Abstract
Introduction
Yttrium‐90 (90Y) microsphere post‐treatment imaging reflects the true distribution characteristics of microspheres in the tumor and liver compartments. However, due to its decay spectra profile lacking a pronounced photopeak, the bremsstrahlung imaging for 90Y has inherent limitations. The absorbed dose calculations for 90Y microspheres radiomicrosphere therapy (RMT) sustain a limitation due to the poor quality of 90Y imaging. The aim of this study was to develop quantitative methods to improve the post‐treatment 90Y bremsstrahlung single photon emission tomography (SPECT)/computed tomography (CT) image analysis for dosimetric purposes and to perform a quantitative comparison with the 99mTc‐MAA SPECT/CT images, which is used for theranostics purposes for liver and tumor dosimetry.
Methods
Pre and post‐treatment SPECT/CT data of patients who underwent RMT for primary or metastatic liver cancer were acquired. A Jasczak phantom with eight spherical inserts of various sizes was used to obtain optimal iteration number for the contrast recovery algorithm for improving 90Y bremsstrahlung SPECT/CT images. Comparison of uptake on 99mTc‐MAA and 90Y microsphere SPECT/CT images was assessed using tumor to healthy liver ratios (TLRs). The voxel dosimetry technique was used to estimate absorbed doses. Absorbed doses within the tumor and healthy part of the liver were also investigated for correlation with administered activity.
Results
Improvement in CNR and contrast recovery coefficients on patient and phantom 90Y bremsstrahlung SPECT/CT images respectively were achieved. The 99mTc‐MAA and 90Y microspheres SPECT/CT images showed significant uptake correlation (r = 0.9, P = 0.05) with mean TLR of 9.4 ± 9.2 and 5.0 ± 2.2, respectively. The correlation between the administered activity and tumor absorbed dose was weak (r = 0.5, P > 0.05), however, healthy liver absorbed dose increased with administered activity (r = 0.8, P = 0.0).
Conclusions
This study demonstrated correlation in mean TLR between 99mTc‐MAA and 90Y microsphere SPECT/CT.
Keywords: 90Y, 99mTc, dosimetry, radiomicrosphere therapy, SPECT/CT
1. INTRODUCTION
Radiomicrosphere therapy (RMT) using Yttrium‐90 (90Y) microspheres is an effective liver‐directed therapy in the management of primary and metastatic liver cancer.1, 2, 3 The biodistribution of 90Y microspheres after treatment is assessed through 90Y bremsstrahlung imaging, preferably SPECT/CT.4, 5, 6 A single time point imaging is used due to the permanent implant of the microspheres where there is no elimination, redistribution and washout phases of the radiopharmaceutical.7
Technetium‐99m macroaggregated albumin (99mTc‐MAA) scanning is performed prior to RMT as a surrogate for 90Y microspheres. In RMT planning, 99mTc‐MAA planar and single photon emission tomography (SPECT)/computed tomography (CT) imaging is used to measure the percentage of particles that shunt to the lungs (lung shunt fraction, LSF), to assess any extrahepatic particle deposition, and to calculate the tumor to normal liver ratio (TLR).1, 8 Various authors have investigated the correlation of uptake and distribution between 99mTc‐MAA and 90Y microspheres SPECT/CT.8, 9 The authors used tumor to normal liver ratio to evaluate the uptake on 99mTc‐MAA and 90Y microsphere SPECT/CT images.
90Y microspheres bremsstrahlung SPECT/CT imaging could potentially identify an extrahepatic uptake. Early detection of such an uptake, thus, could initiate preventative measures early on. The tumor and liver dose estimates obtained from 90Y microspheres bremsstrahlung SPECT/CT imaging could be correlated with tumor response and liver toxicity clinical data.10 The major problem in 90Y bremsstrahlung imaging is the lack of a pronounced photopeak energy due to the continuous and broad energy spectrum of bremsstrahlung photons giving it a poor image quality. As a result, various studies based on phantom and Monte Carlo (MC) simulation have recommended appropriate energy windows in accordance with the collimator used.11, 12, 13, 14 It has been shown, often in conjunction with phantom studies that the incorporation of MC simulations to clinical images gives an optimal accuracy of bremsstrahlung images by compensating for photons through correction for attenuation, scatter, and collimator‐detector response.13, 14, 15 Alternately, studies have demonstrated the feasibility of 90Y PET/CT imaging excelling in contrast and resolution compared to bremsstrahlung SPECT/CT.16, 17 A recent study by Yue et al.17 concluded that both modalities are comparable for post‐RMT dosimetry estimation if appropriate reconstruction for 90Y bremsstrahlung SPECT/CT is applied. However, an appropriate reconstruction method requires incorporation of MC simulation in 90Y bremsstrahlung SPECT/CT for quantitative improvement, but is not easily applied for clinical implementation. Another recent study by Siman et al.18 developed a practical imaging protocol called background compensation for 90Y bremsstrahlung SPECT/CT imaging as an alternative to the MC method. The authors assert that their method doesn't address the main image degrading factors in 90Y bremsstrahlung SPECT/CT imaging, such as object scatter, septal penetration, and backscatter; nonetheless, they reported an improvement in recovery coefficient from 39% to 90% in a 37 mm sphere in a 10 mm volume of interest.
