Abstract
We propose a noninvasive, self-diagnostic device that enables safe tear collection and glucose measurement. The device described herein was manufactured by tight assembly of a lid for tear collection in conjunction with a strip-type glucose sensor. The lid was designed to be in contact with the inferior palpebral conjunctiva for tear collection and was thus designed to possess a proper contact area and rounded boundaries to avoid eye tissue damage. For the strip-type glucose sensor, we employed a commercially available electrochemical sensor (Accu-Chek test strips), which was modified to reduce the volume of the reaction chamber (0.4 μl) for a small amount of collected tear fluid. When tested with in vivo animal models, the device was able to collect tear fluid in a relatively short time (<2 s) without causing eye tissue damage, and the device allowed the collected tear fluid to be delivered to the sensor for measurement of tear glucose concentrations. The blood glucose concentrations estimated with the tear glucose concentrations obtained with the device exhibited a high correlation with those actually measured with a clinically available glucometer (R2 = 0.9617).
Subject terms: Diabetes, Diagnosis
Introduction
Diabetes has been reported to be largely influenced by genetic factors1,2 and thus, early diagnosis of prediabetes, followed by proper treatment, has been shown to prevent progression to Type 2 diabetes, hence improving the quality of patients’ life3,4. In this sense, albeit healthy, the people with a high chance of getting diabetes can benefit from watching the blood glucose level early; however, with the currently-available strategies, blood needs to be withdrawn by pricking the fingertip with a needle for self-assessment of blood glucose5,6. This relatively invasive, painful measurement process often results in discomfort, even causing massive callous formation or loss of sensibility at the site of multiple punctures7. Therefore, this was considered one of the major reasons leading to an insufficient number of blood glucose tests even for diabetic patients8.
Therefore, much effort has been devoted to developing a less invasive, more convenient strategy for glucose measurement9,10. Among them, glucose measurement through tear fluid has drawn a great deal of interest. The glucose level in tears is known to have a demonstrable correlation with that in blood11, and tears can be collected relatively easily from the anterior space of the eye, which can be less invasive compared with conventional blood withdrawal through skin punctures. Therefore, many different measurement strategies, such as biological12, electrochemical13,14, spectroscopic15,16 and colorimetric17 measurements, have been extensively investigated for a more accurate measurement of glucose concentrations in tears.
However, those studies have not yet fully considered a practical methodology in the context of self-diagnosis. First, the two-step measurement procedure is not convenient for patients: tear fluid is collected separately and is then transferred and loaded to the analytical instrument for measurement. A relatively large volume of approximately 100 μl of tear fluid is often collected, which requires a long time of tear collection, as only approximately 6.5 μl of tear fluid is available at the preocular surface and the average rate of tear production is very low (1 μl min−1)18,19. Moreover, tear fluid may be collected using a glass capillary or water-absorbing paper; however, this technique has not yet been fully proven to be appropriate due to its invasiveness to the eye when used by the patients themselves. Therefore, these conventional strategies can still cause uncomfortable eye irritation or even damage and this can also lead to inaccurate measurements due to reflex tear production or an irrelevant increase in tear glucose due to the damaged tissue in the anterior surface of the eye8,20–23.
Therefore, in this work, we propose a device for glucose measurement in tears (i.e., a tear-glucose device), meeting the following design criteria: for the convenience of the patients, the tear-glucose device needs to be equipped with both a tear-collector and a sensor as a single entity, thereby allowing for concurrent collection and measurement of tear fluids. In this device, a tip needs to be designed minimally invasive as it must be in contact with the preocular surface (i.e., an ocular tip). Tear collection should be conducted for a short time, hence resulting in a small volume of collected tear fluid. The sensor also needs to be sensitive enough to accurately measure glucose concentrations in a small quantity of tears. At the same time, we also proposed a concept of device design that could be quick to apply to clinical applications and easy to manufacture.
