Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2020 Mar 1.
Published in final edited form as: Burns. 2018 Oct 14;45(2):450–460. doi: 10.1016/j.burns.2018.09.026

Evaluating clinical observation versus Spatial Frequency Domain Imaging (SFDI), Laser Speckle Imaging (LSI) and thermal imaging for the assessment of burn depth

Adrien Ponticorvo 1, Rebecca Rowland 1, Melissa Baldado 1, David M Burmeister 2, Robert J Christy 2, Nicole P Bernal 3, Anthony J Durkin 1,4,*
PMCID: PMC6420831  NIHMSID: NIHMS1509685  PMID: 30327232

Abstract

While clinical examination is needed for burn severity diagnosis, several emerging technologies aim to quantify this process for added objectivity. Accurate assessments become easier after burn progression, but earlier assessments of partial thickness burn depth could lead to earlier excision and grafting and subsequent improved healing times, reduced rates of scarring/infection, and shorter hospital stays. Spatial Frequency Domain Imaging (SFDI), Laser Speckle Imaging (LSI) and thermal imaging are three non-invasive imaging modalities that have some diagnostic ability for noninvasive assessment of burn severity, but have not been compared in a controlled experiment. Here we tested the ability of these imaging techniques to assess the severity of histologically confirmed graded burns in a swine model. Controlled, graded burn wounds, 3 cm in diameter were created on the dorsum of Yorkshire pigs (n=3, 45–55kg) using a custom-made burn tool that ensures consistent pressure has been employed by various burn research groups. For each pig, a total of 16 burn wounds were created on the dorsal side. Biopsies were taken for histological analysis to verify the severity of the burn. Clinical analysis, SFDI, LSI and thermal imaging were performed at 24 and 72 hours after burn to assess the accuracy of each imaging technique. In terms of diagnostic accuracy, using histology as a reference, SFDI (85%) and clinical analysis (83%) performed significantly better that LSI (75%) and thermography (73%) 24 hours after the burn. There was no statistically significant improvement from 24 to 72 hours across the different imaging modalities. These data indicate that these imaging modalities, and specifically SFDI, can be added to the burn clinicians’ toolbox to aid in early assessment of burn severity.

1. Introduction

There are as many as 11 million cases worldwide annually of people requiring medical attention due to burn injuries[1]. Burn wounds and the subsequent scarring that can result from these injuries are damaging both cosmetically and functionally, necessitating the search for better and more efficient treatments. One immediate way to both improve treatment/outcomes and reduce overall costs related to patient care is to make accurate diagnoses earlier. Delays decision making related to excision and grafting enables burn wound progression[2] and increases length of hospital stay. This gap in health care has led to the development of numerous imaging technologies that have potential for assisting physicians in rapidly and correctly diagnosing burn severity.

Laser Doppler based techniques have been used since the 1980s to assess perfusion in burns alongside clinical evaluation[3]. Laser speckle imaging (LSI)[4, 5], also known as Laser Speckle Contrast Analysis (LASCA)[6], utilizes similar principals to generate perfusion images over a large field of view with fast acquisition times making it ideally suited to study perfusion in burns. Measurements that indicate low blood perfusion may imply the destruction of the underlying vasculature and indicate a need for surgical intervention, while high perfusion measurements suggest that the vascular network is intact and there is a strong chance of spontaneous healing. While several studies have shown that LSI is more accurate than clinical analysis for burn assessment, this accuracy is a highly time dependent phenomenon. It is still common to wait 48–72 hours before excision to ensure accurate clinical observation of burn depth[3]. Other studies have employed laser Doppler measurements earlier than 48 hours with inconclusive results[7].

Thermal imaging has recently been adapted for burn wound diagnostics[4]. In relation to burn wound severity, it is typically used to identify regions of tissue within a burn that demonstrate reduced surface temperature relative to some baseline normal skin surface temperature. The reduction in temperature suggests that in these areas, the underlying blood vessels have been damaged to the point where the lack of perfusion leads to a reduction in thermal emission and this can be empirically related to burn depth[810]. Recent studies suggest that thermography is only as accurate as photographic clinical assessments, and is highly dependent on the ambient temperature during measurements[4]. Other studies have found that LSI outperforms thermal imaging in a clinical setting[8]. However, one difficulty in assessing these imaging techniques in a clinical environment is that “ground truth” is not known.

Multispectral and hyperspectral imaging have also been investigated within the context of assessment of burn wound severity. These techniques utilize images taken at multiple wavelengths of light to measure the reflectance from burnt issue[11, 12]. hese reflectance measurements can be converted into concentrations of chromophores such oxygenated and deoxygenated hemoglobin, but require assumptions about the reduced scattering coefficient[13, 14]. Differences in oxygenated and deoxygenated hemoglobin concentrations of burn wound regions and unburned skin indicate damage to the underlying vasculature which can be indicative of burn severity.

