Abstract
Aims/Introduction
It is reported that interfering substances in the blood might influence the value for measurement of active glucagon‐like peptide‐1 (GLP‐1) in human plasma. Solid phase extraction (SPE) pretreatment is recommended to reduce their influence, but it requires a lot of cost and time. However, there is little investigation about causative inhibitory substances and about methods that can replace solid phase extraction. In the present study, we aimed to seek the candidate of the substances that might interfere with an active GLP‐1 enzyme‐linked immunosorbent assay (ELISA).
Materials and Methods
Two kinds of active GLP‐1 ELISA kits using different antibodies, plural extraction carriers and elution solutions were used to evaluate the SPE method. Active GLP‐1 concentration was compared with or without SPE, and with or without a heterophilic blocking tube.
Results
Active GLP‐1 values were often higher without SPE compared with those with SPE pretreatment. This difference was eliminated by pretreatment with a heterophilic blocking tube or ELISA kits that did not use a mouse monoclonal antibody, and was independent of SPE.
Conclusions
Substances absorbed to a heterophilic blocking tube carrier might interfere with an active GLP‐1 immunoassay. Solid‐phase extraction treatment is required for measurement of active GLP‐1 by an ELISA kit affected by heterophilic antibodies.
Keywords: Active glucagon‐like peptide‐1, Immunoassay, Solid phase extraction
Introduction
Incretin hormones, namely, glucose‐dependent insulinotropic polypeptide (GIP) and glucagon‐like peptide‐1 (GLP‐1), are secreted from enteroendocrine cells on intake of nutrients and enhance insulin secretion in response to glucose1, 2, 3. Among the incretins, GLP‐1 shows an insulin secretion enhancing effect even in the diabetic state, and has inhibitory actions on glucagon concentrations4. After it is secreted from intestinal endocrine cells as an active peptide, GLP‐1 has a very short half‐life, as is rapidly inactivated by dipeptidyl peptidase‐4, which is present in the intestinal tract and blood5. Recently, dipeptidyl peptidase‐4 inhibitors, which increase the concentration of active incretin in the blood, were developed and are widely used clinically6. In this context, the measurement of total incretin in the blood, including active and inactivated forms, is important for evaluating incretin secretion from enteroendocrine cells after nutrient ingestion. In contrast, to evaluate the incretin effect, it seems to be more useful to measure active incretin concentrations. Therefore, it is important to measure active GLP‐1 when evaluating the effect of incretin‐based drugs.
Previous reports showed differences in active GLP‐1 concentrations in the human blood, suggesting the presence of substances that might affect the value based on the measurement system used7, 8. Subsequently, solid phase extraction (SPE) or ethanol extraction was recommended to eliminate this interference9, 10, 11. Furthermore, the active GLP‐1 values in the literature are comparable after extraction pretreatment12, 13, 14. In contrast, SPE is time‐consuming and expensive, which makes measuring active GLP‐1 less desirable. Few reports have evaluated SPE in terms of its influence on interfering substances.
In the present study, we confirmed that active GLP‐1 values in the plasma affected by interfering substances differ widely among individuals. Under different pretreatment conditions, we showed that these differences can be eliminated by SPE or by absorption of heterophilic antibodies. We therefore propose that these pretreatments are not required if using an immunoassay kit that is not affected greatly by heterophilic antibodies.