In this study, our aim was to apply post‐reconstruction techniques to improve the quantitative quality of 90Y bremsstrahlung SPECT/CT images and evaluate the correlation of the uptake pattern on the 90Y microspheres SPECT/CT post‐therapy images vs pretreatment 99mTc‐MAA SPECT/CT images. We developed an iterative deconvolution based algorithm to improve the quantitative quality of 90Y bremsstrahlung SPECT images. This work provides a quantitative comparison of the uptake distribution in the 99mTc‐MAA and 90Y microsphere SPECT images to increase the accuracy of the comparison. Tumor segmentation based on SPECT images is feasible in bremsstrahlung images since the deconvolution correction method results in images with more concentrated uptake in areas where the tumor is (excluding necrotic tumor) vs a dispersed or partially uniform distributions observed in the uncorrected 90Y bremsstrahlung SPECT images. To our knowledge, no work to date, has performed quantitative improvement on 90Y bremsstrahlung SPECT images prior to comparison with 99mTc‐MAA SPECT/CT images. Furthermore, we analyzed the association between administered activity and liver dosimetry in patient study.
2. MATERIALS AND METHODS
2.A. SPECT/CT image acquisition and reconstruction
All data were acquired on a dual‐head Philips Precedence 16P SPECT/CT (Philips Medical Systems Inc., USA).
2.A.1. Patient study
Patient data were acquired as part of a retrospective study approved by the Florida International University Institutional Review Board. Therefore, written patient informed consent was not sought nor documented and image data were handled anonymously. In total, SPECT/CT data of 12 patients who underwent RMT with 90Y‐labelled resin microspheres (SIR‐spheres; Sirtex) were acquired. All patients had undergone 99mTc‐MAA SPECT/CT scan prior to RMT. Mean administered 99mTc‐MAA was 190 MBq. Data were collected for patients whose LSF ≤ 5% to minimize error due to extrahepatic deposition consistent with another group.18 The acquisition energy window was centered at 140 keV ± 15% for 99mTc‐MAA SPECT with 128 × 128 projection matrix over 360°, for 20 s per azimuth on a low‐energy high‐resolution collimator. Image reconstruction was done using the ordered subset expectation maximization (OSEM) algorithm with three iterations and 16 subsets, 132 × 132 matrix and 4.664 × 4.664 × 4.664 mm3 voxel size. 90Y bremsstrahlung SPECT/CT was performed with a medium energy general purpose collimator with energy window centered on 90 keV ± 15%, 35 s per azimuth for 2 × 64 views over 360° (currently a standard protocol at our center). Image reconstruction for 90Y bremsstrahlung SPECT was performed using a built‐in three‐dimensional (3D) OSEM algorithm. 90Y bremsstrahlung SPECT images were reconstructed with four iterations and eight subsets, 132 × 132 matrix and 4.664 × 4.664 × 4.664 mm3 voxel size.
2.A.2. Phantom study
The Jasczak phantom with eight fillable spherical inserts of inner diameter 2, 8, 10, 12, 16, 25, 31, and 34 mm was used. The spheres (multiple sizes, see Table 1) and background (~6 L volume) were filled with Yttrium‐90 (III) chloride solution (PerkinElmer Inc., Waltham, MA, USA) of activity concentrations 0.52 and 0.04 MBq/ml respectively, with an approximate sphere to background activity concentration ratio of 13:1. Total activity was 27 and 255 MBq in the spheres and background, respectively. Table 1 shows activities within the different sphere sizes. 90Y activity was diluted and measured in a 60 ml vial before adding to the spheres. The activity inside each sphere was measured using a Capintec dose calibrator (read out scale factor = 10). Imaging was set identical to patient studies in terms of imaging window, collimator, and image reconstruction.
Table 1.
Activities with the hollow spheres
| Sphere no | Sphere size in diameter (mm) | Activity (MBq) |
|---|---|---|
| S1 | 34 | 11.10 |
| S2 | 31 | 8.88 |
| S3 | 25 | 4.44 |
| S4 | 16 | 1.11 |
| S6 | 12 | 0.56 |
| S7 | 10 | 0.28 |
| S8 | 8 | 0.14 |
| S9 | 2 | ~0 |
2.A.3. Image processing
A MATLAB® algorithm was developed to import and export images for Volumes of interest (VOIs) generation, semi‐automatic tumor segmentation, activity estimation, absorbed dose estimation, and statistical and mathematical analysis. For the patient studies, VOIs were drawn manually on the CT image slices. Binary masks from the CT images were mapped onto the respective SPECT scans. For the phantom study, eight circular VOIs were manually drawn on the CT slices, where knowledge of the phantom composition allowed us to identify the spheres and make sure the volumes in VOIs were consistent with the true measured sphere volumes. Background VOIs consisted of all voxels within the phantom boundary excluding voxels that belong to the spheres’ VOIs.
