Skip to main content
Proceedings of the National Academy of Sciences of the United States of America logoLink to Proceedings of the National Academy of Sciences of the United States of America
. 2019 Feb 19;116(10):4129–4134. doi: 10.1073/pnas.1815329116

Beating the reaction limits of biosensor sensitivity with dynamic tracking of single binding events

Derin Sevenler a,b,1, Jacob Trueb c, M Selim Ünlü a,1
PMCID: PMC6410873  PMID: 30782809

Significance

Low-abundance molecular biomarkers are a promising reservoir of diagnostic information for the early detection and precise treatment of a range of diseases. Yet, most leading assay technologies are currently limited by the affinity and specificity of the molecular-recognition agents they use. This binding “reaction limit” could be beaten by measuring the duration of individual molecular-binding events as they happen in real time. We developed a technology to measure the duration of thousands of individual binding events across a large sensor area. These duration measurements allowed specific analyte binding to be distinguished from nonspecific background binding. The sensor is large enough to accommodate 12 ultrasensitive tests at once.

Keywords: single-molecule detection, biosensor, nanotechnology, nanophotonics, molecular recognition

Abstract

The clinical need for ultrasensitive molecular analysis has motivated the development of several endpoint-assay technologies capable of single-molecule readout. These endpoint assays are now primarily limited by the affinity and specificity of the molecular-recognition agents for the analyte of interest. In contrast, a kinetic assay with single-molecule readout could distinguish between low-abundance, high-affinity (specific analyte) and high-abundance, low-affinity (nonspecific background) binding by measuring the duration of individual binding events at equilibrium. Here, we describe such a kinetic assay, in which individual binding events are detected and monitored during sample incubation. This method uses plasmonic gold nanorods and interferometric reflectance imaging to detect thousands of individual binding events across a multiplex solid-phase sensor with a large area approaching that of leading bead-based endpoint-assay technologies. A dynamic tracking procedure is used to measure the duration of each event. From this, the total rates of binding and debinding as well as the distribution of binding-event durations are determined. We observe a limit of detection of 19 fM for a proof-of-concept synthetic DNA analyte in a 12-plex assay format.


Several of the promises of precision medicine rely on ultrasensitive molecular diagnostic technologies. Liquid biopsies of circulating genomic, transcriptomic, or proteomic biomarkers of cancer promise earlier detection and treatment, as well as improved guidance of targeted therapies in treating minimum residual disease (1, 2). Similarly, sensitive and specific molecular diagnostic tests for infectious pathogens are vital for the identification and management of pre- or asymptomatic individuals (3, 4). Likewise, panel assays of circulating biomarkers could soon improve the accuracy of diagnosis of injuries such as acute liver failure or traumatic brain injury (5, 6).

These pressing clinical needs have motivated the development of a variety of ultrasensitive assay technologies, culminating in technologies capable of single-molecule detection. A unifying characteristic of essentially all of these assay technologies is that they employ molecular-recognition agents (or “capture probes”) such as antibodies, nanobodies, peptides, oligonucleotides, aptamers, or other agents that bind specifically to the molecule of interest. Single-molecule detection technologies commonly then use droplet emulsions (7) or microwell arrays (8) to isolate and then enumerate the precise number of analyte molecules bound to the capture probes.

In terms of signal transduction, it is clear that single-molecule detection is “as good as it gets” (9). However, transducing the amount of captured analyte is only one half of the picture—the analyte must be captured in the first place. Even with single-molecule detection, assay performance is still limited by the affinity of the capture probes. This causes sensitivity and specificity to vary widely between different probe–analyte pairs. For example, it is now relatively routine to quantify some molecular species (e.g., genomic DNA) with single-copy sensitivity and precision (10), while the detection limits of other analytes (e.g., microRNA) are many orders of magnitude worse (1113). Probe affinity can also vary between samples. Variations in extensive properties of the sample such as pH and ion content change the free energy of binding, and variable amounts of nonspecific background binding further complicates quantitation.

Current leading single-molecule detection technologies rely on signal-amplification reactions. These are endpoint assays: the probe molecules are incubated with the sample for a set amount of time, after which the reaction is halted so that amplification and detection can be performed. What is measured is the amount of bound analyte at the instant the incubation is halted.

