Skip to main content
NASA Author Manuscripts logoLink to NASA Author Manuscripts
. Author manuscript; available in PMC: 2019 Mar 4.
Published in final edited form as: Int J Mass Spectrom. 2016 Jun 15;407:51–61. doi: 10.1016/j.ijms.2016.06.005

High-precision molybdenum isotope analysis by negative thermal ionization mass spectrometry

Emily A Worsham 1,*, Richard J Walker 1, Katherine R Bermingham 1
PMCID: PMC6398950  NIHMSID: NIHMS1008018  PMID: 30842700

Abstract

Procedures for the separation, purification, and high-precision analysis of mass-independent isotopic variations in molybdenum (Mo) using negative thermal ionization mass spectrometry are reported. Separation and purification of Mo from silicate and metal matrices are achieved using a two-stage anion exchange chromatographic procedure. Molybdenum is ionized as the MoO3 species using a double filament assembly. The MoO3 ion beams are collected using Faraday cup detectors equipped with a mixed array of amplifiers utilizing 1011 and 1012 Ω resistors, which allows for in situ measurement and correction of oxygen isobars. The long-term external reproducibility of 97Mo/96Mo, the most precisely measured Mo isotope ratio, is ±5.4 ppm (2SD), based on the repeated analyses of the Alfa Aesar Specpure® Mo plasma standard and using 98Mo/96Mo for fractionation correction. The long-term external reproducibilities of 92Mo/96Mo, 94Mo/96Mo, 95Mo/96Mo, and 100Mo/96Mo are ±107, 37, 23, and 32 ppm (2SD), respectively. With this precision, smaller differences in Mo isotopic compositions can be resolved than have been previously possible.

Keywords: Molybdenum isotope, Thermal ionization mass spectrometry, N-TIMS, Oxide correction, 1012 Ω Resistor, Oxygen isotope

1. Introduction

The molybdenum (Mo) isotopic system is uniquely suited to address current issues in cosmochemistry from the origin of isotopic heterogeneity in the solar nebula to the diversity of planetary bodies. Though the cosmochemical applications for Mo isotopes are the primary focus of this study, the application of mass-dependent Mo isotope effects as indicators of paleoredox conditions has been the focus of most Mo isotope studies in geochemistry (e.g., [13]). In cosmochemistry, “mass-independent” nucleosynthetic Mo isotope anomalies, which have been observed in a variety of cosmochemical materials, including at the bulk meteorite scale, have been used primarily to evaluate the extent of mixing and/or thermal processing in the solar nebula, and to test genetic relations amongst solar system materials (e.g., [47]). The use of Mo as a genetic tracing tool is important for the elucidation of nebular and planetary processes, including the characterization of the nebular feeding zones of planetary bodies. By merging groups or incorporating ungrouped or anomalous meteorites into existing groups, the accretionary histories and evolution of meteorite parent bodies can be further developed. As a corollary, by demonstrating that certain solar system objects are unrelated, the parent bodies’ evolutionary histories can be clarified.

Molybdenum has been previously analyzed by thermal ionization mass spectrometry in both positive (P-TIMS – e.g., [811]) and negative mode (N-TIMS – [12,5,6,13]). Molybdenum isotopes in cosmochemical and terrestrial materials have also been measured by multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS) (e.g., [14,15,4,7,16]). The current state-of-the-art precision (2SD) of the measured 97Mo/96Mo isotope ratio is ±13 ppm by N-TIMS [13] and ±21 ppm by MC-ICP-MS [13]. Although the method of Nagai and Yokoyama [13] produces high-precision Mo isotope ratios, 1000 ng–4000 ng of Mo were used in their study. This requires large sample masses of up to 4 g for some chondritic meteorites. Due to the valuable nature of cosmochemical materials, it is desirable to use as little material as possible.

In an effort to increase the resolving power of the Mo isotopic system when ≤1000 ng of Mo are available, we developed procedures for the separation, purification, and high-precision analysis of Mo using a Thermo-Fisher Triton Plus N-TIMS at the University of Maryland. Using N-TIMS has the advantage of more stable signal intensities and lower likelihood of isobaric interferences, as opposed to P-TIMS and MC-ICP-MS. In addition, the sensitivity of N-TIMS is typically higher than that of P-TIMS for elements that form negative ions or oxyanions, such as Mo [17,18]. The method, however, requires correction for interferences from oxides containing the heavier oxygen isotopes (e.g., [17]). One recent advance in the application of N-TIMS is the in situ measurement of certain oxygen isobars, which provides the opportunity to monitor the variability in the oxygen isotopic composition during analysis [1924,13,25]. This study incorporates the use of 1012 Ω resistors to measure and correct for oxygen isobars.

The methods presented here are primarily applicable to the measurement of mass-independent nucleosynthetic effects, but may also be adapted to double-spike methods for the determination of mass-dependentstable isotope effects. To demonstrate the utility of these procedures, the Mo isotopic compositions of standards, a gravimetrically prepared spike-standard mixture, the iron meteorites Toluca and Skookum, and the chondrite Allegan are reported.

2. Sample preparation

Iron meteorite samples were cut into 0.3–2 g pieces using a water-cooled Leco “Vari-cut” saw with a 12.7 cm diamond wafering blade. All iron meteorite pieces were mechanically abraded with different pieces of carborundum paper for each sample, and sonicated in ethanol three times to remove adhering rust, fusion crust, or saw blade contamination. All acids used for digestion and ion exchange chromatography were sub-boiling distilled with a quartz still (HCl and HNO3) or a Savillex Teflon® still (HF) in-house. High-purity water used for dilutions and cleaning was produced using a Milli-Q water purification system (deionized to a resistivity of 18.2 MΩ·cm). Iron meteorite samples were digested using 30–40 ml of 8 M HCl in Savillex Teflon® beakers which were heated at 130–140 °C on a hotplate for at least 48 h. The chondrite sample (Allegan) was powdered in an agate mortar and pestle, and digested in 30 ml of a concentrated HF:HNO3 mixture in 5:1 proportions. The Allegan solution was evaporated to a sludge, 5 ml of concentrated HNO3 was added to dilute the remaining HF, and then 5 ml of concentrated HCl was added and evaporated twice to dissolve fluorides that are insoluble in HNO3. Chondrite and iron meteorite samples were dried to a sludge, and 3 ml of concentrated HNO3:HCl in 2:1 proportions was added to the samples and heated at 100 °C overnight or 115 °C for at least one hour in an effort to drive Mo to its highest oxidation state (MoVI). This step was used in an attempt to mitigate loss of Mo during ion exchange due to the presence of multiple valences of Mo on the column. The acid phase was evaporated to a sludge, at which point, 6 M HCl was added and dried to sludge twice to remove traces of HNO3. Then 1 M HF was added and dried to sludge twice, and finally the samples were taken into the 1 M HF loading solution. At this point, some Mo was likely reduced to a lower oxidation state, but the successive additions of HCl and HF were necessary to ensure that the loading solution was the correct molarity of HF, with minimal traces of HCl and HNO3.

The separation and purification of Mo from natural samples utilized a two-stage anion exchange chromatography method adapted from Pietruszka et al. [26], Scheiderich et al. [27], and Nagai and Yokoyama [28] (Table 1). Eichrom AG 1 × 8 200–400 mesh resin was pre-cleaned using reagent grade 1 M HCl and 6 M HCl, followed by quartz-distilled 6 M HCl and 1 M HCl, added in succession with Milli-Q water between stages of cleaning. The primary column generally follows the same procedure as the primary column of Nagai and Yokoyama [28]. Disposable Biorad Poly-Prep columns with 2 ml resin bed volume capacity were loaded with 1.4 ml of resin, which was further cleaned using the procedures outlined in Table 1. Samples were loaded in 1 ml of 1 M HF for each column, with 1 column per ~100 mg of iron meteorite and ~300 mg of chondrite. Major elements were eluted in an additional 3 ml of 1 M HF; Zr, Ru, Ti, and Hf were eluted in 6 ml of 9 M HCl–0.05 M HF; W was eluted in 12 ml of 9 M HCl–1 M HF; and Mo was collected in 10 ml of 6 M HNO3–3 M HF. Before the addition of 6 M HNO3–3 M HF, 1.5 ml of Milli-Q H2O was added to the column to prevent the formation of aqua regia in the column.

Table 1.

The anion exchange chromatographic procedures for the separation and purification of Mo. The secondary column was repeated one time to improve sample purity.

