Abstract
Background
Bioprosthetic aortic valve degeneration is increasingly common, often unheralded, and can have catastrophic consequences.
Objectives
The authors sought to assess whether 18F-fluoride positron emission tomography (PET)-computed tomography (CT) can detect bioprosthetic aortic valve degeneration and predict valve dysfunction.
Methods
Explanted degenerate bioprosthetic valves were examined ex vivo. Patients with bioprosthetic aortic valves were recruited into 2 cohorts with and without prosthetic valve dysfunction and underwent in vivo contrast-enhanced CT angiography, 18F-fluoride PET, and serial echocardiography during 2 years of follow-up.
Results
All ex vivo, degenerate bioprosthetic valves displayed 18F-fluoride PET uptake that colocalized with tissue degeneration on histology. In 71 patients without known bioprosthesis dysfunction, 14 had abnormal leaflet pathology on CT, and 24 demonstrated 18F-fluoride PET uptake (target-to-background ratio 1.55 [interquartile range (IQR): 1.44 to 1.88]). Patients with increased 18F-fluoride uptake exhibited more rapid deterioration in valve function compared with those without (annualized change in peak transvalvular velocity 0.30 [IQR: 0.13 to 0.61] vs. 0.01 [IQR: −0.05 to 0.16] ms−1/year; p < 0.001). Indeed 18F-fluoride uptake correlated with deterioration in all the conventional echocardiographic measures of valve function assessed (e.g., change in peak velocity, r = 0.72; p < 0.001). Each of the 10 patients who developed new overt bioprosthesis dysfunction during follow-up had evidence of 18F-fluoride uptake at baseline (target-to-background ratio 1.89 [IQR: 1.46 to 2.59]). On multivariable analysis, 18F-fluoride uptake was the only independent predictor of future bioprosthetic dysfunction.
Conclusions
18F-fluoride PET-CT identifies subclinical bioprosthetic valve degeneration, providing powerful prediction of subsequent valvular dysfunction and highlighting patients at risk of valve failure. This technique holds major promise in the diagnosis of valvular degeneration and the surveillance of patients with bioprosthetic valves. (18F-Fluoride Assessment of Aortic Bioprosthesis Durability and Outcome [18F-FAABULOUS]; NCT02304276)
Key Words: aortic valve replacement, bioprosthetic valve degeneration, calcification, histology, positron emission tomography
Abbreviations and Acronyms: CT, computed tomography; PET, positron emission tomography; TBR, target-to-background ratio
Central Illustration
The implantation of bioprosthetic heart valves is increasing rapidly due to patient preference, the increasing prevalence of valve disease in an aging population, and the emergence of transcatheter aortic valve implantation 1, 2, 3, 4. In the United States, 90,000 surgical aortic valve replacements are performed annually, with over three-quarters incorporating bioprosthetic valves (3). In addition, >80,000 transcatheter aortic valve implantation procedures have been performed since U.S. Food and Drug Administration approval in 2011 (4). Given the finite lifespan of these valves and the rapid expansion of the population of patients requiring regular surveillance, bioprosthetic valve degeneration will become a major cause of cardiovascular morbidity and health care burden over the coming decades.
The pathophysiology of bioprosthetic valve degeneration is poorly understood. Although calcification appears to contribute to both progressive valve narrowing and leaflet tears 5, 6, 7, noninvasive methods for detecting this process have been lacking, and the triggers of valve degeneration and calcification are unknown. Current standard of care relies on serial echocardiography and clinical assessment aimed at detecting the valve dysfunction that occurs only toward the end stages of the degeneration process. Unfortunately, many patients present in extremis with unheralded valve failure due to rapid-onset valvular obstruction or regurgitation, with repeat operation a high-risk undertaking. Indeed, emergency repeat aortic valve replacement surgery is associated with a mortality of 22.6% compared with 1.4% for elective repeat surgery (8). Detection of bioprosthetic valve degeneration is therefore highly desirable, allowing at-risk patients to be identified early, offered close tailored monitoring, and optimized timing of repeat elective intervention, thereby avoiding potentially catastrophic valve failure.
18F-fluoride positron emission tomography (PET) has recently been used to image tissue calcification activity in a range of cardiovascular diseases 9, 10, 11, 12. 18F-fluoride preferentially binds to areas of developing microcalcification indicative of tissue degeneration (13) that precede the macrocalcification detectable by computed tomography (CT) 14, 15. Given that calcification is one of the key pathological processes underlying bioprosthetic valve degeneration, we hypothesized that increased 18F-fluoride uptake would identify prosthetic valve degeneration and predict subsequent deterioration in bioprosthetic valve function.
Methods
Ex vivo assessment of degenerated bioprosthetic aortic valves
Explanted degenerated aortic valve bioprostheses were obtained with written consent from patients undergoing repeat surgical aortic valve replacement for bioprosthetic valve failure. Valves were weighed, photographed, and their macroscopic features documented. Ex vivo micro–PET-CT was performed with a nano–PET-CT scanner (Mediso, Budapest, Hungary) and x-ray microtomograph (Bruker, Kontich, Belgium) before undergoing histological evaluation (Online Appendix).
Clinical study design and population
Patients over 40 years of age who had undergone previous surgical aortic valve replacement using a bioprosthetic valve were prospectively recruited into a single-center cohort study according to defined inclusion and exclusion criteria (Online Figure 1). Participants were recruited if they were under routine clinical review and had: 1) known evidence of bioprosthetic valve failure on echocardiography and had been referred for repeat aortic valve intervention 16, 17; or 2) no known evidence of valve dysfunction or degeneration. Written informed consent was obtained from all participants. The study (NCT02304276) was conducted in accordance with the Declaration of Helsinki and was approved by an NHS Scotland Research Ethics Committee (14/SS/1049) and the Administration of Radioactive Substances Advisory Committee.