In contrast to endpoint assays, kinetic assays directly measure probe–analyte interactions during course of the incubation. Kinetic assays collect more information than endpoint assays: they can measure not just concentration but also molecular affinity via rates of association and dissociation. This could allow for intersample variations in probe affinity or nonspecific binding to be identified and mitigated without additional tests. For low concentrations of analyte, kinetic assays could also be capable of distinguishing low-abundance specific binding from a larger background of nonspecific binding, or even measuring analytes below the so-called “critical concentration” at which there is fewer than one analyte molecule bound at equilibrium—a feat impossible for endpoint assays (14).

However, single-molecule kinetic measurements are technically demanding: without amplification reactions, specific binding events are more difficult to discern against a background of nonspecific interactions. Indeed, an exquisitely sensitive transduction mechanism is required to directly detect single binding events at all. A range of scientific apparatuses have been developed to investigate single-molecule binding kinetics. However, none of these techniques are useful for ultrasensitive clinical assays because the sensors are too small (14). To investigate nanoscale phenomena, these devices are themselves nano- or microscale: their active sensors are the size of single nanoparticles or nanowires (1517), or else they require high-magnification and high-numerical aperture optics with a small field of view (0.001–0.01 mm2) (1820). This is problematic because small sensors only have space for a small number of capture probes. Maximizing the number of probes is vital for ultrasensitivity: at low concentrations, the amount of captured analyte at equilibrium is proportional to the number of probe molecules. Single-molecule assay technologies therefore use large sensor areas packed with capture probes. For example, the Quanterix Simoa technology interrogates ∼25,000 beads, each 2.7 µm in diameter, corresponding to a total sensor area of 0.57 mm2 (12).

Here, we describe a kinetic assay that measures the duration of individual binding events over time on a large sensor surface with a low-magnification objective, while retaining the advantages of kinetic analysis such as discrimination between specific and nonspecific events based on duration (i.e., affinity). In this study, we used a 20×, 0.45 NA objective and a 1.1-inch format camera, which yielded a sensor area of 0.38 mm2, comparable with that of ultrasensitive endpoint methods. (This area could be further increased several-fold with different optical instrumentation and stage scanning.)

Results

Detection of Individual Binding Events Across a Large Field of View.

We recently described the development of a “digital microarray” assay technology, which rapidly enumerates individual captured molecules across hundreds of microarray spots (21). This technology uses probe-conjugated gold nanorods as molecular labels and an interferometric reflectance imaging sensor (IRIS) to rapidly detect individual nanorods with a low-magnification (10–20×) objective. The large field of view enabled a similar throughput to commercial fluorescence readers while enhancing the limit of detection (LOD) and dynamic range by a factor of ∼10,000. The assay reaction is based on the bio-barcode assay developed by others (22) and is compatible with a range of different analytes.

We adapted the IRIS digital microarray platform for dynamic measurements by designing a perfusion chamber that consists of an IRIS chip, a patterned silicone gasket, and an antireflection-coated coverglass window (Fig. 1A). Two holes for the chamber inlet and outlet are drilled in the IRIS chips by wafer-scale laser micromachining. The assembly is held by a custom clamp fixture that makes fluidic connections to the inlet and outlet on the bottom of the chip (SI Appendix, Fig. S1).

Fig. 1.

Fig. 1.

Dynamic measurements of single binding events across a large microarray. (A) Rendered image of IRIS chip perfusion chamber for dynamic measurements of molecular interactions. A DNA or antibody microarray is printed on the IRIS chip. Then, the chamber is formed by layering a patterned adhesive gasket and antireflection-coated coverglass viewing window. The IRIS chip has two through-holes for sample perfusion. The entire disposable costs about $5 USD. (B) Nucleic acid assay with IRIS. DNA-conjugated gold nanorods are preincubated with the sample solution and hybridized with complementary nucleic acids. The mixture is flowed over the chip. Complementary nucleic acid strands tether nanorods to the cDNA microarray spot. (C) Schematic of dynamic detection of single nanorods with IRIS. Images are simulated. Nanorods on the chip surface are observed as diffraction-limited spots and automatically detected using purpose-built software. (D) Plots of total nanoparticle binding to six complementary (red) and six noncomplementary (green) DNA spots over time, as measured with dynamic tracking, for one representative experiment where the analyte concentration is 316 fM.

