. 2018 Dec 7;2(4):yty117. doi: 10.1093/ehjcr/yty117

Cusp thrombosis of a self-expandable sutureless aortic valve treated by valve-in-valve transcatheter aortic valve implantation procedure: case report

Ulvi Cenk Oezpeker 1,, Gudrun Feuchtner 2, Nikolaos Bonaros 1
Editors: Georg Goliasch, Dejan Milasinovic, Riccardo Liga, Mohammed Akhtar, Peysh Patel
PMCID: PMC6426046  PMID: 31020193

Abstract

Background

Surgical implantation rates of bioprosthetic valves, especially the use of sutureless or rapid deployment valves, as well as the advent of transcatheter valve implantation (TAVR) have increased during the last decades mainly due to their excellent haemodynamic and clinical results. One common characteristic of all bioprosthetic types of surgical aortic valve replacement (SAVR) and TAVR is the risk of early degeneration, which leads to valve-dysfunction and is associated with higher rates of valve reinterventions. Recent studies have demonstrated that cusp thrombosis may play a role in early valve dysfunction. This case report is, to our knowledge, the first documentation on a successful treatment of early aortic valve (AV) degeneration of a sutureless AV thrombosis with a valve-in-valve (ViV) TAVR implantation.

Case summary

A 77 years old woman was re-evaluated from the heart-team, which considered the following characteristics: severe impairment of mobility and frailty with an STS-score of 10.01% and a EuroSCORE II of 6.9%. Due to the high surgical risk for SAVR, we decided to perform a ViV-TAVR using a balloonexpandable bioprosthesis. The procedure was performed via transfemoral access under general anaesthesia using a 23 mm Edwards-Sapien 3 bioprosthesis without balloon-valvuloplasty and with nominal-volume dilation under rapid-pacing.

Discussion

The differentiation of bioprosthesis valve thrombosis, and hypoattenuating leaflet thickening vs. structural valve degeneration can be difficult, and a multimodality imaging approach, comprising trans-thoracic echocardiogram, transoesophageal echocardiography and computed tomography, useful. These investigations are very important to decide the right strategy of surgical valve replacement vs. TAVR.

Keywords: Case report, Sutureless valve, Bioprosthesis thrombosis, Valve in valve TAVI


Learning points

  • Cusp thrombosis of a bioprosthesis may also affect sutureless valves and is probably related to an elliptic expansion of the prosthesis.

  • Transcatheter valve-in-valve implantation can be used as a therapeutic option in case of failure of anticoagulation treatment.

  • A combination of haemodynamic measurements and multimodality imaging is essential for the diagnosis of bioprosthetic cusp thrombosis and the evaluation of treatment success.

Introduction

Surgical implantation rates for bioprosthetic valves, especially sutureless or rapid deployment valves, as well as the advent of transcatheter valve implantation (TAVI)1 have increased during the last decades mainly due to their excellent haemodynamic and clinical results.2 One common characteristic of all bioprosthetic types of surgical aortic valve replacement (SAVR) and TAVI is the risk of early degeneration,3 which leads to valve dysfunction and is associated with higher reintervention rates. Recent studies have demonstrated that bioprosthetic valve thrombosis (BVT) may play a role in early valve dysfunction. This case report is, to our knowledge, the first documented report on therapy of a BVT of a sutureless valve (SV) with a valve-in-valve (ViV) TAVI.

Timeline

Symptom onset November 2015
Diagnosis of aortic stenosis January 2016 to February 2016
Surgical aortic valve replacement March 2016
Recurrence of heart failure symptoms February 2017
Cusp thrombosis March 2017
Failure of pharmacological thrombus resolution May 2017
Valve-in-valve transcatheter valve implantation June 2017
Last follow-up March 2018

Patient information

A 77-year-old woman presented at our department for treatment of a highly symptomatic low-flow (SVI: 23 mL/m2/1.73 m2 body surface area), low-gradient (mean: 38 mmHg) AV stenosis (AV area 0.52 cm2) with preserved ejection fraction (left-ventricular ejection fraction (LVEF): 58%), and a very hypertrophic left ventricular myocardium. In addition, she suffered from secondary pulmonary hypertension (catheter-measured systolic pulmonary artery pressure 50 mmHg). At that time she was dyspnoeic at New York Heart Association (NYHA) functional status III. Additional comorbidities included arterial hypertension, insulin-dependent diabetes mellitus, permanent pacemaker implantation due to sick-sinus syndrome and reported weight loss of 8 kg since the onset of symptoms.