Study assessments and data collection
All participants underwent baseline clinical assessment including Doppler and 2-dimensional echocardiography, noncontrast CT calcium scoring, contrast-enhanced CT angiography, and in vivo 18F-fluoride PET (Online Appendix, Online Table 1). Participants were invited to return annually for 2 years for repeat clinical assessments and echocardiography to assess changes in bioprosthesis performance. Definitions of bioprosthetic degeneration and dysfunction vary, but in this study, we used definitions from both contemporary international guideline criteria 16, 17 and a recent expert consensus statement (18) (Online Appendix). Bioprosthetic valve failure was defined as the development of severe hemodynamic valve dysfunction combined with patient symptoms, the need for redo valve intervention, or valve-related death (19). Patients were followed up for the development of valve failure beyond the 2-year timeframe using electronic patient record data.
PET and CT image analysis
Fused PET and contrast-enhanced CT images were reconstructed in diastole, coregistered and reoriented to provide en face images of the bioprosthetic valve with corresponding long-axis views (Online Figure 2). Images were analyzed by 2 experienced PET-CT observers (T.R.G.C. and M.R.D.). Using pre-specified criteria, CT scans were adjudicated to be abnormal if there was pannus (circumferential low-attenuation [noncalcific] material with radial thickness ≥2 mm and encroachment on the valve cusps) 16, 20, noncalcific leaflet thickening (focal areas of low attenuation [30 to 200 Hounsfield Units (HU)] cusp thickening ≥2 mm visualized in at least 2 planes) 20, 21, or leaflet calcification (calcium >500 HU localized to valve cusp in ≤2 planes) (22). Leaflet calcification was further subdivided into spotty calcification if maximum diameter was <3 mm and large calcification if maximum diameter was ≥3 mm (23). 18F-fluoride uptake was quantified primarily using the most-diseased-segment mean target-to-background ratio (TBR) method in keeping with prior studies of the native valve, although other measures were also reported (24). PET scans were adjudicated to be abnormal if increased 18F-fluoride uptake (threshold TBR >1.3) (11) was observed originating from the valve cusps on both en face and long-axis views (Online Appendix).
Statistics and data analysis
Baseline characteristics are reported as number (percentages) for categorical variables and mean ± SD or median (interquartile range) for continuous variables depending on whether variables were normally distributed. Categorical data were compared using chi-square or Fisher exact tests. Continuous variables were log-transformed [logN(x+1) for positive values and −logN(−x+1) for negative values] where not normally distributed. The point-biserial correlation coefficient was used to measure the strength and direction of the association between 1 dichotomous variable and 1 continuous variable. One-way analysis of variance was used to compare continuous data across multiple factors, with post hoc analysis using the Bonferroni test where appropriate. The Student’s t-test or Mann-Whitney U test were used to compare continuous outcomes between 2 independent groups depending on whether they were normally distributed. The Student’s t-test or Wilcoxon signed rank test were used to compare paired variables. Two-tailed Pearson’s correlation analysis was performed to investigate the relationship between 18F-fluoride uptake and echocardiographic measures of valve function. Multivariable analysis was performed to assess the predictors of deterioration in bioprosthetic valve dysfunction (annualized change in peak velocity after 2 years). Statistical analysis was undertaken using IBM SPSS Statistics version 23 software (IBM, Armonk, New York), and significance was taken at the 2-sided 5% level (p < 0.05).
Results
Explanted degenerate bioprosthetic aortic valves
Fifteen failed explanted bioprosthetic aortic valves were obtained (Online Appendix) for ex vivo investigation. Micro-CT detected leaflet calcification in 13 valves, which was confirmed on histology. By contrast, all 15 valves demonstrated 18F-fluoride leaflet uptake that correlated with a range of histological markers of bioprosthetic tissue degeneration. In particular, 18F-fluoride activity detected both micro- and macrocalcific deposits within the valve leaflets that colocalized predominantly with regions of pannus (fibrous thickening) and thrombus formation on histology. However, 18F-fluoride uptake was additionally observed in the absence of calcification on histology at sites of leaflet thickening, fluid insudation, and disrupted collagen architecture (Figure 1).
Clinical cohort study population
Eighty participants were recruited to the clinical cohort study and underwent in vivo PET-CT imaging, although 2 patients were unable to complete the baseline scan (claustrophobia) and were excluded. The remaining 78 patients were 75 ± 7 years of age, with a range of bioprosthetic valve models (Online Tables 2 and 3), a high prevalence of coronary artery disease, and associated risk factors (Table 1).
Table 1.