To prepare such a tear-glucose device, we prepared an assembled entity consisting of a lid and a strip-type glucose sensor. We engineered the lid to consist of a minimally invasive ocular tip and a hollow body with a slit, where the strip-type glucose sensor could be inserted for tight assembly. In the ocular tip, a micron-size inlet was prepared, where after its contact with the preocular surface, a tear could be absorbed and delivered to reach the sensor inserted in the lid. For the strip-type glucose sensor, we employed a commercially-available test strip (Accu-Chek Performa, Roche Diagnostics, Switzerland), which is known to measure glucose levels via electrochemical means with high accuracy and selectivity24. However, in this work, we modified this test strip to be suitable specifically for the collection and measurement of a small quantity of tear fluid. Therefore, the lid herein possessing a simple structure would not be difficult to manufacture. The glucose test strip has been already widely used in clinical settings and thus, its manufacturing process is already well developed for a large-scale production.
The tear-glucose device herein was first tested at the range of physiological glucose concentrations present in tears under in vitro simulated environments. To test in vivo usability, the device was placed in contact with the inferior palpebral conjunctiva (IPC) of rabbit eyes to collect and measure tear fluids, since the IPC can be considered less sensitive compared with other regions in the preocular space25. We also validated the correlation between glucose concentrations measured in blood and tears. After application of our device, the IPC tissue was also examined for the purpose of safety evaluation.
Results
Lid
We prepared the lid to serve as an interface between the eye surface and glucose sensor to safely collect and deliver tear fluids. To achieve this, we first designed the lid to consist of two distinct parts: the ocular tip and the hollow body using SolidWorks (Dassault Système SOLIDWORKS Corp., USA). As shown in Fig. 1(a), the ocular tip was shaped to be circular (2 mm in diameter) with rounded boundaries, which has been reported to not damage the IPC tissue during contact according to a previous study conducted by our group26. In the ocular tip, a small inlet (1 × 0.5 × 0.4 mm, W × H × L) was prepared to be able to absorb tear fluids via capillary action after contact with the IPC and to minimize the dead volume of tear fluid trapped inside of the inlet. The hollow body was designed to possess a slit (7.4 × 0.55 mm, W × H) for insertion of the strip-type glucose sensor used herein. The opening (1 × 1 mm, W × H) was also made at the top of the hollow body so that tear fluids could be smoothly absorbed and transferred to the glucose sensor. Figure 1(b) shows the lid fabricated with a 3D printer (ProJet 3500 HD, 3D Systems, USA), using the printing material Visijet M3 Crystal, certified by USP (United States Pharmacopeia) Class VI.
Strip-type glucose sensor
To prepare a strip-type sensor, we employed the commercially available Accu-Chek test strip and modified it for the purpose of glucose measurement in tears. This test strip was equipped with an electrochemical sensor with a low limit of quantification (<45 μM) and high selectivity for glucose measurement27. In this sensor, collecting a fixed volume of fluid sample is important for accurate measurement of glucose, as the electrical current that varies according to the rate of glucose oxidation is also influenced by the volume of fluid in the reaction chamber. The Accu-Chek test strip is originally designed to measure glucose levels in the blood, and thus the reaction chamber with the electrodes is shaped to hold 0.8 μl of fluid.
In this work, we attempted to minimize the volume of collected tear fluid, so we modified the test trip mainly to decrease the volume of the reaction chamber. During this process, the major electrodes for glucose measurement needed to be kept intact. In the reaction chamber of the Accu-Chek test strip, four different electrodes were embedded; the counter, reference, working and fill-sufficiency electrodes. For glucose measurement, the counter, reference and working electrodes must be kept intact. However, the fill-sufficiency electrodes for detection of the presence of liquid might not be necessary to test the prototype device prepared in this work. Given these facts, as shown in Fig. 2, we first cut out the frontal end of the strip to the extent where the counter, reference and working electrodes were not affected, to reduce the length of the reaction chamber. We also rounded off the sharp edges in the frontal end to avoid possible damage to the IPC tissue. Then, we punched out the other end of the reaction chamber, removing the fill-sufficiency electrodes, to further reduce its volume. Thus, after modification, the resulting strip-type glucose sensor became equipped with a reaction chamber containing a volume reduced by half to 0.4 μl.
