Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2020 Feb 1.
Published in final edited form as: IEEE Trans Med Imaging. 2018 Aug 13;38(2):349–359. doi: 10.1109/TMI.2018.2865121

CONtrast Conformed Electrical Properties Tomography (CONCEPT) Based on Multi-channel Transmission and Alternating Direction Method of Multipliers

Yicun Wang 1, Pierre-Francois Van de Moortele 2, Bin He 3
PMCID: PMC6372102  NIHMSID: NIHMS1520727  PMID: 30106715

Abstract

In Magnetic Resonance based Electrical Properties Tomography (EPT), circularly polarized magnetic field B1 from a transmit radiofrequency (RF) coil is measured and utilized to infer the electrical conductivity and permittivity of biological tissues. Compared to a quadrature RF coil, a multi-channel transmit coil provides a plurality of unique transmit B1 patterns that help to alleviate the under-determined-ness of EPT reconstruction problem, and it also allows to circumvent the “transceive phase assumption” that fails at ultra-high-field MRI. Here, a new approach, CONtrast Conformed Electrical Properties Tomography or CONCEPT, is proposed based on multi-channel transmission that retrieves electrical properties (EPs) by solving a linear partial differential equation with discriminated L1 and L2 norm regularization informed by intermediate EP gradient. The theory of CONCEPT and a fast reconstruction algorithm based on Alternating Direction Method of Multipliers are described and evaluated using numerical simulations, phantom experiment and analysis of in vivo human brain data at 7 T MRI. Compared to multi-channel gradient-based EPT (gEPT) method, this new technology does not require receive B1 sensitivity profiles and does not rely on symmetry assumption regarding RF coil design and imaged target. Moreover, it is not dependent on external prior information, such as integration seed point or anatomical MRI, which can be sources of bias in reconstructed EP values. By deriving EPs from transmit B1 profiles only, CONCEPT can be used with RF coils that include receive-only arrays with large channel count which can, in turn, offer substantial gains in Signal to Noise Ratio. It also holds potentials to image unsymmetrical body organs and diseased brain. CONCEPT provides solutions for the practical problems during the implementation of gEPT, thus representing a more generalized framework in the context of multi-channel RF transmission.

Keywords: electrical properties tomography (EPT), magnetic resonance imaging (MRI), B1 mapping, multi-channel transmission, alternating direction method of multipliers (ADMM)

I. INTRODUCTION

MAGNETIC Resonance based Electrical Properties Tomography (MREPT or EPT) is a non-invasive approach to retrieve tissue electrical conductivity and permittivity by fitting the measured radiofrequency magnetic field (B1 field) to time-harmonic Maxwell’s Equations [1]–[4]. The reconstructed electrical properties (EPs) provide fundamentally differential information compared to MR relaxometry, holding promise for additional sensitivity and specificity in clinical diagnoses, such as cancer identification and characterization [5]–[7]. Recent studies also shed light on the associations of EPs with other valuable biomarkers, such as sodium concentration, protein, and water content, that are more challenging for direct quantifications [8], [9]. In addition, subject-specific EP distribution is an essential ingredient for realistic local specific absorption rate (SAR) estimation, especially for ultra-high-field (≥ 7 T) applications where electromagnetic wave behavior becomes prominent [10], [11]. Improvements in the reliability of local SAR estimation would in return unleash the full power of ultra-high-field MRI that is otherwise limited by conservative safety constraints.

The problem of inferring tissue EPs from MRI measurable, circularly-polarized magnetic field B1 manifests as a nonlinear partial differential equation [12], [13]. In order to decrease the condition number and to obtain numerical stability, conventional EPT methods made the assumption that tissue EPs are piece-wise homogeneous, leading to a simplified reconstruction method known as “Helmholtz equation.” This assumption eliminates the EP spatial variation term, and one B1 distribution obtained with a quadrature volume coil suffices for a successful EP reconstruction. Actual EPs in the human body, however, are spatially heterogeneous, yielding boundary artifact and noise amplification that significantly thwart proper interpretations of Helmholtz-equation based EP reconstruction.

Recently, efforts have been made to address boundary artifacts coming from the Helmholtz equation. One potential solution is to utilize structural MR images to inform EP reconstruction on piecewise areas of relatively uniform contrast. This can be done by using a spatially weighted reconstruction kernel or adding regularization terms based on edge information [7], [14]. However, structural MRI contrasts and EP contrast do not rely on the same mechanism, therefore, spatial boundaries defined by their respective spatial variations do not necessarily coincide, which may bias the EP reconstruction process. Moreover, signal intensity in structural MRI is highly dependent on which MR sequences and parameters are used, creating additional sources of potential inconsistency that may propagate to EPs.

Other EPT approaches attempted to alleviate the boundary artifact by incorporating the spatial variation term into the reconstruction. Zhang et al. take EPs and their gradient terms as independent unknowns which are simultaneously retrieved using dual excitations of linear-polarization modes in a quadrature coil, demonstrating improved boundary fidelity [4]. However, the algorithm was based on the Cartesian components of the RF magnetic field that are not directly measurable from MRI. “Convection-reaction MREPT (cr-MREPT)” and its conductivity-only variant take the reciprocal of EPs as unknowns to linearize the reconstruction problem, from which 2D slices of EPs can be reconstructed in a single iteration [15], [16]. Nevertheless, the empirical assumptions made about B1 magnitude and phase, as well as EPs of the imaged target itself may introduce reconstruction errors.

Another category of EP mapping technologies is built upon forward modeling and electromagnetic simulations. In Contrast Source Inversion (CSI), the incident field of an empty coil is simulated, and the contrast source distribution is iteratively updated to minimize the discrepancies between scattering field and the observed field [17]. A similar idea can be implemented in spatial frequency space which retrieves EPs without iterations [18]. In addition, Global Maxwell Tomography (GMT) performs a full-wave simulation with a current guess of EP distribution, which gets updated iteratively to match the measured field [19]. These simulation-based algorithms utilize the integral form of Maxwell equations that naturally incorporates spatial variation of EPs. However, highly accurate electromagnetic simulations need to be performed and carefully registered to experimentally acquired data in order to preserve the delicate EP contrast, which is challenging to implement in realistic scenarios. To the best of our knowledge, there is no experimental validation of these approaches.

Alternatively, a multi-channel RF coil array can be used to effectively address the boundary artifact issue and B1 field assumptions by leveraging a plurality of B1 measurements [20]–[22]. Gradient-based EPT (gEPT) solves a pixel-wise inverse problem using both transmit B1 and receive B1, or B1+ and B1, to retrieve the EP gradient, followed by its spatial integration based on a user-assigned seed point [23]. This method successfully retrieves both conductivity and permittivity in high quality with improved robustness against noise. Nevertheless, two caveats need to be taken into consideration which may cause errors in the reconstruction results: i) symmetry assumption about RF coil structure and imaged target; ii) subjective assignment of integration seeds, i.e. tissue voxels with presumptive EP values.

