Abstract
Objective
To evaluate the initial outcomes of a composite cage with integral fixation using the Redmond titanium (Ti)/polyetheretherketone (PEEK) anterior cervical discectomy and fusion (ACDF) device.
Methods
Data from 50 consecutive patients were prospectively collected from a single senior surgeon cohort. All cages were between 5 and 8 mm in height, and were packed with supercritical CO2 sterilized allograft. Patients were followed up for a minimum of 6 months, and implant complications were assessed.
Results
From the original cohort, three were unavailable for follow‐up. Forty‐seven patients with a total of 58 operative levels were observed for a mean of 7.9 months. A fusion rate of 96% was achieved. Good to excellent outcomes were seen in 92% of patients. There were no cases of implant Ti/PEEK delamination or implant failure, with excellent early fusion rates using supercritical CO2 allograft.
Conclusions
The present study demonstrates the development of a composite ACDF cage design that is a safe and effective treatment option with the potential for early osseointegration and interbody fusion. Supercritical CO2 sterilized allograft was an effective graft material supporting fusion.
Keywords: Anterior cervical discectomy and fusion, Composite titanium/PEEK, Integral fixation ACDF cage
Introduction
Since its description by Robinson and Smith in 1955, anterior cervical discectomy and fusion (ACDF) has been widely used for the treatment of symptomatic cervical spondylosis and disc herniation that is unresponsive to conservative management1. Although the use of autograft is considered the gold standard in achieving fusion, due to associated short‐term and long‐term complications of graft harvest, interbody cage implants with a variety of bone graft substitutes have become the mainstay of ACDF2.
Historically, three key materials have been used in the creation of cervical cages: titanium (Ti) and its alloys, polyetheretherketone (PEEK), and carbon fiber‐PEEK. Due to the synovitis and lymphatic spread of fiber debris associated with radiolucent carbon fiber‐PEEK cages, Ti and PEEK are preferred in current designs. Both materials have advantages and disadvantages3, 4, 5, 6. Although Ti has shown extensive ability to support osseointegration7, PEEK is radiolucent, allowing for easier fusion assessment and has an elastic modulus matched closer to bone, theoretically reducing levels of subsidence8. Hypothetically, the improved osteoconductivity of Ti can be utilized in combination with the elastic modulus and radioopacity of PEEK through the creation of composite Ti/PEEK spacers9, 10, 11. Clinically available composite spacers combine a PEEK body with Ti‐endplates to theoretically augment bone‐implant fusion; however, to the author's knowledge, there are only two clinical studies evaluating their usage12, 13.
There is little investigation in the literature into whether impaction of titanium‐coated PEEK cages into the disc space can result in wear or delamination of the coating. The senior author has observed 1 case of titanium wear with potential delamination (Fig. 1), which represents the first such reported case. Kienle et al.14 report that titanium cages with subtractive surface etching (no coating) are less susceptible to such delamination and failure.
Figure 1.

Potential issues with spray titanium (Ti)/polyetheretherketone (PEEK) cages. (A) Day 1 postoperation C5/6 anterior cervical discectomy and fusion (ACDF) using spray Ti/PEEK integral fixation cage; (B) 9 months postoperation with anterior cage migration (arrow). (C) Incongruous titanium spray surface with potential delamination.
The design modification of ACDF cages to include integral fixation via two or three screws has become increasingly popular due to the avoidance of anterior plate fixation that requires more extensive dissection and bone preparation. Recent evidence supports the use of these devices for anterior cervical fusion over anterior plate constructs15. The aim of the present study is to report early clinical and radiological outcomes of a unique design integral fixation Ti/PEEK cage.
Materials and Methods
Ethics Approval
Approval was obtained from the South Eastern Sydney Local Health District‐Northern Sector (SESLHD‐NS) ethics committee, Ref: HREC 11/183.
Patient Data
Over a 6‐month period from July 2016 to December 2016, 50 patients were operated on and data was collected prospectively from a single senior surgeon cohort (RJM).
Inclusion criteria: (i) patients aged between 18 and 75 years; (ii) patients suffering from cervical traumatic or degenerative disease; and (iii) patients unsuitable for or unresponsive to conservative treatment.
Exclusion criteria: (i) patients suffering from significant comorbidities, including systemic infection and terminal cancer; (ii) patients with posterior longitudinal ligament ossification; and (iii) patients with osteopenia on preoperative bone mineral density were considered acceptable (however, patients with osteoporosis were not included in the current study).