To demonstrate dynamic detection of single molecules, a synthetic ssDNA oligonucleotide was used as a proof-of-concept analyte. The analyte was preincubated with cDNA-conjugated gold nanorods nominally 25 nm × 70 nm for 90 min. The concentration of nanorod labels was kept constant for all experiments at 14 pM, while the concentration of the analyte varied from 10 pM to 10 fM. After preincubation, the mixture was perfused over IRIS chips with DNA microarrays of complementary and noncomplementary probes and nanorod-labeled oligos hybridized to the complementary spots (Fig. 1B).

Images were acquired every 30 s with the IRIS instrument during perfusion. Nanorods on the IRIS chip were visible as faint diffraction-limited blobs in the images, which were detected in each frame independently using custom software (Fig. 1C). The binding rates were then measured determined using a custom dynamic tracking algorithm described in the following section (Fig. 1D and Movie S1).

IRIS detects individual nanorods based on light scattering. Since water has a higher refractive index than air, the polarizability and scattering cross section of the nanorods were reduced compared with dry chips. Additionally, the image suffered from spherical aberrations caused by the air-coverglass interface. Although nanorods were detectable with a 10×, 0.3 NA objective, visibility was much improved with a 20×, 0.45 NA coverglass corrected objective. The resulting field of view of 0.38 mm2 (725 µm × 530 µm) could comfortably accommodate 12 microarray spots, each ∼80 µm in diameter.

Dynamic Tracking of Binding Events over Time.

Under sufficiently high flow rates, the initial rate of binding of analyte is proportional to the bulk analyte concentration. One may estimate the analyte concentration by plotting the number of bound nanorods over time and measuring the initial slope (“naïve counting”; Fig. 2A). However, this approach has several problems. The first is the fundamental issue related to finite probe affinity mentioned earlier. Ultralow analyte concentrations will reach equilibrium with very few (or even fewer than one) bound analyte molecules. In those cases, the initial slope will not be measurable even with perfect error-free readout. The second issue is that some unbound nanorods are visible in each frame as they transiently diffuse through the detection volume. These transient particles result in false positives that increase the overall noise floor of the sensor.

Fig. 2.

Fig. 2.

(A) Dynamic tracking improves sensitivity when the debinding rate is nonzero. Assuming first-order reaction kinetics, equilibrium is reached when the rate of new analyte binding is balanced by the rate of debinding from the surface. Naïvely counting the instantaneous number of bound analyte provides no additional information once equilibrium has been reached. Dynamic tracking distinguishes the binding of new particles from the debinding of old ones. The cumulative number of binding events continues to increase, even at equilibrium. (BD) Diagram of the multistep dynamic tracking algorithm, described in the text.

To address these issues, we developed a postprocessing algorithm that uses the spatial positions of particles to track them individually over the course of the experiment (Fig. 2B). First, particles are detected in each video frame independently of other frames, and their positions in the image are recorded. Second, these positions are compared with those of particles in the next frame of the video (Fig. 2C). Particles in the same position in both frames are “matched,” indicating that they are in fact the same particle. Particle matching includes a clustering algorithm that is robust to small translations between frames. For a video with N time points, this results in N1 lists of matches. Third, these lists are compiled into a single master catalog, which tracks the contiguous series of frames in which each particle was observed (Fig. 2D). This is essentially a table that lists when each particle bound, where it bound, and when (if ever) it debound from the surface. Finally, this catalog is filtered to reduce false positives and false negatives. False negatives occur when a particle is mistakenly not detected in a single frame but was detected in the same position in previous and subsequent frames. This erroneously results in two entries in the catalog. These gaps are repaired by identifying whether the binding of each particle corresponds to the exact same place as the debinding of another particle two frames prior and then merging the two catalog entries (e.g., particles 3 and 7; Fig. 2D). False positives are caused by particles visible in just one frame and are simply removed (e.g., particle 4; Fig. 2D).

This catalog can then be used to plot the cumulative number of new binding events over time. For low-analyte concentrations, most binding sites remain empty at equilibrium, and the rate of new binding events will be constant and proportional to the bulk concentration. At an analyte concentration of 316 fM, for example, the sensor reached equilibrium with about 80 bound nanorods after 1 h (naïve counting; Fig. 3A). However, the cumulative number of new binding events continued to increase linearly even after equilibrium was reached (“total binding”; Fig. 3B). Note that the total number of binding and debinding events are both cumulative measurements and are therefore monotonically increasing over time.