In this study, we report a novel EPT approach, dubbed CONtrast Conformed Electrical Properties Tomography (CONCEPT), based on a linear inverse problem regularized by intermediate EP gradient information, using only B1+ maps from a multi-channel RF coil array. Preliminary ideas about this technology were recently presented in [24]. Compared to gEPT, it does not make symmetry assumptions, nor does it require subjective assignment of integration seed points. Numerical simulations of a digital phantom and a realistic human head model have been performed to illustrate the concept of the proposed approach and to evaluate its performance under various noise levels. A phantom experiment as well as a retrospective in vivo human brain data analysis has been performed at 7 T to demonstrate its applicability in realistic settings. CONCEPT extends the generality of gEPT and holds promise for clinical applications using a generic multi-transmit RF coil.

II. THEORY

In this section, we first derive the central physical equation from the governing electromagnetic theory that relates the measurable B1+ maps to EPs. We then introduce a discriminated regularization strategy using the intermediate EP gradient values, which represents the core of CONCEPT. Lastly, we describe an algorithm based on Alternating Direction Method of Multipliers (ADMM) to rapidly solve the central optimization problem derived from the previous steps. The reconstruction flowchart is shown in Fig. 1.

Fig. 1.

Fig. 1.

Reconstruction flowchart of CONCEPT.

A. Derivation of Data Fidelity Term

An MRI scan involves electromagnetic radiation from an RF coil at the proton resonance frequency or Larmor frequency. Magnitude and phase distributions of the time-harmonic magnetic field reflect electrical properties of the underlying tissue. Combining Ampère’s law and Faraday’s law, we have

2H=ω2μεcHεc××Hεc (1)

where denotes the angular Larmor frequency, μ magnetic permeability, and εc := ε − iσ/ω electrical properties comprised of permittivity ε and electrical conductivity σ. By transforming H in the Cartesian system to a rotating frame using H1+=(Hx+iHy)/2, with assumptions that B1+:=μH1+μ0H1+andHz=0 for a longitudinal coil element placed in z direction, (1) can be written as

2B1+=ω2μ0εcB1++(B1+xiB1+y)gw+B1+zgz (2)

where gw=lnεcx+ilnεcy is the spatial variation of EPs in the xy plane, and gz=lnεcz is that in the z direction.

Note that B1+=|B1+|eiφ+ is a complex quantity with both magnitude and phase components. The magnitude of B1+ can be readily mapped by a variety of well-established “B1 mapping” methods [25]–[27]. However, the absolute phase of B1+ is not directly measureable, since MRI signal phase has contributions from B1+, B1, B0 and chemical shift. Utilizing a multi-channel transmit coil array, signal phase components irrelevant to B1+ can be cancelled out by taking the relative phase with respect to a reference channel r. Liu et al. has shown that the spatial gradient of absolute B1+ phase ∇φ+ can be retrieved simultaneously with gw and gz by solving (2), given that at least four channels are employed [23], [28]. The retrieved ∇φ+ and gz are then re-introduced back into (2) as known variables. As proposed by Hafalir et al. in the cr-MREPT method [15], (2) can be transform into a linear partial differential equation (PDE) in 2D with respect to the scaled complex admittance γc=1ωεc

(2B1+B1+zgz)γc+(B1+xiB1+y).(γcx+iγcy)=ωμ0B1+ (3)

with unknown terms written in red. The spatial gradient term of γc can be discretized using central difference method

γcx+iγcy=γc(x+1,y)γc(x1,y)2dx+iγc(x,y+1)γc(x,y1)2dy (4)

where (x, y) is spatial location indices, dx and dy are step size along x and y directions, respectively. As such, (3) can be written as a linear inverse problem for each transmit channel k

Akϒ=bk (5)

where ϒ ∈ ℂN contains γc values of all N pixels in a region of support Ω in a 2D imaging slice, Ak ∈ ℂN×N is a sparse matrix with five nonzero elements in each row, and bk ∈ ℂN is a vector of scaled B1+. The inverse problems derived from K different transmit channels can be concatenated as

Aϒ=b (6)

where A = [A1; A2;...; AK], b = [b1; b2;...; bK]. In this study, K = 16 sinceB1 maps from all 16 RF coil elements were used. We utilize L2-norm of the residue, Aϒb22, as a metric of data fidelity to the B1+ measurements, which is optimized by the pseudo-inverse of (6)

ϒPDE=(AHA)1AHb (7)

This solution is referred to as “Multi-channel cr-MREPT” in this paper.

B. Discriminated Regularization

Direct inversion of (6) results in spatially pervasive reconstruction errors due to PDE nature of the problem, intensified by noise effects and inaccuracies in B1+ measurement. Furthermore, numerical errors due to the finite difference model of the problem also deteriorate image quality [29]. Therefore, we propose to use the intermediate variable gw as prior information to determine the EP transition area and guide image reconstruction. To start, we hard-threshold |gw| to generate two binary masks

{W=(|gw|κ)ΩW¯=(|gw|<κ)Ω (8)

where κ is an empirical threshold so that approximately 30% of the pixels in Ω are identified as transition area. In the homogeneous area W¯=1, the L2-norm of γ gradient is minimized to simultaneously suppress large spatial variation and error propagation. In the transition area W, the L1-norm of γ gradient is minimized to promote edge sparsity. Therefore, the central reconstruction is formulated as

minϒ12Aϒb22+λ1WΨϒ1+λ2W¯Ψϒ22 (9)

where is the finite difference operator calculating ∇γ in the region of support. Since the regularization is informed by intermediate EP contrast instead of structural MRI, this technology is named as “CONtrast Conformed Electrical Properties Tomography”, or CONCEPT.

C. Solving the Central Problem Using Alternating Direction Method of Multipliers

Equation (9) is a large-scale optimization problem involving simultaneous minimization of L1-norm and L2-norm. Conventional gradient-based methods, such as gradient descent or conjugate gradient method, are inefficient for solving this type of problem due to the non-smoothness of L1-norm and the requirement of evaluating function gradient in each iteration. Alternatively, we propose to use Alternating Direction Method of Multipliers (ADMM) to solve (9). ADMM is based on the philosophy of “divide and conquer” in algorithm design: it splits the original complicated problem into a series of simpler sub-problems, which are sequentially addressed until a consensus is reached [30]. ADMM has received increasing attention due to its high efficiency for large-scale image reconstruction problems, especially those with sparsity and low-rank constraints [31]–[33].

