small accumulation of Li outside the slit of the tube. A thin Ni mesh inside the tube could be used to wick the Li in a future design16.

A laser beam counterpropagating to the Li beam allows the temperature T of the Li beam to be determined spectroscopically. We monitor the laser induced fluorescence collected on a charge-coupled device (CCD) camera from a spatial region along the laser beam. The frequency-dependent fluorescence is fit to a Maxwell-Boltzmann distribution for a 1D beam. The temperature was measured for a number of powers and fit to an empirical model containing conductive (PTTr) and radiative (P~T4Tr4) terms, where Tr = 20(1) °C is the measured room temperature. In the experiments discussed below, this fit was used to determine the Li beam temperatures from the measured power dissipated in the AMD.

All stainless steel components of the vacuum chamber were baked at 450 °C for 21 days prior to assembly to reduce hydrogen outgassing17,18. All vacuum components, including the 3D-printed titanium AMD, were cleaned in alkaline detergent, followed by acetone, then ethanol. No cleaning procedure was performed on the Li beyond selecting pellets with minimal black nitride coating.

The chamber was pumped by a 50 Ls−1 ion pump and monitored by a mass spectrometer (Granville-Phillips Vacuum Quality Monitor, with an ion gauge mounted nearby measuring total pressure19). The AMD was degassed for 3 days by dissipating P = 2.0 W (T ≈ 240 °C). We did not otherwise bake the vacuum chamber to remove water. Upon initial degassing, the pressure increased to 3 · 10−5 Pa, dominated by water. After degassing, the mass spectrometer recorded a significant decrease in all gasses except for m = 28 u. Outgassing that is not a strong function of time is likely due to species that are chemically bound or are diffusing from bulk material. Because Li forms a nitride layer in the presence of air, we suspect the m = 28 peak to be N2 originating from the Li pellets.

Figure 2 shows the increase in background gas composition recorded on the mass spectrometer when operating the AMD under typical conditions (P = 3.0 W, T ≈ 330 °C, total pressure 3.3(6) · 10−7 Pa). The presence of a m = 14 u peak roughly 1/10 the intensity of the m = 28 u peak is consistent with the cracking fraction of N2 in most mass spectrometers. The lack of a visible m = 12 u peak indicates the possible contribution of CO to the m = 28 u peak is small. The magnitude of the observed m = 44 u peak indicates a negligible contribution to the m = 28 u peak from cracking of CO2 into CO (28 u) and O (16 u). In addition, the partial pressure of O2 (32 u) is observed to increase. This too may originate from the Li pellets because Li forms hydroxides and carbonates upon exposure to air.

FIG. 2.

FIG. 2.

(Color online) Background-subtracted (i.e. source on minus source off) mass spectrum of gas composition with T = 330°C. The total pressure including background is 3.3(6) · 10−7 Pa.

We investigated the outgassing rate of the AMD by the throughput method, wherein the pressure p in a vacuum chamber is determined by the total gas flow q from all sources into the chamber and effective pumping speed S, q = pS. The AMD was tested in a small stainless-steel vacuum chamber evacuated by a turbomolecular pump. The effective pumping speed S was limited by the conductance of the vacuum components between the AMD and the turbomolecular pump; we roughly estimate S ≈ 7Ls−1 for N2 at Trm = 20 °C. The outgassing rate for the AMD qAMD is determined by the measured pressure rise pAMD above the background when the source is turned on. We obtain an N2 outgassing rate of qAMD = 5(2) · 10−7 Pa L s−1 for the source operating at 330 °C. Our estimate of qAMD does not take into account possible outgassing of the chamber due to heating from the AMD, and therefore represents an upper limit.

The MOT consists of six independent, circularly polarized laser beams, detuned −18MHz from the F = 2 → F′ = 3 transition of the Li D2 line. Each beam has Gaussian rms width 3.6 (2) mm and power 40 (1)mW. An electro-optic modulator adds 814MHz RF sidebands (≈ 20% in each of the ±1 order sidebands) to the beams in order to drive the F = 1 → F′ = 2 repump transition. A quadrupole magnetic field with axial gradient Bz=3 mT/cm is formed by two arrays of grade N52 permanent magnet bars held in 3D-printed acrylonitrile butadiene styrene mounts around the vacuum chamber.