Fig. 3.

Fig. 3.

Experimental evaluation of kinetic IRIS measurements with dynamic tracking. (A) Instantaneous number of nanorods binding to complementary and noncomplementary control spots over time. The analyte concentration is 316 fM, and the nanorod concentration is 14 pM. The sensor reaches equilibrium within 90 min. (B) Total number of nanorod binding and debinding for the complementary spot in A, as measured with dynamic tracking. The rate of total nanorod binding is constant throughout, and equilibrium is reached when the rates of binding and debinding are equal. (C) Histogram of binding-event durations across all experiments, with a biexponential fit. This biexponential distribution is thought to be caused by differences in affinity between side-immobilized and end-immobilized nanorods.

We compared measured binding rates with the predicted rates of transport of analyte-bound nanorods to the spots and found the measured binding rates were well below the theoretical upper limit predicted by mass transport (SI Appendix). We also experimentally characterized the maximum expected binding rates with a series of positive control experiments, in which the nanorods irreversibly bound directly to the microarray spots without any intermediate analyte (SI Appendix). In these control experiments, the rate of debinding was negligible, so the instantaneous and cumulative binding rates were nearly identical and linearly proportional to nanorod concentration, as expected (SI Appendix, Figs. S4–S5). We also evaluated whether replicate spots downstream of one another would experience reduced binding due to sample depletion. We observed no difference in binding rates between upstream and downstream spots at any concentration, supporting the notion that the system was not transport limited (SI Appendix, Fig. S6).

Measurements of Nanorod Dwell Times.

The duration or “dwell time” of each binding event may be measured using dynamic tracking. Note that the dwell time can only be measured for particles that debind before the end of the experiment. Taken together, these dwell times allow the off rate koff of the nanorods to be determined (Fig. 3C). We found that our experimental results were best explained by a biexponential fit of the form N(t)=A1ek1t+A2ek2t. A histogram of dwell times across all complementary spots was generated for each experiment and fitted independently (SI Appendix, Fig. S3). Ignoring experiments that were too brief or contained too few binding events to have meaningful statistics, the fitting parameters consistent across experiments regardless of analyte concentration (SI Appendix, Fig. S3). The average values were k1=0.53min1 and k2=0.082min1 and A1/A225, corresponding with primary and secondary dissociation time constants τ1=1.9 min and τ2=12 min. At first, we hypothesized this heterogeneity consisted of nanorods tethered by either one versus two or more analyte molecules. However, the relative weights between the two terms A1/A2 was similar across a large range of concentrations, and the ratio did not tend to decrease with lower analyte concentrations. Since the total nanorod concentration was kept constant at 14 pM, the relative number of nanorods with two bound analytes versus one bound analyte would have decreased with decreasing analyte concentration.

Instead, we hypothesize that the biexponential distribution in dwell times was caused by the asymmetry of the nanorods themselves as they bind to the surface. The binding energy is likely greater if the rod is tethered to the surface by one end, rather than by the middle. First, there is electrostatic repulsion between the DNA-functionalized nanorods and the DNA-coated chip surface. A side-tethered nanorod is constrained in a manner that brings the centroid closer to the chip and brings a larger surface area adjacent to the chip. Second, the end-tethered nanorod has a larger number of conformational degrees of freedom (DOF) (three rotational DOF) than a side-tethered one (one rotational DOF), resulting in a lower entropic penalty to binding. Since the rods are functionalized uniformly across their surfaces, nanorods are more likely to capture analyte to their sides rather than ends during preincubation and therefore be side-tethered. This is supported by the observation that the faster dissociation rate was also more prevalent across all concentrations.

It may be important to note that we cannot differentiate between dissociation of the analyte from the surface probe and from the nanorod label. In the model system used here, both surface–analyte and analyte–label duplexes are 25 bp long and have similar GC content. Therefore, their affinities should be similar, and they should be responsible for nanorod dissociation at roughly similar rates.

Detection Below the Critical Concentration.