Spatial Frequency Domain Imaging (SFDI) is a developing imaging technique that can be used to determine the optical properties and chromophore concentrations of tissue[1518]. It is sensitive to both hemodynamic[19, 20] and structural[2123] components in tissue. Our previous research suggests that SFDI can be used to noninvasively characterize changes in tissue structure and function over a range of burn grades in rats and pigs[2123]. In the porcine burn study, we found that changes in tissue scattering properties are acutely related to structural damage to collagen (i.e., the zone of coagulation). While our previous SFDI imaging of burn severity in a porcine model used a device based on 3 near infrared LED wave lengths[21], here we employ a new SFDI device based on 9 wavelengths over the range 471nm – 971nm. Specifically, we compare the histologically validated burn severity prediction accuracy of SFDI, LSI and thermal imaging, to clinical impression. We examined the performance of these techniques within the context of a swine model of graded burn wounds.

2. Methodology

2.1. Porcine Model

All experiments were performed in accordance with the UC Irvine Institutional Animal Care and Use Committee (IACUC protocol #2015–3154). Three Yorkshire pigs (50 kg average weight) were used in this experiment. Animals were allowed to acclimate to their individual housing for 7 days prior to the experiment. An intramuscular injection of tiletamine-zolazepam (Telazol, 6 mg/kg) was used to sedate the animals. Hair was removed from the area of interest with clippers. Animals were intubated and placed on an automatic ventilator with an initial tidal volume at 10 ml/kg, peak pressure at 20 cmH2O and respiration rate at 10 breaths per minute. The ventilator setting was adjusted to maintain an end tidal PCO2 of 38–42 mmHg, and anesthesia was maintained at 1–3% isofluorane. Vital signs were continuously monitored and kept constant by adjusting anesthesia levels. A heating blanket was used to maintain a constant body temperature (36–38°C). Controlled, graded burn wounds were created using a custom-made setup described previously[21]. Briefly, custom made cylindrical brass probes (3 cm diameter) with stainless steel posts were custom fabricated along with a spring loaded device to hold the probes and ensure consistent pressure. The brass probes were heated in a dry bath incubator to 100°C and held in place by a custom made aluminum block. Sixteen burns were created along the dorsal side approximately 1 cm from the spine and approximately 3 cm away from each other. 8 wounds were created on each side of the spine, with contact times of 5, 10, 15, 20, 25, 30, 35, and 40 seconds. Non-adherent gauze (Telfa, Tyco Healthcare, Mansfield, MA) was placed on top of the wound and held in place with Ioban™ (3M, St. Paul, MN). Bandaging materials were removed at 24 hours and 72 hours after the initial burn in order to image the wounds. Pigs were survived for 28 days in order to evaluate burn healing and relate this back to the original assessment of the burn. A diagram of the animal setup is shown in Fig 1a.

Fig. 1.

Fig. 1.

a) Isotemp™ dry bath incubator (Thermo Fisher Scientific Inc., Pittsburgh, PA) and custom spring loaded burn device. b) A description of the burn placement and typical burn size and biopsy location.

2.2. Histological Assessment

Histological assessments were performed during the experiment to determine the true burn severity. At 24 hours after the initial burn injury, a 6mm punch biopsy was used to collect tissue from within each burn region near the edge of the burn (Fig 1b). This biopsy was performed after all of the appropriate imaging was completed. Great care was taken to ensure that biopsy only contained tissue from the burn region. We have considerable practice with this approach in previous work[21, 22]. In addition, the biopsy region was selected in the most homogenous appearing area near the edge. Biopsies were fixed in 10% neutral buffered formalin for 48 hours, dehydrated in alcohol, and embedded in paraffin wax. The biopsies were sectioned into 7μm thick samples and stained with Masson’s Trichrome (ab150686, Abcam, Cambridge, MA), marking healthy collagen blue, and damaged collagen, epidermis, and muscle tissue hair follicle lumens red. To determine burn depth and categorize the burn, the mounted tissue sections were examined for collagen coagulation and hair follicle damage[24, 25]. Burns were either categorized as requiring or not requiring surgical excision and grafting based on the depth of coagulation and the damage to hair follicles and vessels. When damage in these areas did not exceed the thickness of the epidermis or the damage was restricted to the papillary dermis, the burn was classified as one that would heal without surgical intervention. Damage that extended through to the reticular dermis indicated a burn that would typically require surgical intervention, although no intervention was performed. The percent of dermal damage was measured by comparing the depth of thermal damage to the overall depth of the dermis. The results of this histological analysis were used as the gold standard assessment of burn depth.

2.3. Clinical Assessment

A board certified (surgery and critical care) burn surgeon with 15 years of experience assessed the severity of each burn at 24 and 72 hours after the initial injury in order to similarly categorize whether the burns would require a graft. Importantly, no surgical intervention (i.e., excision/grafting) was performed on these wounds in order to allow for inspection at subsequent time points. Time points were chosen to represent a time frame that spans an ideal and realistic period for excision and grafting[3]. The surgeon was instructed to utilize any visual or tactile clues that would normally be used to assess burn severity in a clinical patient. The area of the burn that would later be biopsied was also the focus of the clinical assessment.

2.4. Laser Speckle Imaging (LSI)

LSI data was collected using a PeriCam SI system (Perimed AB, Sweden) at the 24 and 72 hour time points after the initial injury. The device was placed at a height of 25 cm above the burn wound creating a field of view of 15 × 15 cm. Images were collected at a rate of 5 images/second for 60 seconds and then averaged. All image collection, region of interest (ROI) selection and analysis was done using the Perimed PimSoft v2 software that reported data in Perfusion Units (PU)[6].

