Validation of an equation for energy expenditure that does not require the respiratory quotient

Karl J Kaiyala 1,*, Brent E Wisse 2, John R B Lighton 3
Editor: Andrej A Romanovsky4
PMCID: PMC6358081  PMID: 30707737

Abstract

Background

Energy expenditure (EE) calculated from respirometric indirect calorimetry is most accurate when based on oxygen consumption (VO2), carbon dioxide production (VCO2) and estimated protein metabolism (PM). EE has a substantial dependence of ~7% on the respiratory quotient (RQ, VCO2/VO2) and a lesser dependence on PM, yet many studies have instead estimated EE from VO2 only while PM has often been ignored, thus reducing accuracy. In 1949 Weir proposed a method to accurately calculate EE without using RQ, which also adjusts for estimated PM based on dietary composition. This RQ- method utilizes the calorimeter airflow rate (FR), the change in fractional O2 concentration (ΔFO2) and the dietary protein fraction. The RQ- method has not previously been empirically validated against the standard RQ+ method using both VO2 and RQ. Our aim was to do that.

Methods

VO2 and VCO2 were measured repeatedly in 8 mice fed a high protein diet (HPD) during exposure to different temperatures (n = 168 measurements of 24h gas exchange). The HPD-adjusted RQ+ equation was: EE [kcal/time] = VO2 [L/time]×(3.853+1.081RQ) while the corresponding RQ- equation was: EE = 4.934×FR×ΔFO2. Agreement was analyzed using the ratios of the RQ- to RQ+ methods along with regression and Bland-Altman agreement analyses. We also evaluated the standard equation using the dietary food quotient (FQ) of 0.91 as a proxy for RQ (FQ+ method).

Results

Ratio analysis revealed that the mean error of the RQ- method was only 0.11 ± 0.042% while the maximum error was only 0.21%. Error using the FQ+ method was 4 -and 10-fold greater, respectively. Bland-Altman analysis demonstrated that the RQ- method very slightly overestimates EE as RQ decreases. Theoretically, this error can be eliminated completely by imposing an incurrent fractional oxygen concentration at a value only slightly greater than the atmospheric level.

Conclusions

The Weir ‘RQ-free’ method for calculating EE is a highly valid alternative to the ‘gold standard’ method that requires RQ. The RQ- approach permits reduced cost and complexity in studies focused on EE and provides a way to rescue EE measurement in studies compromised by faulty CO2 measurements. Practitioners of respirometry should consider adjusting EE calculations for estimated protein metabolism based on dietary composition.

Introduction

Animal life is powered by catalytic combustion, the intricate “Fire of Life” [1] by which oxygen necessarily participates in the thermochemical transformation of food energy into biological work and heat. In accord with Hess’s law of constant heat sums (a restatement of the first law of thermodynamics) [2], the quantity of heat released by the low temperature biological “fire” is the same as when an equivalent amount food is suddenly combusted in a bomb calorimeter. Accordingly, in the 242 years since Lavoisier first proposed an obligatory metabolic role for “oxygene” [2], technological and conceptual advances have steadily improved the ability to quantify energy expenditure (EE) using mathematical transformations of oxygen consumption (VO2), carbon dioxide production (VCO2), or both to achieve the greatest accuracy [15].

The ratio of VCO2 to VO2, termed the respiratory quotient (RQ), normally ranges between ~0.7 and ~1.0 depending largely on the proportions of carbohydrate, fat and protein being combusted. EE is commonly calculated by multiplying VO2 times a linear transform of the form (A+B×RQ), where A and B are coefficients whose ‘most accurate’ values have a modest inverse dependence on the rate of protein metabolism as estimated from nitrogen excretion or diet composition [6, 7]. By contrast, the per liter energy equivalent of VO2 depends more importantly on RQ: when RQ equals 0.7, each liter of VO2 represents ~4.7 kcal, but when RQ equals 1.0 the value is ~5 kcal. It is, therefore, widely understood that the most accurate estimation of EE from respiratory gas exchange requires measurements of both VO2 and VCO2.

One of the most widely employed equations for calculating EE from respirometric data was introduced in 1949 by Weir in a paper focused principally on adjusting the EE calculation for protein metabolism [6]:

EE=V˙O2(3.941+1.106(RQ))(1+0.082P) [1]

where P is the proportion of total EE accounted for by protein metabolism [6, 7]. EE is in units of kcal/h when VO2 is in L/h, and so forth. VO2 is corrected to standard temperature and pressure, dry (STPD). We denote Eq 1 the RQ+ method.

