Short-acquisition-time JPRESS and its application to paediatric brain tumours

Dominic Carlin 1,2, Ben Babourina-Brooks 1,2, Theodoros N Arvanitis 1,2,3, Martin Wilson 2,4, Andrew C Peet 1,2,5,
PMCID: PMC6424926  PMID: 30460431

Abstract

Objective

To develop and assess a short-duration JPRESS protocol for detection of overlapping metabolite biomarkers and its application to paediatric brain tumours at 3 Tesla.

Materials and methods

The short-duration protocol (6 min) was optimised and compared for spectral quality to a high-resolution (38 min) JPRESS protocol in a phantom and five healthy volunteers. The 6-min JPRESS was acquired from four paediatric brain tumours and compared with short-TE PRESS.

Results

Metabolite identification between the 6- and 38-min protocols was comparable in phantom and volunteer data. For metabolites with Cramer–Rao lower bounds > 50%, interpretation of JPRESS increased confidence in assignment of lactate, myo-Inositol and scyllo-Inositol. JPRESS also showed promise for the detection of glycine and taurine in paediatric brain tumours when compared to short-TE MRS.

Conclusion

A 6-min JPRESS protocol is well tolerated in paediatric brain tumour patients. Visual inspection of a 6-min JPRESS spectrum enables identification of a range of metabolite biomarkers of clinical interest.

Electronic supplementary material

The online version of this article (10.1007/s10334-018-0716-6) contains supplementary material, which is available to authorized users.

Keywords: Magnetic resonance spectroscopy, Metabolism, Brain neoplasms

Introduction

Metabolic reprogramming is an emerging hallmark of cancer [1]. Magnetic resonance spectroscopy (MRS) provides a convenient non-invasive means to investigate the metabolic composition of cancer. A wide range of metabolic changes have been identified in cohort studies of brain tumours [2, 3]; however, unambiguous detection of some overlapping J-coupled metabolites in a single spectrum has been limited. Clinical applications of MRS have, therefore, typically focused on N-acetyl aspartate (NAA), lipids, creatine (Cr) and choline (Cho), which can be identified at short and long TE, and lactate (Lac), which can be distinguished from lipids and macromolecules at long TE.

In addition to the above, there is growing evidence that glutamate (Glu) [3, 4], glutamine (Gln) [5], glycine (Gly) [3, 6, 7], myo-Inositol (mI) [3, 810], scyllo-Inositol (Scy) [4] and taurine (Tau) [2, 911] are altered in paediatric brain tumours. Therefore, these metabolites may provide new markers of diagnosis, prognosis and identification of high-risk patients, which may allow for better treatment stratification and more effective disease management. However, identification of these metabolites is hindered by small peak intensities and spectral overlap leading to reduced reproducibility [12] and increased Cramer–Rao lower bounds (CRLB) [13]. Various methods have been proposed for the detection of Glu, Gln, Gly, mI and Tau. These methods range in complexity from simply using an optimised TE [1416] to more complex acquisition schemes such as TE averaging [17] and quantum filters [1820]. However, these methods are typically optimised for the detection of a specific metabolite and can sacrifice information from other relevant metabolite biomarkers. As use of all quantified metabolite data can improve the accuracy of tumour classification [21], acquisition methods that identify as many metabolites as possible are, therefore, preferred.

Commonly used in vitro, J-resolved spectroscopy (JPRESS) [22, 23] has previously shown promise in brain tumours [24]. JPRESS is acquired by collecting PRESS spectra at multiple echo times, retaining the chemical shift information typical of conventional one-dimensional spectroscopy in F2, the x-dimension, and indirectly encoding the scalar coupling information for each metabolite in F1, the y-dimension. This spreads coupled metabolite resonances in two dimensions, disambiguating the identification of metabolites using the known chemical shift and scalar coupling constants for each metabolite based on their position in the spectrum [25].

One of the key challenges to implementing JPRESS in clinical practice is the associated long acquisition times. Children are often unable to tolerate long scanning sessions, making scan duration, particularly, pertinent in this setting. The aim of this study was to propose and test a short-duration JPRESS protocol, and evaluate its ability to detect and discriminate between overlapping metabolites of interest in paediatric brain tumours.

Materials and methods

The study was approved by the East Midlands—Derby Research Ethics Committee (REC 04/MRE04/41) operating under the rules of Declaration of Helsinki 1975 (and as revised in 1983), and informed consent was obtained from all volunteers and patients. The acquisition protocol was optimised using a combination of simulated, phantom and volunteer data to determine the optimal spacing between TEs (ΔTE), number of TE steps acquired (NTE) and the number of averages (NSA) acquired per TE (NSA/TE). A high-resolution 38.4-min JPRESS protocol was used for comparison to assess the reduction in spectral quality (Table 1).

Table 1.