2.5. Thermal Imaging

Thermal images were collected with a FLIR A300 (FLIR Systems, Inc., Wilsonville, OR; 7.5–13 µm) at the 24 and 72 hour time points after the initial injury. The device was placed at a height of 40 cm above the burn wound creating a field of view of 17 × 22 cm. All image collection, ROI selection and analysis was done using FLIR Tools v5.13.17214.2001, which reported temperature values with a resolution of 0.1 °C and an assumed emissivity of 0.98, previously validated in skin tissue [26].

2.6. Spatial Frequency Domain Imaging (SFDI)

The Reflect RS™ (Modulated Imaging, Inc., Irvine, CA), a commercially available SFDI measurement system capable of imaging optical properties (absorption coefficient, µa and reduced scattering coefficient, µ’s)[16] of tissue over large fields of view (20 × 15 cm) was employed. The device was placed at a height of 32 cm above the burn wound, and centered so each image captured two neighboring burns at a time. Imaging was performed through the MI Acquire v1.34.00 software. Each region of interest was imaged 3 consecutive times requiring a total of approximately 90 seconds. Sinusoidal patterns are projected and images were captured at 8 wavelengths between 471 nm – 850 nm and at 5 spatial frequencies evenly spaced between 0 mm−1 and 0.2 mm−1 according to a protocol that we have previously employed[21]. In addition planar (0 spatial frequency) reflectance images were acquired at each of the 8 spatially modulated wavelengths and at 971 nm.

After each day of imaging, a reference calibration phantom having known optical properties was measured using the SFDI device under the same lighting conditions as the animal measurements, as has previously described in detail[15]. This phantom is made of a silicon based polymer, mixed with India ink and titanium dioxide, to mimic tissue optical properties[27]. The phantom and files of its optical properties are provided with the Reflect RS™ SFDI instrument. After all measurements were complete, the raw image files were processed in the MI-Analyze interface, part of the MI Acquire software package. Through this processing, the reflectance images were calibrated against the phantom measurements, and converted to images of the absorption and reduced scattering coefficient (μs’) at all measured wavelengths. A schematic of the imaging setup for all 3 instruments, a visualization of their projected illumination methods, and their output data is depicted in Fig. 2.

Fig. 2.

Fig. 2.

A schematic of the three imaging modalities used in the study. From left to right, SFDI, thermal imaging, and LSI.

2.7. Color Photography

Color images were acquired at a fixed distance and angle in relationship to the burn in order to ensure reproducibility. The color camera used here was a 14-megapixel digital camera (NEX-3, Sony Corporation of America, New York, NY) and was employed at the conclusion of each imaging session in order to document the clinical appearance of burn wounds at each time point.

2.8. Statistical Analysis

For each imaging modality, a region of interest (ROI) approximately 6 × 6 mm in size within the burn wound and encompassing the entire biopsied region was selected so as to match the histological analysis of the biopsied region and the clinical assessment. Data sets from the different imaging devices were exported to Microsoft Excel (Microsoft Corporation, Redmond, WA) before statistical analysis was completed using R 3.4.3 (The R Foundation). A receiver operator curve (ROC) was generated for each imaging modality at the 24 and 72 hour time points, and the area under the curve (AUC) was calculated to determine the ability of each technique to correctly classify burns in a binary manner as either severe enough to require a graft or not severe enough that a graft would be necessary. A p-value < 0.05 testing the null hypothesis that the AUC = 0.5 was considered significant. In order to compare results between different imaging modalities, McNemar’s test was used with a p-value < 0.05 considered significant. This statistical test determines if there are differences in paired binary data collected from matched subjects. In this case the binary data was the imaging modality determining if a burn wound would need a graft. The main outcome measure of the experiment was to compare the results of the different imaging modalities graded against the histological assessment, which was considered the “ground truth”. The secondary outcome was to compare each of the different modalities at the 24 and 72 hour time points. The Youden index is a technique used to objectively determine the optimal threshold from an ROC curve[28]. This was used on each imaging modality to determine a threshold that would predict whether a particular imaging modality categorized the wound as needing a graft or not needing a graft. Each imaging modality then determined the accuracy, sensitivity, specificity, positive predictive values, negative predictive values at the 24 and 72 hour time points. Sensitivity was determined to be an imaging modality’s ability to correctly determine if a burn wound required a graft when the histological analysis established that the burn wound required a graft. Specificity was determined to be an imaging modality’s ability to correctly determine if a burn wound did not require a graft when the histological analysis established that the burn wound did not required a graft.

3. Results

3.1. Histological Assessment

Burn severity was determined by assessing each stained histologic sample at 24 hours. Burns were either categorized as requiring or not requiring surgical excision and grafting based on the depth of collagen coagulation and the damage to hair follicles, as seen in Fig. 3. When damage in these areas did not exceed the thickness of the epidermis or the damage was restricted to the papillary dermis, the burn was classified as one that would heal without surgical intervention. Damage that extended through to the reticular dermis or affected more than 50% of the reticular dermis indicated a burn that would typically require surgical intervention, although no intervention was performed. Using these criteria, 19 of the 48 burns included in the study would not require surgical intervention, while 29 of the 48 burns were severe enough to require surgical intervention. Table 1 breaks down the number of burns that were severe enough to require surgical intervention as a function of burn contact time and the average percent of dermal damage for each set of contact times.