In the same paper [6] that introduced Eq 1, Weir also proposed a second and theoretically accurate equation for EE that relies only on the airflow rate and change in fractional oxygen concentrations; if P is 0.1, and the excurrent flow rate is measured, the equation is simply:

EE=5×V˙e×ΔFO2 [2]

where Ve is the excurrent flow rate corrected to STPD, and ΔFO2 is the incurrent minus excurrent fractional oxygen difference (FiO2 minus FeO2). Of critical importance to the validity of Eq 2 (and for the calculation of VO2 and RQ in Eq 1), the fractional gas concentrations must be either scrubbed of water vapor or mathematically corrected for its presence using Dalton’s law of partial pressures [35] to avoid dilution by this gas, which otherwise would result in marked overestimation of VO2, a lesser overestimation of VCO2, and a consequent dramatic underestimation of RQ [3, 4] (discussed below).

It is important to emphasize that the constant term in Eq 2 depends on the fraction of total EE accounted for by protein fraction in Eq 1, as well as on the particular EE equation employed to transform gas exchange into EE (e.g., an updated equation has been proposed by Hall and colleagues [7]). Therefore, we denote Eq 2 as being representative of the RQ- method.

The RQ- method is all but unknown, likely overshadowed by Weir’s canonical EE equation [6] and perhaps impeded also by the somewhat arcane theoretical explanations put forward by Weir and others [3, 6, 8, 9]. Perhaps the most important reason, however, is that (to our knowledge) the RQ- method has never been validated in a rigorous empirical test.

If the RQ- method in fact agrees very well with the standard RQ+ method, its adoption could confer a number of important advantages (discussed below). Accordingly, our major aim was to validate the RQ- method. We also present a new and hopefully more rigorous and transparent explanation for why the method should work, and demonstrate that, theoretically, the agreement between the RQ- and RQ+ methods can be made perfect by imposing an incurrent oxygen fraction that is only slightly greater than the normal atmospheric value. Finally, our analysis suggests that protein metabolism should get wider consideration in the application of respirometry.

Materials and methods

Subjects, diet and institutional approval

Male C57Bl/6J mice (N = 8, age ~8 weeks; Jackson Laboratories (Bar Harbor, ME)) were housed in a constant temperature walk-in room housed within the AAALAC-accredited animal care facility at the University of Nevada at Las Vegas. Mice were individually housed within live-in, unsealed, pull-mode metabolic measurement cages (see [35] for discussions of pull- vs. push-mode respirometry). The mice were fed the Labdiet 5001 (LabDiets, St Louis, MO), a high protein diet (HPD; 28.5% of kcals; fat = 13.4% of kcals) having a food quotient (FQ) of 0.906 (calculated using the ‘Indirect Calorimetry Equations’ supplement in [7]). Mice were supplied with water ad libitum, and provided with Bed O'Cobs 1/4” bedding material (Anderson Industrial Products, Maumee, OH).

All study procedures were approved by the Institutional Animal Care and Use Committee (IACUC) at the University of Nevada at Las Vegas. The program is fully accredited by the Association for the Assessment and Accreditation for Laboratory Animal Care International (AAALAC).

Derivation of the RQ- equation for the high protein diet

A valid expression for VO2, easily derived from equation 11.2 on p.126 in [3], is:

V˙O2=V˙e(FiO2FeO2)1FiO2+FiO2RQ [3]

(We emphasize that the flow rate and fractional gas concentrations must be mathematically corrected for or scrubbed of water vapor).

Next, to compute EE in accordance with Eq 1, we multiply VO2 times the Weir RQ+ transform adjusted for the HPD (i.e., 3.853 + 1.081RQ), but write the result as follows:

EE=V˙e(FiO2FeO2)(3.853+1.081RQ)(1FiO2+FiO2RQ) [4]

Note that we intentionally placed the RQ+ transform over the denominator of Eq 3 for VO2 and placed that quotient (Q) to the right in Eq 4 because, as is easily confirmed, Q very nearly equals 4.934 for any value of RQ ranging from 0.7 to 1.0 (note that 4.934 is the value of the RQ+ transform when RQ = 1.0). A formal (and to our knowledge novel) explanation for why this pertains is as follows: Multiply the numerator of Q by 1.0 but express this as the ratio 4.934÷4.934. Next, substitute the atmosphere’s near constant fractional O2 of 0.20939 [10] into the denominator of Q. The result can then be written as:

EE=V˙e(FiO2FeO2)(0.7809+0.2191RQ)(0.7906+0.2094RQ)(4.934)V˙e(FiO2FeO2)(4.934) [5]

Remarkably, the numerator and denominator of the ratio in Eq 5 are nearly identical linear functions of RQ. Therefore, the ratio remains very close to 1.0 across the entire range of RQ; specifically, it equals 1.0 when RQ equals 1.0, and equals 0.9969 when RQ equals 0.7. Accordingly, and as indicated above, Eq 5 can be simplified to exclude any dependence on RQ with almost no loss of accuracy. We employed the simplified expression of Eq 5 to compute RQ- EE in the present work.