Patients were followed up for a minimum of 6 months, with assessments postoperatively at day 1, weeks 2 and 6, and months 3 and 6. All patients were operated on using the composite integral fixation Ti/PEEK Redmond cage (A‐Spine ASIA, Taiwan, China) (Fig. 2). This cage includes features of an integral fixation ×2 screw design, with porous Ti‐endplates, and a PEEK body and is available in the dimensions 14 mm × 15 mm (depth × width), 15 mm × 15 mm, 15 mm × 17 mm and 16 mm × 18.5 mm, ranging in height from 5 mm to 8 mm. Implant size choice was based on: (i) the height of the implant to mirror an adjacent normal disc height; and (ii) the maximum coverage of the endplate to distribute the load evenly.
Figure 2.

Redmond (A‐Spine, Asia, Taiwan, China), composite titanium (Ti)/polyetheretherketone (PEEK) integral fixation spacer featuring ridged titanium alloy endplates in combination with a PEEK body.
Implant
The integral fixation titanium/PEEK cage was constructed with a porous titanium plate architecture rather than a spray coating, to reduce the risk of delamination/wear of the implant surface. The integral fixation design consists of ×2 screws to prevent implant migration with an anterior anti‐backout plate (Fig. 2).
Indications
The indications for ACDF in this patient cohort of 50 patients include 48 cases of degenerative pathologies and 2 cases of trauma. Examples of indications are included in Figs 3 and 4 (foraminal stenosis with radiculopathy/hybrid procedure), Fig. 5 (myelopathy) and Fig. 6 (trauma with bifacetal dislocation and spinal cord injury). Both trauma cases were bifacetal dislocation with neurological deficit, and operative management included a combination anterior and posterior decompression, realignment and fusion (Fig. 6). The degenerative pathologies included 29 patients with radiculopathy due to disc/osteophyte complex with foraminal stenosis (Fig. 3), 10 patients with myelopathy (Fig. 4), 6 patients with discogenic neck pain without radiculopathy, and 3 patients with adjacent segment degeneration with progressive neurological deterioration (Fig. 7). Clinical outcomes were based on the cohort of patients with degenerative pathologies only, with exclusion of the trauma cohort as both patients suffered an ASIA‐B neurological injury.
Figure 3.

Anterior cervical discectomy and fusion (ACDF)/disc replacement hybrid for disc height loss and foraminal stenosis. (A) Saggital MRI of cervical level. (B) Intraoperative X‐ray.
Figure 4.

(A) Myelo/radiculopathy at C5‐6. (B) Trial prosthesis to check cage position and height. (C) Device in position.
Figure 5.

(A) Myelopathy at C4‐5 due to large central disc herniation. (B) Device in position.
Figure 6.

(A) Pre‐operative CT scan showing trauma with (inset) Bifacetal dislocation at C5‐6. (B) Anterior interbody fixation, with posterior lateral mass fixation (Neon 2 Posterior Cervical, Ulrich, Germany).
Figure 7.

Adjacent segment degeneration: 83‐year‐old female with progressive myelopathy and a background of C4‐5 and C5‐6 ACDF performed 11 years prior. (A) Severe canal stenosis C3‐4 with cord signal change. (B) Intraoperative trial of prosthesis to confirm height and depth. (C) Final implant position.
Surgical Technique
All surgeries were conducted by a single senior surgeon using a modified Smith–Robinson technique under general anesthesia. Following a linear right anterolateral incision, retraction of the musculo‐visceral column was achieved using the Trim‐Line retractor system with distraction of the vertebral bodies using modified Caspar retraction pins (A‐Spine ASIA, Taiwan, China). Pathological disc material was then removed using a combination of rongeurs and curettes under the direct observation of an operating microscope (Pentero, C. Zeiss, Germany). Any visible osteophytes were also removed using a high speed 2‐mm drill and the posterior longitudinal ligament was divided and removed in all cases. Complete decompression and visualization of the dura and nerve roots was achieved. Decortication of the vertebral endplates was performed to optimize the bone‐cage/graft interface.
The appropriate size cage was determined through the usage of a trial spacer to confirm the height of the disc space (Fig. 3B). Cages were filled with human allograft (Allovance Cortico‐Cancellous Crunch (1–3.5 mm), Australian Biotechnologies, Sydney, Australia) that was processed and sterilized using a proprietary supercritical carbon dioxide process. The allograft was firmly packed into the cage with the aim of distributing the axial loading through the implant. Cages were inserted using standard instrumentation and tapped into place (Fig. 3A–C).