Acid Volume per column (ml) Step
Primary column - 1–8 Biorad columns filled with 1.4 ml AG 1-X8 200–400 mesh
6 M HC1 10 Cleaning
Milli-Q H2O 10 Cleaning
6M HNO3 - 3 M HF 10 Cleaning
Milli-Q H2O 5 Cleaning
1 M HF 3 Equilibration
1 M HF 1 Load sample
1 M HF 3 Rinse (Fe, matrix)
9 M HC1 – 0.05 M HF 3 + 3 Rinse (Ru, Zr, Ti, Hf)
9 M HC1 – 1 M HF 3 + 3 + 3 + 3 Rinse (W, Re)
Milli-Q H2O 1.5 Remove HO
6 M HNO3 −3 M HF 10 Collect Mo
Secondary column - 1 Teflon® column filled with ~ 0.3 ml AG 1-X8 200–400 mesh
0.01 M HC1 4 Cleaning
1 M HC1 8 Cleaning
6 M HC1 4 Cleaning
6 M HC1 0.5 Load sample
6 M HC1 4 Rinse remaining matrix
0.01 M HC1 – 0.1 M HF 1 Rinse remaining matrix
0.01 M HC1 3 Rinse remaining matrix
1 M HC1 12.5 Collect Mo

The secondary column was a small version of the anion column described by Pietruszka et al. (2006), and was used to further purify the Mo. The columns used for this study were made with Teflon® heat-shrink tubing and were fitted with Teflon® frits. The resin bed for these columns had dimensions of 0.4 × 2.4 cm. The secondary column was filled to the base of the reservoir with Eichrom AG 1 × 8 200–400 mesh resin (~ 300 μl). The resin bed was then cleaned using 0.01 M HCl, 1 M HCl, and 6 M HCl. Prior to loading, samples were dissolved in 2:1 HNO3:HCl and heated overnight or for one hour, as was done prior to the primary column. Again, 6 M HCl was added and dried to remove traces of HNO3 twice, and then the sample was taken up into 0.5 ml of 6 M HCl to be loaded onto the column. Four ml of 6 M HCl was added to the column to elute remaining impurities, such as Fe, Ni, and Co, followed by 1 ml of 0.01 M HCl-0.1 M HF to remove remaining Fe, followed by 3 ml of 0.01 M HCl. Molybdenum was then collected in 12.5 ml of 1 M HCl (Table 1). This column, including the oxidation stage with 2:1 HNO3:HCl, was repeated once to improve sample purity. Molybdenum purity and yield was monitored qualitatively using a Thermo-Fisher Element 2 single-collector ICP-MS prior to loading onto filaments for analysis. Occasional above-background signals on masses corresponding to Na, Al, V, Mn, Zn, and Sn were observed; some of which may have been generated by gas species from the plasma. The largest and most variable signals were observed for 23Na and 27Al, and the ratios of these signals to the corresponding 98Mo signals were typically 0.02–2 (23Na/98Mo) and 0.2–1 (27Al/98Mo). Other species typically had signal ratios, relative to 98Mo, significantly less than ~0.2. The occurrences of these non-analyte species were not correlated with poor ionization of Mo.

Three total analytical blanks were determined for these procedures. Each blank used either one, three, or four primary columns, and resulted in blanks of 0.92, 1.3, and 2.8 ng of Mo, respectively. The total analytical blank using one primary column is, therefore, estimated to be ~0.9 ± 0.3 ng (2SD). These blanks were sufficiently low to be negligible for the quantities of Mo used for the analyses reported.

The total yield for the chemistry described above was ~50–65%. It is possible that the sub-optimal recovery of Mo resulted in mass-dependent fractionation of the Mo isotopes. However, the Mo isotopic compositions reported here were fractionation corrected for mass-dependent isotope effects that occurred naturally and/or that resulted from the column chemistry. Although Rizo et al.[29] showed that fractionation correction of mass-dependently fractionated materials can result in apparent mass-independent anomalies in W isotopes, the agreement of the Mo data reported here with those of previous studies, as discussed below, indicate that fractionation that occurred during the column chemistry was effectively corrected for.

Loss of Mo primarily occurred in the primary column in the loading stage where some Mo was eluted with the Fe. Because the yield for the secondary column was typically 80–100%, and matrix was largely absent at this stage, it is likely that the matrix was the complicating factor in the primary column. The breakthrough of Mo in the primary column was mitigated by loading in 1 M HF, as opposed to the 0.4 M HCl–0.5 M HF loading solution used by Nagai and Yokoyama [28]. The partition coefficient between the resin and the acid phase (Kd) for Mo is relatively low in 0.4 M HCl–0.5 M HF compared to 1 M HF (Kd = 45 in 0.4 M HCl–0.5 M HF vs. Kd = 573 in 1 M HF – Nagai and Yokoyama [28]. These authors chose not to use 1 M HF due to the high Kd values they measured for Zr and Hf, but we observed that Zr was efficiently separated from Mo in the primary and secondary columns. Hafnium was not monitored by ICP-MS, but due to its similar Kd values to Zr in the acids used, it is likely that it was also efficiently separated from Mo. Although Fe does not occupy anion exchange sites in 1 M HF, overloading the column may reduce the tendency of Mo to stick to the resin when Fe is present. The sample sizes loaded onto each primary column were small in order to minimize the effect of Fe on the behavior of Mo on the column. Alternatively, the presence of redox-sensitive Fe or other major elements may change the valence of Mo, resulting in the persistence of mixed valence states in the primary column. One way to mitigate this possible effect is to load the column using H2O2 or another oxidizing agent to better control the oxidation state of Mo (and Fe) (e.g., [30], but this is yet to be incorporated into methods described here.

After the completion of the chromatographic procedures, the purified sample was taken into solution in 2:1 concentrated HNO3:HCl and dried to remove resin-derived organics. This step was repeated seven times for each sample. Following the last addition of 2:1 HNO3:HCl, concentrated HCl was added and dried three times to remove traces of HNO3. Samples were then dissolved in 6 M HCl to load onto filaments.

3. Mass spectrometry

3.1. Sample loading procedure

Samples and standards were loaded onto outgassed Re filaments. The 99.99% pure Re ribbon used for this study was 0.76 mm wide and 0.030 mm thick and was purchased from the H. Cross Company. Rhenium filaments were outgassed at 4.2 A for 40 min at least 24 h before loading. A double filament assembly was found to increase signal intensity and duration of the ionization of Mo, compared to single filaments, in agreement with Turnlund et al. [9] and Giussani et al. [12]. Further, we found that the double filament assembly provided better control of the ionization of Mo than was achieved with single filaments. The ionization of Mo on single filaments typically would reach a point where the growth of the signal and temperature became exponential without manual heating. Soon after, both the signal and temperature would reach a maximum and then the signal would decrease precipitously. The use of an ionization filament and evaporation filament prevented the exponential growth by providing a non-direct heat source, which dramatically improved signal duration and stability. Further, Mo isotopes ionized on single filaments were often fractionated more quickly and more intensely than Mo that was ionized using a double filament assembly. This phenomenon was reflected by the absolute range in the fractionation factor, β, in the exponential law which was used for fractionation correction, discussed below. Over the course of a given analysis, β had an absolute range up to ~0.5 for single filament analyses, but a range of <0.3 for most double filament analyses, which were also ~3.5 times longer in duration.

About 400–1000 ng Mo for samples and 1000 ng Mo for standards were loaded onto the evaporation filament in two ~0.5 μl aliquots, and dried by applying a current of 0.6 A. Before the first aliquot was dry, the second was added so that distinct reservoirs of Mo did not form. The avoidance of distinct reservoirs was found to improve the ionization of Mo on single filaments in the early stages of the development of these procedures. For samples, once the deposit was dried, the current was increased slowly from 0 to ~2 A until a faint red glow was observed or until the sample transitioned from cloudy white to a colorless or bronze-colored deposit with a black rim. This was done to remove any remaining organics not eliminated by successive additions of 2:1 HNO3:HCl prior to loading. Care was taken not to heat the filament beyond a dull glow because the sample can burn off. Ten micrograms of 5 μg/μl La(NO3)3 in 1 M HNO3 were then added as an electron emitter to reduce the work function of Re and promote the ionization of MoO3 (e.g., [17,6]). The La(NO3)3 activator was dried at 0.6 A, but because the solution does not dry completely, the current was then increased from 0 to ~2 A until the solution was observed to boil to dryness. Ten micrograms of La(NO3)3 was also added to the ionization filament, using the same drying procedure.

As this loading procedure was developed, variables such as the type of loading solution, the type and amount of activator, the filament assembly, and the placement of activator on the ionization and/or evaporation filaments were systematically varied in order to optimize the ionization of Mo to facilitate stable, long lasting signals. It was found that loading Mo in a chloride form in HCl led to better ionization than when Mo was loaded in HBr or in HNO3. In agreement with Nagai and Yokoyama [13], the optimal activator was found to be La(NO3)3, although we also trialed Ba(OH)2, Ca(NO3)2, and a mixture of La and Gd. Nagai and Yokoyama [13] also found that a ratio of 5:1 La:Mo was most effective. The results of the initial work with single filaments in this study are consistent with this conclusion, though the optimal ratio of La:Mo for double filaments appears to be higher. Initially, 1ug of Mo was loaded onto the evaporation filament and 5ug of La(NO3)3 was loaded only onto the ionization filament, but it was found that adding an additional 5ug of La(NO3)3 on top of the sample or standard improved ionization. However, when this was done, the problem of exponential growth and subsequent rapid decrease of the Mo signals that was observed with single filaments persisted. When the amount of La(NO3)3 was doubled on both the ionization and evaporation filaments, the temperature of the filaments and the ionization of Mo were better controlled, and, thus, the Mo signal was more stable and long-lived.