It can be shown that the RQ- minus RQ+ EE difference in kcal/h calculated as the simplified version of Eq 5 minus Eq 1 adjusted for the HPD equals the equation:

ΔEE=0.0029V˙O2(1RQ) [6]

Thus, for a given RQ <1.0, the EE difference is predicted to scale directly but only slightly with VO2, and for a given VO2, the difference is predicted to increase linearly, but again only slightly, as RQ decreases.

Respirometry

Metabolic rates were measured using an 8-cage Promethion-C continuous, parallel metabolic phenotyping system (Sable Systems International (SSI), Las Vegas, NV; SSI). This system imposes minimal stress due to handling or other disruptive influences because it uses live-in cages of ~8 L STP internal volume that are transferred from the housing colony to the testing room for studies. Air was pulled from the cages at a controlled mass flow rate of 2 L/min STP. This yielded a time constant of ~4 min. The flow from each cage was sampled by a gas analysis chain consisting of a water vapor analyzer a CO2 analyzer, an O2 analyzer, a barometric pressure sensor, and a subsampling flow control system, all integrated into one gas analysis system (GA3m4: SSI) per bank of 4 cages. Gas flow for each bank was generated by a FR-4b mass flow controlled pull flow generator (SSI). The calorimeter room also incorporated a fluorescent light source controlled by a timer set to a 12:12 light:dark cycle.

The system acquired data on fractional O2 and CO2 concentrations, water vapor pressure (WVP), barometric pressure (BP), ambient temperature and light levels, flow rates, food and water dispenser masses (to 1 mg), body masses (to 1 mg via a weighed enrichment habitat), running wheel revolutions, and X, Y and Z locations together with beam-breaks. Measurements were acquired at a sample rate of 1 sample/sec for all sensors and cages simultaneously via an error-correcting control area network (CAN). The provision for exercise increased variability in EE and RQ, important goals of our study design, but the primary strategy for achieving that end was to systematically vary the temperature of the testing room from 19 to 29°C. Specifically, the 8 mice were each tested at 19°C (total of 3 d), 21°C (9 d), 23°C (3 d), 25°C (2 d), 27°C (1 d), and 29°C (4 d) for a total of N = 168 measurements. Note that the maintenance of thermal homeostasis in mice requires EE to be exquisitely sensitive to even the seemingly mild cold stress imposed by typical laboratory temperatures of ~21°C while thermoneutrality in mice is achieved at ~30–32°C [1114].

We corrected for water vapor dilution of fractional gas concentrations using Dalton's law of partial pressures, an application of elementary chemistry based on measurement of BP and WVP in the gas stream. The equation is simply:

FDRYgas=FDILgas(BP)(BPWVP) [7]

where FDRYgas and FDILgas are the WVP-corrected and diluted fractional gas concentrations, and WVP is in the same units as BP [35].

The calorimeter system switched from measuring the excurrent air pulled from each cage to measuring incurrent air pulled from the cages’ environment for 30 sec every 20 min. This permitted periodic re-spanning of the O2 analyzers to WVP-corrected FiO2 = 0.2094, effectively eliminating O2 drift; this also allowed measurement of FiCO2. During data analysis these brief and infrequent interruptions–each lasting only ~15% of the cage time constant, thus minimizing their effect on the underlying data -were removed by linear interpolation, which rendered them effectively invisible.

We also contrasted the performance of the RQ- equation with the standard Weir equation using the mouse diet’s FQ of 0.91 as a proxy for RQ, denoted the FQ+ method.

Data reduction and statistical analysis

Data were stored in raw, unprocessed form for later analysis using analysis scripts run on ExpeData analytical software (SSI). This allowed complete and traceable control of the analytical process, the equations used, the baselining algorithms employed, and all other aspects of data transformation and final data extraction.

For the major analyses in this paper, calorimeter outcomes were averaged (mean) across each 24h circadian cycle of each day of exposure at the 6 ambient temperatures. A separate time series analysis binned the 1 Hz continuous data into 30 sec bins across a 24h study to make the data and graphical analyses more tractable to analysis.