2.B. Contrast recovery
We propose a new method based on an iterative expectation maximum likelihood deconvolution technique to improve the quantitative quality of the 90Y bremsstrahlung SPECT images. The Richardson–Lucy (RL) technique was used for post‐reconstruction image deconvolution implemented with MATLAB®. The method is an iterative expectation maximum likelihood deconvolution algorithm [Eq. (1)] where images degraded by the point spread function (PSF) of the detector and additive noises are corrected.19, 20 The choice of a maximum likelihood algorithm has the benefit of producing good quality images in the presence of high noise levels by preserving positive values through accounting for fluctuation in the signal and thus limiting noise amplification.20
| (1) |
A is the reconstructed image degraded by the response of the detector, A’ is the contrast recovered image where = A, ⊗ is the 3D (x, y, z) convolution operation and i is the iteration number. In our study the PSF was fixed and the only iterative maximum likelihood estimate was the image. The PSF of the collimator‐detector response was modeled by the Gaussian function according to Eq. (2) where σ was found from the relationship with full‐width at half‐maximum (FWHM) of the detector, × σ.
| (2) |
Improvements in the quantitative quality 90Y bremsstrahlung SPECT/CT images were evaluated using contrast to noise ratio (CNR) and contrast recovery coefficients (QH)16 for the patient and phantom studies respectively as given by Eqs. (3)) and (4). Our analysis of quantitative image improvement is based on matched VOIs between the CT and SPECT images as this is required for dosimetry estimation.
| (3) |
| (4) |
M T is the mean count in tumor VOIs, M B is the mean count in healthy liver VOIs, C S is the mean count in the sphere VOIs, C B is the mean count in the background VOIs and R is the true sphere to background ratio. The block diagram of the employed RL algorithm is shown in Fig. 1.
Figure 1.

Block diagram of the Richardson–Lucy algorithm employed.
The iteration number for the algorithm was chosen at the point of maximum likelihood where the contrast recovery coefficient for the 34 mm sphere was at its maximum value, which corresponded to a decline in the associated root mean square error (RMSE) between two consecutive iterative image estimates, which was found to be at the sixth iteration as shown in Fig. 2. The contrast improvement algorithm using Richardson‐Lucy deconvolution is a new idea integrated in this approach.
Figure 2.

(a) Phantom computed tomography (CT) scan and single photon emission tomography (SPECT) images for iteration numbers of 0, 6, and 15 left to right. (b) Plot of QH (right axis) and root mean square error (RMSE) (left axis) vs iteration number for the 34 mm sphere
2.C. SPECT calibration factor measurement
The calibration factor (CF) was defined as the ratio of the total reconstructed counts to the true activity. Due to the continuous nature of the bremsstrahlung x‐ray spectrum that results in the absence of a photopeak window, a CF calculated using a point source might not quite represent photon counts in the clinical setup. As a result, mostly phantom studies are used for estimating a CF in 90Y bremsstrahlung SPECT/CT imaging. We used two methods for calculating the CF from the 90Y bremsstrahlung SPECT/CT images of the 12 patients. The first approach (method 1) used the entire total reconstructed counts within the field of view (FOV) of the SPECT detector. The second approach (method 2) used counts from the liver segments only. In method 2 the liver was first segmented as described in the image processing section and total counts within the liver VOIs were taken. This method avoids counts due to image artifacts. The motivation for using directly the patients’ liver uptake to calculate the CF is that due to the proximity of the treatment to the target area where the 90Y microspheres are deposited predominantly in the liver, the 90Y activity can be determined with high accuracy.21 Calibration curves were generated using linear regression analysis to derive the relationship between patient administered activity and reconstructed counts. The CFs were calculated from the slopes of the calibration curves.
For the phantom study, the total reconstructed counts within the phantom boundary were considered. The total activity was the sum of the activities inside the spheres and the background. The CF from the phantom study was evaluated to be within the 95% confidence interval (CI) of the calibration curves from the 90Y microsphere patient studies to validate its application on the phantom images.
2.D. Activity estimation and comparison with the administered dose
The 90Y bremsstrahlung SPECT reconstructed pixel values were divided by the CF to convert values in counts per second (cps) into units of activity (MBq). For the patient studies, total liver activity estimation was compared to the administered activity. Activities inside the eight spherical inserts were compared to the activities measured using the dose calibrator for the phantom study. Percent error was calculated as % Error = (True Activity − Estimated Activity)/True Activity × 100%, where true activity is the administered activity. To compare the correlation between uptake distribution on 99mTc‐MAA and 90Y microsphere SPECT/CT images, the two images were co‐registered.
2.E. Co‐registration of 99mTc‐MAA and 90Y microspheres SPECT/CT
Since the SPECT images were separately registered to the CT images we tried three approaches. The first method was taking single CT images as a reference (“fixed image”) and the SPECT images as a source (“moving image”). The second method is CT‐to‐CT registration. In the third method, 90Y microsphere SPECT was taken as the reference image and the 99mTc‐MAA SPECT was the source image (SPECT‐to‐SPECT). Two image registration tools were used, FMRIB's Software Library (FSL) and Statistical Parametric Mapping (SPM), both well‐established tools extensively used for registering anatomical and functional images, and we found visually the best results using SPM.22 We used the spatial normalization function of SPM12 using the standard settings of the toolbox. In order to choose the best registration approach, we calculated the mutual information between the co‐registered 90Y microsphere and 99mTc‐MAA SPECT images from the methods. The mutual information between the two images is defined as:
| (5) |
H (A) and H (B) are the entropy of image A and B, respectively, and H (A, B) is their joint entropy. The method which gave the highest I (A, B) was used for registering the images. Details of the calculation of mutual information is explained by Maes et al.23 Mean CPU time to register two images was about 15 min on a standard PC.