The number of atoms N in the MOT is determined by fluorescence imaging on a CCD camera. We estimate the measured atom number to be accurate to within a factor of 2. The load dynamics of the MOT are well described by the differential equation

dNdt=RγNβn2d3x, (1)

Here, n is the atomic density, R is the trap loading rate, and γ and β are the one- and two-body loss rate coefficients, respectively. Equation 1 has solution

N(t)=NS1eγ0t1+NS2βVReγ0t (2)
γ0=γ1+4βRVγ2, (3)

where NS is the steady-state atom number, and we assume the MOT occupies a fixed volume V determined by the fitted Gaussian widths on the CCD camera, such that n = N/V .

Figure 3 shows the loading rate R and steady-state atom number NS as a function of Li beam temperature T. We observe loading rates as high as 5 · 108 s1, comparable to that of many Zeeman slowers20. The steady state atom number initially increases as T is increased, reaching a maximum NS ≈ 107. For temperatures in excess of 330 °C, the increase in background pressure due to outgassing exceeds the increase in Li production, and the trapped atom number decreases marginally. At the optimal T = 330 °C, we measure an equilibrium N2-equivalent pressure of 1.8(4) · 10−7 Pa on an ion gauge attached to the vacuum chamber.

FIG. 3.

FIG. 3.

(Color online) Loading rate R (black squares), steady-state atom number NS (red circles), and one-body loss rate γ (blue triangles) as a function of temperature T. The inset shows a typical MOT loading curve with T = 330°C. The red line is a fit to Eq.2.

AMDs have proven useful for a variety of cold atom experiments. The 3D-printed AMD presented here loads a Li MOT with comparable atom number, load rate, and background pressure to other MOTs directly loaded from effusive sources2,1113. Reference16 shows that with careful preparation of the Li, similar loading rates to those demonstrated here are attainable at operating pressures an order of magnitude lower (and thus NS an order of magnitude higher). We plan to use this 3D-printed AMD in designing a cold atom vacuum standard based on Li7,8. Our design is generic to a number of atomic species; specifically Sr, Yb, Mg, and Ca all have negligible vapor pressure at room temperature, but can achieve vapor pressures comparable to Li at comparable (±50K) operating temperatures. Thus, this source warrants consideration for other deployable technologies based on cold atoms, including clocks and accelerometers.

ACKNOWLEDGMENTS

The authors thank M. Sefa for initial measurements of 3D-printed titanium outgassing and S. Maxwell and W. McGehee for useful comments on the manuscript.