Tear-glucose devices
The tear-glucose device herein described was prepared by simple assembly of the lid and the modified strip-type glucose sensor, as shown in Fig. 3. After insertion in the slit, the strip-type sensor was tightly fit, and the reaction chamber was properly aligned to the inlet prepared in the ocular tip of the lid. After that, the counter, reference and working electrodes were each connected to the counterpart connection in a potentiostat to apply 150 mV of DC voltage and measure the electrical current. In this work, we used a two-electrode system and thus, the counter and reference electrodes were connected as a single cathode while the working electrode was connected separately as an anode (Fig. 3).
To test the feasibility of the tear-glucose device herein, in vitro calibration was conducted with a glucose solution prepared in phosphate-buffered saline (PBS) with varying concentrations of 0, 0.01, 0.05, 0.1, 0.2, 0.4 and 0.8 mM. As shown in Fig. 4, a highly linear relationship was obtained between the glucose concentration and electrical current (R2 > 0.9864), indicating that the tear-glucose device herein described was valid to measure the in vivo glucose concentration in tear fluid, which was reported to range from 0~0.6 mM28.
In vivo evaluation
To evaluate the feasibility of the tear-glucose device herein described, in vivo experiments were conducted using rabbits, where the blood glucose concentration was elevated through anesthetization13,14,28. In this work, the tear glucose concentrations were measured with the tear-glucose device and the blood glucose concentrations were measured with a blood glucometer approved for clinical use (Accu-Chek Performa, Roche Diagnostics, Switzerland) at scheduled times of 15, 30, 45, and 60 min after anesthetization.
Figure 5(a) shows the plot of the tear and blood glucose concentrations obtained from all eight rabbits tested in this work, showing a linear relation between tear and blood glucose concentrations. However, the data points were spread to some extent and hence the linear correlation appeared not to be very strong (R2 = 0.7640, df = 30, P < 0.05). This discrepancy was also observed in previous reports and was ascribed to the inherent differences among tested animals13,14. Therefore, when the data points averaged at each time of measurement (i.e., at every 15 min) were plotted13, as shown in Fig. 5(b), a strong linear correlation with a statistical significance was obtained (R2 = 0.9617, df = 2, P < 0.05).
In this study, we sought to show that our tear-glucose device herein described can estimate the blood glucose concentration based on the glucose concentration measured in tears. To evaluate this, the estimated and actual blood glucose concentrations were plotted in a Clarke error grid29,30. The estimated blood glucose concentrations were calculated from the tear glucose concentrations using the equation obtained from the graph in Fig. 5(b). The actual blood glucose concentrations were measured from the blood using a glucometer approved for clinical use. As shown in Fig. 6, all data points are in the zones A and B, suggesting that the blood glucose concentrations estimated with the tear-glucose device herein are highly correlated and possibly interchangeable with those conducted with the conventional blood glucose measurement method. Approximately 81.2% of the data points are in the zone A, implying a relatively higher accuracy of blood glucose measurement with the tear-glucose device described herein.
To evaluate the safety of the tear-glucose device, we also examined the IPC tissues at the end point of the experiments, i.e., after each of the four times of contact of the tear-glucose device on the IPC, where the eye surface was stained with a fluorescein solution to assess possible tissue damage31. As shown in Fig. 7, after contact of the tear-glucose device, the IPC tissues were not observed to be stained, as seen with the nontreated, negative control eye, indicating no epithelial tissue damage. In contrast, when the IPC was in contact with a glass capillary tube, conventionally used for tear collection, an apparent sign of staining was observed due to damage of the IPC tissue.
Discussion
Tear glucose measurement has been suggested as a potential, noninvasive strategy of blood glucose prediction8,10,32. Most of the previous studies focused on developing sensors with a higher accuracy since the glucose concentration in tears is known to be lower than that in the blood13,33,34. However, to our knowledge, studies on devices for practical, self-diagnostic applications is scarce. In this context, a device allowing concurrent tear collection and glucose measurement could be useful and convenient for users. Such a device would be more advantageous if the measurement could be reliable even with a small quantity of tear fluid as this would allow for a short time of tear collection, hence less invasiveness on the preocular tissues.