The loading blank using this loading procedure was ~1 ng. It was suspected that the Mo was primarily sourced from the Re ribbon, as Mo was not observed in the La(NO3)3 activator when it was examined by ICP-MS. Because Mo was not eliminated by outgassing, and Mo and Re are chemically similar, Mo may be bound in the structure of the Re ribbon. The effect of the loading blank on the measured isotope ratios was examined using different size loads of the gravimetrically prepared spike-standard mixture (400–1000 ng), and was observed to be negligible.

3.2. Instrumental setup and measurement routine

A Thermo-Fisher Triton Plus solid source mass spectrometer was used to measure the Mo isotopic compositions of samples and standards. The instrument was equipped with nine Faraday cup detectors, seven of which were used to measure Mo isotopes. A gain calibration of the Faraday cup detectors was performed each day. Liquid nitrogen was added to a cryopump each day to maintain a stable working vacuum. Using a variable leak valve, high-purity oxygen was bled into the source can in an attempt to promote the formation and ionization of MoO3. The O2 partial pressure was ~ 1.73 × 10−7 mbar above the baseline pressure with liquid nitrogen, such that the total pressure in the source can was typically 2.0–2.4 × 10−7 mbar. Using single filaments, standards were also analyzed with no O2 bled into the source, and with total pressures up to 4 × 10−7 mbar. With double filaments, total pressures up to 7.4 × 10−7 mbar were tested. This testing did not definitively show that Mo ionization correlates with the amount of O2 that is used. For their single filament method, Nagai and Yokoyama [13] observed that the best results were obtained without bleeding O2 into the source.

The heating protocol for the filaments was consistent for all standards and samples. Because of the close proximity of the two filaments, the ionization and evaporation filaments were initially heated slowly to avoid warping the Re ribbon. The ionization filament was heated first at 50 mA/min to 1000 mA, then at 20 mA/min to 1700 mA, and then at 10 mA/min to 1930 mA, in order to gently warm the evaporation filament. When the ionization filament reached 1700 mA, the evaporation filament was heated at 30 mA/min to 500 mA, then at 20 mA/min. Both filaments were continuously heated, with the rate gradually being decreased during the heating procedure until the ionization and evaporation filaments were heated at 0.1 mA/min and 0.1–0.2 mA/min, respectively. Once the ionization filament reached ~1930 mA and the evaporation filament reached ~630 mA, a mV-level signal was typically registered on the Faraday cups. During the warm-up of samples and standards, the intensity of the signal usually went through periods during which it decreased substantially until the evaporation filament reached 1200–1500 mA. The peak shapes during this period were also usually asymmetrical, likely due to overlapping peaks from ribbon-derived Mo that was not equilibrated with the sample or standard. After the evaporation filament heating current surpassed 1200–1500 mA and the intensity of the signal on 98Mo16O3 (m/z ~146) reached 300–400 mV, the peak shape typically became symmetrical for samples and standards. Once the signal intensity reached 800–1000 mV on 98Mo16O3, the analyses were started. This signal intensity typically corresponded to a current of 1650–1800 mA and 2080–2140 mA applied to the evaporation filament and ionization filament, respectively. At the beginning of analyses for samples, the signal often decreased for the duration of ~50–100 ratios, but would then stabilize or begin to grow. During a single analysis, lasting ~13 h, the typical signal intensity for 98Mo16O3 ranged between 1000 mV and 1600 mV.

Molybdenum was measured as xMoO3 in a static single-line measurement routine. Seven Faraday cup detectors, which were electronically connected to amplifiers equipped with 1011 Ω resistors, were used to collect the xMoO3 beams, with 96Mo16O3 in the center cup (m/z ~ 144–Table 2). An additional Faraday cup (H4) was electronically connected to an amplifier equipped with a 1012 Ω resistor. This detector was used to collect the 100Mo18O16O2 beam, which was used for the in-situ oxide correction. Later in the development of the analytical method, the L3 cup and a 1012 Ω resistor were used to collect the 92Mo17O16O2 beam, as discussed below. The amplifiers equipped with 1011 Ω resistors were electronically rotated to remove amplifier bias using the virtual amplifier capability of the Triton.

Table 2.

Faraday cup configuration for Mo analyses. The resistors equipped to the amplifiers for each cup and the trioxide interferences are also shown.

L4 L3c L2 L1 C H1 H2 H3 H4
Resistor (Ω)a 1011 1012 1011 1011 1011 1011 1011 1011 1012
Species 92Mo16O3 92Mo17O16O2 94Mo16O3 95Mo16O3 96Mo16O3 97Mo16O3 98Mo16O3 100Mo16O3 100Mo18O16O2
Interfering Trioxideb 92Mo50O3 92Mo51O3 92Mo52O3 94Mo51O3 94Mo52O3 96Mo52O3 98Mo52O3
94Mo49O3 94Mo5003 95Mo50O3 95Mo51O3 97Mo51O3
95Mo49O3 96Mo49O3 96Mo50O3 98Mo50O3
97Mo49O3
a

Only amplifiers equipped with 1011 Ω resistors were rotated during analyses to remove amplifier bias.

b

Trioxide masses are the total mass of the O3. For instance, 92Mo50O3 is composed of both 92Mo18O16O2 and 92Mo17O216O.

c

L3 was used to measure 92Mo17O16O2 from standard 35 onward, but was not used to calculate 17O/16O due to the lower internal precision, relative to the 18O/16O measurement.

Although Zr tends to form positive ions (e.g., [31]), and Ru, which forms negative trioxides, does not ionize on Re filaments at the running temperature of Mo [24], potential minor isobaric interferences from Zr and Ru were monitored by scanning the low and high masses surrounding the mass range of xMoO3 (140–148) using the electron multiplier. No interferences were observed corresponding to Zr or Ru. Further, no unaccounted-for signals (those not associated with MoO2 and MoO4) were observed to be greater than ~ 10–200 cps, excluding a ~1000 cps peak at mass 155 which we attribute to LaO. Therefore, no corrections for isobaric interferences from species in the MoO3 mass range were applied, excluding oxide corrections.

The procedures for incorporating the use of 1012 resistors into the measurement routine were adapted from Koornneef et al. [32], Liu and Pearson [33], and Trinquier et al. [25]. Baseline measurement times of 1200 s, with a 30 s settling time, and idle times of 15 s were used. Due to the longer settling time of the 1012 Ω resistor, compared to 1011 Ω resistors, and the lower signal intensity of the 100Mo18O16O2 species (2–4 mV), the optimal time over which the isotope ratio measurements were integrated was 67.109 s. Much of the data for the standards and samples were collected using two 67.109 s integrations such that the isotope ratio measurements were integrated over a total of 134.218 s. It was subsequently determined, however, that this integration time was unnecessarily long, so it was halved and the number or ratios was doubled from standard 35 onward (Table 3). No difference in the measured ratios was evident with the long vs. short integration times, but the internal statistics improved slightly with the shorter integration times. Measurements were taken in 50 blocks with 10 cycles/block for a total of 500 ratios. Baselines were measured every 10 blocks and the pilot mass 96Mo16O3 was centered and automatically focused every 2 blocks.

Table 3.

Molybdenum isotopic compositions of repeated analyses of the Alfa Aesar Specpure® Mo plasma standard using the in situ line by line oxide correction and fractionation correction using the molecular trioxide mass and 98Mo/96Mo. Measured, not fractionation corrected, 18O/16O ratios, averaged over the course of each analysis, are also shown. The 2SE of each analysis is given in ppm.