Agreement between the RQ- and RQ+ equations was analyzed in several ways. One simply involved analysis of the N = 168 ratios of RQ-:RQ+ EE calculations. We also employed the Tukey mean-difference method (widely known as the Bland-Altman technique [15] because this citation classic [16] is considered the ‘gold standard’ for agreement analysis), and we augmented this by regression analysis. We also present scatterplots involving the methods due to their intuitive accessibility. Owing to the study’s repeated measures design, regression analyses were performed using linear mixed model analysis [17].

All statistical and graphical analyses were performed using R (R: A Language and Environment for Statistical Computing. R Core Team. R Foundation for Statistical Computing. Vienna, Austria, 2018. url = https://www.R-project.org/).

Data are reported as means ± SEM unless reported otherwise.

Results

Agreement analysis

Mean ± SD EE outcomes in kcal/h for the three computational methods were essentially the same: RQ+ = 0.421 ± 0.0651, RQ- = 0.421 ± 0.0652, FQ+ = 0.423 ± 0.0667. The mean ± SD ratios were: (RQ-/RQ+) = 1.0011 ± 0.00042, (FQ+/RQ+) = 1.0044 ± 0.0092. The most extreme ratios were: (RQ-/RQ+) = 1.0021, (FQ+/RQ+) = 1.0265. Accordingly, the largest relative errors compared to the ‘gold standard’ RQ+ method were 0.21% for the RQ- method and 2.65% for the FQ+ method.

Scatterplots relating the alternative EE equations are depicted in Fig 1.

Fig 1. Scatterplots comparing the results of equations that do not use RQ compared to the standard Weir equation that does use RQ.

Fig 1

(A) Substituting the FQ of the mouse chow for RQ in the standard Weir equation (y-axis) versus the standard Weir RQ+ equation. (B) Comparison of the Weir RQ- equation to the standard Weir RQ+ method.

Data consist of 168 measurements of 24h EE in n = 8 mice. EE, energy expenditure; RQ, Respiratory Quotient; FQ, Food Quotient.

Note in Fig 1A that the FQ+ proxy method compared to the standard RQ+ equation exhibits more apparent variability and bias than does the plot of the RQ- versus RQ+ equations depicted in Fig 1B. Regression does not, in general, properly address agreement (explained in (23)). Fig 2 depicts agreement between the FQ+ and RQ+ methods (Fig 2A), while agreement between the RQ- and RQ+ equations is presented in Fig 2B.

Fig 2. Tukey-Bland-Altman mean-difference agreement plots comparing EE computed by alternative methods, and error as a function of RQ.

Fig 2

In panels A and B the x-axis depicts the mean of two methods being compared and the y-axis depicts the difference; the standard Weir RQ+ method was subtracted from the other method (refer to text or Fig 1 for equations). The plots indicate the mean bias (solid horizontal lines) ± 2 SDs (dashed lines) and mean EE (solid vertical lines). (A) Agreement between the Weir EE equation using FQ as a proxy for RQ and the standard Weir RQ+ EE. (B) Agreement between the Weir RQ- equation and the standard Weir RQ+ equation. The superiority of this method over the one assuming that RQ equals FQ is striking. (C) Weir RQ- minus Weir RQ+ EE differences as a function of 1 minus RQ. Note that the predicted mean EE difference between methods at RQ = 0.7 is just 0.0014 kcal/h, which represents 0.31% of the corresponding metabolic rate calculated from the RQ+ method at RQ = 0.7. Data consist of 168 measurements of 24h EE in n = 8 mice. EE, energy expenditure; RQ, Respiratory Quotient; FQ, Food Quotient; ΔEE = Weir FQ+ or Weir RQ- minus standard the Weir RQ+ EE difference.

The performance of the RQ- method was strikingly superior to that of the FQ+ method; indeed, the ± 2 SD range of agreement was 18.8-fold greater for the FQ+ equation than for the RQ- method. This is notable, in part, because the data involve 24h averages for which one expects RQ and FQ to be equivalent in unstressed weight stable animals (as was the case in our study). The increased error reflects the fact that RQ is very labile in fed animals (example below).

The RQ- equation exhibited a very slight positive bias (mean bias = 0.00047 kcal/h, representing just 0.11% of the overall mean of the two methods (and 0.11% of the mean RQ+ EE since the overall means were equal); this simply indicates that, as expected, the RQ- method exhibited a very slight tendency to overestimate EE compared to the RQ+ method. The upper bound for agreement (mean bias + 2 SD) was 0.00088 kcal/h and equates to an error of just 0.21% of the mean. Indeed, of the entire data vector, just 5 of the 168 (RQ- minus RQ+) difference values (3%) exceeded the 0.21% limit. Thus agreement between the RQ- and RQ+ equations was highly satisfactory.