In our implementation, we first combine the two L2-norm terms in (9)

minϒ12Eϒc22+λ1WΨϒ1 (10)

where E=[A;λ2W¯Ψ] and c = [b; 0]. Introducing an auxiliary variable z so that (10) is equivalent to

minϒ12Eϒc22+λ1z1s.t.WΨϒz=0 (11)

The augmented Lagrangian can be formed as

Lρ(ϒ,z,u)=12Eϒc22+λ1z1+Re(ρuH(WΨϒz))+ρ2WΨϒz22 (12)

where u is the multiplier and ρ > 0 is the augmented Lagrangian parameter. We can express ADMM updates as

Υk+1=argminϒLρ(ϒ,zk,uk)=argminϒ12Eϒc22+Re(ρ(uk)H(WΨϒzk))+ρ2WΨϒzk22=argminϒ12Eϒc22+ρ2WΨϒzk+uk22=(EHE+ρΨHWHWΨ)1(EHc+ρWΨ(zkuk))
zk+1=argminzLρ(ϒk+1,z,uk)=argminzλ1z1+Re(ρ(uk)H(WΨϒk+1z))+ρ2WΨϒk+1z22=argminzλ1z1+ρ2zWΨϒk+1uk22=Sλ1/ρ(WΨϒk+1+uk)
uk+1=uk+WΨϒk+1zk+1

where Sτ (a) denotes the soft-thresholding function of a with τ. z0 and u0 are initialized with zero vectors. Note that the algorithm does not require initialization of ϒ, since ϒ1 is calculated from z0 and u0 in the first iteration. The stopping criterion is

ϒk+1ϒk2ϒk+12tolerance,ork>kmax (13)

We observed that the algorithm typically converges within 15 iterations using tolerance = 10−4.

III. METHODS

A. Numerical Simulations

Electromagnetic simulations were performed in a Finite-Difference Time-Domain (FDTD) based software (Sim4Life version 3.4, ZMT, Zurich, Switzerland). The 16-channel stripline coil array used in the experiment was numerically modeled and loaded with a cylindrical phantom containing tube anomalies, an extruded (in z-direction) brain slice with a circular anomaly, and a realistic head model (Duke Model, Virtual Family [34]), as shown in Fig. 2. The digital phantom was 160 mm tall, with a long axis of 210 mm and a short axis of 168 mm, and was assigned the mean electrical properties of brain at 298 MHz (0.55 S/m and 52ε0) [35]. Five sets of tube anomalies with diameters of 5 mm, 10 mm and 20 mm were evenly distributed in azimuthal direction, spanning the physiologically relevant EPs range of 0.2–2.5 S/m and 40ε0-100ε0. For the head models, EPs of various tissues were assigned based on Gabriel et al. [35]. The anomaly in Fig. 2b mimics a brain tumor or hemorrhage that disrupts structural symmetry, with EPs of 2 S/m and 76ε0. The imaged objects were positioned in the center of the coil, mimicking actual experimental setup.

Fig. 2.

Fig. 2.

Simulation setup. The 16-channel stripline RF coil array is loaded with (a) a numerical phantom, (b) an extruded brain slice with a 50-mm-diameter anomaly in red, and (c) a realistic 3D Duke head model. The last one is shown as voxels after discretization.

Spatially anisotropic discretization of the models was performed by the software based on their geometry and EP distributions to achieve the optimal performance. A maximal spatial step of 1.5 mm was prescribed inside the phantom/head model to ensure high precision. The 16 channels of the coil were excited sequentially with a normalized input power of 1 W. The simulated complex B1+ field was re-gridded to a homogenous 1.5 mm × 1.5 mm × 1.5 mm grid using cubic splines. Relative B1+ phase was calculated taking each channel as reference. Complex white Gaussian noise was added to B1+ to adjust the signal to noise ratio (SNR) to desired levels.

B. Phantom Experiment

An agar phantom was constructed to validate the performance of CONCEPT in realistic settings. The “background” or main body of the phantom contained deionized water, agar, NaCl, NiCl2· 6H2O with a mass ratio of 1000:15:1.12:0.5. Eps of the phantom at 298 MHz were measured using a dielectric probe (85070E, Keysight Technologies) as 0.294 S/m and 78.6ε0. The main body was poured into a cylindrical jar with 120 mm diameter and 150 mm height. The contrast solutions were prepared based on deionized water, with NaCl to adjust conductivity and polyvinylpyrrolidone (PVP) to adjust permittivity [36]. Three solutions with different EPs were prepared and measured using the probe. Solutions were filled into a 25-mm-diameter plastic tube and a 5-mm-diameter straw, and inserted into the phantom background. The whole phantom was allowed to stand for 24 hours in room temperature to solidify. The insulating materials of the anomaly wall do not significantly affect electromagnetic wave propagation and EP reconstruction at Larmor frequency, as investigated in [37].

Experiments were performed on a 16-channel Siemens 7T system equipped with 16 × 1 kW power amplifiers (CPC, Hauppauge, NY) with their transmit power and phase remotely controlled by a phase/gain control unit, interfaced with a locally developed software toolbox. A 16-channel RF coil array, consisting of stripline elements aligned with the main magnetic field direction, was used to obtain multiple B1+ patterns. The coil array was connected to the MRI system through a custom 16-channel T/R switch box. Each coil element was tuned and matched at the operating frequency (297.2 MHz) when loaded with object under detection. B1+ mafgnitude map from each channel was acquired using a hybrid method of multi-slice gradient-echo (GRE) and 3D Actual Flip-angle Imaging (AFI) method [25], [38], [39]. For the former, each coil element was excited sequentially in small-flip-angle regime (< 15°), with TR = 76 ms, TE = 3.3 ms and 8 averages, duration 23.0 min; For the latter, two shimmed B1+ patterns resembling CP+ mode and CP2+ mode, respectively, were imaged with 1 average each, nominal flip angle = 55, TR = 20/120 ms, TE = 3.2 ms, total duration 19.4 min. The entire B1+ mapping session takes approximately 60 min, with additional ~10 min of pilot single-channel B1+ acquisition for shimming purpose, and ~10 min of pilot AFI acquisition for calibration of flip-angle. Hybrid B1+ reconstruction of each transmit channel was initially performed for the two shimmed patterns separately, then combined with weights of the flip angle map, so as to circumvent inaccuracies due to RF nulling spots present in the CP+ and CP2+B1+ maps. Details are provided in Appendix A. B1+ maps were acquired with in-plane resolution of 1.5 mm × 1.5 mm over a stack of 12 contiguous slices of 3 mm thickness.

C. In Vivo Validation

A retrospective data analysis of human brain was performed. The data was described in [23]. The study was approved by the Institute Review Board at the University of Minnesota. Briefly, single-channel B1+ mapping with relative phase was performed at 7 T in a similar setup to the phantom experiment using the hybrid B1+ mapping method of a single AFI (1 average, 9 min) and a series of small-flip-angle GRE (10 average, 28 min) in a resolution of 1.5 mm × 1.5 mm × 5.0 mm.

D. Image Reconstruction

Simulated noisy data and experimental data were de-noised using a low-pass Gaussian filter with a kernel size of 5×5×3 voxels and standard deviation of 1.2 × 1.2 × 0.8 voxels. The spatial gradient and Laplacian of B1+ were generated based on second order polynomial fitting using a 5 × 5 × 3 Savitzky-Golay filter [40]. Image reconstruction was performed using MATLAB (The MathWorks, MA, USA) codes developed in-house, running on a desktop with a 3.4 GHz GPU and 32 GB RAM.

IV. RESULTS

A. Digital Phantom Simulation

The simulated transmit B1 field and absolute phase gradient from channel 1 are shown in Fig. 3. The spatial variation of B1 phase is smooth except for the area corresponding to low B1. The retrieved phase gradient using (2) is in good agreement with the ground truth.