Patients With Valve Failure | Patients Without Known Valve Degeneration |
|||||
---|---|---|---|---|---|---|
All | 1 Month | 2 Year | 5 Year | 10 Year | ||
Clinical | ||||||
Subjects | 6 | 71 | 9 | 22 | 20 | 20 |
Female | 2 (33.3) | 38 (46.5) | 5 (55.6) | 8 (36.4) | 10 (50.0) | 10 (50.0) |
Age, yrs | 81.3 ± 3.2 | 73.9 ± 7.0 | 73.4 ± 9.0 | 73.0 ± 5.4 | 72.7 ± 5.1 | 76.4 ± 8.9 |
Body mass index, kg/m2 | 27.5 ± 4.5 | 26.9 ± 5.6 | 27.8 ± 4.7 | 27.7 ± 4.8 | 27.8 ± 5.1 | 24.8 ± 6.8 |
Systolic blood pressure, mm Hg | 144.8 ± 18.0 | 153.8 ± 21.9 | 152.4 ± 21.2 | 149.8 ± 22.1 | 158.1 ± 15.2 | 154.5 ± 27.8 |
Diastolic blood pressure, mm Hg | 58.0 ± 8.4 | 79.0 ± 11.0 | 82.0 ± 11.4 | 81.3 ± 12.0 | 78.6 ± 7.4 | 75.4 ± 12.6 |
Heart rate, beats/min | 72.2 ± 17.0 | 71.2 ± 11.8 | 79.8 ± 10.0 | 69.7 ± 11.6 | 73.3 ± 13.4 | 66.6 ± 8.9 |
Medical history | ||||||
Hypertension | 4 (66.7) | 50 (70.4) | 8 (88.9) | 16 (72.7) | 14 (70.0) | 12 (60.0) |
Coronary artery disease | 4 (66.7) | 29 (40.9) | 5 (55.6) | 7 (31.8) | 6 (30.0) | 11 (55.0) |
Coronary bypass surgery | 3 (50.0) | 23 (32.4) | 2 (22.2) | 7 (31.8) | 5 (25.0) | 9 (45.0) |
Diabetes | 1 (16.7) | 4 (5.6) | 0 (0.0) | 0 (0.0) | 2 (10.0) | 2 (10.0) |
Hypercholesterolemia | 5 (83.3) | 55 (77.5) | 8 (88.9) | 18 (81.8) | 13 (65.0) | 16 (80.0) |
Current smoker | 0 (0.0) | 8 (11.4) | 0 (0.0) | 1 (4.6) | 3 (15.0) | 4 (20.0) |
Ex-smoker | 3 (50.0) | 29 (40.8) | 4 (44.4) | 10 (45.5) | 7 (35.0) | 8 (40.0) |
Medication | ||||||
Aspirin | 4 (66.7) | 51 (71.8) | 7 (77.8) | 18 (81.8) | 12 (60.0) | 14 (70.0) |
Clopidogrel | 1 (16.7) | 9 (12.7) | 1 (11.1) | 2 (9.1) | 4 (20.0) | 2 (10.0) |
Warfarin | 2 (33.3) | 4 (5.6) | 0 (0.0) | 1 (4.5) | 3 (15.0) | 0 (0.0) |
Other anticoagulant | 0 (0.0) | 2 (2.8) | 0 (0.0) | 0 (0.0) | 1 (5.0) | 1 (5.0) |
ACEi or ARB | 0 (0.0) | 39 (54.9) | 6 (66.7) | 12 (54.5) | 10 (50.0) | 11 (55.0) |
Beta-blocker | 2 (33.3) | 32 (45.1) | 4 (44.4) | 13 (59.1) | 7 (35.0) | 8 (40.0) |
Statin | 4 (66.7) | 50 (70.4) | 6 (66.7) | 15 (68.2) | 14 (70.0) | 14 (70.0) |
Biochemistry | ||||||
Creatinine, μmol/l | 102.2 ± 37.2 | 84.5 ± 23.4 | 90.3 ± 36.4 | 80.9 ± 17.8 | 89.1 ± 19.8 | 81.3 ± 25.4 |
eGFR, ml/min/1.73 m2 | 48.5 ± 13.0 | 58 ± 7.5 | 53.4 ± 11.8 | 59.8 ± 3.2 | 57.1 ± 6.8 | 59.2 ± 8.5 |
Electrocardiogram | ||||||
Sinus rhythm | 4 (66.7) | 64 (91.4) | 9 (100.0) | 20 (95.2) | 18 (90.0) | 17 (85.0) |
Paced rhythm | 0 (0.0) | 4 (5.7) | 0 (0.0) | 0 (0.0) | 1 (5.0) | 3 (15.0) |
Atrial fibrillation | 2 (33.3) | 2 (2.9) | 0 (0.0) | 1 (4.8) | 1 (5.0) | 0 (0.0) |
LV hypertrophy | 3 (50.0) | 22 (32.4) | 4 (44.4) | 7 (33.3) | 9 (45.0) | 2 (11.1) |
Strain pattern | 3 (50.0) | 14 (20.6) | 4 (44.4) | 2 (9.5) | 6 (30.0) | 2 (11.1) |
Baseline echocardiography | ||||||
LV systolic dysfunction | 4 (66.7) | 11 (15.5) | 1 (11.1) | 3 (13.6) | 4 (20.0) | 3 (15.0) |
Peak valve velocity, m/s | 3.19 (2.84–4.47) | 2.73 (2.38–3.07) | 2.44 (2.27–2.75) | 2.68 (2.28–2.82) | 2.93 (2.55–3.40) | 2.85 (2.36–3.13) |
Mean valve gradient, mm Hg | 20.6 (14.0–39.5) | 15.0 (11.3–19.3) | 13.0 (10.0–15.0) | 13.9 (10.2–17.8) | 18.5 (13.1–23.8) | 17.2 (11.1–20.8) |
Effective orifice area, cm2 | 0.52 (0.52–0.52) | 1.13 (0.94–1.46) | 1.33 (0.95–1.60) | 1.14 (0.95–1.46) | 1.03 (0.92–1.31) | 1.20 (0.86–1.49) |
Dimensionless velocity index | 0.29 (0.20–0.42) | 0.39 (0.33–0.44) | 0.47 (0.38–0.51) | 0.40 (0.36–0.43) | 0.35 (0.31–0.40) | 0.38 (0.31–0.46) |
Prosthetic regurgitation, ≥ moderate | 5 (83.3) | 2 (2.8) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 2 (10.0) |
Acceleration time, ms | * | 80.3 (75.2–87.4) | 87.0 (74.0–93.0) | 80.0 (75.0–85.0) | 79.0 (75.0–86.0) | 85.0 (76.7–94.0) |
Acceleration time/LV ejection time | * | 0.25 (0.24–0.28) | 0.29 (0.25–0.31) | 0.25 (0.24–0.27) | 0.25 (0.23–0.27) | 0.25 (0.22–0.28) |
Computed tomography | ||||||
Abnormal findings | 6 (100.0) | 14 (19.7) | 0 (0.0) | 3 (13.6) | 4 (20.0) | 7 (30.0) |
Spotty calcification | 3 (50.0) | 5 (7.0) | 0 (0.0) | 1 (4.5) | 1 (5.0) | 3 (15.0) |
Large calcification | 3 (50.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) | 0 (0.0) |
Noncalcific leaflet thickening | 2 (33.3) | 5 (7.0) | 0 (0.0) | 1 (4.5) | 3 (15.0) | 1 (5.0) |
Pannus | 1 (16.7) | 7 (9.9) | 0 (0.0) | 2 (9.1) | 0 (0.0) | 5 (25.0) |
18F-fluoride positron emission tomography | ||||||
Leaflet uptake | 6 (100.0) | 24 (33.8) | 1 (11.1) | 6 (27.2) | 8 (40.0) | 9 (45.0) |
SUV MDS max | 4.66 (3.16–7.57) | 1.73 (1.42–2.02) | 1.89 (1.42–1.99) | 1.50 (1.31–1.80) | 1.94 (1.39–2.26) | 1.73 (1.54–2.34) |
SUV MDS mean | 3.50 (2.64–5.90) | 1.48 (1.27–1.84) | 1.59 (1.29–1.82) | 1.31 (1.14–1.57) | 1.72 (1.21–2.05) | 1.44 (1.32–2.04) |
TBR MDS max | 3.61 (2.37–5.82) | 1.31 (1.18–1.70) | 1.26 (1.17–1.35) | 1.26 (1.12–1.47) | 1.52 (1.17–1.82) | 1.47 (1.24–2.12) |
TBR MDS mean | 2.91 (2.10–4.09) | 1.12 (1.04–1.51) | 1.07 (1.06–1.18) | 1.11 (1.00–1.31) | 1.38 (1.01–1.62) | 1.15 (1.06–1.87) |
Values are n, n (%), mean ± SD, or median (interquartile range). *Incomplete data.
