Abstract
Focused ultrasound combined with microbubble for blood-brain barrier disruption (FUS-BBBD) is a promising technique for noninvasive and localized brain drug delivery. This study demonstrates that passive cavitation imaging (PCI) is capable of predicting the location and concentration of nanoclusters delivered by FUS-BBBD. During FUS-BBBD treatment of mice, the acoustic emissions from FUS-activated microbubbles were passively detected by an ultrasound imaging system and processed offline using a frequency-domain PCI algorithm. After the FUS treatment, radiolabeled gold nanoclusters, 64Cu-AuNCs, were intravenously injected into the mice and imaged by positron emission tomography/computed tomography (PET/CT). The centers of the stable cavitation dose (SCD) maps obtained by PCI and the corresponding centers of the 64Cu-AuNCs concentration maps obtained by PET coincided within 0.3 ± 0.4 mm and 1.6 ± 1.1 mm in the transverse and axial directions of the FUS beam, respectively. The SCD maps were found to be linearly correlated with the 64Cu-AuNCs concentration maps on a pixel-by-pixel level. These findings suggest that SCD maps can spatially “paint” the delivered nanocluster concentration, a technique that we named as cavitation dose painting. This PCI-based cavitation dose painting technique in combination with FUS-BBBD opens new horizons in spatially targeted and modulated brain drug delivery.
Introduction
Focused ultrasound in combination with microbubbles for blood-brain barrier disruption (FUS-BBBD) has been established as a promising technique for the delivery of therapeutic agents to a targeted brain location without invasive surgery1–3. The blood-brain barrier (BBB) excludes ~100% of large-molecule drugs and 98% of small-molecule drugs from entering the brain parenchyma4. FUS-BBBD utilizes the mechanical interactions between microbubbles and the surrounding blood vessels to enhance the BBB permeability, allowing various therapeutic agents in the systemic circulation to be delivered across the BBB to the brain tissue. Extensive preclinical studies have been performed to develop the FUS-BBBD technique, including evaluation of its feasibility and safety in smaller animal models (e.g., mice, rats, and rabbits)3,5–10 and larger animal models (e.g., pigs and non-human primates)11,12; optimization of treatment protocols13; assessment of therapeutic efficacy for various diseases5,6,9; and revealing its fundamental mechanisms14–16. Recently, a clinical trial successfully demonstrated the feasibility and safety of FUS-BBBD in patients with Alzheimer’s disease17. Despite these great advancements of the FUS-BBBD technique, there is an unmet need for a treatment monitoring technique that is capable of predicting the location and concentration of therapeutic agents delivered by FUS-BBBD. This capability is critically needed to verify localized drug delivery to the targeted brain location without off-target effects in the brain; assess whether the concentrations of the delivered agents reach the desired therapeutic levels; characterize variations among repeated treatments; and develop real-time feedback control of the FUS parameters to deliver the desired concentrations of drugs to the targeted locations.
Several existing imaging techniques have been used for the assessment of FUS-BBBD treatment outcome. Magnetic resonance imaging (MRI) is the most commonly used imaging modality for the assessment of FUS-mediated BBB permeability changes based on the leakage of MR contrast agents from the blood into the brain parenchyma. Positron emission tomography (PET) and single photon emission computed tomography (SPECT) in combination with radiolabeled therapeutic agents provide noninvasive, sensitive, and quantitative methods for directly assessing the spatial distribution of radiolabeled agents18–20. Several studies have used PET or SPECT to evaluate the spatial distributions of radiolabeled agents in the brain after FUS-BBBD treatment21–23. Our previous study verified that in vivo PET imaging can reliably estimate the concentrations of the 64Cu-labeled gold nanoclusters (64Cu-AuNCs) delivered to the brain using ex vivo gamma counting and inductively coupled plasma mass spectrometry (ICP-MS)20. The 64Cu radioactivity was found to be linearly correlated with the Au concentration quantified using ICP-MS (R2 = 0.94), which confirmed the radiolabel stability of 64Cu for accurate measurement of 64Cu-AuNCs distribution20. However, MRI, PET, and SPECT cannot be used to perform intraprocedural monitoring of the FUS treatment.
Among existing techniques, passive cavitation monitoring, including passive cavitation detection (PCD) and passive cavitation imaging (PCI), is the most commonly used technology for FUS-BBBD treatment monitoring. The underlying concept is that mechanical oscillations of the microbubbles emit ultrasound waves that can be detected using ultrasound sensors to passively “listen” to the secondary acoustic emissions by the microbubbles24 (this differs from active pulse-echo ultrasound imaging). Passive cavitation detection (PCD), using a single-element ultrasound sensor, has been broadly used for real-time FUS-BBBD treatment monitoring to quantify cavitation levels, differentiate between cavitation modes (i.e., stable cavitation and inertial cavitation), investigate bioeffects of cavitation, and develop feedback control algorithms to enhance the FUS treatment safety and efficacy25–27. Stable cavitation (SC, small amplitude microbubble oscillation) can be characterized by the detection of harmonic emissions in the frequency domain, while inertial cavitation (IC, large amplitude microbubble oscillation followed by violent collapse) can be characterized by broadband emissions28–30. PCD has been implemented for real-time monitoring of the FUS-BBBD treatment27,31,32 and controlling FUS exposure parameters using a closed-loop control algorithm27. Although PCD has been shown to be a useful tool for FUS-BBBD monitoring, it cannot detect the spatial distribution of cavitation.