Fig. 3.

Fig. 3.

Transmit B1 field and phase gradient from channel 1 of the phantom simulation. (a) Transmit B1 magnitude and absolute phase. (b) Phase gradient calculated from the absolute phase juxtaposed with the retrieved phase using equation (2).

EP reconstruction results using Helmholtz equation, Multichannel cr-MREPT, gEPT and CONCEPT are compared in Fig. 4. For Helmholtz equation, accurate reconstruction is limited to large homogenous areas, such as the inside of large tubes as well as the background. Severe oscillations can be observed at the interface of different contrast components. In particular, reconstruction results of small structures (5 mm and 10 mm tubes) are dominated by boundary artifact. For Multi-channel cr-MREPT, boundary artifact is significantly mitigated, thanks to the inclusion of the EP variation term. Nevertheless, chessboard-like artifacts still appear due to the use of finite difference as an approximation of spatial derivative (white arrow). This artifact is more pronounced in the tubes with large EP values because of higher contrast to the background. In addition, erroneously reconstructed values can be identified near the phantom boundary close to individual RF coil elements (green arrows). This artifact arises from inaccurate estimation of local high-order spatial derivatives of B1+ due to its rapid attenuation near each coil element. As shown in the profiles in Fig. 4b, without constraints, such artifacts can bleed into the contrast tubes and bias reconstruction results. For gEPT, one seed point in the center of the phantom was used to implement the algorithm (red asterisk). High-frequency chessboard artifact exists in both EP maps, on top of low-frequency bias induced by violation of the “symmetry assumption”. The EP profiles demonstrate spatially dependent bias in the reconstruction result, which tends to be smaller near the seed point and accumulates going outward radially. For CONCEPT, reconstructed EP values and profiles are close to those of the target. The chessboard artifact and spatial bias are minimized thanks to the informed regularization.

Fig. 4.

Fig. 4.

Comparison of different reconstruction methods using simulated phantom data. (a) Reconstructed conductivity and permittivity using Helmholtz equation, Multi-channel cr-MREPT, gEPT and CONCEPT are compared with target maps. Red asterisk labels position of the seed point used by gEPT. Both transmit and receive B1 were employed for gEPT. Other methods do not require seeds or receive B1. White and green arrows denote chessboard artifact and B1 peak related artifact, respectively. (b) Reconstructed EP profiles along the red line on the target maps in (a).

Reconstructed EP images using CONCEPT at various SNR levels are shown in Fig. 5a. The obtained EP values inside the large tubes are fitted against the target values in Fig. 5b. Accurate conductivity and permittivity can be obtained for SNR = 200 and 100 within the ranges of interest, with reconstruction errors being ≤ 0.15 S/m for conductivity and ≤ 5.0ε0 for permittivity. On the conductivity images, the smallest tubes can also be well differentiated from the background and from each other, except for the one with the least conductivity contrast to the background (0.25 S/m). Detection of small structures on the permittivity image is more difficult. Only the two tubes with the largest contrast (≥ 28ε0) are clearly visible. At SNR = 50, the fitting results show global overestimation in both contrasts. Conductivity retains reasonable imaging quality especially for the large tubes, yet permittivity degrades significantly. Notably, overestimation is pronounced for the tube with the largest EPs. Potential reasons for this overshot are discussed in Appendix B.

Fig. 5.

Fig. 5.

CONCEPT reconstruction results at SNR = 200, 100 and 50. (a) Reconstructed conductivity and permittivity maps. (b) Linear fitting results of the retrieved values inside the largest tubes against the target values. R2 values of fitting for SNR = 200, 100, 50 are 0.99, 0.99, 0.97 for conductivity and 0.99, 0.98, 0.86 for permittivity.

B. Extruded Brain Simulation

Reconstructed conductivity maps of the extruded brain model using gEPT and CONCEPT are compared in Fig. 6. As a consequence of asymmetry in the imaged target, gEPT leads to spatial bias in the result even using the symmetric 16-channel RF coil. Notably the conductivity of the simulated anomaly is underestimated. This bias is alleviated using CONCEPT.

Fig. 6.

Fig. 6.

Comparison of reconstructed conductivity using gEPT and CONCEPT in the presence of asymmetry. Red rectangular denotes coils from which B1 is used, and gray denotes coils from which B1 (both transmit and receive) is discarded.

Fig. 6 also shows results using an asymmetric RF coil array by discarding B1+ and B1 information from the coils shown in gray, mimicking dead channels due to malfunction of amplifier/TR switch/coil element or a different design of the RF coil. Due to this hardware asymmetry, gEPT reconstruction error is significantly elevated. In contrast, CONCEPT demonstrates robustness with a mild change in the reconstruction results.

C. Realistic 3D Brain Simulation

As an intermediate step of CONCEPT, the determination of the regularization mask from |gw| is shown in Fig. 7. By solving equation (2), gw can be retrieved which faithfully reflects the boundary of tissues, even in the noisy case (Fig. 7a). Selecting a threshold for the transition area involves a tradeoff between sensitivity to low-contrast or small structures and robustness against measurement noise. 30% percentile seems to give a good balance between the two in this case, meaning that most transition areas can be correctly identified (true positive rate is high) without including too many noisy pixels (true negative rate is high).

Fig. 7.

Fig. 7.

Determination of CONCEPT mask in the 3D Duke simulation. (a) Retrieved |gw| using either noise-less or noisy (SNR = 50) B1. (b) Effects of the threshold. 20%, 30%, 40% are the percentage of pixels considered as transition pixels.

Three axial slices of the reconstructed brain model with a spatial gap of 12 mm are shown in Fig. 8. Note that the low-pass filter was not applied to the noise-less B1+ data, but only to the noisy B1+ (SNR = 50). For both SNR levels, all three slices show successful reconstruction of the contrasts amongst major brain tissues, e.g. white matter (WM), gray matter (GM) and cerebrospinal fluid (CSF). The reconstructed values of these three tissues are summarized in Table I. Consistent with the digital phantom results, we observe overestimation in the permittivity of CSF, even for the noiseless case. This is potentially due to the increased complexity of the real-3D Duke model compared to the 2D-projected phantom model. The magnitude of this overestimation is smaller for the noisy reconstruction because of the smoothing effects of the low-pass filter.

Fig. 8.

Fig. 8.

Reconstructed EPs of a simulated 3D Duke head model. Shown are the target maps and reconstruction results using noise-free B1 data as well as noisy B1 with SNR = 50. The three axial slices are separated by 12 mm.

TABLE I.