Fig. 3.

Fig. 3.

Biopsies of 5, 15, 25, and 40 second burns were taken at 24 hours and stained with Masson’s Trichrome. The depth of denatured collagen is indicated by the white dotted line. Intact hair follicles are designated by black arrows, and damaged hair follicles are marked by orange arrows.

Table 1.

Burns of various contact times listed with the percentage that would require grafts based on histological analysis as well as the average percentage of dermal damage.

Burn ontact Time
(Seconds)
Histological Assessment
(Graft)
Average Percent of Dermal
Damage
5 – 10 0/12 (0%) 14.7%
15 – 20 5/12 (42%) 27.0%
25 – 30 7/12 (58%) 36.8%
35 – 40 12/12 (100%) 53.0%
Total 29/48 (60%)

3.2. Clinical Assessment

After the imaging sessions at 24 hours and 72 hours, a burn surgeon performed a clinical assessment of each wound to classify whether or not surgical intervention would be prescribed. This was compared to the histological assessment to determine diagnostic accuracy. The surgeon was blinded by the randomization of burn contact times for each burn. Table 2 shows how the overall accuracy of the clinical assessment at 24 and 72 hours, as well as a breakdown of the accuracy for different groups of burn contact times. The 24 hour clinical assessments had an overall accuracy of 83% (correct classification of 40/48 burn wounds) with a sensitivity of 100% and a specificity of 58%. he positive and negative predictive values were 78% and 100% respectively. The 72 hour clinical assessments had an overall accuracy of 85% (correct classification of 41/48 burn wounds) with a sensitivity of 100% and a specificity of 63%. The positive and negative predictive values were 81% and 100% respectively. There was no statistically significant difference between the clinical assessments at 24 hours vs. 72 hours. The majority of the incorrect clinical assessments were made on burns with contact times of 15 – 20 seconds.

Table 2.

Correct clinical assessment of burns based on histological results.

Burn Contact Time
(Seconds)
24 hour Assessment
(Correct)
72 hour Assessment
(Correct)
5 – 10 10 / 12 11 / 12
15 – 20 7 / 12 7 / 12
25 – 30 11 / 12 11 / 12
35 – 40 12 / 12 12 / 12
Total 40 / 48 41 / 48

3.3. Laser Speckle Imaging

The averaged perfusion values at 24 and 72 hours, over the appropriate ROI, were used to create the ROC shown in Fig. 4a. The area under the curve (AUC) of the 24 hour perfusion data was 0.79, while the AUC of the 72 hour perfusion data was 0.85. Both values were statistically significant (p<0.001). After choosing a threshold of 39.7 PU, the overall accuracy of the 24 hour perfusion values were 75% (correct classification of 36/48) with a sensitivity of 66% and a specificity of 90%. The positive and negative predictive values were 91% and 63% respectively. The overall accuracy of the 72 hour perfusion values was 83% (correct classification of 40/48) with a sensitivity of 100% and a specificity of 58%. The positive and negative predictive values were 78% and 100% respectively. There was no statistically significant difference between the classifying ability of the perfusion values at 24 hours vs 72 hours. Fig. 4b shows typical perfusion images from 24 and 72 hour time points. Perfusion within the burn region is lower than perfusion for unburned tissue in all burn severities at 24 hours. While at 72 hours, perfusion in the periphery of the 5 second burn is higher than unburned tissue, perfusion in the 20 sec, 25 sec, and 35 sec burns is still lower than unburned tissue by 72 hours (Fig 4b).

Fig. 4.

Fig. 4.

a) ROC curves of LSI perfusion data at 24 and 72 hours post-burn. b) Perfusion images from 5, 20, 25 and 35 second burns at 24 and 72 hours post-burn.

3.4. Thermal Imaging

The ROC curve for the temperature values collected via thermal imaging at 24 and 72 hour time points are shown in Fig. 5a. The AUC of the 24 hour thermography values was 0.76, while the AUC of the 72 hour thermography values was 0.79. Both values were statistically significant (p<0.01). After choosing a threshold of 34.9 °C, the overall accuracy off the 24 hour thermography values were 73% (correct classification of 35/48) with a sensitivity of 66% and a specificity of 84%. The positive and negative predictive values were 86% and 61% respectively. The 72 hour thermography values had an overall accuracy of 75% (correct classification of 36/48) with a sensitivity of 76% and a specificity of 74%. The positive and negative predictive values were 82% and 67% respectively. There was no statistically significant difference between the classifying ability of the thermography values at 24 and 72 hours. An example of typical thermography images collected are shown in Fig. 5b. At 24 hours, temperature within the superficial partial burn (5 sec) regions are higher than unburned tissue. The 20 sec burn demonstrated similar temperature to unburned tissue at 24 hours, and both the 25 and 35 sec burns were lower in temperature than unburned tissue at 24 hours. At 72 hours, the temperature within the 5 sec burn was similar to unburned tissue. The 20 sec, 25 sec and 35 sec burns had lower temperatures than unburned tissue at 72 hours. The temperature measurements of the unburned regions during 72 hours measurements were higher than that recorded at 24 hours.