Therefore, we proposed the tear-glucose device herein as a noninvasive self-diagnostic tool for prediction of blood glucose levels. The lid in the device was shown to not damage the eye tissues due to its design with a proper contact area and rounded boundaries without sharp edges (Fig. 7)26. Although the lid herein was made of a slightly hydrophobic material (Visijet M3 Crystal, 3D Systems, USA), the micron size inlet still allowed for a strong capillary pressure, hence safe and efficient absorption of tear fluid, which could be rapidly transferred to fill the reaction chamber with a reduced volume (0.4 µl) in the strip-type glucose sensor (Fig. 2). With a lid made of a more hydrophilic material, it is expected that a more amount of tear can be collected in a more efficient manner35. Since the lid for tear collection and the strip-type sensor for measurement were assembled as a combined entity, the glucose concentration could be measured almost instantaneously after tear collection without its evaporation. Therefore, the whole process from tear collection to measurement took less than 2 s in this work. This short time of preocular contact would minimize eye irritation and possible measurement errors caused by reflex tear generation8,13,23. For measurement, we applied 150 mV to induce an electric current at 300 nA. Considering the maximum preocular contact time of 2 s, this would give at most 110 nJ, which is far less than that known to be safe when applied to the ocular tissues (12 J)36,37.
Our in vivo findings revealed that the glucose concentrations in tears measured with the tear-glucose device were highly correlated with those in blood measured with a glucometer in clinical use Fig. 5(a). This correlation was more prominent when the relations were plotted using the average glucose concentrations obtained at each measurement time Fig. 5(b). Those results suggested that the tear-glucose device herein could have a similar level of accuracy compared to a conventional glucometer (Fig. 6) and compared to the strategies reported in previous studies13,14. However, due to an inherent variation among individuals, as well as the presence of temporal correlation between the blood and tear glucose levels (Supplementary Fig. 1), there could be some limitations for treating accurate glycaemia events with the diabetic patients. In this sense, our device could be better suited for healthcare purposes, especially for early diagnosis of prediabetes. With ease of use and noninvasiveness, our device herein could be more favored by the people with a high chance of getting diabetes, allowing for watching the blood glucose level early without much of discomfort even under a healthy condition. Considering actual applications in human clinical settings, a strip-type electrochemical sensor with a higher sensitivity may need to be developed. For humans, the tear glucose concentrations may vary according to the tear collection methods, but are approximately 50 times lower than those in blood38,39 and this dilution factor was lower than that reported in rabbits (25 times)28. However, the device herein was able to measure the glucose concentrations at such lowered ranges expected in human tear fluids (Supplementary Fig. 2).
In conclusion, we propose a noninvasive self-diagnostic tear-glucose device that can perform rapid and concurrent tear collection and glucose measurement. The tear glucose concentrations can be measured with an electrochemical, strip-type sensor embedded in the tear-glucose device, which can estimate the blood glucose concentrations that were shown to be highly correlated with those measured with a conventional glucometer in clinical use. The proposed lid in the tear-glucose device can avoid eye tissue damage after contact with the inferior palpebral conjunctiva, hence minimizing irritation and the production of reflex tear fluids. Therefore, we envision that the tear-glucose device, a combined entity consisting of a lid and a sensor, has the potential to predict blood glucose levels, thereby allowing for early diagnosis of prediabetes.
Methods
In vivo experiments
Approval for the in vivo experiments was granted by the Institutional Animal Care and Use Committee (IACUC No. 15-0285) at the Biomedical Research Institute of the Seoul National University Hospital. We confirmed that all experiments were performed in accordance with relevant guidelines and regulations. Male New Zealand white rabbits (2.5~3.0 kg) were raised in a controlled environment (temperature: 21 ± 1 °C, humidity: 55 ± 1%, light/dark cycle: 12 hours, and food and water ad libitum). Hyperglycemia was induced in animal models via anesthetization, as reported in previous studies13,14,28. To achieve this, the rabbits were first anesthetized with a subcutaneous injection of a cocktail of 15 mg kg−1 ketamine (Ketamine: Yuhan, Korea) and 5 mg kg−1 xylazine (Rompun: Bayer, Germany) and after 40 min, a booster shot of 7.5 mg kg−1 ketamine and 2.5 mg kg−1 xylazine was given. The glucose concentrations in both blood and tears were measured at 15, 30, 45, and 60 min after the first shot of the anesthetics. To measure blood glucose concentrations, blood was drawn from the right ear vein of the rabbit using a lancet, and a glucometer was used (Accu-Chek Performa, Roche Diagnostics, Switzerland). To measure the tear glucose concentrations, the lid of the tear-glucose device connected to a potentiostat was in contact with the IPC of the rabbit eyes for 2 s. At the specified endpoints of the experiments, we examined the presence of tissue damage on the IPC surface, following a previously reported protocol31. Briefly, a 5-μl drop of 0.25% w v−1 fluorescein sodium solution was instilled in the eye and after 5 min, the eye was washed thoroughly with normal saline to remove excess fluorescein solution. Then, a fluorescent image of the IPC surface was obtained, using a camera (Galaxy S7, Samsung, Korea) equipped with excitation (475 nm) and emission (542 nm) light filters (Thorlabs, USA).