In all cases, following implant impaction and verification on lateral X‐ray, integral fixation was applied via the use of ×2 self‐tapping screws (Fig. 8D). Prior to wound closure, intraoperative anteroposterior and lateral plain radiographs were obtained to confirm the correct implant and screw positioning. All patients were advised to wear a cervical orthosis postoperatively for a period of 2 weeks. Postoperative pain relief included a low dose of non‐steroidal anti‐inflammatory drug (NSAID) and paracetamol.
Figure 8.

Sequence of anterior cervical discectomy and fusion (ACDF) using the ACDF device: (A) Exposure with Trim Line and Casper retractor with discectomy and decompression of the neurological elements. (B) Trial implant to determine width and height of final prosthesis. (C) Implantation of Redmond titanium (Ti)/polyetheretherketone (PEEK) cage with Awl guide on inserter. (D) ×2 screw integral fixation. (Inset) Final appearance.
Outcome Measures
Radiographic Assessment
Radiographic fusion was assessed by an independent radiology practice (Southern Radiology, Sydney, Australia) with no conflict of interest with regards to the study outcome. Anteroposterior and lateral cervical radiographs were performed at day 1 and week 6 postoperatively for radiographic assessment of the implant to confirm that there was no implant failure, delamination or movement from the original implantation position. Evidence of delamination of the titanium/PEEK interface was determined at each follow‐up time point14. Computed tomography (CT) scans were performed at 3 months to assess the early fusion status of bone growth through the implant and the absence of lucency, and at 6 months if unsure of fusion status. Fusion was considered successful if bridging bone incorporating the graft and adjoining endplates was apparent, with evidence of graft remodeling and new bone formation, restoration of interbody space, and no hardware/implant failure16.
Clinical outcomes were assessed preoperatively and 6 months postoperatively. Patients were asked to quantify neck and arm pain on a visual analogue scale (VAS) ranging from 0 (no pain/discomfort) to 10 (worst pain/discomfort imaginable) preoperatively and postoperatively. Functional outcome was measured using the neck Oswestry disability index (NODI). In addition, patients were assessed according to the Quality of Life 12‐Item Short Form (SF‐12) Health Survey, with their procedure determined using the patient satisfaction index (PSI) as described by Palit et al.17 at final follow‐up.
Statistical Analysis
Descriptive data are represented as means ± SD (range, minimum–maximum). All datasets were tested for normality with the D’Agostino and Pearson omnibus normality test. Unpaired nonparametric data was analyzed using the Mann–Whitney U test and parametric data with an unpaired t test for comparison of the results. The Wilcoxon signed rank test was used for nonparametric, symmetrically distributed data and a paired t test was used for parametric data comparing preoperative and postoperative variables within patient groups. Statistical significance was set at P < 0.05. All analyses and graphs were generated using a commercial software package (GraphPad Prism version 5.01, GraphPad Software, USA).
Results
Patient Demographics
From 50 consecutive patients in the original dataset, 47 patients with 58 operative levels met the inclusion criteria. A total of 37 patients received a single level ACDF, 9 patients double level and 1 patient triple level. There were no patient deaths. There were 27 men and 20 women, with a mean age of 57 ± 14.5 years (range, 28–76 years) and a mean follow‐up period of 7.9 months (range, 6–10.8 months). Nine patients reported that they were smokers. There were 6 diabetics and 7 patients receiving workers’ compensation coverage for their surgery. Neck pain was present in almost all patients, with the main indication for surgery being cervical disc herniation associated with radiculopathy or spinal stenosis and cord compression. Within the non‐trauma cohort, the mean preoperative symptom length was 2.7 years (range, 9 weeks to 18 years). All patients were operated on between the surgical levels C3 and C7, with 10 multiple level ACDF procedures and 37 single level ACDF procedures.
Radiological Outcomes
A fusion rate of 96% (45/47) was achieved, with evidence of bridging bone between endplates on CT scanning at 6 months postoperation. The presence of titanium endplates did not interfere with fusion assessments (Fig. 9). Evidence of graft remodeling at the interface with the endplate was noted.
Figure 9.

Fusion 3 months postoperatively. (A) Preoperative X‐ray. (B) Intraoperative X‐ray. (C, D) Difficult fusion environment with ×3 level anterior cervical discectomy and fusion (ACDF), with features of interbody fusion at all levels.
Clinical Outcomes
Visual Analogue Scale Score
Neck and arm pain scored using the VAS showed significant improvement (P < 0.001) between preoperative and postoperative groups. Overall pain improved from 7.3 ± 1.7 to 2.3 ± 1.5, with an average improvement of 5.0. These results are consistent with reported outcomes for ACDF surgery in the degenerative population18.