2.F. Tumor segmentation and tumor to liver ratio calculation (TLR)
Quantitative uptake analysis on 99mTc‐MAA SPECT/CT for tumor delineation is usually done using an isocontour method with a dedicated software.8 In our study tumor segmentation based on the uptake of 99mTc‐MAA and 90Y microspheres was performed with an active contour segmentation method using the well‐known Chan‐Vese energy paradigm24 to delineate areas of high activity (tumor) from the surrounding low activity (healthy) regions. This method doesn't use a single threshold, which is an advantage in this case since the administered activities of 99mTc‐MAA and 90Y microspheres are quite different. The method uses gradient based information between neighboring voxels in order to segment the higher uptake regions. The global gradient calculation is specific to the image, that is, the 99mTc‐MAA and 90Y microspheres SPECT images are treated separately but in the same slice range. Due to the proximity of the treatment, higher regions of uptake are taken as tumor. The segmented tumor was subtracted from the whole liver volume to get the healthy part. Since counts are proportional to activity concentration, the TLRs were calculated by dividing the mean counts per pixel of the tumor by the mean counts per pixel of the healthy liver.
2.G. Absorbed dose estimation
The voxel S‐value method was used to estimate 3D radiation absorbed dose in 90Y bremsstrahlung SPECT/CT images.25, 26 Cubical voxel S values for the pixel size of 4.664 mm were determined using linear interpolation from 3 and 6 mm pixel sizes calculated by Franquiz et al. and Bolch et al.25, 26 It is expected that errors are associated with using a linear interpolation for estimating S‐values for a specific cubic voxel size from tabulated results.27 For our retrospective study this error is acceptable as explained in MIRD pamphlet no 17.26 Absorbed dose per voxel was estimated by convolution of the activity images with the corresponding voxel S values [Eq. (5)], based on the MIRD Pamphlet No. 17.26
| (6) |
DT is the absorbed dose at the target voxel (mGy), A (MBq) is the cumulated activity from the surrounding source voxels, ⊗ is the 3D convolution and S is the voxel S‐value (mGy/MBq) for each associated source distance to the target voxel. Cumulative dose‐volume histograms (cDVHs) and isodose curves were generated for the tumor and healthy liver VOIs from the SPECT dose map images.
2.H. Statistical analysis
Quantitative parameters are presented as mean ± SD and ranges. Linear regressions were generated between administered activities (independent variable) and cps for the purpose of predicting CFs. We report slope, R 2, standard error and 95% CI of the regression models. Pearson correlation coefficient and its P value were used to test for significance of correlations between TLRs from 99mTc‐MAA and 90Y microsphere SPECT/CT as well as between administered activity and absorbed doses. Statistical analyses were deemed significant as having a P < 0.05. All statistical analyses were performed with Minitab® software package (version 17).
3. RESULTS
3.A. Contrast recovery
Figure 3(a) shows examples of improvements in the CNRs between tumor and the background for patient 90Y bremsstrahlung SPECT/CT images. For the patients shown (Pat 3 and 9), the CNRs were respectively 18.9 and 48.8 before and 24.1 and 51.7 after contrast recovery. For the phantom study improvement in QH ranged between −8.3 to 41.0%. For the smallest spheres (2, 8, and 10 mm), no improvement in QH was found. Figure 3(b) shows line profiles along the 16 and 12 mm spheres. From the profiles, it can be seen that the signal in the 16 mm sphere significantly differs from the background while the signal within the 12 mm sphere and the background are indistinguishable.
Figure 3.

(a) Result of contrast recovery showing the before and after 90Y bremsstrahlung single photon emission tomography (SPECT) images of patients with the respective computed tomography (CT) scans. (b) Line profiles through the 16 mm (lower) and 12 mm (upper right) spheres of the phantom 90Y bremsstrahlung SPECT/CT image. Phantom SPECT images were resized to 512 × 512 for display purpose hence higher pixel position for the line profiles
3.B. SPECT CF
Results of the regression analyses of the relationships between cps and administered activity gave the following results, the slopes being the CFs (cps/MBq); method 1, CF1 = 19 990 cps/MBq with 95% CI 14 081 ≤ β ≤ 25 899 (R 2 = 0.85, standard error (SE) = 2652) and method 2, CF2 = 8316 cps/MBq with 95% CI of 6810 ≤ β ≤ 9822 (R 2 = 0.94, SE = 676). The CF from the phantom study was 9049 cps/MBq, which falls within the 95% CI of CF2.