A standard curve was measured by performing identical experiments with a range of analyte concentrations between 10 fM and 10 pM. Nanorod binding to three replicate complementary spots were independently analyzed using both dynamic tracking and naïve counting (Fig. 4). A modified first-order reaction model was used to capture both the linear behavior at low-analyte concentrations and the saturation behavior at high concentrations, where most nanorods had captured at least one analyte (SI Appendix). The LOD was calculated as three SDs above the mean signal from the blank sample. As expected, the LOD for dynamic tracking (LOD = 19 fM) far surpassed that of naïve counting (LOD = 687 fM), by about 36-fold.

Fig. 4.

Fig. 4.

Rate of particle binding versus analyte concentration for the same experiments analyzed using dynamic tracking (A) or naïve counting (B). A modified first-order kinetic model was fit to the data and used to determine the LOD and critical concentration (c*). Insets show the same data and model on linear axes. Dynamic tracking improved the LOD by 36-fold compared with naïve counting. Notably, the dynamic tracking LOD (19 fM) is 3.6-fold lower than the critical concentration (68 fM), at which just one molecule is bound to the sensor on average.

Notably, dynamic tracking had an LOD 3.6 times lower than the critical concentration of this assay (c*; Fig. 4A). Equivalently, dynamic tracking was able to detect the presence of the analyte even when the average duration of binding events was shorter than the time between them. At the critical concentration, equilibrium is reached with one bound analyte molecule on average. This dissociation rate is simply the weighted average of the two dissociation constants found earlier: k¯off=(A1k1+A2k2)/A1A2=0.52min1=31 particles per hour. The critical concentration was determined by taking intercept of the dynamic tracking regression line with this binding rate: c= 55 fM.

We observed a saturation in binding rate at concentrations above 2–3 pM (Fig. 4A). We hypothesized this could be the result of using a lower concentration of nanorods than expected. We performed spectrophotometry to estimate the concentration of the conjugated nanorods as well as the nanorod stock solution from the manufacturer. We compared the absorbances to that predicted by numerical simulations and estimated that the nanorod concentration was indeed several times lower than we expected (SI Appendix, Fig. S7).

Discussion

A range of endpoint-assay technologies have been developed that have single-molecule readout. For these assays, the limiting factor is the affinity and specificity of the molecular-recognition agents rather than readout sensitivity. The number of capture probes (for example, the number of functionalized beads) can almost always be increased until sensitivity is limited by nonspecific binding rather than insufficient numbers of probes (23), but further improvement must come through careful optimization of washes and reactions. Protocol optimization are particularly challenging for multiplexed test development, since the optimal wash conditions (duration, ion content, surfactants, pH, and so on) are often different for different probe–analyte complexes.

In response to these limitations, we have introduced a kinetic-assay technology that measures the duration of individual binding events across a large sensor area. In this work, we distinguished specific binding events from the background without even a single wash step, which could be used to further improve specificity. Notably, kinetic analysis alleviates the need for a globally optimal wash protocol and therefore makes multiplexed tests straightforward.

“Solid-phase” surface sensors are sometimes criticized for having poor mass-transport kinetics, compared with bead-based assays. We alleviated this effect by using a high flow rate, which makes the depletion layer very thin (SI Appendix). For longer experiments or smaller sample volumes, peristaltic pumping and recirculation may be used.

The unexpected biexponential histogram of dwell times suggested two different conformations of immobilized nanorods, each with different binding free energy: end-tethered and side-tethered. This variability complicates the probe–analyte affinity measurement and could be problematic for heterogeneous samples. This could be further tested and perhaps mitigated by preferentially functionalizing just the ends of nanorods (24, 25).

Because these proof-of-concept experiments were conducted with synthetic analytes and pure buffer solutions, direct comparisons with more mature assay technologies cannot yet be made. However, future applications of this technique are envisioned in response to current clinical needs for multiplexed quantification of microRNAs for cancer diagnostics, mRNAs for phenotypic identification of drug susceptibility, and circulating protein biomarkers of viral infection.

Materials and Methods

Additional materials and methods are available in SI Appendix, Supplemental Materials and Methods.

Perfusion Chamber Assembly.

No. 1 coverslips, 25.4 mm × 12.7 mm, with a broadband antireflection coating on one side were purchased from Abrisa Technologies. Custom patterned silicone gaskets were purchased from Grace Biolabs. Silicone gaskets were 25.4 mm × 12.7 mm, 0.15-mm thickness, with pressure-sensitive adhesive on one side. In preparation, gaskets were adhered to the noncoated side of the coverglass and stored with protective tape in place.

