Electrode Designs for Protection of the Delicate Cochlear Structures

Mark Downing 1,
PMCID: PMC6354533  PMID: 30644381

Abstract

The most recent electrode introductions from Advanced Bionics, the HiFocus Mid-Scala and the HiFocus SlimJ arrays, have common design goals intended to provide sufficient access to the necessary frequency range while avoiding intracochlear trauma. The electrode choice, either a pre-curved (mid-scala) array or straight (lateral) array, can be made by the surgeon based on anatomical considerations and/or their preferred surgical approach. Both arrays offer ease of handling, suitability for a round window and cochleostomy based insertion and control of the insertion speed.

Keywords: Cochlear implant, electrode, HiFocus, Mid-Scala, SlimJ, pre-curved

INTRODUCTION

The most recent electrodes introduced by Advanced Bionics, the HiFocus Mid-Scala and the HiFocus SlimJ arrays (Advanced Bionics AG, Stäfa, Switzerland), have common design goals intended to provide sufficient access to the necessary frequency range while avoiding intracochlear trauma. The electrode choice, either a pre-curved (mid-scala) array or straight (lateral) array, can be made by surgeons based on anatomical considerations and/or their preferred surgical approach. Both of these arrays offer ease of handling, suitability for a round window- or cochleostomy-based insertion, and control of the insertion speed.

Electrode Design Goals

The design and development of the HiFocus Mid-Scala and HiFocus SlimJ electrodes has been documented in the company White Papers [1,2] and includes their histological assessment. HiFocus Mid-Scala has been designed to sit close to, but not against, the medial wall (Figure 1), whereas HiFocus SlimJ is placed closer to the lateral wall (Figure 2). Both electrodes are available on the HiRes Ultra 3D Cochlear Implant (Advanced Bionics AG, Stäfa, Switzerland), with an innovative magnet designed for hassle-free magnetic resonance imaging.

Figure 1.

Figure 1

HiFocus Mid-Scala electrode.

Figure 2.

Figure 2

HiFocus SlimJ electrode.

The design features of the electrodes are shown in Figures 3 and 4. The tip feature of both the electrodes is intended to aid insertion through the round window. Blue markers on the electrodes guide the surgeon with the suggested insertion technique. The pre-curved HiFocus Mid-Scala electrode has a straight apical portion of the array to reduce the risk of tip fold-over during insertion that may be seen with a peri-modiolar design [3, 4].

Figure 3.

Figure 3

HiFocus Mid-Scala electrode showing the pre-curved shape to conform to the cochlea. It is straightened using a stylet that is removed during insertion.

Figure 4.

Figure 4

HiFocus SlimJ electrode has a gentle curvature to ensure movement of the electrode in the apical direction during insertion. The proximal wing feature allows easy handling and insertion.

The designs of both HiFocus Mid-Scala and SlimJ electrodes have demonstrated excellent preservation of the delicate cochlear structures and consistent placement (Figures 5 and 6).

Figure 5.

Figure 5

HiFocus Mid-Scala electrode placement in the scala tympani adjacent to the medial wall and away from the cochlear structures.

Figure 6.

Figure 6

HiFocus SlimJ electrode placement in the scala tympani toward the lateral wall and away from the cochlear structures.

Asymmetric stiffness to limit the vertical movement of the array in the scala tympani is a prominent feature of the design of both the electrodes. These features, together with a reduced profile, have been shown to reduce or eliminate intracochlear trauma [58]. Preservation of the cochlear structures is beneficial, even if there is no aidable residual hearing, because less trauma has been shown to provide better electrical-only hearing [9].

An angular insertion depth of 420° as measured from the round window was chosen for the designs based on the corresponding tonotopic spiral ganglion location [10] that is sufficient to access the frequency ranges necessary for high-level speech recognition without increasing the risk of trauma or perceptual distortions reported with deeper insertions [1113]. Microanatomical examination of the temporal bones using micro-computed tomography (CT) [14] has revealed that the dimensions of the scala tympani markedly reduce after 450°. A more consistent insertion depth may result in a greater efficiency in establishing sound processor fitting parameters and improved overall performance [15] and may reduce the time required for adaptation and learning after device fitting [16].

In temporal bone studies [1, 2], the HiFocus Mid-Scala electrode has demonstrated an average angular insertion depth of 420°, with a standard deviation (SD) of 21°, and the HiFocus SlimJ electrode an average angular insertion depth of 413° with an SD of 42°. Larger variability in the depth of insertion has been reported for other electrode designs [1722], both lateral wall and peri-modiolar.

Both electrode designs are capable of providing a controlled insertion, and this has been shown to be beneficial [23].