SF‐12 Score
Preoperative SF‐12 scores were physical component summary score (PCS) 37.8 ± 6, mental component summary score (MCS) 36.0 ± 7, with mean postoperative improvement of scores of 4.2 for PCS to 42.7 ± 8 and a mean postoperative improvement of 11.0 for MCS to 49.8 ± 8. This was statistically significant (P = 0.025) for MCS but not for PCS.
Neck Oswestry Disability Index Score
Significant mean neck ODI (NODI) scores preoperatively were 47.0 (SD ± 15.2), with a mean improvement of 22 (SD ± 7.8) (P = 0.04).
A summary of VAS, SF‐12 and NODI is provided in table 1.
Table 1.
Patient clinical outcomes (mean ± standard deviation)
| Evaluation criterion | Preoperative score | Postoperative score | Improvement |
|---|---|---|---|
| VAS | 7.3 ± 1.7 | 2.3 ± 1.5 | 5.0 |
| SF‐12 | |||
| PCS | 37 ± 6 | 43 ± 8 | 4 |
| MCS | 36 ± 7 | 47 ± 8 | 11* (P = 0.025) |
| NODI | 47 ± 15.2 | 22 ± 7.8 | 25* (P = 0.04) |
P < 0.05
MCS, mental component summary score; NODI, neck Oswestry disability index.; PCS, physical component summary score; SF‐12, Quality of Life 12‐item Short Form; VAS, visual analogue scale
Odom's Assessment Results
According to Odom's criteria, there were 39 excellent, 7 good, 3 fair, and no poor outcomes, with 92% of patients achieving a good–excellent clinical outcome. The 2 trauma patients were not included in the Odom's assessment, as noted above (Table 2).
Table 2.
Clinical and radiological outcomes of Ti, PEEK, and Ti/PEEK cages19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 29
| Cage material | Good‐to‐excellent clinical outcome (%) | Fusion rate at 3 months (%) | Fusion rate at 6 months (%) | Fusion rate at 12 months (%) | Subsidence (%) |
|---|---|---|---|---|---|
| Titanium20, 22, 27, 28 | 46–95 | — | 37.2–97 | 86.5–99 | 13–45 |
| PEEK19, 20, 23, 24, 25, 26, 27, 28, 29 | 74–100 | — | 61.1–96 | 93–100 | 5–15 |
| Ti/PEEK (current study) | 92 | — | 96 | — | 8.3 (4/48) |
PEEK, polyetheretherketone; Ti, titanium.
Complications
Complications were divided into approach‐related and implant‐related issues. There were no implant complications and there was no evidence of delamination in any cases. Additional attention was focused on 6‐month fine‐cut CT to determine the presence of implant failure at the Ti/PEEK junction, which was not present in any cases. Approach‐related complications included one case of recurrent laryngeal nerve palsy that had improved at the 4‐month follow‐up mark. There was 1 case of postoperative hematoma requiring an additional procedure for clot removal within 12 h of the index procedure. The patient proceeded to have an excellent clinical outcome at 6 months follow‐up.
Discussion
A Cochrane systematic review concluded that interbody fusion techniques employing autograft yielded higher fusion rates than allograft and synthetic bone substitute techniques; however, donor site morbidity associated with autograft has fueled the growing interest in alternative materials30, 31. Furthermore, authors have questioned the current validity of the statement that “autograft is the gold standard”32. In this study, the combination of a Ti/PEEK cage, with integral fixation, with allograft, proved to be an effective and safe ACDF design and material combination, resulting in statistically significant improvements in pain and function at 6 months follow‐up.
Interbody Cage Material Properties
Polyetheretherketone is a semicrystalline polyaromatic linear polymer and thermoplastic material of high molecular weight, which is biologically inert, radiolucent, and non‐resorbable19. For interbody spacers, PEEK provides a hard frame that can resist spinal loading, thereby providing initial stability while having an elastic modulus similar to that of bone to minimize graft subsidence33. Titanium can be modified to improve both ongrowth and ingrowth34, 35, 36. Ongrowth of bone is the direct apposition of bone to the surface. Ingrowth involves the interlocking or bone growth “into” the surface of a material, requiring a 3D structure with pores open to the outside. These modifications are aimed at influencing the way tissues integrate with the implant material37. The presence of titanium at the interface between the host and the device has the potential benefits of the hydrophilic and osteoconductive nature of titanium compared the hydrophobic nature of PEEK alone38. Titanium in the current device, however, is not a coating which minimizes the potential for delamination at the PEEK–titanium interface.