Reconstructed EPs of Duke Head Model

Target Noise-free SNR = 50
WM σ 0.41 0.58 ± 0.10 0.62 ± 0.12
ε 43.8 47.2 ± 4.3 52.0 ± 4.8

GM σ 0.69 0.63 ± 0.30 0.68 ± 0.33
ε 60.1 56.1 ± 9.5 58.9 ± 9.2

CSF σ 2.22 1.70 ± 0.34 1.57 ± 0.29
ε 72.8 105.2 ± 9.6 100.8 ± 7.0

σ in S/m and ε in ε0

D. Phantom Experiment

A structural image of the phantom and B1 images from a single channel are shown in Fig. 9. Local spotty artifacts can be seen in the hybrid B1 map when using a single shim pattern (Fig. 9c white arrows). These artifacts occur at the locations of RF nulls due to complex destructive interference of B1+ between channels [39]. By contrast, B1 image obtained merging two shim patterns is free of such errors, bearing high resemblance to the B1 pattern measured by AFI (Fig. 9b).

Fig. 9.

Fig. 9.

Hybrid transmit B1 mapping based on dual shims. (a) An axial slice of the agar phantom used in the experiment. Scale bar is 10 mm. (b) Transmit B1 magnitude from channel 5 (ch5) measured using AFI. (c) Hybrid B1 mapping results of ch5 based on CP-like shim I and CP2+-like shim II, and their combination. αI and αII are measured using AFI. White arrows point to locations with inaccurate calculated B1 due to RF nulls in the corresponding shimmed B1 pattern.

Based on the combined hybrid B1 images, reconstructed EPs of the phantom are shown in Fig. 10, and summarized in Table II. As an intermediate quantity, the spatial variation or gradient strength of EPs across the phantom can be robustly retrieved, enabling a clean separation of EP contrast boundary (Fig. 10a). Importantly, compared to the Helmholtz based reconstruction, CONCEPT gives accurate conductivity reconstruction results not only in homogenous areas but also around contrast boundaries. In addition, all three 5-mm small tubes are visible from the reconstructed map. Similar to the simulations, there is overshot in the reconstructed permittivity of the tubes with high EP values (tube III). Reconstruction errors of the other tubes are relatively smaller.

Fig. 10.

Fig. 10.

Phantom reconstruction results. (a) Intermediate gradient strength distribution and CONCEPT mask generated by hard-thresholding. (b) Comparison of the reconstruction results using Helmholtz equation and CONCEPT.

TABLE II.

Reconstructed EPs of the Agar Phantom

Probe Measurement Reconstructed
Tube I σ 0.65 0.70 ± 0.05
ε 50.2 65.0 ± 1.6

Tube II σ 1.02 0.98 ± 0.02
ε 62.7 73.6 ± 0.7

Tube III σ 2.14 2.14 ± 0.07
ε 78.8 124.0 ± 4.3

σ in S/m and ε in ε0

E. In Vivo Validation

Reconstructed in vivo images are presented in Fig 11. Large brain compartments, such as the ventricles filled with CSF, and the GM, can be distinguished from white matter with reasonably clear boundaries. However, details of the gyri and sulci are difficult to identify as a result of low-pass filtering. Conductivity in CSF, GM and WM are found to be 1.7±0.58, 0.9 ± 0.42 and 0.5 ± 0.25 in S/m, respectively. Reconstructed permittivity values for the same brain compartments are 93± 25.4, 71 ± 15.2 and 61 ± 3.4 in ε0.

Fig. 11.

Fig. 11.

In vivo EPs imaging result using CONCEPT compared to T1-weighted anatomical image.

V. CONCLUSION AND DISCUSSIONS

In this study, we propose a novel approach, dubbed “CONCEPT”, to retrieve EPs from MRI using multiple RF transmission. CONCEPT leverages on the intermediate EP gradient information for edge detection and discriminated regularization. The resultant large-scale, mixed L1 and L2 norm image reconstruction problem is solved efficiently using ADMM. A systematic investigation using numerical simulations and phantom experiments at 7 T MRI demonstrate its capability to retrieve accurate EP values in large homogeneous areas (≥ 20 mm), as well as to detect small structures down to 5 mm, and to faithfully reflect the EP transition zone. CONCEPT addresses the symmetry assumptions and seed point requirements of the gEPT method, representing a step forward in clinical applications of the multi-channel EPT approach.

The mere Helmholtz-based approaches of MREPT suffer from boundary artifact, spatial bias, and noise amplification effects, due to a substantial simplification of the governing physical equations and to the use of high-order spatial differentiation operators. These adverse effects can be partially mitigated using numerical regularization with prior information such as contrast edges detected by structural MRI [7], [14]. Nevertheless, spurious EP contrast may be introduced, since it is fundamentally different from magnetic relaxation contrasts and does not necessarily spatially coincide with structural MRI. In comparison, CONCEPT exploits EP gradient information directly derived from B1+, thus it is in principle immune to other sources of contrast such as MR relaxation effects. However, CONCEPT is based on the framework of multi-channel RF transmission, which is not yet widely available in clinical settings. Meanwhile, acquisition of multiple high-resolution B1 requires lengthy scans which potentially compromise patient comfort and throughput.

For single-channel PDE-based MREPT methods, a Dirichlet boundary condition is usually enforced to produce a stable result, since the central equation is underdetermined and its possible solutions are infinite [15], [16]. However, boundary conditions may be complex and intractable in practical settings, such as imaging of the entire human brain. Erroneously assigned boundary conditions can adversely affect reconstruction result not only in the periphery of the imaged structure, but in the center of it through the spatial differentiation term. Thanks to the plurality of B1+ provided by a multi-transmit RF coil, the central equation of CONCEPT is overdetermined and it does not rely on any boundary condition. Nevertheless, certain forms of relaxed boundary condition, e.g. constrained-value reconstruction in physiologically relavent ranges, may be enforced to further stablize the result.

The global PDE version of MREPT is prone to reconstruction bias in the center of the imaged structure, so-called “low convection field (LCF) artifact” by Hafalir et al. in the cr-MREPT study [15]. It is attributed to small values of B1+/xiB1+/y in the center of a quadrature coil which weights the spatial derivative of EPs. This artifact can also be observed when using multiple transimit coils, since all channels share a small-B1+ region in the center of the imaged structure, making it both low-SNR and low-convection-field [12]. A couple of strategies can be employed to suppress this artifact. Firstly, the LCF can be shifted by padding dielectric material near the subject [15]. For a multi-transmit coil, this can also be potentially done by manipulating B1 shim patterns. Secondly, regularization and/or a prior information can be employed to correct the bias, e.g. enforcing constrained values in “constrained cr-MREPT” [15] or using regularization in CONCEPT. Last but not the least, spatial propagation of noise should be suppressed by low-pass filering noisy B1 in preprocessing and utilizing noise-robust differential kernels, such as the Savitzky-Golay filters [40].

In this study, each imaging slice was reconstructed separately, and gz was not explicitly emploited except as a part of coefficient in (3) weighted by B1+/z which is a small value in the central part of the longitudinal stripline coil. 3D EPs reconstruction can potentially be performed by incorporating EPs gradient in the z direction into the central equation. However, the following two factors need to be considered. Firstly, the ubiquitous quadrature coil and TEM coils may not be suitable for this purpose. Since EPs gradient in the z direction is weighted by B1+/z small values of B1+/z in these coils make z-gradient terms neglegible in the central equation compared to x- and y- gradient terms weighted by B1+/x and B1+/y (Fig. 12). Non-TEM coils with a larger B1+ variation in the z direction, such as a small loop array, may better pronounce EPs gradient along z. Secondly, the size of A matrix in (6) scales quadratically with ϒ, and it becomes more complicated internally due to incorporation of z differentiation. These factors increase the computational complexity and size of the problem which may lead to instability of the reconstruction result. 999

Fig. 12.