Statistical analysis
Statistical analysis was performed with two-tailed Pearson’s correlations using SPSS (SPSS version 23, IBM, USA). A Student’s t-test was performed to obtain the P values. P < 0.05 was considered a statistically significant difference.
Supplementary information
Acknowledgements
This research was supported by the Basic Science Research Program (2017R1A2B3004830), and Bio & Medical Technology Development Program (2015M3A9E2030129) of the National Research Foundation (NRF) funded by the Ministry of Science, ICT & Future Planning.
Author Contributions
Y.B.C. designed and initiated the research project. S.H.L. and Y.C.C. prepared and characterized the tear-glucose device, and conducted in vivo experiments. The manuscript was written by Y.B.C., S.H.L. and Y.C.C.
Competing Interests
The authors declare no competing interests.
Footnotes
Publisher’s note: Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Seung Ho Lee and Yong Chan Cho contributed equally.
Change history
9/3/2019
An amendment to this paper has been published and can be accessed via a link at the top of the paper.
Supplementary information
Supplementary information accompanies this paper at 10.1038/s41598-019-41066-8.
References
- 1.McCarthy MI. Genomic Medicine Genomics, Type 2 Diabetes, and Obesity. New Engl J Med. 2010;363:2339–2350. doi: 10.1056/NEJMra0906948. [DOI] [PubMed] [Google Scholar]
- 2.Fuchsberger C, et al. The genetic architecture of type 2 diabetes. Nature. 2016;536:41–47. doi: 10.1038/nature18642. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Bartha, J. L., Martinez-Del-Fresno, P., Comino-Delgado, R. J. E. J. O. O., Gynecology & Biology, R. Early diagnosis of gestational diabetes mellitus and prevention of diabetes-related complications. 109, 41–44 (2003). [DOI] [PubMed]
- 4.Clark Charles M., Fradkin Judith E., Hiss Roland G., Lorenz Rodney A., Vinicor Frank, Warren-Boulton Elizabeth. Promoting Early Diagnosis and Treatment of Type 2 Diabetes. JAMA. 2000;284(3):363. doi: 10.1001/jama.284.3.363. [DOI] [PubMed] [Google Scholar]
- 5.Valeri C, Pozzilli P, Leslie D. Glucose control in diabetes. Diabetes-Metab Res. 2004;20:S1–S8. doi: 10.1002/dmrr.512. [DOI] [PubMed] [Google Scholar]
- 6.Meetoo, D. D., McAllister, G. & West, A. Assessing glycaemic control: self-monitoring of blood glucose. Br J Nurs20 (2011). [DOI] [PubMed]
- 7.Heinemann L. Finger pricking and pain: a never ending story. J Diabetes Sci Technol. 2008;2:919–921. doi: 10.1177/193229680800200526. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.La Belle JT, et al. Self-monitoring of tear glucose: the development of a tear based glucose sensor as an alternative to self-monitoring of blood glucose. Chem Commun. 2016;52:9197–9204. doi: 10.1039/c6cc03609k. [DOI] [PubMed] [Google Scholar]
- 9.Tura A, Maran A, Pacini G. Non-invasive glucose monitoring: Assessment of technologies and devices according to quantitative criteria. Diabetes Res Clin Pr. 2007;77:16–40. doi: 10.1016/j.diabres.2006.10.027. [DOI] [PubMed] [Google Scholar]
- 10.Vashist SK. Non-invasive glucose monitoring technology in diabetes management: A review. Anal Chim Acta. 2012;750:16–27. doi: 10.1016/j.aca.2012.03.043. [DOI] [PubMed] [Google Scholar]