ACEi = angiotensin-converting enzyme inhibitor; ARB = angiotensin receptor blockade; CT = computed tomography imaging; eGFR = estimated glomerular filtration rate; LV = left ventricle; MDS = most diseased segment; PET = positron emission tomography; SUV = standardized uptake value; TBR = target-to-background ratio.
Patients with suspected bioprosthetic valve failure
Seven participants were recruited to the cohort with suspected bioprosthetic valve failure. One subject proceeded to repeat surgical aortic valve replacement, 5 underwent valve-in-valve transcatheter valve implantation, and 1 subject was deemed to have severe patient–prosthesis mismatch according to guideline criteria and therefore not considered to have valve failure 16, 17. All 6 subjects with confirmed bioprosthetic valve failure demonstrated abnormalities on both CT and 18F-fluoride PET studies (Figure 2). CT revealed evidence of leaflet calcification in all patients (spotty calcification n = 3, large calcification n = 3) and less commonly noncalcific leaflet thickening (n = 2), circumferential pannus (n = 1), and distorted leaflet morphology (n = 1). Increased 18F-fluoride PET leaflet uptake was observed in the bioprosthetic valves of all these patients, and TBR values were nearly 3 times higher than in patients without known valvular dysfunction (TBR 2.91 [interquartile range (IQR): 1.75 to 4.09] vs. 1.12 [IQR: 1.04 to 1.51]; p < 0.001). In the 1 patient with severe patient–prosthesis mismatch, there was no CT abnormality or 18F-fluoride uptake.
Patients without known bioprosthetic valve dysfunction
In the 71 patients without known bioprosthetic valve dysfunction, surgical aortic valve replacement had been conducted 1 month (n = 9), 2 years (n = 22), 5 years (n = 20), and >10 years (n = 20) previously. The function of these different-aged valves was generally within normal limits at baseline (peak velocity 2.76 ± 0.52 m/s, mean gradient 16.4 ± 7.0 mm Hg) (Table 1).
Contrast-enhanced CT images were not interpretable in 8 patients (11%) due to motion degradation and metallic blooming artefact from the valve struts. Fourteen subjects (19%) had 1 or more leaflet abnormalities on CT: spotty calcification (n = 5), noncalcific leaflet thickening (n = 5), and pannus (n = 7) (Figure 2).
PET scans were interpretable in all patients. Increased 18F-fluoride uptake was more common than abnormalities on CT and was seen in 24 patients (34%) (TBR 1.55 [IQR: 1.44 to 1.88]) (Figure 2, Online Figure 3). Similar to the ex vivo findings, increased 18F-fluoride uptake colocalized with areas of spotty calcification, noncalcific leaflet thickening (suggestive of thrombus), and pannus observed on the CT, but was also observed remote from CT abnormalities (Figure 2).
Prediction of bioprosthetic valve dysfunction
Sixty-seven of the patients without established valve degeneration at baseline underwent repeat echocardiography at 2 years and demonstrated increased peak velocities compared with baseline (2.87 [IQR: 2.52 to 3.13] m/s vs. 2.73 [IQR: 2.38 to 3.07] m/s; p = 0.002). There was no association between hemodynamic progression and either the type or the age of the bioprosthesis (Online Table 4).
The 14 patients with abnormal CT findings appeared to demonstrate deterioration in bioprosthetic valve function after 2 years, but there was no statistical difference in disease progression compared with patients with a normal CT (change in peak velocity: 0.25 [IQR: −0.06 to 0.57] m/s vs. 0.10 [IQR: −0.04 to 0.31] m/s; p = 0.232).
By contrast, the 24 patients with increased 18F-fluoride uptake at baseline demonstrated clear evidence of deteriorating bioprosthesis function after 2 years, whereas patients without uptake displayed no change in valve function (change in peak velocity: 0.30 [IQR: 0.13 to 0.61] vs. 0.01 [IQR: −0.05 to 0.16] m/s/year; p < 0.001). Similarly, valves demonstrated progressive hemodynamic deterioration during follow-up on moving across tertiles of 18F-fluoride uptake (Online Figure 3). Moreover, baseline 18F-fluoride uptake correlated strongly with all echocardiographic measures of hemodynamic progression regardless of the method used to quantify 18F-fluoride uptake (Table 2).