Fig. 12.

Comparison of B1+ gradient components in x, y and z directions using the realistic 3D Duke model.

Compared to the multi-channel gEPT method [23], CONCEPT has the following merits. Firstly, the reconstruction process does not require assignment of integration seed, which is unavailable in realistic settings. Secondly, B1 is no longer needed, leading to a shorter scan time and improved patient compliance. Last but not least, the imaging setup, including both the subject brain and RF coil, does not need to be symmetric, opening the potential to image diseased brain and irregularly shaped body organs using customized RF coil. In fact, since B1 and B1+ are completely disentangled in CONCEPT, one is no longer limited to transceiver RF coil arrays. Instead, separate transmit and receive coil arrays, that are more ubiquitous than customized transceiver arrays, can be utilized. In addition, such coil type generally has a higher count of receive elements with dedicated high-performance electronics in a close proximity to the imaged target, which significantly improves SNR for MR data acquisition. Additional transmit elements could also be used to obtain a larger number of different B1+ patterns to improve the conditioning of the inverse problem.

In its current form, CONCEPT is based on multiple B1+ maps, yet its counterpart using multiple B1 maps can be derived in a similar way. The major advantage of using multiple B1 lies in its high temporal efficiency, since data acquisition can be accomplished in a single acquisition [41]. However, B1 maps are typically weighted by proton density which cannot be removed in a straightforward way. Computing the ratio of B1 maps between different channels cancels out the shared proton density weighting, yet the resultant relative B1 magnitude cannot be directly utilized by CONCEPT. Alternatively, proton density can be taken as unknown variables and reconstructed simultaneously with EPs [28].

The central equation of CONCEPT (9) contains two regularization terms, i.e. L1-norm and L2-norm of spatial gradient of ϒ. The effective areas of these two norms are controlled by a binary mask W that is derived from hard-thresholding the intermediate EP gradient. More advanced regularization strategies may be employed to further improve image quality. Firstly, W could be designed as a continuous function of EP gradient, such as linear or logarithm, to better characterize EP distribution of imaged object and gain robustness against measurement noise. Secondly, the L1-norm of spatial gradient of ϒ, known as total variation (TV), could be replaced by total generalized variation (TGV) containing higher orders of spatial derivative. TGV does not imply the piecewise constant assumption, thus it is capable of alleviating the “staircase” artifact and capturing contrast changes that are naturally continuous [42]. Thirdly, an additional weighting function can be designed and imposed to ϒ to counteract B1+ weighting to the reconstruction results seen in Fig. 8. Lastly, low-rank constraints could be imposed to better represent different types of tissue, as has been shown in other parametric MRI technologies [31]. These features can be added to CONCEPT and solved using ADMM without significant modifications to the current algorithm structure. However, caution needs to be used as a much higher dimension of regularization parameters needs to be explored to obtain the optimal reconstruction outcome.

CONCEPT is based on the 3rd order spatial differentiation of B1+, since it requires the spatial derivative of ∇φ+ which is calculated from an equation involving 2B1+ [23]. This makes the algorithm more sensitive to noise contaminations, flow effects and motion, which can only be partially mitigated by regularization. Data pre-processing using intensive low-pass filters can alleviate the adverse effects, yet at a price of resolution loss which can be deleterious when targeting small structures. As a result, only large brain compartments are visible in the current in vivo brain imaging result. To visualize complex anatomical structures using CONCEPT, improvements in SNR are warranted. This could be achieved by implementation of recently introduced B1 mapping methods such as DREAM [27] or Bloch-Siegert shift [26]. In addition, transmit channels could be aggregated to generate a smaller number of “virtual channels” that have better spatial coverage than a single transmit element. The gained SNR can be traded for shorter scan time or higher resolution as desired.

ACKNOWLEDGMENT

The authors thank Dr. Gregor Adriany for MRI hardware assistance; Drs. Jiaen Liu, Xiaotong Zhang, Long Yu and Mehmet Akçakaya for constructive discussions.

This work was supported in part by NIH R21 EB017069, R01 MH114223, R01 EB021027, R01 NS096761, R01 AT009263, R21 EB014353, P41 EB015894, R01 EB011551, S10 RR026783, NSF CBET-1450956, and WM KECK Foundation.

APPENDIX

A. Hybrid Transmit B1 Mapping based on Dual Shims

At the beginning of the transmit B1 mapping session, pilot single-channel B1 is acquired, which is used to determine two static B1 shim settings resembling CP+ (circularly polarized, phaseincrement360°16=22.5°) and CP2+ (phaseincrement720°16=45°). These two shims are used because they have optimized transmit efficiency and their B1 patterns are complementary to each other [43]. Flip-angle images αI and αII are acquired in these two modes using AFI. Single-channel GRE is acquired once, and combined with αI and αII separately using the original hybrid B1 mapping method [38] to obtain B1,I and B1,II (Fig. 9c). Final single-channel B1 is calculated using weighted sum

B1=αIB1,I+αIIB1,IIαI+αII (14)

The Combined B1 Is Thus Free From Spurious Singularities Due to Small Flip-Angle in Either Shim I or Shim II

B. Analysis of Error Propagation

In this section, we propose and analyze potential reasons why the reconstructed permittivity map has more severe distortion and overshot during CONCEPT reconstruction.

Firstly, as pointed by Marques et al. [41], given that the real and imaginary part of εc can be reconstructed at the same SNR level, a change in permittivity of 10ε0 is commeasurable to a change in conductivity of 0.17 S/m at 7 T. In human brain, the normal physiological range of permittivity is 40ε0 - 70ε0, corresponding to a conductivity range of 0.68 S/m - 1.19 S/m, which is much narrower than the actual physiological range of conductivity 0.41 S/m - 2.22 S/m. Therefore, permittivity is relatively more susceptible to disturbances such as noise or B1 inaccuracy.

Secondly, the overestimation in permittivity may originate from conversion of the reconstructed γc to conductivity and permittivity. As shown in Fig. 13, for pixels with larger conductivity values, the same value of the real part of γc corresponds to two very different permittivity values. The accuracy of reconstructed permittivity is thus highly dependent on the imaginary part of γc, which is unfortunately insensitive to permittivity change in this range. For example, two data points shown in Fig. 13 have the same conductivity value of 2.2 S/m and the same real part of γc. Despite their significant difference in permittivity (142ε0), there is only 21% difference in the imaginary part of γc compared to its dynamic range. Therefore, solving for γc instead of εc can result in unreasonably larger value of permittivity.

Fig. 13.

Fig. 13.

Projection maps from conductivity and permittivity to complex admittance. Shown are the normalized real (a) and imaginary (b) parts of admittance as functions of conductivity and permittivity, contoured by red lines with a step size of 0.05.