Previous studies have compared the efficacy of Ti and PEEK cages in both single and multi‐level ACDF. Studies have also shown PEEK to have better long‐term maintenance of clinical height, lower rates of subsidence, and improved clinical outcomes39. These advantages are attributed to PEEK's elastic modulus; however titanium implants have an exceptional ability to support direct bone ongrowth, and have a surface structure that is comparably resistant to microbial adhesion40, 41. By combining these materials in a composite cage, the advantages of both materials may theoretically be utilized. This study provides data for composite cages, with integral fixation, and demonstrates that they achieve comparable radiological and clinical outcomes to single material devices within a 6‐month follow‐up19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 29.
Graft Choice
Although the literature comparing implant materials and designs in ACDF is vast, information on the clinical impact of interbody graft choice is limited. Successful fusion is dependent on not only osteogenic potential and osteoinductive factors but also the structural scaffold that aids neovascularization and bony ingrowth42. Graft materials should promote osseointegration by providing an osteoconductive scaffold that can participate in the fusion. However, graft material inside a cage may also play an additional role in mechanical load distribution and reduce local stress concentrations43. Subsidence is thought to occur in relation to high pressures delivered through interbody spacers over a small surface area. As a result, PEEK cages, which have an elastic modulus closer to cancellous bone, experience lower rates of subsidence39. There is very little data comparing the rates of subsidence between patients with and without grafting, as well as graft types; however, it is known that autograft, allograft, and synthetic bone graft substitutes each have different mechanical and osteoconductive capacity. The mechanical properties of allograft bone are dependent on its sterilization treatment. While ionizing radiation sterilization of allograft increases brittleness and effects mechanical load bearing, SCCO2 treatment maintains the graft's intrinsic mechanical properties44, 45. All implants in our study were grafted with SCCO2 allograft, which was carefully packed into the cage and can be hypothesized to play a role in improved load distribution at the device–endplate interface and, therefore, prevention of subsidence; however, further studies are required to determine the role of this interbody graft choice.
Integral Fixation Design
Integral fixation aims to reduce the amount of micro‐motion at the graft–host interface, graft settling, and kyphotic deformity, but also add to costs, risks, and operative time46. The role of integrated plate devices as provided by low‐profile designs including the Zero‐P (Synthes CmbH Switzerland, Oberdorf, Switzerland) and the ROI‐C cervical cage (LDR Holding Global Corporation, France) have been designed to assist surgeon work flow, and ease of implantation. By streamlining anterior plating into a stand‐alone device, these designs aim to minimize implant‐to‐soft tissue impact, reducing dysphagia rates and other plate‐related complications, while still reducing the risk of subsidence, pseudoarthrosis and cervical kyphosis11, 47. Early results have shown these designs to achieve good clinical outcomes, with lower levels of dysphagia and shorter operation times47, 48. The current study combines the positive benefits of both integral fixation technology with the dual material combination of titanium and PEEK.
Study Limitations
A primary limitation of this study is the relatively small numbers involved. Incomplete follow‐up data on patients was a difficulty in our study, as is common for clinical studies; in our particular cohort, this was primarily related to patients from rural areas being unable to access medical imaging centers in the required timeframe of the study.
The assessment of interbody fusion and the integration of the Ti endplate remains a challenge. As there are no universally accepted criteria for determining radiological fusion, it is often difficult to arrive at a true assessment of fusion based on plain radiography alone, particularly when synthetic cages are used. Our study utilized fine‐cut CT scans with reconstructions, which has been shown to be more reliable and sensitive for the detection of pseudoarthrosis than plain radiography49, 50. In addition, it was noted that the Ti‐endplates did not interfere with fusion assessments on either radiographs or CT.
Conclusion
In this study, we have found that using an integral fixation Ti/PEEK interbody cage containing allograft in anterior cervical discectomy and fusion was a safe and effective treatment for degenerative and traumatic cervical pathologies. Enhancement of PEEK cages with titanium endplates is likely to assist with early integration of the prosthesis with the surrounding bone and vertebral endplate. Further studies are required to determine if the usage of a composite design improves implant longevity by limiting subsidence as well as stress shielding and associated complications.
Acknowledgment
Thanks to Dr Vedran Lovric for providing information on SCCO2 allograft.
Disclosure: Professor Mobbs is a design surgeon for multiple companies, including A‐Spine ASIA. Professor Walsh is involved in research activities on interbody devices, bone graft materials, and osseointegration for multiple companies.
