Quantifying Amide Proton Exchange rate and Concentration in Chemical Exchange Saturation Transfer Imaging of the Human Brain

Hye-Young Heo 1,2, Zheng Han 1,2, Shanshan Jiang 1, Michael Schär 1, Peter CM van Zijl 1,2, Jinyuan Zhou 1,2
PMCID: PMC6422734  NIHMSID: NIHMS1519197  PMID: 30654175

Abstract

Current chemical exchange saturation transfer (CEST) neuroimaging protocols typically acquire CEST-weighted images, and, as such, do not essentially provide quantitative proton-specific exchange rates (or brain pH) and concentrations. We developed a dictionary-free MR fingerprinting (MRF) technique to allow CEST parameter quantification with a reduced data set. This was accomplished by subgrouping proton exchange models (SPEM), taking amide proton transfer (APT) as an example, into two-pool (water and semisolid macromolecules) and three-pool (water, semisolid macromolecules, and amide protons) models. A variable radiofrequency saturation scheme was used to generate unique signal evolutions for different tissues, reflecting their CEST parameters. The proposed MRF-SPEM method was validated using Bloch-McConnell equation-based digital phantoms with known ground-truth, which showed that MRF-SPEM can achieve a high degree of accuracy and precision for absolute CEST parameter quantification and CEST phantoms. For in-vivo studies at 3 T, using the same model as in the simulations, synthetic Z-spectra were generated using rates and concentrations estimated from the MRF-SPEM reconstruction and compared with experimentally measured Z-spectra as the standard for optimization. The MRF-SPEM technique can provide rapid and quantitative human brain CEST mapping.

Keywords: CEST, APT, exchange rate, concentration, MR fingerprinting

1. Introduction

Chemical exchange saturation transfer (CEST) imaging is an important molecular MRI technique that can generate contrast based on the proton exchange between labeled protons in solutes and free bulk water protons, or on the water exchange between solute-bound water and free bulk water in tissue (Goffeney et al., 2001; Ward et al., 2000). CEST-MRI was developed to indirectly detect various endogenous low-concentration mobile biomolecules with water-exchangeable protons, such as proteins and peptides (van Zijl et al., 2003; Zhou et al., 2003b); glutamate (Cai et al., 2012); creatine (Haris et al., 2014); glycogen (van Zijl et al., 2007); and glycosaminoglycans (Ling et al., 2008); exogenous-specific molecular imaging agents (Aime et al., 2005; Walker-Samuel et al., 2013; Zhang et al., 2001); as well as tissue physiological parameters, including temperature (Zhang et al., 2005) and pH (Sun et al., 2007; Zhou et al., 2003b). Currently, CEST-MRI holds great promise for abundant clinical molecular imaging applications (Jiang et al., 2017; Jones et al., 2017; Li et al., 2014; Zhang et al., 2016a; Zhang et al., 2017a; Zhou et al., 2013b).

Most current CEST-MRI protocols acquire CEST-weighted images that reflect a few other contributions, including residual magnetization transfer contrast (MTC) and residual tissue relaxation characteristics (Xu et al., 2014b; Zaiss et al., 2011; Zhou et al., 2008), thus limiting the assessment of quantitative proton exchange rates and concentrations. In addition, inconsistent and controversial results have been reported by different research groups due to the choice of different CEST metrics (Heo et al., 2017a; Xu et al., 2014b; Zaiss et al., 2011), reference images (Cai et al., 2015; Heo et al., 2016c; Jin et al., 2013; Jones et al., 2013; Zhou et al., 2003b), and different experimental parameters (Heo et al., 2017c; Sun et al., 2013; Zhao et al., 2011; Zhou et al., 2013a). Investigators in the CEST community have a great interest in quantifying label proton concentrations and exchange rates. One of the most promising CEST quantification methods is to fit CEST signals obtained from repeated and serial image acquisition with varied saturation powers, as well as saturation frequency offsets to the steady-state analytical solution of the Bloch-McConnell equation (Geades et al., 2017; Heo et al., 2016c; Liu et al., 2013; McMahon et al., 2006; Woessner et al., 2005; Zhou et al., 2004). The acquisition of multi-sampled Z-spectra, i.e., acquisitions at multiple B1 or saturation time settings, allows analysis by model-based fitting of the CEST process. The simplest CEST model-based analysis consists of two pools, free bulk water protons and single solute protons. However, MTC effects from semisolid macromolecular protons in-vivo interfere with the CEST process (Desmond and Stanisz, 2012; Heo et al., 2016b; Hua et al., 2007; van Zijl and Yadav, 2011). Incorporating multiple pools in the model-based analysis is challenging because it requires more parameters that must be fitted from the data, leading to a higher risk of over-fitting errors, and thus, inaccurate quantification results. In addition, the use of multiple RF saturation frequencies under varied RF saturation powers requires long scan times, which is a major obstacle to clinical translation.

MR fingerprinting (MRF) is a new approach for efficient multiple tissue parameter mapping with varying pulse sequence parameters of interest in a pseudorandom manner (Cohen et al., 2018; Liao et al., 2017; Ma et al., 2013; Ma et al., 2017; Su et al., 2017; Wang et al., 2017; Zhou et al., 2018). Typically, a pattern-matching algorithm is used to find different tissue-type parameters against a pre-calculated database, a so-called “dictionary,” from Bloch equation simulations with a wide variety of tissue parameter combinations. Therefore, MRF has been shown to be robust to severe aliasing artifacts from highly undersampled k-space measurements due to the incoherent sampling and the nature of a matching procedure based on prior information. When multiple components for CEST-MRI are considered, however, it requires even more parameters that must be matched from the dictionary, possibly leading to erroneous quantification results. Furthermore, the size of the database would have to be dramatically increased and an exhaustive search performed, limiting practical application.

In this study, we developed a dictionary-free MRF technique to allow CEST quantification with a reduced data set by using the simplifying assumption of subgrouping proton exchange models (MRF-SPEM). As a first example of application to amide proton transfer (APT) MRI (Zhou et al., 2003a; Zhou et al., 2003b), we assumed two-pool (water and semisolid macromolecules) and three-pool (water, semisolid macromolecules, and amide protons) models. A variable RF saturation scheme was used to generate uncorrelated signal evolutions for different tissue properties. Carefully designed frequency sampling and RF saturation power variation further allowed quantification in terms of group-based average exchange rates and apparent concentrations at a certain amide proton frequency. The MRF-SPEM method was validated using Bloch-McConnell equation-based digital phantoms with known ground-truth values. For in-vivo studies, synthetic Z-spectra were generated by inserting CEST quantities obtained from MRF-SPEM reconstruction into the forward Bloch transform. These were compared with measured Z-spectra as the standard due to the lack of an objective ground-truth or gold standard in-vivo.

2. Methods

2.1. MRF-SPEM Acquisition

In the MRF-SPEM framework, RF saturation frequency offsets (Ω), saturation power (B1), saturation duration (Ts), and repetition time (TR) were varied throughout the acquisition, generating unique signal evolutions for different tissue properties, as shown in Fig. 1. TRs were varied according to Ts, but relaxation delay (Td) and turbo spin echo (TSE) acquisition (Ta) times were fixed (TR = Ts + Ta + Td). For APT quantification, MRF-SPEM images consisted of two distinct datasets: (1) MTC data with far off-resonance frequency offsets between 10 ppm and 50 ppm (black crosses in Fig. 2a); and (2) APT-weighted data with saturation frequency offsets between 3 ppm and 4 ppm (red crosses in Fig. 2a). The far off-resonance frequency offsets were chosen to sample MTC data points and to avoid possible downfield CEST and upfield nuclear Overhauser enhancement (NOE) signal contributions to water saturation originating from mobile proteins and peptides. These data were fitted to a two-pool MTC exchange model, while APT-weighted data acquired at RF saturation frequency offsets around 3.5 ppm were fitted to the three-pool exchange model. The two-pool parameters were incorporated into the three-pool model as prior known information, reducing the number of parameters and fitting uncertainty errors. In addition, fast-exchanging amine protons around 3 ppm in glutamate (Cai et al., 2012), hydroxyl protons around 1 ppm in myoinositol (Haris et al., 2011), glycogen (van Zijl et al., 2007), glycosaminoglycans (Ling et al., 2008), side‐chain amide or guanidinium protons around 2 ppm in proteins (Zhang et al., 2017d; Zhou et al., 2003a), and guanidinium protons in creatine (Cai et al., 2015; Haris et al., 2012) may make contributions to the APT-weighted signals. At lower RF saturation power, CEST effects are dominated by slowly exchanging amide protons, whereas, at higher B1, this is taken over by fast-exchanging amine and hydroxyl protons (Heo et al., 2017b; Jin et al., 2012; van Zijl et al., 2018). Based on this theory, relatively low B1 powers between 0.5 μT and 1.2 μT were applied in an attempt to isolate APT signals from the broad resonances of rapidly exchangeable protons.

Figure 1.