- 11.Baca JT, Finegold DN, Asher SA. Tear glucose analysis for the noninvasive detection and monitoring of diabetes mellitus. Ocul Surf. 2007;5:280–293. doi: 10.1016/S1542-0124(12)70094-0. [DOI] [PubMed] [Google Scholar]
- 12.Liu C, et al. A glucose oxidase-coupled DNAzyme sensor for glucose detection in tears and saliva. Biosens Bioelectron. 2015;70:455–461. doi: 10.1016/j.bios.2015.03.070. [DOI] [PubMed] [Google Scholar]
- 13.Yan QY, et al. Measurement of Tear Glucose Levels with Amperometric Glucose Biosensor/Capillary Tube Configuration. Anal Chem. 2011;83:8341–8346. doi: 10.1021/ac201700c. [DOI] [PubMed] [Google Scholar]
- 14.Peng B, et al. Evaluation of enzyme-based tear glucose electrochemical sensors over a wide range of blood glucose concentrations. Biosens Bioelectron. 2013;49:204–209. doi: 10.1016/j.bios.2013.05.014. [DOI] [PubMed] [Google Scholar]
- 15.Cameron BD, Li Y. Polarization-based diffuse reflectance imaging for noninvasive measurement of glucose. J Diabetes Sci Technol. 2007;1:873–878. doi: 10.1177/193229680700100611. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Stuart DA, et al. Glucose sensing using near-infrared surface-enhanced Raman spectroscopy: Gold surfaces, 10-day stability, and improved accuracy. Anal Chem. 2005;77:4013–4019. doi: 10.1021/ac0501238. [DOI] [PubMed] [Google Scholar]
- 17.Gabriel EFM, Garcia PT, Lopes FM, Coltro WKT. Based Colorimetric Biosensor for Tear Glucose Measurements. Micromachines. 2017;8:104. doi: 10.3390/mi8040104. [DOI] [Google Scholar]
- 18.Scherz W, Doane MG, Dohlman CH. Tear Volume in Normal Eyes and Keratoconjunctivitis Sicca. A Graef Arch Klin Ex. 1974;192:141–150. doi: 10.1007/Bf00410700. [DOI] [PubMed] [Google Scholar]
- 19.Mishima S, Gasset A, Klyce S, Baum J. Determination of tear volume and tear flow. Invest Ophthalmol Vis Sci. 1966;5:264–276. [PubMed] [Google Scholar]
- 20.Vansetten GB, et al. Effects of the Schirmer Test on the Fibrinolytic System in the Tear Fluid. Exp Eye Res. 1990;50:135–141. doi: 10.1016/0014-4835(90)90223-H. [DOI] [PubMed] [Google Scholar]
- 21.Haggerty CM, Larke JR. Human tear protein fractions during waking hours. Ophthalmic Physiol Opt. 1982;2:187–191. doi: 10.1111/j.1475-1313.1982.tb00176.x. [DOI] [PubMed] [Google Scholar]
- 22.Daum KM, Hill RM. Human Tears - Glucose Instabilities. Acta Ophthalmol. 1984;62:530–536. doi: 10.1111/j.1755-3768.1984.tb03964.x. [DOI] [PubMed] [Google Scholar]
- 23.Rohit A, Stapleton F, Brown SHJ, Mitchell TW, Willcox MDP. Comparison of Tear Lipid Profile among Basal, Reflex, and Flush Tear Samples. Optom Vis Sci. 2014;91:1391–1395. doi: 10.1097/OPX.0000000000000411. [DOI] [PubMed] [Google Scholar]
- 24.Cha KH, Jensen GC, Balijepalli AS, Cohan BE, Meyerhoff ME. Evaluation of commercial glucometer test strips for potential measurement of glucose in tears. Anal Chem. 2014;86:1902–1908. doi: 10.1021/ac4040168. [DOI] [PubMed] [Google Scholar]
- 25.Norn M. Conjunctival sensitivity in normal eyes. Acta Ophthalmol. 1973;51:58–66. doi: 10.1111/j.1755-3768.1973.tb08246.x. [DOI] [PubMed] [Google Scholar]