PCI, integrating multi-element ultrasound arrays with beam-forming techniques, can be used to image the spatial distribution of cavitation. Using available ultrasound imaging probes, different 2D PCI algorithms have been introduced, for example, time-domain delay-and-sum beamformer33, time-domain robust Capon beamformer34, frequency domain delay-and-sum beamformer35, robust beamforming by linear programming36, and the angular spectrum method37. Moreover, a recent study demonstrated the feasibility of 3D passive cavitation imaging using a customized 2D array38. However, to our knowledge, only two studies have been published on evaluating the feasibility of using PCI to predict the location of BBBD by correlating PCI with the BBB permeability changes evaluated by post-treatment contrast-enhanced MRI39,40. No study has been reported on using PCI to estimate the spatial distribution (both location and concentration) of therapeutic agents delivered by FUS-BBBD.
The objective of this study was to evaluate the potential of PCI for estimating the location and concentration of therapeutic agents delivered by FUS-BBBD. 64Cu-AuNCs were used as model agents because they are chemically stable, biocompatible, easily functionalized by labeling, and constitute a promising theranostic nanomedicine in their own right41–43. In vivo PET imaging was used to quantify the spatial distribution of 64Cu-AuNCs in the brain after FUS-BBBD treatment and correlate it with the cavitation activity detected by PCI during FUS-BBBD treatment. We demonstrated that PCI was capable of pixel-wise estimation of the delivered 64Cu-AuNCs concentration, a technique that we named as cavitation dose painting.
Results
Using an experimental setup illustrated in Fig. 1, PCI successfully recorded the spatiotemporal dynamics of microbubble cavitation activity in the mouse brain during FUS treatment. Figure 2A displays representative SC level maps (used to quantify the stable cavitation activity at individual time points) acquired at different time points (1 s and 60 s) during the treatment. PCI recorded the spatial distribution of the cavitation activity over time. Figure 2B shows the corresponding SCD map, which was generated by integrating the SC level over the whole FUS treatment duration for each pixel to quantify the total cavitation energy. Figure 2C shows the SC level over time at one representative pixel (indicated by the cross in Fig. 2B). The SC level increased immediately after bolus injection of the microbubbles and then showed the trend of decrease over time. Figure 2D–F present the corresponding IC level map, IC dose (ICD) map, and IC level-time curve, which all showed that the microbubble IC activity was at the noise level. PCI data was acquired successfully in eight out of nine mice. The failure to perform PCI in one mouse was due to the malfunction of the ultrasound imaging system.
Same as our previous publications19,20, PET was used to quantify the concentrations of 64Cu-AuNCs delivered by FUS-BBBD in terms of the percent injected dose per gram of tissue (%ID/g). Successful delivery of 64Cu-AuNCs at the FUS-treated brain region was observed in seven out of the eight mice with successful PCI data acquisition, as indicated by higher radioactivity at the targeted brain location compared with surrounding untreated brain location (Fig. 3). The failure to deliver 64Cu-AuNCs to one mouse was due to an unsuccessful tail vein injection indicated by the accumulation of radioactivity in the mouse tail on the acquired PET image. Results obtained from this mouse was discarded in the analysis. Therefore, we have a total of seven mice for analyzing the correlation between PCI and PET.
Figure 4 displays representative SCD maps (column 1), 64Cu-AuNCs concentration maps acquired at approximately the same imaging plane as the SCD maps (column 2), and overlay of these two maps (column 3) for three representative cases. Although all seven mice showed localized delivery of the 64Cu-AuNCs in the mouse brains after FUS-BBBD treatment using the same protocol, variations in the recorded microbubble cavitation activity and the delivery outcomes were observed among mice as shown by these representative cases. For the three cases, the maximum SCD of each case varied from 108 dB to 116 dB, and the maximum 64Cu-AuNCs concentration of each case varied within the range of 1.15–1.95%ID/g.
For all the treated mice, the offset between the pixel with the maximum SCD (SCDmax) on the PCI maps and the pixel with the highest concentration of 64Cu-AuNCs on the PET images were 0.3 ± 0.4 mm and 1.6 ± 1.1 mm in the transverse and axial directions of the FUS beam, respectively (Fig. 5B). We then evaluated the pixel-by-pixel correlation between the SCD maps and the concentration maps within a region in the brain (the width of this region was defined by a lateral line within the brain passing the SCDmax; the height of this region was defined by −1 dB of the SCDmax in the axial direction to ensure this region was inside the brain). A good linear correlation was obtained based on segmented linear regression (R2 = 0.61). The segmented linear regression was used to model a threshold for BBBD in reference to a previous publication27.