Figure 1

(a) An illustration of turbo spin-echo (TSE)-based CEST sequence with four (i = 4) dynamic scans. A saturation pulse train consists of block pulses of 200 ms duration each with a 10 ms delay between them. (SPIR = spectral pre-saturation with inversion recovery). (b) The evolution of the longitudinal magnetization of the water (Mzw) with varied saturation parameters, where Ωi = 4, 3, 3.5, 10 ppm, B1i = 1.2, 0.8, 2, 3 μT, Tsi = 800, 400, 800, 600 ms, and TRi = 3.5, 3.08, 3.5, 3.29 sec., respectively. (c) Zoomed-in displays (pink boxes in b) showing distinct saturation effects. All initial magnetizations for each dynamic scan are identical due to the fixed relaxation delay time.

Figure 2.

Figure 2

(a) An example of Ω, B1, Ts, and TR schedules for an MRF-SPEM acquisition. Black crosses are for two-pool MTC model-fitting (10 ppm < Ω < 50 ppm and 1.2 μT < B1 < 3 μT) and red crosses (3 ppm < Ω < 4 ppm and 0.5 μT < B1 < 1.2μT) are for three-pool model-fitting. (b) MRF-SPEM signal profiles obtained from five compartments with varied proton exchange rates and concentrations listed in (c). (c) Pixel-wise maps of the proton exchange rate (ksw, Hz) and concentration (M0s, mM) in digital phantoms (SNR = 100) with the variable RF saturation schedules as shown in (a). (c) Correlation plots comparing proton exchange rates (top) and concentrations (bottom) between MRFSPEM reconstruction values and ground-truth (GT) phantom values.

2.2. Bloch-McConnell Simulations

Three-pool, Bloch-McConnell equation-based simulations were performed to simulate MRF-SPEM signal profiles. All longitudinal magnetization evolutions were analytically simulated following a single-shot TSE sequence, as shown in Fig. 1. MRF-SPEM profiles were generated with a schedule of Ω, B1, Ts, and TR, and were used for digital phantoms that consisted of two compartments (C1 and C2) with different exchange rates and concentration parameters for CEST and MTC contributions. Before simulating MRF-SPEM Z-spectra, four dummy scans were employed to achieve steady-state. The parameters (C1/C2) used for the simulation are shown in Table 1 (left two columns). Furthermore, T2 values were assumed equal in both compartments, namely 65 ms for water and amide protons and 20 μs for MTC protons. The chemical shifts of water, MTC, and amide protons were 0 ppm, 0 ppm, and 3.5 ppm, respectively. B0 and B1 fields were assumed to be homogenous. For comparison with magnetization transfer ratio asymmetry (MTRasym) analysis, conventional Z-spectra were also generated at the RF frequency offset ranging from −6 to +6 ppm for saturation powers of 0.5, 1, 1.5, 2, and 3 μT, and a saturation duration of 800 ms (a series of four block RF saturation pulses, 200 ms duration each). A Bland-Altman analysis was performed to evaluate the agreement between MRF-SPEM and ground-truth phantom values. For all simulations, Rician noise was added to generate noisy images or profiles. We estimated an apparent standard deviation of ~855, scaled by a factor of 1/0.655 to account for the Rician noise distribution from a background region, and an average signal of 97138 from a white matter region in saturated MRF-SPEM images at 3.5 ppm (SNR of ~113). For the digital phantom study, we used a SNR level of 100 with the pseudo-random Rician noise samples which were generated by using conventional inverse transform sampling. The concentration of the water protons (110 M) was used to convert the semisolid macromolecular and amide protons concentrations from relative to absolute units for simulation, phantom, and in-vivo human studies.

Table 1.

Ground truth values of the three-pool exchange model and estimated parameters (mean ± standard deviation) from MRF-SPEM in the numerical phantom study.

Ground Truth MRF-SPEM
C1 C2 C1 C2
T1w (sec) 1.20 1.40 1.20 ± 0.00 1.40 ± 0.00
kmw (Hz) 20 20 20 ± 0 20 ± 1
M0m (M) 11.0 4.4 10.9 ± 0.0 4.4 ± 0.2
ksw (Hz) 300 70 301 ± 2 72 ± 2
M0s (mM) 150 250 149 ± 0 246 ± 0

Other simulation parameters are shown in the Method section.

2.3. Phantom Experiments

Ammonium chloride (NH4Cl, Sigma Aldrich, St Louis, MO, USA) phantoms were prepared in 50 mL tubes by dissolving 1.07g (500 mM) or 2.14g (1M) ammonium chloride in 40 mL PBS (pH 7.0), followed by titration to pH of 4.5, 4.6, and 5.0 using NaOH solution and a pH meter. To mimic the MTC pool, 1% agarose (low gelling temperature, Sigma Aldrich, St Louis, MO, USA) was added to tubes, which was heated briefly using microwave to dissolve agarose. Tubes were then bound using tapes and placed in a plastic container. The container was then filled with heated water containing 1% agarose, which was allowed to gel before scan. All studies were performed on a 3 T Philips Achieva MRI system (Best, Netherlands) with a 32-channel element head coil, and all reconstruction algorithms were implemented in MATLAB R2017 (The MathWorks, Inc., Natick, MA). For CEST imaging, three sets of imaging parameters were adopted. First, conventional Z-spectrums were acquired with 161 frequency offsets between −20 and 20 ppm at intervals of 0.25 ppm with RF saturation powers of 1, 1.5, 2, 2.5, and 3 μT. The scan time was 47.6 min. Second, densely sampled Z-spectrums were acquired with 94 frequency offsets between 1.5 and 3.5 ppm at intervals of ~0.0234 ppm and between 8 and 20 ppm at intervals of 1 ppm with RF saturation powers of 1, 1.5, 2, 2.5, 3, and 3.5 μT. The scan time was 33.6 min. Third, MRF-SPEM profiles were acquired at 26 frequency offsets between 1.5 and 3.5 ppm for three-pool CEST fitting and 5 frequency offsets between 10 and 20 ppm for two-pool MTC fitting with RF saturation powers ranging from 1 to 3 μT and RF saturation times ranging from 400 to 800 ms. The scan time was 1.9 min. Unsaturated images were also acquired for signal normalization. Additionally, water saturation shift-referencing (WASSR) (Kim et al., 2009) (26 frequency offsets from −1.2 to 1.2 ppm at intervals of 0.125 ppm, B1 of 0.5 μT) and dual-TR (TR = 20 and 120 ms) data were acquired for B0 and B1 corrections, respectively.

2.4. In-Vivo MRI Experiments

Human studies were approved by the Johns Hopkins Institutional Review Board. Five healthy volunteers (women, n=2; men, n=3; age, 32.6 ± 4.0) were recruited, who provided written, informed consent for the study. MRF-SPEM image data were obtained from a fat-suppressed (spectral pre-saturation with inversion recovery, SPIR), single-shot TSE pulse sequence using the following parameters: TE= 6.4 ms; FOV = 212 × 186 mm2; acquisition and reconstruction resolution = 2.2 × 2.2 × 4.4 and 0.83 × 0.83 × 4.4 mm3; turbo spin-echo factor = 45; and single slice acquisition. WASSR and dual-TR data were also acquired with the same parameters used for the ammonium chloride phantom and incorporated into the MRF-SPEM reconstruction framework for B0 and B1 corrections, respectively. The MRF-SPEM was also compared to the extrapolated semisolid magnetization transfer reference (EMR) method (Heo et al., 2016a; Heo et al., 2016c; Lee et al., 2017). For EMR data acquisitions, the frequency sweep corresponded to full Z-spectra with the following frequency offsets: unsaturated (S0), 0 to ± 6 ppm at intervals of 0.5 ppm; + 7 ppm to + 20 ppm at intervals of 1 ppm; + 25 ppm to + 50 ppm at intervals of 5 ppm; and 60 ppm. Three RF saturation powers (1, 1.5, and 2 μT) were applied. Total EMR scan time was 11.2 min.

2.5. MRF-SPEM Processing

For motion correction, all MRF-SPEM datasets were registered to the first saturated image using a rigid body transformation algorithm with a mutual information cost function and bicubic resampling (Zhang et al., 2016b). Then, we fitted MRF-SPEM data with the above-described two- and three-pool, Bloch-McConnell equation-based proton exchange models using a nonlinear least-squares fitting approach, which was implemented using the Levenberg-Marquardt algorithm. The three-pool exchange model, including the free bulk water proton pool (w), the water exchangeable solute proton pool (s), the semisolid macromolecular proton pool (m), and the magnetization in each pool (Mw, Ms, and Mm) in the presence of exchange and RF irradiation, can be described in matrix format by (Heo et al., 2016c; Woessner et al., 2005; Zhou and van Zijl, 2006; Zhou et al., 2004):

dM(t)dt=A(t)M(t)+B [1]

where

A(ω1)=[DwNwsNwmNswDsNsmNmvNmsDm] [2]
Di=[k2i(ωωi)0(ωωi)k2iω10ω1k1i] [3]
Nij=kij[100010001] [4]
B=[BwBsBm]T [5]
Bi=[00M0iR1i]T [6]

where ω1 is the RF saturation amplitude; ω is the RF saturation frequency; ωi is the resonance frequency of pool i; kij represents the proton exchange rate from pool i to pool j; R1i and R2i are the longitudinal and transverse relaxation rate of pool i, respectively; k1/2w, k1/2s, and k1/2m are given by R1/2w+kws+kwm, R1/2s+ksw, and R1/2m+kmw, respectively; and M0i is the equilibrium magnetization of pool i. The analytical solution of the coupled linear differential equations (Eq. [1]) is:

