Abstract
Intensity-modulated radiotherapy and volumetric modulated arc therapy are modern radiation therapy technologies that can create the desired dose distribution by multileaf collimator movement and dose-rate control. However, the homogeneous dose delivery of small-field irradiation techniques shows disagreement with that of treatment planning system. The removal of the flattening filter by flattening filter free (FFF) beam irradiation increases dose conformity and further reduces treatment delivery time in radiosurgery. This study aims to investigate the dose distribution of FFF and flattened beams for small-field irradiation by using the 3D gel dosimeter. The N-isopropylacrylamide (NIPAM) polymer gel dosimeter was employed to record the 3D dose distribution. In addition, flattened and FFF beams were compared using the gamma evaluation technique. The use of an FFF accelerator resulted in excellent radiation treatments with short delivery times and low doses to normal tissues and organs. Results also showed that the passing rate increased with the decrease of field size (30 × 30, 20 × 20, and 10 × 10 mm2) at post-irradiation times of 24, 48, 72, and 96 h. The passing rates for each field size were retained at the same level when gamma criteria, namely, distance-to-agreement (DTA) = 3 mm/dose difference (DD) = 3%, were used. Nevertheless, the passing rates for each field size decreased slowly after 48 h when DTA = 2 mm/DD = 2%. The Wilcoxon signed-rank test was used to determine statistical difference with a significant level of p < 0.05. The passing rates of flattened and FFF beams showed no significant difference. The edge enhancement effect in the flattened beam was more evident than in the FFF beam. The 3D NIPAM gel dosimeter can be used for dose verification of small field for radiation therapy with high dose rate.
Introduction
The introduction of stereotactic body radiotherapy (SBRT) can provide acceptable control probabilities to patients [1]. This technique is a high-precision radiotherapy used for treating small to moderate extracranial tumors [2]. The SBRT may deliver a greater biologic dose of radiation, thus it can shorten the treatment period. Intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT), as modern radiation therapy technologies, can create the desired dose distribution by multileaf collimator (MLC) movement and dose-rate control. Vassiliev et al. [3] revealed that the FFF photon beam can reduce the out-of-field dose, because the removal of flattened filter can reduce the head scatter and softer spectra. Cashmore [4] also concluded similar results; the head scatter effect can be reduced approximately by 70% and reduction of out-of-field dose by 11% for a 6 MV FFF photon beam. Peguret et al. [5] adopted FFF beam for lung treatment and found that the treatment time was significantly reduced. Stieler et al. [6] used 6 MV standard vs. 6 MV FFF mode radiation beams for brain metastasis treatment. Their results showed that FFF mode treatment can reduce the treatment time to 51.5% with similar MU for VMAT and IMRT radiation. The gamma evaluation can achieve 99.08 ± 1.58% and 93.46 ± 2.41% with gamma criteria 3% dose difference (DD)/3 mm dose to agreement (DTA) and 3% DD/1 mm DTA, respectively. Navarria et al. [7] performed a clinical study to assess the outcome between 3DCRT with flattened beams (FF) and VMAT with FFF beams for SBRT. Both radiation modes can achieve dose conformity. Moreover, VMAT with FFF mode showed statistically significant reduction of ipsilateral lung doses and beam-on-time. In addition, local control rate was as high as 100% with FFF mode but was only 92.5% with FF beams (p = 0.03) at 1-year follow-up.
A region with a high dose gradient appears in the edge of radiation field because of the thin penumbra for the FFF mode. To obtain a true 3D dose distribution of FFF mode radiation, the gel dosimeter is a good choice. Olding et al. [8] adopted a 1D ion chamber, a 2D Gafchromic EBT3 film, and a 3D gel dosimeter to validate the two radiation modes of VMAT stereotactic ablative radiotherapy for a motion phantom. Both measurement dose distributions were consistent. However, this distribution also showed that the most disagreement of dose profile was located at the edge of high dose region. Pavoni et al. [9] proposed a composite gel-alanine phantom to validate a VMAT treatment of multiple brain metastases. The results were consistent with those of planning data. McErlean et al. [10] established a new 3D dosimetry combined with optical computed tomography (optical CT) to perform the dose verification of synchrotron microbeam radiation therapy (MRT) dose distribution. The optical CT and gel dosimeter can provide an excellent MRT visualization. The result can be useful for identifying potential treatment anomalies. Waldenberg et al. [11] adopted the NIPAM gel dosimeter combined with magnetic resonance imaging to investigate the dose dependence of the gel dosimeter. An important observation from their result was that the dose response of the gel is dependent on sequential beam irradiation. In other words, a high dose rate gives a low dose response. However, the sequential irradiation dependence is less pronounced if the time between the sequential beams had been reduced. Yao et al. [12] showed that the treatment planning system (TPS) map can achieve higher conformity to the planning target volume than that of IMRT. Three-dimensional (3D) gel dosimetry was used to validate the dose distribution agreement of VMAT, and the results showed good performance [12]. The passing rate can be remarkably high at approximately 95% agreement compared with that of TPS, with 3%/3 mm criterion [13,14]. The dose verification of single square-field and multifield irradiation of IMRT was also performed by the 3D gel dosimeters [15]. Nevertheless, the homogeneous dose delivery revealed disagreement with TPS when using small-field irradiation techniques [16,17]. The flattening filter free (FFF) beam irradiation increased the dose conformity and further reduced the treatment delivery time in radiosurgery by removing the flattening filter [1,18,19]. When performing stereotactic ablative body radiotherapy, in which the intrafraction organ/patient motion may affect the delivery efficacy, treatment time may be an important issue. Therefore, this study aims to investigate the dose distribution of FFF and flattened beams for small-field irradiation by using a 3D gel dosimeter. FFF and flattened beams were also compared using gamma evaluation. We utilized a simple square field irradiation in the current study because of the sequential irradiation dependence.