To validate the above correlation between PCI and PET identified in vivo, nine additional mice were treated by the same protocol and sacrificed at 24 h after FUS treatment for the ex vivo quantification of the radioactivity using gamma counting and measurement of gold concentration using ICP-MS (Fig. 6A). It was found that the 64Cu-AuNCs concentration quantified ex vivo in the FUS-treated halves and contralateral non-treated halves were well correlated (R2 = 0.62 for gamma counting and R2 = 0.53 for ICP-MS) with the spatial-averaged SCD within the corresponding brain regions using the segmented linear regression (Fig. 6B,C).
Discussion
This study demonstrated that PCI was capable of predicting the location and concentration of 64Cu-AuNCs delivered by FUS-BBBD, a technique that we named as cavitation dose painting. This PCI-based cavitation dose painting technique can be used in the future to guide FUS-BBBD treatment to achieve conformal delivery of therapeutic agents to the targeted brain region and avoid side effects in the non-targeted brain area, and to deliver higher prescribed drug concentrations to high-risk subvolumes while treating lower-risk subvolumes by delivering lower prescribed drug concentrations. This novel strategy can open new horizons in controlled and precise brain drug delivery.
The capability of PCI to measure the location of FUS-BBBD mediated drug delivery was evaluated by quantifying the offsets between the pixel location of the SCDmax and the pixel location with the highest concentration of 64Cu-AuNCs. The spatial resolution of PCI is determined by the diffraction pattern of the imaging array. The lateral resolution of PCI is at the scale of 10x of the axial resolution, which limits the accuracy of PCI in predicting the agent delivery location in the axial direction. Two previous studies evaluated the capability of PCI in estimating the location of the BBB disruption by comparing PCI with post-treatment contrast-enhanced MRI39,40. In one study, the mean offsets were 0.7 mm and 2.2 mm in the transverse and axial directions, respectively40. In the other study, it was reported that the offsets were 0.3 ± 1.5 mm and 0.5 ± 7.5 mm in the transverse and axial directions, respectively39. Although the offsets were close to those found in the current study (0.3 ± 0.4 mm in the transverse direction and 1.6 ± 1.1 mm in the axial direction), the two previous studies used contrast-enhanced MRI to infer the locations of the delivered agents. The physicochemical properties of MR contrast agents are different from the actual agents intended to be delivered by FUS-BBBD. Thus, the biodistribution of MR contrast agents may not accurately represent the distribution of the delivered agent. The present study directly imaged the 64Cu-AuNCs distribution by PET. PET imaging of radiolabeled nanoparticles provides a noninvasive, highly sensitive, and quantitative method for assessing the efficiency of nanoparticle delivery and their spatial distribution. Therapeutic agents beside nanoparticles, e.g., chemotherapy drugs, peptides, and proteins, can be radiolabeled for directly evaluating the agent spatial distribution in the brain using PET44, instead of inferring its spatial distribution based on the leakage of MR contrast agents.
Our study showed for the first time that a linear relationship exists between SCD quantified using PCI (2D) and 64Cu-AuNC concentration quantified using PET at a pixel-by-pixel level. A previous study also found a similar linear correlation between SCD quantified using PCD (1D) and the concentration of the delivered agents quantified using ex vivo fluorescence imaging. As mentioned earlier, our previous study demonstrated that the 64Cu radioactivity was linearly correlated with the Au concentration (R2 = 0.94), indicating that the radioactivity of 64Cu can accurately measure the concentration of the AuNCs20. In the current study, the linear relationship identified using in vivo PET was verified with ex vivo quantification of the radioactivity using gamma counting (Fig. 6B) and ex vivo quantification of the gold concentration using ICP-MS (Fig. 6C). As the ex vivo study measured the average 64Cu-AuNC concentration in each half of the brain slices (Fig. 6A), the values of the measured concentrations (Fig. 6B,C) were lower than those obtained using PET (Fig. 5C).
All the mice were treated using the same FUS parameters; however, the cavitation dose maps and 64Cu-AuNCs delivery outcomes varied (Fig. 4). This variation among repeated FUS-BBBD treatment has been reported before. It can be caused by variation among repeated experiment in parameters that are hard to control, such as the size distribution of microbubbles reaching the targeted brain location45, circulating microbubble concentration in blood46, blood vessel density within the treated region of the brain47, and heterogeneous acoustic property of skull48. This finding justified the need for treatment monitoring to characterize variation among experiments.
Our finding that PCI can quantitatively measure the location and concentration of 64Cu-AuNCs delivered by FUS-BBBD is significant because it suggests that PCI can be used to verify localized drug delivery to the targeted brain site without off-target effects in the brain, assess whether the concentration of the delivered agent reaches the desired therapeutic level, and characterize variations among repeated treatments. Real-time PCI needs to be developed in the future to quantify the cavitation dose during the FUS-BBBD treatment and provide feedback control of the FUS parameters during the treatment. This PCI-feedback-controlled FUS-BBBD technique will allow controlling drug delivery location and spatially tailoring the delivered drug concentration to match the needed drug delivery doses at different subvolumes of the diseased brain region. The cavitation dose painting technique is not limited to the delivery of nanoparticles. In future clinical translation, therapeutic agents (e.g., chemotherapeutic drugs, peptides, and proteins) can be radiolabeled and injected to the patient at a low dose (lower than the therapeutic level) after FUS-BBBD treatment. The relationship between SCD and the delivered drug concentration as quantified by PET can be identified to develop the cavitation dose painting technique, which can then be used to estimate and control the delivery of the actual drugs without radiolabeling.