M=(M0+BA1)eAtBA1 [7]

where

M0=[M0wM0sM0m]T [8]
M0i=[00M0i]T [9]

The longitudinal magnetization evolution was generated based on a pulse-train CEST sequence with a single-shot TSE readout. Crusher gradients of 10-ms duration were applied between block RF saturation pulses to suppress residual transverse magnetization by setting zero transverse magnetization components (spoiler matrix). For the image acquisition, the magnetization vector evolution was calculated by multiplying the previous magnetization vectors with the rotation matrices to consider a 90° excitation pulse followed by a 180° refocusing pulse and multiple refocusing pulses forming echo trains. During the relaxation delay period, the magnetization vector was calculated solely by relaxation recovery process in the absence of B1. The three-pool APT exchange model can be reduced to the two-pool MTC exchange model (free bulk water + semisolid macromolecular proton pools) by assuming the concentration of amide proton pools to be zero. The RF saturation parameters and repetition times corresponding to saturation schedules were updated for each dynamic scan. The three-pool APT exchange model was analytically solved with the a priori fitted two-pool symmetric MTC parameters obtained from far off-resonance frequency offsets and the parameter fitting was performed using the minimum norm estimate. Our model fitting approach substantially reduced the risk of over-fitting and the uncertainties of direct three-pool model parameter estimates. The quality of the estimated parameters was evaluated by the root of the sum of the squared difference between the fitted and acquired MRF-SPEM data, and the χ2 goodness-of-fit metric. In the fitting process, the values of the parameters were constrained within upper and lower bounds of [0.1, 3.5 sec] for T1w, [0.01, 2 sec] for T2w, [1, 200 μsec] for T2m, [1, 200 Hz] for kmw, [1, 20 M] for M0m, [0.1, 500 msec] for T2s, [1, 600 Hz] for ksw, [1, 600 mM] for M0s for in-vivo studies, but adjusted for simulation and phantom studies. B0 shifts and B1 scaling factors obtained from the additional WASSR and B1 calibration scans were applied pixel-wise during the fitting.

In addition, Monte-Carlo simulation studies were performed to compare MRFSPEM and MRF (one-step fitting) methods with simulated MRF-SPEM signal profiles. 1000 sets of noisy images (SNR of ~100) were generated with different parameters of the semisolid macromolecular proton and amide proton pools. For each trial, the pool parameters were randomly chosen within the range of kmw from 20 to 60 Hz at intervals of 0.25 Hz, the range of M0m from 4 to 12 M at intervals of 0.05 M, the range of ksw from 50 to 500 Hz at intervals of 2.5 Hz, and the range of M0s from 100 to 400 mM at intervals of 2 mM.

2.6. Validation of MRF-SPEM using synthetic CEST data

A “true” gold standard does not currently exist for absolute CEST quantification of invivo brain tissue. Instead, synthetic Z-spectra (or CEST signals) were generated using CEST parameters obtained from the MRF-SPEM methodology and then, were compared with experimental measurements. Virtual scanner settings (e.g., TR, TE, TSE factor, excitation flip angles, Ω, B1, Ts, etc.) that corresponded to the experimental measurements were used. For a direct comparison with the standard, synthesized APT# signal intensities were calculated by:

APT#(syn)=Zref(+3.5ppm)Zlab(+3.5ppm) [10]

where Zref and Zlab are the reference image signal intensity and label image signal intensity normalized with respect to the image signal intensity measured without RF saturation, respectively. For APT#(syn) calculation, Zref and Zlab were calculated by using the Bloch-McConnell equation with two-pool and three-pool exchange models, respectively, with parameters obtained from MRF-SPEM. By assuming the concentration of a solute proton pool to be zero (M0s= 0), the three-pool model (Zlab) can be reduced to the two-pool model (Zref).

For the validation standard, reference and labeled Z-spectra were taken from the extrapolated semisolid MT reference (EMR) signal (namely, ZEMR) and the B0-corrected experimental Z-spectra (Zexp), respectively:

APT#(meas)=ZEMR(+3.5ppm)Zexp(+3.5ppm) [11]

The EMR calculation was performed according to the procedures based on recently published papers (Heo et al., 2018; Heo et al., 2016c). Wide-offset experimental Z-spectra with MTC data points between 8 and 60 ppm were chosen to avoid the interference of possible CEST and upfield NOE signals from mobile proteins and peptides and were fitted to a two-pool MTC model with a symmetric Lorentzian lineshape. The experimental MTC datasets acquired at three RF saturation powers (1, 1.5, and 2 μT) were simultaneously fitted to improve the EMR fitting quality. Then, the MTC and direct water saturation contributions were extrapolated to obtain baseline signals (ZEMR) at the respective RF saturation frequencies of interest (e.g., 3.5 ppm). To compare APT#(syn) results with the experimental measurement, APT#(meas), two ROIs enclosing the gray matter (GM) and white matter (WM) were analyzed. These ROIs were carefully drawn on the unsaturated S0 image. Data in graphs and tables are presented as mean ± standard deviation. Statistical analysis was performed using a one-way analysis of variance, followed by Tukey’s post hoc test. Statistical significance was considered at p < 0.05.

3. Results

3.1. Bloch-McConnell Simulation Results

The longitudinal magnetization of free bulk water (Mzw) was changed dynamically under a variable RF saturation scheme (Ω, B1, Ts, and TR), generating unique magnetization signals. Fig. 2a shows the pattern of RF saturation parameters and TR values varied deliberately throughout the acquisition, producing dynamic CEST-weighted images (red crosses) and MTC-weighted images (black crosses) where the magnitude of the MR signal of each voxel changed at every dynamic step (Fig. 2b). Based on our digital phantom studies (Fig. 2c) at an appropriate SNR level of 100 (SNR for WM measured is 113), excellent agreement was observed for MRF-SPEM and the known exchange rate and concentration (ground-truth values). As shown in Fig. 2d, the results show a strong linear correlation (R2 = 0.985 for ksw; R2 = 0.996 for M0s), with slopes of 0.995 for ksw and 1.000 for M0s. The y-intercepts were 1.82 Hz for ksw and −0.222 mM for M0s. For ksw, the 95% limits of agreement were −34 Hz to 35 Hz, with a mean bias of 0.38 Hz. For M0s, the 95% limits of agreement were −9 mM to 8.7 mM, with a mean bias of −0.17 mM (also see Fig. 3). A Bland-Altman analysis was performed to assess the agreement between MRF-SPEM reconstruction and ground-truth values at different SNR levels (Fig. 3). The analysis revealed no significant bias for the measurement of ksw and M0s over all SNR levels. However, the 95% limits of agreement became narrow and fewer outliers were present at higher SNRs. Monte-Carlo simulations were used to assess the precision and accuracy of two-pool MTC and solute proton parameter estimation by MRF-SPEM and MRF methods as shown in Fig. 4. The parameters estimated from MRF-SPEM were more accurate (close to ground truth) and precise (narrow distribution) than MRF method.

Figure 3.

Figure 3

Bland-Altman plots comparing proton (a) exchange rates and (b) concentrations between MRF-SPEM reconstruction values and ground-truth phantom values at different SNR levels. The 95% limits of agreement are indicated by dotted black lines, and the mean bias is indicated by the solid black line.

Figure 4.

Figure 4

Monte-Carlo simulation results (1000 trials) comparing the MRF-SPEM and MRF methods.

Fig. 5 shows another Bloch simulation result for the digital phantom with two compartments (C1 and C2 mimicked normal tissue and lesion, respectively). The parameters of the compartments are shown in Table 1. Interestingly, there was a negligible MTRasym(3.5ppm) signal difference between the two compartments at an RF saturation power of 1 μT (Fig. 5b and 5d), despite the different water relaxation rates, MTC (except for semisolid macromolecular proton exchange rate), and CEST parameters. Furthermore, the strong RF saturation power dependency of MTRasym(3.5ppm) image contrast can be seen clearly in Fig. 5b and 5d. A positive MTRasym(3.5ppm) image contrast between compartments C2 and C1 was observed at low RF power (0.5 μT) due to small MTC and direct saturation effects and the smaller exchange rate in C2 having close to maximal saturation. The MTRasym(3.5ppm) image contrast between two compartments became negative at higher RF saturation powers (> 1 μT) because the solute exchange rate of C1 was faster than that of C2 and the small APT effect in C2 was overwhelmed by the MTC and direct saturation effects. When using the variable RF saturation scheme, image voxels with different relaxation rates, water-exchangeable proton concentrations, and exchange rates evolved differently, thereby generating unique magnetization trajectories, as shown in Fig. 5c. Accurate and precise parameters (Fig. 5e) were successfully decoded, which were in excellent agreement with the ground-truth values (Table 1).

Figure 5.