- 26.Lee Seung Ho, Cho Yong Chan, Nam Dong Yeon, Huh Beom Kang, Kim Se-Na, Lee Dong Hoon, Park Chun Gwon, Choy Young Bin. Designing Minimally Invasive Preocular Contact Tips for Potential Application in Tear Collection. Cornea. 2018;37(9):1163–1168. doi: 10.1097/ICO.0000000000001603. [DOI] [PubMed] [Google Scholar]
- 27.Cha KH, Qin Y, Meyerhoff ME. Origin of Low Detection Limit and High Selectivity of Roche Accu‐Chek Test Strips that Enables Measurement of Tear Glucose Levels. Electroanal. 2015;27:670–676. doi: 10.1002/elan.201400576. [DOI] [Google Scholar]
- 28.Chu MX, et al. Soft contact lens biosensor for in situ monitoring of tear glucose as non-invasive blood sugar assessment. Talanta. 2011;83:960–965. doi: 10.1016/j.talanta.2010.10.055. [DOI] [PubMed] [Google Scholar]
- 29.Clarke WL, Cox D, Gonderfrederick LA, Carter W, Pohl SL. Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood-Glucose. Diabetes Care. 1987;10:622–628. doi: 10.2337/diacare.10.5.622. [DOI] [PubMed] [Google Scholar]
- 30.Clarke WL. The original Clarke error grid analysis (EGA) Diabetes Technol Ther. 2005;7:776–779. doi: 10.1089/dia.2005.7.776. [DOI] [PubMed] [Google Scholar]
- 31.Korb DR, Herman JP, Finnemore VM, Exford JM, Blackie CA. An evaluation of the efficacy of fluorescein, rose bengal, lissamine green, and a new dye mixture for ocular surface staining. Eye Contact Lens. 2008;34:61–64. doi: 10.1097/ICL.0b013e31811ead93. [DOI] [PubMed] [Google Scholar]
- 32.Zhang J, Hodge W, Hutnick C, Wang X. Noninvasive diagnostic devices for diabetes through measuring tear glucose. J Diabetes Sci Technol. 2011;5:166–172. doi: 10.1177/193229681100500123. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Baca JT, et al. Mass spectral determination of fasting tear glucose concentrations in nondiabetic volunteers. Clin Chem. 2007;53:1370–1372. doi: 10.1373/clinchem.2006.078543. [DOI] [PubMed] [Google Scholar]
- 34.Taormina CR, Baca JT, Asher SA, Grabowski JJ, Finegold DN. Analysis of tear glucose concentration with electrospray ionization mass spectrometry. J Am Soc Mass Spectr. 2007;18:332–336. doi: 10.1016/j.jasms.2006.10.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Collins RE, Cooke CE. Fundamental Basis for the Contact Angle and Capillary Pressure. T Faraday Soc. 1959;55:1602–1606. doi: 10.1039/tf9595501602. [DOI] [Google Scholar]
- 36.Al-Achi, A., Gupta, M. R. & Stagner, W. C. Integrated Pharmaceutics: Applied Preformulation, Product Design, And Regulatory Science. 575 (John Wiley & Sons, 2013).
- 37.Patane MA, et al. Evaluation of Ocular and General Safety Following Repeated Dosing of Dexamethasone Phosphate Delivered by Transscleral Iontophoresis in Rabbits. J Ocul Pharmacol Th. 2013;29:760–769. doi: 10.1089/jop.2012.0175. [DOI] [PubMed] [Google Scholar]
- 38.Lane JD, Krumholz DM, Sack RA, Morris C. Tear glucose dynamics in diabetes mellitus. Curr Eye Res. 2006;31:895–901. doi: 10.1080/02713680600976552. [DOI] [PubMed] [Google Scholar]
- 39.Unwin N, Shaw J, Zimmet P, Alberti KGMM. Impaired glucose tolerance and impaired fasting glycaemia: the current status on definition and intervention. Diabetic Med. 2002;19:708–723. doi: 10.1046/j.1464-5491.2002.00835.x. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.