Materials and methods
Gel preparation
N-isopropylacrylamide (NIPAM) polymer gel was used in the current study. The NIPAM polymer gel dosimeters were prepared by a method similar to that described by Senden et al. [20]. Nonetheless, the original formulation was modified to obtain high sensitivity and linear gel recipe [21]. As shown in Table 1, the gel recipe includes 5% gelatin (300 Bloom Type A, Sigma-Aldrich, St. Louis, MO, USA), 5% NIPAM (97% pure; Wako, Osaka, Japan), 3% N,N′-methylene bisacrylamide (Bis, Sigma-Aldrich, St. Louis, MO, USA), and 5 mM tetrakis (hydroxymethyl) phosphonium chloride (THPC, TCI, Sigma-Aldrich, St. Louis, MO, USA). In the beginning, the gel phantom was heated to 45°C gradually in a water bath and stirred continously until the gelation solution became clear and transparent. Then, 3 wt% Bis was added and dissolved at 45°C by continuous stirring for 15 min. Subsequently, 5 wt% monomer was added and dissolved at 45°C and kept stirring for approximately 15 min. Finally, 5 mM THPC was added to the solution and dissolved at 42°C by continuous stirring for 2 min until the resulting gel finally became clear and transparent. The acrylic phantom with the dimension of 10 cm in diameter, 10 cm in height, and 3 mm in wall thickness, was filled with gel. Then, the gel phantom was immersed in a refrigerator with a fixed temperature of 6°C for at least 6 h until complete solidification. Further details on the gel preparation are provided in the work of Hsieh et al. [22]. To avoid the oxygen effect on the gel polymerization, the acrylic phantom cap was sealed with parafilm to prevent oxygen from entering into the phantom. The entire gel phantom was wrapped entirely by an aluminum foil to prevent photopolymerization.
Table 1. Gel recipe used in this study [11].
| Composition | Weight (%) | Amounta |
|---|---|---|
| Gelatin | 5 | 5±0.001 g |
| N-isopropylacrylamide (NIPAM) | 5 | 5±0.001 g |
| N,Nʹ-methylene bisacrylamide | 3 | 3±0.001 g |
| Tetrakis (hydroxymethyl) phosphonium chloride | 5 | 8.96×10−2±1×10−3 mL |
| Distilled water | 87 | 87±0.1 mL |
aThe composition was determined according to the preparation of 100 mL NIPAM dosimeter.
Irradiation and TPS
This study adopted Elekta Versa HD linear accelerator (Elekta AB, Sweden), which was provided by Changhua Christian Hospital. This accelerator provided two modes of irradiation including the flattened photon beam mode and the FFF photon beam mode. The gel phantom was aligned to the center line of the treatment head. The pinnacle TPS (Pinnacle3 version 9.8, Philips, USA) was used. The pinnacle TPS for small-field irradiation was calibration by using a beam data obtained from small field 10 × 10 mm2 and 20 × 20 mm2 radiation, which was measured by a pinpoint ion chamber (0.015c.c)(PinPoint Ionization Chambers 30014, PTW, Germany). The target dose was 5 Gy at 5 cm depth of the NIPAM gel dosimeter. The source-to-surface distance was 95 cm, and the gantry angle was 0°. The dose rate in the FF mode was 600 MU/min, with 6 MV photon energy. In the FFF mode, the dose rate was 1,400 MU/min, with 6 MV photon energy. The field sizes were 30 × 30, 20 × 20, and 10 × 10 mm2. The parameters of each batch are shown in Table 2. Eighteen gel dosimeters were prepared for three field sizes and two irradiation modes. Eighteen batches of gel phantoms were prepared.