The perfusion chamber was assembled by aligning the gasket-window assembly to the IRIS chip, loading it into the clamp fixture, removing the protective tape, and engaging the clamp to form seals between the chip and the gasket as well as with the sample inlet and outlet. The volume of the chamber was ∼8 µL.

IRIS Digital Microarray Instrument for Dynamic Detection.

The operating principle of IRIS is thoroughly described elsewhere (26). Briefly, the IRIS instrument consists of a reflectance microscope with a single high-powered LED for illumination (M660L4 LED with FB650-10 bandpass filter; Thorlabs) and a monochrome machine vision camera (Grasshopper GS3-U3-123S6M-C; Point Gray Research). The digital microarray implementation of IRIS is optimized for rapidly detecting individual gold nanorods based on their anisotropic light-scattering properties. The design, optimization, and implementation of the optical system has been described in detail elsewhere (21). For dry IRIS chips, this system can detect single gold nanorods with a 10×, 0.3 NA objective. For dynamic experiments the system was entirely the same except that a 20×, 0.45 NA coverslip-corrected air immersion objective (Nikon CFI S Plan Fluor ELWD 20×) was used. The higher light-collection efficiency compensated for the decreased intensity of nanorod light scattering due to immersion of the rods in water, and the collar allowed correction of spherical aberrations from the coverslip–air interface.

Assay Protocol.

The assay protocol was identical for all experiments, except that the concentration of the analyte was changed. First, the DNA–nanorod conjugates and synthetic DNA oligos were premixed in a “hybridization buffer” consisting of 10 mM phosphate (pH 7.4), 600 mM Na+, 0.1% Tween-20, and 1 mM EDTA; 100 µL of nanorods stored at 140 pM were mixed with 900 µL of hybridization buffer containing the analyte ssDNA. The final nanorod concentration was 14 pM for all experiments. The mixture was vortexed briefly and sonicated for 10 s before storing at room temperature in a microcentrifuge tube. After 90 min, the sample was aspirated with a disposable 1-mL needle-tipped syringe. The needle tip was removed, and the syringe was connected to a Luer fitting on the end of the inlet tube. The outlet waste tube was left in conical vial. The syringe was mounted in a syringe pump, and the sample was dispensed at 10 µL/min for up to 90 min. The instrument was refocused as soon as the liquid sample filled the chamber. Video acquisition began 1–3 min after the sample first contacted the chip surface.

Image Acquisition.

Image acquisition was automated using the Micromanager (27) microscope control software with custom scripts. Scripts have been made freely available online at https://www.github.com/derinsevenler/IRIS-API. Time points were taken every 30 s at each time point, and a z-stack of nine images was acquired with a step size of 2 μm (i.e., a span of 16 μm). At each z-position, four images were acquired and averaged pixelwise before saving to reduce shot noise.

Image Processing and Particle Detection.

The video data from each experiment consisted of an image hyperstack of 180 (t) × 9 (z) × 12.4 MP (x,y). A 86 µm × 86 µm (500 × 500 pixels) region of the video was cropped around each microarray spot. Nanorods in each region and timepoint were detected independently. The particle detection method described here is a refinement of methods described earlier (21, 28), and has three steps: preprocessing, key point detection, and key point filtering. First, a sparse pseudomedian filter is applied to each frame of the z-stack (made available online by others at imagejdocu.tudor.lu/doku.php?id=plugin:filter:fast_filters:start) to estimate the image background. True median filtering is effective for removing punctate features but computationally expensive for larger kernels. We found the sparse pseudomedian algorithm preferable due to its speed. Next, the normalized intensity image was calculated by pixelwise division of the original frame from the background. Finally, the normalized intensity range (NIR) image was measured by projecting the maximum difference (i.e., maximum − minimum) at each pixel of the normalized intensity stack. Although not every nanorod is visible in every normalized intensity image in the stack, each particle is clearly visible in the resulting NIR image.

Key points in the NIR images were detected by applying a global threshold to binarize the image. Blobs in the binary image (i.e., regions brighter than the threshold) were enumerated and then filtered based on size and shape. Specifically, a minimum area, maximum area, and minimum area-to-perimeter ratio were specified. The detection threshold and key point-filtering parameters were manually selected and then kept constant for all experiments.