This proof-of-concept study demonstrated the concept of cavitation dose painting. Future work is needed to address several limitations of this study. First, PCI imaged cavitation activity in one coronal plane while PET imaged the radioactivity of the delivered 64Cu-AuNCs in 3D. Future studies are needed to improve the spatial co-registration of PCI and PET by developing 3D PCI techniques. Second, the pixel size of the PET images was 0.8×0.8 mm2 and the pixel size of the PCI was 0.2 × 0.2 mm2. Downsampling of the PCI was performed in order to analyze the pixel-by-pixel correlation between PCI and PET. Future studies using larger animal models (e.g., pigs and non-human primates) will allow us to better characterize the pixel-wise correlation between these two imaging modalities. Third, although almost the whole brain was included in our pixel-by-pixel analysis (Fig. 5C), we did not use the whole brain region as our region of interest because the axial resolution of PCI was low and including regions outside the −1 dB of the SCDmax in the axial direction decreased the correlation coefficient to R2 = 0.45. Future studies need to improve the axial resolution of PCI by developing advanced PCI imaging algorithms36.
Materials and Methods
FUS-BBBD treatment
All animal procedures were reviewed and approved by the Institutional Animal Care and Use Committee of Washington University in St. Louis, in accordance with the National Institutes of Health Guidelines for animal research. C57BL/6 female mice (6–8 weeks, 20–25 g body weight, n = 18 in total) were purchased from Charles River Laboratory (Wilmington, MA, USA). Following a procedure described in our previous publication20, the animals were prepared for FUS sonication by removing the fur on the mouse head and coupled to a water container using ultrasound gel.
The prepared mice were sonicated using a commercially available ultrasound image-guided FUS system (VIFU 2000, Alpinion, Bothell, WA, USA) that integrated a FUS system with a software programmable ultrasound imaging system. This system used a FUS transducer with a center frequency of 1.5 MHz, a focal depth of 60 mm, an aperture of 60 mm, and a circular central opening of 38 mm. A linear array (L8–17, Alpinion, Seoul, Korea) with a bandwidth of 8–17 MHz and a center frequency of 12 MHz was inserted into the FUS transducer center opening. The ultrasound imaging plane was aligned with the FUS focal plane.
The ultrasound imaging system was used for both treatment planning and PCI. Treatment planning was performed with the assistant of a metal grid using a previously published method to align the FUS transducer focus at the left brainstem20. Size-isolated microbubbles with a median diameter of 4–5 μm were prepared in-house according to a previously described protocol45 and diluted using sterile saline to a final concentration of about 8 × 108 number of microbubbles per ml. The diluted microbubbles (volume = 25 μL) were administered by a bolus injection via the tail vein. Immediately after injection (~9 s), the mice were treated by the FUS system using the following parameters: center frequency = 1.5 MHz, peak negative pressure = 0.61 MPa, pulse length = 6.7 ms, pulse repetition frequency = 5 Hz, and sonication duration = 120 s. After treatment, mice were put back into their cages for them to recover from anesthesia.
The pressure amplitude and beam dimensions of the FUS transducer were calibrated using a needle hydrophone (HGL-0200, Onda, CA, USA) in a degassed water tank before the experiment. The pressures reported here were the measured hydrophone peak negative pressures corrected for 18% mouse skull attenuation3. The full width at half maximum (FWHM) dimensions of the axial and lateral beams were 6.04 mm and 0.62 mm, respectively.
PCI data acquisition and processing
The ultrasound imaging system was programmed to first operate in the pulse-echo mode to acquire a B-mode image for identifying the location of the mouse skull during PCI post-processing. The system was then operated at the passive mode for acquiring two sets of PCI data. The first PCI data acquisition was performed with FUS sonication on but before the injection of microbubbles to define the background cavitation noise level. The second PCI data acquisition was performed immediately after microbubble injection and in synchronized with the FUS sonication. During the sonication of each FUS pulse, one PCI frame was acquired with an acquisition duration of 400 μs.
In reference to a PCI method used by Haworth et al.35, a frequency-domain PCI algorithm was written in Matlab (Mathworks inc., Natick, MA, USA) to proccess the acquired PCI data offline through the following procedure: (1) Apply Butterworth bandpass filtering around the imaging probe bandwidth of 8–17 MHz to reduce contributions from FUS source signals; (2) Apply a phase shift to the signal acquired by each element in the frequency domain (equivalent to a time delay in temporal domain) based on the propagation times between the receiving element and the spatial location the pixel represents; (3) Sum the phase-shifted waveforms across all the elements for each pixel and its energy is computed by calculating the square of the summed waveforms; Parallel computation using Graphics Processing Units (GPUs) was integrated into the Matlab code to accelerate the calculation speed; (4) Calculate the mean amplitude of the spectrum within selected harmonic bandwidths (0.3 MHz window around all harmonics and superharmonics) for each pixel to generate SC maps; the mean amplitudes of braodband signal falling within frequency bands between the harmonic and ultraharmonic bands were used to calculate the IC levels and generate the IC maps; (5) The SC and IC maps obtained from each FUS treatment were integrated over time to obtain the SCD and ICD maps, respectively.