Acquisition protocols in phantom and volunteer for JPRESS protocol and processing development

TR (ms) Starting TE (ms) NSA/TE TE spacing (ms) #TEs Maximum TE (ms) Acquisition time (min) Volunteers scanned
Protocol 1 2000 36 8 5 128 676 38.4 0 (2 Phantoms)
Protocol 2 2000 36 16 10 64 186 38.4 1
Protocol 3 2000 36 8 15 18 291 6 4 (2 Phantoms)

All data were acquired with one dummy scan per TE and a water-unsuppressed spectrum collected with NSA 1 per TE

Experimental

Simulations

JPRESS spectra of common brain metabolites [25] were simulated using VESPA [26]. 1D MRS of each metabolite was simulated using the VESPA pulse sequence option ‘PRESS Ideal’ at a range of echo times from 35 ms to 675 ms. A fixed TE1 (20 ms) was used with TE2 varied as required for each simulated TE. The standard PRESS sequence was repeated for each TE and the 1D MRS was combined with JPRESS datasets with ΔTEs of 5, 10, 15 and 20 ms. The JPRESS spectra were then visually assessed to determine the optimal TE spacing. The PRESS Ideal pulse sequence assumes ideal hard RF pulses which flip all spins during an infinitely short time with the same phase and angle. Experimental acquisitions may be subject to signal cancellation due to phase variation and chemical shift displacement artefact for coupled spins such as lactate. In particular, the limited RF pulse bandwidth leads to spatially dependent evolution of J-coupling and consequently, additional artefact peaks at J = 0 Hz due to J-refocusing [27]. In general, however, the simulated JPRESS serves as a good approximation of metabolite peak patterns.

The 2D JPRESS spectra were further used as a reference library to aid metabolite identification and are presented in Online Resource 1; however, the spectrum for NAAG, which is typical of low abundance, does not include its associated coupled resonances. Each metabolite spectrum was displayed with 40 contour lines up to a level of 50% of the maximum signal intensity for that metabolite and no minimum threshold.

Phantom

MRS was acquired using a Philips Achieva 3T scanner (Philips Healthcare, Best, Netherlands). JPRESS was acquired as a TE series and TE1 was fixed to 20 ms for experimentally acquired data. All averages for a given TE were acquired before acquiring data for the next linear TE, with the Philips function to mitigate the frequency drift applied.

Phantom data were acquired from two phantoms. A “braino” phantom (GE Medical Systems, Milwaukee, WI, USA) of common brain metabolites at physiological levels containing 12.5 mM NAA, 12.5 mM Glu, 10 mM Cr, 7.5 mM, mI, 5 mM Lac and 3 mM Cho was scanned. The second phantom contained 5 mM Glu, 5 mM Gln, 5 mM Gly and 5 mM mI to assess the resolution of key overlapping metabolites under experimental conditions.

1 g of sodium azide was added to each phantom as a biocide to prevent the growth of bacterial organisms. The phantoms were pH adjusted to 7.2 and 0.5 mM and gadopentetate dimeglumine (Magnevist®, Bayer HealthCare Pharmaceuticals, Berlin, Germany) was added to each phantom as a relaxation agent to shorten relaxation times to the physiological range.

Volunteer

Data were obtained from five healthy adults (three male and two female) with a mean age of 25 ± 2 years for the volunteer data. MRS of parietal grey matter was acquired from cubic voxels of size 30 × 30 × 30 mm3. The acquisition protocols for the phantom and volunteer data are presented in Table 1. Protocol 2 was assumed to be the gold standard for metabolite detection due to its greater number of NSA/TE and the long echo train length. The spectral quality of the other protocols was assessed with reference to Protocol 2.

Protocol development

TE spacing

The spectral width of F1 is set by 1/ΔTE. Phantom data acquired with ΔTE = 5 ms (Protocol 1) were spliced to generate additional subsets with ΔTE of 10 ms, 15 ms and 20 ms with the same NSA/TE and range of TEs.

NSA/TE

Volunteer data acquired with 8 NSA/TE (Protocol 3) and 16 NSA/TE (Protocol 2) were acquired. The two spectra were compared to determine if the major resonances of Glu, Gln, and mI could be identified with fewer NSA/TE.

Number of TEs

Phantom and volunteer data with final TEs of 290 ms (Protocol 3) and 665 ms (Protocols 1 and 2) were compared to assess the effect that truncating the number of TEs acquired had on spectral resolution.

Clinical JPRESS of paediatric brain tumours

Four patients with paediatric brain tumours were investigated (Table 2). The optimised JPRESS protocol was acquired following conventional imaging.

Table 2.

Patient details of paediatric brain tumour patients studied with JPRESS

Age (years) Sex Tumour type
9.8 Male Medulloblastoma
8.2 Male Pilocytic astrocytoma
6.5 Female Optic pathway glioma
10.6 Male Diffuse intrinsic pontine glioma (DIPG)

JPRESS was acquired from cubic voxels of size 30 × 30 × 30 mm3 for all the cases with TR = 2000 ms, TEmin = 42 ms, NTE = 18, ΔTE = 15 ms and NSA/TE = 8. The total acquisition time was 6 min. PRESS with an echo time of 35 ms, TR of 2000 ms and 128 NSA was also acquired from the cases with medulloblastoma and diffuse intrinsic pontine glioma.

Data processing and analysis

The data were processed in two steps. First, the 1D MRS of each acquired TE was extracted and analysed with TARQUIN v4.3.8 using the 1H Brain basis set [28]. TARQUIN estimates of the water peak position, ref, in the water-suppressed spectra and CRLBs were recorded.