Table 2. Parameters of each batch.
| Irradiation mode | Flattened photon beam | Flattening filter free (FFF) photon beam | ||||
|---|---|---|---|---|---|---|
| Dose rate (MU/min) | 600 | 1400 | ||||
| Field size (FS, mm2) | 30×30 | 20×20 | 10×10 | 30×30 | 20×20 | 10×10 |
| Field mode | MLC | MLC | MLC | MLC | MLC | MLC |
| Monitor unit (MU) | 562 | 587 | 653 | 547 | 573 | 625 |
| Time of beam on (s) | 56.2 | 58.7 | 65.3 | 23.4 | 24.6 | 27.9 |
| Batch | 1, 13, 15 | 3, 7, 11 | 5, 9, 17 | 2, 14, 16 | 4, 8, 12 | 6, 10, 18 |
Gel phantom scanning and image reconstruction
We developed the proposed charge coupled device (CCD)-based parallel-beam optical-CT scanner (CT-s2) shown in Fig 1 [15,23]. The CT-s2 was used as dose readout tool after the polymer gel phantom was irradiated. The CT-s2 comprises (1) a laser, (2) spatial filter, (3) pinhole, (4) rotating diffuser, (5) collimating lens, (6) NIPAM gel phantom, (7) aquarium, (8) collimating lens, (9) CCD camera, and (10) stepping motor. The laser beam was converted into a plane light source by using a spatial filter, a microscope objective lens, and a pinhole. A significantly uniform parallel light source can be obtained by speckle reduction technique [24]. The gel phantom was mounted on the stepping-motor-driven rotating table immersed in an aquarium. The aquarium filled with matching liquid with best-fit refractive index of 1.346 was used to reduce the refraction and reflection effects at the interface between materials with different refractive indices [25]. Totally, 360 projection images were acquired by the CCD camera. Then, the image reconstruction was performed using a filtered back-projection algorithm with our proposed program written in MATLAB [15,23–25].
Fig 1. Image of the optical computed tomography scanner CT-s2.
1. laser, 2. spatial filter, 3. pinhole, 4. rotating diffuser, 5. collimating lens, 6. NIPAM gel, 7. aquarium, 8. collimating lens, 9. CCD camera, and 10. stepping motor.
Gamma evaluation
The quantitative gamma evaluation proposed by Low et al. [26,27] was used for performance analysis. Given that the acceptance criterion was varied in low- and high-gradient regions, direct comparison between the measured and calculated dose distribution values in various planes was unsuitable. In gamma evaluation, the measured and calculated doses are compared using a combination of a dose-difference (DD) criterion and a distance-to-agreement (DTA) criterion [26]. Point-by-point calculation was performed by comparing the dose distribution from the calculated TPS and that measured by the gel dosimeter. To perform a comparison, we used 2D level gamma calculation. After calculation, the gamma map and gamma passing rate can be obtained. In this study, two gamma criteria for gamma evaluation were used, namely, 3% DD/3 mm DTA and 2% DD/2 mm DTA [27]. The gamma passing rate was calculated in the percentage area with γ < 1.
Results and discussion
Spatial uniformity of the non-irradiated gel phantom
Differences in the temperature history experienced by a normoxic polymer gel formulation after its fabrication may influence the spatial uniformity of the gel phantom [28,29]. Fig 2 shows the scanning results of the reference image of the gel phantom (Fig 2A). The image is a reconstruction of a transverse slice. Fig 2B presents the line profiles of the scanning results of the non-irradiated gel phantom. These profiles illustrated the variations in the optical densities at different depths of 30, 40, 45, 50, and 60 mm. The maximum standard deviations between different depths in the central 60 pixel/mm region were less than 0.01%. The maximum standard deviation for all 18 batches was less than 0.3%. This result indicated high spatial uniformity for these batches of non-irradiated NIPAM gel dosimeters.
Fig 2.
(a) Reference image of the gel phantom. (b) Line profiles of the gel phantom at depths of 30, 40, 45, 50, and 60 mm.
Comparison of gamma passing rates between various irradiation modes
A quantitative gamma evaluation was performed between the dose map calculated from TPS and that measured by an optical-CT scanner. Table 3 shows the average gamma passing rate and standard deviation with gamma criteria of DTA = 3 mm/DD = 3% for various irradiation modes and FS at various depths and different post-irradiation periods. At FS = 30 × 30 mm2 and post-irradiation time of 24 h, 12 gamma passing rates corresponding to three batches of the NIPAM gel dosimeter and four depths at 30, 40, 50, and 60 mm were calculated. The means and standard deviations of flattened and FFF beams at post-irradiation time of 24 h were 97.41 ± 1.58 and 97.49 ± 1.95, respectively. The gamma passing rates were almost the same for the two types of irradiation mode. At FS = 30′30 mm2 and post-irradiation times of 48, 72, and 96 h, the means and standard deviations of gamma passing rates for flattened and FFF beams were 95.44 ± 2.25 and 96.30 ± 2.43, 96.77 ± 2.42 and 95.95 ± 2.12, and 96.31 ± 2.26 and 96.75 ± 2.47, respectively. The overall means and standard deviations of gamma passing rates for flattened and FFF beams were 96.48 ± 2.29 and 96.62 ± 2.35, respectively. We used the Wilcoxon signed-rank test to determine whether or not the passing rates of flattened and FFF beams show significant statistical difference. This non-parametric statistical hypothesis test is used to compare two related samples, matched samples, or repeated measurements on a single sample to assess differences in their population mean ranks. The Wilcoxon signed-rank test can be used as an alternative to the paired t-test, t-test for matched pairs, or the t-test for dependent samples when the population is assumed to be not normally distributed [30]. p < 0.05 was considered statistically significant. The passing rates of flattened and FFF beams were not significantly different (p = 0.566). The corresponding p-values when FS were 20 × 20 and 10 × 10 mm2 are shown in Table 3. The same conclusion can be obtained for two irradiation modes. Previous studies [31,32] reported similar findings and revealed that the dosimetric data show no statistically significant difference (p > 0.05) for the target coverage of flattened and FFF beams at 6 MV energy.