In vivo MicroPET/CT image acquisition and processing
The mice were imaged by the Inveon PET/CT system (Siemens, Knoxville, TN) at 24 h after intravenous injection of 64Cu-AuNCs at a concentration ~10 MBq following the FUS-treatment. Details on the 64Cu-AuNCs manufacture and characterization were reported in our previous publications49. The 64Cu-AuNCs had a homogeneous size distribution with the hydrodynamic diameter = 5.60 ± 0.50 nm and zeta potential = −0.40 ± 0.11 mV. In our previous study, we performed PET imaging at 1 h, 4 h, and 24 h post treatment and found the 64Cu-AuNCs were almost cleared from the blood circulation at the 24 h time point20. Therefore, in the current study, we chose to perform the microPET/CT scans at the 24 h time point as it allowed us to more accurately quantify the 64Cu-AuNCs concentrations in the brain tissue compared with earlier time points by eliminating the interference of the 64Cu-AuNCs in the blood.
The acquired PET images were corrected for attenuation, scatter, normalization, and camera dead time and co-registered with CT images. The PET images were reconstructed with the maximum a posteriori (MAP) algorithm. Images were analyzed using Inveon Research Workplace (Siemens, Knoxville, TN) and the Matlab program. Partial volume correction was performed following an established method to eliminate the spillover of signals from tissue outside the brain50. Decay correlation was applied to compensate for the decay of 64Cu over time. The pixel intensity of the PET image was quantified as percent injected dose per gram tissue (%ID/g).
Analysis of the correlation between PCI and PET
Since the ultrasound imaging probe was co-axially aligned with the FUS transducer, the focal point of the FUS transducer was located in the ultrasound imaging plane. To investigate the correlation between PCI and PET, we first identified the PET voxel location with the highest radioactivity inside the brain, which was considered to be the location of the FUS focus. Then the PET images acquired in the coronal plane passing through this voxel was selected and considered to be the corresponding imaging plane for the B-mode/PCI. This selection of the corresponding PET imaging plane was verified in all the cases as the same anatomic features of the mouse skulls were observed in the B-mode images and corresponding CT images (Fig. 1). Then six controlling points were selected on the B-mode and corresponding CT images (Fig. 1). The CT images were translated and rotated to fit the same controlling points on B-mode images. Through this process, the transformation matrix was acquired and applied to PET images to register the PET images with PCI images.
For each mouse, the capability of SCD in predicting 64Cu-AuNC delivery location was characterized by quantifying the offsets between the pixel location of the SCDmax and the location with the maximum 64Cu-AuNC concentration. The offsets of these two locations were decomposed to the lateral direction and axial direction. The capability of SCD in predicting 64Cu-AuNCs concentration was determined by evaluating the pixel-by-pixel correlation between the SCD map and the concentration map. The pixel size of the PET scanner was 0.8 mm × 0.8 mm, corresponding to the intrinsic sampling size of the scanner. The SCD maps had a pixel size of 0.2 mm × 0.2 mm. To perform a pixel by pixel comparison between these two maps, we first aligned the two images by aligning the pixel location of SCDmax and the pixel with the highest concentration. Then we decreased the pixel size of the SCD map by downsampling to match that of the PET. The correlations between these two maps within the 2D region (Fig. 5A) were evaluated by the segmented linear regression using GraphPad Prism (Version 6.04, La Jolla, CA, USA), and the goodness of fit was assessed by the correlation coefficient, R2.
Ex vivo quantification
To validate the correlation found between PCI and in vivo PET imaging, nine additional mice were used to perform ex vivo quantification of 64Cu-AuNCs radioactivity using gamma counting and Au concentration by ICP-MS20. These mice were treated using the same FUS treatment protocol as described in 4.1. After FUS treatment, they were sacrificed by transcardial perfusion at 24 h after FUS treatment; their brains were collected and sliced coronally into 2-mm slices. Then the slices containing the brainstem was cut into two halves (left and right representing FUS-treated and non-treated samples, respectively) and the radioactivity was counted using a Beckman 8000 gamma counter (Beckman, Fullerton, CA). The count rate (counts per minute, CPM) for each tissue sample was corrected by automatic background subtraction. Decay correction was applied (compensated for the decay of 64Cu radioactivity over time). The corrected CPM from each tissue sample was normalized both to the mass of the tissue sample (in grams, g) and to the injected dose (ID). The concentration of 64Cu-AuNCs in each tissue sample was then calculated as %ID/g. The correlation between the radioactivity (%ID/g) of 64Cu-AuNCs in the FUS-treated and contralateral non-treated brainstem and the corresponding spatial-averaged SCD within the same regions was evaluated using the segmented linear regression.