The full multi-TE dataset for each experiment was then transferred to a personal computer and processed using software written in-house in Python (v 2.7.3) to produce the 2D JPRESS spectra. The water component of the free induction decays was removed using Hankel singular value decomposition and the baseline was corrected for any displacement to reduce interference. Baseline displacement correction was performed by taking the average value of a region of the spectrum containing only noise and no signal, and subsequently subtracting this value from the spectrum. Estimates of the water peak position, water_ref, in the water-suppressed spectra, were collected for each TE from the TARQUIN result file. Small frequency drifts between TEs caused broader resonances in the JPRESS F2 dimension. The water peak was, therefore, used as a chemical shift reference and was aligned to 4.65 ppm for all spectra in a dataset. Peak alignment was performed by multiplying each point, k, of the corresponding FID by exp(− 2πi × ref × k×(128⁄2000)), where ref = 4.65 ppm—water_ref, 128 MHz is the central 1H NMR frequency and 2000 Hz is the sampling frequency [29]. TARQUIN fits to NAA, Cho and Cr singlets which were used to simulate spectra at TE > 297 ms to reduce truncation artefacts caused by undersampling. Simulated spectra were appended to the dataset with a relaxation penalty of exp(− TE/T2) [30], applied to successive echo times and a T2 of 200 ms assumed for all metabolites [31]. Using this method, the dataset was extended from 18 TEs (297 ms) to 30 TEs (477 ms) improving spectral resolution (Fig. 1).

Fig. 1.

Fig. 1

Representative examples of a acquired TE = 297-ms volunteer MRS from JPRESS acquisition and simulated TE = 312-ms MRS based on TARQUIN singlet fits to acquired TE = 297-ms MRS data. Simulated MRS was used to extend the JPRESS dataset. The F1 linewidths of b non-extended JPRESS were larger than those of c extended JPRESS for macromolecular resonances (0.5–1.5 ppm) and for coupled metabolites (2–3 ppm and 3.2–3.9 ppm)

The 2D Fourier-transformed JPRESS matrix was displayed as a contour plot in magnitude mode. Lower and upper bounds were used with thresholds adjusted to ensure spectra were clean and free of noise where possible. In each case, the thresholds were examined to determine if the metabolites could be identified in a more noisy spectrum. Spectral quality was determined through visual inspection of the display for artefacts. The thresholds of the display were further optimised manually to ensure that all desired peaks were present and fully resolved. The full data processing pipeline is shown in Fig. 2.

Fig. 2.

Fig. 2

Optimised acquisition protocol and processing pipeline for JPRESS of paediatric brain tumours

CRLBs were used to assess confidence in the presence of metabolites in short-TE PRESS.

Results

Protocol development and optimisation

The clinical JPRESS protocol was optimised for short acquisition times by optimising the parameter settings of ΔTE, NTE and NSA/TE. The ability to detect key metabolites using short-duration 6-min JPRESS was compared with a 38.4-min JPRESS protocol, with the 38.4-min protocol assumed to be the gold standard for metabolite detection in this study.

Phantom

Figure 3 shows short-TE MRS and JPRESS acquired in a braino phantom (Fig. 3a–c) and a phantom containing equal quantities of Glu, Gln, Gly and mI (Fig. 3d–f). The position of JPRESS resonance peaks can be described by their J-coupling (Hz) values and chemical shift (ppm) positions on the F1 and F2 axes, respectively. The Glu resonances with the greatest signal intensity are located at F2/F1 = 2.4 ppm/0 Hz and 2.35 ppm/− 7 Hz. mI presents as a complex arrangement of peaks between 3.5 and 3.6 ppm with characteristic resonances off the F1 = 0 Hz axis at F2/F1 = 3.50 ppm/− 10 Hz, 3.50 ppm/8 Hz and 3.60 ppm/5 Hz. The Gln resonance with the highest signal intensity is located at 2.40 ppm/− 7 Hz and Gly presents as a single resonance at 3.55 ppm/0 Hz.

Fig. 3.

Fig. 3

a 35-ms PRESS, b 38.4-min JPRESS (Protocol 1) and c 6-min JPRESS of the braino phantom (Protocol 3). d 35-ms PRESS, e 38.4-min JPRESS (Protocol 1) and f 6-min JPRESS (Protocol 3) of a phantom containing 5 mM Glu, 5 mM Gln, 5 mM Gly and 5 mM mI. For comparison of spectral quality, b and e show only the F1 spectral range from − 33 Hz to + 33 Hz, with the whole spectrum inset top right and dashed lines showing F1 = ± 33 Hz

A 38.4-min-long JPRESS protocol (Protocol 1) was compared with a shortened 6-min JPRESS protocol (Protocol 3). In both phantoms, all metabolite resonances fell within the F1 spectral range of ± 33 Hz and were visible using both protocols. Simulated JPRESS spectra of common brain metabolites are presented in Online Resource 1, and all metabolite resonances simulated also fall within this range, which corresponds to a ΔTE of 15 ms.

Improved resolution in the F1 dimension was found in the 38.4-min protocol (Fig. 3b) with the tails of the NAA, Cr and Cho singlets longer in the shorter 6-min protocol (Fig. 3c). There was no observable difference in resolution of the F2 dimension between the two protocols.

The main Glu and Gln resonances were resolved in both protocols. The Gly resonance at F2/F1 = 3.54 ppm/0 Hz was indistinct from mI; however, mI could be detected from its off − 0 Hz resonances.

Volunteer

Figure 4 compares a 38.4-min JPRESS protocol (Fig. 4b) with a representative short 6-min protocol (Fig. 4d), and mI, Glu and Gln were identified by their characteristic resonance patterns in both the short and longer protocols; however, improved resolution of Glu and Gln was found in phantom data when compared to in vivo data. The 35-ms PRESS MRS (Fig. 4c) was unable to detect Gln, Tau or Gly; however, the corresponding short 6-min protocol discriminated Gln.

Fig. 4.