3.C. Activity estimation and comparison with the administered dose
The total activity inside the liver was estimated for each patient using the two CFs. For CF1, total liver activity estimation resulted in mean percent error of 59 ± 5%. Applying CF2 gave the smallest error (−5 ± 13%), thus it was used for subsequent analysis. Table 2 shows results of total activity estimation within the liver VOIs using CF2. For most of the patients the estimated total liver activity percent errors were within ±10% giving an overall satisfactory results. As our study is retrospective based on anonymized data, we couldn't provide possible clinical reasons for the larger deviations of the estimated activities in some of the patients. Results of activity estimates inside the spheres and the background for the phantom study gave total mean percent error of −23 ± 41%.
Table 2.
Administered activities and total estimated activities inside liver VOI
| Patient no. | Administered activity (MBq) | Estimated activity (MBq) | Error (%) |
|---|---|---|---|
| Pat_1 | 555 | 480 | 14 |
| Pat_2 | 570 | 652 | −14 |
| Pat_3 | 759 | 967 | −27 |
| Pat_4 | 781 | 788 | −1 |
| Pat_5 | 966 | 1251 | −30 |
| Pat_6 | 999 | 1009 | −1 |
| Pat_7 | 1203 | 1202 | 0 |
| Pat_8 | 1236 | 1221 | 1 |
| Pat_9 | 1262 | 1288 | −2 |
| Pat_10 | 1436 | 1350 | 6 |
| Pat_11 | 1517 | 1572 | −4 |
| Pat_12 | 2072 | 2188 | −6 |
| Mean | 1113 | 1164 | −5 |
| SD | 438 | 448 | 13 |
3.D. Co‐registration of 99mTc‐MAA and 90Y microspheres SPECT/CT
Figure 4 shows results of the three co‐registration approaches where the CT scan of the post‐treatment as a reference image gave the best alignment between the images. Using the 90Y microsphere SPECT image as a reference also gave a good result visually comparable to the aforementioned result [Fig. 4(c)]. But the mutual information is higher or equal in most patients where the CT scan was the reference image [Fig. 5(a)]. In other patients, for example, patient 7, the 99mTc‐MAA and 90Y microspheres SPECT images didn't co‐register correctly with the CT image. From the SPECT scans of this patient, we observed that the patient had a hepatic tumor with a necrotic core with minimal uptake inside the liver which reduced the mutual information required for co‐registering the SPECT and CT scans [Fig. 5(b)]. Patient 4 showed the smallest of all mutual information between 99mTc‐MAA and 90Y microsphere SPECT scans and similar to the previously mentioned patient, the SPECT scan didn't co‐register well with the CT scan. We observed that the 99mTc‐MAA and 90Y microsphere SPECT scans of this patient have exclusively localized activity in the left lobe. For these patients image analyses on the 99mTc‐MAA and 90Y microsphere SPECT/CT were performed separately.
Figure 4.

Patient 6 (left) and 8 (right) co‐registered computed tomography (CT) and single photon emission tomography (SPECT) images of 99mTc‐MAA (blue) and 90Y microsphere (red) for reference images of CT of post‐treatment (a), CT‐CT registration (b) and 90Y SPECT (c). The arrows in (b) show the misalignment between the CT and SPECT images showing activity distributions outside of the liver boundary in the coronal and sagittal views.
Figure 5.

(a) Mutual information between co‐registered 90Y microsphere and 99mTc‐MAA SPECT images. (b) Patient 7 computed tomography (CT) (left), 99mTc‐MAA (middle) and 90Y microsphere (right) SPECT scans showing a minimal uptake within the liver. (c) Patient 4 CT (left), 99mTc‐MAA (middle) and 90Y microsphere (right) SPECT that showed the smallest mutual information between the co‐registered pre and post‐treatment images.
3.E. Tumor segmentation and TLR comparison
For 99mTc‐MAA, tumor volumes ranged between 160 and 1010 (695 ± 275) ml. 90Y‐microshpere uptake gave tumor volumes in the range of 207–1868 (786 ± 462) ml. Figure 6 shows the Box‐Whisker plot of the volumes including whole liver. 90Y‐microsphere SPECT/CT images gave an overall higher tumor volume for most patients compared to the 99mTc‐MAA SPECT/CT images. A paired t‐test of the two volumes gave insignificant difference (t = −0.82, P = 0.429; Table 3).
Figure 6.

Box plot of whole liver and segmented volumes tumor.
Table 3.
Result of paired t‐test between pre and post‐treatment tumor volumes
| N | Mean | SD | SE mean | |
|---|---|---|---|---|
| Paired T for 99mTc‐MAA vs Y–90 microsphere | ||||
| 99mTc‐MAA | 12 | 695 | 275 | 79 |
| Y–90 microsphere | 12 | 786 | 462 | 133 |
| Difference | 12 | −91 | 383 | 110 |
T‐test of mean difference = 0 (vs ≠ 0): T = −0.82.
P = 0.429.
The total mean TLR was 9.4 ± 9.2 and 5.0 ± 2.2 on 99mTc‐MAA and 90Y microsphere SPECT/CT, respectively. Figure 7 shows the scatterplot of mean TLRs from the two images displaying a significant correlation (r = 0.9, P = 0.00). From the plot, one patient appears to be an outlier (Grubbs’ outlier test, P = 0.00), and taking this patient out of the analysis gave a reduced correlation (r = 0.6, P < 0.05).