Methods
Selection of extraction column and elution solvent
Iodine‐labeled active GLP‐1; that is, [125‐I]GLP‐1 (7‐36 amide; catalog no. H‐6804; Peninsular Laboratories, LLC, Bachem Group, San Carlos, California, USA) with non‐labeled GLP‐1 (7‐36 amide; catalog no. 4344‐v; Peptide Institute, Inc., Osaka, Japan) adjusted to 20,000–30,000 c.p.m. was added to three types of extraction columns to select the appropriate column for SPE: (i) a C18 column (VARIAN Bond elute 200 mg; Agilent Technologies Japan Ltd., Tokyo, Japan); (ii) a C18 particle column (C18 125A; Wako, Osaka, Japan); and (iii) a silica particle column (Silica GelODS‐Q3; Waters Corporation, Milford, Massachusetts, USA). The radioactivity of each sample eluted using the same eluent, namely, 0.5% ammonia (v/v) and 75% ethanol (v/v), from the three columns was measured using a gamma counter (1470WIZARD Gamma Counter; Perkin Elmer Co., Ltd., Waltham, Massachusetts, USA), and the recovery ratios were compared (Table S1). For selection of the eluate, labeled active GLP‐1 was added to a C18 column and eluted with five different solutions (Table S2). Each eluate was measured with a gamma counter, and the recovery ratios were calculated.
SPE
As shown in Tables S1 and S2, the C18 column was selected as the separation column, and a 0.5% ammonia (v/v) and 75% ethanol (v/v) solution was selected as the eluant based on the high recovery rate. The procedure was carried out as follows: the C18 column was sequentially equilibrated with 2 mL of methanol and distilled water. Plasma samples, threefold diluted in phosphate‐buffered saline, were added to the column. The system was washed twice with 1 mL of 10% methanol, and samples were eluted twice with 0.75 mL of eluant. Eluted samples were dried under nitrogen gas flow at 40°C and re‐dissolved in assay buffer. Figure S1 shows the extraction protocol.
Plasma samples
A total of 63 human plasma samples collected into BD P700 Blood Collection Tubes (catalog no. 366473; Becton Dickinson, Franklin Lakes, New Jersey, USA) from 49 volunteers (33 participants with normal glucose tolerance, eight participants with impaired glucose tolerance and eight participants with type 2 diabetes mellitus15). The plasma was stored at −80°C until use. The study was designed in compliance with the considerations of the Helsinki Declaration and ethics committee affiliated with Kyoto University (registry no. C362). All participants provided informed consent.
Measurement of active GLP‐1
Active GLP‐1 was measured using two active GLP‐1 enzyme‐linked immunosorbent assay (ELISA) kits: (i) the Millipore ELISA kit (catalog no. EGLP‐35K; Millipore, Billerica, Massachusetts, USA) that has 100% cross‐reactivity with GLP‐1 (7–36) amide, 99.5% cross‐reactivity with GLP‐1 (7–37), 0.2% cross‐reactivity with GLP‐1 (1–36) amide and 0.2% cross‐reactivity with GLP‐1 (1–37), and does not detect GLP‐1 (9–36) amide and GLP‐1 (9–37), intra‐assay coefficient of variation (CV) 6–9%, interassay CV <1–13%, and percentage recovery in human plasma 78–94%; and (b) the IBL ELISA kit (catalog no. 27784; IBL, Gunma, Japan) that recognizes GLP‐1 isoforms, has 0.04% cross‐reactivity with GLP‐1 (1–37), 100% cross‐reactivity with GLP‐1 (7–37) and GLP‐1 (7–36) amide, and 0.02% cross‐reactivity with GLP‐1 (9–36) amide, and does not detect GLP‐1 (9–39), intra‐assay CV 4.5–5.5%; interassay CV 7.1–10.9%, and percentage recovery in human plasma 82.0–88.4%.
Comparison SPE and heterophilic antibody blocking
To investigate the effect of removing the heterophilic antibody, heterophilic blocking tubes (HBT; catalog no. 3IX762; Scantibodies Laboratory Inc., Santee, California, USA) were used. To compare the effects of SPE and HBT, sample pretreatments were carried out as follows: both pretreatments (SPE[+]HBT[+]), SPE alone (SPE[+]HBT[−]), HBT alone (SPE[−]HBT[+]) and no pretreatment (SPE[−]HBT[−]). In each case, active GLP‐1 was measured twice using each ELISA kit (Millipore ELISA kit and IBL ELISA kit). Thus, from the combinations of four pretreatments and two ELISA kits, eight different values were obtained for each sample (Figure 1).