Table 3. Average gamma passing rate and standard deviation of various irradiation modes and FS at different post-irradiation periods with distance-to-agreement (DTA) of 3 mm and dose distance (DD) of 3%.
| Post-irradiation time | 24 h | 48 h | 72 h | 96 h | Overall mean ±SD |
Wilcoxon Test (p-value*) |
|||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Irradiation mode | FS = 30×30 mm2 | ||||||||||
| Flattened photon beam | 97.41 | ±1.58 | 95.44 | ±2.25 | 96.77 | ±2.42 | 96.31 | ±2.26 | 96.48 | ±2.29 | 0.566 |
| FFF photon beam | 97.49 | ±1.95 | 96.30 | ±2.43 | 95.95 | ±2.12 | 96.75 | ±2.47 | 96.62 | ±2.35 | |
| Irradiation mode | FS = 20×20 mm2 | ||||||||||
| Flattened photon beam | 98.40 | ±1.24 | 98.45 | ±1.69 | 97.32 | ±1.86 | 97.69 | ±2.30 | 97.97 | ±1.90 | 0.622 |
| FFF photon beam | 98.01 | ±1.64 | 98.15 | ±1.15 | 97.34 | ±1.11 | 98.07 | ±0.87 | 97.89 | ±1.28 | |
| Irradiation mode | FS = 10×10 mm2 | ||||||||||
| Flattened photon beam | 100 | ±0.0 | 100 | ±0.0 | 100 | ±0.0 | 100 | ±0.0 | 100 | ±0.0 | No difference exists. |
| FFF photon beam | 100 | ±0.0 | 100 | ±0.0 | 100 | ±0.0 | 100 | ±0.0 | 100 | ±0.0 | |
*p > 0.05 represents no significant difference.
Gamma criteria of DTA = 2 mm/DD = 2% were used to calculate the passing rates, and the results are shown in Table 4. The passing rates with DTA = 2 mm/DD = 2% were lower than that with DTA = 3 mm/DD = 3%. The overall means and standard deviations of flattened and FFF beams at FS of 30 × 30, 20 × 20, and 10 × 10 mm2 were 85.11 ± 3.45 and 86.14 ± 4.85, 89.59 ± 4.11 and 88.88 ± 3.82, and 94.94 ± 2.12 and 94.73±2.14, respectively. When the FS was 30 × 30, 20 × 20, and 10 × 10 mm2, the corresponding p-values were 0.272, 0.406, and 0.626, respectively. For both gamma criteria DTA = 3 mm/DD = 3% and DTA = 2 mm/DD = 2%, the passing rates of flattened and FFF beams showed no significant difference (p-value > 0.05).
Table 4. Average gamma passing rate and standard deviation of various irradiation modes and FS at different post-irradiation periods with DTA = 2 mm/DD = 2%.
| Post-irradiation time | 24 h | 48 h | 72 h | 96 h | Overall mean ±SD |
Wilcoxon Test (p-value*) |
|||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Irradiation mode | FS = 30×30 mm2 | ||||||||||
| Flattened photon beam | 85.18 | ±4.10 | 84.22 | ±3.46 | 83.85 | ±2.88 | 81.06 | ±2.61 | 85.11 | ±3.45 | 0.272 |
| FFF photon beam | 85.05 | ±4.83 | 85.30 | ±4.78 | 82.75 | ±4.64 | 84.47 | ±4.63 | 86.14 | ±4.85 | |
| Irradiation mode | FS = 20×20 mm2 | ||||||||||
| Flattened photon beam | 90.65 | ±2.50 | 90.68 | ±3.37 | 87.68 | ±4.61 | 88.32 | ±4.73 | 89.59 | ±4.11 | 0.406 |
| FFF photon beam | 89.54 | ±4.07 | 89.20 | ±3.04 | 88.17 | ±4.68 | 88.64 | ±2.91 | 88.88 | ±3.82 | |
| Irradiation mode | FS = 10×10 mm2 | ||||||||||
| Flattened photon beam | 95.76 | ±1.71 | 94.60 | ±2.21 | 94.82 | ±2.24 | 94.57 | ±1.96 | 94.94 | ±2.12 | 0.626 |
| FFF photon beam | 95.26 | ±2.05 | 94.99 | ±2.16 | 94.29 | ±2.27 | 94.41 | ±1.83 | 94.73 | ±2.14 | |
*p > 0.05 represents no significant difference.