Fig. 5.

Fig. 5.

a) ROC curves of thermography data from 24 and 72 hours post-burn. b) Thermography images from 5, 20, 25 and 35 second burns at 24 and 72 hours post-burn.

3.5. SFDI measured reduced scattering changes

The averaged reduced scattering coefficient, (μs’) values at 24 and 72 hours were used to create the ROC shown in Fig. 6a. The AUC of the 24 hour μs’ values was 0.90, while the AUC of the 72 hour AUC values was 0.91. Both values were statistically significant (p<0.001). After choosing a threshold of 1.23 mm−1, the overall accuracy off the 24 hour perfusion values 85% (41/48) with a sensitivity of 93% and a specificity of 74%. The positive and negative predictive values were 84% and 88% respectively. The 72 hour μs’ values also had an overall accuracy of 85% (correct classification of 41/48) with a sensitivity of 83% and a specificity of 90%. The positive value was 92% and the negative predictive value was 77%. There was no statistically significant difference between the classifying ability of the μs’ values at 24 and 72 hours. Fig. 6b shows images of μs’ typical perfusion images at 24 hours and 72 hours.

Fig. 6.

Fig. 6.

a) ROC curves of SFDI scattering data from 24 and 72 hours post-burn. b) Images of reduced scattering at 659 nm from 5, 20, 25 and 35 second burns at 24 and 72 hours post-burn.

3.6. Summary

A summary of the statistics for all the imaging modalities at 24 and 72 hours are shown in Table 3.

Table 3.

Summary of statistics obtained using different measurement modalities

Accuracy (%) Sensitivity
(%)
Specificity
(%)
PPV (%)  NPV (%)
Clinical
Assessment
(24 hours)
83 100 58 78 100
LSI (24 hours) 75 66 90 91 63
Thermography
(24 hours)
73 66 84 86 62
SFDI (24
hours)
85 93 74 84 88
Clinical
Assessment
(72 hours)
85 100 63 81 100
LSI (72 hours) 83 100 58 78 100
Thermography
(72 hours)
75 76 74 82 67
SFDI (72
hours)
85 83 90 92 77

A comparison of the classifiers at 24 hours, as a function of measurement modality, using McNemar’s test, showed a statistically significant difference when comparing SFDI with LSI and thermal imaging. Additionally, there was a statistically significant difference when comparing clinical analysis with LSI and thermal imaging. Results for all modalities are shown in Table 4.

Table 4.

McNemar’s test p-values for comparing modalities at 24 hours.

Clinical LSI Thermography SFDI
Clinical N/A <0.001* <0.001* 0.07
LSI <0.001* N/A 1.00 0.003*
Thermography <0.001* 1.00 N/A 0.004*
SFDI 0.07 0.003* 0.004* N/A
*

denotes p<0.05

4. Discussion

4.1. Histological Assessment

As one of the gold standards for determining burn severity[24, 25], histology illustrates detailed changes within tissue after the burn wound is created, as seen in Fig. 3. While histology represents a conventional and relatively objective means to determine burn depth, taking and processing biopsies is a time consuming process. In addition, it is an extremly invasive procedure which is not practical from a clinical standpoint. Furthermore, the relatively small region of tissue sampled by histology is likely not representative of the severity of the entire burn region, which in our experience may be several hundreds of square centimeters in extent. Indeed, spatial heterogeneity of burns are well documented, even in controlled studies like the current one that aim for homogenous sounds (as seen in scattering values in Fig. 6b). The rapid collagen changes associated with the zone of coagulation within burns may explain why techniques like SFDI, which is sensitive to structural changes[21, 22], may provide information as early as a few hours post-burn[23]. hysiological changes such as perfusion and temperature typically occur 48–72 hours post-burn[3].

4.2. Laser Speckle Imaging

Other investigators have found that LSI technology becomes compelling as a tool with which to assess burn wound severity in the 48–72 hour time window[29]. When comparing the accuracy of clinical observation and laser Doppler imaging techniques, time points of 24 hours and 72 hours have been used to highlight the advantage of laser Doppler imaging over clinical assessments[3]. The 24 hour time point provides the benefit of being clinically relevant in terms of when patients often present to clinical staff for burn wound treatment. Commercialized LSI devices have frequently been used to study burns in animal models and also in a clinical setting on human patients[30]. They have typically demonstrated an ability to categorize which burns require surgical intervention more accurately than clinicians[3, 8, 29]. LSI typically has an accuracy of about 80% 24 hours after the burn that increases to approximately 95% at 72 hours after the burn[3]. While we saw an accuracy of 75% using LSI at 24 hours, there was not a significant increase in accuracy at 72 hours. additionally, as seen in Fig. 4b, there was no increase in perfusion for burns that do not require intervention. This increase is well documented in the clinical setting and likely leads to the increased classification accuracy at the later time points. It is possible that this response is delayed in the animal model studied. Other studies that have focused on animal models have similarly seen decreased perfusion that can last for 7 to 14 days depending on burn severity. This delayed perfusion response could help explain the lack of a larger increase in diagnostic accuracy at 72 hours. The same study investigated the classification ability of LSI and reported the AUC of perfusion measurements taken 1 hour after burn to be 78%, which was similar to the AUC of 79% for our 24 hour measurements[31].