Figure 5

(a) Two homogenous compartments (C1 and C2) in a digital phantom (SNR = 100) having different water T1 relaxation times, semisolid macromolecular proton concentrations, solute proton exchange rates, and concentration values, as shown in Table 1. (b) Conventional Z-spectra and MTR asymmetry curves at different RF saturation powers. (c) Unique MRF-SPEM signal evolution profiles obtained from two compartments. (d) MTRasym(3.5ppm) image contrasts under different RF saturation powers. (e) water, MTC, and solute proton quantitative maps. Note that the semisolid macromolecular proton exchange rates in both compartments are identical.

3.2. CEST Phantom Imaging Results

Fig. 6 shows ammonium chloride phantom results. The phantom consists of four compartments with either varying NH4Cl concentration or pH (Fig. 6a). RF saturation power dependencies of the direct water saturation, semisolid MTC, and CEST signals can be seen clearly as shown in Figs. 6b–c. Instead of using a repeated and serial image acquisition with varied saturation powers and saturation frequency offset, a variable RF saturation and acquisition created unique signal evolutions (Fig. 6d) for the compartments with different CEST concentrations or pH. Fig. 6e shows the image contrast of MTRasym(2.5ppm) between vials. However, it is not clear what the contribution (proton exchange rate or concentration) is of the underlying contrast on the observed CEST-weighted images. Quantitative CEST parameter maps obtained from the conventional Bloch equation fitting method which is a two-step fitting approach (two-pool MTC model fitting first, followed by three-pool model fitting) using the densely sampled Z-spectra (Fig. 6c) and MRF-SPEM reconstruction using the MRF-SPEM signal profile (Fig. 6d) created by the variable RF saturation are shown in Fig. 6f–g, respectively. The CEST parameter values estimated by MRF-SPEM were in good agreement with values estimated by using the reference measurement in spite of concurrent variation of CEST concentration and pH in each vial (Table 2).

Figure 6.

Figure 6

CEST phantom validation experiments. (a) a phantom with four compartments: (i) pH 4.5, 0.5 M NH4Cl + 1% agarose + PBS, (ii) pH 5.0, 0.5 M NH4Cl + 1% agarose + PBS, (iii) pH 4.6, 1 M NH4Cl + 1% agarose + PBS, and (iv) pH 7.0, 1% agarose + PBS. (b) Z-spectra and MTRasym curves, (c) densely sampled Z-spectra, and (d) MRF-SPEM signals obtained from four ROIs. (e) MTRasym(2.5 ppm) maps with RF saturation powers of 1, 1.5, 2, 2.5, and 3 μT. Quantitative CEST parameter maps from (f) the conventional three-pool Bloch-equation fitting method using the densely sampled Z-spectra as shown in (c) and (g) MRF-SPEM reconstruction using the MRF-SPEM signal profiles as shown in (d).

Table 2.

Estimated CEST parameters (exchange rates and concentrations) in the presence of MTC (1% agarose) pool from the ammonium chloride phantom study

Reference* MRF-SPEM
ksw (Hz): mean ± STD 95% CI ROI1 2599 ± 51 2592 ± 44
2585 − 2613 2586 − 2598
ROI2 3876 ± 79 3911 ± 28
3870 − 3883 3906 − 3916
ROI3 2692 ± 81 2771 ± 37
2674 − 2710 2767 − 2775
M0s (mM): mean ± STD 95% CI ROI1 446 ± 240 460 ± 154
445 − 448 451 − 470
ROI2 496 ± 74 486 ± 80
495 − 496 485 − 488
ROI3 1045 ± 197 1075 ± 79
1043 − 1046 1072 − 1079
*

Reference: a three-pool Bloch equation fitting method using densely sampled Z-spectra with multiple RF saturation powers.

3.3. Human Volunteer Imaging Results

To compare the proposed MRF-SPEM with conventional CEST measurements as a standard for validation of in-vivo CEST parameter quantification, Z-spectra were synthesized using estimated parameters from MRF-SPEM and compared with experimentally measured Z-spectra with three different RF saturation powers (1, 1.5, and 2 μT). Fig. 7a shows average ROI-based MRF-SPEM signal evolution profiles (+), and two-pool fitted (solid line), three-pool fitted (dashed line) curves of gray and white matter. The reference signals (solid lines) were calculated by the two-pool MTC model fitting with semisolid MTC data (black crosses in Fig. 7a). Shaded areas (dark gray and light gray) indicate APT signal effects by subtracting three-pool MRF-SPEM label signals from two-pool MRF-SPEM reference signals. The APT signal intensities were higher in the gray matter than in the white matter. Using CEST parameters estimated from MRF-SPEM, synthetic Z-spectra (Fig. 7b) were reconstructed with saturation/imaging parameters identical to those used in the standard scan (Fig. 7c–d). Synthesized (Fig. 7b) and experimentally measured (Fig. 7c) signals were in excellent agreement at positive frequency offsets (red vs. black solid lines and + signs in Fig. 7d). Note that the synthesized Zlab signals (Fig. 7b) evidently differed from the result of the experimental measurements (Fig. 7c) at negative frequency offsets (upfield from water) because only a three-pool exchange model (free bulk water, semisolid MTC, and downfield amide protons pools) was considered in MRF-SPEM and because the MTC effect was considered symmetric. As shown in Fig. 8a–b, both synthetic Zref(3.5ppm) values in gray matter and white matter were in excellent agreement with the EMR results. Interestingly, even though the difference was not statistically significant, the APT#(syn) values seemed slightly higher than the APT#(meas) values, as shown in Fig. 8c, probably because multiple CEST components contaminated the signal at the amide proton frequency (more detail provided in the Discussion section).

Figure 7.

Figure 7

(a) Average ROI-based MRF-SPEM signal evolution profiles (+), and two-pool fitted (solid lines) and three-pool fitted (dashed lines) curves from white matter and gray matter (n = 5). Note that shaded areas indicate APT effects by subtracting three-pool MRF-SPEM label signals from two-pool MRF-SPEM reference signals. Black and red crosses are two-pool MTC and three-pool APT data acquisitions, respectively. (b) Synthetic two-pool (red solid lines) and three-pool (red +) Z-spectra using quantitative parameters estimated from MRF-SPEM profiles. (c) Experimentally measured conventional Z-spectra with three different RF saturation powers as a standard for validating MRF-SPEM quantification. (d) Experimentally measured conventional Z-spectra (black +) and extrapolated two-pool MTC-fitted curves (ZEMR, black solid lines). Synthetic two-pool (red solid lines) and three-pool (red +) Z-spectra are shown for comparison.

Figure 8.

Figure 8

(a) Average synthesized and experimentally measured Z-spectra from gray matter and white matter. (b) Average experimentally measured ZEMR(3.5ppm) and synthesized Zref(3.5ppm). (c) Experimentally measured APT#(meas) (calculated by Eq. [11]) and synthesized APT#(syn) (calculated by Eq. [10]) image intensities obtained from gray matter and white matter. Error bars depict standard deviations. No values were statistically significant (p > 0.05).

Quantitative parameter maps and values of the healthy volunteer brain are shown in Fig. 9 and Table 3. The gray matter and white matter have very different semisolid macromolecular proton and amide proton concentrations and exchange rates. The semisolid macromolecular proton exchange rates for gray and white matter were ~40 Hz (95% confidence interval: 34 Hz – 46 Hz) and ~29 Hz (95% confidence interval: 24 Hz – 34 Hz), respectively, and the concentrations of ~6 M (95% confidence interval: 5.4 M – 7.1 M) and ~11 M (95% confidence interval: 10.3 M – 12.1 M), respectively, were in good agreement with previous observations (Mougin et al., 2010; Sled and Pike, 2001; Stanisz et al., 2005). The amide proton concentration in gray matter (266 ± 22 mM; 95% confidence interval: 238 mM – 293 mM) was somewhat higher than that of the white matter (212 ± 22 mM; 95% confidence interval: 185 mM – 240 mM), which is in line with observed APT effects (shaded areas in Fig. 7a). In addition, the amide proton exchange rate (~365 Hz; 95% confidence interval: 342 Hz – 389 Hz) of gray matter was significantly faster than that of white matter (~162 Hz; 95% confidence interval: 142 Hz – 182 Hz). All differences in the amide proton exchange rates and concentrations between the gray matter and white matter were significant (p < 0.05, see Table 3). Synthesized APT# images were in good agreement with the experimentally measured APT# images as shown in Fig. 9b–c. However, the intrinsic water T1 relaxation time of the white matter calculated from MRF-SPEM reconstruction was longer than the observed water T1 relaxation time of the white matter obtained from a modified look-locker inversion recovery experiment due to the effect of coupling to a semisolid macromolecular proton pool (Henkelman et al., 1993).

Figure 9.

Figure 9

(a) An unsaturated image with color coded ROIs (green: gray matter, red: white matter) and B0, B1, quantitative MTC (kmw and M0m), APT (ksw and M0s) maps of a representative healthy volunteer human brain. (b) experimentally measured T1 map (T1Wobs) from a modified look-locker inversion recovery and APT#(meas) images with RF saturation powers of 1, 1.5, and 2 μT. (c) Synthetic T1 map (T1w(syn)) and APT#(syn) images with RF saturation powers of 1, 1.5, and 2 μT.

Table 3.

Estimated free bulk water, semisolid MTC, and APT parameters for white matter and gray matter of the healthy volunteer human brain.