References
- 1. Mobbs RJ, Rao P, Chandran NK. Anterior cervical discectomy and fusion: analysis of surgical outcome with and without plating. J Clin Neurosci, 2007, 14: 639–642. [DOI] [PubMed] [Google Scholar]
- 2. McConnell JR, Freeman BJ, Debnath UK, Grevitt MP, Prince HG, Webb JK. A prospective randomized comparison of coralline hydroxyapatite with autograft in cervical interbody fusion. Spine, 2003, 28: 317–323. [DOI] [PubMed] [Google Scholar]
- 3. Bartels RH, Donk RD, Feuth T. Subsidence of stand‐alone cervical carbon fiber cages. Neurosurgery, 2006, 58: 502–508. [DOI] [PubMed] [Google Scholar]
- 4. Vavruch L, Hedlund R, Javid D, Leszniewski W, Shalabi A. A prospective randomized comparison between the cloward procedure and a carbon fiber cage in the cervical spine: a clinical and radiologic study. Spine, 2002, 27: 1694–1701. [DOI] [PubMed] [Google Scholar]
- 5. Gercek E, Arlet V, Delisle J, Marchesi D. Subsidence of stand‐alone cervical cages in anterior interbody fusion: warning. Eur Spine J, 2003, 12: 513–516. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Parsons JR, Bhayani S, Alexander H, Weiss AB. Carbon fiber debris within the synovial joint a time‐dependent mechanical and histologic study. Clin Orthop Relat Res, 1985, 196: 69–76. [PubMed] [Google Scholar]
- 7. Svehla M, Morberg P, Zicat B, Bruce W, Sonnabend D, Walsh WR. Morphometric and mechanical evaluation of titanium implant integration: comparison of five surface structures. J Biomed Mater Res, 2000, 51: 15–22. [DOI] [PubMed] [Google Scholar]
- 8. Rao PJ, Pelletier MH, Walsh WR, Mobbs RJ. Spine interbody implants: material selection and modification, functionalization and bioactivation of surfaces to improve osseointegration. Orthop Surg, 2014, 6: 81–89. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Olivares‐Navarrete R, Gittens RA, et al Osteoblasts exhibit a more differentiated phenotype and increased bone morphogenetic protein production on titanium alloy substrates than on poly‐ether‐ether‐ketone. Spine J, 2012, 12: 265–272. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Han CM, Lee EJ, Kim HE, et al The electron beam deposition of titanium on polyetheretherketone (PEEK) and the resulting enhanced biological properties. Biomaterials, 2010, 31: 3465–3470. [DOI] [PubMed] [Google Scholar]
- 11. Chong E, Pelletier MH, Mobbs RJ, Walsh WR. The design evolution of interbody cages in anterior cervical discectomy and fusion: a systematic review. BMC Musculoskelet Disord, 2015, 16: 99. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Chong E, Mobbs RJ, Pelletier MH, Walsh WR. Titanium/Polyetheretherketone cages for cervical arthrodesis with degenerative and traumatic pathologies: early clinical outcomes and fusion rates. Orthop Surg, 2016, 8: 19–26. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Mobbs RJ, Phan K, Assem Y, Pelletier M, Walsh WR. Combination Ti/PEEK ALIF cage for anterior lumbar interbody fusion: early clinical and radiological results. J Clin Neurosci, 2016, 34: 94–99. [DOI] [PubMed] [Google Scholar]
- 14. Kienle A, Graf N, Wilke HJ. Does impaction of titanium‐coated interbody fusion cages into the disc space cause wear debris or delamination? Spine J, 2016, 16: 235–242. [DOI] [PubMed] [Google Scholar]
- 15. Nambiar M, Phan K, Cunningham JE, Yang Y, Turner PL, Mobbs R. Locking stand‐alone cages versus anterior plate constructs in single‐level fusion for degenerative cervical disease: a systematic review and meta‐analysis. Eur Spine J, 2017, 26: 2258–2266. [DOI] [PubMed] [Google Scholar]
- 16. van Jonbergen HP, Spruit M, Anderson PG, Pavlov PW. Anterior cervical interbody fusion with a titanium box cage: early radiological assessment of fusion and subsidence. Spine J, 2005, 5: 645–649. [DOI] [PubMed] [Google Scholar]
- 17. Palit M, Schofferman J, Goldthwaite N, et al Anterior discectomy and fusion for the management of neck pain. Spine (Phila Pa 1976), 1999, 24: 2224–2228. [DOI] [PubMed] [Google Scholar]