Fig. 4

a 35-ms PRESS and b 38.4-min JPRESS (Protocol 2) of healthy grey matter in a female volunteer. c 35-ms PRESS and d 6-min JPRESS (Protocol 3) of a healthy grey matter in a male volunteer. For comparison of spectral quality, b shows only the F1 spectral range from − 33 Hz to + 33 Hz, with the whole spectrum inset top right and dashed lines showing F1 = ± 33 Hz

JPRESS of paediatric brain tumours

Figure 5 shows the JPRESS spectrum and a short-TE 35-ms PRESS spectrum acquired in the same scanning session from a medulloblastoma prior to treatment. The JPRESS resonance at F2/F1 = 3.54 ppm/0 Hz is most consistent with Gly, with no mI cross-peaks observed. The corresponding short-TE MRS resonance was assigned to Gly (CRLB = 12.9%) with no mI detected. Resonances were also present at F2/F1 = 3.40 ppm/0 Hz and 3.34 ppm/− 7 Hz. While the location of these resonances is most consistent with Tau, the F2/F1 = 3.40 ppm, 0 Hz resonance was twice as large as that at 3.34 ppm/− 7 Hz. The corresponding resonances of the simulated Tau JPRESS were of equal intensity. There is, therefore, an unassigned resonance at F2/F1 = 3.40 ppm, 0 Hz with no obvious associated resonances. The short-TE CRLB for Tau was 12.8%; however, there was a small residual peak in this region. The lactate doublet (short-TE CRLB = 112%) and lipid peak at 1.3 ppm (short-TE CRLB = 124%) were well separated in JPRESS. Glu (short-TE CRLB = 54.2%) and Gln (short-TE CRLB = 94.2%) were not detected in the JPRESS spectrum.

Fig. 5.

Fig. 5

a JPRESS spectrum of medulloblastoma, b 35-ms PRESS of medulloblastoma collected in the same session. TARQUIN residual and fits (FWHM = 7 Hz) for Gly and Tau included

Figure 6 shows the JPRESS spectrum and the extracted short-TE (42 ms) MRS of a pilocytic astrocytoma (SNR 12.8). The resonances between 3.5 and 3.6 ppm were assigned by TARQUIN to Gly (short-TE CRLB = 41.7%) and mI (short-TE CRLB = 53.5%) with the peaks at 3.55 ppm being of equal intensity. The intensities of the JPRESS resonances at F2/F1 = 3.54 ppm/0 Hz and 3.54 ppm/− 11 Hz are approximately equal, which is consistent with a combination of mI and Gly. Scy was also present in the JPRESS spectrum at 3.34 ppm/0 Hz and in the short-TE analysis (CRLB = 118%). Glu (CRLB = 67.1%) and Gln (not detected in short-TE MRS) were not detected in the JPRESS spectrum.

Fig. 6.

Fig. 6

a JPRESS spectrum of pilocytic astrocytoma, b 1D MRS extracted from the pilocytic astrocytoma JPRESS dataset with a TE of 42-ms. TARQUIN fits (FWHM = 7 Hz) for mI, Gly and Scy included

The JPRESS and corresponding short-TE (35 ms) MRS of an optic pathway glioma (OPG) are shown in Fig. 7. The short-TE assignments to the peaks at F1 = 0 Hz, F2 = 3.50, 3.6 and 3.8 ppm were to mI, mI and Glth with short-TE CRLBS of 67.3% for mI and 67.1% for Glth. The JPRESS shows the characteristic mI peak at F1/F2 = 3.50 ppm/− 10 Hz and also shows the splitting of the Lac doublet at 1.3 ppm. No Lac was detected in the short-TE MRS analysis. Glu (CRLB = 106%) and Gln (CRLB = 164%) were not detected in the JPRESS spectrum.

Fig. 7.

Fig. 7

a JPRESS spectrum of optic pathway glioma (OPG), b 35-ms PRESS of OPG collected in the same session. TARQUIN residual and fits (FWHM = 4 Hz) for Lac, mI and Glth included

The JPRESS and short-TE PRESS spectra (SNR = 12.2) of a DIPG post-radiotherapy are shown in Fig. 8. The resonances between − 20 Hz and 10 Hz at approximately 3.54 ppm are most consistent with mI. The vast majority of this region in the short-TE PRESS was assigned to mI (CRLB = 13.2%) with a small amount of Gly (CRLB = 112%) present. Glu (CRLB = 134%) and Gln (CRLB = 110%) were not detected in the JPRESS spectrum. Since mI is a complex multiplet, a large FWHM leads to individual peaks merging and, therefore, a different overall lineshape being formed when compared with the cases where the FWHM is small and the individual peaks are better resolved. This effect is seen in the DIPG spectrum where the FWHM is 8 Hz (Fig. 8) compared to the OPG spectrum where the FWHM is 4 Hz (Fig. 7).

Fig. 8.

Fig. 8

a JPRESS spectrum of diffuse intrinsic pontine gliomas (DIPG), b 35-ms PRESS of DIPG collected in the same session. TARQUIN residual and fits (FWHM = 8 Hz) for mI and Gly included

Discussion

This study presents a 6-min clinical protocol for JPRESS developed for visual detection of key metabolites found in paediatric brain tumours and its application to a series of cases. The protocol was optimised using a braino phantom and volunteer data with the short-duration protocol’s efficacy assessed by comparing with a longer protocol which was taken as the reference standard.