Comparison of temporal stability at various FS
As shown in Fig 3, the gamma passing rates of flattened and FFF beams at various post-irradiation periods and FS were plotted. The solid line represents the flattened beam, and the dashed line represents the FFF beam. The passing rate increased when the FS decreased from 30 × 30 mm2 to 10 × 10 mm2 at post-irradiation times of 24, 48, 72, and 96 h. For each FS, the passing rates retained the same level with DTA = 3 mm/DD = 3%. However, for each FS, the passing rates decreased slowly after 48 h when DTA = 2 mm/DD = 2%. Similar phenomenon was reported by Chang [16], who used an FS of 40 × 40 mm2.
Fig 3.
Gamma passing rate curve of various irradiation modes and FS at different post-irradiation periods with two gamma criteria: (a) DTA = 3 mm/DD = 3% and (b) DTA = 2 mm/DD = 2%. The solid line represents the flattened beam, and the dashed line represents the FFF beam.
Comparison of gamma map between various irradiation modes and FS
Figs 4 and 5 show the gamma maps of the flattened and FFF beams at 24 h after irradiation at a depth of 50 mm. Fig 4 displays that the rejected area, where gamma value γ > 1, was mostly located at the edge of the irradiation field for DTA = 3 mm/DD = 3% (Fig 6) and DTA = 2 mm/DD = 2% (Fig 5). The edge of the irradiation field is generally a high-dose gradient region, where edge enhancement may occur [33]. Furthermore, the rejected areas decreased when the FS decreased from 30 × 30 mm2 to 10 × 10 mm2 for both gamma criteria. Nevertheless, the edge enhancement effect was highly evident when the gamma criteria of 2 mm/2% were adopted (Fig 5). The FFF beam with a high dose rate (1,400 MU/min) showed rejected regions similar to those of the flattened beam with a low dose rate (600 MU/min). Therefore, the passing rates of the flattened and FFF beams presented no significant difference.
Fig 4. Various FS of the gamma map of the NIPAM gel dosimeter at a depth of 50 mm for 24 h post-irradiation time.
The gamma criteria used were DTA = 3 mm/DD = 3%. The left image represents the flattened beam, and the right image represents the FFF beam. (a) FS = 30 × 30 mm2, (b) FS = 20 × 20 mm2, and (c) FS = 10 × 10 mm2.
Fig 5. Various FS of the gamma map of NIPAM gel dosimeter at a depth of 50 mm for 24 h post-irradiation time.
The gamma criteria used were DTA = 2 mm/DD = 2%. The left image represents the flattened beam, and the right image represents the FFF beam. (a) FS = 30 × 30 mm2, (b) FS = 20 × 20 mm2, and (c) FS = 10 × 10 mm2.
Fig 6. Comparison of the 100%, 90%, 80%,70%, 60%, and 50% isodose lines between the gel measurement (dotted lines) and treatment planning system (solid lines) at a depth of 50 mm for 24 h post-irradiation time.
The left image represents the flattened beam, and the right image represents the FFF beam. (a) FS = 30 × 30 mm2, (b) FS = 20 × 20 mm2, and (c) FS = 10 × 10 mm2.
Comparison of dose distributions and isodose lines
A comparison of the percentage isodose lines of NIPAM gel dosimeters at a depth of 50 mm and post-irradiation time of 24 h is shown in Fig 6. The dose percentages of 100%, 90%, 80%, 70%, 60%, and 50% of the gel measurement (dotted lines) and TPS (solid lines) for various FS are illustrated in the plots. The isodose maps of flattened and FFF beams were in the left and right panels of Fig 6, respectively. A comparison of the isodose lines of the TPS and the measured dose can be used to investigate the origin of dose deviations. The edge of the irradiation field is generally a high-dose gradient region and may cause edge enhancement effect in the gel dosimeter [33]. Gels are prone to induce measurement errors at high-dose gradient regions [34]. Free radical diffusion may occur near the high dose gradient regions and cause dose deviations. Fig 6 illustrates high consistency for small FS between the TPS and the measured dose at dose percentages of 50%–100%. These results showed that the edge enhancement effect in the large FS was more evident than in small FS. The isodose line showed excellent consistency between the measured data and TPS calculated data, even in high-dose gradient region near the edge of the irradiation field at FS of 10 × 10 mm2. Evidently, the penumbra in the flattened beam (left panel) was thicker than in the FFF beam (right panel). The edge enhancement effect in the flattened beam was also more evident than in the FFF beam. Previous research [34] also reported that filter removal can significantly reduce the amount of head scatter photons, thereby reducing doses to normal tissues and organs. Treatments can also benefit from an increased dose rate.