Figure 1.

Comparison study of solid phase extraction and heterophilic antibody blocking. In each pretreatment condition, two kinds of active glucagon‐like peptide‐1 enzyme‐linked immunosorbent assay (ELISA) kits were used. (1), Millipore ELISA Kit; (2), IBL ELISA kit. HBT, heterophilic blocking tube.
Statistical analysis
Pearson's correlation test was used to test for correlations between the values. A P‐value of <0.05 was considered statistically significant.
Results
SPE and value of plasma active GLP‐1 concentration
We examined differences in the values of the active GLP‐1 concentration in the plasma obtained using the Millipore ELISA kit with or without SPE. Figure 2 shows a comparison of values obtained for the active GLP‐1 concentration in the plasma of each individual (Figure 2). Although the values after the extraction were generally lower compared with those without extraction, the difference was large between individuals, and the ratio of the measured value after extraction to that before extraction ranged widely from 0.27 to 1.19.
Figure 2.

Differences in the active glucagon‐like peptide‐1 values between measurements with and without solid phase extraction. Plasma active glucagon‐like peptide‐1 values of measurements using a Millipore (Billerica, MA, USA) enzyme‐linked immunosorbent assay kit directly or after solid phase extraction in each individual. White bars, without extraction; black bars, with solid phase extraction.
HBT pretreatment eliminates the difference of active GLP‐1 values by SPE
The Millipore ELISA kit includes a mouse monoclonal antibody, therefore, we focused on the effects of heterophilic antibodies, and examined the effect of HBT and ELISA kits using antibodies other than mouse monoclonal antibody. Table 1 shows the correlation of active GLP‐1 values measured with the two types of ELISA kits (Millipore and IBL) under various pretreatment conditions. Using the Millipore ELISA kit, active GLP‐1 values after SPE correlated significantly with those without extraction (r = 0.594, P < 0.0001; Figure 3a). A similar correlation was observed between active GLP‐1 values with and without HBT pretreatment (r = 0.628, P < 0.0001; Figure 3b). It should be noted that the correlation in the cross‐study of the pretreatments (SPE and HBT) was greatly increased between SPE(+)/HBT(−) and SPE(−)/HBT(+) (r = 0.940, P < 0.0001; Figure 3c).
Table 1.
Correlation of active glucagon‐like peptide‐1 values measured with the two types of enzyme‐linked immunosorbent assay kits (Millipore and IBL) under various pretreatment conditions
| Millipore kit | IBL kit | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Extraction (+) | Extraction (−) | Extraction (+) | Extraction (−) | |||||||
| HBT (+) | HBT (−) | HBT (+) | HBT (−) | HBT (+) | HBT (−) | HBT (+) | HBT (−) | |||
| Millipore kit | Extraction (+) | HBT (+) | 1 | 0.941 | 0.91 | 0.498 | 0.591 | 0.531 | 0.503 | 0.506 |
| P < 0.0001 | P < 0.0001 | P < 0.0001 | P < 0.0001 | P < 0.0001 | P < 0.0001 | P < 0.0001 | ||||
| HBT (−) | 0.941 | 1 | 0.94 | 0.594 | 0.643 | 0.606 | 0.579 | 0.582 | ||
| P < 0.0001 | P < 0.0001 | P < 0.0001 | P < 0.0001 | P < 0.0001 | P < 0.0001 | P < 0.0001 | ||||
| Extraction (−) | HBT (+) | 0.91 | 0.94 | 1 | 0.628 | 0.661 | 0.628 | 0.618 | 0.628 | |
| P < 0.0001 | P < 0.0001 | P < 0.0001 | P < 0.0001 | P < 0.0001 | P < 0.0001 | P < 0.0001 | ||||
| HBT (−) | 0.498 | 0.594 | 0.628 | 1 | 0.373 | 0.414 | 0.463 | 0.482 | ||
| P < 0.0001 | P < 0.0001 | P < 0.0001 | P = 0.0026 | P = 0.0008 | P = 0.0001 | P = 0.0001 | ||||