Lastly, reconstruction error on contrast boundary may perfuse to the whole image due to the nature of partial differential equation. Spatially constrained regularization may mitigate this effect but it cannot be eliminated.

Contributor Information

Yicun Wang, Department of Biomedical Engineering, University of Minnesota, MN 55455, USA..

Pierre-Francois Van de Moortele, Center for Magnetic Resonance Research, University of Minnesota, MN 55455, USA..

Bin He, Department of Biomedical Engineering, University of Minnesota, MN 55455, USA; and is with the Department of Biomedical Engineering, Carnegie Mellon University, PA 15213, USA. (bhe1@andrew.cmu.edu)..

REFERENCES

  • [1].Haacke EM, Petropoulos LS, Nilges EW, and Wu DH, “Extraction of conductivity and permittivity using magnetic resonance imaging,” Phys. Med. Biol, vol. 36, no. 6, pp. 723–734, June 1991. [Google Scholar]
  • [2].Wen H, “Noninvasive quantitative mapping of conductivity and dielectric distributions using RF wave propagation effects in high-field MRI,” in Proc. SPIE, San Diego, USA, 2003, pp. 471–477. [Google Scholar]
  • [3].Katscher U, Voigt T, Findeklee C, Vernickel P, Nehrke K, and Dossel O, “Determination of electric conductivity and local SAR via B1 mapping,” IEEE Trans. Med. Imaging, vol. 28, no. 9, pp. 1365–1374, September 2009. [DOI] [PubMed] [Google Scholar]
  • [4].Zhang X, Zhu S, and He B, “Imaging electric properties of biological tissues by RF field mapping in MRI,” IEEE Trans. Med. Imaging, vol. 29, no. 2, pp. 474–481, February 2010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [5].Shin J, Kim MJ, Lee J, Nam Y, Kim M, Choi N, Kim S, and Kim DH, “Initial study on in vivo conductivity mapping of breast cancer using MRI,” J. Magn. Reson. Imaging, vol. 42, no. 2, pp. 371–378, August 2015. [DOI] [PubMed] [Google Scholar]
  • [6].Liu J, Shao Q, Wang Y, Adriany G, Bischof J, Van de Moortele P-F, and He B, “In vivo imaging of electrical properties of an animal tumor model with an 8-channel transceiver array at 7T using electrical properties tomography,” Magn. Reson. Med, vol. 78, no. 6, pp. 2157–2169, January 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [7].Tha KK, Katscher U, Yamaguchi S, Stehning C, Terasaka S, Fujima N, Kudo K, Kazumata K, Yamamoto T, Cauteren MV, and Shirato H, “Noninvasive electrical conductivity measurement by MRI: a test of its validity and the electrical conductivity characteristics of glioma,” Eur. Radiol, vol. 28, no. 1, pp. 348–355, January 2018. [DOI] [PubMed] [Google Scholar]
  • [8].Liao Y, Romanzetti S, Gras V, Huang D, and Shah NJ, “The dielectric properties of brain tissues: variation in electrical conductivity with tissue sodium concentration and tissue water content at 3T/4T,” in Proc. Intl. Soc. Mag. Reson. Med. 23, Toronto, Canada, 2015, p. 3302. [Google Scholar]
  • [9].Mandija S, de Bruin P, Webb A, Luijten P, and van den Berg C, “Investigating the relation between electrical conduction and tissue composition with proton and sodium MRI,” in Proc. Intl. Soc. Mag. Reson. Med. 25, Honolulu, USA, 2017, p. 3639. [Google Scholar]
  • [10].Vaughan J, Garwood M, Collins C, Liu W, DelaBarre L, Adriany G, Andersen P, Merkle H, Goebel R, Smith M. b., and Ugurbil K, “7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head images,” Magn. Reson. Med, vol. 46, no. 1, pp. 24–30, July 2001. [DOI] [PubMed] [Google Scholar]
  • [11].Zhang X, Schmitter S, Van de Moortele P, Liu J, and He B, “From complex mapping to local SAR estimation for human brain MR imaging using multi-channel transceiver coil at 7T,” IEEE Trans. Med. Imaging, vol. 32, no. 6, pp. 1058–1067, June 2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [12].Liu J, Wang Y, Katscher U, and He B, “Electrical properties tomography based on B1 maps in MRI: principles, applications, and challenges,” IEEE Trans. Biomed. Eng, vol. 64, no. 11, pp. 2515–2530, November 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [13].Katscher U and van den Berg CAT, “Electric properties tomography: biochemical, physical and technical background, evaluation and clinical applications,” NMR Biomed, vol. 30, no. 8, p. e3729, August 2017. [DOI] [PubMed] [Google Scholar]
  • [14].Ropella KM and Noll DC, “A regularized, model-based approach to phase-based conductivity mapping using MRI,” Magn. Reson. Med, vol. 78, no. 5, pp. 2011–2021, December 2016. [DOI] [PubMed] [Google Scholar]
  • [15].Hafalir FS, Oran OF, Gurler N, and Ider YZ, “Convection-Reaction Equation Based Magnetic Resonance Electrical Properties Tomography (cr-MREPT),” IEEE Trans. Med. Imaging, vol. 33, no. 3, pp. 777–793, March 2014. [DOI] [PubMed] [Google Scholar]
  • [16].Gurler N and Ider YZ, “Gradient-based electrical conductivity imaging using MR phase,” Magn. Reson. Med, vol. 77, no. 1, pp. 137–150, January 2017. [DOI] [PubMed] [Google Scholar]
  • [17].Balidemaj E, van den Berg C, Trinks J, Lier A, Nederveen A, Stalpers L, Crezee H, and Remis R, “CSI-EPT: A Contrast Source Inversion Approach for Improved MRI-Based Electric Properties Tomography,” IEEE Trans. Med. Imaging, vol. 34, no. 9, pp. 1788–1796, February 2015. [DOI] [PubMed] [Google Scholar]
  • [18].Schmidt R and Webb A, “A new approach for electrical properties estimation using a global integral equation and improvements using high permittivity materials,” J. Magn. Reson, vol. 262, pp. 8–14, January 2016. [DOI] [PubMed] [Google Scholar]
  • [19].Serralles J, Polimeridis A, Vaidya M, Haemer G, White J, Sodickson D, Daniel L, and Lattanzi R, “Global Maxwell Tomography: a novel technique for electrical properties mapping without symmetry assumptions or edge artifacts,” in Proc. Intl. Soc. Mag. Reson. Med. 24, Singapore, 2016, p. 2993. [Google Scholar]
  • [20].Katscher U, Findeklee C, and Voigt T, “B1-based specific energy absorption rate determination for nonquadrature radiofrequency excitation,” Magn. Reson. Med, vol. 68, no. 6, pp. 1911–1918, December 2012. [DOI] [PubMed] [Google Scholar]