As predicted by Eq 6, Fig 2B depicts a slight positive relationship between the EE difference and corresponding mean EE calculation. Mixed model regression quantified this as 0.0017 ± 0.00023 kcal/h per kcal/h (p<0.0001).

Another view of the impact of RQ on the Weir RQ- minus Weir RQ+ EE difference is depicted in Fig 2C (note rescaling of y-axis); this illustrates that, as predicted, the mean difference between the RQ- and RQ+ methods increases as RQ decreases (see Materials and Methods); however, the practical consequence is trivial as the predicted mean EE difference at RQ = 0.7 is just 0.003 kcal/h, representing <0.02% of the corresponding RQ+ EE calculated for RQ = 0.7 using Eq 1.

The foregoing data indicate that the Weir RQ- method works very well for analyzing EE data averaged over longer time durations, but we also wanted to confirm that this method holds up for continuous time series data in single mice. The data depicted in Fig 3 demonstrate that it does.

Fig 3. Time plots of continuous EE data calculated using the Weir RQ- method and its correspondence to the standard Weir RQ+ method and RQ in a single mouse.

Fig 3

The thick solid lines indicate the dark photoperiod. (A) EE as computed by the RQ- equation (see text or Fig 1). Median, 97.5th and 25th percentiles are indicated. (B) The RQ- minus RQ+ EE difference. Mean ± 2 SD limits are indicated. (C) RQ. Median, 97.5th and 25th percentiles are indicated. Note that the Δ EE trace indicates remarkably little error between the RQ- and RQ+ equations, but that it tends to rise to its most extreme values as RQ decreases. This fits with the predicted behavior of the RQ- equation. EE, energy expenditure; RQ, Respiratory Quotient; Δ = Weir (RQ-) minus standard the Weir (RQ+) EE difference.

Of particular note in Fig 3A, when EE tended to be lowest during the light photoperiods, the RQ- method tended to deviate furthest from the RQ+ method (Fig 3B), and this alignment corresponds to the periods of lowest RQ (Fig 3C). Indeed, the EE difference time series in Fig 3B and the RQ time series in Fig 3C embody symmetrical mirror image-like profiles. This is fully in keeping with the predicted behavior of the RQ+ method as explained above.

We also performed the time series analysis using the standard Weir equation assuming the mouse diet’s FQ of 0.91; the difference between the upper and lower agreement limits was more than 30-fold greater than for the RQ- method (0.0275 vs. 0.0009 kcal/h). This emphasizes the fact that the difference between FQ and RQ in ad-lib fed animals varies markedly across the circadian cycle and so reduces the accuracy of EE calculations when using FQ in place of measured RQ.

It is important to emphasize the importance of correcting fractional gas concentrations for WV dilution. Fig 4 demonstrates that failure to correct for WV artifactully increases EE by a large amount. Failure to correct for WVP will also badly distort RQ calculations, and constitutes one reason that RQ can differ from FQ. In the present work, mean 24h RQ calculated after correction for WVP was 0.895 ± 0.0007 and did not differ from the calculated FQ of 0.906 (p = 0.15). Failure to correct for WVP resulted in a mean RQ of 0.43 ± 0.004. We believe that correcting for WVP is ideally done using Dalton’s law of partial pressures because physically or chemically removing WV from airstreams involves considerable potential for error and increases calorimeter response times [3] (discussed below).

Fig 4. Effect of not correcting for water vapor dilution on mean EE and agreement between the RQ- and RQ+ methods for calculating EE.

Fig 4

Data consist of 168 measurements of 24h EE in n = 8 mice.

Accuracy of RQ- method depends on the incurrent oxygen fraction

Eq 5 immediately reveals two additional insights: The first is that if FiO2 equals the coefficient of the RQ term in the numerator of the ratio in Eq 5 (i.e., 0.2191), then that ratio equals exactly 1.0 across the entire range of RQ because the numerator and denominator will be identical linear equations. Therefore, manipulating FiO2 can, in theory, result in perfect agreement between the RQ- and RQ+ equations. The second insight, a corollary of the first, is that if FiO2 deviates substantially from normoxia, then Eq 5 will agree less well with the standard method. It should be noted in this context that FiO2 can be generated at a ‘customized’ value by blending gases with precision flow controllers, e.g., as in [1826].

Discussion

Agreement

The present work demonstrates clearly, and to our knowledge for the first time, that the ‘RQ-free’ method published seven decades ago by Weir [6] calculates EE with almost no error compared to Weir’s standard method that uses both VO2 and RQ. Indeed, our finding that the maximum observed relative error of the RQ- method was just 0.21% may seem difficult to reconcile with the fact that RQ is universally acknowledged as an important variable for transforming VO2 into EE (e.g., [2, 3, 6, 7]). Our data by no means challenge that view, rather the reconciliation, as mathematically demonstrated in the Materials and Methods, is that both VO2 and the EE transform that constitute Eq 1 contain functions of RQ that very nearly cancel each other out. This serendipitous property of aerobic biomathematics means that the product of VO2 and the transform equation can be simplified into an expression that excludes any explicit dependence on either RQ or VO2.