Table 2.
Annualized Change | SUV Mean (LogN) | SUV Max (LogN) | TBR Mean (LogN) | TBR Max (LogN) |
---|---|---|---|---|
Peak valve velocity, LogN | r = 0.58 p < 0.001* |
r = 0.62 p < 0.001* |
r = 0.72 p < 0.001* |
r = 0.72 p < 0.001* |
Mean valve gradient, LogN | r = 0.42 p < 0.001* |
r = 0.45 p = 0.001* |
r = 0.52 p < 0.001* |
r = 0.52 p < 0.001* |
Effective orifice area, cm2/yr | r = −0.45 p < 0.001* |
r = −0.52 p < 0.001* |
r = −0.57 p < 0.001* |
r = −0.60 p < 0.001* |
Doppler velocity index | r = −0.46 p < 0.001* |
r = −0.51 p < 0.001* |
r = −0.59 p < 0.001* |
r = −0.60 p < 0.001* |
In 71 volunteers with bioprosthetic aortic valves without known valve dysfunction, baseline 18F-fluoride uptake measured by positron emission tomography using any conventional method of quantification (SUV or TBR) was associated with subsequent deterioration in each echocardiographic measure of valve function. Note all values are based upon the most-diseased-segment approach. The TBR mean values were used as the primary comparison and are referred to elsewhere in the paper simply as the TBR abbreviation. *Statistically significant.
Abbreviations as in Table 1.
On the basis of international guideline criteria 16, 17, 10 patients developed new bioprosthetic valve dysfunction during follow-up: 2 with valve regurgitation, 6 with valve stenosis, and 2 with mixed dysfunction. Median time from valve implantation to their assessment in the study was 7.5 (5 to 10) years. Of these patients, 5 had an abnormal baseline CT, whereas all 10 patients had increased 18F-fluoride uptake (TBR 1.89 [IQR: 1.46 to 2.59]), and this included the 7 patients with the highest TBR values in the cohort. Two proceeded to urgent valve reintervention, and another died of valve failure. Using the alternative criteria suggested by the recent international consensus statement (18), 16 patients met the definition of structural valve degeneration, 15 of whom demonstrated increased 18F-fluoride uptake (TBR 1.54 [IQR: 1.38 to 1.96] vs. 1.08 [IQR: 1.02 to 1.19]; p < 0.001). Seven patients fulfilled criteria for stage 2 or 3 structural valve degeneration, signifying the development of valve dysfunction, and they all exhibited increased baseline 18F-fluoride uptake (TBR 1.89 [IQR: 1.47 to 3.15]) (Figure 3). Moreover, 18F-fluoride activity increased in a stepwise fashion across these progressive stages of structural degeneration (TBR no degeneration: 1.08 [IQR: 1.02 to 1.19], stage 1: 1.48 [IQR: 1.34 to 1.64], stage 2: 1.72 [IQR: 1.42 to 1.94], stage 3: 4.23 [IQR: 3.15 to 5.36]; post hoc linear trend p < 0.001) (Figure 3). Two patients developed overt bioprosthetic valve failure during 2-year follow-up, and a further 2 patients developed valve failure on subsequent follow-up, all demonstrating high-intensity 18F-fluoride uptake (TBR 2.57 [IQR: 1.96 to 3.70]) (19) (Figure 4).
On univariable analysis, the only predictors of deterioration in bioprosthetic valve function (annualized change in bioprosthetic valve peak velocity) were current smoking habit (p = 0.047) and 18F-fluoride PET uptake, irrespective of whether the latter was considered as a categorical or continuous variable (both p < 0.001) (Online Table 4). On multivariable analysis incorporating age, sex, duration of valve implantation, baseline peak prosthetic valve velocity, and CT findings, 18F-fluoride uptake emerged as the only predictor of deterioration in bioprosthetic valve function (unstandardized coefficient 0.79 [95% confidence interval: 0.62 to 0.96]; p < 0.001) (Figure 3, Online Table 5).
Irrespective of the definition used 16, 17, 18, similar results were observed when the development of bioprosthetic valve dysfunction was considered as a categorical variable, with baseline 18F-fluoride uptake emerging each time as an independent predictor on multivariable analyses (international guideline criteria 16, 17: unstandardized coefficient 7.57 [standard error 2.66]; p = 0.004); expert consensus statement [18]: unstandardized coefficient 6.81 [standard error 2.92]; p = 0.02) (Online Table 6).
Discussion
In this multimodality prospective imaging study, we have identified 18F-fluoride PET-CT as the first noninvasive technique capable of detecting early bioprosthetic valve degeneration and of predicting future valve dysfunction (Central Illustration). We provide extensive validation of the technique against state-of-the-art ex vivo imaging and histology. This consistently demonstrated increased 18F-fluoride uptake in each of the failed bioprosthetic aortic valves examined, with PET activity colocalizing to areas of calcification, pannus, thrombus, and disrupted tissue architecture on histology. When applied to patients in the clinical setting, 18F-fluoride PET identified early valve degeneration beyond the resolution of conventional assessments and predicted the development of new valvular dysfunction and overt valve failure within the 2-year follow-up period. Indeed, 18F-fluoride uptake was an independent predictor of deteriorating bioprosthetic valve performance, outperforming all other variables, including valve type and age, and echocardiographic and CT findings. 18F-fluoride PET-CT, therefore, provides a readily applicable measure of valve degeneration with the potential to transform how we monitor and treat the expanding population of patients living with bioprosthetic valves.