| IBL kit | Extraction (+) | HBT (+) | 0.591 | 0.643 | 0.661 | 0.373 | 1 | 0.957 | 0.883 | 0.9 |
| P < 0.0001 | P < 0.0001 | P < 0.0001 | P = 0.0026 | P < 0.0001 | P < 0.0001 | P < 0.0001 | ||||
| HBT (−) | 0.531 | 0.606 | 0.628 | 0.414 | 0.957 | 1 | 0.898 | 0.911 | ||
| P < 0.0001 | P < 0.0001 | P < 0.0001 | P = 0.0008 | P < 0.0001 | P < 0.0001 | P < 0.0001 | ||||
| Extraction (−) | HBT (+) | 0.503 | 0.579 | 0.618 | 0.463 | 0.883 | 0.898 | 1 | 0.98 | |
| P < 0.0001 | P < 0.0001 | P < 0.0001 | P = 0.0001 | P < 0.0001 | P < 0.0001 | P < 0.0001 | ||||
| HBT (−) | 0.506 | 0.582 | 0.628 | 0.482 | 0.9 | 0.911 | 0.98 | 1 | ||
| P < 0.0001 | P < 0.0001 | P < 0.0001 | P = 0.0001 | P < 0.0001 | P < 0.0001 | P < 0.0001 | ||||
HBT, heterophilic blocking tube.
Figure 3.

Effect of solid phase extraction (SPE) and heterophilic blocking tube (HBT) on active glucagon‐like peptide‐1 measurement by enzyme‐linked immunosorbent assay kit (Millipore, Billerica, MA, USA) using mouse monoclonal antibody detection. Relationships of measured plasma active glucagon‐like peptide‐1 values with or without pre‐treatments. (a), SPE(+) and SPE(−), (b), HBT(+) and HBT (−), and (c), SPE(+) and HBT(+); cross comparison of 63 independent samples from 49 individuals. Pearson's correlation tests were carried out to calculate the correlation coefficient (r) and P‐values.
ELISA kit that did not use a mouse monoclonal antibody eliminates the differences of active GLP‐1 values by SPE
In comparison among measurements with the IBL ELISA kit, which includes rabbit immunoglobulin G as the capture antibody, the correlations between active GLP‐1 values with and without the use of pretreatments increased dramatically; that is, SPE(+) vs SPE(−) (r = 0.911, P < 0.0001), HBT(+)vs HBT(−) (r = 0.980, P < 0.0001; Figure 4a,b). Similarly to the results obtained using the Millipore ELISA kit, the cross‐study of two pretreatments with the IBL ELISA kit showed a good correlation between SPE(+)/HBT(−)vs SPE(−)/HBT(+) (r = 0.898, P < 0.0001; Figure 4c).
Figure 4.

Effect of solid phase extraction (SPE) and heterophilic blocking tube (HBT) on active glucagon‐like peptide‐1 measurement by enzyme‐linked immunosorbent assay kit (IBL, Gunma, Japan) using rabbit antibody detection. Relationships of measured plasma active glucagon‐like peptide‐1 values with or without pre‐treatments. (a) SPE(+) and SPE(−), (b) HBT(+) and HBT (−), and (c), SPE(+) and HBT(+); cross comparison of 63 independent samples from 49 individuals. Pearson's correlation tests were carried out to calculate the correlation coefficient (r) and P‐values.
Discussion
In the present study, we confirmed the following: (i) by comparing extraction carriers and extraction solutions, we established an SPE method for measuring active GLP‐1 concentrations in the plasma; and (ii) active GLP‐1 values are overestimated because of the influence of interfering substances, and SPE pretreatment eliminates this interference. Additionally, pretreatment produces a large difference in active GLP‐1 values; and, (iii) by comparing values obtained by various combinations of pretreatments and ELISA kits, we showed that pretreatment is not required when using an ELISA kit that is not greatly affected by heterophilic antibodies. The reproducibility of these two kits in the present study is important as one of prerequisites for these observations. Considering the measured values of active GLP‐1 in this study, the performance of these kits can be considered to adequately withstand comparison of measurement values between various conditions.