Supplementary Material

Supplementary File
Supplementary File
Download video file (507.8KB, avi)

Acknowledgments

The authors are grateful to Fulya Ekiz Kanik, Joeseph Greene, and Abdul Bhuiya for feedback on the image acquisition and analysis software and perfusion chamber design; to Celalettin Yurdakul for help designing the chip photolithography pattern; to Matthew Geib for help preparing IRIS chips and spotting; and to Ahmad Khalil for fruitful discussions about these results.

Footnotes

The authors declare no conflict of interest.

This article is a PNAS Direct Submission. D.R.W. is a guest editor invited by the Editorial Board.

This article contains supporting information online at www.pnas.org/lookup/suppl/doi:10.1073/pnas.1815329116/-/DCSupplemental.

References

  • 1.Cohen JD, et al. Combined circulating tumor DNA and protein biomarker-based liquid biopsy for the earlier detection of pancreatic cancers. Proc Natl Acad Sci USA. 2017;114:10202–10207. doi: 10.1073/pnas.1704961114. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Alix-Panabières C, Pantel K. Clinical applications of circulating tumor cells and circulating tumor DNA as liquid biopsy. Cancer Discov. 2016;6:479–491. doi: 10.1158/2159-8290.CD-15-1483. [DOI] [PubMed] [Google Scholar]
  • 3.Shurtleff AC, Whitehouse CA, Ward MD, Cazares LH, Bavari S. Pre-symptomatic diagnosis and treatment of filovirus diseases. Front Microbiol. 2015;6:108. doi: 10.3389/fmicb.2015.00108. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Alsdurf H, Hill PC, Matteelli A, Getahun H, Menzies D. The cascade of care in diagnosis and treatment of latent tuberculosis infection: A systematic review and meta-analysis. Lancet Infect Dis. 2016;16:1269–1278. doi: 10.1016/S1473-3099(16)30216-X. [DOI] [PubMed] [Google Scholar]
  • 5.Antoine DJ, et al. Mechanistic biomarkers provide early and sensitive detection of acetaminophen-induced acute liver injury at first presentation to hospital. Hepatology. 2013;58:777–787. doi: 10.1002/hep.26294. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Papa L, et al. Time course and diagnostic accuracy of glial and neuronal blood biomarkers GFAP and UCH-L1 in a large cohort of trauma patients with and without mild traumatic brain injury. JAMA Neurol. 2016;73:551–560. doi: 10.1001/jamaneurol.2016.0039. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Vogelstein B, Kinzler KW. Digital PCR. Proc Natl Acad Sci USA. 1999;96:9236–9241. doi: 10.1073/pnas.96.16.9236. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Cohen L, Walt DR. Single-molecule arrays for protein and nucleic acid analysis. Annu Rev Anal Chem (Palo Alto Calif) 2017;10:345–363. doi: 10.1146/annurev-anchem-061516-045340. [DOI] [PubMed] [Google Scholar]
  • 9.Walt DR. Optical methods for single molecule detection and analysis. Anal Chem. 2013;85:1258–1263. doi: 10.1021/ac3027178. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Schoepp NG, et al. Rapid pathogen-specific phenotypic antibiotic susceptibility testing using digital LAMP quantification in clinical samples. Sci Transl Med. 2017;9:eaal3693. doi: 10.1126/scitranslmed.aal3693. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Rissin DM, et al. Polymerase-free measurement of microRNA-122 with single base specificity using single molecule arrays: Detection of drug-induced liver injury. PLoS One. 2017;12:e0179669. doi: 10.1371/journal.pone.0179669. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Cohen L, Hartman MR, Amardey-Wellington A, Walt DR. Digital direct detection of microRNAs using single molecule arrays. Nucleic Acids Res. 2017;45:e137. doi: 10.1093/nar/gkx542. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Wu D, Katilius E, Olivas E, Dumont Milutinovic M, Walt DR. Incorporation of slow off-rate modified aptamers reagents in single molecule array assays for cytokine detection with ultrahigh sensitivity. Anal Chem. 2016;88:8385–8389. doi: 10.1021/acs.analchem.6b02451. [DOI] [PubMed] [Google Scholar]