Because the RQ- approach constitutes a method rather than a single equation, we should stress how easy it is to adapt it for use with a different RQ+ equation, for instance the equation derived by Hall and associates [7]. Adjusted for a typical contribution of dietary protein to human metabolic rate (15% of EE) the Hall equation is:

EE=VO2(3.85+1.07RQ) [8]

Note that Eq 8 indicates that 1 L of O2 corresponds to 4.92 kcal when RQ = 1.0; therefore, this is the constant for the RQ- equivalent:

EE=4.92×V˙e×ΔFO2 [9]

When Eqs 5 or 9 are derived to use the incurrent flow rate Vi instead of Ve, the result is simply to replace Vi for Ve.

Adjusting respirometry for protein metabolism

Protein oxidation has been widely ignored in research involving respirometric EE estimation. Depending on study design and goals, the potential impact of protein metabolism may or may not be significantly problematic. To illustrate, we note that the predicted ‘error’ in EE calculated from the RQ+ or RQ- methods without adjustment for the high dietary protein percentage of our study diet would be a non-trivial 2.3%. Accordingly, it would seem problematic to ignore protein when comparing EE in groups fed diets that differ markedly in protein content, or in studies involving interventions that might alter protein metabolism.

One reason that protein metabolism has been ignored is the assumed need to measure nitrogen excretion (6.25 g of protein metabolized per g of excreted nitrogen [7]). This requires a specialized metabolic cage to collect urine for nitrogen determination. By contrast, the simple approach of estimating protein metabolism from diet composition has been implemented in excellent work involving the measurement and mathematical modeling of EE [7, 2729], and we note that protein is taken into account in calculating FQ, which, in turn, is expected to agree with RQ, as it did in the present study. The point is that investigators should consider adjusting respirometric EE calculations for protein metabolism based on diet composition.

Advantages of the RQ- method

Our findings do more than simply validate the use of the RQ- method for research involving the need to compute only EE. In particular, measuring VCO2 is technically problematic and adds cost. CO2 calibration gases of <1% accuracy are not widely available and most are accurate to only ±2–5%. CO2 is generally measured using an optical absorption method with non-linear properties [3], and CO2 analyzers vary widely in the extent to which they successfully compensate for this. Some investigators use multiple span gases to address residual non-linearity, but this introduces further calibration uncertainty because each span gas has an independent and unknown error percentage. Operational issues also arise: a researcher might employ an unreliable or poorly calibrated CO2 analyzer, or the tank of CO2 span gas may prove to be empty immediately prior to an important experiment.

By contrast, it is considerably easier to accurately measure fractional O2 concentrations within a range narrowly centered on the normal atmospheric value. One reason whose importance is hard to overstate is that O2 calibration is anchored to the atmosphere’s near-constant FO2 of 20.939 ± 0.0003% after correcting for WVP and variations in BP [10]. Another crucial advantage is that the O2 calibration curve is linear. Finally, modern high precision O2 sensors are exquisitely sensitive to fractional O2 concentrations.

Our results and comments regarding CO2 should make it obvious that the RQ- method provides an excellent means by which to rescue high accuracy EE data in studies compromised by faulty CO2 sensing. Moreover, the RQ- method may also be developed to provide a way to determine whether CO2 sensing is potentially compromised. Specifically, if CO2 sensing is accurate, then mean EE as calculated by the RQ- method will be only very slightly higher than EE calculated by the RQ+ method (0.11% in the present study), whereas if, for example, CO2 sensing is too low and therefore results in artifactually low RQ values, then mean EE as calculated by the RQ- method may be notably higher than EE calculated by the RQ+ method. To take advantage of this quality control approach, it would be important to first determine typical RQ- to RQ+ EE ratios in studies where the gas sensors are known to be functioning well.

Another potentially important application of the RQ- approach is that eliminating the CO2 sensor and related components would help minimize the weight, volume and expense of small ‘wearable’ calorimeters designed to measure human EE during occupational, recreational and ‘everyday living’ tasks.

Finally, our study also lends credence to the use of the Weir RQ- method in previous research involving measurement of EE during exposure to nitrous oxide, a gas that interferes with CO2 sensing [1826].

Supporting information

S1 Dataset. 24h_mean_EE.xlsx.

(XLSX)

S2 Dataset. Mouse6.