- 18. Buttermann GR. Anterior cervical discectomy and fusion outcomes over 10 years: a prospective study. Spine (Phila Pa 1976), 2018, 43: 207–214. [DOI] [PubMed] [Google Scholar]
- 19. Mastronardi L, Ducati A, Ferrante L. Anterior cervical fusion with polyetheretherketone (PEEK) cages in the treatment of degenerative disc disease. Preliminary observations in 36 consecutive cases with a minimum 12‐month follow‐up. Acta Neurochir, 2006, 148: 307–312. [DOI] [PubMed] [Google Scholar]
- 20. Chou YC, Chen DC, Hsieh WA, et al Efficacy of anterior cervical fusion: comparison of titanium cages, polyetheretherketone (PEEK) cages and autogenous bone grafts. J Clin Neurosci, 2008, 15: 1240–1245. [DOI] [PubMed] [Google Scholar]
- 21. Moreland DB, Asch HL, Clabeaux DE, et al Anterior cervical discectomy and fusion with implantable titanium cage: initial impressions, patient outcomes and comparison to fusion with allograft.[Erratum appears in Spine J. 2004 May–Jun;4(3):following table of contents]. Spine J, 2004, 4: 184–191; discussion 91. [DOI] [PubMed] [Google Scholar]
- 22. Majd ME, Vadhva M, Holt RT. Anterior cervical reconstruction using titanium cages with anterior plating. Spine, 1999, 24: 1604–1610. [DOI] [PubMed] [Google Scholar]
- 23. Eastlack RK, Garfin SR, Brown CR, Meyer SC. Osteocel plus cellular allograft in anterior cervical discectomy and fusion: evaluation of clinical and radiographic outcomes from a prospective multi‐center study. Spine (Phila Pa 1976), 2014, 39: E1331–E1337. [DOI] [PubMed] [Google Scholar]
- 24. Faldini C, Chehrassan M, Miscione MT, et al Single‐level anterior cervical discectomy and interbody fusion using PEEK anatomical cervical cage and allograft bone. J Orthop Traumatol, 2011, 12: 201–205. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Faldini C, Miscione MT, Acri F, et al Single level cervical fusion by an anterior approach using autologous bone graft influences the adjacent levels degenerative changes: clinical and radiographic results at 10‐year minimum follow‐up. Eur Spine J, 2012, 21 (Suppl. 1): S90–S93. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Ha SK, Park JY, Kim SH, Lim DJ, Kim SD, Lee SK. Radiologic assessment of subsidence in stand‐alone cervical polyetheretherketone (PEEK) cage. J Korean Neurosurg Soc, 2008, 44: 370–374. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Liao JC, Niu CC, Chen WJ, Chen LH. Polyetheretherketone (PEEK) cage filled with cancellous allograft in anterior cervical discectomy and fusion. Int Orthop, 2008, 32: 643–648. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Niu CC, Liao JC, Chen WJ, Chen LH. Outcomes of interbody fusion cages used in 1 and 2‐levels anterior cervical discectomy and fusion: titanium cages versus polyetheretherketone (PEEK) cages. J Spinal Disord Tech, 2010, 23: 310–316. [DOI] [PubMed] [Google Scholar]
- 29. Yang JJ, Yu CH, Chang BS, Yeom JS, Lee JH, Lee CK. Subsidence and nonunion after anterior cervical interbody fusion using a stand‐alone polyetheretherketone (PEEK) cage. Clin Orthop Surg, 2011, 3: 16–23. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Vaccaro AR, Singh K, Haid R, et al The use of bioabsorbable implants in the spine. Spine J, 2003, 3: 227–237. [DOI] [PubMed] [Google Scholar]
- 31. Jacobs WC, Anderson PG, Limbeek J, Willems PC, Pavlov P. Single or double‐level anterior interbody fusion techniques for cervical degenerative disc disease. Cochrane Database Syst Rev, 2004, 18: CD004958. [DOI] [PubMed] [Google Scholar]
- 32. Maharaj MM, Phan K, Mobbs RJ. Anterior cervical discectomy and fusion (ACDF) autograft versus graft substitutes: what do patients prefer?—a clinical study. J Spine Surg, 2016, 2: 105–110. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Cho DY, Liau WR, Lee WY, Liu JT, Chiu CL, Sheu PC. Preliminary experience using a polyetheretherketone (PEEK) cage in the treatment of cervical disc disease [erratum appears in Neurosurgery 2003;52:693]. Neurosurgery, 2002, 51: 1343–1349; discussion 9‐50. [PubMed] [Google Scholar]