No. T1w (sec) T2w (msec) T2m (μsec) kmw (Hz) M0m (M) T2s (msec) ksw (Hz) M0s (mM)
WM GM WM GM WM GM WM GM WM GM WM GM WM GM WM GM
1 0.99 1.10 35.0 70.5 58 61 23 33 11.9 7.0 2.5 0.9 151 353 198 299
2 1.05 1.29 39.4 67.0 62 64 26 36 10.8 6.8 2.2 1.3 141 380 235 241
3 1.07 1.34 42.0 77.0 61 67 31 44 11.9 5.4 1.8 0.8 181 390 234 256
4 1.06 1.05 44.8 77.0 68 71 31 42 10.3 5.7 2.0 0.9 168 346 185 276
5 1.05 1.30 37.8 65.1 67 72 33 44 11.0 6.5 1.9 1.7 171 357 209 256
mean ± STD 1.05 ± 0.03 1.21 ± 0.13 39.8 ± 3.8 71.3 ± 5.6 63 ± 4 67 ± 5 29 ± 4 40 ± 5 11.2 ± 0.7 6.3 ± 0.7 2.1 ± 0.3 1.1 ± 0.4 162 ± 16 365 ± 19 212 ± 22 266 ± 22
95% CI 1.01 − 1.08 1.05 − 1.38 35.1 − 44.5 64.4 − 78.3 58 − 68 61 − 73 24 − 34 34 − 46 10.3 − 12.1 5.4 − 7.1 1.7 − 2.4 0.7 − 1.6 142 − 182 342 − 389 185 − 240 238 − 293
p-value 0.0330 < 0.001 0.0097 < 0.001 < 0.001 0.0112 < 0.001 0.0458

4. Discussion

We developed a novel MRF concept to allow CEST quantification by SPEM. Numerical phantom studies demonstrated that MRF-SPEM can enable a high degree of accuracy and precision in absolute CEST quantification at 3T magnetic field strength. The same model used in-vivo allowed fast quantification of apparent exchange rates and concentrations at the amide proton frequency. The present MRF-SPEM imaging scan took just 2 min 50 sec (including B0 and B1 mapping) for quantitative APT mapping (as compared to 11 min 12 sec for EMR image acquisition with three RF saturation powers).

Our proposed method cannot be systematically assessed in-vitro and in-vivo due to the lack of an objective ground-truth. To enable effective validation, in lieu of ground-truth, we compared MRF-SPEM with a conventional Bloch equation fitting method with high-resolution Z-spectra using ammonium chloride-agarose phantoms and observed great agreement between the two measurements for the proton quantification. However, the estimated proton exchange rates were higher than the value reported in a previous ammonium chloride phantom study with similar pH ranges (Desmond and Stanisz, 2012). We observed broadened CEST peaks with significant asymmetry in the Z-spectra due to coalescence of the fast exchanging proton peaks with the bulk water peak. Presumably, a change in pH during the phantom preparation (e.g., boiling for the agarose) may lead to a change in the lineshape of Z-spectra. In the human study, synthesized Z-spectra and APT# images generated with CEST parameters estimated from MRF-SPEM were in good agreement with the experimental measurements in white matter and gray matter regions of the healthy volunteer brain at 3 T. A fast exchange rate of semisolid macromolecular and amide protons was observed in voxels located partly in the cerebral ventricles (cerebrospinal fluid) with long T1 relaxation times. However, previous studies showed that APT signal of the CSF is almost zero and independent of RF saturation power, thus the APT effect can be assumed to be negligible in the ventricles (Paech et al., 2014; Zhao et al., 2011; also see simulation results in Supporting Fig. S1). In addition, there should be no MTC in the ventricles and these APT and MTC effects are only apparent and due to partial volume effects with tissue and perhaps CSF flow-related effects.

We investigated the spatial distribution of the magnitude of MTC and APT effects in the human brain. While the estimated semisolid macromolecular proton concentration was significantly larger in white matter than in gray matter, the exchange rate was significantly faster in gray matter than in white matter. The estimated semisolid macromolecular proton parameters by MRF-SPEM were consistent with previous quantitative MT studies (Mougin et al., 2010; Sled and Pike, 2001; Stanisz et al., 2005). The APT signals were found to be slightly higher in gray matter than in white matter, in line with previous experiments (Jin et al., 2013; Xu et al., 2016), presumably as a result of the higher content of mobile proteins and peptides, or the greater magnetization exchange between protons in the water and macromolecular pools in white matter, leaving fewer water protons for exchange with mobile proteins and peptides. The apparent amide proton exchange rates of the gray matter and white matter estimated from MRF-SPEM were around 162 Hz and 365 Hz, respectively. Our APT quantification values were high compared to the values of the exchange rates and concentration (~ 28 Hz and 72 mM) reported in previous rat animal studies at 4.7 T using a water-exchange spectroscopy (WEX) approach (Mori et al., 1997; Zhou et al., 2003b; Zhou et al., 2004). One possible reason for this lack of consensus is that the amide protons have a wide range of exchange rates, while the model used here assumed that the amide protons had a single exchange rate. Amide protons in proteins and peptides are known to exhibit a large range of exchange rates at a physiologically relevant pH range (Wuthrich, 1986). The WEX approach may be less suitable for measuring very fast exchange rates due to the finite duration (several ms) of the water labeling preparation period. In the previous study, furthermore, the amide proton concentration (72 mM) was retrospectively calculated from a simplified two-pool exchange model-based APT ratio equation with the exchange rate of 28 Hz measured from WEX experiment. Another less likely possibility is that multiple CEST components can influence the signal at the amide proton frequency offset. At 3T, even with the low RF saturation power pulse, there may still be some contamination from broad coalesced signals of amine protons from glutamate (Cai et al., 2012; Zhang et al., 2018), guanidinium protons from creatine and mobile protein/peptide (Haris et al., 2012; Zhang et al., 2017c), and hydroxyl protons from myo-inositol (Haris et al., 2011), glycogen (van Zijl et al., 2007), and glycosaminoglycans (Ling et al., 2008) over the amide proton frequency. Note that a previous human study at 7T (Liu et al., 2013) measured the amide proton exchange rate of ~280 Hz, which is comparable to that measured here, while the concentration was somewhat lower than our estimation. A human study using the frequency labeled exchange approach measured a rate on the order of 350–400 Hz (Yadav et al., 2013). Another human study (Geades et al., 2017) measured the amide proton concentration of 220 mM (0.2% relative to water, 110M), which is consistent with that measured in this work.

Finding a unique (or sparse) solution in an ill-posed inverse problem is always challenging, particularly when multiple components for CEST MRI are considered. To overcome this issue, we shrank the exchange model and reduced fitting parameters by subgrouping proton exchange models. Additional information/constraints, such as different RF saturation powers, saturation times, and frequency offsets were added to find a unique solution (CEST parameters). Nevertheless, the coupled effect from the exchange rate and concentration was often observed, particularly as shown in kmw and M0m maps (Fig. 9a). The high kmw (bright) and low M0m (dark) values were shown in the anterior and posterior edges whereas the T1wobs, APT# (Fig. 9b) and kmw·M0m images (see Supporting Fig. S2) showed structures. These CEST-related parameters appear multiplied together in the Bloch equations and thus, it may be difficult to completely disentangle the coupled parameters. In future work, more MTC image acquisitions with varied RF saturation parameters may be required. In addition, the application of a varied inter-pulse delay time (Xu et al., 2014a) may be added to limit CEST signals only to those from biomolecules with a specific range of proton exchange rate to MRF-SPEM. This may yield more reliable CEST parameter mapping, especially when incorporating fast exchangeable amine and guanidinium proton pools in the analysis. Upfield NOE-related proton pools can be easily incorporated into MRF-SPEM because NOE signals are easily discriminated from downfield CEST signals due to the difference in saturation frequency offsets. Nevertheless, including additional proton pools in the model inevitably results in increasing acquisition images and scan time. Currently, an acquisition schedule with 32 image acquisitions was chosen, which is the similar acquisition number used in previous works (Cohen et al., 2018; Zhou et al., 2018). However, future studies are needed to optimize acquisition schedules for reducing acquisition times and maximizing specificity of CEST components.

In the present study, MRF-SPEM data were reconstructed by solving a nonlinear least squares problem, rather than a dictionary-match approach used in the original MRF approach. The advantage of this approach is that an exhaustive brute-force dictionary generation for high precision and search can be avoided, enabling straightforward extensions of the proton exchange model to incorporate multiple CEST parameters without a new dictionary construction. In addition, the use of highly undersampled k-space and short TRs for transient-state signal acquisition, as used in the original MRF may be limited in CEST MRI due to inherently low SNR. Currently, dictionary-free or undersampled/fully sampled Cartesian MRF methods are being explored (Anderson et al., 2018; Buonincontri and Sawiak, 2016; Cohen et al., 2018; Sbrizzi et al., 2017; Zhang et al., 2017b; Zhou et al., 2018). In MRF-SPEM reconstruction, CEST parameter mapping with two-dimensional data (2562) took about six hours. Furthermore, the three-pool model might not be enough for in-vivo. Additional CEST pools could be incorporated in MRF-SPEM reconstruction, but it inevitably leads to longer reconstruction times. Recently, much effort has been focused on the compression of the dictionary computation and acceleration of the dictionary matching (Asslander et al., 2018; Cao et al., 2017; Cauley et al., 2015; Cline et al., 2017; McGivney et al., 2014). Such an accelerated dictionary-based, pattern-matching approach could benefit fast three-dimensional MRF-SPEM reconstruction due to the inherent robustness of the dictionary matching to aliasing artifacts.