Standard # 92Mo/96Mo 2SE 94Mo/96Mo 2SE 95Mo/96Mo 2SE 97Mo/96Mo 2SE 100Mo/96Mo 2SE 18O/16O
1 0.883168 9.5 0.552441 5.9 0.953191 4.5 0.573952 4.2 0.581460 6.6 0.002067
2 0.883225 9.3 0.552453 6.8 0.953205 3.9 0.573950 3.4 0.581472 5.8 0.002072
3 0.883106 6.6 0.552430 4.4 0.953188 2.9 0.573955 2.2 0.581446 3.7 0.002063
4 0.883189 20 0.552443 8.1 0.953197 5.0 0.573953 3.6 0.581465 8.4 0.002074
5 0.883174 8.2 0.552441 4.7 0.953198 3.9 0.573954 3.0 0.581452 4.5 0.002060
6 0.883169 6.3 0.552441 3.9 0.953193 3.0 0.573950 2.3 0.581459 3.5 0.002068
7 0.883211 6.8 0.552450 3.7 0.953206 3.3 0.573951 2.7 0.581468 4.6 0.002066
8 0.883176 11 0.552443 5.0 0.953196 3.3 0.573953 2.6 0.581457 4.8 0.002060
9 0.883189 18 0.552447 8.5 0.953201 5.5 0.573952 3.9 0.581462 8.1 0.002055
10 0.883195 5.6 0.552449 3.3 0.953203 2.8 0.573954 2.2 0.581460 3.4 0.002054
11 0.883135 8.9 0.552437 4.5 0.953195 2.9 0.573953 2.1 0.581450 3.9 0.002058
12 0.883148 8.7 0.552438 5.3 0.953192 3.7 0.573953 2.6 0.581456 4.6 0.002084
13 0.883129 12 0.552434 7.8 0.953185 4.9 0.573953 3.7 0.581455 6.8 0.002089
14 0.883113 7.6 0.552432 5.3 0.953188 3.7 0.573953 2.9 0.581451 4.8 0.002071
15 0.883262 8.9 0.552461 5.4 0.953206 3.9 0.573949 3.5 0.581481 5.0 0.002044
16 0.883197 7.6 0.552448 5.1 0.953198 3.6 0.573952 5.4 0.581473 5.0 0.002062
17 0.883164 10 0.552440 6.5 0.953191 3.9 0.573953 3.8 0.581468 5.8 0.002078
18 0.883180 6.7 0.552443 4.8 0.953190 3.2 0.573952 2.7 0.581469 4.1 0.002064
19 0.883238 9.9 0.552455 5.5 0.953203 4.1 0.573952 2.8 0.581478 5.8 0.002069
20 0.883211 8.7 0.552449 6.4 0.953192 4.7 0.573948 3.3 0.581479 5.9 0.002052
21 0.883186 12 0.552438 7.6 0.953191 6.3 0.573951 5.0 0.581476 7.5 0.002056
22 0.883189 6.9 0.552443 4.8 0.953194 3.6 0.573949 3.1 0.581471 4.3 0.002072
23 0.883194 9.5 0.552441 6.4 0.953191 4.1 0.573950 3.4 0.581476 5.5 0.002053
24 0.883135 7.6 0.552432 5.1 0.953186 3.5 0.573954 3.1 0.581465 5.1 0.002055
25 0.883182 12 0.552444 9.4 0.953186 5.8 0.573952 5.5 0.581479 9.5 0.002058
26 0.883154 5.0 0.552435 3.9 0.953186 2.5 0.573952 2.4 0.581471 3.2 0.002062
27 0.883189 7.5 0.552446 4.1 0.953197 4.1 0.573954 3.7 0.581475 5.2 0.002043
28 0.883138 9.8 0.552434 5.4 0.953188 4.2 0.573952 3.0 0.581465 5.1 0.002056
29 0.883140 6.4 0.552436 3.6 0.953187 3.8 0.573953 2.9 0.581465 4.5 0.002057
30 0.883108 7.0 0.552427 4.1 0.953177 3.1 0.573952 2.4 0.581459 4.2 0.002067
31 0.883093 8.0 0.552424 4.4 0.953175 2.9 0.573954 2.5 0.581460 4.3 0.002064
32 0.883149 8.3 0.552438 5.7 0.953185 4.3 0.573953 3.1 0.581468 5.1 0.002047
33 0.883218 9.9 0.552450 7.4 0.953200 4.5 0.573951 3.9 0.581473 6.4 0.002041
34 0.883100 6.7 0.552428 4.9 0.953176 3.3 0.573954 2.5 0.581461 4.2 0.002059
35 0.883112 5.0 0.552430 3.4 0.953182 2.4 0.573953 1.9 0.581454 3.1 0.002055
36 0.883074 4.7 0.552420 3.3 0.953172 2.4 0.573954 1.8 0.581454 3.1 0.002055
37 0.883200 5.4 0.552444 3.8 0.953189 2.8 0.573951 2.2 0.581475 3.3 0.002056
38 0.883158 5.5 0.552437 3.6 0.953183 2.5 0.573953 2.0 0.581469 3.1 0.002059
39 0.883170 4.9 0.552439 3.8 0.953184 2.5 0.573953 2.0 0.581468 3.4 0.002054
40 0.883155 5.1 0.552437 3.3 0.953184 2.3 0.573952 1.8 0.581465 3.1 0.002053
41 0.883066 5.1 0.552418 3.5 0.953169 2.5 0.573953 2.0 0.581453 3.2 0.002061
42 0.883089 5.8 0.552422 3.8 0.953167 2.5 0.573952 1.8 0.581462 3.3 0.002058
43 0.883090 5.9 0.552424 4.0 0.953170 2.8 0.573951 2.5 0.581452 3.7 0.002056
44 0.883102 5.6 0.552424 4.0 0.953170 2.9 0.573951 2.2 0.581458 3.8 0.002060
45 0.883055 6.6 0.552413 4.6 0.953163 3.5 0.573951 2.9 0.581451 4.4 0.002059
46 0.883172 7.4 0.552437 5.6 0.953181 3.6 0.573950 2.7 0.581475 4.9 0.002049
47 0.883194 5.3 0.552442 3.6 0.953186 2.8 0.573950 2.2 0.581480 3.2 0.002042
48 0.883138 7.1 0.552432 4.6 0.953174 3.2 0.573950 2.3 0.581469 4.0 0.002055
Average 0.883157 0.552438 0.953188 0.573952 0.581465 0.002060
2SD (ppm) 107 37 23 5.4 32 0.96%

3.3. Fractionation correction

The isotope ratios were oxide corrected, as described below, and then fractionation corrected for natural and instrumental mass-dependent fractionation using the exponential law [34]. Instrumental mass-dependent fractionation in thermal ionization is largely due to the preferential evaporation of the lighter isotopes. Traditionally, the normalizing ratio of 98Mo/96Mo has been used [4,7], where 98Mo/96Mo = ≡1.453171 [35]. The exponential law is commonly used to correct a given isotope ratio (Ryx) by normalizing to a second ratio (Ryx) of two stable and non-radiogenic isotopes [34], where:

Rwyxy=(Ryx×Myx)β

and Rwyxy is the fractionation corrected ratio, Ryx is the oxide-corrected, measured ratio, Myx is the isotope mass ratio of the isotopes being fractionation corrected, and β is the fractionation factor:

β=ln(Ryw/Ry  Acceptedw)ln(Myw)

where Ryw is the measured, oxide-corrected ratio, RyAcceptedw is the accepted value (98Mo/96Mo = 1.453171), and Myw is the mass ratio of the isotopes used for normalization. Because Mo is measured as a trioxide, the molecular mass ratios were used in addition to the isotope mass ratios (e.g., 140/144 vs. 92/96). These results are compared below (Table 4). The molecular mass was required to fractionation correct the 100Mo18O16O2 species because it has no corresponding isotope mass, as 100Mo is the heaviest Mo isotope.

Table 4.

Comparison of the various data reduction methods. The best external precision (2SD) is obtained for the 97Mo/96Mo ratio using the line by line oxide correction and the molecular mass fractionation correction.

Fractionation correction Oxygen correction External Precision (ppm)
92Mo/96Mo 94Mo/96Mo 95Mo/96Mo 97Mo/96Mo 100Mo/96Mo
Isotope mass Nier 175 73 45 67 90
Measured single value 169 50 34 16 42
Line by line 166 56 30 6.7 45
Line by line (FC) 158 76 21 20 76
Molecular mass Nier 117 63 40 66 87
Measured single value 110 31 28 15 35
Line by line 107 37 23 5.4 32
Line by line (FC) 100 54 17 23 55

3.4. Oxide corrections

Because Mo was measured as a trioxide, the intensity measured for each xMoO3, except the lowest mass 92Mo16O3, must be corrected for isobaric interferences from the heavier oxygen isotope trioxides of lower mass Mo isotopes (Table 2). For example, 94Mo16O3 at mass 142 must be corrected for the interference caused by 92Mo18O16O2 at mass 142. Other oxide species which generate significant interferences are Mo17O 16O2, Mo17O2 16O, Mo18O2 16O , Mo18O 17O 16O, Mo17O3, and Mo17O218O. Additional species (Mo17O18O2 and Mo18O3) were also formed, but in such low abundance that they do not generate significant interferences [36]. The relative abundances of the oxide species were calculated as in Creaser et al. [37] and Harper and Jacobsen [38] using the isotopic composition of oxygen.

Oxide corrections commonly have been applied using a single oxygen isotopic composition, such as that from Nier [39] (17O/16O = 0.0003749 and 18O/16O = 0.0020439), for every sample or standard. However, systematic variations in the isotope ratios from standard to standard are likely due to variability in the oxygen isotopic composition, as has been the case for other measurements of Mo by N-TIMS [13] and for measurements of Os, B, W, and Ru by N-TIMS (e.g., [22,26,1,37,7,39]). Therefore, 1012 Ω resistors were installed in order to measure 100Mo18O16O2, and later 92Mo17O16O2, from which the oxygen isotope composition can be calculated. The 1012 Ω resistor reduces the signal to noise ratio and is most advantageous for signals between ~5 and 20 mV (e.g., [32]), which is on the order of the typical signal intensity for 100Mo18O16O2 (2–4 mV). The ratio of the intensities of 100Mo18O16O2/100Mo16O3 divided by 3 gives the 18O/16O ratio. By dividing by 3, the different permutations that 18O can be configured in a trioxide are accounted for [36].