References

  • 1.Wang R, Liu M, Minardi F, and Kasevich M, Phys. Rev. A 75, 013610 (2007). [Google Scholar]
  • 2.Ladouceur K, Klappauf BG, Dongen JV, Rauhut N, Schuster B, Mills AK, Jones DJ, and Madison KW, J. Opt. Soc. Am. B 26, 210 (2009). [Google Scholar]
  • 3.Hogan JM, Johnson DMS, Dickerson S, Kovachy T, Sugarbaker A, Chiow S.-w., Graham PW, Kasevich MA, Saif B, Rajendran S, Bouyer P, Seery BD, Feinberg L, and KeskiKuha R, General Relativity and Gravitation 43, 1953 (2011). [Google Scholar]
  • 4.Müntinga H, Ahlers H, Krutzik M, Wenzlawski A, Arnold S, Becker D, Bongs K, Dittus H, Duncker H, Gaaloul N, Gherasim C, Giese E, Grzeschik C, Hänsch TW, Hellmig O, Herr W, Herrmann S, Kajari E, Kleinert S, Lämmerzahl C, Lewoczko-Adamczyk W, Malcolm J, Meyer N, Nolte R, Peters A, Popp M, Reichel J, Roura A, Rudolph J, Schiemangk M, Schneider M, Seidel ST, Sengstock K, Tamma V, Valenzuela T, Vogel A, Walser R, Wendrich T, Windpassinger P, Zeller W, van Zoest T, Ertmer W, Schleich WP, and Rasel EM, Phys. Rev. Lett 110, 093602 (2013). [DOI] [PubMed] [Google Scholar]
  • 5.Beloy K, Hinkley N, Phillips NB, Sherman JA, Schioppo M, Lehman J, Feldman A, Hanssen LM, Oates CW, and Ludlow AD, Phys. Rev. Lett 113, 260801 (2014). [DOI] [PubMed] [Google Scholar]
  • 6.Marti GE, Hutson RB, Goban A, Campbell SL, Poli N, and Ye J, ArXiv e-prints (2017), arXiv:1711.08540 [physics.atom-ph].
  • 7.Scherschligt J, Fedchak JA, Barker DS, Eckel S, Klimov N, Makrides C, and Tiesinga E, Metrologia 54, S125 (2017). [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Jousten K, Hendricks J, Barker D, Douglas K, Eckel S, Egan P, Fedchak J, Flgge J, Gaiser C, Olson D, Ricker J, Rubin T, Sabuga W, Scherschligt J, Schdel R, Sterr U, Stone J, and Strouse G, Metrologia 54, S146 (2017). [Google Scholar]
  • 9.Perez MA, Nguyen U, Knappe S, Donley EA, Kitching J, and Shkel AM, Sensors and Actuators A: Physical 154, 295 (2009), 21st IEEE International Conference on Micro Electro Mechanical Systems, Tucson, AZ, USA, 13–17 January 2008. [Google Scholar]
  • 10.Mhaskar R, Knappe S, and Kitching J, Applied Physics Letters 101, 241105 (2012), 10.1063/1.4770361. [DOI] [Google Scholar]
  • 11.Moore KL, Purdy TP, Murch KW, Leslie S, Gupta S, and Stamper-Kurn DM, Review of Scientific Instruments 76, 023106 (2005), 10.1063/1.1841852. [DOI] [Google Scholar]
  • 12.Muhammad R, Ramirez-Serrano J, Magalhes K, Paiva R, Shiozaki R, Weiner J, and Bagnato V, Optics Communications 281, 4926 (2008). [Google Scholar]
  • 13.Scherer DR, Fenner DB, and Hensley JM, Journal of Vacuum Science & Technology A: Vacuum, Surfaces, and Films 30, 061602 (2012), 10.1116/1.4757950. [DOI] [Google Scholar]
  • 14.Gans AR, Jobbins MM, Lee DY, and Kandel SA, Journal of Vacuum Science & Technology A: Vacuum, Surfaces, and Films 32, 023201 (2014), 10.1116/1.4846195. [DOI] [Google Scholar]
  • 15.Vovrosh J, Voulazeris G, Petrov P, Zou J, Gaber Y, Benn L, Woolger D, Attallah MM, Boyer V, Bongs K, and Holynski M, ArXiv e-prints (2017), arXiv:1710.08279 [physics.app-ph]. [DOI] [PMC free article] [PubMed]
  • 16.Gunton W, Semczuk M, and Madison KW, Phys. Rev. A 88, 023624 (2013). [Google Scholar]
  • 17.Santeler DJ, Journal of Vacuum Science Technology 10, 1879 (1992). [Google Scholar]
  • 18.Sefa M, Fedchak JA, and Scherschligt J, Journal of Vacuum Science & Technology A: Vacuum, Surfaces, and Films 35, 041601 (2017), 10.1116/1.4983211. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Any mention of commercial products within this work is for information only; it does not imply recommendation or endorsement by NIST,.
  • 20.Peixoto FM, Enhanced Loading of a Lithium 7 Magneto Optical Trap using Transverse Cooling and Frequency Spread Light, Ph.D. thesis, Yale University; (2002). [Google Scholar]

RESOURCES