Our study has several major strengths. This is a comprehensive multimodal imaging study using ultrasound, CT, and PET in patients with bioprosthetic aortic valves of varying implant ages. Uniquely, we have been able to correlate imaging findings with state-of-the-art ex vivo imaging and histological characterization of explanted valves, enabling us to validate our imaging technique and provide novel insights into the pathogenesis of bioprosthetic valve degeneration. Moreover, our systematic prospective study design has allowed us to confirm the utility of baseline 18F-fluoride PET in identifying patients at risk of developing future bioprosthetic dysfunction and overt valve failure.
A clinical challenge
Although bioprosthetic valves offer important advantages over mechanical valves, they have only limited durability 25, 26. Indeed, bioprosthetic valve degeneration has become a major clinical issue with the growing rate of bioprosthetic valve implantation and improved long-term patient survival. Although international guidelines recommend serial echocardiography for the detection of bioprosthetic degeneration 16, 17, visualization of the valve is often poor due to acoustic artefacts from the prosthesis, limiting its sensitivity. This means that valve degeneration is frequently well advanced before clinically overt valve dysfunction is apparent. Moreover, valve obstruction often evolves rapidly, whereas the catastrophic development of valvular regurgitation due to leaflet tears is frequently unheralded and unpredictable. Novel techniques capable of detecting the earlier stages of valve degeneration and allowing personalized and better planned management strategies for patients with bioprosthetic valves are therefore highly desirable.
Insights into the mechanism of bioprosthetic valve degeneration
We and others have used 18F-fluoride PET to assess cardiovascular calcification activity and tissue degeneration in a range of conditions including coronary heart disease, stroke, abdominal aortic aneurysms, and aortic stenosis 9, 10, 11, 12. We now extend these observations to bioprosthetic valves and confirm that 18F-fluoride localizes to regions of developing leaflet calcification on histology and high-resolution micro–PET-CT imaging. Calcification is believed to be the final common response to bioprosthetic valve injury and a major driver toward valve dysfunction, causing both progressive cusp stiffness and obstruction as well as leaflet fragility and tears. In this study, all bioprosthetic valves with established degeneration and valve failure demonstrated increased 18F-fluoride uptake. Moreover, in both our ex vivo and in vivo studies, we observed a close spatial interaction of calcification with both leaflet thrombosis and pannus, suggesting these may be potential upstream triggers. By contrast, a minority of bioprosthetic valves appear to degenerate without any histological evidence of calcification. In these cases, the remarkable histological features were gross leaflet thickening, fluid insudation, and disruption of normal collagen architecture. Interestingly, these valves also exhibited high levels of 18F-fluoride uptake. We hypothesize that up-regulation of matrix metalloproteinases may be implicated in the degradation of bioprosthetic leaflet tissue and that these proteins have the potential to bind 18F-fluoride 27, 28. Further investigation of the mechanism of uptake is required.
Recent reports have described focal noncalcific leaflet thickening as a marker of thrombus formation 20, 29. Although these areas can cause acute valve obstruction, more commonly, they do not result in any immediate hemodynamic disturbance, leading some to question their clinical relevance. In our study, we also observed low-attenuation noncalcific leaflet thickening on CT, suggestive of nonobstructive thrombus, that was associated with both increased 18F-fluoride activity and delayed deterioration in valve performance over the following 2 years. Similar findings were observed in explanted valves, where areas of thrombus colocalized with calcium on Von Kossa staining and exhibited increased 18F-fluoride PET activity. This raises the hypothesis that bioprosthetic valve thrombosis may act as one potential trigger to valve degeneration that might be prevented with prompt detection and anticoagulation. Interestingly, recent observational data have suggested that bioprosthetic durability may be enhanced by concomitant anticoagulation therapy (30).
Clinical implications
Our findings have several major implications for clinical practice. First, in addition to micro- and macrocalcification, 18F-fluoride PET-CT can identify a range of degenerative processes including fibrosis, thrombosis, and leaflet degradation with disrupted collagen architecture and fluid insudation. This has provided important insights into the pathogenesis of bioprosthetic valve failure and highlights potential targets for novel therapies aimed at improving valve durability. Second, 18F-fluoride PET can detect and quantify the early stages of valvular degeneration and identify patients otherwise thought to have normal valves but who are in fact at high risk of subsequent valve failure. Given the high mortality associated with emergency repeat valve replacement (8), these patients are likely to benefit from both close tailored surveillance and early elective intervention before the onset of abrupt valve failure. Conversely, patients without PET uptake, who here demonstrated very stable valve function during follow-up, could undergo less intensive surveillance. Our data suggest that 5 years post-implantation may be an appropriate stage at which to offer a PET-CT scan and guide follow-up, with further studies needed. Third, 18F-fluoride PET-CT appears of value in determining the presence of valvular degeneration in cases where the diagnosis and clinical management is uncertain, for example in patients with suspected patient–prosthesis mismatch. Finally, 18F-fluoride PET-CT provides a readily applicable measure of valve durability that may prove useful in assessing novel prosthetic designs such as transcatheter valves before these are extended into younger and healthier patient populations.
Study limitations
It was initially envisaged that serial CT calcium scoring of bioprosthetic aortic valves would also provide a marker of valve degeneration. However, in our hands, we observed major artefact related to motion and the valve frame that precluded accurate analysis of CT calcium scores. Leaflet pathology was instead more successfully identified by contrast-enhanced CT, where superior anatomic detail allowed differentiation between pannus ingrowth, thrombosis, or calcification, although a minority of scans still could not be accurately adjudicated. Alongside the robust information provided by 18F-fluoride PET this is an important advantage of the hybrid PET/CT technique.
Although our study population is relatively large compared with other cardiovascular PET studies, external validation in a larger study population with longer follow-up would be welcome. Future studies may wish to investigate 18F-fluoride activity in novel bioprostheses, in particular aortic and mitral transcatheter valves, as well as assessing whether the bioprosthetic 18F-fluoride signal is modifiable with adjuvant therapies, such as anticoagulation in patients with associated thrombus or drugs that modify calcium metabolism (e.g., bisphosphonates, denosumab).