4.3. Thermal Imaging

Thermal imaging was able to classify burns at 24 and 72 hours using temperature measurements, but ultimately performed worse than clinical assessment, LSI and SFDI. A 2017 study recently compared real-time clinical assessments and passive thermal imaging of burns on clinical patients[8]. The study found that thermal imaging at 24 hours was 56% accurate, and at 72 hours was 71% accurate. Realtime clinical assessments outperformed thermal imaging and was 88% accurate at 24 hours. We found similar results in our study as real-time clinical assessments (83% accuracy) outperformed thermal imaging at 24 hours (73% accuracy) and 72 hours (75% accuracy).

4.4. SFDI

In terms of overall accuracy, SFDI (85%) was the only imaging modality to perform better than clinical assessment (83%) at 24 hours. It also performed significantly better than LSI and thermal imaging at classifying burns by whether they would require surgical intervention. While previous work using SFDI in rat burn experiments[23] as well as pig burn experiments[21, 22, 32] have focused on the ability to distinguish different categories of burns, here we have focused on directly comparing it to clinical observation along with other popular imaging modalities for burn care.key difference bet ween SFDI and the other imaging modalities is that the reduced scattering coefficient changes are largely in response to changes in tissue structure that occur immediately after the burn, whereas LSI and thermal imaging rely on perfusion changes that typically occur days later. While early measurements from all devices perform well, LSI and thermal imaging did not exceed what could be done by clinical analysis alone. SFDI outperformed the other imaging modalities and has the added benefit of measuring multiple parameters meaning future work can focus on optimizing SFDI measurements.

4.5. Summary

In this experiment, 16 burns were created on 3 pigs yielding 48 burns that were treated as independent data points. While the different burn wounds in an individual pig may not be completely independent of each other, this assumption is necessary in order to develop enough data points for any significant analysis similar to what other groups have done[13, 31, 3336]. Based on the size of the burn wounds relative to the total body surface area (~1%) there should be no systemic effects and the distance between each wound (~3 cm) should ensure that these wounds represent unique samples from a diagnostic imaging standpoint.

While we have examined the results from these imaging techniques in a controlled experiment, it is important to consider the strengths and weaknesses of each technique in a clinical setting. Laser speckle imaging techniques have the advantage of being well established for examining burn wounds based on a long history that began with laser Doppler techniques that evolved into laser Doppler imaging[37]. LSI has considerably reduced acquisition times from previous scanning laser Doppler techniques. This increase in speed reduces movement artifacts and enables real time feedback to clinicians. The clinical accuracy of LSI over clinical observation is well documented[29], but the expense of current commercial systems does limit widespread use[5]. Thermal imaging systems also have fast acquisition times and can provide real time feedback to clinicians. Hand-held systems have recently been commercialized and these are relatively inexpensive[9]. However thermal imaging can be very sensitive to fluctuations in ambient temperature[10] and studies that have directly compared laser Doppler techniques and thermal imaging have found that laser Doppler can be more accurate in terms of categorizing burn severity [8]. SFDI is a newer technology and commercial realization of devices is in early stages. Thus, interfaces and data collection procedures, while appropriate for research studies, are still being refined to enable clinical data collection. SFDI collects a larger data set (multiple wavelength images at multiple spatial frequencies) than LSI or thermal imaging systems. This means that data acquisition can take longer (up to a minute at the current state of development employed here) and data analysis is still being refined in order to optimize the information content of the results. However the additional information provided by SFDI allows it to outperform the other systems in terms of burn assessment accuracy in these controlled experiments. Future studies have the potential to optimize the information content provided by SFDI in terms that are most relevant for burn wounds. This will enable reduction in data set sizes and thus shorter acquisition times that present only the most relevant data to clinicians.

5. Conclusion

SFDI, LSI and thermal imaging are all non-invasive modalities that according to the literature can help discriminate differences in burn severity and clinical decision making[1, 8]. While new technologies have the capability to aid the decision making capabilities of burn clinicians, they must be tested in a controlled environment to gauge their efficacy. These techniques have been studied individually and sometimes comparatively in clinical and research settings, but here we present those modalities in relationship to clinical analysis in a controlled animal model in which burn severity has been verified using histological assessment. We found that SFDI was the only modality to have a diagnostic accuracy higher than clinical assessment.

Highlights.