  • [21].Zhang X, de Moortele P-FV, Schmitter S, and He B, “Complex B1 mapping and electrical properties imaging of the human brain using a 16-channel transceiver coil at 7T,” Magn. Reson. Med, vol. 69, no. 5, pp. 1285–1296, May 2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [22].Sodickson D, Alon L, Deniz C, Ben-Eliezer N, Cloos M, Sodickson L, Collins C, Wiggins G, and Novikov D, “Generalized Local Maxwell Tomography for Mapping of Electrical Property Gradients and Tensors,” in Proc. Intl. Soc. Mag. Reson. Med. 21, Salt Lake City, USA, 2013, p. 3372. [Google Scholar]
  • [23].Liu J, Zhang X, Schmitter S, Van de Moortele P-F, and He B, “Gradient-based electrical properties tomography (gEPT): A robust method for mapping electrical properties of biological tissues in vivo using magnetic resonance imaging,” Magn. Reson. Med, vol. 74, no. 3, pp. 634–646, September 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [24].Wang Y, Van de Moortele P-F, and He B, “CONtrast Conformed Electrical Properties Tomography (CONCEPT) Based on Multi-Channel Transmission,” in Proc. Intl. Soc. Mag. Reson. Med. 25, Honolulu, USA, 2017, p. 3640. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [25].Yarnykh VL, “Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field,” Magn. Reson. Med, vol. 57, no. 1, pp. 192–200, January 2007. [DOI] [PubMed] [Google Scholar]
  • [26].Sacolick LI, Wiesinger F, Hancu I, and Vogel MW, “B1 mapping by Bloch-Siegert shift,” Magn. Reson. Med, vol. 63, no. 5, pp. 1315–1322, May 2010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [27].Nehrke K and Börnert P, “DREAM—a novel approach for robust, ultrafast, multislice B1 mapping,” Magn. Reson. Med, vol. 68, no. 5, pp. 1517–1526, November 2012. [DOI] [PubMed] [Google Scholar]
  • [28].Liu J, de Moortele PFV, Zhang X, Wang Y, and He B, “Simultaneous Quantitative Imaging of Electrical Properties and Proton Density From Maps Using MRI,” IEEE Trans. Med. Imaging, vol. 35, no. 9, pp. 2064–2073, September 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [29].Mandija S, Sbrizzi A, Katscher U, Luijten PR, and van den Berg CAT, “Error analysis of helmholtz-based MR-electrical properties tomography,” Magn. Reson. Med, vol. 80, no. 1, pp. 90–100, November 2017. [DOI] [PubMed] [Google Scholar]
  • [30].Boyd S, Parikh N, Chu E, Peleato B, and Eckstein J, “Distributed Optimization and Statistical Learning via the Alternating Direction Method of Multipliers,” Found Trends Mach Learn, vol. 3, no. 1, pp. 1–122, January 2011. [Google Scholar]
  • [31].Zhao B, Lu W, Hitchens TK, Lam F, Ho C, and Liang Z-P, “Accelerated MR parameter mapping with low-rank and sparsity constraints,” Magn. Reson. Med, vol. 74, no. 2, pp. 489–498, August 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [32].Akçakaya M, Nam S, Basha TA, Kawaji K, Tarokh V, and Nezafat R, “An Augmented Lagrangian Based Compressed Sensing Reconstruction for Non-Cartesian Magnetic Resonance Imaging without Gridding and Regridding at Every Iteration,” PLoS ONE, vol. 9, no. 9, p. e107107, September 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [33].Assländer J, Cloos MA, Knoll F, Sodickson DK, Hennig J, and Lattanzi R, “Low rank alternating direction method of multipliers reconstruction for MR fingerprinting,” Magn. Reson. Med, vol. 79, no. 1, pp. 83–96, January 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [34].Christ A, Kainz W, Hahn EG, Honegger K, Zefferer M, Esra Neufeld W. Rascher, Janka R, Bautz W, Chen J, Kiefer B, Schmitt P, Hollenbach Hans-Peter, Shen J, Oberle M, Szczerba D, Kam A, Guag JW, and Kuster N, “The Virtual Family—development of surface-based anatomical models of two adults and two children for dosimetric simulations,” Phys. Med. Biol, vol. 55, no. 2, p. N23, 2010. [DOI] [PubMed] [Google Scholar]
  • [35].Gabriel S, Lau RW, and Gabriel C, “The dielectric properties of biological tissues: III. Parametric models for the dielectric spectrum of tissues,” Phys. Med. Biol, vol. 41, no. 11, pp. 2271–2293, November 1996. [DOI] [PubMed] [Google Scholar]
  • [36].Ianniello C, de Zwart JA, Duan Q, Deniz CM, Alon L, Lee J-S, Lattanzi R, and Brown R, “Synthesized tissue-equivalent dielectric phantoms using salt and polyvinylpyrrolidone solutions,” Magn. Reson. Med, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [37].Kim HJ, Jeong WC, Sajib SZK, Kim M-O, Kwon OI, Je Woo E, and Kim D-H, “Simultaneous imaging of dual-frequency electrical conductivity using a combination of MREIT and MREPT,” Magn. Reson. Med, vol. 71, no. 1, pp. 200–208, 2014. [DOI] [PubMed] [Google Scholar]
  • [38].Van de Moortele PF, Snyder C, DelaBarre L, Adriany G, Vaughan JT, and Ugurbil K, “Calibration Tools for RF Shim at Very High Field with Multiple Element RF Coils: from Ultra Fast Local Relative Phase to Absolute Magnitude B1+ Mapping,” in Proc. Intl. Soc. Mag. Reson. Med. 15, Berlin, Germany, 2007, p. 1676. [Google Scholar]
  • [39].Van de Moortele P-F, Akgun C, Adriany G, Moeller S, Ritter J, Collins CM, Smith MB, Vaughan JT, and Ug K ˘urbil, “B1 destructive interferences and spatial phase patterns at 7 T with a head transceiver array coil,” Magn. Reson. Med, vol. 54, no. 6, pp. 1503–1518, December 2005. [DOI] [PubMed] [Google Scholar]
  • [40].Lee S-K, Bulumulla S, and Hancu I, “Theoretical investigation of random noise-limited signal-to-noise ratio in MR-based electrical properties tomography,” IEEE Trans. Med. Imaging, vol. 34, no. 11, pp. 2220–32, November 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [41].Marques JP, Sodickson DK, Ipek O, Collins CM, and Gruetter R, “Single acquisition electrical property mapping based on relative coil sensitivities: A proof-of-concept demonstration,” Magn. Reson. Med, vol. 74, no. 1, pp. 185–195, August 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [42].Knoll F, Bredies K, Pock T, and Stollberger R, “Second order total generalized variation (TGV) for MRI,” Magn. Reson. Med, vol. 65, no. 2, pp. 480–491, February 2011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [43].Orzada S, Maderwald S, Poser BA, Johst S, Kannengiesser S, Ladd ME, and Bitz AK, “Time-interleaved acquisition of modes: An analysis of SAR and image contrast implications,” Magn. Reson. Med, vol. 67, no. 4, pp. 1033–1041, 2012. [DOI] [PubMed] [Google Scholar]

RESOURCES