5. Conclusion

A fast quantitative CEST imaging technique based on MRF-SPEM was developed, validated in numerical phantoms, and demonstrated in-vivo using a synthetic CEST analysis. This quantitative approach could provide significant insights into the origin of the conventional CEST-weighted image contrast in normal human brains and in many pathologies, such as cancer, stroke, and various psychiatric and neurodegenerative diseases.

Supplementary Material

1
  1. A new MR fingerprinting concept was proposed to allow CEST quantification.

  2. A varied RF saturation was designed to generate CEST signal evolutions.

  3. Synthetic CEST MRI was used for validation of in-vivo CEST quantification.

  4. The MRF-SPEM technique can provide rapid and quantitative human brain CEST mapping.

Acknowledgment

The authors thank Drs. Nirbhay Yadav and Yi Zhang for helpful discussions and Dr. Guanshu Liu for the support in the construction of the phantom. This work was supported in part by grants from the National Institutes of Health (R01EB009731, R01CA166171, R01CA228188, R01EB015032, R21CA227783, UG3NS106931, and P41EB015909).

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

Data and Code Availability Statement

Data from this study are available on request from the corresponding author.

References

  1. Aime S, Calabi L, Biondi L, De Miranda M, Ghelli S, Paleari L, Rebaudengo C, Terreno E, 2005. Iopamidol: Exploring the potential use of a well-established x-ray contrast agent for MRI. Magn Reson Med 53, 830–834. [DOI] [PubMed] [Google Scholar]
  2. Anderson CE, Wang CY, Gu Y, Darrah R, Griswold MA, Yu X, Flask CA, 2018. Regularly incremented phase encoding - MR fingerprinting (RIPE-MRF) for enhanced motion artifact suppression in preclinical cartesian MR fingerprinting. Magn Reson Med 79, 2176–2182. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Asslander J, Cloos MA, Knoll F, Sodickson DK, Hennig J, Lattanzi R, 2018. Low rank alternating direction method of multipliers reconstruction for MR fingerprinting. Magn Reson Med 79, 83–96. [DOI] [PMC free article] [PubMed] [Google Scholar]
  4. Buonincontri G, Sawiak SJ, 2016. MR fingerprinting with simultaneous B1 estimation. Magn Reson Med 76, 1127–1135. [DOI] [PMC free article] [PubMed] [Google Scholar]
  5. Cai K, Singh A, Poptani H, Li W, Yang S, Lu Y, Hariharan H, Zhou XJ, Reddy R, 2015. CEST signal at 2ppm (CEST@2ppm) from Z-spectral fitting correlates with creatine distribution in brain tumor. NMR Biomed 28, 1–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Cai KJ, Haris M, Singh A, Kogan F, Greenberg JH, Hariharan H, Detre JA, Reddy R, 2012. Magnetic resonance imaging of glutamate. Nat Med 18, 302–306. [DOI] [PMC free article] [PubMed] [Google Scholar]
  7. Cao X, Liao C, Wang Z, Chen Y, Ye H, He H, Zhong J, 2017. Robust sliding-window reconstruction for Accelerating the acquisition of MR fingerprinting. Magn Reson Med 78, 1579–1588. [DOI] [PubMed] [Google Scholar]
  8. Cauley SF, Setsompop K, Ma D, Jiang Y, Ye H, Adalsteinsson E, Griswold MA, Wald LL, 2015. Fast group matching for MR fingerprinting reconstruction. Magn Reson Med 74, 523–528. [DOI] [PMC free article] [PubMed] [Google Scholar]
  9. Cline CC, Chen X, Mailhe B, Wang Q, Pfeuffer J, Nittka M, Griswold MA, Speier P, Nadar MS, 2017. AIR-MRF: Accelerated iterative reconstruction for magnetic resonance fingerprinting. Magn Reson Imaging 41, 29–40. [DOI] [PubMed] [Google Scholar]
  10. Cohen O, Huang S, McMahon MT, Rosen MS, Farrar CT, 2018. Rapid and quantitative chemical exchange saturation transfer (CEST) imaging with magnetic resonance fingerprinting (MRF). Magn Reson Med, 10.1002/mrm.27221. [DOI] [PMC free article] [PubMed] [Google Scholar]
  11. Desmond KL, Stanisz GJ, 2012. Understanding quantitative pulsed CEST in the presence of MT. Magn Reson Med 67, 979–990. [DOI] [PubMed] [Google Scholar]
  12. Geades N, Hunt BAE, Shah SM, Peters A, Mougin OE, Gowland PA, 2017. Quantitative analysis of the z-spectrum using a numerically simulated look-up table: Application to the healthy human brain at 7T. Magn Reson Med 78, 645–655. [DOI] [PubMed] [Google Scholar]
  13. Goffeney N, Bulte JWM, Duyn J, Bryant LH, van Zijl PCM, 2001. Sensitive NMR detection of cationic-polymer-based gene delivery systems using saturation transfer via proton exchange. J. Am. Chem. Soc 123, 8628–8629. [DOI] [PubMed] [Google Scholar]
  14. Haris M, Cai K, Singh A, Hariharan H, Reddy R, 2011. In vivo mapping of brain myo-inositol. Neuroimage 54, 2079–2085. [DOI] [PMC free article] [PubMed] [Google Scholar]
  15. Haris M, Nanga RP, Singh A, Cai K, Kogan F, Hariharan H, Reddy R, 2012. Exchange rates of creatine kinase metabolites: feasibility of imaging creatine by chemical exchange saturation transfer MRI. NMR Biomed 25, 1305–1309. [DOI] [PMC free article] [PubMed] [Google Scholar]
  16. Haris M, Singh A, Cai K, Kogan F, McGarvey J, Debrosse C, Zsido GA, Witschey WR, Koomalsingh K, Pilla JJ, Chirinos JA, Ferrari VA, Gorman JH, Hariharan H, Gorman RC, Reddy R, 2014. A technique for in vivo mapping of myocardial creatine kinase metabolism. Nat Med 20, 209–214. [DOI] [PMC free article] [PubMed] [Google Scholar]
  17. Henkelman RM, Huang X, Xiang QS, Stanisz GJ, Swanson SD, Bronskill MJ, 1993. Quantitative interpretation of magnetization transfer. Magn Reson Med 29, 759–766. [DOI] [PubMed] [Google Scholar]
  18. Heo HY, Jones CK, Hua J, Yadav N, Agarwal S, Zhou J, van Zijl PC, Pillai JJ, 2016a. Whole-brain amide proton transfer (APT) and nuclear overhauser enhancement (NOE) imaging in glioma patients using low-power steady-state pulsed chemical exchange saturation transfer (CEST) imaging at 7T. J Magn Reson Imaging 44, 41–50. [DOI] [PMC free article] [PubMed] [Google Scholar]
  19. Heo HY, Lee DH, Zhang Y, Zhao X, Jiang S, Chen M, Zhou J, 2017a. Insight into the quantitative metrics of chemical exchange saturation transfer (CEST) imaging. Magn Reson Med 77, 1853–1865. [DOI] [PMC free article] [PubMed] [Google Scholar]
  20. Heo HY, Zhang Y, Burton TM, Jiang S, Zhao Y, van Zijl PCM, Leigh R, Zhou J, 2017b. Improving the detection sensitivity of pH-weighted amide proton transfer MRI in acute stroke patients using extrapolated semisolid magnetization transfer reference signals. Magn Reson Med 78, 871–880. [DOI] [PMC free article] [PubMed] [Google Scholar]
  21. Heo HY, Zhang Y, Jiang S, Lee DH, Zhou J, 2016b. Quantitative assessment of amide proton transfer (APT) and nuclear overhauser enhancement (NOE) imaging with extrapolated semisolid magnetization transfer reference (EMR) signals: II. Comparison of three EMR models and application to human brain glioma at 3 Tesla. Magn Reson Med 75, 1630–1639. [DOI] [PMC free article] [PubMed] [Google Scholar]
  22. Heo HY, Zhang Y, Jiang S, Zhou J, 2018. Influences of experimental parameters on chemical exchange saturation transfer (CEST) metrics of brain tumors using animal models at 4.7T. Magn Reson Med, 10.1002/mrm.27389. [DOI] [PMC free article] [PubMed] [Google Scholar]
  23. Heo HY, Zhang Y, Lee DH, Hong X, Zhou J, 2016c. Quantitative assessment of amide proton transfer (APT) and nuclear overhauser enhancement (NOE) imaging with extrapolated semi-solid magnetization transfer reference (EMR) signals: Application to a rat glioma model at 4.7 tesla. Magn Reson Med 75, 137–149. [DOI] [PMC free article] [PubMed] [Google Scholar]