  • Clinical examination is the standard for diagnosing burn severity

  • Accurate assessments at early time points are difficult but could provide benefit

  • We compared three imaging technologies to clinical assessment in a pig model

  • Only one technique outperformed clinical assessment

  • Spatial Frequency Domain Imaging could be a useful tool to aid clinical diagnosis

Acknowledgments

We gratefully acknowledge support from NIGMS grant R01GM108634 and the NIBIB, including P41EB015890 (A Biomedical Technology Resource). The content is solely the responsibility of the authors and does not necessarily represent the official views of the NIGMS, NIBIB or NIH. We also acknowledge the Arnold and Mabel Beckman Foundation. We would also like to thank Earl Steward for facilitating the animal work involved in this study.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  • [1].Peck MD. Epidemiology of burns throughout the world. Part I: Distribution and risk factors. Burns 2011;37:1087–100. [DOI] [PubMed] [Google Scholar]
  • [2].Shupp JW, Nasabzadeh TJ, Rosenthal DS, Jordan MH, Fidler P, Jeng JC. A review of the local pathophysiologic bases of burn wound progression. Journal of burn care & research : official publication of the American Burn Association 2010;31:849–73. [DOI] [PubMed] [Google Scholar]
  • [3].Hoeksema H, Van de Sijpe K, Tondu T, Hamdi M, Van Landuyt K, Blondeel P, et al. Accuracy of early burn depth assessment by laser Doppler imaging on different days post burn. Burns 2009;35:36–45. [DOI] [PubMed] [Google Scholar]
  • [4].Burke-Smith A, Collier J, Jones I. A comparison of non-invasive imaging modalities: Infrared thermography, spectrophotometric intracutaneous analysis and laser Doppler imaging for the assessment of adult burns. Burns 2015;41:1695–707. [DOI] [PubMed] [Google Scholar]
  • [5].Kaiser M, Yafi A, Cinat M, Choi B, Durkin AJ. Noninvasive assessment of burn wound severity using optical technology: a review of current and future modalities. Burns 2011;37:377–86. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [6].Krezdorn N, Limbourg A, Paprottka FJ, Konneker, Ipaktchi R, Vogt PM Assessing burn depth in tattooed burn lesions with LASCA Imaging. Ann Burns Fire Disasters 2016;29:223–7. [PMC free article] [PubMed] [Google Scholar]
  • [7].Nguyen K, Ward D, Lam L, Holland AJ. Laser Doppler Imaging prediction of burn wound outcome in children: is it possible before 48 h? Burns 2010;36:793–8. [DOI] [PubMed] [Google Scholar]
  • [8].Wearn C, Lee KC, Hardwicke J, Allouni A, Bamford A, Nightingale P, et al. Prospective comparative evaluation study of Laser Doppler Imaging and thermal imaging in the assessment of burn depth. Burns 2018;44:124–33. [DOI] [PubMed] [Google Scholar]
  • [9].Burmeister DM, Cerna C, Becerra SC, Sloan M, Wilmink G, Christy RJ. Noninvasive Techniques for the Determination of Burn Severity in Real Time. Journal of burn care & research : official publication of the American Burn Association 2017;38:e180–e91. [DOI] [PubMed] [Google Scholar]
  • [10].Miccio J, Parikh S, Marinaro X, Prasad A, McClain S, Singer AJ, et al. Forward-looking infrared imaging predicts ultimate burn depth in a porcine vertical injury progression model. Burns 2016;42:397–404. [DOI] [PubMed] [Google Scholar]
  • [11].Paluchowski LA, Nordgaard HB, Bjorgan A, Hov H, Berget SM, Randeberg LL. Can spectral-spatial image segmentation be used to discriminate experimental burn woundsq? Journal of biomedical optics 2016;21:101413. [DOI] [PubMed] [Google Scholar]
  • [12].Calin MA, Parasca SV, Savastru R, Manea D. Characterization of burns using hyperspectral imaging technique - a preliminary study. Burns 2015;41:118–24. [DOI] [PubMed] [Google Scholar]
  • [13].King DR, Li W, Squiers JJ, Mohan R, Sellke E, Mo W, et al. Surgical wound debridement sequentially characterized in a porcine burn model with multispectral imaging. Burns 2015;41:1478–87. [DOI] [PubMed] [Google Scholar]
  • [14].Chin MS, Babchenko O, Lujan-Hernandez J, Nobel L, Ignotz R, Lalikos JF. Hyperspectral Imaging for Burn Depth Assessment in an Animal Model. Plastic and reconstructive surgery Global open 2015;3:e591. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [15].Cuccia DJ, Bevilacqua F, Durkin AJ, Ayers FR, Tromberg BJ. Quantitation and mapping of tissue optical properties using modulated imaging. Journal of biomedical optics 2009;14:024012. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [16].Cuccia DJ, Bevilacqua F, Durkin AJ, Tromberg BJ. Modulated imaging: quantitative analysis and tomography of turbid media in the spatial-frequency domain. Optics letters 2005;30:1354–6. [DOI] [PubMed] [Google Scholar]