Due to moderate frailty with a 3.8% STS score and a 1.3% EuroSCORE II she was assigned for minimally invasive SAVR and received a self-expandable SV [PERCEVAL-S®, sizeM, (Livanova, Saluggia, Italy)]. Both the intra- and post-operative course were uneventful. The initial anticoagulation protocol was maintained with low-molecular weight heparin s.c. at a doses of 2× bodyweight in milligram with anti-Xa monitoring and addition of aspirin 100 mg/d, followed by acenocoumarol (INR target 2.0–3.0). Discharge echocardiography showed a mean pressure gradient (MPG) across the AV of 7 mmHg without AV regurgitation. She was readmitted 10 months after the index procedure with symptoms of recurrent heart failure, dyspnoea, pulmonary congestion (oxygen saturation 94%), peripheral oedema, mild end-diastolic murmur, paced heart rate 78 b.p.m., NYHA IV, and increased NT-proBNP.

Diagnostic assessment

Transthoracic echocardiography demonstrated recurrent AV stenosis (AV area 0.92 cm2), with a MPG of 12 mmHg with severe regurgitation and a preserved LV-EF (Figure 1A). Transoesophageal echocardiography (TOE) showed an immobile right-coronary cusp (RCC) due to BVT (Figure 1B). Cardiac computed tomography (CT) scan revealed a 12 × 4.9 mm thrombus at the ventricular aspect of the RCC and the right non-coronary commissure (Figure 2) with an elliptic expansion shape (22.6 × 18.8 mm) of the bioprosthesis at the annulus level (Figure 3A). Any relevant coronary stenosis (>50%) was excluded by angiography prior to SAVR and coronary CT scan prior to ViV-TAVI.

Figure 1.

Figure 1

(A and B) Transoesophageal echocardiography with aortic valve gradient and thrombus pre-valve-in-valve transcatheter valve implantation.

Figure 2.

Figure 2

Aortic valve thrombus of PERCEVAL-S® seen by computed tomography scan.

Figure 3.

Figure 3

(A and B) Computed tomography scan pre- (elliptical shape A) and post-valve-in-valve transcatheter valve implantation (circular shape B).

Acquired hypercoagulable disorders or systemic lupus erythematodes were excluded. Neoplastic diseases with paraneoplastic syndrome were excluded in the CT scan, and the patient’s initial gastrointestinal symptoms were thoroughly investigated with gastroscopy, colonoscopy, and intrauterine ultrasound.

Interventions

Pharmacological therapy for thrombus resolution was initiated with systemic heparinization (aPTT 50–60 s) followed by acenocoumarol (target INR: 2.5–3.5) for 3 months, according to the current guidelines/recommendations.4,5 Because of further clinical deterioration and failure of thrombus resolution, the patient was considered for redo-SAVR or ViV-TAVI. She was re-evaluated by the heart-team, which considered the following characteristics: severe frailty with a 10.01% STS score and a 6.9% EuroSCORE II. Due to this high surgical risk, we performed a transfemoral ViV-TAVI under general anaesthesia using a 23 mm Edwards-Sapien 3 bioprosthesis. The lower stent row of the TAVI was positioned under the lowest part of the SV frame to achieve stability of the implanted valve and to engage the thrombus. Intraoperative valve function was excellent showing minimal residual transvalvular gradient (Figure 4) without paravalvular regurgitations on TOE, confirmed by aortic root angiography. She was extubated immediately after the procedure with an uneventful post-operative course. She was discharged home on Day 7 after TAVI with the following anticoagulation regime: acenocoumarol (target INR 2–3) and 100 mg aspirin once per day.

Figure 4.

Figure 4

(A and B) Valve-in-valve transcatheter valve implantation angiography and intra-OP gradient post-transcatheter valve implantation.

Follow-up and outcome

Post-operative echocardiography 5 days, 6, and 12 weeks after TAVI showed regular AV function with a constant minimal MPG (≤10 mmHg) and excluded valvular regurgitation. Computed tomography scan on Day 5 after TAVI confirmed perfect anatomical positioning of the AV with a completely circular shape of the annulus [diameter 20.4 × 20.1 mm (Figure 3B)]. The course of NT-proBNP levels was 2260 ng/L before initial aortic valve replacement, 2892 ng/L before the ViV procedure and 1475 ng/L immediately thereafter (normal range <376 ng/L). The calculated systolic PAP decreased from 40 mmHg before ViV-TAVI to 30 mmHg. The patient presented with NYHA I 1 month after ViV-TAVI.