The methods for measuring oxygen in situ were adapted from Liu et al. [19] and Luguet et al. [20]. In agreement with Liu et al. [19] and Luguet et al. [20], the 18O/16O and 17O/16O ratios were observed to increase over the course of a run, and the average 18O16O and 17O/16O ratio was found to vary from standard to standard (Table 3) and sample to sample. Because of the changing oxygen isotope ratio within analyses, the oxide correction was done for each cycle (referred to as line by line oxide correction). Data were also corrected using the oxygen isotopic composition of Nier [39] (Nier oxide correction), the measured oxygen isotopic composition averaged over the entire run (measured single value oxide correction), and the line by line oxygen correction with “fractionation corrected” 100Mo18O16O2 (line by line (FC) oxide correction).

For the measured single value, line by line, and line by line (FC) oxide corrections, the data reduction required two iterations [20]. The first was to correct the measured 100Mo18O16O2 intensity for the interference from 98Mo18O 162O using the oxygen isotopic composition of Nier [39], termed the first-order oxide correction. The second iteration used the average oxide-corrected 100Mo18O16O2 intensity for the measured single value oxide correction, or the oxide-corrected 100Mo18O16O2 intensities for each line for the line by line oxide correction to calculate the 18O/16O ratio(s). For the line by line (FC) oxide correction, the oxide-corrected 100Mo18O16O2 was fractionation corrected as described above.

The 17O/16O was calculated from the 18O/16O ratio using the slope of the terrestrial fractionation line (TFL). The 17O/16O was also measured in situ from standard 35 onward. However, the signal intensity for the 92Mo17O16O2 species was at the limit of the effective range of the 1012 Ωresistor (<2 mV), and the internal precision of these measurements was not as precise as the measurement of 18O/16O. A 1013 Ω resistor, with a lower effective range (~ 0.3–20 mV – [41]), may be advantageously employed to measure this species, but this has not yet been tested. Only negligible differences in the external reproducibility of the 14 standards were observed when the oxide correction was done using the calculated vs. measured 17O/16O, although the absolute values of the Mo isotope ratios were different (Table 5). To be consistent and to characterize the long-term external reproducibility, only Mo isotope data that was oxide corrected using the calculated 17O/16O is reported here.

Table 5.

Molybdenum isotopic compositions of standards 35–48, where the oxide corrections were done using the measured 17O/16O, as opposed to the 17O/16O calculated using the TFL. Also shown are the measured, non-fractionation corrected 17O/16O and 18O/16O, averaged over the course of the analysis and the average results of standards 35–48 calculated using the TFL-calculated 17O/16O..

Standard # 92Mo/96Mo 2SE 94Mo/96Mo 2SE 95Mo/96Mo 2SE 97Mo/96Mo 2SE 100Mo/96Mo 2SE 17O/16O 18O/16O
Measured 17O/16O
35 0.883175 5.1 0.552459 3.4 0.953203 2.4 0.573932 1.9 0.581451 3.1 0.0003872 0.002055
36 0.883136 4.7 0.552448 3.3 0.953193 2.4 0.573934 1.8 0.581451 3.1 0.0003871 0.002055
37 0.883262 5.4 0.552472 3.8 0.953210 2.8 0.573930 2.2 0.581472 3.3 0.0003870 0.002056
38 0.883219 5.6 0.552465 3.7 0.953203 2.5 0.573932 2.0 0.581466 3.1 0.0003872 0.002059
39 0.883231 4.9 0.552467 3.8 0.953205 2.5 0.573933 2.0 0.581465 3.4 0.0003869 0.002054
40 0.883218 5.1 0.552466 3.3 0.953205 2.3 0.573932 1.8 0.581462 3.1 0.0003869 0.002053
41 0.883128 5.1 0.552446 3.5 0.953190 2.5 0.573933 2.0 0.581450 3.2 0.0003876 0.002061
42 0.883152 5.9 0.552450 3.8 0.953188 2.5 0.573931 1.8 0.581459 3.3 0.0003875 0.002058
43 0.883151 5.9 0.552452 4.0 0.953191 2.8 0.573930 2.5 0.581449 3.7 0.0003872 0.002056
44 0.883163 5.6 0.552452 4.0 0.953190 2.8 0.573930 2.3 0.581455 3.8 0.0003875 0.002060
45 0.883117 6.6 0.552441 4.6 0.953184 3.4 0.573930 2.8 0.581448 4.4 0.0003875 0.002059
46 0.883233 7.5 0.552465 5.6 0.953201 3.7 0.573930 2.7 0.581471 4.9 0.0003865 0.002049
47 0.883256 5.3 0.552470 3.6 0.953207 2.8 0.573929 2.3 0.581476 3.2 0.0003858 0.002042
48 0.883201 7.2 0.552461 4.6 0.953195 3.2 0.573929 2.3 0.581466 4.0 0.0003874 0.002055
Average 0.8831S9 0.552458 0.953198 0.573931 0.581460 0.0003871 0.002055
2SD (ppm) 110 36 17 5.2 33 0.25% 0.49%
TFL calculated17O/16O
Average 0.883127 0.552430 0.953177 0.573952 0.581463
2SD (ppm) 111 36 18 5.2 33

4. Results and discussion

The standard data for all Mo isotopes using the preferred data reduction method is shown in Table 3 and Fig. 1. The external reproducibilities of the standards for the 97Mo/96Mo measurement, which is the Mo isotope ratio that is measured to the best statistical precision, using the various oxide and fractionation correction methods are compared in Table 4. The magnitudes of the oxide interferences for each Mo isotope are presented in Table 6. The results for spike-standard mixtures and samples are reported in μ notation (Tables 7 and 8), which gives the deviation of xMo/96Mo of a sample relative to the average xMo/96Mo of the repeated analyses of the Alfa Aesar Mo plasma standard over a given period of time, in parts per million. For example, for μ97Mo:

μ97Mosample=[(97Mo96Mo)sample(97Mo96Mo)standard1]×106

Fig. 1.

Fig. 1.

The μ97Mo for repeated analyses of the Alfa Aesar Mo plasma standard and the gravimetrically prepared spike-standard mixture. The collection period was 13 months. Data were reduced using in situ 18O/16O measurements and the line by line oxide correction. Data were fractionation corrected with 98Mo/96Mo = 1.453171 [35] and using the molecular mass instead of the isotope mass as described in text. Dotted lines represent two standard deviations of the analyses (2SD). For standards, 2SD = ± 5.4 ppm and for spike standard mixes, 2SD = ±2.9 ppm. The spike-standard mixture was gravimetrically predicted to have a μ97Mo value of + 16.7 ppm. This compares favorably with the measured ratio of μ97Mo = + 15.4 ± 2.9. Error bars for the individual analyses are the internal standard error of the mean (2SE).

Table 6.

The magnitude of the oxide interference corrections for each Mo isotope, given as percent relative to the oxide-corrected signal intensity of the Mo trioxide. Also shown are the percentages of the most abundant 17O- and 18O-bearing species relative to the total oxide interference signal. Other species incorporating 17O and 18O are minor and not taken into account for these calculations, such that the totals are generally not 100%.

94 95 96 97 98 100
Total oxide signal (% relative to oxide-corrected signal) 0.99 0.066 0.45 1.2 0.47 1.5
Total 17O16O2 signal (%) - 99.3 23.9 16.1 9.5 -
Total 18O16O2 signal (%) 100.0 - 75.9 83.9 90.4 99.8
Total 100.0 99.3 99.7 100.0 99.9 99.8

Table 7.

μ97Mo values for the spike-standard mixes. Mixes 6 and 7 were analyzed with only 400 ng of Mo, whereas mixes 1–5 were analyzed with 1000 ng of Mo.

μ97Mo 2SE
16.7 ppm mix 1 16.4 2.0
16.7 ppm mix 2 13.8 2.3
16.7 ppmmix3 16.5 1.9
16.7 ppm mix 4 15.4 1.8
16.7 ppm mix 5 17.5 2.2
16.7 ppm mix 6 13.9 3.1
16.7 ppm mix 7 14.3 2.0
Average 15.4 2.9 (2SD)

Table 8.

Molybdenum isotopic compositions of natural samples (in ppm). Different pieces of the same meteorite are numbered and different analyses of the same solution from the same piece of meteorite are numbered and lettered.