  • [17].Torabzadeh M, Park IY, Bartels RA, Durkin AJ, Tromberg BJ. Compressed single pixel imaging in the spatial frequency domain. Journal of biomedical optics 2017;22:30501. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [18].Wilson RH, Crouzet C, Torabzadeh M, Bazrafkan A, Farahabadi H, Jamasian B, et al. High-speed spatial frequency domain imaging of rat cortex detects dynamic optical and physiological properties following cardiac arrest and resuscitation. Neurophotonics 2017;4:045008. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [19].Ponticorvo A, Taydas E, Mazhar A, Scholz T, Kim HS, Rimler J, et al. Quantitative assessment of partial vascular occlusions in a swine pedicle flap model using spatial frequency domain imaging. Biomedical optics express 2013;4:298–306. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [20].Yafi A, Vetter TS, Scholz T, Patel S, Saager RB, Cuccia J, et al. Postoperative quantitative assessment of reconstructive tissue status in a cutaneous flap model using spatial frequency domain imaging. Plastic and reconstructive surgery 2011;127:117–30. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [21].Ponticorvo A, Burmeister DM, Yang B, Choi B, Christy RJ, Durkin AJ. Quantitative assessment of graded burn wounds in a porcine model using spatial frequency domain imaging (SFDI) and laser speckle imaging (LSI). Biomedical optics express 2014;5:3467–81. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [22].Burmeister DM, Ponticorvo A, Yang B, Becerra SC, Choi B, Durkin AJ, et al. Utility of spatial frequency domain imaging (SFDI) and laser speckle imaging (LSI) to non-invasively diagnose burn depth in a porcine model. Burns 2015;41:1242–52. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [23].Nguyen JQ, Crouzet C, Mai T, Riola K, Uchitel D, Liaw LH, et al. Spatial frequency domain imaging of burn wounds in a preclinical model of graded burn severity. Journal of biomedical optics 2013;18:66010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [24].Watts AM, Tyler MP, Perry ME, Roberts AH, McGrouther DA. Burn depth and its histological measurement. Burns 2001;27:154–60. [DOI] [PubMed] [Google Scholar]
  • [25].Thomsen S Pathologic analysis of photothermal and photomechanical effects of laser-tissue interactions. Photochemistry and photobiology 1991;53:825–35. [DOI] [PubMed] [Google Scholar]
  • [26].Boylan A, Martin CJ, Gardner GG. Infrared emissivity of burn wounds. Clin Phys Physiol Meas 1992;13:125–7. [DOI] [PubMed] [Google Scholar]
  • [27].Ayers F, Grant A, Kuo D, Cuccia DJ, Durkin AJ. Fabrication and characterization of silicone-based tissue phantoms with tunable optical properties in the visible and near infrared domain. Proc Spie 2008;6870.
  • [28].Fluss R, Faraggi D, Reiser B. Estimation of the Youden Index and its associated cutoff point. Biom J 2005;47:458–72. [DOI] [PubMed] [Google Scholar]
  • [29].Shin JY, Yi HS. Diagnostic accuracy of laser Doppler imaging in burn depth assessment: Systematic review and meta-analysis. Burns 2016;42:1369–76. [DOI] [PubMed] [Google Scholar]
  • [30].Lindahl F, Tesselaar E, Sjoberg F. Assessing paediatric scald injuries using Laser Speckle Contrast Imaging. Burns 2013;39:662–6. [DOI] [PubMed] [Google Scholar]
  • [31].Ganapathy P, Tamminedi T, Qin Y, Nanney L, Cardwell N, Pollins A, et al. Dual-imaging system for burn depth diagnosis. Burns 2014;40:67–81. [DOI] [PubMed] [Google Scholar]
  • [32].Mazhar A, Saggese S, Pollins AC, Cardwell NL, Nanney L, Cuccia DJ. Noncontact imaging of burn depth and extent in a porcine model using spatial frequency domain imaging. Journal of biomedical optics 2014;19:086019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [33].Heredia-Juesas J, Thatcher JE, Lu Y, Squiers JJ, King D, Fan W, et al. Burn-injured tissue detection for debridement surgery through the combination of non-invasive optical imaging techniques. Biomedical optics express 2018;9:1809–26. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [34].Gaines C, Poranki D, Du W, Clark RA, Van Dyke M. Development of a porcine deep partial thickness burn model. Burns 2013;39:311–9. [DOI] [PubMed] [Google Scholar]
  • [35].Sowa MG, Leonardi L, Payette JR, Cross KM, Gomez M, Fish JS. Classification of burn injuries using near-infrared spectroscopy. Journal of biomedical optics 2006;11:054002. [DOI] [PubMed] [Google Scholar]
  • [36].Gnyawali SC, Barki KG, Mathew-Steiner SS, Dixith S, Vanzant D, Kim J, et al. High-resolution harmonics ultrasound imaging for non-invasive characterization of wound healing in a pre-clinical swine model. PloS one 2015;10:e0122327. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [37].Stewart CJ, Frank R, Forrester KR, Tulip J, Lindsay R, Bray RC. A comparison of two laser-based methods for determination of burn scar perfusion: laser Doppler versus laser speckle imaging. Burns 2005;31:744–52. [DOI] [PubMed] [Google Scholar]
  • [38].Ponticorvo A, Rowland R, Baldado M, Burmeister DM, Christy RJ, Bernal NP, Durkin AJ, Evaluating clinical observation versus Spatial Frequency Domain Imaging (SFDI), Laser Speckle Imaging (LSI) and thermal imaging for the assessment of burn depth, Burns, 10.1016/j.burns.2018.09.026 [DOI] [PMC free article] [PubMed]

RESOURCES