The main metabolites of interest in this study were Glu, Gln, Gly, mI, Scy and Tau due to their roles in cancer metabolism. Metabolite identification was assessed in phantom and volunteer JPRESS spectra acquired with 8 NSA/TE and 16 NSA/TE. Glu, Gln and mI could all be identified in the 8 NSA/TE volunteer spectra and these were selected to reduce the scan duration. The absence of Gly, Scy and Tau was expected in volunteers due to their relative low abundance in healthy brain [25] but they are of sufficiently high concentration in some brain tumours that the same NSA/TE should be appropriate for these metabolites [2]. Further reduction of the protocol duration was obtained by considering the final TE acquired. Due to their relatively small peak intensities, a final TE of 297 ms was sufficient for strongly coupled resonances to be indistinguishable from the noise following T2 relaxation decay, but for singlets, approximately 20% of the metabolite signal remains assuming typical T2 relaxation times [29].

Not acquiring MRS with a TE greater than 297 ms reduced acquisition time and did not cause major artefacts in the JPRESS spectral when the singlet decay at longer echo times was simulated. The ΔTE increment was increased to 15 ms for a J-resolved bandwidth of 67 Hz, which was considered to be the bandwidth required to observe all coupled resonances of common metabolites [25]. While the ∆TE increment used here is sufficient for metabolite detection, smaller ∆TE increments are also used for JPRESS applications [22] as these can help avoid artefacts caused by signal folding due to insufficient water suppression or no or non-optimal apodisation.

In the four paediatric brain tumour cases, metabolites of interest were confirmed or excluded based on the JPRESS spectrum compared with the short-TE MRS. In the medulloblastoma case, Gly and Tau were identified with peak positions and J-coupled resonances consistent with metabolite simulations. The short-TE analysis was consistent with the interpretation of the JPRESS spectrum, with a small residual peak between 3.30 and 3.40 ppm. Based on the JPRESS appearance, this could be due to an unassigned singlet peak. The medulloblastoma JPRESS was consistent with ex vivo HR-MAS analysis [32]. Interpretation of JPRESS, with the aid of simulated data, also corroborated the TARQUIN assignments of Gly, mI and Scy in the short-TE MRS pilocytic astrocytoma patient case. TARQUIN analysis of the short-TE DIPG MRS indicated that mI was the predominant metabolite at 3.50 ppm, which was consistent with the JPRESS spectrum. In the optic pathway glioma patient, the JPRESS indicated that Glth, mI and Lac were present in the voxel, whilst TARQUIN’s assignments in the short-TE PRESS had CRLB > 50% for Glth and mI and Lac was not fitted. The discrimination of these metabolites of interest has clinical relevance for both diagnosis and prognosis. DIPG and OPG patients rarely have biopsies and, therefore, imaging plays a vital role in their tumour diagnosis. The discrimination of Gly and mI is of particular importance as Gly has been found to be a survival marker in paediatric brain tumours [33], while Scy is also a marker of survival [5]. The confidence in the assignment of these metabolites is, therefore, of value and JPRESS can be used in a short scan time to achieve this, with the features seen in the low-grade optic pathway glioma and pilocytic astrocytomas reflecting those commonly identified ex vivo [34].

One of the key advantages of JPRESS over short-TE MRS is that visual inspection increases confidence in metabolite assignment. CRLBs are the standard method of assessing confidence in 1D MRS metabolite assignments. While for reliable metabolite quantification, the consensus CRLB threshold is 20%, there is consensus that metabolites can be detected with high confidence when CRLBs are less than 50% [35]. However, in the short-TE analysis of patients in this study, several key metabolites of clinical interest had CRLBs > 50% but were clearly present in the JPRESS spectrum. These included Lac in the medulloblastoma and OPG, mI in the PA and OPG and Scy in the PA, with Gly also having a relatively high CRLB of 41.7% in the PA.

A range of methods are available for the detection of metabolite biomarkers of disease and prognosis, including optimised TE 1D MRS and other spectral-editing approaches. JPRESS was chosen ahead of these as 2D methods do not suppress or enhance metabolite resonances, allowing a wide range of metabolites to be potentially identified. JPRESS was also chosen ahead of L-COSY [36] as it is more readily available on clinical 3 T MR systems without modification and has regulatory approval (CE-marking) on the scanner used in this study; however, the analysis itself is not CE-marked. JPRESS also offers the facility to extract the short- and long-TE 1D MRS spectra that radiologists are already familiar with interpreting, aiding its translation into the clinical environment; however, these extracted spectra will have lower SNR than a typical 1D MRS acquisition. Further improvement of JPRESS could be conferred by optimising the apodisation function or by the use of a weighted averaging acquisition [37].

While the presented JPRESS protocol offers some clear advantages over conventional short-TE PRESS, there are a number of limitations to this study. First, with visual detection of key metabolite biomarkers being the aim of this study, no attempt was made to quantify metabolite concentrations. However, while quantification was not performed here, a recent study of short-duration JPRESS observed a test–retest reliability < 10% for NAA, Cho, Glth, Glu and mI [38]. 2D JPRESS can also be quantified using the designated quantification package ProFit [39], which could yield the same, or better results, as those achieved here by visual inspection. While ProFit was not used in this study, a comparison of the two approaches would be a logical extension to this work and will be examined in future work.

Second, a voxel of size 30 × 30 × 30 mm3 was used to ensure sufficient SNR. Though this voxel is large, this JPRESS protocol would be suitable for brain tumours which have not been resected, with metabolite-specific pulse sequences used to follow-up response to treatment for pre-selected metabolites. Recent advances, such as the combination of JPRESS with fast echo-planar spectroscopic imaging approaches [40, 41], are likely to mitigate the partial volume problems associated with large voxels. These methods typically require the use of a research mode on clinical scanners, but should also be investigated.