(XLSX)

Acknowledgments

We thank Thomas Foerster PhD for assistance with carrying out the metabolic measurements at UNLV. We thank Douglas S. Ramsay, DMD, PhD, Katherine Rafferty, PhD, Stephan Guyenet, PhD, and Stephen C. Woods for their very helpful comments during internal review.

Data Availability

All relevant data are within the manuscript and its Supporting Information files.

Funding Statement

KJK received salary support from the National Mouse Metabolic Phenotyping Centers (Working Group Award by NIH-NIDDK U24DK076169). The funder had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific role of this author is articulated in the ‘author contributions’ section. BEW received support from the National Institutes of Health (DK074758, DK090320, and DK076126). Sable Systems International Inc., provided support in the form of salaries for author JRBL, but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific role of this author is articulated in the ‘author contributions’ section.

References

  • 1.Kleiber M. The fire of life—an introduction to animal energetics. Robert E. Krieger Publishing Co., Inc., NY, NY: 1975. [Google Scholar]
  • 2.McLean JA, Tobin G. Animal and human calorimetry Cambridge Cambridgeshire; New York: Cambridge University Press; 1987. xiii, 338 p. p. [Google Scholar]
  • 3.Lighton JRB. Measuring metabolic rates: a manual for scientists Oxford; New York: Oxford University Press; 2008. xiii, 201 p. p. [Google Scholar]
  • 4.Lighton JR, Halsey LG. Flow-through respirometry applied to chamber systems: pros and cons, hints and tips. Comparative biochemistry and physiology. 2011. March;158(3):265–75. 10.1016/j.cbpa.2010.11.026 . Epub 2010/12/08. eng. [DOI] [PubMed] [Google Scholar]
  • 5.Melanson EL, Ingebrigtsen JP, Bergouignan A, Ohkawara K, Kohrt WM, Lighton JR. A new approach for flow-through respirometry measurements in humans. Am J Physiol Regul Integr Comp Physiol. 2010. June;298(6):R1571–9. 10.1152/ajpregu.00055.2010 . Pubmed Central PMCID: 2886696. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Weir JB. New methods for calculating metabolic rate with special reference to protein metabolism. J Physiol. 1949. August;109(1–2):1–9. PubMed . eng. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Hall KD, Chen KY, Guo J, Lam YY, Leibel RL, Mayer LE, et al. Energy expenditure and body composition changes after an isocaloric ketogenic diet in overweight and obese men. The American journal of clinical nutrition. 2016. August;104(2):324–33. 10.3945/ajcn.116.133561 . Pubmed Central PMCID: 4962163. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Koteja P. Measuring energy metabolism with open-flow respirometric systems: which design to choose? Functional Ecology. 1996;10:675–7. [Google Scholar]
  • 9.Arch JR, Hislop D, Wang SJ, Speakman JR. Some mathematical and technical issues in the measurement and interpretation of open-circuit indirect calorimetry in small animals. International journal of obesity (2005). 2006. September;30(9):1322–31. 10.1038/sj.ijo.0803280 . eng. [DOI] [PubMed] [Google Scholar]
  • 10.Tohjima Y, Machida T, Watai T, Akama I, Amari T, Moriwaki Y. Preparation of gravimetric standards for measurements of atmospheric oxygen and reevaluation of atmospheric oxygen concentration. J Geophys Res. 2005;110:D11302. [Google Scholar]
  • 11.Abreu-Vieira G, Xiao C, Gavrilova O, Reitman ML. Integration of body temperature into the analysis of energy expenditure in the mouse. Molecular metabolism. 2015. June;4(6):461–70. 10.1016/j.molmet.2015.03.001 . Pubmed Central PMCID: 4443293. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Deem JD, Muta K, Ogimoto K, Nelson JT, Velasco KR, Kaiyala KJ, et al. Leptin Regulation of Core Body Temperature Involves Mechanisms Independent of the Thyroid Axis. Am J Physiol Endocrinol Metab. 2018. June 26 10.1152/ajpendo.00462.2017 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Kaiyala KJ, Ogimoto K, Nelson JT, Muta K, Morton GJ. Physiological role for leptin in the control of thermal conductance. Molecular metabolism. 2016. October;5(10):892–902. 10.1016/j.molmet.2016.07.005 . Pubmed Central PMCID: 5034509. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Gordon CJ. The mouse thermoregulatory system: Its impact on translating biomedical data to humans. Physiol Behav. 2017. October 1;179:55–66. 10.1016/j.physbeh.2017.05.026 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986. February 8;1(8476):307–10. PubMed . eng. [PubMed] [Google Scholar]