Figure 7.

(a) Scatterplot illustrating correlation between mean tumor to healthy liver ratios in 90Y microspheres and 99mTc‐MAA SPECT arrow indicating an outlier. (b) Outlier in (A) removed showing a reduced correlation between 90Y microsphere and 99mTc‐MAA SPECT TLRs.
3.F. Absorbed dose estimation
The mean absorbed doses for the tumor and healthy liver from 90Y bremsstrahlung SPECT/CT images were 62.6 ± 20.2 (range: 38.4 to 117.2 Gy) and 12.4 ± 4.7 (range: 6 to 23.7 Gy), respectively. Figure 8 demonstrates the relationship between the administered activities and absorbed doses. Tumor absorbed dose showed non‐significant correlation with the administered activity (r = 0.5, P > 0.05) however, there was a strong correlation between healthy liver absorbed dose and the administered activity (r = 0.8, P = 0.00). Figure 9 shows the cDVH of 90Y microsphere uptake in the tumor segments for all the patients. Figure 10(a) shows dose distribution of 90Y microsphere SPECT/CT for patient 6 with pre and post‐treatment TLR of 2.7 and 2.6, respectively, Fig. 10(b) shows patient 8 with pre and post‐treatment TLR of 10.2 and 5.3, respectively and Fig. 10(c) shows dose distribution for patient 12 with pre and post‐treatment TLR of 14.7 and 5.0, respectively. The isodose lines represent percent of maximum absorbed dose uptakes.
Figure 8.

Correlation of administered activity with healthy liver and tumor absorbed doses.
Figure 9.

cDVH of tumor volumes for all patients.
Figure 10.

Coronal slice through single photon emission tomography (SPECT) dose map with isodose lines fused with computed tomography showing 90Y microsphere dose distribution for patients 6(a), 8(b), and 12(c).
4. DISCUSSION
The primary objective of this study was to develop a post‐reconstruction algorithm to quantitatively improve the 90Y bremsstrahlung SPECT/CT images. The second objective was to compare the uptake distribution of pretreatment 99mTc‐MAA SPECT/CT and 90Y microspheres SPECT/CT images. In retrospect, the contribution of this study is in developing a clinically applicable method to improve the quantitative quality of 90Y bremsstrahlung SPECT/CT images for a more accurate quantitative comparison with 99mTc‐MAA SPECT/CT images. The proposed algorithm for 90Y bremsstrahlung SPECT/CT image improvement uses the Richardson–Lucy technique for contrast recovery. The challenges for our method were image degradation due to object scatter, septal penetration, and backscatter. Despite the inherent shortcomings, we achieved a meaningful improvement in CNR and QH which were measured in clinical and phantom 90Y bremsstrahlung SPECT/CT images, respectively. The small sphere volumes (2, 8, and 10 mm) showed indifferent results before and after contrast recovery. The challenge to draw the ROIs introduces extra errors in addition to the fact that the volumes might be highly influenced by noise. The smallest volume that gave an acceptable result was the 16 mm sphere with substantial signal difference from the background. Our result of the highest QH for the 34 mm was ~41% using a matching VOI. Applying similar methods used by other authors to evaluate QH, 16, 28 that is using an ROI in the slice through the center of the spheres, we found a higher value of QH = 80% for the 34 mm sphere. For a true dose assessment, however, this calculation should base on equal sized VOIs between the CT and SPECT image of the sphere.18 In addition, using an activity similar to clinically administered 90Y microsphere is beneficial in increasing accuracy of the calculation due to an increase in sample size of the counting statistics.
With the clinically available reconstruction method and application of contrast recovery algorithm, total liver activity estimation gave percent error of −5 ± 13 using the CF found from counts within the liver VOIs. Siman et al.18 generated a global CF derived from patient studies where entire counts within the FOV were considered. The global CF was applied to an IEC phantom with a 37 mm sphere insert and the authors reported an error of −25% with respect to the true activity. For our phantom study, we used CF estimated from the total reconstructed counts, which was within 95% CI of the patient calibration curve, and obtained total mean percent error of −23 ± 41%. Sphere volumes ≤12 mm resulted in the highest error. Further research is required for clinical application in patients with high LSF values.
The predictive accuracy of 99mTc‐MAA regarding the actual 90Y microspheres dose distribution and its impact on patient outcomes is a source of controversy. Ilhan et al.9 have studied the relationship between 99mTc‐MAA and 90Y microspheres uptake in a retrospective study which involved 502 patients of various liver cancer types. This study found a weak correlation between the mean tumor to background (healthy liver) ratios of 99mTc‐MAA SPECT/CT and 90Y microsphere SPECT/CT images. We believe a comparison of the pre and post‐treatment images of 99mTc‐MAA and 90Y microspheres may help evaluate the discordance between MAA distribution vs actual microsphere distribution. There are known factors that contribute to discrepancies. First and foremost is the variable size and shape of the albumin particles and clusters. Additional factors may include the embolizing effect of 90Y microspheres and differences in positioning of the catheter tip between the 99mTc‐MAA and 90Y microsphere procedures.