Third, while both Glu and Gln could be identified in healthy volunteers, they were not detected in any of the four paediatric brain tumours investigated with JPRESS. This is likely due to the lower abundance of these metabolites in brain tumours compared with age-matched normal brain [3]. In the corresponding short-TE MRS, Glu and Gln consistently had CRLBs > 50%; however, a greater NSA/TE might be better suited for the detection of these metabolites. While we did not thoroughly explore the effect of varying the NSA/TE, we chose to acquire eight averages per TE to maximise SNR. Reducing the NSA/TE may be able to achieve similar spectral quality in a shorter timeframe, while scanning more TEs with a smaller NSA/TE could potentially be used to improve spectral quality.

Finally, while JPRESS has the potential to identify or confirm the presence of novel metabolite biomarkers, metabolite assignment can still be a non-trivial task. It is hoped that the simulated JPRESS spectra presented in Online Resource 1 will aid others in basic metabolite assignment. High-resolution experiments using ex vivo tissue could help to elucidate metabolite spectra further.

Conclusions

A 6-min JPRESS protocol was well tolerated in a paediatric setting and was able to detect metabolite biomarkers of disease and prognosis in childhood brain tumours. Visual inspection of JPRESS confirmed the presence of metabolites with high CRLBs in short-TE MRS and would be suitable for use in patients with large tumours.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Acknowledgements

This study was funded by a Birmingham Children’s Hospital Research Foundation (BCHRF287), the National Institute for Health Research (NIHR) via a Research Professorship (13-0053), and an EU Marie Curie: International Incoming Fellowship (PIIF-GA-2012-332278), CR UK and EPSRC Cancer Imaging Programme at the Children’s cancer and Leukaemia Group (CCLG) in association with the MRC and Department of Health (England) (C7809/A10342), Medical Research Council G0601327, and Experimental Cancer Medicine Centre Paediatric Network C8232/A25261.

Author contributions

TNA: study conception and design. BBB: analysis and interpretation of data, drafting of manuscript, and critical revision. DC: study conception and design, acquisition of data, analysis and interpretation of data, drafting of manuscript, and critical revision. ACP: study conception and design, analysis and interpretation of data, drafting of manuscript, and critical revision. MW: study conception and design, acquisition of data, analysis and interpretation of data, drafting of manuscript, and critical revision.

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical approval

All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. Informed consent was obtained from all individual participants involved in the study.