Samplesa μ92Mo 2SD μ94Mo 2SD μ95Mo 2SD μ97Mo 2SD μ100Mo 2SD
Toluca 1 −15 111 2.6 36 −9.7 18 2.0 5 3.2 33
Toluca 2 12 90 9.1 29 −5.9 15 1.3 6 0.8 24
Toluca 3 30 92 19 29 −3.9 14 0.02 5 0.4 34
Toluca mean 9 45 10 17 −6.5 6 1.1 2 1.46 3.0
Skookum la 302 92 173 29 116 14 51 5 119 34
Skookum lb 278 92 170 29 114 14 55 5 107 34
Skookum 2a 267 92 164 29 109 14 54 5 109 34
Skookum 2b 146 90 123 29 97 15 57 6 63 24
Skookum mean 248 139 157 46 109 17 54 5 100 50
Allegan 43 111 52 36 22 18 19 5.2 21 33
a

Toluca 1 and Allegan were processed using the procedures described here. Toluca 2 and Skookum 2 were digested and processed using the same methods, with the exception that these pieces were loaded onto the primary column using 0.4 M HCl-0.5 M HF as the loading solution. Toluca 3 was obtained as a byproduct of the chemistry used to separate and purify W [40], and was purified using the secondary anion column described here. Skookum 1 was processed using a different primary anion column that was more similar to the secondary column [26].

The best external reproducibility (2SD) for all oxide (except Nier, as described below) and fractionation correction methods (molecular vs. isotope mass) was obtained for 97Mo because it is bracketed by the normalizing ratio isotopes.

4.1. Oxide corrections comparison and molecular vs. isotopic fractionation correction

The average 18O/16O measured for standards was0.002060 ± 0.000020 (2SD, n = 48), which is within uncertainty of the ratio reported by Nagai and Yokoyama [13] (0.002064 ± 0.000005, n = 21). The long-term external reproducibility of 97Mo/96Mo using the Nier, measured single value, line by line, and line by line (FC) oxide corrections and the molecular mass fractionation correction were ± 66 ppm, 15 ppm, 5.4 ppm, and 23 ppm, respectively (Table 4). For the Nier oxide correction, 95Mo had the best precision of the Mo isotopes (40 ppm). This effect is likely because the 95MoO3 beam has the smallest oxide interference compared to the other Mo isotope beams, especially 97MoO3 and 100MoO3 (Table 6). Therefore, the difference between the Nier [39] and measured oxygen isotopic compositions was magnified with the Nier oxide correction on 97MoO3, compared to 95MoO3. All oxide corrections using the in situ measured oxygen isotopic composition resulted in improved precision relative to the Nier oxide correction, demonstrating that the variability of the oxygen isotopic composition between analyses adversely affects precision. Further, better precision for 97Mo/96Mo resulted from using the line by line correction vs. the measured single value correction, which demonstrates that the variability of the oxygen isotopic composition within a single analysis also affects external precision. The line by line (FC) correction exhibits worse external reproducibility for most Mo isotopes, including 97Mo/96Mo, relative to the non-fractionation corrected line by line method. This is likely because the fractionation correction using the 100Mo18O16O2 species does not sufficiently account for oxygen isotope fractionation that originates in the tank from which O2 is bled into the source can, or during formation of the trioxide [20]. Therefore, the preferred oxide correction method is the line by line correction.

The fractionation correction of Mo using the isotope mass, as opposed to the molecular mass, gives a long-term external reproducibility of 6.7 ppm for 97Mo/96Mo. The reason for the difference between the two methods is unclear; however, the discrepancy may indicate that Mo forms the molecular trioxide before evaporation on the filament fractionates Mo. Thus, using the molecular mass for fractionation correction provides a more accurate correction.

The best external reproducibility for 97Mo/96Mo was obtained using the line by line oxide correction and the fractionation correction using the molecular mass. Using these correction methods, the long-term external reproducibility (2SD) over 13 months of periodic measurement campaigns (Fig. 1 and Table 3, n = 48) was ±107 ppm for 92Mo, 37 ppm for 94Mo, 23 ppm for 95Mo, 5.4 ppm for 97Mo, and 32 ppm for 100Mo. The short-term external reproducibility for one ~2 week period (n = 8) was ± 112, 37, 17, 3.7, and 28 ppm for 92Mo, 94Mo, 95Mo, 97Mo, and 100Mo, respectively. The internal precision (2SE) was typically <4 ppm for 97Mo. The findings that a line by line oxide correction and fractionation correction using the molecular mass produce the most precise measurements of the Mo isotope ratios is in agreement with Nagai and Yokoyama [13], who used a single filament assembly and measured both 17O/16O and 18O/16O from the 100Mo17O16O2 and 100Mo18O16O2 species using 1011 Ω resistors. However, the external reproducibility of 97Mo/96Mo reported in this study is improved by a factor of ~2.4 relative to Nagai and Yokoyama [13], the precision of 100Mo/96Mo is roughly equivalent to slightly better on the short term, and the other isotope ratios are less precise (by a factor of ~2.3).

The methods reported here produce more or equivalently precise ratios for the heavy Mo isotopes and less precise ratios for the lighter Mo isotopes, relative to Nagai and Yokoyama [13]. The reason for this is unknown, but may be related to the differences in fractionation of Mo on single vs. double filaments and/or to the measurement vs. calculation of the oxygen isotope ratios. Nagai and Yokoyama [13] reported an absolute range in the fractionation factor, β, of ~0.6, similar to the range we obtained using single filaments. As noted above, the range in β over the course of an analysis is ~0.3 for double filaments. Therefore, if fractionation of Mo isotopes were the dominant cause for the difference in precision between this study and Nagai and Yokoyama [13], one might expect that more strongly fractionated measurements would have poorer precision on the light Mo isotopes. This is because the light Mo isotopes are evaporated more readily than the heavier isotopes and are farther in atomic mass units from the normalizing ratio, which could cause the fractionation correction to be inadequate. However, this is not observed, likely because Mo fractionation is coupled with the effects of the fractionating oxygen isotope composition, since the trioxides evidently form before Mo is fractionated by evaporation.

The magnitude of the oxide correction is largest on 97Mo and 100Mo, and the contribution of 18O to the total oxide interference signal becomes larger relative to 17O from 95Mo to 100Mo (Table 6). Because this study utilized a 1012 Ω resistor to measure the 100Mo18O16O2 species, the variance of the 18O/16O ratio may be better accounted for relative to measurements using 1011 Ω resistors [13]. Therefore, the oxide correction applied here for 97Mo and 100Mo, which is larger in magnitude than for other isotopes, may be more accurate, resulting in higher precision for these isotope ratios.

It is possible that some of the low mass Mo isotopes (i.e., 95Mo and 96Mo – Table 6) are less accurately corrected, and therefore less precise, because the contribution of species incorporating 17O is larger on these isotopes than on the heavier Mo isotopes. Because 96Mo is affected by this, and because the data are normalized to 96Mo, the effect of an inaccurate correction on 96Mo may be magnified on 94Mo and 92Mo, which are far from the normalizing ratio in terms of atomic mass units. The low mass Mo isotopes are measured less precisely in this study when both the measured 17O/16O and the calculated 17O/16O are used for the oxide correction. This may be because of the lower internal precision of the 17O/16O measurements or that the canonical TFL slope is not the optimal slope by which to calculate 17O/16O in this system. Bermingham et al.[7] and Trinquier [42] found that the oxygen isotopic compositions of RuO3 and U+/UO2+, respectively, do not always fractionate along the canonical TFL slope (i.e., δ18O vs. δ17O = 0.52 or 18O/16O vs. 17O/16O = 0.0954). Instead, for RuO3, which is run in negative mode like Mo, oxygen fractionation may result from mixing isotopically distinct reservoirs of oxygen from the oxygen tank and from the filament.

To test for this possibility during analysis of Mo isotopes, a 97Mo-100Mo double spike was measured, such that sufficiently intense signals were generated for the 100Mo17O16O2 (2–4 mV) and 100Mo18O16O2 species (10–20 mV) when using 1012 Ω resistors (Fig. 2). For the longer of the two analyses, the slope of 18O/16O vs. 17O/16O was 0.1128 ± 0.015 (2SD), which is marginally higher than the canonical slope of 0.0954. This indicates that mixing of two isotopically distinct reservoirs of oxygen may result in an inaccurate calculation of 17O/16O when using the TFL. This effect is evidently minor, however, as the standard data that was oxide corrected using the measured 17O/16O produces nearly identical precision. Alternatively, the mixing effect may simply not be observed when 17O/16O is measured using 92Mo17O16O2 when its signal intensity is <2 mV. Because the effective range of 1013 Ω resistors includes signals of <2 mV, future methods using a 1013 Ω resistor to measure the 92Mo17O16O2 or 100 Mo17O16O2 species and a 1012 Ω or 1013 Ω resistor to measure the 100Mo18O16O2 will likely be necessary to optimize the precision on all the Mo isotope ratios. It is also worth noting that the intensity of the 92Mo17O16O2 species is higher than the 100Mo17O16O2 species, due to the larger abundance of 92Mo, and it requires no first-order oxide correction. Thus, the use of the 92Mo17O16O2 species to calculate the 17O/16O ratio is advantageous.