Real-Time Monitoring During Electrode Insertion

A promising new area of exploration in the field of cochlear implants is the use of real-time electrocochleography (ECochG) recording during electrode insertion to monitor potentials in the cochlea. It has been reported that this real-time monitoring may provide feedback that can improve surgical technique and hearing preservation [24, 25]. The Advanced Bionics ECochG research system (Advanced Bionics AG, Stäfa, Switzerland) can make these measurements with both the HiFocus Mid-Scala and the HiFocus SlimJ electrodes and has been used extensively to conduct research on real-time ECochG [2527]. This research continues to grow and more firmly indicate that real-time ECochG is an important tool for cochlear implant surgery and should be used in combination with electrodes, e.g., HiFocus Mid-Scala and the HiFocus SlimJ, that have been successfully designed for atraumatic insertion.

CONCLUSION

Modern electrode designs, either pre-curved or straight arrays, can offer sufficient cochlear coverage while avoiding the risk of damage to the delicate cochlear structures. The real-time monitoring of cochlear potentials during insertion will lead to better patient outcomes with current electrode designs and may influence electrode designs of the future.

Footnotes

Peer-review: Externally peer-reviewed.

Conflict of Interest: The author is an employee of Advanced Bionics LLC.

Financial Disclosure: The author is an employee of Advanced Bionics LLC.