References

  • 1.Ward PS, Thompson CB. Metabolic reprogramming: a cancer hallmark even warburg did not anticipate. Cancer Cell. 2012;21(3):297–308. doi: 10.1016/j.ccr.2012.02.014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Panigrahy A, Krieger MD, Gonzalez-Gomez I, Liu X, McComb JG, Finlay JL, Nelson MD, Jr, Gilles FH, Bluml S. Quantitative short echo time 1H-MR spectroscopy of untreated pediatric brain tumors: preoperative diagnosis and characterization. Am J Neuroradiol. 2006;27(3):560–572. [PMC free article] [PubMed] [Google Scholar]
  • 3.Davies NP, Wilson M, Harris LM, Natarajan K, Lateef S, Macpherson L, Sgouros S, Grundy RG, Arvanitis TN, Peet AC. Identification and characterisation of childhood cerebellar tumours by in vivo proton MRS. NMR Biomed. 2008;21(8):908–918. doi: 10.1002/nbm.1283. [DOI] [PubMed] [Google Scholar]
  • 4.Wilson M, Gill SK, MacPherson L, English M, Arvanitis TN, Peet AC. Noninvasive detection of glutamate predicts survival in pediatric medulloblastoma. Clin Cancer Res. 2014;20(17):4532–4539. doi: 10.1158/1078-0432.CCR-13-2320. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Wilson M, Cummins CL, Macpherson L, Sun Y, Natarajan K, Grundy RG, Arvanitis TN, Kauppinen RA, Peet AC. Magnetic resonance spectroscopy metabolite profiles predict survival in paediatric brain tumours. Eur J Cancer. 2013;49(2):457–464. doi: 10.1016/j.ejca.2012.09.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Gill SK, Wilson M, Davies NP, MacPherson L, English M, Arvanitis TN, Peet AC. Diagnosing relapse in children’s brain tumors using metabolite profiles. Neuro Oncol. 2014;16(1):156–164. doi: 10.1093/neuonc/not143. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Davies NP, Wilson M, Natarajan K, Sun Y, MacPherson L, Brundler MA, Arvanitis TN, Grundy RG, Peet AC. Non-invasive detection of glycine as a biomarker of malignancy in childhood brain tumours using in vivo H-1 MRS at 1.5 Tesla confirmed by ex vivo, high-resolution magic-angle spinning NMR. Nmr in Biomedicine. 2010;23(1):80–87. doi: 10.1002/nbm.1432. [DOI] [PubMed] [Google Scholar]
  • 8.Panigrahy A, Nelson MD, Jr, Finlay JL, Sposto R, Krieger MD, Gilles FH, Bluml S. Metabolism of diffuse intrinsic brainstem gliomas in children. Neuro Oncol. 2008;10(1):32–44. doi: 10.1215/15228517-2007-042. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Harris LM, Davies NP, Macpherson L, Lateef S, Natarajan K, Brundler MA, Sgouros S, English MW, Arvanitis TN, Grundy RG, Peet AC. Magnetic resonance spectroscopy in the assessment of pilocytic astrocytomas. Eur J Cancer. 2008;44(17):2640–2647. doi: 10.1016/j.ejca.2008.08.012. [DOI] [PubMed] [Google Scholar]
  • 10.Bluml S, Margol AS, Sposto R, Kennedy RJ, Robison NJ, Vali M, Hung LT, Muthugounder S, Finlay JL, Erdreich-Epstein A, Gilles FH, Judkins AR, Krieger MD, Dhall G, Nelson MD, Asgharzadeh S. Molecular subgroups of medulloblastoma identification using noninvasive magnetic resonance spectroscopy. Neuro Oncol. 2016;18(1):126–131. doi: 10.1093/neuonc/nov097. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Kovanlikaya A, Panigrahy A, Krieger MD, Gonzalez-Gomez I, Ghugre N, McComb JG, Gilles FH, Nelson MD, Bluml S. Untreated pediatric primitive neuroectodermal tumor in vivo: quantitation of taurine with MR spectroscopy. Radiology. 2005;236(3):1020–1025. doi: 10.1148/radiol.2363040856. [DOI] [PubMed] [Google Scholar]
  • 12.Mullins PG, Chen H, Xu J, Caprihan A, Gasparovic C. Comparative reliability of proton spectroscopy techniques designed to improve detection of J-coupled metabolites. Magn Reson Med. 2008;60(4):964–969. doi: 10.1002/mrm.21696. [DOI] [PubMed] [Google Scholar]
  • 13.Opstad KS, Wright AJ, Bell BA, Griffiths JR, Howe FA. Correlations between in vivo (1)H MRS and ex vivo (1)H HRMAS metabolite measurements in adult human gliomas. J Magn Reson Imaging. 2010;31(2):289–297. doi: 10.1002/jmri.22039. [DOI] [PubMed] [Google Scholar]
  • 14.Choi C, Ganji SK, DeBerardinis RJ, Dimitrov IE, Pascual JM, Bachoo R, Mickey BE, Malloy CR, Maher EA. Measurement of glycine in the human brain in vivo by 1H-MRS at 3 T: application in brain tumors. Magn Reson Med. 2011;66(3):609–618. doi: 10.1002/mrm.22857. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Schubert F, Gallinat J, Seifert F, Rinneberg H. Glutamate concentrations in human brain using single voxel proton magnetic resonance spectroscopy at 3 Tesla. Neuroimage. 2004;21(4):1762–1771. doi: 10.1016/j.neuroimage.2003.11.014. [DOI] [PubMed] [Google Scholar]
  • 16.Snyder J, Wilman A. Field strength dependence of PRESS timings for simultaneous detection of glutamate and glutamine from 1.5 to 7T. J Magn Reson. 2010;203(1):66–72. doi: 10.1016/j.jmr.2009.12.002. [DOI] [PubMed] [Google Scholar]
  • 17.Hurd R, Sailasuta N, Srinivasan R, Vigneron DB, Pelletier D, Nelson SJ. Measurement of brain glutamate using TE-averaged PRESS at 3T. Magn Reson Med. 2004;51(3):435–440. doi: 10.1002/mrm.20007. [DOI] [PubMed] [Google Scholar]
  • 18.Kim H, Wild JM, Allen PS. Strategy for the spectral filtering of myo-inositol and other strongly coupled spins. Magn Reson Med. 2004;51(2):263–272. doi: 10.1002/mrm.10697. [DOI] [PubMed] [Google Scholar]
  • 19.Lei H, Peeling J. Simultaneous spectral editing for gamma-aminobutyric acid and taurine using double quantum coherence transfer. J Magn Reson. 2000;143(1):95–100. doi: 10.1006/jmre.1999.1958. [DOI] [PubMed] [Google Scholar]
  • 20.Thompson RB, Allen PS. A new multiple quantum filter design procedure for use on strongly coupled spin systems found in vivo: its application to glutamate. Magn Reson Med. 1998;39(5):762–771. doi: 10.1002/mrm.1910390514. [DOI] [PubMed] [Google Scholar]