Discussion
In current study, we performed the dose comparison between TPS and measured data from NIPAM gel dosimeter. In order to verify the dose accuracy of the Pinnacle3 TPS, we have conducted a commissioning procedure for Pinnacle3 TPS before experiment. As mentioned in “Irradiation and TPS” section, calibration was conducted by using the beam data obtained from small field 10 × 10 mm2 and 20 × 20 mm2 radiation, which was measured by a pinpoint ion chamber (0.015c.c) (PinPoint Ionization Chambers 30014, PTW, Germany). It ensured the comparison of the gel dosimeter and TPS was meaningful.
The passing rates between flattened and FFF beams for various irradiation FS with two gamma criteria were compared. Statistical analysis indicated no significant difference in passing rates between the flattened and FFF beams. Thus, the dose distributions irradiated by the flattened and FFF beams were the same, but the time of the FFF beam can be reduced to half of that of the flattened beam, as shown in Table 2. Therefore, the patients can reduce their unsatisfactory experiences during radiation therapy. Table 2 shows that the dose rate of FFF beam was 2.33 times higher than that of the flattened beam. Initially, we considered that severe edge enhancement may be observed in the FFF beam because high dose rate may induce fast generation of free radicals. However, the edge enhancement effect on the flattened beam was more evident than on the FFF beam. Mcauley [33] proposed simple free-radical solution polymerization systems, which indicated that when new radicals are not generated, the rate of polymerization quickly decreases to zero, because no free radicals exist. The time of the FFF beam was only half of that of the flattened beam in the current study. The similar results can be found in previous research [5,6]. Therefore, the polymerization time for FFF beam is much shorter than that of the flattened beam, thereby causing a small amount of excess polymer generated near the edge. In previous research, Olding et al. [8] claimed that the most disagreement of dose profile was located at the edge of high dose region. Moreover, limited polymerization time reduces the diffusion time of polymer. This condition results in less edge enhancement for the FFF beam. Previous results [22] also suggested that the amount of additional polymer near the edge increases with increased post-irradiation time. The same trend can be observed on the change in passing rates with post-irradiation time, as shown in Table 4 and Fig 3.
Conclusion
Quantitative dosimetric comparison of FF and FFF radiation therapy was conducted using gamma evaluation. Three batches of NIPAM gel dosimeters for two irradiation modes (flattened and FFF beams) and three FS (30 × 30, 20 × 20, and 10 × 10 mm2) were prepared. Eighteen batches of NIPAM gel dosimeters were analyzed. The gamma passing rates at four depths and four post-irradiation time periods were calculated. The overall means and standard deviations of gamma passing rates with DTA = 3 mm/DD = 3% for flattened and FFF beams were 96.48 ± 2.29 and 96.62 ± 2.35, 97.97 ± 1.90 and 97.89 ± 1.28, and 100 ± 0.0 and 100 ± 0.0 for FS of 30 × 30, 20 × 20, and 10 × 10 mm2, respectively. In addition, the overall means and standard deviations of gamma passing rates with DTA = 2 mm/DD = 2% for flattened and FFF beams were 85.11 ± 3.45 and 86.14 ± 4.85, 89.59 ± 4.11 and 88.88 ± 3.82, and 94.94 ± 2.12 and 94.73 ± 2.14 for FS of 30 × 30, 20 × 20, and 10 × 10 mm2, respectively. The paired sample t-test for the two irradiation modes showed no significant difference between the flattened and FFF beams at 6 MV energy. The gamma maps illustrated that low passing rates were obtained from the edge of the irradiation regions, where high-dose gradient was located. The present results revealed that the 3D NIPAM gel dosimeter can be used for dose verification of small FS for radiation therapy with high dose rate. Therefore, the 3D NIPAM gel dosimeter can be potentially used for clinical dose verification of modern radiation technologies.
Acknowledgments
This study was supported by the Ministry of Science and Technology, Executive Yuan, Taipei, Taiwan (grant number: MOST 106-2314-B-035-001, website: https://www.most.gov.tw/?l=en), China Medical University (grant number: CMU107-S-12 and CMU106-S-04, website: http://english.cmu.edu.tw/), and Changhua Christian Hospital (grant number: CTU106-CH-006, website: http://www2.cch.org.tw/cch_english/).
Data Availability
All relevant data are within the paper.
Funding Statement
This study was supported by the Ministry of Science and Technology, Executive Yuan, Taipei, Taiwan (grant number: MOST 106-2314-B-035-001, website: https://www.most.gov.tw/?l=en) and Changhua Christian Hospital (grant number: CTU106-CH-006, website: http://www2.cch.org.tw/cch_english/). The funders had no role in the study design, data collection, analysis, decision to publish, or preparation of the manuscript.