After the gamma counting, these samples were then digested using a high-pressure microwave digestion system (Milestone Inc. Monroe, CT) and the gold concentrations in the digested brain tissue samples were determined using ICP-MS (Elan DRC-e, PerkinElmer, Germany). Au standard was used to generate the standard curve. The Au concentration was expressed as the percentage of the Au in the brainstem over the total injected Au normalized by the tissue weight (%ID/g). The Au concentration (%ID/g) in FUS-treated and non-treated brainstem was correlated with the corresponding spatial-averaged SCD within the same regions using the segmented linear regression.
Conclusions
This paper has demonstrated the potential of PCI in predicting the spatial location and concentration of therapeutic agents delivered by FUS-BBBD. We call this new strategy PCI-based cavitation dose painting. The accuracy of this strategy was evaluated by correlating PCI with in vivo PET/CT imaging of radiolabeled nanoparticles delivered by FUS-BBBD. This study laid the foundation for the future development of PCI-based real-time feedback control of the FUS-BBBD treatment to spatially control and modulate the delivered drug concentrations.
Acknowledgements
This work was in part supported by the Children’s Discovery Institute of Washington University and St. Louis Children’s Hospital (grant number MC-II-2017-661), American Cancer Society (grant number IRG-15-170-58), and the National Institutes of Health (NIH) grant R01MH116981 and R01EB027223. The authors wish to thank Dr. Kevin Haworth (PhD, University of Cincinnati) for insightful discussions of passive cavitation imaging.
Author Contributions
H.C. and Y.Y. developed the idea. Y.Y., X.Z., D.Y., Y.L. and H.C. designed and performed the experiment. R.L. and G.W. assisted with the data analysis and ideal development. Y.Y. and H.C. wrote the manuscript and analyzed the data. All authors reviewed the manuscript.
Data Availability
The datasets generated in this study are available from the corresponding author upon request.
Competing Interests
The authors declare no competing interests.
Footnotes
Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Yaoheng Yang and Xiaohui Zhang contributed equally.
References
- 1.Burgess A, Hynynen K. Noninvasive and targeted drug delivery to the brain using focused ultrasound. ACS Chem. Neurosci. 2013;4:519–526. doi: 10.1021/cn300191b. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Aryal M, Arvanitis CD, Alexander PM, McDannold N. Ultrasound-mediated blood-brain barrier disruption for targeted drug delivery in the central nervous system. Adv. Drug Deliv. Rev. 2014;72:94–109. doi: 10.1016/j.addr.2014.01.008. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Choi JJ, Pernot M, Small SA, Konofagou EE. Noninvasive, transcranial and localized opening of the blood-brain barrier using focused ultrasound in mice. Ultrasound Med. Biol. 2007;33:95–104. doi: 10.1016/j.ultrasmedbio.2006.07.018. [DOI] [PubMed] [Google Scholar]
- 4.Pardridge WM. The blood-brain barrier: Bottleneck in brain drug development. NeuroRx. 2005;2:3–14. doi: 10.1602/neurorx.2.1.3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Aryal M, Vykhodtseva N, Zhang YZ, Park J, McDannold N. Multiple treatments with liposomal doxorubicin and ultrasound-induced disruption of blood-tumor and blood-brain barriers improve outcomes in a rat glioma model. J. Control. Release. 2013;169:103–111. doi: 10.1016/j.jconrel.2013.04.007. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Burgess A, et al. Alzheimer disease in a mouse model: MR imaging–guided focused ultrasound targeted to the hippocampus opens the blood-brain barrier and improves pathologic abnormalities and behavior. Radiology. 2014;273:736–745. doi: 10.1148/radiol.14140245. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.McDannold N, Zhang Y, Vykhodtseva N. The effects of oxygen on ultrasound-induced blood–brain barrier disruption in mice. Ultrasound Med. Biol. 2017;43:469–475. doi: 10.1016/j.ultrasmedbio.2016.09.019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Mead BP, et al. Targeted gene transfer to the brain via the delivery of brain-penetrating DNA nanoparticles with focused ultrasound. J. Control. Release. 2016;223:109–117. doi: 10.1016/j.jconrel.2015.12.034. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Wang S, Olumolade OO, Sun T, Samiotaki G, Konofagou EE. Noninvasive, neuron-specific gene therapy can be facilitated by focused ultrasound and recombinant adeno-associated virus. Gene Ther. 2014;22:104–110. doi: 10.1038/gt.2014.91. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Hynynen K, McDannold N, Vykhodtseva N, Jolesz FA. Noninvasive MR imaging–guided focal opening of the blood-brain barrier in rabbits. Radiology. 2001;220:640–646. doi: 10.1148/radiol.2202001804. [DOI] [PubMed] [Google Scholar]