Prior to quantitative comparison, the pre and post‐treatment SPECT/CT images were co‐registered. The accuracy of the comparison strongly depends on the proper matching of the two images in all three dimensions. To minimize outlier effects, we used the ratio of the mean activities between the tumorous and healthy livers to calculate the TLRs. The mean TLRs between 99mTc‐MAA and 90Y microsphere SPECT/CT images showed a significant correlation with one patient as an outlier. Ilhan et al.9 state that the TLRs from 99mTc‐MAA SPECT/CT is higher in all patients compared to the respective 90Y microsphere SPECT/CT scans, which the authors believe to be due to the poor image quality of 90Y bremsstrahlung SPECT/CT. In our study, we found that after 90Y bremsstrahlung SPECT/CT image improvement the TRLs were higher in 99mTc‐MAA SPECT/CT for most of the patients. The standard deviation of the TLRs on 90Y microspheres SPECT/CT images is lower than 99mTc‐MAA SPECT/CT which could be associated with the embolizing effect of 90Y microsphere particles.8, 9
Dosimetry of 90Y microsphere distribution has the benefit of providing the real dose‐response relationship for further treatment if required, instead of the predicted dose‐response using 99mTc‐MAA. From our dosimetry analysis, the administered activity showed a strong correlation with the absorbed dose in the healthy liver, however, the relationship was weak with tumor absorbed dose. There are several factors that could contribute to the correlation between the absorbed doses and the administered activity. The higher non‐uniform microsphere distribution within the tumor vasculature compared to the healthy liver could contribute to the more proportional dose deposition in the healthy liver parenchyma vs to the tumor region.3
5. CONCLUSION
In this study, we proposed a technique for quantitative image improvement of 90Y bremsstrahlung SPECT/CT with the objective of quantitative comparison with 99mTc‐MAA SPECT/CT images and estimation of liver dosimetry. Within the limitation of our image correction method, the results of this study showed meaningful quantitative improvement of 90Y bremsstrahlung SPECT/CT images in phantom and patient studies. We identified a correlation in mean TLRs between 99mTc‐MAA and 90Y microspheres SPECT/CT uptake distribution. In addition, a linear relationship was observed between administered activity and absorbed dose in the healthy liver while the relation was insignificant with tumor absorbed dose.
CONFLICT OF INTEREST
Adjouadi M. received research grant from National Science Foundation (CNS‐1532061). The other authors declare no conflicts of interest.
ETHICAL APPROVAL
This retrospective study was performed in accordance with the 1964 Declaration of Helsinki and was approved by the Institutional Review Board of Florida International University (Number: 104882). Informed consent was waived for all patients included in this retrospective study.
ACKNOWLEDGMENT
The authors would like to thank Baptist Hospital of Miami for providing access to imaging devices.
REFERENCES
- 1. Hamami ME, Poeppel TD, Müller S, et al. SPECT/CT with 99mTc‐MAA in radioembolization with 90Y microspheres in patients with hepatocellular cancer. J Nucl Med. 2009;50:688–692. [DOI] [PubMed] [Google Scholar]
- 2. Stubbs R, Wickremesekera S. Selective internal radiation therapy (SIRT): a new modality for treating patients with colorectal liver metastases. HPB. 2004;6:133–139. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Gulec SA, Mesoloras G, Dezarn WA, McNeillie P, Kennedy AS. Safety and efficacy of Y‐90 microsphere treatment in patients with primary and metastatic liver cancer: the tumor selectivity of the treatment as a function of tumor to liver flow ratio. J Transl Med. 2007;5:15. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Kennedy A, Nag S, Salem R, et al. Recommendations for radioembolization of hepatic malignancies using yttrium‐90 microsphere brachytherapy: a consensus panel report from the radioembolization brachytherapy oncology consortium. Int J Radiat Oncol Biol Phys. 2007;68:13–23. [DOI] [PubMed] [Google Scholar]
- 5. Gulec S, Selwyn R, Weiner R. Nuclear medicine guidelines for radiomicrosphere therapy using Y‐90 microspheres in patients with primary and metastatic liver cancer. J Intervent Oncol. 2009;2:26–39. [Google Scholar]
- 6. Walrand S, Flux GD, Konijnenberg MW, et al. Dosimetry of yttrium‐labelled radiopharmaceuticals for internal therapy: 86Y or 90Y imaging? Eur J Nucl Med Mol Imaging. 2011;38:57–68. [DOI] [PubMed] [Google Scholar]