References
- 1.Navarria P, Ascolese A, Mancosu P, Alongi F, Clerici E, Tozzi A, et al. Volumetric modulated arc therapy with flattening filter free (FFF) beams for stereotactic body radiation therapy (SBRT) in patients with medically inoperable early stage non small cell lung cancer (NSCLC). Radiother Oncol. 2013; 107:414–418. 10.1016/j.radonc.2013.04.016 [DOI] [PubMed] [Google Scholar]
- 2.Nagata Y, Takayama K, Matsuo Y, Norihisa Y, Mizowaki T, Sakamoto T, et al. Clinical outcomes of a phase I/II study of 48 Gy of stereotactic body radiotherapy in 4 fractions for primary lung cancer using a stereotactic body frame. Int J Radiat Oncol Biol Phys. 2005; 63:1427–31. 10.1016/j.ijrobp.2005.05.034 [DOI] [PubMed] [Google Scholar]
- 3.Cashmore J. The characterization of unflattened photon beams from a 6 MV linear accelerator. hys. Med. Biol. 2008; 53(7):1933. [DOI] [PubMed] [Google Scholar]
- 4.Peguret N, Dahele M, Cuijpers JP, Slotman BJ, Verbakel WFAR. Frameless high dose rate stereotactic lung radiotherapy: Intrafraction tumor position and delivery time. Radiother Oncol. 2013;107:419–22. 10.1016/j.radonc.2013.04.019 [DOI] [PubMed] [Google Scholar]
- 5.Stieler F, Fleckenstein J, Simeonova A, Wenz F, Lohr F. Intensity modulated radiosurgery of brain metastases with flattening filter-free beams, Radiother Oncol. 2013;109:448–451. [DOI] [PubMed] [Google Scholar]
- 6.Navarria P, Ascolese AM, Mancosu P, Alongi F, Clerici E, Tozzi A, et al. Volumetric modulated arc therapy with flattening filter free (FFF) beams for stereotactic body radiation therapy (SBRT) in patients with medically inoperable early stage non small cell lung cancer (NSCLC), Radiother Oncol. 2013;107:414–418. [DOI] [PubMed] [Google Scholar]
- 7.Ferrari A, Fasso A, Ranft J, Sala PR. FLUKA:AMulti-Particle Transport Code 2005; CERN, Geneva, Switzerland. [Google Scholar]
- 8.Olding T, Alexander KM, Jechel C, Nasr AT and Joshi C. Delivery validation of VMAT stereotactic ablative body radiotherapy at commissioning. J Phys Conf Ser. 2015;573:012019. [Google Scholar]
- 9.Pavoni F, Neves-Junior WFP, Silveira MA, Ramos PAMM, McErlean CM, Bräuer-Krisch E, et al. Assessment of optical CT as a future QA tool for synchrotron x-ray microbeam therapy. Phys. Med. Biol. 2016;61:320–337. 10.1088/0031-9155/61/1/320 [DOI] [PubMed] [Google Scholar]
- 10.Waldenberg C, Hauer AK, Gustafsson C and Ceberg S. Dose integration and dose rate characteristics of a NiPAM polymer gel MRI dosimeter system, J Phys Conf Ser. 2017;847:012063. [Google Scholar]
- 11.Yao CH, Chang DH, Tsai MJ, Lai YC, Chen YA, Chang YJ, et al. Dose verification of volumetric modulation arc therapy by using a NIPAM gel dosimeter combined with a parallel-beam optical computed tomography scanner. J Radioanal Nucl Chem. 2017; 311:1277–1286. [Google Scholar]
- 12.Hayashi N, Malmin RL, Watanabe Y. Dosimetric verification for intensity-modulated arc therapy plans by use of 2D diode array, radiochromic film and radiosensitive polymer gel. J Radiat Res. 2014; 55:541–552. 10.1093/jrr/rrt139 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Olding T, Alexander KM, Jechel C, Nasr AT, Joshi C. Delivery validation of VMAT stereotactic ablative body radiotherapy at commissioning. J Phys Conf Ser. 2015; 573:012019. [Google Scholar]
- 14.Huang WT, Chen CH, Hung CN, Tuan CC, Chang YJ. Implementation of a parallel-beam optical-CT apparatus for three-dimensional radiation dosimetry using a high-resolution CCD camera. Nucl Instrum Methods Phys Res, Sect A. 2015;784:590–596. [Google Scholar]
- 15.Chang YJ, Chen CH, Hsieh BT (2014) Characterization of long-term dose stability of N-isopropylacrylamide polymer gel dosimetry. J Radioannal Nucl Chem. 301:765–780 [Google Scholar]
- 16.Shih TY, Wu J, Shih CT, Lee YT, Wu SH, Yao CH, et al. Small-Field Measurements of 3D Polymer Gel Dosimeters through Optical Computed Tomography. PLoS One 2016;11(3):e0151300 10.1371/journal.pone.0151300 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Sangeetha S and Sureka C S. Comparison of Flattening Filter (Ff) and Flattening-Filter-Free (FFF) 6 mv photon beam characteristics for small field dosimetry using egsnrc monte carlo code. Radiat Phys Chem, 2017; 135:63–75. [Google Scholar]