  24. Heo HY, Zhang Y, Lee DH, Jiang S, Zhao X, Zhou J, 2017c. Accelerating chemical exchange saturation transfer (CEST) MRI by combining compressed sensing and sensitivity encoding techniques. Magn Reson Med 77, 779–786. [DOI] [PMC free article] [PubMed] [Google Scholar]
  25. Hua J, Jones CK, Blakeley J, Smith SA, van Zijl PC, Zhou J, 2007. Quantitative description of the asymmetry in magnetization transfer effects around the water resonance in the human brain. Magn Reson Med 58, 786–793. [DOI] [PMC free article] [PubMed] [Google Scholar]
  26. Jiang S, Eberhart CG, Zhang Y, Heo H-Y, Wen Z, Blair L, Qin H, Lim M, Quinones-Hinojosa A, Weingart JD, Barker PB, Pomper MG, Laterra J, van Zijl PCM, Blakeley JO, Zhou J, 2017. Amide proton transfer-weighted MR image-guided stereotactic biopsy in patients with newly diagnosed gliomas. Eur. J. Cancer 83, 9–18. [DOI] [PMC free article] [PubMed] [Google Scholar]
  27. Jin T, Wang P, Zong X, Kim SG, 2013. MR imaging of the amide-proton transfer effect and the pH-insensitive nuclear overhauser effect at 9.4 T. Magn Reson Med 69, 760–770. [DOI] [PMC free article] [PubMed] [Google Scholar]
  28. Jin T, Wang P, Zong XP, Kim SG, 2012. Magnetic resonance imaging of the Amine-Proton EXchange (APEX) dependent contrast. Neuroimage 59, 1218–1227. [DOI] [PMC free article] [PubMed] [Google Scholar]
  29. Jones CK, Huang A, Xu J, Edden RA, Schar M, Hua J, Oskolkov N, Zaca D, Zhou J, McMahon MT, Pillai JJ, van Zijl PC, 2013. Nuclear Overhauser enhancement (NOE) imaging in the human brain at 7T. Neuroimage 77, 114–124. [DOI] [PMC free article] [PubMed] [Google Scholar]
  30. Jones KM, Randtke EA, Yoshimaru ES, Howison CM, Chalasani P, Klein RR, Chambers SK, Kuo PH, Pagel MD, 2017. Clinical Translation of Tumor Acidosis Measurements with AcidoCEST MRI. Mol Imaging Biol 19, 617–625. [DOI] [PMC free article] [PubMed] [Google Scholar]
  31. Kim M, Gillen J, Landman BA, Zhou J, van Zijl PC, 2009. Water saturation shift referencing (WASSR) for chemical exchange saturation transfer (CEST) experiments. Magn Reson Med 61, 1441–1450. [DOI] [PMC free article] [PubMed] [Google Scholar]
  32. Lee DH, Heo HY, Zhang K, Zhang Y, Jiang S, Zhao X, Zhou J, 2017. Quantitative assessment of the effects of water proton concentration and water T1 changes on amide proton transfer (APT) and nuclear overhauser enhancement (NOE) MRI: The origin of the APT imaging signal in brain tumor. Magn Reson Med 77, 855–863. [DOI] [PMC free article] [PubMed] [Google Scholar]
  33. Li C, Peng S, Wang R, Chen H, Su W, Zhao X, Zhou J, Chen M, 2014. Chemical exchange saturation transfer MR imaging of Parkinson’s disease at 3 Tesla. Eur. Radiol 24, 2631–2639. [DOI] [PMC free article] [PubMed] [Google Scholar]
  34. Liao C, Bilgic B, Manhard MK, Zhao B, Cao X, Zhong J, Wald LL, Setsompop K, 2017. 3D MR fingerprinting with accelerated stack-of-spirals and hybrid sliding-window and GRAPPA reconstruction. Neuroimage 162, 13–22. [DOI] [PMC free article] [PubMed] [Google Scholar]
  35. Ling W, Regatte RR, Navon G, Jerschow A, 2008. Assessment of glycosaminoglycan concentration in vivo by chemical exchange-dependent saturation transfer (gagCEST). Proc. Natl. Acad. Sci. (USA) 105, 2266–2270. [DOI] [PMC free article] [PubMed] [Google Scholar]
  36. Liu D, Zhou J, Xue R, Zuo Z, An J, Wang DJJ, 2013. Quantitative characterization of nuclear Overhauser enhancement and amide proton transfer effects in the human brain at 7 Tesla. Magn. Reson. Med 70, 1070–1081. [DOI] [PMC free article] [PubMed] [Google Scholar]
  37. Ma D, Gulani V, Seiberlich N, Liu K, Sunshine JL, Duerk JL, Griswold MA, 2013. Magnetic resonance fingerprinting. Nature 495, 187–192. [DOI] [PMC free article] [PubMed] [Google Scholar]
  38. Ma D, Jiang Y, Chen Y, McGivney D, Mehta B, Gulani V, Griswold M, 2017. Fast 3D magnetic resonance fingerprinting for a whole-brain coverage. Magn Reson Med, 10.1002/mrm.26886. [DOI] [PMC free article] [PubMed] [Google Scholar]
  39. McGivney DF, Pierre E, Ma D, Jiang Y, Saybasili H, Gulani V, Griswold MA, 2014. SVD compression for magnetic resonance fingerprinting in the time domain. IEEE Trans Med Imaging 33, 2311–2322. [DOI] [PMC free article] [PubMed] [Google Scholar]
  40. McMahon MT, Gilad AA, Zhou J, Sun PZ, Bulte JWM, van Zijl PCM, 2006. Quantifying exchange rates in chemical exchange saturation transfer agents using the saturation time and saturation power dependencies of the magnetization transfer effect on the magnetic resonance imaging signal (QUEST and QUESP): pH calibration for poly-L-lysine and a starburst dendrimer. Magn. Reson. Med 55, 836–847. [DOI] [PMC free article] [PubMed] [Google Scholar]
  41. Mori S, Abeygunawardana C, Berg JM, van Zijl PCM, 1997. NMR study of rapidly exchanging backbone amide protons in staphylococcal nuclease and the correlation with structural and dynamic properties. J. Am. Chem. Soc 119, 6844–6852. [Google Scholar]
  42. Mougin OE, Coxon RC, Pitiot A, Gowland PA, 2010. Magnetization transfer phenomenon in the human brain at 7T. Neuroimage 49, 272–281. [DOI] [PubMed] [Google Scholar]
  43. Paech D, Zaiss M, Meissner JE, Windschuh J, Wiestler B, Bachert P, Neumann JO, Kickingereder P, Schlemmer HP, Wick W, Nagel AM, Heiland S, Ladd ME, Bendszus M, Radbruch A, 2014. Nuclear overhauser enhancement mediated chemical exchange saturation transfer imaging at 7 Tesla in glioblastoma patients. PLoS One 9, e104181. [DOI] [PMC free article] [PubMed] [Google Scholar]
  44. Sbrizzi A, Bruijn T, van der Heide O, Luijten P, van den Berg CAT, 2017. Dictionary-free MR Fingerprinting reconstruction of balanced-GRE sequences [Google Scholar]
  45. Sled JG, Pike GB, 2001. Quantitative imaging of magnetization transfer exchange and relaxation properties in vivo using MRI. Magn. Reson. Med 46, 923–931. [DOI] [PubMed] [Google Scholar]
  46. Stanisz GJ, Odrobina EE, Pun J, Escaravage M, Graham SJ, Bronskill MJ, Henkelman RM, 2005. T-1, T-2 relaxation and magnetization transfer in tissue at 3T. Magn. Reson. Med 54, 507–512. [DOI] [PubMed] [Google Scholar]
  47. Su P, Mao D, Liu P, Li Y, Pinho MC, Welch BG, Lu H, 2017. Multiparametric estimation of brain hemodynamics with MR fingerprinting ASL. Magn Reson Med 78, 1812–1823. [DOI] [PMC free article] [PubMed] [Google Scholar]
  48. Sun PZ, Lu J, Wu Y, Xiao G, Wu R, 2013. Evaluation of the dependence of CEST-EPI measurement on repetition time, RF irradiation duty cycle and imaging flip angle for enhanced pH sensitivity. Phys Med Biol 58, N229–240. [DOI] [PMC free article] [PubMed] [Google Scholar]
  49. Sun PZ, Zhou J, Sun W, Huang J, van Zijl PCM, 2007. Detection of the ischemic penumbra using pH-weighted MRI. J. Cereb. Blood Flow Metab 27, 1129–1136. [DOI] [PubMed] [Google Scholar]
  50. van Zijl PCM, Jones CK, Ren J, Malloy CR, Sherry AD, 2007. MRI detection of glycogen in vivo by using chemical exchange saturation transfer imaging (glycoCEST). Proc. Natl. Acad. Sci. (USA) 104, 4359–4364. [DOI] [PMC free article] [PubMed] [Google Scholar]
  51. van Zijl PCM, Lam WW, Xu J, Knutsson L, Stanisz GJ, 2018. Magnetization Transfer Contrast and Chemical Exchange Saturation Transfer MRI. Features and analysis of the field-dependent saturation spectrum. Neuroimage 168, 222–241. [DOI] [PMC free article] [PubMed] [Google Scholar]