Conclusions
18F-fluoride PET-CT detects early bioprosthetic valve degeneration, providing powerful prediction of subsequent deterioration in valve performance and highlighting patients at imminent risk of valve failure. This novel imaging approach has the potential to transform our understanding of bioprosthetic valve failure and the way in which we monitor and treat patients with bioprosthetic valves.
Perspectives.
COMPETENCY IN PATIENT CARE AND PROCEDURAL SKILLS: 18F-fluoride PET-CT imaging can detect early bioprosthetic valve degeneration and predict subsequent valve dysfunction, identifying patients at high risk of earlier valve failure.
TRANSLATIONAL OUTLOOK: Pragmatic studies are needed to define the role of 18F-fluoride PET-CT imaging in protocols for the surveillance of patients with bioprosthetic valves.
Footnotes
This work was supported by the British Heart Foundation, London, United Kingdom (FS/13/77/30488). The Edinburgh Clinical Research Facility and Edinburgh Imaging Facility are supported by NHS Research Scotland. Dr. Rudd is supported in part by the NIHR Cambridge Biomedical Research Centre, the British Heart Foundation, HEFCE, the EPSRC and the Wellcome Trust. Dr. van Beek is supported by the Scottish Imaging Network–a Platform for Scientific Excellence Chair of Clinical Radiology. Dr. Newby is supported by the British Heart Foundation (CH/09/002, RE/13/3/30183, RM/13/2/30158); and is the recipient of a Wellcome Trust Senior Investigator Award (WT103782AIA). Dr. Dweck is supported by the British Heart Foundation FS/14/78/31020; and is the recipient of the Sir Jules Thorn Award for Biomedical Research 2015 JTA/15. Dr. van Beek has received research support from Seimens Healthineers; has been a consultant for Imbio, Aidence NV, InHealth, and Mentholatum; and owns QCTIS Ltd. Dr. Leipsic’s institution has served as an institutional core lab for Edwards Lifesciences, Abbott, and Medtronic; and has been a consultant for and holds stock options in Circle CVI and HeartFlow. All other authors have reported that they have no relationships relevant to the contents of this paper to disclose.
Listen to this manuscript's audio summary by Editor-in-Chief Dr. Valentin FusteronJACC.org.
Appendix
References
- 1.Iung B., Vahanian A. Epidemiology of valvular heart disease in the adult. Nat Rev Cardiol. 2011;8:162–172. doi: 10.1038/nrcardio.2010.202. [DOI] [PubMed] [Google Scholar]
- 2.Nkomo V.T., Gardin J.M., Skelton T.N., Gottdiener J.S., Scott C.G., Enriquez-Sarano M. Burden of valvular heart diseases: a population-based study. Lancet. 2006;368:1005–1011. doi: 10.1016/S0140-6736(06)69208-8. [DOI] [PubMed] [Google Scholar]
- 3.Brown J.M., O’Brien S.M., Wu C., Sikora J.A., Griffith B.P., Gammie J.S. Isolated aortic valve replacement in North America comprising 108,687 patients in 10 years: changes in risks, valve types, and outcomes in the Society of Thoracic Surgeons National Database. J Thorac Cardiov Sur. 2009;137:82–90. doi: 10.1016/j.jtcvs.2008.08.015. [DOI] [PubMed] [Google Scholar]
- 4.Brennan J.M., Thomas L., Cohen D. Transcatheter versus surgical aortic valve replacement: propensity-matched comparison. J Am Coll Cardiol. 2017;70:439–450. doi: 10.1016/j.jacc.2017.05.060. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Pibarot P., Dumesnil J.G. Prosthetic heart valves: selection of the optimal prosthesis and long-term management. Circulation. 2009;119:1034–1048. doi: 10.1161/CIRCULATIONAHA.108.778886. [DOI] [PubMed] [Google Scholar]
- 6.Singhal P., Luk A., Butany J. Bioprosthetic heart valves: impact of implantation on biomaterials. ISRN Biomaterials. 2013;2013:728791. [Google Scholar]
- 7.Rodriguez-Gabella T., Pierre Voisine P., Puri R., Pibarot P., Rodés-Cabau J. Aortic bioprosthetic valve durability: incidence, mechanisms, predictors, and management of surgical and transcatheter valve degeneration. J Am Coll Cardiol. 2017;70:1013–1028. doi: 10.1016/j.jacc.2017.07.715. [DOI] [PubMed] [Google Scholar]
- 8.Vogt P.R., Brunner-LaRocca H.P., Sidler P. Reoperative surgery for degenerated aortic bioprostheses: predictors for emergency surgery and reoperative mortality. Eur J Cardio-Thorac. 2000;17:134–139. doi: 10.1016/s1010-7940(99)00363-2. [DOI] [PubMed] [Google Scholar]
- 9.Dweck M.R., Jones C., Joshi N.V. Assessment of valvular calcification and inflammation by positron emission tomography in patients with aortic stenosis. Circulation. 2012;125:76–86. doi: 10.1161/CIRCULATIONAHA.111.051052. [DOI] [PubMed] [Google Scholar]
- 10.Forsythe R.O., Dweck M.R., McBride O.M.B. 18F–Sodium fluoride uptake in abdominal aortic aneurysms: the SoFIA3 study. J Am Coll Cardiol. 2018;71:513–523. doi: 10.1016/j.jacc.2017.11.053. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Joshi N.V., Vesey A.T., Williams M.C. 18F-Fluoride positron emission tomography for identification of ruptured and high-risk coronary atherosclerotic plaques: a prospective clinical trial. Lancet. 2014;383:705–713. doi: 10.1016/S0140-6736(13)61754-7. [DOI] [PubMed] [Google Scholar]