Discussion

To our knowledge this is the first report of a ViV-TAVI in a patient with a BVT of a self-expandable SV. There are sparse reports of documented SV thrombosis.6,7 The prevalence of BVT after SAVR or TAVI ranges between 7% and 23%.8–11 Notably, some studies reported no association with clinical symptoms9,12 and revealed complete resolution of BVT after vitamin K antagonist treatment. Egbe et al.13 identified a 50% increase in MPG over 5 years as an independent predictor of BVT in 397 patients.

Patient deterioration is probably associated with a combination of stenosis and regurgitation due to BVT as well as with the cardiac low-flow condition. The latter is a real challenge in heart failure management with preserved EF and is probably even more challenging when setting continuous right ventricular pacemaker stimulation resulting in a left bundle branch block. In our experience, these patients are extremely difficult to treat, as they have a high perioperative risk and often return with recurrent symptoms.

The evolution of cardiac CT14 and the advent of three-dimensional TOE have provided appropriate imaging evidence of hypoattenuating leaflet thrombosis, and the reported data are steadily increasing.

There are several technical aspects to be considered when planning and performing ViV-TAVI. First the exact brand, size and position of the index bioprosthesis volume rendering by CT angiography (Figure 3A) should be known for precise planning of the new prosthesis type and size. Dysfunctioning SV are considered a special condition, as they are self-expanding and the thrombosis may be associated with the initial anatomical valve position. Remaining calcifications of the native AV or a severely elliptic annulus may prevent proper expansion of the valve and thus pave the way for turbulence with increased shear stress or low-flow phenomena within the SV, which is prone to thrombosis. In our case, the SV was correctly sized and positioned. The form of the PERCEVAL-S® probably makes the use of high-profile TAVI less beneficial due to its difficult positioning (valve trapping or stent engagement) during implantation. We opted for a balloon-expandable valve to increase the circularity of the annulus after the procedure and to avoid tangling with the high frame of the PERCEVAL-S®.

Moreover, the exact position of the thrombus and its mobility should be thoroughly examined due to the risk of embolism inherent in wire manipulations, which should not be underestimated during the procedure.

Conclusion

To summarize, we present a case of BVT of a SV treated by ViV-TAVI as an alternative to high-risk SAVR with excellent results in the short-term follow-up.15 However, mid- and long-term results are still outstanding, making this therapy’s perspective unclear. This strategy is feasible when specific valve type and anatomical conditions are exhaustively considered.

Slide sets: A fully edited slide set detailing this case and suitable for local presentation is available online as Supplementary data.

Consent: The author/s confirm that written consent for submission and publication of this case report including image(s) and associated text has been obtained from the patient in line with COPE guidance.

Conflict of interest: none declared.