REFERENCES

  • 1.HiFocus Mid-Scala Electrode White Paper, 02-M224-03 Rev C
  • 2.HiFocus SlimJ Electrode White Paper, 027-N086-02
  • 3.Aschendorff A1, Briggs R, Brademann G, Helbig S, Hornung J, Lenarz T, et al. Clinical Investigation of the Nucleus Slim Modiolar Electrode. Audiol Neurotol. 2017;22:169–79. doi: 10.1159/000480345. [DOI] [PubMed] [Google Scholar]
  • 4.McJunkin JL, Durakovic N, Herzog J, Buchman CA. Early Outcomes with a Slim, Modiolar Cochlear Implant Electrode Array. Otol Neurotol. 2018;39:e28–e33. doi: 10.1097/MAO.0000000000001652. [DOI] [PubMed] [Google Scholar]
  • 5.Boyle PJ. The Rationale for a Mid-Scala Electrode Array. Euro Ann Otorhinolaryngol. 2016;133S:S61–S62. doi: 10.1016/j.anorl.2016.05.002. [DOI] [PubMed] [Google Scholar]
  • 6.Hassepass F, Bulla S, Maier W, Laszig R, Arndt S, Beck R, et al. The New Mid-Scala Electrode Array: A Radiologic and Histologic Study in Human Temporal Bones. Otol Neurotol. 2014;35:1415–20. doi: 10.1097/MAO.0000000000000412. [DOI] [PubMed] [Google Scholar]
  • 7.Rebscher SJ, Hetherington A, Bonham B, Wardrop P, Whinney D, Leake PA. Considerations for the Design of Future Cochlear Implant Electrode Arrays: Electrode Array Stiffness, Size, and Depth of Insertion. J Rehab Res Develop. 2008;45:731–48. doi: 10.1682/JRRD.2007.08.0119. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Wright CG, Roland PS, Kuzma J. Advanced Bionics Thin Lateral and Helix II Electrodes: A Temporal Bone Study. Laryngoscope. 2005;115:2041–5. doi: 10.1097/01.MLG.0000181461.63392.49. [DOI] [PubMed] [Google Scholar]
  • 9.Aschendorff A, Kromeier J, Klenzner T, Laszig R. Quality Control after Insertion of the Nucleus Contour and Contour Advance Electrode in Adults. Ear Hear. 2007;28(Suppl 2):75S–9S. doi: 10.1097/AUD.0b013e318031542e. [DOI] [PubMed] [Google Scholar]
  • 10.Stakhovskaya O, Sridhar D, Bonham BH, Leake PA. Frequency Map for the Human Cochlear Spiral Ganglion: Implications for Cochlear Implants. J Assoc Res Otolaryngol. 2007;8:220–33. doi: 10.1007/s10162-007-0076-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Adunka O, Kiefer J. Impact of Electrode Insertion Depth on Intra-Cochlear Trauma. Otolaryngol Head Neck Surg. 2006;135:374–82. doi: 10.1016/j.otohns.2006.05.002. [DOI] [PubMed] [Google Scholar]
  • 12.Baumann U, Nobbe A. The Cochlear Implant Electrode-Pitch Function. Hear Res. 2006;213:34–42. doi: 10.1016/j.heares.2005.12.010. [DOI] [PubMed] [Google Scholar]
  • 13.Gani M, Valentini G, Sigrist A, Kós MI, Boëx C. Implications of Deep Electrode Insertion on Cochlear Implant Fitting. J Assoc Res Otolaryngol. 2007;8:69–83. doi: 10.1007/s10162-006-0065-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Avci E, Nauwelaers T, Lenarz T, Hamacher V, Kral A. Variations in Microanatomy of the Human Cochlea. J Comp Neurol. 2014;522:3245–61. doi: 10.1002/cne.23594. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Başkent D, Shannon RV. Interactions between Cochlear Implant Electrode Insertion Depth and Frequency-Place Mapping. J Acoust Soc Am. 2005;117:1405–16. doi: 10.1121/1.1856273. [DOI] [PubMed] [Google Scholar]
  • 16.Reiss LA, Turner CW, Erenberg SR, Gantz BJ. Changes in Pitch with a Cochlear Implant Over Time. J Assoc Res Otolaryngol. 2007;8:241–57. doi: 10.1007/s10162-007-0077-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Adunka O, Kiefer J, Unkelbach MH, Lehnert T, Gstoettner W. Development and Evaluation of an Improved Cochlear Implant Electrode Design for Electric Acoustic Stimulation. Laryngoscope. 2004;114:1237–41. doi: 10.1097/00005537-200407000-00018. [DOI] [PubMed] [Google Scholar]
  • 18.James CJ, Fraysse B, Deguine O, Lenarz T, Mawman D, Ramos A, et al. Combined Electroacoustic Stimulation in Conventional Candidates for Cochlear Implantation. Audiol Neurotol. 2006;11(Suppl 1):57–62. doi: 10.1159/000095615. [DOI] [PubMed] [Google Scholar]
  • 19.Radeloff A, Mack M, Baghi M, Gstoettner WK, Adunka OF. Variance of Angular Insertion Depth in Free-Fitting and Peri-Modiolar Cochlear Implant Electrodes. Otol Neurotol. 2008;29:131–6. doi: 10.1097/MAO.0b013e318157f0ea. [DOI] [PubMed] [Google Scholar]
  • 20.Stöver T, Issing P, Graurock G, Erfurt P, ElBeltagy Y, Paasche G, et al. Evaluation of the Advance Off-Stylet Insertion Technique and the Cochlear Insertion Tool in Temporal Bones. Otol Neurotol. 2005;26:1161–70. doi: 10.1097/01.mao.0000179527.17285.85. [DOI] [PubMed] [Google Scholar]
  • 21.Tykocinski M, Saunders E, Cohen LT, Treaba C, Briggs RJ, Gibson P, et al. The Contour Electrode Array: Safety Study and Initial Patient Trials of a New Peri-Modiolar Design. Otol Neurotol. 2001;22:33–41. doi: 10.1097/00129492-200101000-00007. [DOI] [PubMed] [Google Scholar]
  • 22.Wardrop P, Whinney D, Rebscher SJ, Roland JT, Jr, Luxford W, Leake PA. A Temporal Bone Study of Insertion Trauma and Intracochlear Position of Cochlear Implant Electrodes. I: Comparison of Nucleus Banded and Contour Electrodes. Hear Res. 2005;203:54–67. doi: 10.1016/j.heares.2004.11.006. [DOI] [PubMed] [Google Scholar]
  • 23.Todt I, Mittmann P, Ernst A. Intracochlear Fluid Pressure Changes Related to the Insertion Speed of a CI Electrode. Biomed Res Int. 2014:16. doi: 10.1155/2014/507241. 507241. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Mandalà M, Colletti L, Tonoli G, Colletti V. Electrocochleography during Cochlear Implantation for Hearing Preservation. Otolaryngol Head Neck Surg. 2012;146:774–81. doi: 10.1177/0194599811435895. [DOI] [PubMed] [Google Scholar]
  • 25.Koka K, Riggs WJ, Dwyer R, Holder JT, Noble JH, Dawant BM, et al. Intra-Cochlear Electrocochleography During Cochlear Implant Electrode Insertion Is Predictive of Final Scalar Location. Otol Neurotol. 2018;39:654–9. doi: 10.1097/MAO.0000000000001906. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.O’Connell BP, Holder JT, Dwyer RT, Gifford RH, Noble JH, Bennett ML, et al. Intra- and Posoperative Electrocochleography May Be Predictive of Final Electrode Position and Postoperative Hearing Preservation. Front Neurosci. 2017;11:291. doi: 10.3389/fnins.2017.00291. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Harris MS, Riggs WJ, Giardina CK, O’Connell BP, Holder JT, Dwyer RT, et al. Patterns Seen During Electrode Insertion Using Intracochlear Electrocochleography Obtained Directly Through a Cochlear Implant. Otol Neurotol. 2018;38:1415–20. doi: 10.1097/MAO.0000000000001559. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from The Journal of International Advanced Otology are provided here courtesy of The European Academy of Otology and Neurotology

RESOURCES