- 12.Vesey A.T., Jenkins W.S.A., Irkle A. 18F-Fluoride and 18F-fluorodeoxyglucose positron emission tomography after transient ischemic attack or minor ischemic stroke: case–control study. Circ-Cardiovasc Imag. 2017;10:e004976. doi: 10.1161/CIRCIMAGING.116.004976. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Irkle A., Vesey A.T., Lewis D.Y. Identifying active vascular microcalcification by (18)F-sodium fluoride positron emission tomography. Nat Commun. 2015;6:7495. doi: 10.1038/ncomms8495. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Dweck M.R., Jenkins W.S., Vesey A.T. 18F-sodium fluoride uptake is a marker of active calcification and disease progression in patients with aortic stenosis. Circ-Cardiovasc Imag. 2014;7:371–378. doi: 10.1161/CIRCIMAGING.113.001508. [DOI] [PubMed] [Google Scholar]
- 15.Jenkins W.S., Vesey A.T., Shah A.S. Valvular (18)F-Fluoride and (18)F-fluorodeoxyglucose uptake predict disease progression and clinical outcome in patients with aortic stenosis. J Am Coll Cardiol. 2014;66:1200–1201. doi: 10.1016/j.jacc.2015.06.1325. [DOI] [PubMed] [Google Scholar]
- 16.Lancellotti P., Pibarot P., Chambers J. Recommendations for the imaging assessment of prosthetic heart valves: a report from the European Association of Cardiovascular Imaging. Eur Heart J-Card Img. 2016;17:589–590. doi: 10.1093/ehjci/jew025. [DOI] [PubMed] [Google Scholar]
- 17.Zoghbi W.A., Chambers J.B., Dumesnil J.G. Recommendations for evaluation of prosthetic valves with echocardiography and Doppler ultrasound. J Am Soc Echocardiog. 2009;22:975–1014. doi: 10.1016/j.echo.2009.07.013. [DOI] [PubMed] [Google Scholar]
- 18.Dvir D., Bourguignon T., Otto C.M. Standardized definition of structural valve degeneration for surgical and transcatheter bioprosthetic valves. Circulation. 2018;137:388–399. doi: 10.1161/CIRCULATIONAHA.117.030729. [DOI] [PubMed] [Google Scholar]
- 19.Capodanno D., Petronio A.S., Prendergast B. Standardized definitions of structural deterioration and valve failure in assessing long-term durability of transcatheter and surgical aortic bioprosthetic valves: a consensus statement from the European Association of Percutaneous Cardiovascular Interventions. Eur Heart J. 2017;38:3382–3390. doi: 10.1093/eurheartj/ehx303. [DOI] [PubMed] [Google Scholar]
- 20.Makkar R.R., Fontana G., Jilaihawi H. Possible subclinical leaflet thrombosis in bioprosthetic aortic valves. N Engl J Med. 2015;373:2015–2024. doi: 10.1056/NEJMoa1509233. [DOI] [PubMed] [Google Scholar]
- 21.Pache G., Schoechlin S., Blanke P. Early hypo-attenuated leaflet thickening in balloon-expandable transcatheter aortic heart valves. Eur Heart J. 2016;37:2263–2271. doi: 10.1093/eurheartj/ehv526. [DOI] [PubMed] [Google Scholar]
- 22.Fujita B., Kutting M., Seiffert M. Calcium distribution patterns of the aortic valve as a risk factor for the need of permanent pacemaker implantation after transcatheter aortic valve implantation. Eur Heart J Cardiovasc Imaging. 2016;17:1385–1393. doi: 10.1093/ehjci/jev343. [DOI] [PubMed] [Google Scholar]
- 23.Motoyama S., Kondo T., Sarai M. Multislice computed tomographic characteristics of coronary lesions in acute coronary syndromes. J Am Coll Cardiol. 2007;50:319–326. doi: 10.1016/j.jacc.2007.03.044. [DOI] [PubMed] [Google Scholar]
- 24.Pawade T.A., Cartlidge T.R., Jenkins W.S. Optimization and reproducibility of aortic valve 18F-fluoride positron emission tomography in patients with aortic stenosis. Circ Cardiovasc Imag. 2016;9:e005131. doi: 10.1161/CIRCIMAGING.116.005131. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Foroutan F., Guyatt G.H., O’Brien K. Prognosis after surgical replacement with a bioprosthetic aortic valve in patients with severe symptomatic aortic stenosis: systematic review of observational studies. BMJ. 2016;354:i5065. doi: 10.1136/bmj.i5065. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Bloomfield P., Wheatley D.J., Prescott R.J., Miller H.C. Twelve year comparison of a Bjork-Shiley mechanical heart valve with porcine bioprostheses. N Engl J Med. 1991;324:573–579. doi: 10.1056/NEJM199102283240901. [DOI] [PubMed] [Google Scholar]
- 27.Simionescu A., Simionescu D., Deac R. Biochemical pathways of tissue degeneration in bioprosthetic cardiac valves. The role of matrix metalloproteinases. ASAIO J. 1996;42:561–567. doi: 10.1097/00002480-199609000-00049. [DOI] [PubMed] [Google Scholar]
- 28.Kato M.T., Bolanho A., Zarella B.L., Salo T., Tjäderhane L., Buzalaf M.A.R. Sodium fluoride inhibits MMP-2 and MMP-9. J Den Res. 2014;93:74–77. doi: 10.1177/0022034513511820. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Chakravarty T., Sondergaard L., Friedman J. Subclinical leaflet thrombosis in surgical and transcatheter bioprosthetic aortic valves: an observational study. Lancet. 2017;389:2383–2392. doi: 10.1016/S0140-6736(17)30757-2. [DOI] [PubMed] [Google Scholar]
- 30.Del Trigo M., Muñoz-García A.J., Latib A. Impact of anticoagulation therapy on valve hemodynamic deterioration following transcatheter aortic valve replacement. Heart. 2018;104:814–820. doi: 10.1136/heartjnl-2017-312514. [DOI] [PubMed] [Google Scholar]