Supplementary Material

Supplementary Data

References

  • 1. Mohr FW, Holzhey D, Möllmann H, Beckmann A, Veit C, Figulla HR, Cremer J, Kuck K-H, Lange R, Zahn R, Sack S, Schuler G, Walther T, Beyersdorf F, Böhm M, Heusch G, Funkat A-K, Meinertz T, Neumann T, Papoutsis K, Schneider S, Welz A, Hamm CW.. The German Aortic Valve Registry: 1-year results from 13,680 patients with aortic valve disease. Eur J Cardiothorac Surg 2014;46:808–816. [DOI] [PubMed] [Google Scholar]
  • 2. Dunning J, Gao H, Chambers J, Moat N, Murphy G, Pagano D, Ray S, Roxburgh J, Bridgewater B.. Aortic valve surgery: marked increases in volume and significant decreases in mechanical valve use—an analysis of 41,227 patients over 5 years from the Society for Cardiothoracic Surgery in Great Britain and Ireland National database. J Thorac Cardiovasc Surg 2011;142:776–782.e3. [DOI] [PubMed] [Google Scholar]
  • 3. Head SJ, Celik M, Kappetein AP.. Mechanical versus bioprosthetic aortic valve replacement. Eur Heart J 2017;38:2183–2191. [DOI] [PubMed] [Google Scholar]
  • 4. Jander N, Kienzle RP, Kayser G, Neumann FJ, Gohlke-Baerwolf C, Minners J.. Usefulness of phenprocoumon for the treatment of obstructing thrombus in bioprostheses in the aortic valve position. Am J Cardiol 2012;109:257–262. [DOI] [PubMed] [Google Scholar]
  • 5. Nishimura RA, Otto CM, Bonow RO, Carabello BA, Erwin JP 3rd, Fleisher LA, et al. 2017 AHA/ACC focused update of the 2014 AHA/ACC Guideline for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines. Circulation 2017;135:e1159–e1195. [DOI] [PubMed] [Google Scholar]
  • 6. Votsch A, Weihs W, Asslaber M, Dapunt O.. Perceval sutureless valve dysfunction caused by valvular thrombosis. Ann Thorac Surg 2016;102:e309–e311. [DOI] [PubMed] [Google Scholar]
  • 7. Yanagawa B, Verma S, Mazer CD.. Subclinical valve thrombosis in sutureless bioprosthetic valves. J Am Heart Assoc 2017;6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Marwan M, Mekkhala N, Göller M, Röther J, Bittner D, Schuhbaeck A, Hell M, Muschiol G, Kolwelter J, Feyrer R, Schlundt C, Achenbach S, Arnold M.. Leaflet thrombosis following transcatheter aortic valve implantation. J Cardiovasc Comput Tomogr 2018;12:8–13. [DOI] [PubMed] [Google Scholar]
  • 9. Pache G, Schoechlin S, Blanke P, Dorfs S, Jander N, Arepalli CD, Gick M, Buettner H-J, Leipsic J, Langer M, Neumann F-J, Ruile P.. Early hypo-attenuated leaflet thickening in balloon-expandable transcatheter aortic heart valves. Eur Heart J 2016;37:2263–2271. [DOI] [PubMed] [Google Scholar]
  • 10. Chakravarty T, Søndergaard L, Friedman J, De Backer O, Berman D, Kofoed KF, Jilaihawi H, Shiota T, Abramowitz Y, Jørgensen TH, Rami T, Israr S, Fontana G, de Knegt M, Fuchs A, Lyden P, Trento A, Bhatt DL, Leon MB, Makkar RR, Ramzy D, Cheng W, Siegel RJ, Thomson LM, Mangat G, Hariri B, Sawaya FJ, Iversen HK.. Subclinical leaflet thrombosis in surgical and transcatheter bioprosthetic aortic valves: an observational study. Lancet 2017;389:2383–2392. [DOI] [PubMed] [Google Scholar]
  • 11. Makkar RR, Fontana G, Jilaihawi H, Chakravarty T, Kofoed KF, De Backer O, Asch FM, Ruiz CE, Olsen NT, Trento A, Friedman J, Berman D, Cheng W, Kashif M, Jelnin V, Kliger CA, Guo H, Pichard AD, Weissman NJ, Kapadia S, Manasse E, Bhatt DL, Leon MB, Søndergaard L.. Possible subclinical leaflet thrombosis in bioprosthetic aortic valves. N Engl J Med 2015;373:2015–2024. [DOI] [PubMed] [Google Scholar]
  • 12. Leetmaa T, Hansson NC, Leipsic J, Jensen K, Poulsen SH, Andersen HR, et al. Early aortic transcatheter heart valve thrombosis: diagnostic value of contrast-enhanced multidetector computed tomography. Circ Cardiovasc Interv 2015;8. [DOI] [PubMed] [Google Scholar]
  • 13. Egbe AC, Pislaru SV, Pellikka PA, Poterucha JT, Schaff HV, Maleszewski JJ, Connolly HM.. Bioprosthetic valve thrombosis versus structural failure: clinical and echocardiographic predictors. J Am Coll Cardiol 2015;66:2285–2294. [DOI] [PubMed] [Google Scholar]
  • 14. Gunduz S, Ozkan M, Kalcik M, Gursoy OM, Astarcioglu MA, Karakoyun S, et al. Sixty-four-section cardiac computed tomography in mechanical prosthetic heart valve dysfunction: thrombus or pannus. Circ Cardiovasc Imaging 2015;8. [DOI] [PubMed] [Google Scholar]
  • 15. Dvir D, Webb J, Brecker S, Bleiziffer S, Hildick-Smith D, Colombo A, Descoutures F, Hengstenberg C, Moat NE, Bekeredjian R, Napodano M, Testa L, Lefevre T, Guetta V, Nissen H, Hernández J-M, Roy D, Teles RC, Segev A, Dumonteil N, Fiorina C, Gotzmann M, Tchetche D, Abdel-Wahab M, De Marco F, Baumbach A, Laborde J-C, Kornowski R.. Transcatheter aortic valve replacement for degenerative bioprosthetic surgical valves: results from the global valve-in-valve registry. Circulation 2012;126:2335–2344. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplementary Data

Articles from European Heart Journal: Case Reports are provided here courtesy of Oxford University Press

RESOURCES