Although measurement of active GLP‐1 with SPE pretreatment was reported previously9, 10, no previous studies have examined detailed methods for establishing SPE. Similar to previous reports7, 9, 10, a relatively higher value of active GLP‐1 was detected, which appeared to be due to interfering substances, and this was eliminated in the present study by SPE. This finding confirms that a pretreatment, such as SPE, to remove interfering substances is necessary for reliable measurement of active GLP‐116, 17, 18, 19. Furthermore, the use of an ELISA kit that is not greatly affected by heterophilic antibodies reduces the apparent difference between active GLP‐1 values with or without SPE and HBT treatment. This suggests that heterophilic antibodies, especially human anti‐mouse antibodies (HAMA), are responsible, at least in part, for the interference. In the Millipore ELISA kit, where mouse monoclonal antibody is used as a capture antibody, the measured values are affected depending on the HAMA in each individual. In contrast, when we used the IBL kit, in which a rabbit polyclonal antibody is used as the capture antibody, active GLP‐1 values were not affected by HAMA. Based on these characteristics of the two ELISA kits used in the present study, a difference in the active GLP‐1 values for each individual was observed with the Millipore kit, but not with the IBL kit, with and without HBT pretreatment. Therefore, HAMA is considered to be a strong candidate as a responsible interfering substance. Furthermore, active GLP‐1 measurement values obtained with HBT pretreatment, which results in the removal of mouse immunoglobulin G, mouse serum, non‐specific monoclonal antibodies and agglutinated immunoglobulin G, were similar to those with SPE pretreatment. This finding supports the possibility that the interfering substance is a heterophilic antibody, such as HAMA.
HAMA is a human antibody that recognizes mouse antibodies as an antigen and has long been known to be an interfering substance that affects measurement values in assay systems that use immunological techniques20. In the thyroid‐stimulating hormone immunoassays in which the mouse monoclonal antibody is used, it is well known that the measurement value is affected by HAMA. The mechanism that elicits pseudo high values is not different from the active GLP‐1 immunoassay. However, the clinical problem frequency in thyroid‐stimulating hormone measurement is quite low. One of the reasons is that the balance between plasma HAMA concentration and the amount of the antigen (hormone) to be measured is related to the interferences with the measurement value. The amount of active GLP‐1 that we try to see is very small, therefore, interference by HAMA with the measurement system could be particularly conspicuous. HAMA production is induced by various stimuli, including administration of mouse‐derived vaccines, which was a common practice when purification techniques were insufficient. HAMA is also produced after administration of molecular targeted drugs21. In the present study, we found a difference in measured values in many individuals, possibly due to heterophilic antibodies, even without any history of administration of molecular targeted drugs. It is difficult to prove that the interfering substance is HAMA, as it is necessary to determine the epitope of the interfering substance and to show that it is a common antigen in the measurement system. Nevertheless, HAMA is a strong candidate.
Notably, the results of the present study showed that the difference in the measured values before and after pretreatment for removal of interfering substances, carried out for the measurement of activated GLP‐1, were dramatically reduced by using an ELISA kit that is not greatly affected by heterophilic antibodies. This means that by using an ELISA kit that is not greatly affected by heterophilic antibodies, plasma active GLP‐1 values can be measured more reliably, without requiring pretreatment, such as SPE or HBT. SPE requires extensive equipment, is time‐consuming and expensive. In addition, SPE removes the impurities non‐specifically, which can be a disadvantage from the viewpoint of measurement stability. Taking these points into consideration, this finding is very important.