- 18.Wiant DB, Terrell JA, Maurer JM, Yount C, Sintay BJ. Commissioning and validation of BrainLAB cones for 6X FFF and 10X FFF beams on a Varian TrueBeam STx. J. Appl. Clin Med Phys. 2013; 14(6): 293–306. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Senden RJ, De Jean P, McAuley KB and Schreiner LJ. Polymer gel dosimeters with reduced toxicity: a preliminary investigation of the NMR and optical dose-response using different monomers. Phys Med Biol. 2006; 51:3301–3314. 10.1088/0031-9155/51/14/001 [DOI] [PubMed] [Google Scholar]
- 20.Chang YJ, Hsieh BT, Liang JA. A systematic approach to determine optimal composition of gel used in radiation therapy. Nucl Instrum Methods Phys Res Sect A. 2011; 652:783–785 [Google Scholar]
- 21.Hsieh BT, Chang YJ, Han RP, Wu J, Hsieh LL, Chang CJ. A study on dose response of NIPAM-based dosimeter used in radiotherapy. J Radioanal Nucl Chem. 2011; 290:141–148 [Google Scholar]
- 22.Chang YJ. Development of CCD-based optical computed tomography and comparison with single-beam optical CT scanner. J Phys Conf Ser. 2015; 573:012060. [Google Scholar]
- 23.Chang YJ. Use of a speckle reduction technique to improve the reconstruction image quality of CCD-based optical computed tomography scanner. Nucl Instrum Methods Phys Res Sect A. 2015; 784:585–589 [Google Scholar]
- 24.Chen CH, Wu J, Hsieh BT, Chen DS, Wang TH, Chien SH, et al. Best fit refractive index of matching liquid for 3D NIPAM gel dosimeters using optical CT. Radiat Phys Chem 2014;104:192–197 [Google Scholar]
- 25.Low DA, Harms WB, Mutic S, Purdy JA. A technique for the quantitative evaluation of dose distributions. Med. Phys. 1998;25(5):656–661. 10.1118/1.598248 [DOI] [PubMed] [Google Scholar]
- 26.Low DA, Dempsey JF. Evaluation of the gamma dose distribution comparison method. Med. Phys. 2003;30(9):2455–2464. 10.1118/1.1598711 [DOI] [PubMed] [Google Scholar]
- 27.Hsieh BT, Wu J, Chang YJ. Verification on dose Profile variation of 3D polymer gel dosimeter. IEEE Trans Nucl Sci.2013; 60:560–565 [Google Scholar]
- 28.De Deene Y, Pittomvils G, Visalatchi S. The influence of cooling rate on the accuracy of normoxic polymer gel dosimeters. Phys Med Biol. 2007; 52:2719–2728 10.1088/0031-9155/52/10/006 [DOI] [PubMed] [Google Scholar]
- 29.Kumar L, Yadav G, Samuvel KR, Bhushan M, Kumar P, Suhail M, et al. Dosimetric influence of filtered and flattening filter free photon beam on rapid arc (RA) radiotherapy planning in case of cervix carcinoma, Rep Pract Oncol Radiother. 2017; 22:10–18 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Lowry R. Concepts & Applications of Inferential Statistics. Publication date: 31 Dec 2015, http://www.freetechbooks.com/concepts-applications-of-inferential-statistics-t1228.html, Retrieved 18 October 2018.
- 31.Barbiero S, Rink A, Matteucci F. Fedele D, Paiar F, Pasqualetti F, Avanzo M. Single-fraction flattening filter-free volumetric modulated arc therapy for lung cancer: dosimetric results and comparison with flattened beams technique, Med. Dosim. 2017; 41(4):334–338. [DOI] [PubMed] [Google Scholar]
- 32.McAuley KB. Fundamentals of Polymer Gel Dosimeters. J Phys Conf Ser. 2006; 56:35. [Google Scholar]
- 33.Massillon JG, Minniti R, Soares CG, Maryanski MJ and Robertson S. Characteristics of a new polymer gel for high-dose gradient dosimetry using a micro optical CT scanner. Appl Radiat Isot. 2010; 68:144–154. 10.1016/j.apradiso.2009.08.016 [DOI] [PubMed] [Google Scholar]
- 34.Vassiliev ON, Titt U, P¨onisch F, Kry SF, Mohan R and Gillin MT. Dosimetric properties of photon beams from a flattening filter free clinical accelerator. Phys Med Biol. 2006;51(7):1907–17 10.1088/0031-9155/51/7/019 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
All relevant data are within the paper.