- 11.Marquet F, Tung Y-S, Teichert T, Ferrera VP, Konofagou EE. Noninvasive, transient and selective blood-brain barrier opening in non-human primates in vivo. PLoS One. 2011;6:e22598. doi: 10.1371/journal.pone.0022598. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.McDannold N, Arvanitis CD, Vykhodtseva N, Livingstone MS. Temporary disruption of the blood-brain barrier by use of ultrasound and microbubbles: safety and efficacy evaluation in rhesus macaques. Cancer Res. 2012;72:3652–63. doi: 10.1158/0008-5472.CAN-12-0128. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Liu HL, et al. Blood-brain barrier disruption with focused ultrasound enhances delivery of chemotherapeutic drugs for glioblastoma treatment. Radiology. 2010;255:415–425. doi: 10.1148/radiol.10090699. [DOI] [PubMed] [Google Scholar]
- 14.Sheikov N, McDannold N, Vykhodtseva N, Jolesz F, Hynynen K. Cellular mechanisms of the blood-brain barrier opening induced by ultrasound in presence of microbubbles. Ultrasound Med. Biol. 2004;30:979–989. doi: 10.1016/j.ultrasmedbio.2004.04.010. [DOI] [PubMed] [Google Scholar]
- 15.Cho EE, Drazic J, Ganguly M, Stefanovic B, Hynynen K. Two-photon fluorescence microscopy study of cerebrovascular dynamics in ultrasound-induced blood-brain barrier opening. J. Cereb. Blood Flow Metab. 2011;31:1852–1862. doi: 10.1038/jcbfm.2011.59. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Tung YS, Vlachos F, Feshitan JA, Borden MA, Konofagou EE. The mechanism of interaction between focused ultrasound and microbubbles in blood-brain barrier opening in mice. J. Acoust. Soc. Am. 2011;130:3059–3067. doi: 10.1121/1.3646905. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Lipsman N, et al. Blood–brain barrier opening in Alzheimer’s disease using MR-guided focused ultrasound. Nat. Commun. 2018;9:2336. doi: 10.1038/s41467-018-04529-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Lee H, et al. 64Cu-MM-302 positron emission tomography quantifies variability of enhanced permeability and retention of nanoparticles in relation to treatment response in patients with metastatic breast cancer. Clin. Cancer Res. 2017;23:4190–4202. doi: 10.1158/1078-0432.CCR-16-3193. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Sultan D, et al. Focused ultrasound enabled trans-blood brain barrier delivery of gold nanoclusters: Effect of surface charges and quantification using positron emission tomography. Small. 2018;14:1703115. doi: 10.1002/smll.201703115. [DOI] [PubMed] [Google Scholar]
- 20.Ye D, et al. Focused ultrasound-enabled delivery of radiolabeled nanoclusters to the pons. J. Control. Release. 2018;283:143–150. doi: 10.1016/j.jconrel.2018.05.039. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Kim H, et al. PET/CT imaging evidence of FUS-mediated (18) F-FDG uptake changes in rat brain. Med. Phys. 2013;40:33501. doi: 10.1118/1.4789916. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Lin KJ, et al. Quantitative micro-SPECT/CT for detecting focused ultrasound-induced blood-brain barrier opening in the rat. Nucl. Med. Biol. 2009;36:853–61. doi: 10.1016/j.nucmedbio.2009.04.011. [DOI] [PubMed] [Google Scholar]
- 23.Yang FY, et al. Pharmacokinetic analysis of 111In-labeled liposomal doxorubicin in murine glioblastoma after blood-brain barrier disruption by focused ultrasound. PLoS One. 2012;7:1–7. doi: 10.1371/journal.pone.0045468. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Hilgenfeldt S, Lohse D, Zomack M. Response of bubbles to diagnostic ultrasound: A unifying theoretical approach. Eur. Phys. J. B. 1998;4:247–255. doi: 10.1007/s100510050375. [DOI] [Google Scholar]
- 25.Chen H, Konofagou EE. The size of blood-brain barrier opening induced by focused ultrasound is dictated by the acoustic pressure. J. Cereb. Blood Flow Metab. 2014;34:1197–204. doi: 10.1038/jcbfm.2014.71. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.O’Reilly MA, Hynynen K. Feedback-controlled focused ultrasound disruption by using an acoustic emissions-based controller. Radiology. 2012;263:96–106. doi: 10.1148/radiol.11111417. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Sun T, et al. Closed-loop control of targeted ultrasound drug delivery across the blood–brain/tumor barriers in a rat glioma model. Proc. Natl. Acad. Sci. 2017;114:E10281–E10290. doi: 10.1073/pnas.1713328114. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Sun T, et al. Acoustic cavitation-based monitoring of the reversibility and permeability of ultrasound-induced blood-brain barrier opening. Phys. Med. Biol. 2015;60:9079–94. doi: 10.1088/0031-9155/60/23/9079. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Arvanitis CD, Livingstone MS, Vykhodtseva N, McDannold N. Controlled ultrasound-induced blood-brain barrier disruption using passive acoustic emissions monitoring. PLoS One. 2012;7:e45783. doi: 10.1371/journal.pone.0045783. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.McDannold N, Vykhodtseva N, Hynynen K. Targeted disruption of the blood–brain barrier with focused ultrasound: association with cavitation activity. Phys. Med. Biol. 2006;51:793–807. doi: 10.1088/0031-9155/51/4/003. [DOI] [PubMed] [Google Scholar]