The present study had the following limitations. First, comparison of plasma active GLP‐1 values was carried out using two ELISA kits, but the specificity of the active GLP‐1 measurement of these kits themselves was not investigated. We also did not clarify which kit provides a more accurate measurement. The purpose of the present study, however, was to evaluate the SPE method and to identify substances that affect the measurement. Therefore, if the reproducibility in each kit is stable, the findings remain valid despite this limitation. Second, the samples were derived from a single facility. Further investigations with a larger number of participants from multiple facilities are required.
In studies of the physiology of glucose metabolism and the pathology of diabetes, there are many issues to be clarified regarding the secretion and actions of incretins2, 5, 13, 14. Stable, easy and inexpensive measurements of active GLP‐1 should be useful for studies of incretins.
Disclosure
Nobuya Inagaki receives fees from Takeda Pharmaceutical Co., Ltd., Mitsubishi Tanabe Pharma Corporation, Merck Sharp & Dohme (MSD), Sanofi, Novartis Pharma, Dainippon Sumitomo Pharma, Kyowa Hakko Kirin Co., Ltd., Eli Lily Japan, Shiratori Pharmaceutical, Roche Diagnostics, Japan Tobacco (JT), Nippon Boehringer Ingelheim Co., Ltd., Astellas Pharma Inc., Daiichi Sankyo Company, Ltd., Ono Pharmaceutical Co., Ltd. and Taisho Toyama Pharmaceutical Co., Ltd. outside the submitted work. Tomonori Hasegawa and Miho Komagata are employees of SRL Inc. The other authors declare no conflict of interest.
Supporting information
Figure S1 | Procedure of solid phase extraction. A C18 column was used as the extraction column. Samples were dissolved in the assay buffer for each enzyme‐linked immunosorbent assay kit.
Table S1 | Comparison of relative ratios of active glucagon‐like peptide‐1 values after treatment with extraction carriers.
Table S2 | Extraction rates of iodine‐labeled active glucagon‐like peptide‐1 when using various extract solutions.
Acknowledgments
This study was supported by grants from the Ministry of Education, Culture, Sports, Science and Technology (MEXT); the Japan Society for the Promotion of Science (JSPS), Ministry of Health, Labor and Welfare; the Ministry of Agriculture, Forestry, and Fisheries; and the Japan Diabetes Foundation.
J Diabetes Investig 2019; 10: 302–308
References
- 1. Nauck M, Stöckmann F, Ebert R, et al Reduced incretin effect in type 2 (non‐insulin‐dependent) diabetes. Diabetologia 1986; 29: 46–52. [DOI] [PubMed] [Google Scholar]
- 2. Bagger JI, Knop FK, Lund A, et al Impaired regulation of the incretin effect in patients with type 2 diabetes. J Clin Endocrinol Metab 2011; 96: 737–745. [DOI] [PubMed] [Google Scholar]
- 3. Harada N, Hamasaki A, Yamane S, et al Plasma gastric inhibitory polypeptide and glucagon‐like peptide‐1 levels after glucose loading are associated with different factors in Japanese subjects. J Diabetes Investig 2011; 2: 193–199. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Holst JJ. The physiology of glucagon‐like peptide 1. Physiol Rev 2007; 87: 1409–1439. [DOI] [PubMed] [Google Scholar]
- 5. Seino Y, Fukushima M, Yabe D. GIP and GLP‐1, the two incretin hormones: Similarities and differences. J Diabetes Investig 2010; 1: 8–23. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Deacon CF, Holst JJ. Dipeptidyl peptidase‐4 inhibitors for the treatment of type 2 diabetes: comparison, efficacy and safety. Expert Opin Pharmacother 2013; 14: 2047–2058. [DOI] [PubMed] [Google Scholar]
- 7. Dai H, Gustavson SM, Preston GM, et al Non‐linear increase in GLP‐1 levels in response to DPP‐IV inhibition in healthy adult subjects. Diabetes Obes Metab 2008; 10: 506–513. [DOI] [PubMed] [Google Scholar]