  • 14.Squires TM, Messinger RJ, Manalis SR. Making it stick: Convection, reaction and diffusion in surface-based biosensors. Nat Biotechnol. 2008;26:417–426. doi: 10.1038/nbt1388. [DOI] [PubMed] [Google Scholar]
  • 15.Baaske MD, Foreman MR, Vollmer F. Single-molecule nucleic acid interactions monitored on a label-free microcavity biosensor platform. Nat Nanotechnol. 2014;9:933–939. doi: 10.1038/nnano.2014.180. [DOI] [PubMed] [Google Scholar]
  • 16.Besteman K, Lee J-O, Wiertz FGM, Heering HA, Dekker C. Enzyme-coated carbon nanotubes as single-molecule biosensors. Nano Lett. 2003;3:727–730. [Google Scholar]
  • 17.Sorgenfrei S, et al. Label-free single-molecule detection of DNA-hybridization kinetics with a carbon nanotube field-effect transistor. Nat Nanotechnol. 2011;6:126–132. doi: 10.1038/nnano.2010.275. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Halpern AR, Wood JB, Wang Y, Corn RM. Single-nanoparticle near-infrared surface plasmon resonance microscopy for real-time measurements of DNA hybridization adsorption. ACS Nano. 2014;8:1022–1030. doi: 10.1021/nn405868e. [DOI] [PubMed] [Google Scholar]
  • 19.Piliarik M, Sandoghdar V. Direct optical sensing of single unlabelled proteins and super-resolution imaging of their binding sites. Nat Commun. 2014;5:4495. doi: 10.1038/ncomms5495. [DOI] [PubMed] [Google Scholar]
  • 20.Lamichhane R, Solem A, Black W, Rueda D. Single-molecule FRET of protein-nucleic acid and protein-protein complexes: Surface passivation and immobilization. Methods. 2010;52:192–200. doi: 10.1016/j.ymeth.2010.06.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Sevenler D, Daaboul GG, Ekiz Kanik F, Ünlü NL, Ünlü MS. Digital microarrays: Single-molecule readout with interferometric detection of plasmonic nanorod labels. ACS Nano. 2018;12:5880–5887. doi: 10.1021/acsnano.8b02036. [DOI] [PubMed] [Google Scholar]
  • 22.Hill HD, Mirkin CA. The bio-barcode assay for the detection of protein and nucleic acid targets using DTT-induced ligand exchange. Nat Protoc. 2006;1:324–336. doi: 10.1038/nprot.2006.51. [DOI] [PubMed] [Google Scholar]
  • 23.Yelleswarapu VR, Jeong H-H, Yadavali S, Issadore D. Ultra-high throughput detection (1 million droplets per second) of fluorescent droplets using a cell phone camera and time domain encoded optofluidics. Lab Chip. 2017;17:1083–1094. doi: 10.1039/c6lc01489e. [DOI] [PubMed] [Google Scholar]
  • 24.Caswell KK, Wilson JN, Bunz UHF, Murphy CJ. Preferential end-to-end assembly of gold nanorods by biotin-streptavidin connectors. J Am Chem Soc. 2003;125:13914–13915. doi: 10.1021/ja037969i. [DOI] [PubMed] [Google Scholar]
  • 25.Pramod P, Joseph STS, Thomas KG. Preferential end functionalization of Au nanorods through electrostatic interactions. J Am Chem Soc. 2007;129:6712–6713. doi: 10.1021/ja071536o. [DOI] [PubMed] [Google Scholar]
  • 26.Sevenler D, Avci O, Ünlü MS. Quantitative interferometric reflectance imaging for the detection and measurement of biological nanoparticles. Biomed Opt Express. 2017;8:2976–2989. doi: 10.1364/BOE.8.002976. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Edelstein AD, et al. Advanced methods of microscope control using μManager software. J Biol Methods. 2014;1:e10. doi: 10.14440/jbm.2014.36. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Trueb J, Avci O, Sevenler D, Connor JH, Ünlü MS. Robust visualization and discrimination of nanoparticles by interferometric imaging. IEEE J Sel Top Quantum Electron. 2017;23:1–10. doi: 10.1109/JSTQE.2016.2639824. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary File
Supplementary File
Download video file (507.8KB, avi)

Articles from Proceedings of the National Academy of Sciences of the United States of America are provided here courtesy of National Academy of Sciences

RESOURCES