  • 16.Bland JM, Altman DG. Agreed statistics: measurement method comparison. Anesthesiology. 2012. January;116(1):182–5. 10.1097/ALN.0b013e31823d7784 . [DOI] [PubMed] [Google Scholar]
  • 17.Fitzmaurice GM, Laird NM, Ware JH. Applied longitudinal analysis 2nd ed Hoboken, N.J.: Wiley; 2011. xxv, 701 p. p. [Google Scholar]
  • 18.Kaiyala KJ, Ramsay DS. Concentration-related metabolic rate and behavioral thermoregulatory adaptations to serial administrations of nitrous oxide in rats. PLoS ONE. 2018;13(4):e0194794 10.1371/journal.pone.0194794 . Pubmed Central PMCID: 5909668. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Ramsay DS, Al-Noori S, Shao J, Leroux BG, Woods SC, Kaiyala KJ. Predicting Addictive Vulnerability: Individual Differences in Initial Responding to a Drug's Pharmacological Effects. PLoS ONE. 2015;10(4):e0124740 10.1371/journal.pone.0124740 . Pubmed Central PMCID: 4400068. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Ramsay DS, Woods SC, Kaiyala KJ. Repeated nitrous oxide exposure in rats causes a thermoregulatory sign-reversal with concurrent activation of opposing thermoregulatory effectors. Temperature. 2014. Oct-Dec;1(3):257–67. 10.4161/23328940.2014.944809 . Pubmed Central PMCID: 4414259. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Ramsay DS, Kaiyala KJ, Woods SC. Correctly identifying responses is critical for understanding homeostatic and allostatic regulation. Temperature. 2014;1(3):157–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Kaiyala KJ, Woods SC, Ramsay DS. Persistence of a hyperthermic sign-reversal during nitrous oxide inhalation despite cue-exposure treatment with and without a drug-onset cue. Temperature. 2014. Oct-Dec;1(3):268–75. 10.4161/23328940.2014.944811 . Pubmed Central PMCID: 4416485. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Kaiyala KJ, Chan B, Ramsay DS. Robust thermoregulatory overcompensation, rather than tolerance, develops with serial administrations of 70% nitrous oxide to rats. J Therm Biol. 2012. January 1;37(1):30–40. 10.1016/j.jtherbio.2011.10.004 . Pubmed Central PMCID: 3255088. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Kaiyala KJ, Butt S, Ramsay DS. Systems-level adaptations explain chronic tolerance development to nitrous oxide hypothermia in young and mature rats. Psychopharmacology (Berl). 2007. January 10;191:233–42. 10.1007/s00213-006-0655-1 . Eng. [DOI] [PubMed] [Google Scholar]
  • 25.Kaiyala KJ, Butt S, Ramsay DS. Direct evidence for systems-level modulation of initial drug (in)sensitivity in rats. Psychopharmacology (Berl). 2007. January 20;191:243–51. 10.1007/s00213-006-0657-z . Eng. [DOI] [PubMed] [Google Scholar]
  • 26.Kaiyala KJ, Ramsay DS. Assessment of heat production, heat loss, and core temperature during nitrous oxide exposure: a new paradigm for studying drug effects and opponent responses. Am J Physiol Regul Integr Comp Physiol. 2005. March;288(3):R692–701. 10.1152/ajpregu.00412.2004 . [DOI] [PubMed] [Google Scholar]
  • 27.Hall KD, Bemis T, Brychta R, Chen KY, Courville A, Crayner EJ, et al. Calorie for Calorie, Dietary Fat Restriction Results in More Body Fat Loss than Carbohydrate Restriction in People with Obesity. Cell metabolism. 2015. September 1;22(3):427–36. 10.1016/j.cmet.2015.07.021 . Pubmed Central PMCID: 4603544. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Guo J, Hall KD. Predicting changes of body weight, body fat, energy expenditure and metabolic fuel selection in C57BL/6 mice. PLoS ONE. 2011;6(1):e15961 10.1371/journal.pone.0015961 . Pubmed Central PMCID: 3016341. Epub 2011/01/20. eng. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Guo J, Hall KD. Estimating the continuous-time dynamics of energy and fat metabolism in mice. PLoS Comput Biol. 2009. September;5(9):e1000511 10.1371/journal.pcbi.1000511 . Pubmed Central PMCID: 2731929. Epub 2009/09/19. eng. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

S1 Dataset. 24h_mean_EE.xlsx.

(XLSX)

S2 Dataset. Mouse6.

(XLSX)

Data Availability Statement

All relevant data are within the manuscript and its Supporting Information files.


Articles from PLoS ONE are provided here courtesy of PLOS

RESOURCES