- 8. Gustavson SM, Dai H, Preston GM, et al Effects of multiple doses of the DPP‐IV inhibitor PF‐734200 on the relationship between GLP‐1 and glucose in subjects with type 2 diabetes mellitus. Diabetes Res Clin Pract 2011; 91: e45–e49. [DOI] [PubMed] [Google Scholar]
- 9. Yabe D, Watanabe K, Sugawara K, et al Comparison of incretin immunoassays with or without plasma extraction: incretin secretion in Japanese patients with type 2 diabetes. J Diabetes Investig 2012; 3: 70–79. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Masuda K, Aoki K, Terauchi Y. Comparison of plasma active glucagon‐like peptide‐1 (GLP‐1) levels assayed with or without plasma extraction in non‐diabetic men. Endocr J 2012; 59: 435–438. [DOI] [PubMed] [Google Scholar]
- 11. Kuhre RE, Wewer Albrechtsen NJ, Hartmann B, et al Measurement of the incretin hormones: glucagon‐like peptide‐1 and glucose‐dependent insulinotropic peptide. J Diabetes Complications 2015; 29: 445–450. [DOI] [PubMed] [Google Scholar]
- 12. Deacon CF, Holst JJ. Immunoassays for the incretin hormones GIP and GLP‐1. Best Pract Res 2009; 23: 425–432. [DOI] [PubMed] [Google Scholar]
- 13. Calanna S, Christensen M, Holst JJ, et al Secretion of glucagon‐like peptide‐1 in patients with type 2 diabetes mellitus: systematic review and meta‐analyses of clinical studies. Diabetologia 2013; 56: 965–972. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Calanna S, Christensen M, Holst JJ, et al Secretion of glucose‐dependent insulinotropic polypeptide in patients with type 2 diabetes: systematic review and meta‐analysis of clinical studies. Diabetes Care 2013; 36: 3346–3352. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15. Seino Y, Nanjo K, Tajima N, et al Report of the committee on the classification and diagnostic criteria of diabetes mellitus. J Diabetes Investig 2010; 1: 212–228. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16. Yabe D, Kuroe A, Watanabe K, et al Early phase glucagon and insulin secretory abnormalities, but not incretin secretion, are similarly responsible for hyperglycemia after ingestion of nutrients. J Diabetes Complications 2015; 29: 413–421. [DOI] [PubMed] [Google Scholar]
- 17. Wewer Albrechtsen NJ, Bak MJ, Hartmann B, et al Stability of glucagon‐like peptide 1 and glucagon in human plasma. Endocr Connect 2015; 4: 50–57. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Bak MJ, Albrechtsen NW, Pedersen J, et al Specificity and sensitivity of commercially available assays for glucagon and oxyntomodulin measurement in humans. Eur J Endocrinol 2014; 170: 529–538. [DOI] [PubMed] [Google Scholar]
- 19. Bak MJ, Wewer Albrechtsen NJ, Pedersen J, et al Specificity and sensitivity of commercially available assays for glucagon‐like peptide‐1 (GLP‐1): implications for GLP‐1 measurements in clinical studies. Diabetes Obes Metab 2014; 16: 1155–1164. [DOI] [PubMed] [Google Scholar]
- 20. Prince HE, Yeh C. Reactivity of human IgM binding murine monoclonal 6B6C1 (IgG2a) with other murine monoclonal IgG antibodies. J Clin Lab Anal 2013; 27: 27–30. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Nakano K, Yasuda K, Shibuya H, et al Transient human anti‐mouse antibody generated with immune enhancement in a carbohydrate antigen 19‐9 immunoassay after surgical resection of recurrent cancer. Ann Clin Biochem 2016; 53: 511–515. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Figure S1 | Procedure of solid phase extraction. A C18 column was used as the extraction column. Samples were dissolved in the assay buffer for each enzyme‐linked immunosorbent assay kit.
Table S1 | Comparison of relative ratios of active glucagon‐like peptide‐1 values after treatment with extraction carriers.
Table S2 | Extraction rates of iodine‐labeled active glucagon‐like peptide‐1 when using various extract solutions.