- 31.Tsai C-H, Zhang J-W, Liao Y-Y, Liu H-L. Real-time monitoring of focused ultrasound blood-brain barrier opening via subharmonic acoustic emission detection: implementation of confocal dual-frequency piezoelectric transducers. Phys. Med. Biol. 2016;61:2926–2946. doi: 10.1088/0031-9155/61/7/2926. [DOI] [PubMed] [Google Scholar]
- 32.O’Reilly MA, Hynynen K. Blood-brain barrier: Real-time feedback-controlled focused ultrasound disruption by using an acoustic emissions-based controller. Radiology. 2012;263:96–106. doi: 10.1148/radiol.11111417. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Gyöngy M, Coussios CC. Passive cavitation mapping for localization and tracking of bubble dynamics. J. Acoust. Soc. Am. 2010;128:EL175–EL180. doi: 10.1121/1.3467491. [DOI] [PubMed] [Google Scholar]
- 34.Coviello C, et al. Passive acoustic mapping utilizing optimal beamforming in ultrasound therapy monitoring. J. Acoust. Soc. Am. 2015;137:2573. doi: 10.1121/1.4916694. [DOI] [PubMed] [Google Scholar]
- 35.Haworth KJ, Bader KB, Rich KT, Holland CK, Mast TD. Quantitative frequency-domain passive cavitation imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control. 2016;64:177–191. doi: 10.1109/TUFFC.2016.2620492. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Lyka, E., Coviello, C. M., Paverd, C., Gray, M. D. & Coussios, C. C. Passive acoustic mapping using data-adaptive beamforming based on higher-order statistics. IEEE Trans. Med. Imaging PP, 1–1 (2018). [DOI] [PubMed]
- 37.Arvanitis CD, Crake C, McDannold N, Clement GT. Passive acoustic mapping with the angular spectrum method. IEEE Trans. Med. Imaging. 2016;36:983–993. doi: 10.1109/TMI.2016.2643565. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Acconcia CN, Jones RM, Goertz DE, O’Reilly MA, Hynynen K. Megahertz rate, volumetric imaging of bubble clouds in sonothrombolysis using a sparse hemispherical receiver array. Phys. Med. Biol. 2017;62:L31–L40. doi: 10.1088/1361-6560/aa84d7. [DOI] [PubMed] [Google Scholar]
- 39.Wu SY, et al. Efficient blood-brain barrier opening in primates with neuronavigation-guided ultrasound and real-time acoustic mapping. Sci. Rep. 2018;8:7978. doi: 10.1038/s41598-018-25904-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Arvanitis CD, Livingstone MS, McDannold N. Combined ultrasound and MR imaging to guide focused ultrasound therapies in the brain. Phys. Med. Biol. 2013;58:4749–4761. doi: 10.1088/0031-9155/58/14/4749. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Zhao, Y. et al. Copper-64-alloyed gold nanoparticles for cancer imaging: Improved radiolabel stability and diagnostic accuracy. Angew. Chemie - Int. Ed. 53, 156–159 (2014). [DOI] [PubMed]
- 42.Zhao Y, et al. Gold nanoclusters doped with 64Cu for CXCR4 positron emission tomographyimaging of breast cancer and metastasis. ACS Nano. 2016;10:5959–5970. doi: 10.1021/acsnano.6b01326. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Bentzen SM, Gregoire V. Molecular Imaging-Based Dose Painting: A Novel Paradigm for Radiation Therapy Prescription. Semin. Radiat. Oncol. 2011;21:101–110. doi: 10.1016/j.semradonc.2010.10.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Gustafsson S, et al. Combined PET and microdialysis for in vivo estimation of drug blood-brain barrier transport and brain unbound concentrations. Neuroimage. 2017;155:177–186. doi: 10.1016/j.neuroimage.2017.04.068. [DOI] [PubMed] [Google Scholar]
- 45.Choi JJ, et al. Microbubble-size dependence of focused ultrasound-induced blood-brain barrier opening in mice in vivo. IEEE Trans. Biomed. Eng. 2010;57:145–154. doi: 10.1109/TBME.2009.2034533. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Morel DR, et al. Human pharmacokinetics and safety evaluation of SonoVueTM, a new contrast agent for ultrasound imaging. Invest. Radiol. 2000;35:80. doi: 10.1097/00004424-200001000-00009. [DOI] [PubMed] [Google Scholar]
- 47.Wu SY, et al. Characterizing focused-ultrasound mediated drug delivery to the heterogeneous primate brain in vivo with acoustic monitoring. Sci. Rep. 2016;6:37094. doi: 10.1038/srep37094. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 48.Aldiabat H, O’Brien PD, Liu D, Ebbini ES. Wideband transskull refocusing of ultrasound beams using dual-mode ultrasound arrays: Ex vivo results. J. Acoust. Soc. Am. 2018;143:1731. doi: 10.1121/1.5035645. [DOI] [Google Scholar]
- 49.Ye D, et al. Focused ultrasound combined with microbubble-mediated intranasal delivery of gold nanoclusters to the brain. J. Control. Release. 2018;286:145–153. doi: 10.1016/j.jconrel.2018.07.020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Izquierdo-Garcia D, et al. Comparison of methods for magnetic resonance-guided [18-F]fluorodeoxyglucose positron emission yomography in human carotid arteries: reproducibility, partial volume correction, and correlation between methods. Stroke. 2009;40:86–93. doi: 10.1161/STROKEAHA.108.521393. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The datasets generated in this study are available from the corresponding author upon request.