Fig. 2.

Fig. 2.

In-run 18O/16O vs. 17O/16O measured in situ for two Mo double spike analyses using the 100Mo17O16O2 and 100Mo18O16O2 species and 1012 Ω two resistors. The intensities were oxide corrected using the Nier first-order oxide correction. The average 18O/16O and 17O/16O measured in situ for 14 standards, where 17O/16O was measured using the 92Mo17O16O2 species and a 1012 Ω resistor, is shown as the black diamond (18O/16O = 0.002055 and 17O/16O = 0.0003871–Table 5). Also shown are the oxygen isotopic compositions of [39], the “best measured” value reported by IAPC ([43]; citing [44] and [45]), and the “Mean O” value reported by Nagai and Yokoyama [13]. The black line represents the terrestrial fractionation line, which is forced through the UMD mean O value and is calculated using the canonical slope of 0.52.

In addition to the in-run oxygen isotopic compositions of the double spikes, the average measured oxygen isotopic composition of the standards for which 17O/16O and 18O/16O were measured is shown in Fig. 2 (UMD mean O). The oxygen isotopic composition of the spikes and the standards are similar to the mean oxygen isotopic composition reported by Nagai and Yokoyama [13]. The UMD mean O is not similar to the Nier [39] value, or the value recommended by IUPAC. This dissimilarity suggests that the use of the Nier [39] value for the first-order correction may be inappropriate. However, when the UMD mean O is used for the first-order correction instead, the precision of each ratio and the isotope ratios, excluding 92Mo/96Mo, are within ~1–2 ppm of those calculated with the Nier [39] value.

4.2. Gravimetrically prepared spike-standard mixture

To test the accuracy of the methods described above, an isotopically-enriched mixture was analyzed in the same way as samples and standards using the line by line oxygen correction and molecular mass fractionation correction. The mixture was gravimetrically prepared using a 97Mo-enriched spike (Oak Ridge National Lab) and the Alfa Aesar Mo plasma standard. The concentration of the 97Mo-enriched spike solution was calibrated prior to preparing the mixture. Once the spike and standard were combined in the appropriate proportions, the mixture was allowed to equilibrate on a hotplate at ~120 °C overnight. The mixture was then evaporated to dryness and 2:1 HNO3:HCl was added and dried one time. Finally, the mixture was taken into solution in 6 M HCl to load onto the filaments. The predicted “anomaly” for the mixture was μ97Mo = + 16.7. Repeated measurements of the spike-standard mixture yielded μ97Mo = + 15.4 ± 2.9 (2 SD, n = 7), which is in excellent agreement with the predicted value (Table 7, Fig. 1). The very good reproducibility indicates that over short-term, differences in μ97Mo < 10 ppm may be resolved. It was suspected that Mo derived from the Re ribbon resulted in relatively large Mo signals that interfered with Mo signals from the samples and standards. To test for this, five of the analyses of the spike-standard mixture were for loads of ~1000 ng and two of the analyses were for loads of ~ 400 ng Mo. The smaller loads showed μ97Mo values within uncertainties of the larger loads, so any effect of the loading blank was not reflected in the measurements. Further, any differences in fractionation behavior caused by the different amounts of Mo on the filament were not evident. These data show that the measurement procedure and data reduction provide accurate and precise data that can be routinely used to resolve differences in μ97Mo of ~10–15 ppm.

4.3. Natural samples

Data for natural samples are shown in Table 8 and Fig. 3. One piece of Toluca and Allegan were digested and processed using the methods described above. Two other pieces of Toluca and two pieces of Skookum were processed during the development of the method using different chemical separation procedures; resulting in variable yields for some samples (see Table 8 for details).

Fig. 3.

Fig. 3.

The μ97Mo for the Toluca and Skookum iron meteorites, and the Allegan chondrite, compared to some literature values. For Toluca and Skookum, uncertainties are the 2SD of the separate analyses. For Allegan, the uncertainty is the 2SD external reproducibility of the standards, represented also by the grey bar (±5.4 ppm). The μ97Mo for all three samples are in good agreement with data from meteorites of the same groups [7]. The datum for Toluca is not within uncertainty of that of Yin et al. [5].

The Mo yield for the three digestions of Toluca ranged from ~15–55%. The yield for one Toluca digestion was low (15%), potentially due to loss of Mo during the loading stage of the primary column using the 0.4 M HCl-0.5 M HF loading solution as described above. The yields for the two digestions of Skookum were similar at around 45%. The μ97Mo values obtained for the three digestions of Toluca with variable yields are within uncertainty of one another and the Alfa Aesar Mo standard. The μ97Mo for Toluca is not within uncertainty of the Mo isotopic composition reported by Yin et al. [5], but is in good agreement with data reported by Dauphas et al. [4] and Burkhardt et al. [7] for meteorites from the IAB group. Likewise, μ97Mo for the two digestions of Skookum are in agreement with data reported for irons of the IVB group [5,4,7]. These data demonstrate that the analyses are not affected by variable yield or different chromatographic processing. The data for Allegan are in agreement with H5 chondrite data from Burkhardt et al. [7]. This indicates that the methods described above are applicable to metallic and silicate matrices.

5. Conclusions

The procedures for measuring Mo by N-TIMS using the Thermo-Fisher Triton Plus are shown to be accurate and precise. The use of a double filament assembly improved signal stability and duration over those achieved using single filaments. Repeated analyses of the Alfa Aesar Specpure® Mo plasma standard using the oxygen isotope composition measured in situ and a line by line oxide correction yields a long-term external reproducibility of ±5.4 ppm for μ97Mo. The repeated measurement of the gravimetrically prepared 97Mo spike-standard mixture yielded μ97Mo that is in agreement with the predicted value, which demonstrates the accuracy of the technique and that differences in μ97Mo ≥ 10 ppm can be routinely resolved. The precision for 97Mo/96Mo was increased by a factor of ~3.9 compared to ICP-MS [7], and by a factor of ~2.4 compared to other TIMS methods [13]. The analyses of iron meteorite and chondrite samples show that different dissolutions and chromatographic procedures do not produce analytical artifacts related to variable yields, blanks, or matrices.

Acknowledgements

This work was supported by NASA Cosmochemistry grant NNX13AF83G which is gratefully acknowledged. We also acknowledge the Smithsonian Institute National Museum of Natural History for providing samples. We thank Igor Puchtel and Richard Ash for assistance in the lab, analytical guidance, and discussion. We also thank Anne Trinquier and Greg Archer for helpful discussion of the oxide correction methods.