- 34. Daugaard H, Elmengaard B, Bechtold JE, Soballe K. Bone growth enhancement in vivo on press‐fit titanium alloy implants with acid etched microtexture. J Biomed Mater Res A, 2008, 87: 434–440. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Jemat A, Ghazali MJ, Razali M, Otsuka Y. Surface modifications and their effects on titanium dental implants. Biomed Res Int, 2015, 2015: 791725. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Ma R, Tang T. Current strategies to improve the bioactivity of PEEK. Int J Mol Sci, 2014, 15: 5426–5445. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37. Pelletier MH, Cordaro N, Punjabi VM, Waites M, Lau A, Walsh WR. Peek versus Ti interbody fusion devices: resultant fusion, bone apposition, initial and 26‐week biomechanics. Clin Spine Surg, 2016, 29: E208–E214. [DOI] [PubMed] [Google Scholar]
- 38. Walsh WR, Bertollo N, Christou C, Schaffner D, Mobbs RJ. Plasma‐sprayed titanium coating to polyetheretherketone improves the bone‐implant interface. Spine J, 2015, 15: 1041–1049. [DOI] [PubMed] [Google Scholar]
- 39. Chen Y, Wang X, Lu X, et al Comparison of titanium and polyetheretherketone (PEEK) cages in the surgical treatment of multilevel cervical spondylotic myelopathy: a prospective, randomized, control study with over 7‐year follow‐up. Eur Spine J, 2013, 22: 1539–1546. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Johansson CB, Röser K, Bolind P, Donath K, Albrektsson T. Bone‐tissue formation and integration of titanium implants: an evaluation with newly developed enzyme and immunohistochemical techniques. Clin Implant Dent Relat Res, 1999, 1: 33–40. [DOI] [PubMed] [Google Scholar]
- 41. Cabraja M, Oezdemir S, Koeppen D, Kroppenstedt S. Anterior cervical discectomy and fusion: comparison of titanium and polyetheretherketone cages. BMC Musculoskelet Disord, 2012, 13: 172. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42. Bruce C, Chin KR, Cumming V, Crawford NR. Stabilizing effects of a particulate demineralized bone matrix in the L4 interbody space with and without PEEK Cage: a literature review and preliminary results of a cadaveric biomechanical study. West Indian Med J, 2013, 62: 748–751. [DOI] [PubMed] [Google Scholar]
- 43. McCaffrey KA, Walsh W, Pelletier MH, Lovric V. Endplate pressure distribution is influenced by bone graft substitute within interbody cages. Spine J, 2014, 14: S164. [Google Scholar]
- 44. Currey JD, Foreman J, Laketić I, Mitchell J, Pegg DE, Reilly GC. Effects of ionizing radiation on the mechanical properties of human bone. J Orthop Res, 1997, 15: 111–117. [DOI] [PubMed] [Google Scholar]
- 45. Russell N, Rives A, Bertollo N, Pelletier MH, Walsh WR. The effect of sterilization on the dynamic mechanical properties of paired rabbit cortical bone. J Biomech, 2013, 46: 1670–1675. [DOI] [PubMed] [Google Scholar]
- 46. Rhee JM, Park JB, Yang JY, Riew DK. Indications and techniques for anterior cervical plating. Neurol India, 2005, 53: 433–439. [DOI] [PubMed] [Google Scholar]
- 47. Miao J, Shen Y, Kuang Y, et al Early follow‐up outcomes of a new zero‐profile implant used in anterior cervical discectomy and fusion. J Spinal Disord Tech, 2013, 26: E193–E197. [DOI] [PubMed] [Google Scholar]
- 48. Wang Z, Jiang W, Zhang Z, et al Comparison of ROI‐C and traditional cage with anterior plating for anterior cervical discectomy and fusion. Zhonghua Wai Ke Za Zhi, 2014, 52: 425–430. [PubMed] [Google Scholar]
- 49. Carreon L, Glassman S, Djurasovic M. Reliability and agreement between fine‐cut CT scans and plain radiography in the evaluation of posterolateral fusions. Spine J, 2007, 7: 39–43. [DOI] [PubMed] [Google Scholar]
- 50. Santos ER, Goss DG, Morcom RK, Fraser RD. Radiologic assessment of interbody fusion using carbon fiber cages. Spine (Phila Pa 1976), 2003, 28: 997–1001. [DOI] [PubMed] [Google Scholar]