  • 21.Zarinabad N, Wilson M, Gill SK, Manias KA, Manias KA, Davies NP, Peet AC. Multiclass imbalance learning: Improving classification of pediatric brain tumors from magnetic resonance spectroscopy. Magn Reson Med. 2016 doi: 10.1002/mrm.26318. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Schulte RF, Lange T, Beck J, Meier D, Boesiger P. Improved two-dimensional J-resolved spectroscopy. NMR Biomed. 2006;19(2):264–270. doi: 10.1002/nbm.1027. [DOI] [PubMed] [Google Scholar]
  • 23.Ryner LN, Sorenson JA, Thomas MA. 3D localized 2D NMR spectroscopy on an MRI scanner. J Magn Reson B. 1995;107(2):126–137. doi: 10.1006/jmrb.1995.1068. [DOI] [PubMed] [Google Scholar]
  • 24.Thomas MA, Ryner LN, Mehta MP, Turski PA, Sorenson JA. Localized 2D J-resolved 1H MR spectroscopy of human brain tumors in vivo. J Magn Reson Imaging. 1996;6(3):453–459. doi: 10.1002/jmri.1880060307. [DOI] [PubMed] [Google Scholar]
  • 25.Govindaraju V, Young K, Maudsley AA. Proton NMR chemical shifts and coupling constants for brain metabolites. NMR Biomed. 2000;13(3):129–153. doi: 10.1002/1099-1492(200005)13:3&#x0003c;129::AID-NBM619&#x0003e;3.0.CO;2-V. [DOI] [PubMed] [Google Scholar]
  • 26.Soher B, Semanchuk P, Todd D, Steinberg J, Young K VeSPA: Integrated applications for RF pulse design, spectral simulation and MRS data analysis. In: International Society for Magnetic Resonance in Medicine, Montreal, Quebec, Canada, 2011 [DOI] [PMC free article] [PubMed]
  • 27.Edden RAE, Barker PB. If J doesn’t evolve, it won’t J-resolve: j-PRESS with bandwidth-limited refocusing pulses. Magn Reson Med. 2011;65(6):1509–1514. doi: 10.1002/mrm.22747. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Wilson M, Reynolds G, Kauppinen RA, Arvanitis TN, Peet AC. A constrained least-squares approach to the automated quantitation of in vivo (1)H magnetic resonance spectroscopy data. Magn Reson Med. 2011;65(1):1–12. doi: 10.1002/mrm.22579. [DOI] [PubMed] [Google Scholar]
  • 29.Hoch JC, Stern AS. NMR data processing. New York City: Wiley; 1996. [Google Scholar]
  • 30.Hancu I. Optimized glutamate detection at 3T. J Magn Reson Imaging. 2009;30(5):1155–1162. doi: 10.1002/jmri.21936. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Ganji SK, Banerjee A, Patel AM, Zhao YD, Dimitrov IE, Browning JD, Brown ES, Maher EA, Choi C. T2 measurement of J-coupled metabolites in the human brain at 3T. NMR Biomed. 2012;25(4):523–529. doi: 10.1002/nbm.1767. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Kohe SE, Bennett CD, Gill SK, Wilson M, McConville C, Peet AC. Metabolic profiling of the three neural derived embryonal pediatric tumors retinoblastoma, neuroblastoma and medulloblastoma, identifies distinct metabolic profiles. Oncotarget. 2018;9(13):11336–11351. doi: 10.18632/oncotarget.24168. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Babourina-Brooks B, Kohe S, Gill SK, Wilson M, MacPherson L, Davies NP, Peet AC Glycine, a marker of survival in paediatric brain tumours measured with non-invasive Magnetic Resonance Spectroscopy: A five-year survival analysis. In: 25th Annual Meeting of the International Society of Magnetic Resonance in Medicine, Honolulu, HA, USA, 2017
  • 34.Wilson M, Davies NP, Brundler MA, McConville C, Grundy RG, Peet AC. High resolution magic angle spinning 1H NMR of childhood brain and nervous system tumours. Mol Cancer. 2009;8:6. doi: 10.1186/1476-4598-8-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Oz G, Alger JR, Barker PB, Bartha R, Bizzi A, Boesch C, Bolan PJ, Brindle KM, Cudalbu C, Dincer A, Dydak U, Emir UE, Frahm J, Gonzalez RG, Gruber S, Gruetter R, Gupta RK, Heerschap A, Henning A, Hetherington HP, Howe FA, Huppi PS, Hurd RE, Kantarci K, Klomp DW, Kreis R, Kruiskamp MJ, Leach MO, Lin AP, Luijten PR, Marjanska M, Maudsley AA, Meyerhoff DJ, Mountford CE, Nelson SJ, Pamir MN, Pan JW, Peet AC, Poptani H, Posse S, Pouwels PJ, Ratai EM, Ross BD, Scheenen TW, Schuster C, Smith IC, Soher BJ, Tkac I, Vigneron DB, Kauppinen RA. Clinical proton MR spectroscopy in central nervous system disorders. Radiology. 2014;270(3):658–679. doi: 10.1148/radiol.13130531. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Thomas MA, Hattori N, Umeda M, Sawada T, Naruse S. Evaluation of two-dimensional L-COSY and JPRESS using a 3 T MRI scanner: from phantoms to human brain in vivo. NMR Biomed. 2003;16(5):245–251. doi: 10.1002/nbm.825. [DOI] [PubMed] [Google Scholar]
  • 37.Kuhn B, Dreher W, Leibfritz D, Heller M. Homonuclear uncoupled H-1-spectroscopy of the human brain using weighted accumulation schemes. Magn Reson Imaging. 1999;17(8):1193–1201. doi: 10.1016/S0730-725X(99)00027-2. [DOI] [PubMed] [Google Scholar]
  • 38.Jensen JE, Auerbach RP, Pisoni A, Pizzagalli DA. Localized MRS reliability of in vivo glutamate at 3 T in shortened scan times a feasibility study. NMR Biomed. 2017 doi: 10.1002/nbm.3771. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Fuchs A, Boesiger P, Schulte RF, Henning A. ProFit Revisited. Magn Reson Med. 2014;71(2):458–468. doi: 10.1002/mrm.24703. [DOI] [PubMed] [Google Scholar]
  • 40.Thomas MA, Nagarajan R, Huda A, Margolis D, Sarma MK, Sheng K, Reiter RE, Raman SS. Multidimensional MR spectroscopic imaging of prostate cancer in vivo. NMR Biomed. 2014;27(1):53–66. doi: 10.1002/nbm.2991. [DOI] [PubMed] [Google Scholar]
  • 41.Wilson NE, Iqbal Z, Burns BL, Keller M, Thomas MA. Accelerated five-dimensional echo planar J-resolved spectroscopic imaging: implementation and pilot validation in human brain. Magn Reson Med. 2016;75(1):42–51. doi: 10.1002/mrm.25605. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials


Articles from Magma (New York, N.y.) are provided here courtesy of Springer

RESOURCES