References

  • [1].Barling J, Arnold GL, Anbar AD, Natural mass-dependent variations in the isotopic composition of molybdenum, Earth Planet. Sci. Lett 193 (2001) 447–457. [Google Scholar]
  • [2].Siebert C, Nägler TF, von Blanckenburg F, Kramers JD, Molybdenum isotope records as a potential new proxy for paleoceanography, Earth Planet. Sci. Lett 211 (2003) 159–171. [Google Scholar]
  • [3].Anbar AD, Knab KA, Barling J, Precise determination of mass-dependent variations in the isotopic composition of molybdenum using MC-ICPMS, Anal. Chem 73 (2001) 1425–1431. [DOI] [PubMed] [Google Scholar]
  • [4].Dauphas N, Marty B, Reisberg L, Molybdenum evidence for inherited planetary scale isotope heterogeneity of the protosolar nebula, Astrophys. J 565 (2002) 640–644. [Google Scholar]
  • [5].Yin QZ, Jacobsen SB, Yamashita K, Diverse supernova sources of pre-solar material inferred from molybdenum isotopes in meteorites, Nature 415 (2002) 881–883. [DOI] [PubMed] [Google Scholar]
  • [6].Becker H, Walker RJ, Efficient mixing of the solar nebula from uniform Mo isotopic composition of meteorites, Nature 425 (2003) 152–155. [DOI] [PubMed] [Google Scholar]
  • [7].Burkhardt C, Kleine T, Oberli F, Pack A, Bourdon B, Wieler R, Molybdenum isotope anomalies in meteorites: constraints on solar nebula evolution and origin of the Earth, Earth Planet. Sci. Lett 312 (2011) 390–400. [Google Scholar]
  • [8].Lu Q, Masuda A, High accuracy measurement of isotope ratios of molybdenum in some terrestrial molybdenites, J. Am. Soc. Mass Spectrom 3 (1992) 10–17. [DOI] [PubMed] [Google Scholar]
  • [9].Turnlund JR, Keyes WR, Peiffer GL, Isotope ratios of molybdenum determined by thermal ionization mass spectrometry for stable isotope studies of molybdenum metabolism in humans, Anal. Chem 65 (1993) 1717–1722. [DOI] [PubMed] [Google Scholar]
  • [10].Wieser ME, De Laeter JR, Thermal ionization mass spectrometry of molybdenum isotopes, Int. J. Mass Spectrom 197 (2000) 253–261. [Google Scholar]
  • [11].Wieser ME, De Laeter JR, Molybdenum isotope mass fractionation in iron meteorites, Int. J. Mass Spectrom 286 (2009) 98–103. [Google Scholar]
  • [12].Giussani A, Hansen Ch., Nüsslin F, Werner E, Application of thermal ionization mass spectrometry to investigations of molybdenum absorption in humans, Int. J. Mass Spec. Ion Processes 148 (1995) 171–178. [Google Scholar]
  • [13].Nagai Y, Yokoyama T, Molybdenum isotopic analysis by negative thermal ionization mass spectrometry (N-TIMS): Effects on oxygen isotopic composition, J. Anal. At. Spectrom 29 (2016) 1090–1096. [Google Scholar]
  • [14].Lee DC, Halliday AN, Precise determinations of the isotopic compositions and atomic weights of molybdenum, tellurium, tin and tungsten using ICP magnetic sector multiple collector mass spectrometry, Int. J. Mass. Ion Processes 31 (1995) 35–46. [Google Scholar]
  • [15].Dauphas N, Reisberg L, Marty B, Solvent extraction, ion chromatography, and mass spectrometry of molybdenum isotopes, Anal. Chem 73 (2001) 2613–2616. [DOI] [PubMed] [Google Scholar]
  • [16].Burkhardt C, Hin RC, Kleine T, Bourdon B, Evidence for Mo isotope fractionation in the solar nebula and during planetary differentiation, Earth Planet. Sci. Lett 391 (2014) 201–211. [Google Scholar]
  • [17].Heumann KG, Eisenhut S, Gallus S, Hebeda EH, Nusko R, Vengosh A, Walczyk T, Recent developments in thermal ionization mass spectrometric techniques for isotope analysis, Analyst 120 (1995) 1291–1299. [Google Scholar]
  • [18].Wieser ME, De Laeter JR, Varner MD, Isotope fractionation studies of molybdenum, Int. J. Mass Spectr 265 (2007) 40–48. [Google Scholar]
  • [19].Liu Y, Huang M, Masuda A, Inoue M, High-precision determination of osmium and rhenium isotope ratios by in situ oxygen isotope ratio correction using negative thermal ionization mass spectrometry, Int. J. Mass Spectrom 173 (1998) 163–175. [Google Scholar]
  • [20].Luguet A, Nowell GM, Pearson G, 184Os/188Os and 186Os/188Os measurements by negative thermal ionization mass spectrometry (N-TIMS): Effects of interfering element and mass fractionation corrections on data accuracy and precision, Chem. Geol 248 (2008) 342–362. [Google Scholar]
  • [21].Aggarwal SK, Wang B-S, You C-F, Chung C-H, hodology for precise and accurate isotopic analysis of boron by negative thermal ionization mass spectrometry based on BO2− ions and using the 18O/16O ratio from ReO4− for internal normalization, Anal. Chem 81 (2009) 7420–7427. [DOI] [PubMed] [Google Scholar]
  • [22].Chatterjee R, Lassiter JC, High precision Os isotopic measurement using N-TIMS: Quantification of various sources of error in 186Os/188Os measurements, Chem. Geol 396 (2015) 112–123. [Google Scholar]
  • [23].Chu Z-Y, Li C-F, Chen Z, Xu J-J, Di Y-K, Guo J-H, High-precision measurement of 186Os/188Os and 187Os/188Os: isobaric oxide corrections with in-run measured oxygen isotope ratios, Anal. Chem 87 (2015) 8765–8771. [DOI] [PubMed] [Google Scholar]
  • [24].Bermingham KR, Walker RJ, Worsham EA, Refinement of high precision Ru isotope analysis using negative thermal ionization mass spectrometry, Int. J. Mass Spectrom 403 (2016) 15–26. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [25].Trinquier A, Touboul M, Walker RJ, High-precision tungsten isotopic analysis by multicollection negative thermal ionization mass spectrometry based on simultaneous measurement of W and 18O/16O isotope ratios for accurate fractionation correction, Anal. Chem 88 (2016) 1542–1546. [DOI] [PubMed] [Google Scholar]
  • [26].Pietruszka AJ, Walker RJ, Candela PA, Determination of mass-dependent molybdenum isotopic variations by MC-ICP-MS: An evaluation of matrix effects, Chem. Geol 225 (2006) 121–136. [Google Scholar]
  • [27].Scheiderich K, Helz GR, Walker RJ, Century-long record of Mo isotopic composition in sediments of a seasonally anoxic estuary (Chesapeake Bay), Earth Planet. Sci. Lett 289 (2010) 189–197. [Google Scholar]
  • [28].Nagai Y, Yokoyama T, Chemical separation of Mo and W from terrestrial and extraterrestrial samples via anion exchange chromatography, Anal. Chem 86 (2014) 4856–4863. [DOI] [PubMed] [Google Scholar]
  • [29].Rizo H, Walker RJ, Carlson RW, Touboul M, Horan MF, Puchtel IS, Boyet M, Rosing MT, Early Earth differentiation investigated through 142Nd, 182W, and highly siderophile element abundances in samples from Isua, Greenland, Geochim. Cosmochim. Acta 175 (2016) 319–336. [Google Scholar]
  • [30].Bandi WR, Buyok EG, Lewis LL, Melnick LM, Anion exchange separation of zirconium, titanium, niobium, tantalum, tungsten, and molybdenum, Anal. Chem 33 (1961) 1275–1278. [Google Scholar]
  • [31].Sahoo SK, Masuda A, Precise measurement of zirconium isotopes by thermal ionization mass spectrometry, Chem. Geol 141 (1997) 117–126. [Google Scholar]
  • [32].Koornneef JM, Bouman C, Schwieters JB, Davies GR, Use of 1012 Ω current amplifiers in Sr and Nd isotope analyses by TIMS for application to sub-nanogram samples, J. Anal. At. Spectrom 28 (2013) 749–754. [Google Scholar]
  • [33].Liu J, Pearson D, Rapid, precise and accurate Os isotope ratio measurements of nanogram to sub-nanogram amounts using multiple Faraday collectors and amplifiers equipped with 1012 Ω resistors by N-TIMS, Chem. Geol 363 (2014) 301–311. [Google Scholar]
  • [34].Russell WA, Papanastassiou DA, Tombrello TA, Ca isotope fractionation in the Earth and other solar system materials, Geochim. Cosmochim. Acta 42 (1978) 1075–1090. [Google Scholar]
  • [35].Lu Q, Masuda A, The isotopic composition and atomic weight of molybdenum, Int. J. Mass Spectrom. Ion Processes 130 (1994) 65–72. [Google Scholar]
  • [36].Yin QZ, N-TIMS Technique for the Re-Os and Ru Isotopic Systems and Its Application to Selected Geochemical and Cosmochemical Problems. PhD Thesis, University of Mainz and Max Planck Institute for Chemistry, 1995. [Google Scholar]
  • [37].Creaser RA, Papanastassiou DA, Wasserburg GJ, Negative thermal ion mass spectrometry of osmium, rhenium and iridium, Geochim. Cosmochim. Acta 55 (1991) 397–401. [Google Scholar]
  • [38].Harper CL, Jacobsen SB, Evidence for 182Hf in the early Solar System and constraints on the timescale of terrestrial accretion and core formation, Geochim. Cosmochim. Acta 60 (1996), 1311–1153. [Google Scholar]
  • [39].Nier AO, A redetermination of the relative abundances of the isotopes of carbon, nitrogen, oxygen, argon, and potassium, Phys. Rev 77 (1950) 789–793. [Google Scholar]
  • [40].Touboul M, Walker RJ, High precision tungsten isotope measurement by thermal ionization mass spectrometry, Int. J. Mass Spectrom 309 (2012) 109–117. [Google Scholar]
  • [41].Koornneef JM, Bouman C, Schwieters JB, Davies GR, Measurement of small ion beams by thermal ionization mass spectrometry using new 1013 Ohm resistors, Anal. Chim. Acta 819 (2014) 49–55. [DOI] [PubMed] [Google Scholar]
  • [42].Trinquier A, Fractionation of oxygen isotopes by thermal ionization mass spectrometry inferred from simultaneous measurement of 17O/16O and 18O/16O ratios and implications for the 182Hf −182W systematics, Anal. Chem 88 (2016) 5600–5604, 10.1021/acs.analchem.6b01101. [DOI] [PubMed] [Google Scholar]
  • [43].Meija J, Coplen TB, Berglund M, Brand WA, De Bièvre P, Gröning M, Holden NE, Irrgeher J, Loss RD, Walczyk T, Prohaska T, Isotopic compositions of the elements 2013 (IUPAC technical report), Pure Appl. Chem 88 (2016) 293–306. [Google Scholar]
  • [44].Baertschi P, Absolute 18O content of standard mean ocean water, Earth Planet. Sci. Lett 31 (1976) 341–344. [Google Scholar]
  • [45].Li W-J, Jin D, Chang T-L, Kexue Tinboa 33 (1988) 1610. [Google Scholar]

RESOURCES