  52. van Zijl PCM, Yadav NN, 2011. Chemical exchange saturation transfer (CEST): What is in a name and what isn’t? Magn. Reson. Med 65, 927–948. [DOI] [PMC free article] [PubMed] [Google Scholar]
  53. van Zijl PCM, Zhou J, Mori N, Payen JF, Wilson D, Mori S, 2003. Mechanism of magnetization transfer during on-resonance water saturation. A new approach to detect mobile proteins, peptides, and lipids. Magn Reson Med 49, 440–449. [DOI] [PubMed] [Google Scholar]
  54. Walker-Samuel S, Ramasawmy R, Torrealdea F, Rega M, Rajkumar V, Johnson SP, Richardson S, Goncalves M, Parkes HG, Arstad E, Thomas DL, Pedley RB, Lythgoe MF, Golay X, 2013. In vivo imaging of glucose uptake and metabolism in tumors. Nat Med 19, 1067–1072. [DOI] [PMC free article] [PubMed] [Google Scholar]
  55. Wang CY, Liu Y, Huang S, Griswold MA, Seiberlich N, Yu X, 2017. (31) P magnetic resonance fingerprinting for rapid quantification of creatine kinase reaction rate in vivo. NMR Biomed 30, e3786. [DOI] [PMC free article] [PubMed] [Google Scholar]
  56. Ward KM, Aletras AH, Balaban RS, 2000. A new class of contrast agents for MRI based on proton chemical exchange dependent saturation transfer (CEST). J. Magn. Reson 143, 79–87. [DOI] [PubMed] [Google Scholar]
  57. Woessner DE, Zhang S, Merritt ME, Sherry AD, 2005. Numerical solution of the Bloch equations provides insights into the optimum design of PARACEST agents for MRI. Magn. Reson. Med 53, 790–799. [DOI] [PubMed] [Google Scholar]
  58. Wuthrich K, 1986. NMR of proteins and nucleic acids, 2 ed John Wiley & Sons, New York. [Google Scholar]
  59. Xu J, Yadav NN, Bar-Shir A, Jones CK, Chan KW, Zhang J, Walczak P, McMahon MT, van Zijl PCM, 2014a. Variable delay multi-pulse train for fast chemical exchange saturation transfer and relayed-nuclear overhauser enhancement MRI. Magn Reson Med 71, 1798–1812. [DOI] [PMC free article] [PubMed] [Google Scholar]
  60. Xu J, Zaiss M, Zu Z, Li H, Xie J, Gochberg DF, Bachert P, Gore JC, 2014b. On the origins of chemical exchange saturation transfer (CEST) contrast in tumors at 9.4 T. NMR Biomed 27, 406–416. [DOI] [PMC free article] [PubMed] [Google Scholar]
  61. Xu X, Yadav NN, Zeng H, Jones CK, Zhou J, van Zijl PCM, Xu J, 2016. Magnetization transfer contrast-suppressed imaging of amide proton transfer and relayed nuclear overhauser enhancement chemical exchange saturation transfer effects in the human brain at 7T. Magn Reson Med 75, 88–96. [DOI] [PMC free article] [PubMed] [Google Scholar]
  62. Yadav NN, Jones CK, Hua J, Xu J, van Zijl PCM, 2013. Imaging of endogenous exchangeable proton signals in the human brain using frequency labeled exchange transfer imaging. Magn Reson Med 69, 966–973. [DOI] [PMC free article] [PubMed] [Google Scholar]
  63. Zaiss M, Schmitt B, Bachert P, 2011. Quantitative separation of CEST effect from magnetization transfer and spillover effects by Lorentzian-line-fit analysis of z-spectra. J. Magn. Reson 211, 149–155. [DOI] [PubMed] [Google Scholar]
  64. Zhang H, Kang H, Zhao X, Jiang S, Zhang Y, Zhou J, Peng Y, 2016a. Amide proton transfer (APT) MR imaging and magnetization transfer (MT) MR imaging of pediatric brain development. Eur. Radiol 26, 3368–3376. [DOI] [PMC free article] [PubMed] [Google Scholar]
  65. Zhang H, Wang W, Jiang S, Zhang Y, Heo HY, Wang X, Peng Y, Wang J, Zhou J, 2017a. Amide proton transfer-weighted MRI detection of traumatic brain injury in rats. J. Cereb. Blood Flow Metab 37, 3422–3432. [DOI] [PMC free article] [PubMed] [Google Scholar]
  66. Zhang S, Winter P, Wu K, Sherry AD, 2001. A novel europium(III)-based MRI contrast agent. J. Am. Chem. Soc 123, 1517–1578. [DOI] [PubMed] [Google Scholar]
  67. Zhang SR, Malloy CR, Sherry AD, 2005. MRI thermometry based on PARACEST agents. J. Am. Chem. Soc 127, 17572–17573. [DOI] [PMC free article] [PubMed] [Google Scholar]
  68. Zhang X, Zhou Z, Chen S, Li R, Hu X, 2017b. MR fingerprinting reconstruction with Kalman filter. Magn Reson Imaging 41, 53–62. [DOI] [PubMed] [Google Scholar]
  69. Zhang XY, Wang F, Li H, Xu J, Gochberg DF, Gore JC, Zu Z, 2017c. Accuracy in the quantification of chemical exchange saturation transfer (CEST) and relayed nuclear Overhauser enhancement (rNOE) saturation transfer effects. NMR Biomed 30, e3716. [DOI] [PMC free article] [PubMed] [Google Scholar]
  70. Zhang XY, Wang F, Xu J, Gochberg DF, Gore JC, Zu Z, 2018. Increased CEST specificity for amide and fast-exchanging amine protons using exchange-dependent relaxation rate. NMR Biomed 31, e3863. [DOI] [PMC free article] [PubMed] [Google Scholar]
  71. Zhang XY, Xie J, Wang F, Lin EC, Xu J, Gochberg DF, Gore JC, Zu Z, 2017d. Assignment of the molecular origins of CEST signals at 2 ppm in rat brain. Magn Reson Med 78, 881–887. [DOI] [PMC free article] [PubMed] [Google Scholar]
  72. Zhang Y, Heo HY, Lee DH, Zhao X, Jiang S, Zhang K, Li H, Zhou J, 2016b. Selecting the reference image for registration of CEST series. J Magn Reson Imaging 43, 756–761. [DOI] [PMC free article] [PubMed] [Google Scholar]
  73. Zhao X, Wen Z, Huang F, Lu S, Wang X, Hu S, Zu D, Zhou J, 2011. Saturation power dependence of amide proton transfer image contrasts in human brain tumors and strokes at 3 T. Magn. Reson. Med 66, 1033–1041. [DOI] [PMC free article] [PubMed] [Google Scholar]
  74. Zhou J, Blakeley JO, Hua J, Kim M, Laterra J, Pomper MG, van Zijl PCM, 2008. Practical data acquisition method for human brain tumor amide proton transfer (APT) imaging. Magn. Reson. Med 60, 842–849. [DOI] [PMC free article] [PubMed] [Google Scholar]
  75. Zhou J, Hong X, Zhao X, Gao JH, Yuan J, 2013a. APT-weighted and NOE-weighted image contrasts in glioma with different RF saturation powers based on magnetization transfer ratio asymmetry analyses. Magn Reson Med 70, 320–327. [DOI] [PMC free article] [PubMed] [Google Scholar]
  76. Zhou J, Lal B, Wilson DA, Laterra J, van Zijl PCM, 2003a. Amide proton transfer (APT) contrast for imaging of brain tumors. Magn. Reson. Med 50, 1120–1126. [DOI] [PubMed] [Google Scholar]
  77. Zhou J, Payen J, Wilson DA, Traystman RJ, van Zijl PCM, 2003b. Using the amide proton signals of intracellular proteins and peptides to detect pH effects in MRI. Nature Med 9, 1085–1090. [DOI] [PubMed] [Google Scholar]
  78. Zhou J, van Zijl PCM, 2006. Chemical exchange saturation transfer imaging and spectroscopy. Progr. NMR Spectr 48, 109–136. [Google Scholar]
  79. Zhou J, Wilson DA, Sun PZ, Klaus JA, Van Zijl PCM, 2004. Quantitative description of proton exchange processes between water and endogenous and exogenous agents for WEX, CEST, and APT experiments. Magn Reson Med 51, 945–952. [DOI] [PubMed] [Google Scholar]
  80. Zhou J, Zhu H, Lim M, Blair L, Quinones-Hinojosa A, Messina AA, Eberhart CG, Pomper MG, Laterra J, Barker PB, van Zijl PCM, Blakeley JO, 2013b. Three-dimensional amide proton transfer MR imaging of gliomas: Initial experience and comparison with gadolinium enhancement. J. Magn. Reson. Imaging 38, 1119–1128. [DOI] [PMC free article] [PubMed] [Google Scholar]
  81. Zhou Z, Han P, Zhou B, Christodoulou AG, Shaw JL, Deng Z, Li D, 2018. Chemical exchange saturation transfer fingerprinting for exchange rate quantification. Magn Reson Med 80, 1352–1363. [DOI] [PMC free article] [PubMed] [Google Scholar]