- 7. Dieudonné A, Garin E, Laffont S, et al. Clinical feasibility of fast 3‐dimensional dosimetry of the liver for treatment planning of hepatocellular carcinoma with 90Y‐microspheres. J Nucl Med. 2011;52:1930. [DOI] [PubMed] [Google Scholar]
- 8. Garin E, Lenoir L, Rolland Y, et al. Dosimetry based on 99mTc‐macroaggregated albumin SPECT/CT accurately predicts tumor response and survival in hepatocellular carcinoma patients treated with 90Y‐loaded glass microspheres: preliminary results. J Nucl Med. 2012;53:255–263. [DOI] [PubMed] [Google Scholar]
- 9. Ilhan H, Goritschan A, Paprottka P, et al. Predictive Value of 99mTc‐MAA SPECT for 90Y‐labeled resin microsphere distribution in radioembolization of primary and secondary hepatic tumors. J Nucl Med. 2015;56:1654–1660. [DOI] [PubMed] [Google Scholar]
- 10. Kappadath SC, Mikell J, Balagopal A, Baladandayuthapani V, Kaseb A, Mahvash A. Hepatocellular carcinoma tumor dose response following 90Y‐radioembolization with glass microspheres using 90Y‐SPECT/CT based voxel dosimetry. Int J Radiat Oncol Biol Phys. 2018;102:451–461. [DOI] [PubMed] [Google Scholar]
- 11. Shen S, DeNardo GL, Yuan A, DeNardo DA, DeNardo SJ. Planar gamma camera imaging and quantitation of yttrium‐90 bremsstrahlung. Energy. 1994;395:1381–1389. [PubMed] [Google Scholar]
- 12. Ito S, Kurosawa H, Kasahara H, et al. 90Y bremsstrahlung emission computed tomography using gamma cameras. Ann Nucl Med. 2009;23:257–267. [DOI] [PubMed] [Google Scholar]
- 13. Minarik D, Sjogreen Gleisner K, Ljungberg M. Evaluation of quantitative (90)Y SPECT based on experimental phantom studies. Phys Med Biol. 2008;53:5689–5703. [DOI] [PubMed] [Google Scholar]
- 14. Rong X, Du Y, Ljungberg M, Rault E, Vandenberghe S, Frey EC. Development and evaluation of an improved quantitative 90Y bremsstrahlung SPECT method. Med Phys. 2012;39:2346–2358. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Rault E, Staelens S, Van Holen R, De Beenhouwer J, Vandenberghe S. Fast simulation of yttrium‐90 bremsstrahlung photons with GATE. Med Phys. 2010;37:2943–2950. [DOI] [PubMed] [Google Scholar]
- 16. Elschot M, Vermolen BJ, Lam MG, de Keizer B, van den Bosch MA, de Jong HW. Quantitative comparison of PET and Bremsstrahlung SPECT for imaging the in vivo yttrium‐90 microsphere distribution after liver radioembolization. PLoS ONE. 2013;8:e55742. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Yue J, Mauxion T, Reyes DK, et al. Comparison of quantitative Y‐90 SPECT and non‐time‐of‐flight PET imaging in post‐therapy radioembolization of liver cancer. Med Phys. 2016;43:5779–5790. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Siman W, Mikell JK, Kappadath SC. Practical reconstruction protocol for quantitative (90)Y bremsstrahlung SPECT/CT. Med Phys. 2016;43:5093. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Richardson WH. Bayesian‐based iterative method of image restoration. J Opt Soc Am. 1972;62:55–59. [Google Scholar]
- 20. Fish D, Walker J, Brinicombe A, Pike E. Blind deconvolution by means of the Richardson–Lucy algorithm. J Opt Soc Am A. 1995;12:58–65. [Google Scholar]
- 21. Balagopal A, Kappadath SC. Characterization of 90Y‐SPECT/CT self‐calibration approaches on the quantification of voxel‐level absorbed doses following 90Y‐microsphere selective internal radiation therapy. Med Phys. 2018;45:875–883. [DOI] [PubMed] [Google Scholar]
- 22. Penny WD, Friston KJ, Ashburner JT, Kiebel SJ, Nichols TE. Statistical Parametric Mapping: The Analysis of Functional Brain Images. Amsterdam, Netherlands: Elsevier; 2011. [Google Scholar]
- 23. Maes F, Vandermeulen D, Suetens P. Medical image registration using mutual information. Proc IEEE. 2003;91:1699–1722. [DOI] [PubMed] [Google Scholar]
- 24. Chan TF, Vese LA. Active contours without edges. IEEE Trans Image Process. 2001;10:266–277. [DOI] [PubMed] [Google Scholar]
- 25. Franquiz J, Chigurupati S, Kandagatla K. Beta voxel S values for internal emitter dosimetry. Med Phys. 2003;30:1030–1032. [DOI] [PubMed] [Google Scholar]
- 26. Bolch WE, Bouchet LG, Robertson JS, et al. MIRD pamphlet no. 17: the dosimetry of nonuniform activity distributions—radionuclide S values at the voxel level. J Nucl Med. 1999;40:11S–36S. [PubMed] [Google Scholar]
- 27. Lanconelli N, Pacilio M, Meo SL, et al. A free database of radionuclide voxel S values for the dosimetry of nonuniform activity distributions. Phys Med Biol. 2012;57:517. [DOI] [PubMed] [Google Scholar]
- 28. Elschot M, Lam MG, van den Bosch MA, Viergever MA, de Jong HW. Quantitative Monte Carlo–based 90Y SPECT reconstruction. J Nucl Med. 2013;54:1557–1563. [DOI] [PubMed] [Google Scholar]